K Number
K163610

Validate with FDA (Live)

Date Cleared
2017-05-19

(148 days)

Product Code
Regulation Number
888.4580
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Piezotome® M+ is an ultrasonic surgical system consisting of handpieces and associated tips, for cutting bone, bone substitutes and metal. The system can be used for osteotomy, osteoplasty, decorticating, drilling, shaping, and smoothing of bones and teeth, in a variety of surgical procedures, including general orthopaedic, otolaryngological, maxillofacial, oral, hand, foot, neurosurgical spine, and plastic/reconstructive surgery.

Piezotome® M+ is to supply utilities to and serve as a base for dental tools such as ultrasonic scaler, bone cutting instrument and accessories for use by qualified dental practitioners in periodontics, scaling, prosthesis and oral surgery.

Device Description

PIEZOTOME M+ is an Electromedical Device that uses micro-vibrations of associated Tips, to perform the clinical procedures defined in Intended Use.

PIEZOTOME M+ consists of a Console, a Multifunction footswitch, two reusable Handpieces, a range of Dental Tips, a range of Bone Surgery Tips, two Irrigation tubings and Flat Wrenches for Tips.

The Console is the heart of the PIEZOTOME M+. The Console contains the display board and the motherboard. Ultrasonic Handpieces and Footswitch are connected to the LCD Color Touch screen present on the front panel is used to define the settings of the PIEZOTOME M+ (modes, ultrasonic power, irrigation flow values).

The Multifunction footswitch has different functions. The access to a number of functions on the control footswitch allows the practitioner to work in a perfectly sterile environment, avoiding the risk of crosscontamination. The Footswitch is classified IP X8 (Protective Index) for operating theater applications.

The Ultrasonic Handpieces are held in the Surgeron's hand. The Ultrasonic Handpieces are connected to the Console via a Handpiece Cord. A Handpiece is dedicated to Conventional dental procedures (NEWTRON LED Handpiece) and a Handpiece is dedicated to Bone Surgery procedures (PIEZOTOME M+ Handpiece). The Ultrasonic Handpieces are equipped with a Piezoelectric Transducer. The Piezoelectric transducer converts the Electrical Signal delivered by the Console into mechanical microvibrations. The Ultrasonic Handpieces are reusable and Sterilizable by autoclaving.

The Pump housings are designed to accommodate Irrigation Tubing cassettes. Irrigation Tubings are Single Use and delivered under Sterile State. Irrigation Solution is intended to cool the clinical site and eliminate by rinsing the fragments such as bone or teeth.

Tips are fixed by means of a Wrench at the extremity of the Ultrasonic Handpieces. Tips are in direct contact to the patient. The ultrasonic mechanical vibrations transmitted to the Tip permits the realization of the clinical procedures defined in Intended Use. Bone Surgery Tips are Single Use and delivered under Sterile State. Dental Tips are reusable.

AI/ML Overview

The provided document is a 510(k) Premarket Notification for a medical device called PIEZOTOME M+, issued by the FDA. It declares the device to be "substantially equivalent" to legally marketed predicate devices.

This document describes a medical device (PIEZOTOME M+, an ultrasonic surgical system) and its performance data to demonstrate substantial equivalence to existing devices. Therefore, the questions related to acceptance criteria and study design for an AI/algorithm-based device (such as sample size for test/training sets, data provenance, expert ground truth, adjudication methods, MRMC studies, standalone performance, and effect sizes of AI assistance) are not applicable to this document.

The "performance data" discussed in the document refers to typical engineering and safety tests for a physical medical device, not an AI/algorithmic system. The information provided is about electrical safety, electromagnetic compatibility, software verification (specifically stating the software is identical to a predicate device, not a new algorithm), and performance testing bench (meaning physical tests to confirm the device functions as intended, not AI model performance).

Here's how the provided information relates to the request, focusing on what can be extracted and what is not applicable:


1. A table of acceptance criteria and the reported device performance

The document does not present "acceptance criteria" in the typical format for an AI/algorithm where specific metrics (e.g., accuracy, sensitivity, specificity) and thresholds are defined. Instead, it demonstrates equivalency through a comparison of technological characteristics and verification against established electrical and safety standards.

