K Number
K170200

Validate with FDA (Live)

Date Cleared
2017-06-01

(129 days)

Product Code
Regulation Number
862.1160
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Teco Carbon Dioxide Reagent Set is a device which is intended for measurement of Carbon Dioxide level in human serum, in vitro diagnostic use only. Test results may provide information regarding the status in the assessment of acid-base balance of metabolic alkalosis or respiratory acidosis.

Device Description

Teco Carbon Dioxide Reagent Set is a single reagent kit. Reagent contains Good's buffer, phosphoenolpyruvate (PEP), phosphoenolpyruvate carboxylase (PEPC), malate dehydrogenase (MDH), magnesium ions, NADH analog, nonreactive stabilizer, preservative, and Buffer. Approximately 90% of carbon dioxide present in serum is in the form of bicarbonate. The measurement of bicarbonate is useful in the assessment of disturbances of acid-base balance resulting from metabolic or respiratory causes.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Teco Carbon Dioxide Reagent Set, based on the provided document:

Acceptance Criteria and Device Performance

Acceptance Criteria CategorySpecific CriteriaReported Device Performance (Teco Carbon Dioxide Reagent Set)
Linearity/Reportable RangeRange over which the assay demonstrates a linear relationship to the analyte concentration. (Predicate: 2.0 to 40.0 mmol/L)8.7 to 40.0 mmol/L (linear up to 64 mmol/L)
Accuracy/CorrelationComparison to a predicate device, typically measured by linear regression with a high R-squared value. (Predicate: $y = 0.965x + 1.200$)$y = 0.9785x + 0.2636$, $R^2 = 0.9925$, and $R = 0.9962$
Detection Limit (LoQ)The lowest concentration at which the analyte can be quantitatively determined with acceptable precision (<=10%CV). (Predicate: 2.0 mmol/L)LoQ: 1.77 mmol/L
Analytical SpecificityNo significant bias (defined as ±10% difference from control results) in the presence of specified interferents at certain concentrations.Hemoglobin: 1,000 mg/dL (no interference) Total Bilirubin: 60 mg/dL (no interference) Conjugate Bilirubin: 30 mg/dL (no interference) Triglycerides: 2,000 mg/dL (no interference) Intralipid: 2,000 mg/dL (no interference) Ascorbic Acid: 6 mg/dL (no interference)
Precision/ReproducibilityDefined by within-laboratory precision (SD and %CV) and repeatability (SD and %CV) across different concentration levels.QC Level 1 (30.6 mmol/L): Repeatability SD 0.1782, %CV 0.582; Within-Laboratory Precision SD 1.62, %CV 5.31 QC Level 2 (20.5 mmol/L): Repeatability SD 0.7065, %CV 3.426; Within-Laboratory Precision SD 1.87, %CV 9.07 Serum Level 1 (12.0 mmol/L): Repeatability SD 0.1275, %CV 1.065; Within-Laboratory Precision SD 0.69, %CV 5.75 Serum Level 2 (26.6 mmol/L): Repeatability SD 0.1521, %CV 0.571; Within-Laboratory Precision SD 0.85, %CV 3.21 Serum Level 3 (37.4 mmol/L): Repeatability SD 0.1949, %CV 0.521; Within-Laboratory Precision SD 0.59, %CV 1.58
Storage Temperature2-8 °C2-8 °C (Accelerated stability: at least 21 months; real-time stability verification ongoing)

Study Details

  1. Sample sizes used for the test set and data provenance:

    • Accuracy/Method Comparison: 176 patient serum samples.
    • Linearity: Undisclosed number of unaltered human serum samples (prepared by mixing a diluted patient serum sample (2 mmol/L) and a spiked patient serum sample (64 mmol/L) to obtain 11 different concentrations).
    • Precision/Reproducibility: Unaltered human serums spiked with carbon dioxide concentrations and commercial serum-based controls.
    • Detection Limit (LoB, LoD, LoQ):
      • LoB: One artificial serum base matrix pool.
      • LoD: Four (4) separate pools.
      • LoQ: Five (5) separate pools.
    • Analytical Specificity: Two (2) different concentrations of sodium bicarbonate (20 mmol/L and 35 mmol/L) human serum samples. Pooled human serum samples with added potential interferents.
    • Data Provenance: The document does not explicitly state the country of origin for the patient samples. All studies appear to be prospective as they were conducted as part of the device's validation.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
    This document describes an in vitro diagnostic (IVD) reagent set. For such devices, ground truth is typically established by reference methods or established analytical techniques, not by human expert interpretation in the way a diagnostic imaging AI might be evaluated. Therefore, no information is provided regarding human experts for establishing ground truth.

