K Number
K170180
Device Name
Will3D
Manufacturer
Date Cleared
2017-05-08

(109 days)

Product Code
Regulation Number
892.2050
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Will3D is a software application used for the display and 3D visualization of medical image files from scanning devices, such as Dental CT scanner. It is intended for use by radiologists, clinicians, referring physicians and other qualified individuals to retrieve, process, render, review, store, print, assist in diagnosis, and distribute images utilizing standard PC hardware. Additionally, Will3D is a preoperative software application used for the simulation and evaluation of dental implants, orthodontic planning and surgical treatments.

Device Description

The Will3D is one of the components of a PACS (Picture Archiving and Communications System). The Will3D is the software application that provides image viewing and manipulation in the diagnostic imaging setting.

AI/ML Overview

The provided document is a 510(k) premarket notification for the medical device "Will3D," a software application for dental imaging. It outlines the device's intended use, technological characteristics, and comparison to predicate devices to establish substantial equivalence.

However, the document states: "No clinical studies were considered necessary and performed."

Therefore, the requested information regarding acceptance criteria, device performance, sample sizes, expert ground truth establishment, adjudication methods, multi-reader multi-case studies, standalone performance, and training set details cannot be provided from this document because no clinical studies were conducted to prove the device meets acceptance criteria derived from such studies. The equivalence is established purely on technological characteristics and intended use compared to predicate devices.

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol resembling an abstract human figure or a caduceus, composed of three interconnected profiles.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 8, 2017

HDX WILL CORP. Seung-Il Chun General Manager #105.201.202.203.204.38. Osongsaengmyeong 4-ro, Osong-eup, Heungdeok-gu, Cheongju-si. Chungcheongbuk-do. 28161 REPUBLIC OF KOREA

Re: K170180

Trade/Device Name: Will3D Regulation Number: 21 CFR 892.2050 Regulation Name: Picture Archiving and Communications System Regulatory Class: Class II Product Code: LLZ Dated: March 7, 2017 Received: March 14, 2017

Dear Seung-Il Chun:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply

{1}------------------------------------------------

with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Michael D'Hara

For

Robert A. Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K170180

Device Name Will3D

Indications for Use (Describe)

Will3D is a software application used for the display and 3D visualization of medical image files from scanning devices, such as Dental CT scanner. It is intended for use by radiologists, clinicians, referring physicians and other qualified individuals to retrieve, process, render, review, store, print, assist in diagnosis, and distribute images utilizing standard PC hardware. Additionally, Will3D is a preoperative software application used for the simulation and evaluation of dental implants, orthodontic planning and surgical treatments.

Type of Use (Select one or both, as applicable)|X Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for HDX Will. The logo consists of a blue square with a white "W" inside, followed by the text "HDX Will" in a bold, blue font. The "W" in the square is stylized with a green curve above it.

510(k) Summary

[As required by 21 CFR 807.92]

1. Date Prepared [21 CFR 807.92(a)(a)]

May 8, 2017

2. Submitter's Information [21 CFR 807.92(a)(1)]

Name of Sponsor:HDX WILL CORP.
Address:#105, 201, 202, 203, 204, 38,
Osongsaengmyeong 4-ro, Osong-eup, Heungdeok-gu,
Cheongju-si, Chungcheongbuk-do, 28161, Korea
Contact Name:Seung-Il Chun / General Manager
Telephone No.:+82-43-710-7318
Fax No.:+82-43-710-7312
Email Address:sichun@iwillmed.com
• Name of Manufacturer:Same as sponsor
- Address:Same as sponsor

3. Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)]

Trade Name:Will3D
-------------------------
  • . Common Name: PACS Software
  • Classification: .
Classification NameSystem, Image Processing, Radiological
Classification Regulation21 CFR 892.2050
Product CodeLLZ
Device ClassII
Classification PanelRadiology

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for HDX Will. The logo consists of a green square with a white "W" inside, followed by the text "HDX Will" in blue. The text is in a bold, sans-serif font.

4. Identification of Predicate Device(s) [21 CFR 807.92(a)(3)[

The identified predicate devices within this submission are shown as follow:

Primary Predicate

510(k) Number:K070803
Applicant:Anatomage Inc.
Classification Name:Imaging Processing System
Trade Name:InVivoDental™
Reference Predicate
510(k) Number:K113543
  • Applicant: CyberMed Inc. ● Classification Name: Imaging Processing System
  • Trade Name: OnDemand3DG2 ●

5. Description of the Device [21 CFR 807.92(a)(4)]

The Will3D is one of the components of a PACS (Picture Archiving and Communications System). The Will3D is the software application that provides image viewing and manipulation in the diagnostic imaging setting.

No.Feature/FunctionalityDescription
1FileRegister the required data in the database and search the data
2MPRDisplay the image visualized in 2D/3D and show the image from aselected direction
3Reconfigure the panoramic and cross-sectional images and draw thePanoramanerve in the panoramic image
4Display the Implant Model registered in a database and virtually placeImplantimplants in the ideal position
5TMIDisplay the 2D/ 3D images for a TMJ field
бOrthodonticDisplay a 2D image for horizontal symmetry check
73D CephProvide the cephalometric analysis and surgical simulation functions
8Map a 3D photo to the entered volume and compare the resultantFaceimage of a surgical simulation conducted in 3D Ceph against theSimulationexisting images
9SuperimpositionDisplay two different volumes on one screen
10EndoscopyProvide a virtual endoscopic function for airway observation
11ReportCreate the patient report

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for HDXWill. The logo consists of a green and blue square with a white "W" inside, followed by the text "HDXWill" in blue. The text is bold and slightly italicized.

