K Number
K170025

Validate with FDA (Live)

Date Cleared
2017-10-03

(273 days)

Product Code
Regulation Number
870.1650
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Angiography syringe is used to inject contrast media into the heart, great vessels, and coronary arteries during angiographic procedures.

Device Description

The Angiography Syringe are plastic , single-use, disposable syringes to be offered in 6ml, 8ml, 10ml, 12ml size. As manual control syringe, each configuration includes bilateral, external finger rings located on the proximal barrel shaft. The proximal thumb rings located on the plunger, in tandem with the external finger rings on the barrel, allows for single handed movement to create aspiration and/or expulsion of fluids. The angiography syringe has two tip fitting: fixed adapter and rotating adapter

AI/ML Overview

The provided text is a 510(k) summary for a medical device (Angiography Syringe) and does not describe the specific type of study, acceptance criteria, or expert involvement typically associated with AI/ML-based medical devices or diagnostic tools.

The document discusses "acceptance criteria of design verification and validation as specified by applicable standards, guidance, and test protocols," but these are for the mechanical and material properties of the syringe itself, not for diagnostic performance in the way a clinical study for AI would be evaluated.

Therefore, many of the requested details cannot be extracted from this document, as they are not applicable to the type of device and study presented.

Here's a breakdown of what can and cannot be answered based on the provided text:

1. A table of acceptance criteria and the reported device performance

The document lists standards that the device complies with, which implicitly serve as acceptance criteria for those specific tests. However, it does not provide quantitative "reported device performance" against these criteria in a table format, other than stating compliance.

Acceptance Criteria (Standards Complied With)Reported Device Performance
ISO 594-1:1986 (Conical Fittings)Complies
ISO 594-2:1998 (Lock Fittings)Complies
ISO 7886-1:1993 (Sterile Hypodermic Syringes)Complies
Biocompatibility (ISO 10993-x; In Vitro Hemolytic, Cytotoxicity, Skin Irritation, Skin Sensitization, Acute Systemic Toxicity, Pyrogen)Complies

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

  • Sample Size: Not specified for non-clinical bench tests. The document states "Bench tests were conducted."
  • Data Provenance: Not applicable. These are laboratory bench tests, not clinical data. The manufacturer is Shanghai Kindly Medical Instruments Co., Ltd., China.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

  • Not applicable. This is a medical device (syringe), not an imaging or diagnostic AI/ML device requiring expert ground truth for its performance evaluation. The ground truth for its mechanical and biological safety is established by adherence to engineering and biocompatibility standards.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

  • Not applicable. There is no human interpretation or adjudication involved in the performance testing of a syringe against engineering or biocompatibility standards.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not applicable. This is not an AI/ML device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not applicable. This is not an AI/ML device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • Not applicable in the context of diagnostic performance. The "ground truth" for this device's performance is established by the specifications in the cited ISO standards for its physical and material properties (e.g., fit of luer connections, sterility, biocompatibility).

8. The sample size for the training set

  • Not applicable. This is not an AI/ML device.

9. How the ground truth for the training set was established

  • Not applicable. This is not an AI/ML device.

Summary of the Study Proving the Device Meets Acceptance Criteria (as described in the document):

The study to prove the device meets acceptance criteria was primarily non-clinical bench testing. The manufacturer conducted various tests to verify that the Angiography Syringe met all device specifications and was substantially equivalent (SE) to a predicate device.

  • Tests Performed:

    • Compliance with ISO 594-1:1986 (Conical Fittings)
    • Compliance with ISO 594-2:1998 (Lock Fittings)
    • Compliance with ISO 7886-1:1993 (Sterile Hypodermic Syringes)
    • Biocompatibility Testing (referencing ISO 10993 standards):
      • In Vitro Hemolytic Properties
      • In Vitro Cytotoxicity
      • Skin Irritation
      • Skin Sensitization
      • Acute Systemic Toxicity
      • Pyrogenicity
  • Conclusion: The test results demonstrated that the proposed devices comply with the listed standards, indicating that they met the predetermined acceptance criteria for design verification and validation. The document explicitly states, "Clinical Test: It is not applicable."

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name encircling a stylized emblem. The emblem depicts a white abstract image of an eagle with three heads. The logo is black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 3, 2017

Shanghai Kindly Medical Instruments Co.,ltd. Jeffery Hui Official Correspondent No. 925 Jinyuan Yi Rd Shanghai. 201803 CN

Re: K170025

Trade/Device Name: Angiography Syringe (Fixed Adapter), Angiography Syringe (Rotating Adapter) Regulation Number: 21 CFR 870.1650 Regulation Name: Angiographic Injector And Syringe Regulatory Class: Class II Product Code: DXT Dated: August 3, 2017 Received: September 7, 2017

Dear Jeffery Hui:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Kenneth J. Cavanaugh -S

for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Form Approved: OMB No. 0910-0120

Expiration Date: January 31, 2017

See PRA Statement below.

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K170025

Device Name

Angiography Syringe (Fixed Adapter), Angiography Syringe (Rotating Adapter)

Indications for Use (Describe)

The Angiography syringe is used to inject contrast media into the heart, great vessels, and coronary arteries during angiographic procedures.

Type of Use (Select one or both, as applicable)

|X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Section III-510(k) Summary

This 510(k) Summary is being submitted in accordance with the requirements of the guidance The 510(k) Program and 21 CFR 807.92.

