(158 days)
PHT-65LHS is intended to produce panoramic, cephalometric or 3D digital x-ray images. It provides diagnostic details of the dento-maxillofacial, ENT, sinus and TMJ for adult and pediatric patients. The system also utilizes carpal images for orthodontic treatment. The device is to be operated by healthcare professionals.
Green16 / Green18 (Model: PHT-65LHS) is an advanced 5-in-1 digital X-ray imaging system that incorporates PANO, CEPH (Optional), CBCT, 3D MODEL Scan and 3D PHOTO (Optional) imaging capabilities into a single system. PHT-65LHS, a digital radiographic imaging system, acquires and processes multi-FOV diagnostic images for dentists and ENT specialists. PHT-65LHS is a complete digital X-ray system equipped with imaging viewers, X-ray generator and a dedicated SSXI detector. The digital CBCT system is based on a CMOS digital X-ray detector. The CMOS CT detector is used to capture 3D radiographic images of head, neck, oral surgery, implant and orthodontic treatment. PHT-65LHS can also acquire 2D diagnostic image data in conventional panoramic and cephalometric modes.
The provided text is a 510(k) Summary for the Vatech Green16/Green18 (Model: PHT-65LHS) dental X-ray system. It describes the device, its intended use, and compares it to a predicate device (Green Smart, PHT-35LHS) to establish substantial equivalence.
However, the document does not describe a study that proves the device meets specific acceptance criteria related to a device's performance in analyzing medical images or making diagnostic predictions (e.g., an AI/ML device). Instead, it describes:
- Acceptance criteria for the imaging system itself: This refers to the fundamental image quality metrics (like MTF, DQE, NNPS, Contrast, Noise, CNR) and safety standards required for an X-ray imaging device to be cleared by the FDA for its intended purpose (producing panoramic, cephalometric, or 3D digital X-ray images).
- Performance data comparing the new device to a predicate device: This is to show that the new device's imaging capabilities are equivalent to or better than a previously cleared device.
Therefore, many of the requested items (e.g., sample size for training/test sets, number of experts for ground truth, adjudication methods, MRMC studies, standalone algorithm performance, types of ground truth for AI/ML) are not applicable to this document as it pertains to the clearance of an imaging hardware system, not an AI/ML-driven diagnostic device.
Here's a breakdown of what can be extracted and what is not present, based on the provided text:
1. Table of acceptance criteria and reported device performance:
| Performance Parameter | Acceptance Criteria (Implied: Equivalent or Better than predicate) | Reported Device Performance (Subject Device vs. Predicate) |
|---|---|---|
| PHT-65LHS Detectors (Xmaru1314CF & Xmaru1515CF) | ||
| Mod. Transfer Function (MTF) | Equivalent to Xmaru1404CF-Plus (predicate) | Similar resolution performance at all spatial frequencies |
| Detective Quantum Efficiency (DQE) | Equivalent to Xmaru1404CF-Plus (predicate) | At low spatial frequency (~0.5 lp/mm), Xmaru1314CF & Xmaru1515CF: 49% (4x4binning) vs. Xmaru1404CF-Plus: 52% (4x4binning). However, overall "similar" performance claimed. |
| Normalized Noise Power Spectrum (NNPS) | Equivalent to Xmaru1404CF-Plus (predicate) | Similar image quality at all spatial frequencies |
| Pixel Resolution (CT&PANO) | Equivalent to predicate (5 lp/mm -2x2 binning detector spec, 2.5 lp/mm -4x4 binning system spec) | Identical for 2x2 binning. Same diagnostic image quality as predicate using 4x4 binning. |
| Pixel Size (CT&PANO) | Equivalent to predicate (99 µm -2X2 binning, 198 µm - 4X4 binning) | Identical |
| Pixel Resolution (CEPH) | Equivalent to predicate (5 lp/mm-Non binning, 2.5 lp/mm -2x2 binning) | Identical |
| Pixel Size (CEPH) | Equivalent to predicate (100 um- Non binning, 200 um -2X2 binning) | Identical |
| CT Image Quality (measured with iterative reconstruction algorithm) | Adherence to IEC 61223-3-5 standard; Equivalent or Better than predicate | Demonstrated "equivalent or better" general image quality |
| Contrast | Equivalent or Better than predicate | Achieved |
| Noise | Equivalent or Better than predicate | Achieved |
| CNR | Equivalent or Better than predicate | Achieved |
| MTF | Equivalent or Better than predicate | Achieved |
| Dosimetric Performance (DAP) | Equivalent to predicate under same exposure conditions | CEPH/PANO: same. CBCT: FOV 5x5 mode equivalent. |
| General Image Quality (Clinical judgment) | Equivalent or Better than predicate | Demonstrated "equivalent or better" |
2. Sample size used for the test set and data provenance:
- The document mentions "Non-Clinical Test results" and "Clinical consideration and image quality evaluation report" but does not specify sample sizes for these tests.
