K Number
K163653

Validate with FDA (Live)

Manufacturer
Date Cleared
2017-10-13

(294 days)

Product Code
Regulation Number
888.3020
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The NOVOS-NAIL Limb Lengthening System is indicated for limb lengthening of the femur and tibia.

Device Description

The Smith & Nephew NOVOS-NAIL Limb Lengthening System (NOVOS-NAIL LLS) is composed of a modular implantable intramedullary rod, locking screws, and an external actuator. The modular implantable rod is available in different configurations, lengths, and diameters to accommodate a variety of patient anatomies. The system uses locking screws and nail caps (optional) previously cleared with the TRIGEN Nail system (K981529).

The NOVOS-NAIL implant includes an enclosed rare earth magnet, telescoping lead screw/nut assembly and gearing. The subject device is supplied sterile by gamma sterilization. The external actuator is supplied non sterile.

AI/ML Overview

This is a set of acceptance criteria for a medical device called the NOVOS-NAIL Limb Lengthening System (LLS). The acceptance criteria and supporting studies are presented below:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device PerformanceStudy that Proves the Device Meets the Acceptance Criteria
Structural StrengthEvaluated for retrograde femoral and antegrade tibia loading configurationsFinite Element Analysis (FEA) of 8mm and 11mm NOVOS Limb Lengthening Nails
Construct Fatigue PerformanceEquivalent to the predicate deviceConstruct Fatigue Testing on 8mm and 11mm NOVOS-NAIL and predicate device
Bending Fatigue PerformanceSimilar to the predicate device, identified potential failure modesBending Fatigue Evaluation of 8mm and 11mm NOVOS-NAIL
Distraction AccuracyMet the acceptance criteriaDistraction Accuracy Testing for 8mm and 11mm nails
UsabilityLow residual use-related risk, outweighed by benefitsUsability Testing (Human Factors Validation)
Bacterial Endotoxin LimitsMet acceptable endotoxin limitsBacterial Endotoxin Testing per FDA Guidance and ANSI/AAMI ST72
Static Bend PerformanceDemonstrated adequate static bending performance during expected clinical useStatic Bend Strength Testing of 8mm and 11mm NOVOS-NAIL in a static three-point bend test

2. Sample size used for the test set and the data provenance:

  • Finite Element Analysis (FEA): The sample size refers to the digital models created. Specific numbers are not provided, but it states "8mm and 11mm NOVOS Limb Lengthening Nails" were analyzed.
  • Construct Fatigue Testing: "8mm and 11mm NOVOS-NAIL and the predicate device" were tested.
  • Bending Fatigue Evaluation: "8mm and 11mm NOVOS- NAIL" were tested.
  • Distraction Accuracy: "8mm and 11mm nail" were tested.
  • Usability Testing: The document mentions "A human factors validation was conducted," but does not specify the number of participants.
  • Bacterial Endotoxin Testing: Not explicitly stated, but typically involves a batch or representative samples of the sterilized device.
  • Static Bend Strength: "8mm and 11mm NOVOS-NAIL" were tested.

The provenance of the data is not explicitly stated in terms of country of origin or whether it was retrospective or prospective. It refers to pre-clinical testing, which is generally conducted in a laboratory setting.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This information is not provided. The studies outlined are primarily mechanical and usability tests, which typically rely on objective measurements and established standards rather than expert consensus for ground truth establishment. For the usability study, "human factors validation" implies involvement of human factors specialists, but their specific qualifications or number are not detailed.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

No adjudication method is mentioned for any of the described tests. The nature of these pre-clinical tests (mechanical and engineering analyses) usually involves direct measurement against pre-defined specifications rather than subjective assessment requiring adjudication.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

No MRMC comparative effectiveness study was done. This device is a mechanical implant for limb lengthening and does not involve AI or human "readers" in its primary function, so such a study would not be applicable.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

This question is not applicable as the device is a mechanical implant and does not involve an algorithm. The "external actuator" mentioned is a mechanical control, not an AI algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

The ground truth for these pre-clinical tests is established by:

  • Engineering Specifications and Standards: For structural strength, fatigue, bending, distraction accuracy, and static bend strength, the ground truth is derived from established mechanical engineering principles, material science, and relevant industry standards (e.g., for fatigue testing, bending strength).
  • Regulatory Guidance: For bacterial endotoxin testing, the ground truth is "acceptable endotoxin limits as stated in the FDA Guidance, 'Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile,' and 'Pyrogen and Endotoxin Testing: Questions and Answers,' and ANSI/AAMI ST72."
  • Human Factors Principles: For usability testing, the ground truth is based on principles of human factors engineering and user experience to identify and mitigate use-related risks.