  3. Adjudication method for the test set:
    Not applicable to an IVD reagent set. Ground truth is determined analytically.

  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
    Not applicable. This is an IVD reagent set, not an AI-assisted diagnostic tool for human readers.

  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
    Yes, all performance studies described are for the standalone Teco Carbon Dioxide Reagent Set. There is no human-in-the-loop component mentioned. The device directly measures carbon dioxide levels.

  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
    The ground truth for this device's performance studies is established by:

    • Reference methods and quantitative analysis: This is evident in the comparison to a predicate device (Pointe Scientific Inc., Carbon Dioxide Reagent) and the use of standards (NIST SRM 351 traceable CO2 standard).
    • Known concentrations: For linearity, precision, and detection limit studies, samples were prepared with known or measured concentrations of carbon dioxide.
    • Spiked samples: Used for linearity and precision to cover a range of concentrations.
  7. The sample size for the training set:
    Not applicable. This is an IVD reagent set, not a machine learning or AI algorithm that requires a "training set." Its function is based on established enzymatic reactions, not learned patterns from data.

  8. How the ground truth for the training set was established:
    Not applicable, as there is no "training set."

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized caduceus, a symbol often associated with medicine and healthcare. The caduceus consists of a staff with two snakes coiled around it, topped by a pair of wings.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 1, 2017

TECO DIAGNOSTICS, INC. HYUN BIN SONG RESEARCH SCIENTIST 1268 NORTH LAKEVIEW AVENUE ANAHEIM CA 92807

Re: K170200

Trade/Device Name: Carbon Dioxide Reagent Set Regulation Number: 21 CFR 862.1160 Regulation Name: Bicarbonate/carbon dioxide test system Regulatory Class: II Product Code: KHS Dated: April 26, 2017 Received: April 27, 2017

Dear Hyun Bin Song:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Kellie B. Kelm -S

for Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Form Approved: OMB No.0910-0120
Expiration Date:January 31, 2017
See PRA Statement below.
510(k) Number (if known)K170200
Device NameCarbon Dioxide Reagent Set
Indications for Use (Describe)Teco Carbon Dioxide Reagent Set is a device which is intended for measurement of Carbon Dioxide level in human serum, in vitro diagnostic use only. Test results may provide information regarding the status in the assessment of acid-base balance of metabolic alkalosis or respiratory acidosis.

Type of Use (Select one or both, as applicable)

</span> Prescription Use (Part 21 CFR 801 Subpart D)□ Over-The-Counter Use (21 CFR 801 Subpart C)
-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows a green hexagon with the letters "TC" in silver inside. The letters are slightly overlapping and have a subtle gradient effect, giving them a three-dimensional appearance. The hexagon has a white outline, which makes it stand out against the background.

1268 N. Lakeview Ave., Anaheim, CA 92807, U.S.A. Phone: (714) 463-1111 Toll Free: 1-800-222-9880 Fax: (714) 463-1169 www.tecodiag.com

Section 5 - 510(k) SUMMARY

This Summary of 510(k) Safety and Effectiveness is being submitted in accordance with the requirements of 21 CFR § 807.92.

Owner's Name: Teco Diagnostics, Inc.

Address and Contact information:

1268 North Lakeview Avenue Anaheim, CA 92807 Phone: 714-463-1111 Fax: 714-463-1169

Contact:

Hyun Bin Song

Date Prepared:

January 20, 2017

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows a green hexagon with the letters 'TC' in silver. The letters are stacked on top of each other, with 'T' on top and 'C' on the bottom. The letters are slightly offset to the left. The background of the hexagon is a solid green color.

1268 N. Lakeview Ave., Anaheim, CA 92807, U.S.A. Phone: (714) 463-1111 Toll Free: 1-800-222-9880 Fax: (714) 463-1169 www.tecodiag.com

Section 5 - 510(k) SUMMARY

A. 510(k) Number:

K170200

B. Analytes:

Carbon Dioxide

C. Type of Test:

Enzymatic, quantitative assay

D. Applicant:

Teco Diagnostics, Inc.

