Intended Use

The VersaOne™ Fascial Closure System Bladed Trocar with fixation cannula is intended for use in a variety of minimally invasive gynecologic, general, and urologic procedures to create and maintain a port of entry for laparoscopic instruments.

The VersaOne™ Fascial Closure System Bladeless Trocar with fixation cannula is intended for use in a variety of minimally invasive gynecologic, general, and urologic procedures to create and maintain a port of entry for laparoscopic instruments.

The VersaOne™ Fascial Closure System Optical Trocar with fixation cannula is intended for use in a variety of minimally invasive gynecologic, general, and urologic procedures to create and maintain a port of entry for laparoscopic instruments.

Device Description

The VersaOne™ Fascial Closure System Bladed Trocar (12mm) is available in standard length (100mm) transparent cannula. The Bladed Trocar obturator has a dolphin nose (conical) shaped tip with a sharp linear blade and a spring-loaded locking shield. The blade is used to cut the tissues as the trocar is inserted into the body cavity. The shield advances to cover the blade upon entry into a free space. The cannula assembly is composed of a seal system, trocar housing, transparent cannula sleeve, and a stopcock. The trocar housing contains internal seals to prevent loss of pneumoperitoneum when instruments are inserted into a port or withdrawn completely from a port. The cannula seal system accommodates instruments ranging 5mm up to 12mm. The distal end of the trocar cannula has two windows, 180 degrees apart, covered by a pierce-able membrane. A guide is provided which fits inside the trocar cannula to guide a suture passer through the wall of the cannula and soft tissue to be approximated. The stopcock valve is for insufflation and desufflation. A reusable suture passer is provided, which delivers suture through the trocar cannula to aid in port site closure.

The VersaOne™ Fascial Closure System Bladeless Trocar (12mm) is available in standard length (100mm) transparent cannula. Bladeless Trocar obturator has a dolphin nose (conical) shaped bladeless tip. The conical shape is for smooth insertion into the body cavity by spreading the tissue layers during insertion. The cannula assembly is composed of a seal system, trocar housing, transparent cannula sleeve, and a stopcock. The trocar housing contains internal seals to prevent loss of pneumoperitoneum when instruments are inserted into a port or withdrawn completely from a port. The cannula seal system accommodates instruments ranging 5mm up to 12mm. The distal end of the trocar cannula has two windows, 180 degrees apart, covered by a pierce-able membrane. A guide is provided which fits inside the trocar cannula to guide a suture passer through the wall of the cannula and soft tissue to be approximated. The stopcock valve is for insufflation and desufflation. A reusable suture passer is provided, which delivers suture through the trocar cannula to aid in port site closure.

The VersaOne™ Fascial Closure System Optical Trocar (12mm) is available in standard length (100mm) transparent cannula. The Optical Trocar with transparent cannula and obturator allows for visualization of tissue layers during insertion. The obturator housing contains a scope retention mechanism. The cannula assembly is composed of a seal system, trocar housing, transparent cannula sleeve, and a stopcock. The trocar housing contains internal seals to prevent loss of pneumoperitoneum when instruments are inserted into a port or withdrawn completely from a port. The cannula seal system accommodates instruments ranging 5mm up to 12mm. The distal end of the trocar cannula has two windows. 180 degrees apart, covered by a pierce-able membrane. A guide is provided which fits inside the trocar cannula to guide a suture passer through the wall of the cannula and soft tissue to be approximated. The stopcock valve is for insufflation and desufflation. A reusable suture passer is provided, which delivers suture through the trocar cannula to aid in port site closure.

AI/ML Overview

The provided document is a 510(k) premarket notification for a medical device (VersaOne Fascial Closure System Trocar). It describes the device, its intended use, and comparisons to predicate devices to demonstrate substantial equivalence. Crucially, this document does NOT describe the performance of an AI/ML-driven medical device. It details bench testing, in-vivo testing (on animals), biocompatibility, sterilization, and cleaning studies for a physical surgical instrument.

Therefore, I cannot provide the information requested in your prompt (Acceptance Criteria and Study for an AI/ML device) because the provided text is about a surgical trocar, not an AI/ML system.

The prompt specifically asks for details relevant to AI/ML device validation, such as:

  • A table of acceptance criteria and reported device performance (for AI metrics like sensitivity, specificity, AUC).
  • Sample size and data provenance for the test set (relevant for imaging datasets for AI).
  • Number of experts and their qualifications for ground truth establishment.
  • Adjudication methods.
  • MRMC studies for human reader improvement with AI.
  • Standalone AI performance.
  • Type of ground truth (e.g., pathology, outcomes data).
  • Training set sample size and ground truth establishment for AI.

