K Number
K170010

Validate with FDA (Live)

Manufacturer
Date Cleared
2017-08-28

(237 days)

Product Code
Regulation Number
876.4620
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

This device is intended to establish and maintain nephrostomy and internal drainage following pyeloplasty for pediatic patients. This device is intended for patients 6 months and older.

Device Description

The Salle Intraoperative Pyeloplasty Stent is a single-lumen ureteral stent inserted following pyeloplasty in order to provide internal and external drainage. It is a prolonged indwelling device not to exceed 4 weeks in the body.

The Salle Intraoperative Pyeloplasty Stent Set is comprised of a stent, wire guide, connecting tube, adaptor, and retention disc with pull tie. The Salle Intraoperative Pyeloplasty Stent is a double pigtail stent that advances from the renal pelvis to the bladder with a flexible extended portion that is brought out percutaneously through the kidney. The Salle stent is available in outer diameters of 4.0 or 4.7 French with working lengths of 12 to 18 cm. The stent's proximal pigtail forms in the renal pelvis, while the distal pigtail forms in the bladder. Drainage can occur internally (from the kidney to the bladder) or externally (from renal pelvis to the outside of the patient). The distal (bladder) pigtail coil is available in a multi-length stent configuration with drainage holes. The proximal (kidney) pigtail coil also contains sideports and ink marks to facilitate the visualization of accurate placement. The proximal end of the stent has an internal stylet which acts as a positioning stylet.

The set will be supplied sterile and is intended for one-time use. The set is packaged in a peel-open pouch with a three-year shelf life.

AI/ML Overview

Here's a summary of the acceptance criteria and the study proving the device meets them, based on the provided text:

Device: Salle Intraoperative Pyeloplasty Stent Set


1. Table of Acceptance Criteria and Reported Device Performance

TestAcceptance CriteriaReported Device Performance
Biocompatibility Testing (Performed per ISO 10993-1)Not explicitly detailed as numerical criteria in the table, but the successful completion of the listed tests is the criterion.All listed tests successfully completed.
- CytotoxicityPass
- SensitizationPass
- IntracutaneousPass
- Acute Systemic ToxicityPass
- Material-Mediated PyrogenicityPass
- Implantation - 2 WeekPass
- Implantation - 4 WeekPass
- Subacute/Subchronic ToxicityPass
- GenotoxicityPass
- Toxicological Risk Assessment of Extractable ChemicalsPass
Retention Strength and Break Strength Testing (Simulated use – Time Zero and post aging)Retention Strength: The minimum retention strength shall be greater than or equal to 0.03 N, but less than or equal to 3.9 N (90% coverage, 90% confidence). Break Strength: The minimum break strength shall be greater than 3.9 N (90% coverage, 90% confidence).All acceptance criteria were met. Pass
Retention Strength and Break Strength Testing (Following a 30 Day Artificial urine Soak – Time Zero and post aging)Retention Strength: The minimum retention strength shall be greater than or equal to 0.03 N, but less than or equal to 3.9 N (90% coverage, 90% confidence). Break Strength: The minimum break strength shall be greater than 3.9 N (90% coverage, 90% confidence).All acceptance criteria were met. Pass
Wire Guide Compatibility, Kink Radius, Flow Rate and Leakage Post Kinking TestingWire Guide Compatibility: The appropriate size wire guide shall pass through the inner lumen of each test article. Kink Radius: The kink radius of each test article shall be less than 30 mm based on worst-case analysis. Flow Rate and Leakage Post Kinking: Flow rate performed for characterization purposes only. If leakage detected, test article deemed a failure.All acceptance criteria were met. Pass
Radiopacity Testing (Per ASTM F640-12)The Salle Stent is visible under fluoroscopy. Visibility of the Salle Stent is equal to or greater than the visibility of the user-defined standard.All acceptance criteria were met. Pass
MR TestingMagnetically Induced Displacement Force: Deflection Angle <45° (Testing per ASTM F2052-15) Magnetically Induced Torque: $τ_{mag} < τ_{grav}$ (Testing per ASTM F2213-06 (2011)) Electrical Conductivity: <1 S/m MR Image Artifacts: For information only (Testing per ASTM F2119-07 (2013))All acceptance criteria were met. Pass

