K Number
K170170

Validate with FDA (Live)

Manufacturer
Date Cleared
2017-02-16

(28 days)

Product Code
Regulation Number
878.4400
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Valleylab FT10 Energy Platform is a high frequency electrosurgical generator intended for use with monopolar and bipolar accessories for cutting and coagulating tissue. When used with compatible sealing devices, it is indicated for sealing vessels up to and including 7 mm, tissue bundles, and lymphatics. The generator can also be used with compatible resectoscopes for endoscopically controlled removal or coagulation of tissue using 0.9% NaCl solution as the irrigation medium.

The tissue fusion function has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use this function for these procedures.

Device Description

The Valleylab™ FT10 Energy Platform (VLFT10) provides radio frequency (RF) energy for monopolar and bipolar surgical applications, and tissue-fusion and vessel-sealing applications (LigaSure/vesselsealing function). It is a combination of a full-featured general-surgery electrosurgical unit and a bipolar vessel sealing system. The monopolar and bipolar sections, including the LigaSure section of the system, are isolated outputs that provide the appropriate power for cutting, desiccating, and fulgurating tissue during monopolar and bipolar surgery. The LigaSure/vessel-sealing section of the system provides power for vessel sealing.

The VLFT10 is used in hospitals and other health care facilities where surgical procedures are carried out.

The VLFT10 can be used with a variety of legally marketed accessories including monopolar and bipolar instruments and a newly introduced bipolar adapter, footswitches, and return electrode pads. The VLFT10 connects to electrical mains and operates at an input line frequency of 47-63 Hz.

AI/ML Overview

This document is a 510(k) premarket notification for the Valleylab FT10 Electrosurgical Platform, which is an electrosurgical generator. It describes modifications made to the device, primarily software updates, and the testing conducted to demonstrate that these changes do not affect the device's safety or performance.

However, the provided text does not contain the information requested regarding acceptance criteria and a study proving a device meets criteria, in the context of an AI/ML-driven medical device. The document is for an electrosurgical platform, not an AI/ML device. Therefore, it does not include details about:

  • A table of acceptance criteria and reported device performance (for AI/ML metrics like sensitivity, specificity, etc.)
  • Sample sizes for test sets (in the context of image data)
  • Data provenance (country of origin, retrospective/prospective)
  • Number of experts and their qualifications for ground truth establishment
  • Adjudication methods
  • Multi-reader multi-case (MRMC) comparative effectiveness studies
  • Standalone algorithm performance
  • Type of ground truth (pathology, outcomes, expert consensus for AI)
  • Sample size for training sets
  • How ground truth for training sets was established

Instead, the document focuses on engineering and performance characteristics of an electrosurgical device, such as:

  • Safety and essential performance (IEC 60601-1, IEC 60601-2-2)
  • Electromagnetic compatibility (IEC 60601-1-2)
  • Ex vivo testing (using porcine tissue for thermal effects)
  • System verification (non-IEC electrical, instrument compatibility, basic functionality)
  • Software verification and validation (FDA guidance for software in medical devices)
  • Usability testing (IEC 62366)

The document explicitly states: "This premarket submission did not rely on the assessment of clinical performance data to demonstrate substantial equivalence." This further confirms that the type of study and data requested (which is typical for AI/ML device submissions) was not performed or required for this particular submission.

Therefore, I cannot extract the requested information from the provided text.

{0}------------------------------------------------

Public Health Service

Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with the profiles overlapping each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 16, 2017

Covidien LLC Ms. Sharon McDermott Sr. Specialist, Regulatory Affairs 5920 Longbow Drive Boulder, Colorado 80301

Re: K170170

Trade/Device Name: Vallevlab FT10 Electrosurgical Platform Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: January 18, 2017 Received: January 19, 2017

Dear Ms. McDermott:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

Jennifer R. Stevenson -S

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K170170

Device Name Valleylab FT10 Energy Platform

Indications for Use (Describe)

The Valleylab FT10 Energy Platform is a high frequency electrosurgical generator intended for use with monopolar and bipolar accessories for cutting and coagulating tissue. When used with compatible sealing devices, it is indicated for sealing vessels up to and including 7 mm, tissue bundles, and lymphatics. The generator can also be used with compatible resectoscopes for endoscopically controlled removal or coagulation of tissue using 0.9% NaCl solution as the irrigation medium.

The tissue fusion function has not been shown to be effective for tubal coagulation for sterilization procedures. Do not use this function for these procedures.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/2 description: The image contains the logo for Covidien. The logo consists of a blue square with a white square inside, and the word "COVIDIEN" is written in blue to the right of the square. The font is sans-serif and the overall design is simple and modern.