- Data provenance (country of origin, retrospective/prospective) is not mentioned beyond the manufacturer being VATECH Co., Ltd. in Korea.
3. Number of experts used to establish the ground truth for the test set and qualifications:
- Not applicable as this is not an AI/ML diagnostic device requiring expert-established ground truth for a test set. The "Clinical consideration and image quality evaluation report" implies human evaluation, but details are absent.
4. Adjudication method for the test set:
- Not applicable.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, and effect size:
- Not applicable. This is not an AI/ML CAD device for human readers.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not applicable. This is an X-ray imaging system, not an algorithm, so its "standalone" performance is its fundamental imaging capability, which is reported via metrics like MTF, DQE, NNPS.
7. The type of ground truth used:
- For the technical performance of the imaging system, the "ground truth" is established by physical phantoms and measurement protocols defined by international standards (e.g., IEC 61223-3-5).
- For "Clinical consideration and image quality evaluation," the "ground truth" would be the subjective assessment of image quality by healthcare professionals, but no details are provided on how this evaluation was structured.
8. The sample size for the training set:
- Not applicable. This device is not an AI/ML product developed with a training set.
9. How the ground truth for the training set was established:
- Not applicable.
{0}------------------------------------------------
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 16, 2017
Vatech Co., Ltd. % Dave Kim Medical Device Regulatory Affairs Mtech Group 8310 Buffalo Speedway Houston, Texas 77025
Re: K170066
Trade/Device Name: Green16/Green18 (Model: PHT-65LHS) Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: Class II Product Code: OAS Dated: December 1. 2016 Received: January 9, 2017
Dear Dave Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Michael D. O'Hara
For
Robert A. Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known)
Device Name Green16/Green18 (Model: PHT-65LHS)
Indications for Use (Describe)
PHT-65LHS is intended to produce panoramic, cephalometric or 3D digital x-ray images. It provides diagnostic details of the dento-maxillofacial, ENT, sinus and TMJ for adult and pediatric patients. The system also utilizes carpal images for orthodontic treatment. The device is to be operated by healthcare professionals.
Type of Use (Select one or both, as applicable)
| ✖ Prescription Use (Part 21 CFR 801 Subpart D) |
|---|
| ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary
1. 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
2. Date 510K Summary prepared: May 15, 2017
3. Administrative Information
| Official Correspondent: | Dave Kim / Mtech Group8310 Buffalo Speedway, Houston, TX 77025Tel: +713-467-2607Fax: +713-583-8988Contact person: Mr. Dave Kim (davekim@mtech-inc.net) |
|---|---|
| 510(k) Submitter: | VATECH Co., Ltd.Address: 13, Samsung 1-ro 2-gil, Hwaseong-si, Gyeonggi-do, 18449, KoreaTel: +82-31-379-9492Fax: +82-31-379-9400Contact person: Daniel Kim / Manager (daniel.kim@vatech.co.kr) |
4. Device Information
| Type of 510(k) Submission: | Special |
|---|---|
| Trade or Proprietary Name: | Green16/Green18 (Model: PHT-65LHS) |
| Common or Usual Name: | Dental Computed Tomography X-ray System |
| Regulation Classification: | Computed tomography x-ray system (21 CFR 892.1750) |
| Product Code: | OAS |
| Class of Device: | Class II |
| Panel: | Radiology |
5. Predicate Device Information
| Manufacturer: | VATECH Co., Ltd. |
|---|---|
| Predicate device: | Green Smart (PHT-35LHS) / K162660 |
| Common or Usual Name: | X-Ray, Tomography, Computed, Dental |
| Regulation Classification: | Computed tomography x-ray system (21 CFR 892.1750) |
| Product Code: | OAS |
| Class of Device: | Class II |
| Panel: | Radiology |
※ This predicate has not been subject to a design-related recall. No reference devices were used in this submission.
