K Number
K170027

Validate with FDA (Live)

Device Name
Inflation Device
Date Cleared
2017-10-06

(276 days)

Product Code
Regulation Number
870.1650
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The inflation device is intended to be used in PTCA or PTA procedures to create and monitor pressure in the balloon and to deflate the balloon dilatation catheter.

Device Description

The inflation device consists of barrel, threaded plunger assemble with a handle, a trigger, pressure gauge, an outer shell assembly that retains the internal components, and a pressure connecting tubing with rotating adapter. Also enclosed is a stopcock to aid in preparation and use of device. The inflation device is 20 mL disposable device capable of producing a maximum pressure of 30 atm.

The inflation device is sterilized by EO.

AI/ML Overview

The provided text describes a 510(k) premarket notification for an "Inflation Device". This document focuses on demonstrating substantial equivalence to a predicate device through non-clinical bench testing, rather than presenting a clinical study with detailed acceptance criteria for diagnostic performance. Therefore, many of the requested elements are not applicable to this type of submission.

Here's a breakdown of the available information:

1. Table of Acceptance Criteria and Reported Device Performance

The submission does not present acceptance criteria in terms of diagnostic performance metrics (e.g., sensitivity, specificity, AUC) or a specific clinical outcome. Instead, it refers to compliance with established international standards for medical devices and device specifications.

Acceptance Criteria CategorySpecific Standard/RequirementReported Device Performance
Functional PerformanceISO 594-1:1986 (Conical Fittings, General Requirements)Device complies with the standard.
ISO 594-2:1998 (Conical Fittings, Lock Fittings)Device complies with the standard.
Device specifications (e.g., volume, pressure range, mechanism of operation)Verified through bench tests; performance is equivalent to predicate device (20 mL, 0-30 ATM).
BiocompatibilityISO 10993 series StandardsConforms to requirements (No hemolysis, No Acute Systemic Toxicity, No cytotoxicity, No intracutaneous reactivity, No Sensitization).
SterilizationSterility Assurance Level (SAL)10-6
Sterilization MethodEO Sterilization
Sterilization ValidationConforms to ISO 11135
PackagingPackage IntegrityConforms to ISO 11607
Material ResidualsEO ResidualConforms to ISO 10993-7
Other SafetyPyrogenicityNon-pyrogenic
LabelingFDA RequirementsConforms to FDA Requirements

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: The document does not specify a "test set" in the context of clinical data. The tests performed are bench tests on the device itself (e.g., for mechanical properties, material testing). The number of units tested per bench test is not detailed.
  • Data Provenance: Not applicable as no human data is used. The tests are performed on the manufactured device.

3. Number of Experts Used to Establish Ground Truth and Qualifications

  • Not applicable. There is no ground truth established by experts in this context as it's a device performance study, not a diagnostic accuracy study relying on human interpretation.

4. Adjudication Method

  • Not applicable. No expert review or adjudication process is described as it's not a clinical study involving human assessment.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No. This is a 510(k) submission for a physical medical device, not an AI software or diagnostic imaging tool. Therefore, a MRMC comparative effectiveness study is not performed or relevant.

6. Standalone Performance

  • Yes, in a way. The "standalone" performance here refers to the device's ability to meet its technical specifications and regulatory standards independently. The non-clinical bench tests assess the device's functional and safety characteristics in a standalone manner.

7. Type of Ground Truth Used

  • The "ground truth" for this device's performance is compliance with established international standards (ISO) and the manufacturer's own validated specifications. For example, for conical fittings, the ground truth is adherence to ISO 594-1 and ISO 594-2 standards. For biocompatibility, it's the outcome of tests against ISO 10993 standards.