E. Trade Name:

Carbon Dioxide Reagent Set

F. Common Name:

Enzymatic Method, Carbon Dioxide

G. Regulatory Information:

    1. Regulation Classification section:
      Class II; 21 CFR § 862.1160-Bicarbonate/carbon dioxide test system
    1. Product Code:
      KHS
    1. Panel:
      75, Clinical Chemistry

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image is a logo with a green hexagon as the background. Inside the hexagon are the letters 'TC' in a silver, metallic-like font. The letters are large and centered within the hexagon, making them the focal point of the logo. The overall design is simple, modern, and professional.

H. Intended Use:

1. Intended used:

See Indications for Use below.

2. Indication(s) for Use:

Teco Carbon Dioxide Reagent Set is a device which is intended for measurement of Carbon Dioxide level in human serum, in vitro use only. Test results may provide information regarding the status in the assessment of acid-base balance of metabolic alkalosis or respiratory acidosis.

3. Special conditions for use statement(s):

This is an in vitro diagnostic test for professional use only.

4. Special instrument requirements:

Not applicable

I. Device Description

Teco Carbon Dioxide Reagent Set is a single reagent kit. Reagent contains Good's buffer, phosphoenolpyruvate (PEP), phosphoenolpyruvate carboxylase (PEPC), malate dehydrogenase (MDH), magnesium ions, NADH analog, nonreactive stabilizer, preservative, and Buffer. Approximately 90% of carbon dioxide present in serum is in the form of bicarbonate. The measurement of bicarbonate is useful in the assessment of disturbances of acid-base balance resulting from metabolic or respiratory causes.

J. Substantial Equivalence Information:

    1. Predicate device name(s): Pointe Scientific Inc., Carbon Dioxide Liquid Stable Reagent
    1. Predicate K number(s): K070251
    1. Compare with predicate:

Table 5.1 Teco Carbon Dioxide Reagent Set vs Pointe Scientific Inc., Carbon Dioxide Liquid Stable Reagent

Teco Carbon Dioxide Reagent Set(Proposed Device)Pointe Scientific Inc.,Carbon Dioxide LiquidStable Reagent (PredicateDevice)
ManufacturerTeco DiagnosticsPointe Scientific Inc.

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows a green hexagon with the letters "TC" in silver inside. The letters are stacked on top of each other, with "T" on top and "C" on the bottom. The hexagon has a slight gradient, with the top being slightly lighter than the bottom. The letters also have a gradient, with the top being slightly lighter than the bottom.

1268 N. Lakeview Ave., Anaheim, CA 92807, U.S.A. Phone: (714) 463-1111 Toll Free: 1-800-222-9880 Fax: (714) 463-1169 www.tecodiag.com

Section 5 - 510(k) SUMMARY

Intended UseThis product is to be used for thequantitative determination ofcarbon dioxide in human serumby spectrophotometric analysis.For in vitro diagnostic use only.Same
Measured ParametersBicarbonateSame
Intended SpecimenSerumSame
Test MethodologyEnzymatic; malate dehydrogenaseand phosphoenolpyruvatecarboxylaseSame
Instrument ModeAutomatic InstrumentSame
LinearitySerum: 8.7 to 40.0 mmol/LSerum: 2.0 to 40.0 mmol/L
Wavelength405 nmSame
InterferencesNo interference was observed byconjugated bilirubin up to 30mg/dL, total bilirubin up to 60mg/dL, hemoglobin up to 1,000mg/dL, triglycerides up to 2,000mg/dL, ascorbic acid up to6mg/dL, and intralipid up to 2,000mg/dL.No interference wasobserved by bilirubin up to20.0 mg/dL, hemoglobinup to 400 mg/dL andlipemia (intralipid) up to1,000 mg/dL.
Accuracy/Correlation(Serum)$y = 0.9785x + 0.2636$ ,$R^2 = 0.9925$ and $R = 0.9962$$y = 0.965x + 1.200$
Detection Limit(LoQ)1.77 mmol/L (LoQ)2.0 mmol/L
Storage Temperature2-8 °CSame

Teco Carbon Dioxide Reagent Set and the Pointe Scientific Carbon Dioxide Liquid Stable Reagent share the same technological characteristics including the testing parameters, test methodology, and storage.

K. Standard/Guidance Document Referred (if applicable):

  1. CLSI EP5-A3, Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline Third Edition.

  2. CLSI EP6-A, Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guidelines.

  3. CLSI EP7-A2, Interference Testing in Clinical Chemistry; Approved Guideline Second Edition.

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows a green hexagon with the letters 'TC' in silver inside. The letters are stylized and appear to be slightly three-dimensional. The background is plain white.