Acceptance Criteria (Implied by Substantial Equivalence and Standards Compliance)Reported Device Performance
Electrical Safety: Compliance with IEC 60601-1:2005 (Medical electrical equipment - Part 1: General requirements for basic safety and essential performance), including USA and Canadian Deviations."The Electrical Safety Tests have been performed according to IEC 60601-1:2005... included USA and Canadian Deviations."
Electromagnetic Compatibility (EMC): Compliance with IEC 60601-1-2:2007 (Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests)."The Electromagnetic Compatibility Tests have been performed according to IEC 60601-1-2:2007..."
Software Verification & Validation: Adherence to "Guidance for Industry and FDA Staff - Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (May 11, 2005)."Software Verification activities were conducted and documented according to the document named 'Guidance for Industry and FDA Staff - Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices' dated May 11, 2005. Moreover, the Software System included in PIEZOTOME M+ is strictly the same Software as the Predicate Device PIEZOTOME 2 (K091331, cleared December 11, 2009)."
Performance (Physical Device Function): Similar technical characteristics for bone surgery and dental applications (piezoelectric and irrigation performance) to predicate devices."The Technical Characteristics for Bone Surgery Application of PIEZOTOME M+ (Piezoelectric Performances and Irrigation Performances) are similar as the Predicate Devices Piezoelectric System (K100410, cleared April 28, 2010) and PIEZOTOME 2 (K091331, cleared December 11, 2009). The Technical Characteristics for Dental Application of PIEZOTOME M+ (Piezoelectric Performances and Irrigation Performances) are similar as the Predicate Device PIEZOTOME 2 (K091331, cleared December 11, 2009)."
Sterilization Validation: Compliance with applicable standards for reusable and single-use accessories."The Sterilizability Tests have been performed according to applicable Standards for Re-Usable accessories and Single Use Accessories provided under Sterile State."
Biocompatibility Validation: Compliance with ISO 10993-1 and 510(k) Memorandum - #G95-1 Table 1 requirements for body contact."The Biocompatibility Tests have been performed according to applicable Standards for Accessories in contact to the Patient. Tests were conducted according to the ISO 10993-1 Standard and 510(k) Memorandum - #G95-1 Table 1 'Initial Evaluation Tests for Consideration requirements (Body Contact: External communicating Device - Type of Tissue / Bone Dentin Communicating - Contact Duration: A limited less than 24 hours)."
Similarity to Predicate Devices: Overall technological characteristics being similar or identical to predicate devices.A comprehensive comparison table (Table 02 on pages 5-9) is provided, detailing numerous identical or substantially equivalent characteristics across various categories (intended use, product codes, dimensions, power supply, safety classifications, internal/external constitution, software, piezoelectric performance, irrigation performance, user interface, footswitch, environmental conditions, handpieces, tips, sterilization, biocompatibility, safety/EMC standards). Minor differences (e.g., footswitch actuators, storage temperature range, specific tip designs, handpiece color) are explicitly stated to have "no impact" on substantial equivalence.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Not Applicable. This document pertains to a physical medical device. The "performance data" refers to engineering and safety tests of the device itself and its components (e.g., electrical safety, EMC, sterilization, biocompatibility), not an AI/algorithm dataset. There is no mention of a "test set" in the context of data used for an algorithm.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Not Applicable. This document does not describe an AI/algorithm that requires expert-established ground truth from medical images or clinical data.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not Applicable. See point 3.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not Applicable. This is not an AI-assisted diagnostic or clinical decision support device. It's an ultrasonic surgical instrument.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not Applicable. There is no standalone algorithm. The "software" mentioned is identical to a predicate device's software, suggesting a control system for the physical device, not an analytical algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • Not Applicable. The "ground truth" for this device's performance is compliance with established engineering standards (e.g., IEC, ISO) and benchmarks against predicate devices' physical characteristics and function.

8. The sample size for the training set

  • Not Applicable. There is no "training set" as this is not an AI/machine learning device.

9. How the ground truth for the training set was established

  • Not Applicable. See point 8.

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an image of three faces in profile, stacked on top of each other, with a stylized representation of a human form below them.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - W O66-G609 Silver Spring, MD 20993-0002

May 19, 2017

SATELEC % Ms. Argie Zoubroulis Quality Manager ACTEON, Inc. 124 Gaither Drive, Suite #140 Mt. Laurel, New Jersey 08054

Re: K163610

Trade/Device Name: PIEZOTOME M+, PIEZOTOME M+ Handpiece Regulation Number: 21 CFR 888.4580 Regulation Name: Sonic Surgical Instrument and Accessories/Attachments Regulatory Class: Class II Product Code: JDX, HWE, DZI, ERL, HBE Dated: April 19, 2017 Received: April 21, 2017

Dear Ms. Zoubroulis :

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Actinclude requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR

{1}------------------------------------------------

Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerelv.