8. The sample size for the training set:

Not applicable. This document describes pre-clinical testing for a physical medical device, not a machine learning model, so there is no concept of a "training set" in this context.

9. How the ground truth for the training set was established:

Not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

October 13, 2017

Image /page/0/Picture/1 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

Smith & Nephew Allison Chan Regulatory Affairs Specialist II 1450 E Brooks Rd. Memphis, Tennessee 38116

Re: K163653

Trade/Device Name: NOVOS-NAIL Limb Lengthening System (LLS) Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary fixation rod Regulatory Class: Class II Product Code: HSB Dated: September 22, 2017 Received: September 25, 2017

Dear Allison Chan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevicesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICESForm Approved: OMB No. 0910-0120
Food and Drug AdministrationExpiration Date: January 31, 2017
Indications for UseSee PRA Statement below.
510(k) Number (if known)K163653
Device NameNOVOS-NAIL Limb Lengthening System
Indications for Use (Describe)The NOVOS-NAIL Limb Lengthening System is indicated for limb lengthening of the femur and tibia.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."

:

Page 1 of 1

·

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image contains the Smith & Nephew logo. The logo consists of an orange flower-like symbol on the left, followed by the text "We are" in gray. To the right of that, the text "smith&nephew" is written in orange.

Submitted by:Smith & Nephew, Inc.Advanced Surgical Division1450 East Brooks RoadMemphis, Tennessee 38116
Date of Summary:October 10,2017
Contact Person and Address:Allison ChanRegulatory Affairs Specialist IIT 901-399-1098F 901-566-7022
Name of Device:Smith & Nephew, Inc. NOVOS-NAIL Limb LengtheningSystem
Common Name:Intramedullary Nail
Device Classification Name andReference:21 CFR 888.3020, Intramedullary fixation rod
Device Class:Class II
Panel Code:Orthopaedics/87
Product Code:HSB (Rod, Fixation, Intramedullary and Accessories)

Device Description

The Smith & Nephew NOVOS-NAIL Limb Lengthening System (NOVOS-NAIL LLS) is composed of a modular implantable intramedullary rod, locking screws, and an external actuator. The modular implantable rod is available in different configurations, lengths, and diameters to accommodate a variety of patient anatomies. The system uses locking screws and nail caps (optional) previously cleared with the TRIGEN Nail system (K981529).

The NOVOS-NAIL implant includes an enclosed rare earth magnet, telescoping lead screw/nut assembly and gearing. The subject device is supplied sterile by gamma sterilization. The external actuator is supplied non sterile.

Intended Use

The NOVOS-NAIL Limb Lengthening System is indicated for limb lengthening of the femur and tibia.

{4}------------------------------------------------

Comparison of Technology Characteristics with the Predicate Device

The NOVOS-NAIL Limb Lengthening System is substantially equivalent in design and fundamental scientific technology to the defined predicate and does not raise any new issues of safety and efficacy. At a high level, the subject and predicate devices are based on the following same technological elements

  • Limb lengthening for tibia and femurs
  • Implanted in the intramedullary canal .

The following technological differences exist between the subject and predicate device.