  1. CLSI EP9-A3, Measurement Procedure Comparison and Bias Estimation Using Patient Samples; Approved Guideline Third edition.

  2. CLSI EP17-A2, Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures: Approve Guideline Second Edition.

  3. CLSI EP25-A. Evaluation of Stability of In Vitro Diagnostic Reagents; Approved Guideline.

L. Test Principle:

The carbon dioxide in the human serum or serum based control sample reacts with phosphoenolpyruvate (PEP) in the presence of phosphoenolpyruvate carboxylase (PEPC) and magnesium to yield oxaloacetic acid (OAA) and phosphate. In the second reaction, in the presence of malate dehydrogenase (MDH), the reduced cofactor is oxidized by OAA. The reduced cofactor absorbs strongly at 405 nm, whereas its oxidized form does not. The difference in absorbance between the final reading and the blank, which is monitored at 405 nm/415 nm, is directly proportional to the total carbon dioxide concentration in the sample.

M. Performance Characteristics:

1. Analytical performance:

  • a. Precision/Reproducibility:
    In the accordance with CLSI EP5-A2 guideline, samples were tested in duplicate, 2 lots, and 2 runs per day for 20 days. The samples were unaltered human serums that were spiked with carbon dioxide concentrations and commercial serum based controls. The results were tabulated, and the data was analyzed for means, standard deviations (SDs), and percent coefficients of variation (%CVs). The precision result from one lot is summarized below.
Mean(mmol/L)RepeatabilityWithin-Laboratory Precision
SampleSD%CVSD%CV
QC Level 130.60.17820.5821.625.31
QC Level 220.50.70653.4261.879.07
Serum Level 112.00.12751.0650.695.75
Serum Level 226.60.15210.5710.853.21
Serum Level 337.40.19490.5210.591.58

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows a green hexagon with the letters "TC" in white inside. The letters are in a sans-serif font and appear to be slightly three-dimensional due to a subtle shading effect. The green hexagon has a slight gradient, with the top being a lighter shade of green than the bottom.

  • b. Linearity/assay reportable range:
    Linearity studies of Teco Carbon Dioxide Reagent Set were performed in accordance with the guidelines set forth in CLSI EP6-A. Linearity was tested using two lots of reagent with unaltered human serum samples. The samples were prepared by mixing a diluted patient serum sample (2 mmol/L) and a spiked patient serum sample (64 mmol/L) in order to obtain 11 different concentrations. Each sample assay was tested in 3 replicates using TC-MATRIX analyzer. The reagent revealed that the assay is linear up to 64 mmol/L. The linearity of Carbon Dioxide (Enzymatic) was analyzed over a measured range from 8.70 to 40.0 mmol/L. As a result, the linear regression analyses were obtained as below:

y= 1.0192x + 0.0351; R2=0.9993

The reportable range of the CO2 assay is 8.7 - 40.0 mmol/L

  • c. Traceability, Stability, Expected values (controls, calibrators, or methods): The CO2 method is traceable to NIST SRM 351. The CO2 assay is calibrated by comparing the change in absorbance of the unknown sample to the change in absorbance of the 30 mEq/L CO2 standard (from Integrated Diagnostics Group). The CO2 standard was previously cleared under K936245.
    It is recommended using an aqueous CO2 standard (30 mmol/L) or an appropriate serum calibrator. To monitor the reliability of results, it is recommended using two levels of control sera with known CO2 values to be run along with patient sample.

The stability claim was verified using an accelerated stability protocol. Accelerated stability of the CO2 reagent was determined based on Arrehnius Model equation. Based on the study, accelerated stability of the CO2 reagent was demonstrated to be at least 21 months if the CO2 reagent is tightly capped and stored at 2-8°C. Reagents in unopened vials are stable at 2-8°C until the expiry date printed on the label. The assay was calibrated using a valid calibrator and measured as sample in duplicates. The verification of the real time stability is on going.

  • d. Quality Control materials
    Users are instructed in the labeling to use quality control samples at least once a day and after each calibration and every time a new bottle of reagent is used. It is recommended to use at least two levels (low and high) of controls. Available controls are Lyphochek® Unassayed Chemistry Control (Human)

{9}------------------------------------------------

Image /page/9/Picture/1 description: The image shows a green hexagon with the letters "TC" in white inside. The letters are slightly shadowed, giving them a three-dimensional appearance. The hexagon is a solid green color and appears to be a logo or icon.