None of this information is present or applicable to the conventional medical device described in the provided text.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing to the right, stacked on top of each other. The profiles are rendered in a dark color, creating a sense of depth and unity. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the image.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 8, 2017

Covidien LLC Mr. Gary LeMere Sr. Regulatory Affairs Specialist 60 Middletown Avenue North Haven, Connecticut 06473

Re: K163654

Trade/Device Name: VersaOne Fascial Closure System Bladed Trocar. VersaOne Fascial Closure System Bladeless Trocar, VersaOne Fascial Closure System Optical Trocar Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: GCJ Dated: March 1, 2017 Received: March 3, 2017

Dear Mr. LeMere:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Jennifer R. Stevenson -S

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known)

K163654

Device Name

VersaOne™ Fascial Closure System Bladed Trocar, VersaOne™ Fascial Closure System Bladeless Trocar, VersaOne™ Fascial Closure System Optical Trocar

Indications for Use (Describe)

The VersaOne™ Fascial Closure System Bladed Trocar with fixation cannula is intended for use in a variety of minimally invasive gynecologic, general, and urologic procedures to create and maintain a port of entry for laparoscopic instruments.

The VersaOne™ Fascial Closure System Bladeless Trocar with fixation cannula is intended for use in a variety of minimally invasive gynecologic, general, and urologic procedures to create and maintain a port of entry for laparoscopic instruments.

The VersaOne™ Fascial Closure System Optical Trocar with fixation cannula is intended for use in a variety of minimally invasive gynecologic, general, and urologic procedures to create and maintain a port of entry for laparoscopic instruments.

Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Traditional 510(k) Summary

SUBMITTER:Covidien Ilc60 Middletown AvenueNorth Haven, CT 06473 USA
CONTACT PERSON:Gary LeMereSr. Regulatory Affairs SpecialistCovidien IlcPhone: (203) 492-5572Fax: (203) 492-5029e-mail: gary.f.lemere@medtronic.com
DATE PREPARED:December 20, 2016
PRODUCT CODE:GCJ
REGULATION NUMBER:21 CFR 876.1500
REVIEW PANEL:Gastroenterology-urology devices
TRADE/PROPRIETARY NAME:VersaOne™ Fascial Closure System Bladed TrocarVersaOne™ Fascial Closure System Bladeless TrocarVersaOne™ Fascial Closure System Optical Trocar
COMMON/USUAL NAME:Surgical Trocar
CLASSIFICATION NAME:Endoscope and Accessories
PREDICATE DEVICES:(K151548) - VersaOne™ Bladed Trocar 12mm,VersaOne™ Bladeless Trocar 12mm(K130435) - Special 510(k) for Versaport™ V2Bladeless Optical trocar
REFERENCE DEVICE:(K980123) - Carter-Thomason Needle-Point SuturePasser Instrument Set [Carter-ThomasonCloseSure System]

DEVICE DESCRIPTION:

The VersaOne™ Fascial Closure System Bladed Trocar (12mm) is available in standard length (100mm) transparent cannula. The Bladed Trocar obturator has a dolphin nose (conical) shaped tip with a sharp linear blade and a spring-loaded locking shield. The blade is used to cut the tissues as the trocar is inserted into the body cavity. The shield advances to cover the blade upon entry into a free space. The cannula assembly is composed of a seal system, trocar housing, transparent cannula sleeve, and a stopcock. The trocar housing contains internal seals to prevent loss of pneumoperitoneum when instruments are inserted into a port or withdrawn completely from a port. The cannula seal system accommodates instruments ranging 5mm up to 12mm. The distal end of the trocar cannula has two windows, 180 degrees apart, covered by a pierce-able

{4}------------------------------------------------

membrane. A quide is provided which fits inside the trocar cannula to quide a suture passer through the wall of the cannula and soft tissue to be approximated. The stopcock valve is for insufflation and desufflation. A reusable suture passer is provided, which delivers suture through the trocar cannula to aid in port site closure.

The VersaOne™ Fascial Closure System Bladeless Trocar (12mm) is available in standard length (100mm) transparent cannula. Bladeless Trocar obturator has a dolphin nose (conical) shaped bladeless tip. The conical shape is for smooth insertion into the body cavity by spreading the tissue layers during insertion. The cannula assembly is composed of a seal system, trocar housing, transparent cannula sleeve, and a stopcock. The trocar housing contains internal seals to prevent loss of pneumoperitoneum when instruments are inserted into a port or withdrawn completely from a port. The cannula seal system accommodates instruments ranging 5mm up to 12mm. The distal end of the trocar cannula has two windows, 180 degrees apart, covered by a pierce-able membrane. A quide is provided which fits inside the trocar cannula to quide a suture passer through the wall of the cannula and soft tissue to be approximated. The stopcock valve is for insufflation and desufflation. A reusable suture passer is provided, which delivers suture through the trocar cannula to aid in port site closure.