510(k) Number: K170025

    1. Date of Submission: Aug.03, 2017

2. Submitter

Shanghai Kindly Medical Instruments Co., Ltd. No. 925, Jinyuan yi Road, Shanghai, 201803, China Establishment Registration Number: 3009605245 Contact Person: Xu Jianhai Position: RA Supervisor Tel.: +086-021-59140056 Fax: +086-021-59140056 Email: xujianhai@kdlchina.net

3. Proposed Device

Proposed Device Name: Angiography Syringe (Fixed Adapter), Angiography Syringe (Rotating Adapter) Proposed Device Model: 6mL, 8mL, 10mL, 12mL. Classification: Class II Classification Name: Angiographic injector and syringe Product Code: DXT Regulation Number: 21 CFR 870.1650 Review Panel: Cardiovascular

4. Predicate device

{4}------------------------------------------------

Product Name: Medline Angiographic Control Syringe

Manufacturer: Medline Industries Inc.

5. Device description

The Angiography Syringe are plastic , single-use, disposable syringes to be offered in 6ml, 8ml, 10ml, 12ml size. As manual control syringe, each configuration includes bilateral, external finger rings located on the proximal barrel shaft. The proximal thumb rings located on the plunger, in tandem with the external finger rings on the barrel, allows for single handed movement to create aspiration and/or expulsion of fluids. The angiography syringe has two tip fitting: fixed adapter and rotating adapter

6. Intended Use Statement:

The Angiography syringe is used to inject contrast media into the heart, great vessels, and coronary arteries during angiographic procedures.

7. Non-clinical Test Conclusion

Bench tests were conducted to verify that the proposed device met all device specification as were substantially equivalent (SE) to the predicate device. The test result demonstrates that the proposed devices comply with the following standards:

ISO 594-1:1986 Conical Fittings With A 6% (Luer) Taper For Syringes, Needles And Certain Other Medical Equipment - Part 1: General Requirements.

ISO 594-2:1998 Conical Fittings With A 6% (Luer) Taper For Syringes, Needles And Certain Other Medical Equipment - Part 2: Lock Fittings

ISO 7886-1:1993 Sterile hypodermic syringes for single use - Part 1: Syringes for manual.

Biocompatibility testing and reference standards:

  1. In Vitro Hemolytic Properties of Materials.

  2. In Vitro Cytotoxicity - Device - Part 5: Tests for In Vitro Cytotoxicity.

-- ISO 10993-10:2010, Biocompatibility Evaluation of Medical 3) Skin Irritation Device - Part 10: Tests for Irritation and Shin Sensitization.

{5}------------------------------------------------

--4) Skin Sensitization ISO 10993-10:2010, Biocompatibility Evaluation of Medical Device - Part 10: Tests for Irritation and Shin Sensitization.

  1. Acute Systemic Toxicity -- ISO 10993-11:2006, Biocompatibility Evaluation of Medical Device - Part 11: Tests for systemic toxicity.

ISO 10993-11:2006, Biocompatibility Evaluation of Medical -- 6) Pyrogen Device - Part 11: Tests for systemic toxicity.

8. Clinical Test

It is not applicable.

9. Summary Comparing the Technological Characteristics

The proposed device and predicate device have the same classification information, same material, principle of operation and syringe tip style. The proposed device has four kinds of volumes, while the predicate has various sizes. So the volume range of the proposed device is included in the volume range of the predicate device. Therefore, they are considered to be substantial equivalent in the volume aspect.

{6}------------------------------------------------

Premarket Notification 510(k) SubmissionSection III-SummaryFile No.2016000-
ItemProposed DevicePredicate Device(K093830)Remark
Product CodeDXTDXTSE
Regulation No.870.1650870.1650SE
ClassIIIISE
Intended UseThe Angiography syringe is used to inject contrastmedia into the heart, great vessels, and coronaryarteries during angiographic procedures.An angiography syringe is a device that consist of asyringe which is used to inject contrast material into theheart, great vessels, and coronary arteries duringangiographic or CT procedures.SE
Principle ofOperationIt is designed for single manual use. The plunger is freeto move. When pulled the plunger to aspirate the fluidor pushed it to injection.It is designed for single manual use. The plunger is freeto move. When pulled the plunger to aspirate the fluidor pushed it to injection.SE
ComponentsBarrel, piston, plunger, plunger cap, rotating adapter,O-ring, thumb ringBarrel, piston, plunger, plunger cap, Rotating connector,O-ring thumb ringSE
Conical fittingComply with ISO ISO 594-1:1986 ISO 594-2:1998Comply with ISO ISO 594-1:1986 ISO 594-2:1998SE
Where usedHospitalHospitalSE
Size6mL,8mL,10mL,12mLVarious sizeSimilar
Syringe tip styleFixed adapter and rotating adapterLL connector and rotating connectorSE

{7}------------------------------------------------

10.Conclusion

The proposed device, angiography syringe, met all the predetermined acceptance criteria of design verification and validation as specified by applicable standards, guidance, and test protocols. The angiography syringe is substantially equivalent to the legally market predicate device.

§ 870.1650 Angiographic injector and syringe.

(a)
Identification. An angiographic injector and syringe is a device that consists of a syringe and a high-pressure injector which are used to inject contrast material into the heart, great vessels, and coronary arteries to study the heart and vessels by x-ray photography.(b)
Classification. Class II (special controls). The device, when it is a non-patient contacting balloon inflation syringe intended only to inflate/deflate balloon catheters and monitor pressure within the balloon, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.