Levels 1 and 2 (Lot number: 27661 and 27662) and were previously cleared under K882486.

e. Detection limit:

Detection limit studies have been carried out in accordance with CLSI EP17-A2 guideline. Limit of Blank (LoB), Limit of Detection (LoD) and Limit of Quantification (LoQ) studies were performed on two lots of reagents tested on one TC-MATRIX analyzer (K073370). For the LoB and LoD studies, four (4) samples were tested in five (5) replicates over 3 days. For LoQ study, five (5) samples were tested in three (3) replicates over 3 days. The LoD pools consisted of 4 separate pools, and LoQ pools consisted of 5 separate pools near the current bottom of the serum assay ranges. The 9 LoD / LoQ serum pools were prepared by diluting patient serum with artificial serum. The LoB pool consisted of artificial serum base matrix with no carbon dioxide added. LoQ is the lowest concentration that can be detected with ≤ 10%CV. LoB, LoD and LoQ results are summarized in the following table:

AnalyteLoBLoDLoQ
CO2 (mmol/L)0.921.51.77

f. Analytical specificity:

Interference studies have been carried out in accordance with CLSI EP7-A2 guideline. Two (2) different concentrations of sodium bicarbonate (20 mmol/L and 35 mmol/L) human serum samples were used to evaluate interference testing using one TC-MATRIX analyzer (K073370) and two reagent lots. Pooled human serum samples with added potential interferents were tested in replicates of ten (10) and compared to a sample without interferent. Teco Carbon Dioxide Reagent Set was evaluated for interferences tested by bilirubin (Total and Conjugated), hemoglobin, triglyceride, ascorbic acid, and intralipid in serum samples. The following substances demonstrated no significant bias as defined as ±10% difference from the control results.

Interferentconcentration at which no significantinterference was observed
Hemoglobin1,000 mg/dL
Total Bilirubin60 mg/dL
Conjugate Bilirubin30 mg/dL
Triglycerides2,000 mg/dL
Intralipid2,000 mg/dL
Ascorbic Acid6 mg/dL

{10}------------------------------------------------

Image /page/10/Picture/1 description: The image is a logo featuring a green hexagon with the letters "TC" in silver. The letters are stylized with a slight 3D effect, giving them a metallic appearance. The green hexagon provides a solid background, making the letters stand out prominently.

  • g. Assay cut-off: Not applicable

2. Comparison studies:

  • a. Method comparison
    Serum CO2 correlation determination of the Pointe Scientific Inc. Carbon Dioxide Reagent (X) and Teco Carbon Dioxide Reagent Set (Y) was performed by using 176 patient serum samples ranging from 8.7 to 37.2 mmol/L of CO2 on TC-MATRIX analyzer (K073370), to obtain the following results

y = 0.9785x + 0.2636 mmol/L, R2 = 0.9925 and R=0.9962.

  • b. Matrix comparison Not applicable.
    1. Clinical studies: Not applicable
    1. Clinical cut-off: Not applicable
    1. Expected values/Reference range:

Reference interval of normal range (Serum): 23 - 33 mmol/L

The provided normal ranges are derived from Tietz NW. Clinical Guide to Laboratory Tests, 3rd ed. Philadelphia, PA: WB Saunders Company; 1995:110-111.

N. Conclusion:

The performance characteristics of the Carbon Dioxide Reagent Set were verified by comparison, precision, linearity, detection limit, specificity and interference, shelf life, and stress studies. Testing results indicate that the Carbon Dioxide Reagent Set performs satisfactorily when used appropriately, as outlined in the package insert. The result of performance studies demonstrates a substantial equivalency on performance between Teco Carbon Dioxide Reagent Set and the predicate device, Pointe Scientific Liquid Stable Reagent.

In summary, the submitted information in this premarket notification is complete and supports a substantial equivalence decision.

§ 862.1160 Bicarbonate/carbon dioxide test system.

(a)
Identification. A bicarbonate/carbon dioxide test system is a device intended to measure bicarbonate/carbon dioxide in plasma, serum, and whole blood. Bicarbonate/carbon dioxide measurements are used in the diagnosis and treatment of numerous potentially serious disorders associated with changes in body acid-base balance.(b)
Classification. Class II.