The VersaOne™ Fascial Closure System Optical Trocar (12mm) is available in standard length (100mm) transparent cannula. The Optical Trocar with transparent cannula and obturator allows for visualization of tissue layers during insertion. The obturator housing contains a scope retention mechanism. The cannula assembly is composed of a seal system, trocar housing, transparent cannula sleeve, and a stopcock. The trocar housing contains internal seals to prevent loss of pneumoperitoneum when instruments are inserted into a port or withdrawn completely from a port. The cannula seal system accommodates instruments ranging 5mm up to 12mm. The distal end of the trocar cannula has two windows. 180 degrees apart, covered by a pierce-able membrane. A quide is provided which fits inside the trocar cannula to quide a suture passer through the wall of the cannula and soft tissue to be approximated. The stopcock valve is for insufflation and desufflation. A reusable suture passer is provided, which delivers suture through the trocar cannula to aid in port site closure.

INTENDED USE:

The VersaOne™ Fascial Closure System Bladed Trocar with fixation cannula is intended for use in a variety of minimally invasive gynecologic, general, and urologic procedures to create and maintain a port of entry for laparoscopic instruments.

The VersaOne™ Fascial Closure System Bladeless trocar with fixation cannula is intended for use in a variety of minimally invasive gynecologic, general, and urologic procedures to create and maintain a port of entry for laparoscopic instruments.

The VersaOne™ Fascial Closure System Optical Trocar with fixation cannula is intended for use in a variety of minimally invasive gynecologic, general, and urologic procedures to create and maintain a port of entry for laparoscopic instruments.

SUMMARY OF TECHNOLOGICAL CHARACTERISTICS:

The VersaOne™ Fascial Closure System combines the existing technologies of the Covidien's VersaOne™ Bladed and Bladeless and Versaport Bladeless Optical Trocar systems with manual port site closure using a guide and suture passer. The VersaOne™

{5}------------------------------------------------

Fascial Closure System is available in three different models (reorder codes) each differing only by the obturator supplied; Bladed, Bladeless or Optical. The suture passer is a universal component to all three models. Below is a summary of characteristics for each of the three models.

ProposedDevicePredicateDeviceSimilaritiesDifferences
VersaOne™FascialClosureSystemBladedTrocarK151548VersaOne™BladedTrocar12mmTrocar Cannulamaintains a port ofentry for laparoscopicinstruments.Both use the exactsame bladed obturatorThe VersaOne™ Fascial ClosureSystem Bladed Trocar cannula has twowindows near the distal end to allow thesuture passer to exit. The predicateVersaOne™ Bladed Trocar does nothave any openings along the trocarcannula
Both use the sameself-adjusting sealhousingBoth are single use,EtO sterilizedThe VersaOne™ Fascial ClosureSystem Bladed Trocar cannula has aline and arrow printed on the side of thecannula for tissue alignment. Thepredicate VersaOne™ Bladed Trocar12mm has no printing on the trocarcannula
Both have three maincomponents: trocarcannula, seal housingand a bladed obturatorThe predicate VersaOne™ BladedTrocar 12mm has application ingynecologic, general, thoracic andurologic procedures. The VersaOne™Fascial Closure System Bladed Trocarhas the same applications with theexception of thoracic.
With the addition of two new features,the VersaOne™ Fascial Closure SystemBladed Trocar cannula also serves tofacilitate closure of the port site. A guidethat is inserted into the cannula providesa pathway for the suture passer todeliver suture into the body cavity. Thepredicate trocar does not have thisfeature. This feature is similar to theReference Device (K980123)

{6}------------------------------------------------

ProposedDevicePredicateDeviceSimilaritiesDifferences
VersaOne™FascialClosureSystemBladelessTrocarK151548VersaOne™BladelessTrocar12mmTrocar Cannulamaintains a port ofentry for laparoscopicinstruments.Both use the exactsame bladelessobturatorThe VersaOne™ Fascial ClosureSystem Bladeless Trocar cannula hastwo windows near the distal end to allowthe suture passer to exit. The predicateVersaOne™ Bladeless Trocar does nothave any openings along the trocarcannula
Both use the sameself-adjusting sealhousingBoth are single use,EtO sterilizedThe VersaOne™ Fascial ClosureSystem Bladeless Trocar cannula has aline and arrow printed on the side of thecannula for tissue alignment. Thepredicate VersaOne™ Bladed Trocar12mm has no printing on the trocarcannula
Both have three maincomponents: trocarcannula, seal housingand a bladelessobturatorThe predicate VersaOne™ BladelessTrocar 12mm has application ingynecologic, general, thoracic andurologic procedures. The VersaOne™Fascial Closure System BladelessTrocar has the same applications withthe exception of thoracic.
With the addition of two new features,the VersaOne™ Fascial Closure SystemBladeless Trocar cannula also serves tofacilitate closure of the port site. A guidethat is inserted into the cannula providesa pathway for the suture passer todeliver suture into the body cavity. Thepredicate trocar does not have thisfeature. This feature is similar to theReference Device (K980123).