  • . Ability to shorten nail if the rod has been distracted too far
  • Use of external device to control interaction of rare earth magnet ●
  • o Manufactured from 20Cr-15Ni-40Co-7Mo-16Fe

Summary of Pre-Clinical Testing

The following performance data was provided in support of the substantial equivalence determination:

  • Finite Element Analysis (FEA) FEA of the 8mm and 11mm NOVOS Limb . Lengthening Nails using Retrograde Femoral and Antegrade Tibial Loading Modes- Finite Element analysis evaluated the structural strength of the subject devices in retrograde femur and antegrade tiba loading configurations.
  • Construct Fatigue Testing- Construct fatigue testing was conducted on the ● 8mm and 11mm NOVOS-NAIL and the predicate device. Results of the testing have determined that the subject devices would have equivalent construct fatigue performance compared to the predicate.
  • . Bending Fatigue Evaluation- Testing was conducted to evaluate the bending fatigue performance of the 8mm and 11mm NOVOS- NAIL and to determine potential failure modes for the nail. Results of the testing have determined that the subject devices would have similar bending fatigue performance as compared to the predicate.
  • . Distraction Accuracy-Testing evaluating the accuracy of the actuator and the distraction mechanism in the NOVOS-NAIL was conducted for both the 8mm and 11mm nail. Results of the testing have shown the subject devices have met the acceptance criteria for distraction accuracy.
  • Usability Testing- A human factors validation was conducted to evaluate the usability of the NOVOS-NAIL LLS. Results of the usability study have shown that any residual use-related risk that remain after human factors testing are low and are outweighed by the benefits derived from the device.
  • Bacterial Endotoxin Testing: Testing was completed and found to have the . met the acceptable endotoxin limits as stated in the FDA Guidance, " Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile," and "Pyrogen and Endotoxin Testing: Questions and Answers," and ANSI/AAMI ST72.
  • Static Bend Strength: Testing was conducted to evaluate the static bend . performance of the 8mm and 11mm NOVOS-NAIL in a static three point bend test. Results of the testing have shown the subject devices demonstrate adequate static bending performance during the expected clinical use.

{5}------------------------------------------------

Substantial Equivalence Information

The substantial equivalence of the NOVOS-NAIL Limb Lengthening Nail System is based on similarities in intended use, indications for use, overall design, and performance to the predicate systems listed in the following table.

Table 5.1: Substantially Equivalent Predicates to the NOVOS Limb Lengthening Nail System

DesignAspectReviewedNOVOS-NAIL LimbLengthening Nail SystemIntramedullarySkeletalKineticDistractor(ISKD)-PrimaryPredicatePreciceIntramedullaryLimbLengtheningSystem-ReferencePredicate
510(k)NumberSubject 510(k)K010322K101997
ManufacturerSmith & NephewOrthoDyne,Inc.EllipseTechnologies, Inc.
SimilarIndicationsfor UseYesYesYes
SterilizationMethodsGammaEthylene OxideGamma
NailDiameter8mm, 11mm10.7mm8.5mm, 10.7mm,and 12.5mm
CrossSectionShapeCircularCircularCircular
Rate ofLengtheningPhysician to determine rate.Dependent upon patientactivity level;1mm/dayPhysician todetermine rate.Dependentupon patientactivitylevel; .75-1.25mm/dayPhysician todetermine rate,dependent uponpatient activitylevel ~1mm /day.
Control ofLengtheningExternal ActuatorPatient'sactivity level,l.e. rotationaloscillations ofthe limbExternal remotecontroller
MonitoringofDistractionX-ray for confirmationExternal handheld monitorX-ray forconfirmation

{6}------------------------------------------------

DesignAspectReviewedNOVOS-NAIL LimbLengthening Nail SystemIntramedullarySkeletalKineticDistractor(ISKD)-PrimaryPredicatePreciceIntramedullaryLimbLengtheningSystem-ReferencePredicate
SafetyFeaturesLength of rod is able to beshortened by rotating theexternal actuatorAutomatic stopwhenpredeterminelength isachieved; one-way clutchdesignLength of rod isable to beshortened fromthe externalremote controller
Material20Cr-15Ni-40Co-7Mo-16FeTi-6Al-4VTi-6Al-4V
Method ofFixation4.5mm & 5.0mm Internal HexCaptured Screws (K981529)4.0mm &4.8mmdiameterlocking screwsLocking Pegs

Conclusion

As previously noted, this Traditional 510(k) Premarket Notification is being submitted to request clearance for the NOVOS-NAIL Limb Lengthening Nail System. Based on the similarities to the predicate components and a review of the mechanical testing performed, the devices are substantially equivalent to above predicate systems.

§ 888.3020 Intramedullary fixation rod.

(a)
Identification. An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.(b)
Classification. Class II.