{4}------------------------------------------------
6. Device Description
Green16 / Green18 (Model: PHT-65LHS) is an advanced 5-in-1 digital X-ray imaging system that incorporates PANO, CEPH (Optional), CBCT, 3D MODEL Scan and 3D PHOTO (Optional) imaging capabilities into a single system.
PHT-65LHS, a digital radiographic imaging system, acquires and processes multi-FOV diagnostic images for dentists and ENT specialists. PHT-65LHS is a complete digital X-ray system equipped with imaging viewers, X-ray generator and a dedicated SSXI detector.
The digital CBCT system is based on a CMOS digital X-ray detector. The CMOS CT detector is used to capture 3D radiographic images of head, neck, oral surgery, implant and orthodontic treatment.
PHT-65LHS can also acquire 2D diagnostic image data in conventional panoramic and cephalometric modes.
Key components of the device
-
- Green16/Green18 (Model: PHT-65LHS) digital x-ray equipment
-
- SSXI detector: Xmaru1314CF, Xmaru1515CF, Xmaru2602CF
| Description | |||
|---|---|---|---|
| Item | CBCT/PANO | Green18 | CEPH |
| Green16 | Green16/Green18 | ||
| Model | Xmaru1314CF | Xmaru1515CF | Xmaru2602CF |
| Detector Type | CMOS photodiode array | CMOS photodiode array | CMOS photodiode array |
| Pixel Size | 99 μm -2X2 binning(detector spec)198 μm - 4X4 binning(system spec) | 99 μm -2X2 binning(detector spec)198 μm - 4X4 binning(system spec) | 100 μm- Non binning(detector spec)200 µm -2X2 binning(system spec) |
| Active Area | CBCT-127.5 x 135.8 mmPANO-5.9 x 135.8 mm | CBCT-155.2 x 145.7 mmPANO-5.9 x 135.8 mm | 15.6 x 259 mm |
| Frame Rate | ~108 fps (4x4 Binning) | ~95 fps (4x4 Binning) | ~330 fps (2x2 Binning) |
| Analogue-DigitalConversion | 14 bits | 14 bits | 14 bits |
| Converter | Csl:Ti | Csl:Ti | Csl:Ti |
| Energy Range | 50 ~ 120 kV | 50 ~ 120 kV | 50 ~ 120 kV |
| Readout Type | Charge amplifier array | Charge amplifier array | Charge amplifier array |
| Video Output | Optic | Optic | Optic |
3) X-ray generator
| Item | Description | ||
|---|---|---|---|
| High VoltageGenerator | Model | DG-07E22T2 | |
| Rated output power | 1.6 kW | ||
| Type | Inverter | ||
| Normal/Pulse | kV | ||
| mA | |||
| Cooling | Air (Optional fan cooling, ≥ 60 °C) | ||
| Total filtration | Min. 2.5 mm Al |
{5}------------------------------------------------
| Item | Description | |
|---|---|---|
| Added filtration | 1.5 mm Al (Fixed) / PANO and CEPH mode1.5 mm Al (Fixed) + 3.0 mm Al (Automatically added) / CBCT mode | |
| X-ray Tube | Manufacturer | Toshiba |
| Model | D-052SB (Stationary Anode type) | |
| Focal spot size | 0.5 x 0.5 mm | |
| Target Angle | 5 degree | |
| Inherent Filtration | At least 0.8 mm Al equivalent at 50 kV | |
| Anode Heat Content | 35 kJ | |
| Duty Cycle | 1:60 or more (Exposure time : Interval time) |
4) PC system
| Item | Description |
|---|---|
| Operating System | Windows 10 Professional 64-Bit OS |
| CPU | Intel i7-6700 |
| RAM | 16GB |
| HDD | 500GB SATA 1st HDD |
| Graphics board | NVIDIA Geforce GTX1060 6GB or greater |
| Ethernet interface | 10/100/1000 Mbps, RJ-45, 2 Port |
| Serial Port (RS232) | HP Serial Port Adapter kit |
| Power Supply | ≥ 700 Watts (90% efficient) |
| Slots | 1 PCI Express Gen3 x16 slot |
| 1 PCI Express Gen2 x4 Slot | |