{7}------------------------------------------------

ProposedDevicePredicateDeviceSimilaritiesDifferences
VersaOne™FascialClosureSystemOpticalTrocarK130435Versaport™V2 BladelessOpticaltrocarTrocar Cannulamaintains a port ofentry for laparoscopicinstruments.Both use the exactsame (bladeless)optical obturatorThe VersaOne™ Fascial ClosureSystem Optical Trocar cannula has twowindows near the distal end to allow thesuture passer to exit. The predicateVersaport V2 Bladeless Optical Trocardoes not have any openings along thetrocar cannula
Both use the sameself-adjusting sealhousingBoth are single use,EtO sterilizedThe VersaOne™ Fascial ClosureSystem Optical Trocar cannula has aline and arrow printed on the side of thecannula for tissue alignment. Thepredicate Versaport™ V2 BladelessOptical Trocar 12mm has no printing onthe trocar cannula
Both have three maincomponents: trocarcannula, seal housingand an opticalobturatorThe predicate VersaOne™ BladelessOptical Trocar 12mm has application ingynecologic, general, thoracic andurologic procedures. The VersaOne™Fascial Closure System Optical Trocarhas the same applications with theexception of thoracic.
With the addition of two new features,the VersaOne™ Fascial Closure SystemOptical Trocar cannula also serves tofacilitate closure of the port site. A guidethat is inserted into the cannula providesa pathway for the suture passer todeliver suture into the body cavity. Thepredicate trocar does not have thisfeature. This feature is similar to theReference Device (K980123).

COMPATIBILITY DATA:

Biocompatibility testing was completed in accordance International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation Dated June 6, 2016 and Testing within a Risk Management Process" as recognized by FDA and FDA Guidance for Industry: Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process dated March 17, 2015.

{8}------------------------------------------------

The following tests were conducted. which are appropriate to the patient contact profile of the proposed devices:

  • . Cytotoxicity
  • Sensitization .
  • . Intracutaneous irritation
  • Pvrogenicitv ●

STERILIZATION:

Sterilization studies were completed for single use devices sterilized by ethylene oxide in accordance with ISO 11135-1 Sterilization of health care products - Ethylene oxide - Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices and ISO 10993-7 Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals as recognized by FDA

CLEANING AND STERILIZATION:

Cleaning and sterilization studies were completed for the reusable suture passer in accordance with Guidance for Industry and Food and Drug Administration Staff: Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling dated March 17, 2015 and AAMI TIR30 A compendium of processes, materials, test methods, and acceptance criteria for cleaning reusable medical devices and AAMI TIR12 Designing, testing, and labeling reusable medical devices for reprocessing in health care facilities: A guide for medical device manufacturers as recognized by FDA

PERFORMANCE DATA:

Performance studies were conducted to demonstrate the proposed device(s), VersaOne™ Fascial Closure System trocars are substantially equivalent to the predicate devices. In-vitro (bench), in-vivo (live animal) and comparison testing to predicate support the intended use of this device and are as follows:

VersaOne™ Fascial Closure System In-Vitro (bench) Testing:

  • Visual Examination .
  • Suture Passer Puncture Force ●
  • Suture Passer handle open force ●
  • Guide insertion force ●
  • Guide torque to lock ●
  • . Suture Passer, suture Retention force
  • Trocar Cannula Leak Test
  • Cannula Leak test with quide ●
  • o Trocar Cannula Fixation Force

VersaOne™ Fascial Closure System In-Vivo (live animal) Testing:

  • Trocar penetration force ●
  • Trocar fixation force
  • Functional Test ●

RELIABILITY TESTING:

Accelerated life cycle testing was performed for the proposed VersaOne™ Fascial Closure System Reusable Suture Passer.

HUMAN FACTORS:

{9}------------------------------------------------

Human Factors evaluations were completed using

IEC 62366-1 Medical Devices - Part 1: Application of Usability Engineering to Medical Devices and FDA guidance Applying Human Factors and Usability Engineering to Medical Devices Dated February 3, 2016. While submission of Human Factors evaluation is not a requirement for premarket submission, a summative usability report is included as part of the objective evidence demonstrating safety and effectiveness.

CONCLUSION:

Supporting data enclosed in this submission demonstrates that the proposed VersaOne™ Fascial Closure System Bladed Trocar, VersaOne™ Fascial Closure System Bladeless Trocar and VersaOne™ Fascial Closure System Optical Trocar devices are as safe, as effective, and performs as well as, or better than their predicate devices and do not present any new issues of safety or effectiveness.

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.