Jennifer R. Stevenson -S

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known) K163610

Device Name PIEZOTOME M+

Indications for Use (Describe)

Piezotome® M+ is an ultrasonic surgical system consisting of handpieces and associated tips, for cutting bone, bone substitutes and metal. The system can be used for osteotomy, osteoplasty, decorticating, drilling, shaping, and smoothing of bones and teeth, in a variety of surgical procedures, including general orthopaedic, otolaryngological, maxillofacial, oral, hand, foot, neurosurgical spine, and plastic/reconstructive surgery.

Piezotome® M+ is to supply utilities to and serve as a base for dental tools such as ultrasonic scaler, bone cutting instrument and accessories for use by qualified dental practitioners in periodontics, scaling, prosthesis and oral surgery.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Additional Information - February 17, 2017

Summary of Safety And Effectiveness [As Required by 21 CFR 807.92.c]

I - SUBMITTER

01.a. 510(k) Submitter:

SATELEC 17, Avenue Gustave Eiffel BP 30216 33708 Merignac Cedex France 011-33-556-340-607 Telephone: 011-33-556-349-292 email: philippe.girard@acteongroup.com

01.b. Contact Person:

Argie ZOUBROULIS SATELEC c/o ACTEON, Inc. 124 Gaither Drive, Suite 140 Mt. Laurel, NJ 08054 Telephone: 800-289-6367 Ext. 390 856-222-4726 Fax: Email: argie.zoubroulis@acteongroup.com

01.c. Establishment Registration Number:8044015
01.d. Date Prepared:November 29, 2016
01.e. Type of 510(k) submission:Traditional 510(k) Submission.

II - DEVICE

02.a. Trade Name of Device:PIEZOTOME M+
02.b. Common Name of Device:Sonic surgical instrument and accessories / attachments
02.c. Classification Regulation:21 CFR 888.4580
02.d. Regulation Identification:A sonic surgical instrument is a hand-held device with various accessories or attachments, such as a cutting tip that vibrates at high frequencies, and is intended for medical purposes to cut bone or other materials, such as acrylic
02.e. Medical Device Class:II
02.f. Panel:Orthopedic

Pre-Market Notification 510(k) Submission for PIEZOTOME M+ By SATELEC K163610 - Traditional 510(k) - Confidential Document - April 10, 2017 - Version 02

Image /page/3/Picture/10 description: The image shows the logo for ACTEON. The logo consists of the word "ACTEON" in a dark gray, sans-serif font. A light blue curved line is located to the left of the word, starting below the "A" and extending upwards and to the left. The logo is simple and modern.

006 - 1

{4}------------------------------------------------

Additional Information - February 17, 2017

JDX 02.g. Product Code:

HWE, DZI, ERL, HBE 02-h. Subsequent Product Codes:

III - PREDICATE DEVICES

The Substantial Equivalence (SE) of PIEZOTOME M+ is based on the Predicate Devices identified in the Table 01.

Trade NameManufacturerProduct Code510(k) numberDate Cleared
Piezoelectric SystemSATELECJDX, DZI, ERL,HBE, HWEK100410April 28, 2010
PIEZOTOME 2SATELECDZIK091331Dec 11, 2009

|--|

SATELEC Piezoelectric System Predicate Device has not been subject to a design-related recall. SATELEC PIEZOTOME 2 Predicate Device has not been subject to a design-related recall.

IV - DEVICE DESCRIPTION

PIEZOTOME M+ is an Electromedical Device that uses micro-vibrations of associated Tips, to perform the clinical procedures defined in Intended Use.

PIEZOTOME M+ consists of a Console, a Multifunction footswitch, two reusable Handpieces, a range of Dental Tips, a range of Bone Surgery Tips, two Irrigation tubings and Flat Wrenches for Tips,

The Console is the heart of the PIEZOTOME M+. The Console contains the display board and the motherboard. Ultrasonic Handpieces and Footswitch are connected to the LCD Color Touch screen present on the front panel is used to define the settings of the PIEZOTOME M+ (modes, ultrasonic power, irrigation flow values)

The Multifunction footswitch has different functions. The access to a number of functions on the control footswitch allows the practitioner to work in a perfectly sterile environment, avoiding the risk of crosscontamination. The Footswitch is classified IP X8 (Protective Index) for operating theater applications.