2. Sample size used for the test set and the data provenance
The document describes "bench testing" and "biocompatibility testing." While it mentions "each test article" for some bench tests, specific numerical sample sizes for these tests are not provided. The data provenance is not specified beyond being "Cook Inc.'s design control system."


3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This is a medical device (stent) and the testing described is primarily bench and biocompatibility testing. There is no mention of experts establishing ground truth as would be required for an AI/algorithm-based diagnostic device.


4. Adjudication method for the test set
Not applicable. This is not an study involving human interpretation of data requiring adjudication.


5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is a physical medical device, not an AI or imaging diagnostic tool.


6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
Not applicable. This is a physical medical device.


7. The type of ground truth used
For the bench tests, the ground truth is based on defined physical and performance standards (e.g., minimum retention strength in Newtons, maximal kink radius in mm, visibility under fluoroscopy, specific deflection angles/torques for MR compatibility). For biocompatibility, the ground truth is established by the successful completion of tests defined in ISO 10993-1.


8. The sample size for the training set
Not applicable. There is no training set mentioned, as this is not an algorithm-based device.


9. How the ground truth for the training set was established
Not applicable. There is no training set.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an image of three stylized human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 28, 2017

Cook Incorporated Minjin Choi Regulatory Affairs Specialist 750 Daniels Wav. P.O. Box 489 Bloomington, IN 47404

Re: K170010

Trade/Device Name: Salle Intraoperative Pyeloplasty Stent Set Regulation Number: 21 CFR§ 876.4620 Regulation Name: Ureteral Stent Regulatory Class: II Product Code: FAD Dated: July 14, 2017 Received: July 14, 2017

Dear Minjin Choi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Image /page/1/Picture/8 description: The image shows the name "Benjamin R. Fisher -S" in a large, bold font. The text is black and appears to be on a white background. The letters are clearly legible and evenly spaced.

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K170010

Device Name Salle Intraoperative Pyeloplasty Stent Set

Indications for Use (Describe)

This device is intended to establish and maintain nephrostomy and internal drainage following pyeloplasty for pediatic patients. This device is intended for patients 6 months and older.

Type of Use (Select one or both, as applicable)

√ Over-The-Counter Use (21 CFR 801 Subpart C) Prescription Use (Part 21 CFR 801 Subpart D)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda. hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Cook Medical. The logo is set against a red background. The word "COOK" is written in large, white, sans-serif letters. Below "COOK" is the word "MEDICAL" in smaller, white, sans-serif letters, set against a red background that is angled slightly.

COOK INCORPO 750 DANIELS BLOOMINGTON, IN 4740 9.2235 TOLL FREE: 800.45 WWW.COOKMEDICAL.COM

2.0 510(k) Summary

Salle Intraoperative Pyeloplasty Stent Set 21 CFR §807.92 Date Prepared: December 30, 2016

Submitted By:

Submission:Traditional 510(k) Premarket Notification
Applicant:Cook Incorporated
Contact:Minjin Choi
Applicant Address:Cook Incorporated750 Daniels WayBloomington, IN 47404
Contact Phone:(812) 339-2235 x104901
Contact Fax:(812) 332-0281

Device Information:

Trade Name:Salle Intraoperative Pyeloplasty Stent Set
Common Name:Stent, Ureteral
Classification Name:Ureteral Stent
Classification Regulation:21 CFR §876.4620, Product Code FAD
Device Class/Classification Panel:Class II, Gastroenterology/Urology

Predicate Device:

  • " Primary predicate device:
    Expel™ Nephroureteral Drainage Stent with Twist-Loc™ Hub System (K141344)

  • י Secondary predicate device:
    Pediatric Ureteral Stent (K060673)

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Cook Medical. The logo is set against a red background. The word "COOK" is written in white, block letters in the upper portion of the logo. Below the word "COOK" is the word "MEDICAL" in smaller, white block letters.