| 1 PCI Express Gen2 x1 Slot |
5) Imaging software
| Item | Description |
|---|---|
| 2D Image Viewing Program | EasyDent (Cleared under K162660) |
| EzDent-i (K161117) | |
| 3D Image Viewing Program | Ez3D Plus (Cleared under K162660) |
| Ez3D-i (K161246) |
7. Indication for use
Green16 / Green18 (Model: PHT-65LHS) is intended to produce panoramic, cephalometric or 3D digital x-ray images. It provides diagnostic details of the dento-maxillofacial, ENT, sinus and TMJ for adult and pediatric patients. The system also utilizes carpal images for orthodontic treatment. The device is to be operated by healthcare professionals.
8. Comparison of Technological characteristics with the predicate device
| Subject Device | Predicate Device | |||
|---|---|---|---|---|
| Device Name | Green 16 / Green 18(Model: PHT-65LHS) | Green Smart(Model: PHT-35LHS) | ||
| Applicant Name | VATECH Co., Ltd. | VATECH Co., Ltd. | ||
| 510(k) Number | N/A | K162660 | ||
| Device Classification Name | X-Ray, Tomography, Computed, Dental | X-Ray, Tomography, Computed, Dental | ||
| Classification Product Code | OAS | OAS | ||
| Regulation Number | Subject Device21 CFR 892.1750 | Predicate Device21 CFR 892.1750 | ||
| Indications for Use | PHT-65LHS is intended to producepanoramic, cephalometric or 3Ddigital x-ray images. It providesdiagnostic details of the dento-maxillofacial, ENT, sinus and TMJ foradult and pediatric patients. Thesystem also utilizes carpal images fororthodontic treatment. The device isto be operated by healthcareprofessionals. | PHT-35LHS is a computed tomographyx-ray system intended to producepanoramic, cephalometric or cross-sectional images of the oral anatomyby computer reconstruction of x-rayimage data from the same axial planetaken at different angles. It providesdiagnostic details of the maxillofacialareas for dental treatments in adultand pediatric dentistry. The systemalso utilizes carpal images fororthodontic treatment. The device isoperated and used by physicians,dentists and x-ray technicians. | ||
| Performance Specification | Panoramic, Cephalometric and computedtomography | Panoramic, Cephalometric and computedtomography | ||
| Input Voltage | AC 100 - 240 V | AC 100 - 240 V | ||
| X-Ray source | D-052SB | D-052SB | ||
| Tube Voltage | 60 - 99 kV | 60 - 99 kV | ||
| Tube Current | 4 - 16 mA | 4 - 16 mA | ||
| Focal Spot Size | 0.5 x 0.5 mm | 0.5 x 0.5 mm | ||
| Scan Time | Max. 14.1 s | Max. 18 s | ||
| Slice Width | Min. 0.1 mm | Min. 0.1 mm | ||
| Total Filtration | Min. 2.5 mm Al | Min. 2.5 mm Al | ||
| Mechanical | Compact design | Compact design | ||
| Electrical | LDCP logic circuit | LDCP logic circuit | ||
| Software | DICOM 3.0 Format compatible | DICOM 3.0 Format compatible | ||
| 2D Image Viewing Program | EasyDent (Cleared under K162660)EzDent-i (K161117) | EasyDent (Cleared under K162660)EzDent-i (K161117) | ||
| 3D Image Viewing Program | Ez3D Plus (Cleared under K162660)Ez3D-i (K161246) | Ez3D Plus (Cleared under K162660)Ez3D-i (K161246) | ||
| Anatomical Sites | Maxillofacial | Maxillofacial | ||
| ImageReceptor | CT&PANO | Xmaru1515CFXmaru1314CF | Xmaru1404CF-Plus | |
| CEPH | Xmaru2602CF | Xmaru2602CF | ||
| Size of Imaging Volume | Max. 180 x 100 mmXmaru1515CFMax. 160 x 90 mmXmaru1314CF | Max. 100 x 8.5 mm | ||