The Ultrasonic Handpieces are held in the Surgeron's hand. The Ultrasonic Handpieces are connected to the Console via a Handpiece Cord. A Handpiece is dedicated to Conventional dental procedures (NEWTRON LED Handpiece) and a Handpiece is dedicated to Bone Surgery procedures (PIEZOTOME M+ Handpiece). The Ultrasonic Handpieces are equipped with a Piezoelectric Transducer. The Piezoelectric transducer converts the Electrical Signal delivered by the Console into mechanical microvibrations. The Ultrasonic Handpieces are reusable and Sterilizable by autoclaving.

The Pump housings are designed to accommodate Irrigation Tubing cassettes. Irrigation Tubings are Single Use and delivered under Sterile State. Irrigation Solution is intended to cool the clinical site and eliminate by rinsing the fragments such as bone or teeth.

Tips are fixed by means of a Wrench at the extremity of the Ultrasonic Handpieces. Tips are in direct contact to the patient. The ultrasonic mechanical vibrations transmitted to the Tip permits the realization of the clinical procedures defined in Intended Use. Bone Surgery Tips are Single Use and delivered under Sterile State. Dental Tips are reusable.

Pre-Market Notification 510(k) Submission for PIEZOTOME M+ By SATELEC K163610 - Traditional 510(k) - Confidential Document - April 10, 2017 - Version 02

Image /page/4/Picture/18 description: The image shows the logo for ACTEON. The logo is in a dark gray color and is in all caps. There is a blue swoosh that starts at the bottom left of the logo and curves up and around the left side of the word.

006 - 2

www.acteondroub.c

{5}------------------------------------------------

Additional Information - February 17, 2017

V - INDICATION FOR USE

Piezotome® M+ is an ultrasonic system consisting of handpieces and associated tips, for cutting bone, bone substitutes and metal. The system can be used for osteoplasty, decorticating, drilling, shaping, and smoothing of bones and teeth, in a variety of surgical procedures, including general orthopaedic, otolaryngological, maxillofacial, oral, hand, foot, neurosurgical spine, and plastic/reconstructive surgery.

Piezotome® M+ is to supply utilities to and serve as a base for dental tools such as ultrasonic scaler, bone cutting instrument and accessories for use by qualified dental practitioners in periodontics, endodontics, scaling, prosthesis and oral surgery.

VI - COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES (SUBSTANTIAL EQUIVALENCE)

A comparison Between PIEZOTOME M+ et The Predicate Devices is present in Table 02.