K1700
Page 2
COOK INCORPORATED
750 DANIELS WAY
BLOOMINGTON, IN 47404 USA
PHONE: 812.339.2235 TOLL FREE: 800.457.4500
WWW.COOKMEDICAL.COM

Device Description:

The Salle Intraoperative Pyeloplasty Stent is a single-lumen ureteral stent inserted following pyeloplasty in order to provide internal and external drainage. It is a prolonged indwelling device not to exceed 4 weeks in the body.

The Salle Intraoperative Pyeloplasty Stent Set is comprised of a stent, wire guide, connecting tube, adaptor, and retention disc with pull tie. The Salle Intraoperative Pyeloplasty Stent is a double pigtail stent that advances from the renal pelvis to the bladder with a flexible extended portion that is brought out percutaneously through the kidney. The Salle stent is available in outer diameters of 4.0 or 4.7 French with working lengths of 12 to 18 cm. The stent's proximal pigtail forms in the renal pelvis, while the distal pigtail forms in the bladder. Drainage can occur internally (from the kidney to the bladder) or externally (from renal pelvis to the outside of the patient). The distal (bladder) pigtail coil is available in a multi-length stent configuration with drainage holes. The proximal (kidney) pigtail coil also contains sideports and ink marks to facilitate the visualization of accurate placement. The proximal end of the stent has an internal stylet which acts as a positioning stylet.

The set will be supplied sterile and is intended for one-time use. The set is packaged in a peel-open pouch with a three-year shelf life.

Indications for Use:

The Salle Intraoperative Pyeloplasty Stent Set is intended to establish and maintain nephrostomy and internal drainage following pyeloplasty for pediatric patients. This device is intended for patients 6 months and older.

Comparison to Predicate Devices:

The Salle Intraoperative Pyeloplasty Stent Set and its primary predicate device, the Expel Nephroureteral Drainage Stent with Twist-Loc Hub System (K141344), are substantially equivalent in that these devices have similar intended uses, fundamental technological characteristics, and design. The modifications from the predicate device include:

  • I Indications for Use
  • I Stent material
  • I Placement method
  • I External end of stent

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the Cook Medical logo. The logo is set against a red background. The word "COOK" is written in white, block letters. Below "COOK" is a smaller rectangle with the word "MEDICAL" written in white letters.

K1700
Page 3

COOK INCORPORATED

750 DANIELS WAY

BLOOMINGTON, IN 47404 USA
PHONE: 812.339.2235 TOLL FREE: 800.457.4500

WWW.COOKMEDICAL.COM

The proposed Salle Intraoperative Pyeloplasty Stent Set is also similar in indications for use, dimensions, and methods of operation to the Pediatric Ureteral Stent (K060673). Differences between the proposed device and the predicate device include:

  • 트 Stent sizes
  • 트 Stent material
  • 트 Placement method
  • I External end of stent

Differences between the characteristics of the proposed device sets and the predicate devices are supported by testing.

Performance Data:

The biocompatibility evaluation for the Salle Intraoperative Pyeloplasty Stent Set was conducted in accordance with the FDA's Use of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process" June 16, 2016 and International Standard ISO 10993-1 "Biological evaluation of medical devices – Part 1 Evaluation and testing within a risk management process" (ISO 10993-1:2009).

The following biocompatibility testing was successfully completed;

  • Cytotoxicity
  • Sensitization
  • I Intracutaneous
  • 트 Acute Systemic Toxicity
  • Material-Mediated Pyrogenicity
  • Implantation - 2 Week
  • 트 Implantation - 4 Week
  • I Subacute/Subchronic Toxicity
  • 트 Genotoxicity
  • Toxicological Risk Assessment of Extractable Chemicals

A summary of the bench testing completed is detailed in Table 1.