6. Intended Use/Indications for Use [21 CFR 807.92(a)(5)]

Will3D is a software application used for the display and 3D visualization of medical image files from scanning devices, such as Dental CT scanner. It is intended for use by radiologists, clinicians, referring physicians and other qualified individuals to retrieve, process, render, review, store, print, assist in diagnosis, and distribute images utilizing standard PC hardware. Additionally, Will3D is a preoperative software application used for the simulation and evaluation of dental implants, orthodontic planning and surgical treatments.

7. Technological Characteristics [21 CFR 807.92(a)(6)]

There are no significant differences in the technological characteristics of this device compared to the predicate devices which adversely affect safety or effectiveness. Provided below is a table summarizing and comparing the technological characteristics of the Will3D and the predicate devices:

Proposed DevicePrimary PredicateReference Predicate
K NumberNot knownK070803K113543
Model NameWill3DInVivoDental™OnDemand3DG2
ManufacturerHDX WILL CORP.Anatomage Inc.CyberMed Inc.
ClassificationSystem, ImageSystem, ImageSystem, Image
NameProcessing, RadiologicalProcessing, RadiologicalProcessing, Radiological
Regulatory21 CFR 892.205021 CFR 892.205021 CFR 892.2050
Number
Product CodeLLZLLZLLZ
ClassificationIIIIII
Intended UseWill3D is a softwareInVivoDental is intendedThe OnDemand3D62 is
application used for thefor use as a front-endintended for use as a
display and 3Dsoftware interface forsoftware package which
visualization of medicalthe transfer of imagingloads DICOM images
image files frominformation from afrom CT, MR, X-Ray,
scanning devices, suchmedical scanner such asstores those and
as Dental CT scanner. Ita Dental CT scanner. It isprovides 3D
is intended for use byalso intended for use asvisualization and
radiologists, clinicians,a planning and20 analysis, various MPR
referring physicians andsimulation software in(Multi-Planar
other qualifiedthe placement of dentalReconstruction)
individuals to retrieve,implants, orthodonticsfunctions for further
process, render, review,and surgical treatment.rapid and precise
store, print, assist indiagnosis.
diagnosis, and distribute1) Functions for implant
images utilizingsurgery planning and
standard PC hardware.fabricating a surgical
Additionally, Will3D is aguide according to the
preoperative softwareplan
application used for the2) Functions to combine

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for HDX Will. The logo consists of a square with a green top half and a blue bottom half, with a white "W" in the center. To the right of the square is the text "HDX Will" in a bold, blue font. The text is slightly slanted to the right, giving the logo a dynamic appearance.

Proposed DevicePrimary PredicateReference Predicate
simulation andand show the image data
evaluation of dentalfrom different
implants, orthodonticmodalities such as CT,
planning and surgicalMRI and PET in one
treatments.window at the same
time
3) Functions to create
orthodontic analysis
using 3D volume data
4) Functions to sort and
save DICOM files, Project
Files and 2D image files
from different modality
[Panorama,
Cephalometric,
intraoral] under a single
patient name
Type of UsePrescription UsePrescription UsePrescription Use
ComponentStandalone softwareStandalone softwareStandalone software
Modality SupportDental CT scannerDental CT scannerCT, MR, and X-Ray
Operating SystemWindow 7 or higher andWindows XP andWindow XP/Vista/7/8
Mac os Yosemite orWindows 7
higher
Image
CommunicationDICOMDICOMDICOM
Standard
Feature/File,Section (MPR),File,
FunctionalityMPR,Volume RenderMPR,
Panorama,(Endoscopy),Panorama,
Implant,Arch SectionImplant,
TMJ,(Panorama),TMJ,
Orthodontic,Implant,3D Ceph,
3D Ceph,Restoration,Orthodontic,
Face Simulation,TML.Face Simulation and
Superimposition,Super Pano,Report
Endoscopy, andSuper Ceph,
ReportSuperimposition,
Gallery,
Model,
3D Analysis (3D Ceph),
Orthodontic,
MD studio,
Report, and
Stitching

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image contains the logo for HDXWill. The logo consists of a green square with a white "W" inside, followed by the text "HDXWill" in a bold, blue font. The text is slightly slanted to the right, giving the logo a dynamic appearance.

Non-Clinical Test Summarv

Will3D contains MODERATE level of concern software. Software was designed and developed according to a software development process and was verified and validated. Software information is provided in accordance with FDA guidance: the content of premarket submissions for software contained in medical devices, issued on May 11, 2005.

Clinical Test Summary:

No clinical studies were considered necessary and performed.

8. Substantial Equivalence [21 CFR 807.92(b)[1] and 807.92]

Will3D has the similar technical characteristics with the predicate devices. These devices are a dental imaging software program which provides 3D visualization that are reconstructed data from DICOM files.

The subject and primary predicate device share the same intended use, type of use, modality support and image communication standard. They also share equivalent operating systems. However, in terms of features and functionality, the subject device differs from the primary predicate device, and includes features and functionality not found in the primary predicate, such as File and Face Simulation. To address these differences, OnDemand3Di2 was introduced as the reference predicate.

In addition, these technological differences do not raise different questions of safety and effectiveness

9. Conclusion [21 CFR 807.92(b)(3)]

In according with the Federal Food & Drug and cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification HDX WILL CORP, concludes that the Will3D is substantially equivalent in safety and effectiveness to the predicate devices as described herein.

N/A