New DevicePredicate Device #1(PD#1)Predicate Device #2(PD#2)
Trade /Device NamePIEZOTOME M+Piezoelectric SystemPIEZOTOME 2Equivalence
KnumberUnknownK100410K091331
Intended UsePiezotome® M+ is anultrasonic surgicalsystem consisting ofhandpiecesandassociated tips, forcutting bone, bonesubstitutes and metal.The system can beused for osteotomy,osteoplasty,decorticating, drilling,shaping,andsmoothing of bonesand teeth, in a varietyof surgicalprocedures, includinggeneral orthopaedic,otolaryngological,maxillofacial,oral,hand,foot,neurosurgicalspine,andplastic/reconstructivesurgery.Piezotome® M+ is tosupply utilities to andserve as a base fordental tools such asultrasonicscaler,bonecuttinginstrumentandaccessories for use byqualifieddentalpractitionersinperiodonticsThe PiezoelectricSystem, distributed bySynthes, is anultrasonic surgicalsystem consisting ofhandpieces andassociated tips, forcutting bone, bonesubstitutes and metal.The system can beused for osteotomy,osteoplasty,decorticating, drilling,shaping, andsmoothing of bonesand teeth, in a varietyof surgicalprocedures, includinggeneral orthopaedic,otolaryngological,maxillofacial, oral,hand, foot,neurosurgical spine,andplastic/reconstructivesurgery.The intended use ofthe SATELECPIEZOTOME 2 is tosupply utilities to andserve as a base fordental tools such asultrasonic scaler,bone cuttinginstrument andaccessories for use byqualified dentalSubstantiallyequivalent toPD#1 for BoneSurgery clinicalapplication - Noimpact.Substantiallyequivalent toPD#2 for Dentalclinicalapplication - Noimpact
New DevicePredicate Device #1(PD#1)Predicate Device #2(PD#2)Equivalence
Trade /Device NamePIEZOTOME M+Piezoelectric SystemPIEZOTOME 2
KnumberUnknownendodontics, scaling,prosthesis and oralsurgery.K100410K091331practitioners inperiodontics,endodontics, scaling,prosthesis and oralsurgery.
Code ProductJDXJDXDZIIdentical toPD#1
Subsequent productCodesDZI, ERL, HBE, HWEDZI, ERL, HBE, HWE/Identical toPD#1
Part888 - OrthopedicDevice888 - OrthopedicDevice872 - Dental devicesIdentical toPD#1
Regulation Number21 CFR 888.458021 CFR 888.458021 CFR 872.4120Identical toPD#1
RegulationIdentificationA sonic surgicalinstrument is a hand-held device withvarious accessories orattachments, such asa cutting tip thatvibrates at highfrequencies, and isintended for medicalpurposes to cut boneor other materials,such as acrylicA sonic surgicalinstrument is a hand-held device withvarious accessories orattachments, such asa cutting tip thatvibrates at highfrequencies, and isintended for medicalpurposes to cut boneor other materials,such as acrylicA bone cuttinginstrument andaccessories is a metaldevice intended foruse in reconstructiveoral surgery to drill orcut into the upper orlower jaw and may beused to prepare boneto insert a wire, pin, orscrew. The deviceincludes the manualbone drill and wiredriver, powered bonedrill, rotary bonecutting handpiece,and AC-poweredbone sawIdentical toPD#1
Regulation NameSonic surgicalinstrument andaccessoriesattachmentsSonic surgicalinstrument andaccessoriesattachmentsBone CuttingInstrument andAccessoriesIdentical toPD#1
Regulatory ClassIIIIIIIdentical toPD#1 and PD#2
Principle of operation
PrincipleAn electrical signal emitted by the medical device is supplied to the dentalpiezo-ultrasonic Handpiece. It comprises a piezoelectric ceramictransducer, which converts the electrical signal into ultrasonic vibrations.Mechanical vibrations are transmitted to a tip attached to the end of theultrasonic HandpieceIdentical toPD#1 and PD#2
Dimensions of the Console
Height / Depth /Witdh (mm)149.5 / 339.9 / 472.9149.5 / 339.9 / 472.9149.5 / 339.9 / 472.9Identical toPD#1 and PD#2
Power Supply
Supply Voltage(VAC)100 to 230100 to 230100 to 230Identical toPD#1 and PD#2
Frequency (Hz)50 to 6050 to 6050 to 60Identical toPD#1 and PD#2
Power Consumption(VA)150150150Identical toPD#1 and PD#2