8. Sample Size for the Training Set

  • Not applicable. There is no "training set" as this is not an AI/machine learning model.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol, composed of three overlapping profiles facing to the right. The profiles are white against a dark background, creating a silhouette effect. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 6, 2017

Shanghai Kindly Medical Instruments Co., Ltd. Jeffery Hui Official Correspondent No. 925 Jinyuan Yi Rd Shanghai, 201803 CN

Re: K170027

Trade/Device Name: Inflation Device Regulation Number: 21 CFR 870.1650 Regulation Name: Angiographic Injector And Syringe Regulatory Class: Class II Product Code: MAV Dated: September 5, 2017 Received: September 7, 2017

Dear Jeffery Hui:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Kenneth J. Cavanaugh -S

for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K170027

Device Name Inflation Device

Indications for Use (Describe)

The inflation device is intended to be used in PTCA or PTA procedures to create and monitor pressure in the balloon and to deflate the balloon dilatation catheter.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary - K170027

This 510(k) Summary is being submitted in accordance with the requirements of the guidance The 510(k) Program and 21 CFR 807.92.

510(k) Number: K170027

    1. Date of Submission: Aug.03, 2017

2. Submitter

Shanghai Kindly Medical Instruments Co., Ltd. No. 925, Jinyuan vi Road, Shanghai, 201803, China Establishment Registration Number: 3009605245 Contact Person: Xu Jianhai Position: RA Supervisor Tel .: +086-021-59140056 Fax: +086-021-59140056 Email: xujianhai@kdlchina.net

3. Proposed Device

Proposed Device Name: Inflation Device Proposed Device Model: AI25 Classification: Class II Classification Name: Angiographic Injector And Syringe Regulation Number: 21 CFR 870.1650 Review Panel: Cardiovascular Product Code: MAV

4. Predicate device

  • 510(k) Number: K102648 a. Product Name: ANT Inflation Device/ ANT Inflation Device Compact Pack Manufacturer: Shenzhen ANT Hi-Tech Industrial Co., Ltd.

5. Device description

The inflation device consists of barrel, threaded plunger assemble with a handle, a trigger, pressure gauge, an outer shell assembly that retains the internal components, and a pressure connecting tubing with rotating adapter. Also enclosed is a stopcock to aid in preparation and use of device. The inflation device is 20 mL disposable device capable of producing a maximum pressure of 30 atm.

The inflation device is sterilized by EO.

{4}------------------------------------------------

6. Indications for Use Statement:

The inflation device is intended to be used in PTCA or PTA procedures to create and monitor pressure in the balloon and to deflate the balloon dilatation catheter.

7. Non-clinical Test Conclusion

Bench tests were conducted to verify that the proposed device met all device specification as were substantially equivalent (SE) to the predicate device. The test result demonstrates that the proposed devices comply with the following standards:

ISO 594-1:1986, Conical Fittings With A 6% (Luer) Taper For Syringes, Needles And Certain Other Medical Equipment - Part 1: General Requirements.

ISO 594-2 1998, Conical Fittings With A 6% (Luer) Taper For Syringes, Needles And Certain Other Medical Equipment - Part 2: Lock Fittings.

8. Clinical Test

It is not applicable.

9. Summary Comparing the Technological Characteristics

The proposed device and predicate device have the same classification information, intended use, principle of operation, design and Volume and pressure range. The proposed device and the predicate's material have a little difference, and the whole device has been tested for material biocompatibility, the test result demonstrated that the device comply with the requirements of a series ISO 10993 standards.

{5}------------------------------------------------

510(k) Summary

ItemProposed DevicePredicate DeviceK102648Remark
Product CodeMAVMAVSE
Regulation No.21 CFR.870.165021 CFR.870.1650SE
ClassClass IIClass IISE
Intended UseThe inflation device is intended to be used in PTCA orPTA procedures to create and monitor pressure in theballoon and to deflate the balloon dilatation catheter.The ANT inflation device is intended for use duringvascular procedures in conjunction withinterventional device such as balloon catheters tocreate and monitor pressure in the balloon catheter.SE

