Osteo-Site Bone Biopsy Needle, Osteo-Site Bone Biopsy Needle Set

K170008 · Cook Incorporated · KNW · Feb 28, 2017 · Gastroenterology, Urology

Device Facts

Record IDK170008
Device NameOsteo-Site Bone Biopsy Needle, Osteo-Site Bone Biopsy Needle Set
ApplicantCook Incorporated
Product CodeKNW · Gastroenterology, Urology
Decision DateFeb 28, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1075
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Osteo-Site® Bone Biopsy Needle and Osteo-Site® Bone Biopsy Needle Set is intended for vertebral body access, biopsy and infusion during a vertebroplasty procedure.

Device Story

Osteo-Site Bone Biopsy Needle and Set are manual surgical instruments used for percutaneous access to the vertebral body. The device consists of a stainless steel cannula, inner stylet, and obturator, with an optional inner biopsy needle. The physician inserts the cannula into the vertebral body; the stylet is removed to allow for biopsy or infusion. The device is used in a clinical setting (e.g., OR or interventional suite) by a physician to facilitate vertebroplasty. The device provides a pathway for bone tissue sampling and subsequent cement infusion. Benefits include minimally invasive access to the vertebral body for diagnostic and therapeutic purposes.

Clinical Evidence

Bench testing only. Performance testing included compatibility and dimension analysis, tensile testing, torque strength, torsional override, compression strength, resistance to breakage, and biopsy performance testing. Biocompatibility testing included cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, hemolysis, partial thromboplastin time, and complement activation. All tests met predetermined acceptance criteria.

Technological Characteristics

Cannula, stylet, obturator, and inner biopsy needle constructed of stainless steel. Needle holder made of ABS/Polycarbonate. Available in 11, 13, 15 gauge (cannula) and 14, 16 gauge (biopsy needle). Tip configurations include tapered, bevel, Franseen, scoop, trocar, diamond, and trephine. Sterilization via EtO to a 10-6 SAL. Manual, non-powered device.

Indications for Use

Indicated for vertebral body access, biopsy, and infusion during vertebroplasty procedures in patients requiring bone tissue sampling or cement delivery.