New DevicePredicate Device #1(PD#1)Predicate Device #2(PD#2)
Trade /Device NamePIEZOTOME M+Piezoelectric SystemPIEZOTOME 2Equivalence
KnumberUnknownK100410K091331
Electrical Safety Classification
EquipmentClassificationClass 1Class 1Class 1Identical toPD#1 and PD#2
Electrical TypeBF TypeBF TypeBF TypeIdentical toPD#1 and PD#2
Additional earthEquipotential plugEquipotential plugEquipotential plugIdentical toPD#1 and PD#2
Applied Parts on the Console
ofQuantityconnectorsforApplied Parts222Identical toPD#1 and PD#2
LocalizationRight and left of thefront panelRight and left of thefront panelRight and left of thefront panelIdentical toPD#1 and PD#2
Internal constitution of the Console
Power supplyPower SwitchingSupply (PSU)Power SwitchingSupply (PSU)Power SwitchingSupply (PSU)Identical toPD#1 and PD#2
Mother BoardSame Mother boardSame Mother boardSame Mother boardIdentical toPD#1 and PD#2
Display BoardSame Display BoardSame Display BoardSame Display BoardIdentical toPD#1 and PD#2
Display Board andMother Board FireaspectsUL94 Printed CircuitBoardUL94 Printed CircuitBoardUL94 Printed CircuitBoardIdentical toPD#1 and PD#2
Software Constitution
MotherBoardSoftwareBO209034BO209034BO209034Identical toPD#1 and PD#2
DisplayBoardSoftwareBO209284BO209274BO209284Identical toPD#2
External constitution of the Console
Fireaspects(forCasing)UL 94 V0ThermoplasticMaterialUL 94 V0ThermoplasticMaterialUL 94 V0ThermoplasticMaterialIdentical toPD#1 and PD#2
PicturesImage: [device]Image: [device]Image: [device]Difference ofcolors -Identical Shape- No Impact
Piezoelectric Performances for Bone Surgery
OutputUltrasonicfrequencyMin 28 kHzMin 28 kHzMin 28 kHzIdentical toPD#1 and PD#2
MaximumHandpiececurrentrange270 mA270 mA270 mAIdentical toPD#1 and PD#2
Available modesD1, D2, D3, D4D1, D2, D3, D4D1, D2, D3, D4Identical toPD#1 and PD#2
FrequencyModulation for D1,D2, D360 Hz60 Hz60 HzIdentical toPD#1 and PD#2
FrequencyModulation for D430 Hz30 Hz30 HzIdentical toPD#1 and PD#2
Piezoelectric Performances for Dental Application
OutputUltrasonicfrequencyMin 28 kHz-Min 28 kHzIdentical toPD#2
New DevicePredicate Device #1(PD#1)Predicate Device #2(PD#2)
Trade /Device NamePIEZOTOME M+Piezoelectric SystemPIEZOTOME 2Equivalence
KnumberUnknownK100410K091331
MaximumHandpiece currentrange100 mA-100 mAIdentical toPD#2
Available modesSOFT / MEDIUM /HIGH / BOOST-SOFT / MEDIUM /HIGH / BOOSTIdentical toPD#2
Irrigation Performances
Irrigation Pumps222Identical toPD#1 and PD#2
Minimum IrrigationFlow Rate (nominal)10 ml/min10 ml/min10 ml/minIdentical toPD#1 and PD#2
Maximum IrrigationFlow Rate (nominal)120 ml/min120 ml/min120 ml/minIdentical toPD#1 and PD#2
User Interface for setting
User interfaceforSetting5.1" color touchscreen LCD on frontpanel5.1" color touchscreen LCD on frontpanel5.1" color touchscreen LCD on frontpanelIdentical toPD#1 and PD#2
UltrasonicPowerSettingAccessible on 5.1"color touch screenLCD on front panelAccessible on 5.1"color touch screenLCD on front panelAccessible on 5.1"color touch screenLCD on front panelIdentical toPD#1 and PD#2
IrrigationFlowSettingAccessible on 5.1"color touch screenLCD on front panelAccessible on 5.1"color touch screenLCD on front panelAccessible on 5.1"color touch screenLCD on front panelIdentical toPD#1 and PD#2
Footswitch
Type of FootswitchMultifunctionMultifunctionMultifunctionIdenticaltoPD#1 and PD#2
Number of actuatorson Footswitch355No impact
Environmental
Where usedOperating TheatreOperating TheatreDental officeIdentical toPD#1
OperatingTemperature+10°C to +30°C+10°C to +30°C+10°C to +40°CIdentical toPD#1
StorageTemperature0°C to +50°C-20°C to +70°C-20°C to +70°CNo impact
Anatomical siteGeneralOral SphereGeneralOral sphereSubstantiallyequivalenttoPD#2 for Dentalclinicalapplication - NoimpactSubstantiallyequivalenttoPD#1 for BoneSurgery clinicalapplication - Noimpact.