510(k) Summary Date summary prepared: 02/13/2017

510(k) Submitter/Holder

Covidien 5920 Longbow Drive Boulder, CO 80301

Contact

Sharon McDermott Senior Specialist, Regulatory Affairs Telephone: 303-581-6789 Fax: 303-530-6313 Email: sharon.1.mcdermott@medtronic.com

Name of Device

Trade Name:Valleylab FT10 Energy Platform
Catalog Numbers:VLFT10GEN
Common Name:Electrosurgical Generator
Classification Name:Electrosurgical cutting and coagulation device and accessories (21 CFR 878.4400, Class II, GEI)

Predicate Device

Trade Name:Valleylab FT10 Energy Platform
Catalog Number:VLFT10GEN
Common Name:Electrosurgical Generator
510(k) Number:K151649 (cleared 10 Sept 2015)
Manufacturer:Covidien
Recalls:This device has not been subject to a design-related recall

Device Description

The Valleylab™ FT10 Energy Platform (VLFT10) provides radio frequency (RF) energy for monopolar and bipolar surgical applications, and tissue-fusion and vessel-sealing applications (LigaSure/vesselsealing function). It is a combination of a full-featured general-surgery electrosurgical unit and a bipolar vessel sealing system. The monopolar and bipolar sections, including the LigaSure section of the system, are isolated outputs that provide the appropriate power for cutting, desiccating, and fulgurating tissue during monopolar and bipolar surgery. The LigaSure/vessel-sealing section of the system provides power for vessel sealing.

The VLFT10 is used in hospitals and other health care facilities where surgical procedures are carried out.

The VLFT10 can be used with a variety of legally marketed accessories including monopolar and bipolar instruments and a newly introduced bipolar adapter, footswitches, and return electrode pads. The VLFT10 connects to electrical mains and operates at an input line frequency of 47-63 Hz.

{4}------------------------------------------------

Indications for Use

The Valleylab FT10 Energy Platform is a high frequency electrosurgical generator intended for use with monopolar and bipolar accessories for cutting and coagulating tissue. When used with compatible sealing devices, it is indicated for sealing vessels up to and including 7 mm, tissue bundles, and lymphatics. The generator can also be used with compatible resectoscopes for endoscopically controlled removal or coagulation of tissue using 0.9% NaCl solution as the irrigation medium.

The tissue fusion function has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use this function for these procedures.

Comparison of Technological Characteristics with the Predicate Device

The Valleylab FT10 with the proposed software modifications has the similar technological and performance characteristics as the predicate Valleylab FT10 cleared in K151649. Both versions of the energy platform are a combination of a full-featured general-surgery electrosurgical system and a vessel sealing system. The VLFT10 with software version 2.0 provides incremental improvements over the predicate while maintaining the same basic functionality and intended use. These are summarized below.

  • . Monopolar capabilities have been expanded with the addition of a Shared Coag mode, a mode that has been available in other electrosurgical generators such as Covidien's Force FX Electrosurgical Generator
  • Enabling a second standard bipolar port via insertion of an adapter into the existing ● LigaSure/Bipolar port
  • Updates to existing Auto-bipolar mode to improve user experience
  • GUI changes to support second bipolar port and instruments under development

Performance Characteristics

The modifications made to the Valleylab FT10 Energy Platform were found to not affect safety or performance through design verification testing. which confirmed the conformance to applicable technical design specifications and performance requirements, including requirements associated with industry safety and performance standards, as follows:

  • Basic safety and essential performance in accordance with IEC 60601-1:2005/A1:2012 and IEC ● 60601-2-2:2009
  • . Electromagnetic compatibility in accordance with IEC 60601-1-2:2014
  • . Ex vivo testing using porcine tissue showed comparable performance with regard to thermal effects
  • . System verification (Non-IEC electrical, instrument compatibility, basic functionality) showed that the VLFT10 has all required functionality and that it meets system specifications.
  • Software verification and validation testing was conducted in accordance with FDA's, Guidance . for the Content of Premarket Submissions for Software Contained in Medical Devices
  • Usability testing in accordance with IEC 62366: 2015Medical devices Application of . usability engineering to medical devices showed that representative surgical tasks with the updated software.

{5}------------------------------------------------

Clinical Studies

This premarket submission did not rely on the assessment of clinical performance data to demonstrate substantial equivalence.

Conclusions

Verification and validation activities demonstrate the Valleylab FT10 with software version 2.0 is substantially equivalent to the predicate Valleylab FT10 Energy Platform cleared in K151649. The intended use of the Valleylab FT10 Energy Platform was not altered. Moreover, the results of testing demonstrate that the software modifications do not affect the safety or performance of the energy platform.

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.