Regulatory Classification

Identification

A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of three stylized human profiles facing to the right, with flowing lines beneath them. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 February 28, 2017 Cook Incorporated Ms. Erum Nasir, MS Regulatory Affairs Team Lead 750 Daniels Way Bloomington, Indiana 47404 Re: K170008 Trade/Device Name: Osteo-Site Bone Biopsy Needle. Osteo-Site Bone Biopsy Needle Set Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-urology biopsy instrument Regulatory Class: Class II Product Code: KNW Dated: December 30, 2016 Received: January 3, 2017 Dear Ms. Nasir: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and Part 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely Yours, # Jennifer R. Stevenson -S For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K170008 Device Name Osteo-Site® Bone Biopsy Needle and the Osteo-Site® Bone Biopsy Needle Set Indications for Use (Describe) The Osteo-Site® Bone Biopsy Needle and Osteo-Site® Bone Biopsy Needle Set is intended for vertebral body access, biopsy and infusion during a vertebroplasty procedure. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## 2.0 510(k) SUMMARY # Osteo-Site® Bone Biopsy Needle and the Osteo-Site® Bone Biopsy Needle Set 21 CFR §807.92 Date Prepared: December 30, 2016 #### Submitted By: | Submission: | Traditional 510(k) Premarket Notification | |----------------------------|------------------------------------------------------------------------------| | Applicant: | Cook Incorporated | | Contact: | Erum Nasir, MS | | Applicant Address: | Cook Incorporated<br>750 Daniels Way<br>Bloomington, IN 47404 | | Contact Phone Number: | (812) 335-3575 x102607 | | Contact Fax Number: | (812) 332-0281 | | Device Information: | | | Trade Name: | Osteo-Site® Bone Biopsy Needle and the Osteo<br>Site® Bone Biopsy Needle Set | | Common Name: | Instrument, Biopsy | | Classification Name: | Gastroenterology-Urology Biopsy Instrument | | Classification Regulation: | 21 CFR §876.1075, Product Code KNW | ## Device Class/Classification Panel: Class II, Gastroenterology/Urology ### Predicate Devices: - Stryker Bone and Vertebral Body Biopsy Kit (K032943) ● ### Device Description: The Osteo-Site® Bone Biopsy Needle and Osteo-Site® Bone Biopsy Needle Set consists of a cannula with an accompanying inner stylet and an obturator. The device may include an additional inner biopsy needle and obturator. The cannula is available in several sizes (11, 13 or 15 gauge) and lengths (10 or 15 cm) with various cannula/stylet tip configurations. The additional inner biopsy needle is available in two sizes (14 or 16 gauge) and lengths (17.1 or 22.1 cm) with a trephine tip and functions coaxially with the cannula once the stylet is removed. A plastic, non-radiopaque needle holder is also available to aid with needle placement. {4}------------------------------------------------ # Intended Use: The Osteo-Site® Bone Biopsy Needle and Osteo-Site® Bone Biopsy Needle Set is intended for vertebral body access, biopsy and infusion during a vertebroplasty procedure. # Comparison to Predicate Devices: The Osteo-Site® Bone Biopsy Needle and Osteo-Site® Bone Biopsy Needle Set is substantially equivalent to the predicate device, the Stryker Bone and Vertebral Body Biopsy Kit (K032943), in that these devices have similar indications for use, technological characteristics, design, and method of placement. The differences between the subject devices and the predicate devices are shown in the table below and include the materials, dimensions, cannula/stylet tip configurations, sterilization method, and included components. These minor differences were appropriately assessed and do not raise new questions regarding safety or effectiveness. | | Stryker Bone and Vertebral<br>Biopsy Kit (K032943) | Osteo-Site® Bone Biopsy Needle<br>and Set -<br>Proposed Device | |---------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------| | Regulation | 876.1075 | Identical | | Product Code | KNW | Identical | | Classification | II | Identical | | Intended Use | Stryker's bone and vertebral<br>body biopsy kits can be used<br>as a biopsy tool to remove a<br>sample of bone tissue from a<br>vertebral body for diagnostic<br>purposes using a coring,<br>cutting or aspiration technique. | This device is intended for vertebral<br>body access, biopsy and infusion<br>during a vertebroplasty procedure. | | Ancillary Device<br>Information | Stryker Introduction<br>Cannula/Stylet is sold<br>separately, but required to<br>access the vertebral body to<br>perform a biopsy with the<br>Stryker Bone and Vertebral<br>Biopsy Kit | N/A<br>The Cannula/Stylet combination is<br>included as a component in the Osteo-<br>Site® Bone Biopsy Needle Set and is<br>subject of this submission. | | Anatomical Site | Vertebral Body | Identical | | Method of Placement | Percutaneous | Identical | | Biopsy technique | Coring/Cutting/Suction | Not specified | | Biopsy Component | Inner Biopsy Needle | Cannula/Inner Biopsy Needle* | #### Substantial Equivalence Table * The proposed set includes the inner biopsy needle and obturator. {5}------------------------------------------------ | Substantial Equivalence Table (cont'd) | | | | |----------------------------------------|----------------------|----------------------------------------------------|----------------------------------------------------------------| | | | Stryker Bone and Vertebral<br>Biopsy Kit (K032943) | Osteo-Site® Bone Biopsy Needle<br>and Set -<br>Proposed Device | | Cannula** | Material | Stainless Steel | Identical | | | Gauge | 10, 11, 13 | 11, 13, 15 | | | Length (cm) | 12.7 | 10, 15 | | | Tip<br>Configuration | Four-Facet | Tapered, Bevel, Franseen | | Stylet** | Material | Stainless Steel | Identical | | | Tip<br>Configuration | Diamond, Bevel | Scoop, Trocar, Bevel, Diamond | | Inner<br>Biopsy<br>Needle* | Material | Stainless Steel | Identical | | | Gauge | 10, 11, 13 | 14, 16 | | | Length (cm) | 12.7, 22.9 | 17.1, 22.1 | | | Tip<br>Configuration | Four-Facet | Trephine | | Obturator* | Material | Stainless Steel | Identical | | | Number | One | One or Two | | Handle Material | | Unknown | ABS/Polycarbonate | | Accessories | | Syringe | Needle Holder | | Obturator Material | | Stainless Steel | Identical | | Sterilization Method | | Gamma Irradiation,<br>VDmax25 | EtO | | Sterilization Assurance Level | | 10-6 | 10-6 | | Packaging | | Peel pouch | Tyvek peel pouch | #### Substantial Equivalence Table (cont'd) * The proposed set includes the inner biopsy needle and obturator. ** Stryker Introduction Cannula/Stylet is listed under product code OAR and sold separately, but is required to access the vertebral body to perform a biopsy with the Stryker Bone and Vertebral Biopsy Kit cleared under K032943. Predicate IFU and brochure is provided in Appendix B (Page B-6 and B-9, respectively) for reference purpose only. # Performance Data: The following tests were performed to demonstrate that the proposed Osteo-Site® Bone Biopsy Needle and Osteo-Site® Bone Biopsy Needle Set met the applicable design and performance requirements and support a determination of substantial equivalence. Where applicable, testing was done per applicable ISO and other international standards. - . Compatibility and Dimension Analysis (Cannula, Stylet, Obturator, Inner biopsy needle, Inner biopsy needle obturator) - Testing demonstrated that the devices function as intended. {6}------------------------------------------------ - . Tensile Testing (Cannula, Stylet, Obturator, Inner biopsy needle, Inner biopsy needle obturator, and Needle holder) - Testing demonstrated that the devices withstand peak load values greater than the predetermined acceptance criterion. - Torque Strength Testing (Cannula, stylet, Inner biopsy needle) Testing . demonstrated that the devices withstand torque forces greater than the predetermined acceptance criterion. - Torsional Override Testing (Cannula) Testing demonstrated that the devices ● withstand torsional load greater than the predetermined acceptance criterion. - Compression Strength Testing (Cannula and stylet assembly, Inner biopsy needle) . – Testing demonstrated that the devices withstand compression forces greater than the predetermined acceptance criterion. - . Resistance to Breakage Testing (Cannula and stylet assembly)- Testing demonstrated that the devices function as intended. - Biopsy Testing (Cannula and inner biopsy needle) Testing demonstrated that the . devices function as intended. - . Biocompatibility Testing - Testing (cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, hemolysis, partial thromboplastin time, and complement activation) shows that the devices are biocompatible. # Conclusion: The results of these tests support a conclusion that the Osteo-Site® Bone Biopsy Needle and Osteo-Site® Bone Biopsy Needle Set met the design input requirements based on the intended use and support the conclusion that this device does not raise new questions of safety or effectiveness and is substantially equivalent to the predicate device, Stryker Bone and Vertebral Body Biopsy Kit (K032943).
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%