Handpieces
Bone SurgeryHandpiecePIEZOTOME M+Handpiece (greycolor)PiezoelectricHandpiece(black color)PIEZOTOMEHandpiece(grey color)SubstantiallyEquivalent - NoImpact
Dental HandpieceNEWTRONHandpiece-NEWTRONHandpieceIdentical toPD#2
Type of Applied PartUltrasonic HandpieceUltrasonic HandpieceUltrasonic HandpieceIdentical toPD#1 and PD#2
Trade /Device NameNew DevicePredicate Device #1(PD#1)Predicate Device #2(PD#2)Equivalence
KnumberPIEZOTOME M+Piezoelectric SystemPIEZOTOME 2
Technology used forHandpiecesUnknownK100410K091331
PiezoelectrictechnologyPiezoelectrictechnologyPiezoelectrictechnologyIdentical toPD#1 and PD#2
Dimensions of the Bone Surgery Handpieces
Length (mm)130130130Identical toPD#1 and PD#2
Maximum Diameter(mm)232323Identical toPD#1 and PD#2
Dimensions of the Dental Handpieces
Length (mm)112112Identical toPD#2
Maximum Diameter(mm)2121Identical toPD#2
Irrigation Tubing
Length (mm)355035502550Identical toPD#1
Principle of irrigationPeristaltic tapePeristaltic tapePeristaltic tapeIdentical toPD#1 and PD#2
Dimensions of the Footswitch
Width / Height /Depth (mm)311 / 181 / 209311 / 181 / 209311 / 181 / 209Identical toPD#1 and PD#2
Bone Ultrasonic Tips
Tips dedicated forOsteotomy,osteoplasty,shaping, andsmoothingof bonesand teethBS1LSaw 20.1 x 21.4 x 4.0x 0.6 mm -03.000.402S-SubstantiallyEquivalent - NoImpact
BS4Scalpel Rnd 22.45 x12.6 x 3.9 mm dia x0.7 mm -03.000.405S-
SL1Diamond 24.6 x 12.85x 2.6 x 0.6 mm -03.000.409S-
SL2Diamond Rnd 21.9 x12.4 x 1.8 mm -03.000.410S-
BS1 XXLSaw Slim 104 mm -03.000.412S-
BS2L XLSaw L Long -03.000.418S-
BS2R XLSaw R Long -03.000.419S-
BS6 XXLScalpel Flat 15° 129mm - 03.000.421S -03.000.411S-
BS1 RDSaw 20 mm -03.000.424S-
Tips dedicated forDecorticatingSL3Sinus Lift 22.9 x 10.1x 5 dia x 0.4 mm -03.000.411S
Cleaning, Disinfection
DentalUltrasonicTipsRe-usable,Re-usable,Identical toPD#2
Cleanable,Cleanable,
DisinfectableDisinfectable
New DevicePredicate Device #1(PD#1)Predicate Device #2(PD#2)
Trade /Device NamePIEZOTOME M+Piezoelectric SystemPIEZOTOME 2Equivalence
KnumberUnknownK100410K091331
BoneUltrasonicTipsSingle UseSingle UseRe-usable,Cleanable,DisinfectableIdentical toPD#1
DentalUltrasonicHandpieceRe-usable,Cleanable,Disinfectable-Re-usable,Cleanable,DisinfectableIdentical toPD#2
BoneUltrasonicHandpieceRe-usable,Cleanable,DisinfectableRe-usable,Cleanable,DisinfectableRe-usable,Cleanable,DisinfectableIdentical toPD#1 and PD#2
Sterilization / Sterile State
DentalUltrasonicTipsSterilizable 132°C / 4 min - Re-usable-Sterilizable 132°C / 4 min - Re-usableIdentical toPD#2
BoneUltrasonicTipsProvided underSterile StateSingle UseProvided underSterile StateSingle UseSterilizable 132°C / 4 minRe-usableIdentical toPD#1
DentalUltrasonicHandpieceSterilizable 132°C / 4 min-Sterilizable 132°C / 4 minIdentical toPD#1
BoneUltrasonicHandpieceSterilizable 132°C / 4 minSterilizable 132°C / 4 minSterilizable 132°C / 4 minIdentical toPD#1 and PD#2
Irrigation TubingProvided underSterile State - Single UseProvided underSterile State - Single UseProvided underSterile StateSingle UseIdentical toPD#1 and PD#2
Biocompatibility
Tipmaterial incontact to thePatientStainless Steel orMedical Material usedin Medical ApplicationStainless Steel orMedical Material usedin Medical ApplicationStainless Steel orMedical Material usedin Medical ApplicationIdentical toPD#1 and PD#2
UltrasonicHandpiece materialin contact to PatientPolyphenylsulfone(PPSU) used inMedical ApplicationPolyphenylsulfone(PPSU) used inMedical ApplicationPolyphenylsulfone(PPSU) used inMedical ApplicationIdentical toPD#1 and PD#2
Safety / EMC and Additional Safety Mark for USA and CANADA
Safety andEMCStandardIEC 60601-1IEC 60601-1-2IEC 60601-1IEC 60601-1-2IEC 60601-1IEC 60601-1-2Identical toPD#1 and PD#2
USElectricalStandardANSI/AAMI ES60601-1:2005/(R)2012UL 60601-1: 2003No impact