Table III-1 General information Comparison

Table III--2 Performance Comparison
ItemProposed DevicePredicate DeviceK102648Remark
PrincipleofOperationThe plunger is free to move when the trigger is pusheddown. In this position you may pull the plunger toaspirate or push it to inject for a quick fill. Release thetrigger to lock the plunger. To add and hold pressure inincrements, turn the handle clockwise with the triggerin the up position. Press the trigger down and pull backon the plunger to deflate.The plunger is free to move when the trigger ispushed down. In this position you may pull thehandle to aspirate or push it to inject for a quick fill.Release the trigger to lock the plunger. To add andhold pressure in increments, turn the handleclockwise with the trigger in the up position. Pressthe trigger down and pull back on the handle todeflate.SE
ComponentsA barrel, a plunger, a trigger, shell, a piston, a pressuregauge, a tubing, a rotating adapter and stopcocksA barrel, a plunger with handle, a trigger, shell, apiston, a pressure gauge, a tubing, a rotating adapterand stopcocksSimilarAnalysis 1

{6}------------------------------------------------

Premarket Notification 510(k) SubmissionSection Ⅲ-510(k) SummaryFile No.2016000
Conical fittingConforms to ISO 594-1:1986 or ISO 594-2:1998ANT Inflation Device compact Pack: a hemostasis,a torque device and a wire introducer.Conforms to ISO 594-1:1986 or ISO 594-2:1998SE
Design1. The 20mL barrel is made of polycarbonate for clarityand added strength against high system pressure.2. The threaded plunger assembly with a handle is madeof high impact resistant polyamides.3. Controlled inflation and deflation is achieved bymanually pushing handle and press the trigger.4. High-pressure polyurethane tubing with polyamidesprovides links from the inflation barrel to theangiographic catheter using a 6% conical lockingconnector.1. The 20mL barrel is made of polycarbonate forclarity and added strength against high systempressure.2. The pressurization system and the plunger aremade of high impact resistant polypropylene.3. Controlled inflation and deflation is achieved bymanually pushing or pulling the plunger and pressthe trigger.4. High-pressure polyurethane tubing provides linksfrom the inflation barrel to the angiographiccatheter using a 6% conical locking connector.SE
Where usedPTCAPTCASE
Volumeandpressure range20 mL, 0~30ATM20 mL, 0~30 ATMSE

Table III-3 Safety Comparison

ItemProposed DevicePredicate DeviceK102648Remark
MaterialPC, PA+PU, ABS, PC alloy, Steel,Silicon rubber, PA, rubberPA, PC, PU etc.SimilarAnalysis 2
BiocompatibilityConforms to the requirements of ISO10993 series StandardsConforms to the requirements of ISO 10993 seriesStandardsSE
In Vitro HemolyticNo hemolysisNo hemolysisSE
Acute Systemic ToxicityNo Acute Systemic ToxicityNo Acute Systemic ToxicitySE
In Vitro CytotoxicityNo cytotoxicityNo cytotoxicitySE
Skin IrritationNo intracutaneous reactivityNo intracutaneous reactivitySE
Skin SensitizationNo SensitizationNo SensitizationSE
SterilizationSAL10-610-6SE
MethodEO SterilizationEO SterilizationSE
ValidationConforms to ISO 11135Conforms to ISO 11135SE

{7}------------------------------------------------

Premarket Notification 510(k) Submission510(k) Summary
PackageIntegrityConforms to ISO 11607Conforms to ISO 11607SE
EOResidualConforms to ISO 10993-7Conforms to ISO 10993-7SE
PyrogenNon-pyrogenicNon-pyrogenicSE
Label and labelingConforms to FDA RequirementsConforms to FDA RequirementsSE

10.Conclusion

The proposed device, Inflation device, met all the predetermined acceptance criteria of design verfication and validation as specified by applicable standards and test protocols. The inflation device is substantially equivalent to the legally market predicate device.

§ 870.1650 Angiographic injector and syringe.

(a)
Identification. An angiographic injector and syringe is a device that consists of a syringe and a high-pressure injector which are used to inject contrast material into the heart, great vessels, and coronary arteries to study the heart and vessels by x-ray photography.(b)
Classification. Class II (special controls). The device, when it is a non-patient contacting balloon inflation syringe intended only to inflate/deflate balloon catheters and monitor pressure within the balloon, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.