Testing was successfully performed per Cook Inc.'s design control system in line with 21 CFR 820.30. The results from performance testing demonstrated that the subject device will remain sterile and will perform as intended for its 3 year shelf life.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Cook Medical. The word "COOK" is in large, white, sans-serif font on a red background. Below the word "COOK" is the word "MEDICAL" in a smaller, white, sans-serif font on a red background.

Table 1: Summary of the Bench Testing Performed

TestAcceptance CriteriaResult
Retention Strength andBreak Strength Testing(Simulated use – TimeZero and post aging)Retention StrengthThe minimum retention strength shall be greater thanor equal to 0.03 N, but less than or equal to 3.9 N(90% coverage, 90% confidence).Break StrengthThe minimum break strength shall be greater than3.9 N (90% coverage, 90% confidence).All acceptancecriteria were met.Pass
Retention Strength andBreak Strength TestingFollowing a 30 DayArtificial urine Soak(Simulated use – TimeZero and post aging)Retention StrengthThe minimum retention strength shall be greater thanor equal to 0.03 N, but less than or equal to 3.9 N(90% coverage, 90% confidence).Break StrengthThe minimum break strength shall be greater than3.9 N (90% coverage, 90% confidence).All acceptancecriteria were met.Pass
Wire Guide Compatibility,Kink Radius, and FlowRate and Leakage PostKinking TestingWire Guide CompatibilityThe appropriate size wire guide shall pass throughthe inner lumen of each test article.Kink RadiusThe kink radius of each test article shall be less than30 mm based on worst-case analysis.Flow Rate and Leakage Post KinkingFlow rate performed for characterization purposesonly. If leakage detected, test article deemed afailure.All acceptancecriteria were met.Pass
Radiopacity TestingTesting Per ASTM F640-121The Salle Stent is visible under fluoroscopy.Visibility of the Salle Stent is equal to or greater thanthe visibility of the user-defined standard.All acceptancecriteria were met.Pass
MR TestingMagnetically Induced Displacement ForceTesting per ASTM F2052-152Deflection Angle <45°Magnetically Induced TorqueTesting per ASTM F2213-063 (2011)$τmag< τgrav$Electrical Conductivity<1 S/mMR Image ArtifactsTesting per ASTM F2119-074 (2013)For information onlyAll acceptancecriteria were met.Pass

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the Cook Medical logo. The word "COOK" is in white, bold, sans-serif font on a red background. Below the word "COOK" is the word "MEDICAL" in white, sans-serif font on a dark red background.

COOK INCORPORATED 750 DANIELS WAY BLOOMINGTON, IN 47404 USA HE: 812.339.2235 TOLL FREE: 800.457.4500 WWW.COOKMEDICAL.COM

Conclusion:

The results of these tests support a conclusion that the Salle Intraoperative Pyeloplasty Stent Set will perform as intended. The proposed device set does not raise new questions of safety or effectiveness as compared to the predicate devices.

1 ASTM F640-12 Standard Test Methods for Determining Radiopacity for Medical Use

2 ASTM F2052-15 Standard test method for measurement of magnetical devices in the magnetic resonance environment

3 ASTM F2213-06 (2011) Standard test method for magnetically induced torque on passive implants in the magnetic resonance environment.

4 ASTM F2119-07 (2013) Standard test method for evaluation of MR image artifacts from passive implants

§ 876.4620 Ureteral stent.

(a)
Identification. A ureteral stent is a tube-like implanted device that is inserted into the ureter to provide ureteral rigidity and allow the passage of urine. The device may have finger-like protrusions or hooked ends to keep the tube in place. It is used in the treatment of ureteral injuries and ureteral obstruction.(b)
Classification. Class II (performance standards).