| PixelResolution | CT&PANO | 5 lp/mm -2x2 binning (detector spec)Xmaru1515CF2.5 lp/mm -4x4 binning (system spec) | 5 lp/mm -2x2 binning (detector spec) | |
| 5 lp/mm -2x2 binning (detector spec)Xmaru1314CF2.5 lp/mm -4x4 binning (system spec) | 2.5 lp/mm -4x4 binning (system spec) | |||
| CEPH | 5 lp/mm-Non binning (detector spec)2.5 lp/mm -2x2 binning (system spec) | 5 lp/mm-Non binning (detector spec)2.5 lp/mm -2x2 binning (system spec) | ||
| Subject Device | Predicate Device | |||
| Pixel Size | CT&PANO | Xmaru1515CF | 99 µm -2X2 binning(detector spec)198 μm - 4X4 binning(system spec) | 99 µm -2X2 binning (detector spec)198 µm - 4X4 binning (system spec) |
| Xmaru1314CF | 99 um -2X2 binning(detector spec)198 μm - 4X4 binning(system spec) | |||
| CEPH | 100 um- Non binning (detector spec)200 um -2X2 binning (system spec) | 100 um- Non binning (detector spec)200 um -2X2 binning (system spec) |
{6}------------------------------------------------
|
{7}------------------------------------------------
9. Performance Data
Summary of Performance Testing
The Green 16/Green18(Model: PHT-65LHS) digital X-ray system described in this 510(k) is similar to the predicate device in terms of indications for use, materials, safety characteristics, and X-ray source.
The following information further substantial equivalence between the subject device and predicate device:
The fundamental technological characteristics of the subject and predicate device are similar. The imaging modes are similar; PANO, CEPH (Optional), CBCT, 3D MODEL Scan and 3D PHOTO (Optional) All viewing software programs have been cleared with previous 510k submissions; EasyDent (K162660), Ez3D Plus (K162660), EzDent-i (K161117) and Ez3D-i (K161246).
The sponsor tested the subject device in a laboratory and provided a non-clinical performance report. The same test protocol was used to test the performance of the subject and the predicate device for comparison. The sponsor certifies that adequate design and development controls (according to 21 CFR 820.30) were in place for manufacturing the subject device. The CT image reconstruction algorithm (Iterative Recon) and MAR (Metal Artifact Reduction) algorithm are the same for the subject device and the predicate device.
The differences are as follows.
- । The subject device is equipped with new detectors, Xmaru1314CF and Xmaru1515CF.
- । The availability of diagnostics for ENT treatment
Green16/Green18 (Model: PHT-65LHS), a digital radiographic imaging system is equipped with Xmaru1314CF and Xmaru1515CF. Xmaru1314CF and Xmaru1515CF is a new SSXI detector, which is used to capture an image in panoramic, CBCT and Model Scan mode.
Based on Non-Clinical Test results of Xmaru1314CF and Xmaru1515CF for the subject device, the CMOS panel of Xmaru1314CF and Xmaru1515CF is exactly same to that of the Xmaru1404CF-Plus. Therefore, the testing image patterns of the new sensor show no aliasing phenomenon throughout the same spatial frequency as the predicate device.
All detectors are technically capable of 2x2 binning. In practice, both subject (PHT-65LHS) and predicate (PHT-35LHS) X-ray system capture images in 4x4 binning mode, which provides the same diagnostic image quality for the subject and predicate device.
Pixel resolutions and pixel sizes in 2x2 binning for subject device and predicate device detector are identical. Moreover, the Xmaru1515CF have performed similarly to the Xmaru1404CF-
{8}------------------------------------------------
Plus in terms of the Modulation Transfer Function (MTF), Detective Quantum Efficiency (DQE) and Normalized Noise Power Spectrum (NNPS). At a low spatial frequency (~0.5 lp/mm), both Xmaru1314CF an Xmaru1515CF have the same DQE of 49% (4x4binning) whereas Xmaru1404CF-Plus has 52% (4x4binning). The MTF and NNPS curves show that the Xmaru1314CF and Xmaru1515CF have similar resolution performance and image quality compared to Xmaru1404CF-Plus at all spatial frequencies. The performance parameters for all detectors have shown similar results.
The acceptance test was performed according to the requirements of 21 CFR Part 1020.30. 1020.33 and IEC 61223-3-5, international performance standard for computed tomography X-ray system. Contrast, Noise, CNR, and MTF, the representative indicators for CT image quality were measured with iterative reconstruction algorithm for the new X-ray equipment. The results demonstrated that the general image quality of the subject device is equivalent or better than the predicate device.
In addition, the dosimetric performance of the subject device and the predicate device was compared in terms of DAP. With the identical FDD(Focal Spot to Detector Distance), detector specifications, DAP measurement in the CEPH and PANO mode of each device under the same X-ray exposure conditions (exposure time, tube voltage, tube current) was the same.
In CBCT mode, the direct comparison of the dosimetric performance for each mode available in the subject and predicate device is difficult due to different exposure conditions such as the exposure time. DAP of the FOV 5x5 mode of the subject device was equivalent to the predicate device. Any user adjustment of the exposure setting in normal and fast mode of the subject device should consider the patient exposure level to be as low as possible.
Moreover, PANO/CEPH/CBCT images from the subject and predicate devices are evaluated in the Clinical consideration and image quality evaluation report. The results demonstrated that the general image quality of the subject device is equivalent or better than the predicate device.
Software Verification and Validation Testing
Software verification and validation were conducted and documented as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level of concern, since a failure or latent flaw in the software would not directly result in serious injury or death to the patient or operator.
Green 16/Green18(Model: PHT-65LHS) provides the following imaging viewer programs;
- -2D Image viewing program: EasyDent(K162660), EzDent-i(K161117)
- । 3D Image viewing program: Ez3D Plus(K162660), Ez3D-i(K161246)
Safety, EMC and Performance Data
Electrical, mechanical, environmental safety and performance testing according to standard IEC 60601-1(Ed. 3, 2005), IEC 60601-1-3 (Ed. 2, 2008), IEC 60601-2-63 (Ed. 1, 2012) were performed, and EMC testing were conducted in accordance with standard IEC 60601-1-2.
The manufacturing facility is in conformance with the relevant EPRC standards as specified in 21 CFR 1020.30, 31, and 33 and the records are available for review.
Green 16/ Green18(Model: PHT-65LHS) conforms to the provisions of NEMA PS 3.1-3.18, Digital Imaging and Communications in Medicine (DICOM) Set.
{9}------------------------------------------------
Non-clinical consideration report according to FDA Guidance "Guidance for the submissions of 510(k)'s for Solid State X-ray Imaging Devices" was provided.
Bench testing according to FDA Guidance "Format for Traditional and Abbreviated 510(k)s, section 18, Performance Testing - Bench" were performed.
Acceptance test and CT image evaluation report according to IEC 61223-3-4 and IEC 61223-3-5 were also performed.
All test results were satisfactory.
10. Conclusions
The proposed device and the predicate device have similar indications for use and demonstrated similar technical characteristics. As demonstrated in the performance test, the Xmaru1314CF and Xmaru1515CF performed similar or better in comparison with the predicate device in various performance parameters such as DQE, MTF and NNPS. In addition, the CT image evaluation of Contrast, Noise, CNR, and MTF further demonstrated the performance equivalency between the subject and predicate device. Quality assurance procedures are adhered to, and the specifications and functional requirements were met as the test results indicated.
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification, VATECH Co., Ltd. concludes that Green 16/Green18|Model: PHT-65LHS) is substantially equivalent to the predicate device as described herein.
N/A