Table 02 - Comparison Between PIEZOTOME M+ et the Predicate Devices

Pre-Market Notification 510(k) Submission for PIEZOTOME M+ By SATELEC K163610 - Traditional 510(k) - Confidential Document - April 10, 2017 - Version 02

Image /page/5/Picture/10 description: The image shows the logo for ACTEON. The logo is in a dark gray color and is in all capital letters. There is a blue curved line that starts on the left side of the logo and goes up and over the top of the letters.

006 - 3

{6}------------------------------------------------

Additional Information – February 17, 2017

Pre-Market Notification 510(k) Submission for PIEZOTOME M+ By SATELEC K163610 - Traditional 510(k) - Confidential Document - April 10, 2017 - Version 02

Image /page/6/Picture/4 description: The image shows the logo for ACTEON. The logo is in a dark gray color and features a stylized letter "A" with a curved line extending from the bottom left of the "A" and wrapping around the word. The word "ACTEON" is written in all capital letters.

006 - 4

{7}------------------------------------------------

Additional Information – February 17, 2017

Pre-Market Notification 510(k) Submission for PIEZOTOME M+ By SATELEC K163610 - Traditional 510(k) - Confidential Document - April 10, 2017 - Version 02

Image /page/7/Picture/4 description: The image shows the logo for ACTEON. The logo is in black and has a blue swoosh underneath it. The text is in all caps and is a sans-serif font.

006 - 5

{8}------------------------------------------------

Additional Information – February 17, 2017

Pre-Market Notification 510(k) Submission for PIEZOTOME M+ By SATELEC K163610 - Traditional 510(k) - Confidential Document - April 10, 2017 - Version 02

Image /page/8/Picture/4 description: The image shows the logo for ACTEON. The logo is in a dark gray color and is in all caps. There is a blue swoosh underneath the logo.

006 - 6

{9}------------------------------------------------

Additional Information – February 17, 2017

Pre-Market Notification 510(k) Submission for PIEZOTOME M+ By SATELEC K163610 - Traditional 510(k) - Confidential Document - April 10, 2017 - Version 02

Image /page/9/Picture/4 description: The image shows the logo for ACTEON. The logo consists of the word "ACTEON" in bold, sans-serif font. A blue curved line is located to the left of the word, starting below the "A" and extending upwards and to the left. The word "ACTEON" is in a dark gray color.

006 - 7

{10}------------------------------------------------

Additional Information - February 17, 2017

A lot of features and characteristics of PIEZOTOME M+ are identical to the Predicate Device #1 and Predicate Device #2. The Identified differences have no impact on the Substantial Equivalence.

Pre-Market Notification 510(k) Submission for PIEZOTOME M+ By SATELEC K163610 - Traditional 510(k) - Confidential Document – April 10, 2017 - Version 02

006 - 8

Image /page/10/Picture/5 description: The image shows the logo for Acteon. The logo consists of the word "ACTEON" in a bold, sans-serif font. A blue curved line is below the word, starting from the left and extending upwards, partially encircling the word. The logo is simple and modern.

{11}------------------------------------------------

Additional Information - February 17, 2017

VII - PERFORMANCE DATA

Electromagnetic Compatibility Test: The Electromagnetic Compatibility Tests have been performed according to IEC 60601-1-2:2007, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility -Requirements and tests (Edition 3). (Recognition Number 5-64).

Electrical Safety Tests: The Electrical Safety Tests have been performed according to IEC 60601-1:2005, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (General). (Recognition Number 5-71) included USA and Canadian Deviations.

Additional Safety Mark: PIEZOTOME M+ is in accordance with ANSI / AAMI ES 60601-1: 2005 / (R) 2012 and bears the *c BVS **us marking, related to the file number GSAA.

Software Verification and Validation: Software Verification activities were conducted and documented according to the document named "Guidance for Industry and FDA Staff - Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" dated May 11, 2005. Moreover, the Software System included in PIEZOTOME M+ is strictly the same Software as the Predicate Device PIEZOTOME 2 (K091331, cleared December 11, 2009).

Performance Testing bench: The comparison of performances between PIEZOTOME M+ and Predicate Devices has been performed.

The Technical Characteristics for Bone Surgery Application of PIEZOTOME M+ (Piezoelectric Performances and Irrigation Performances) are similar as the Predicate Devices Piezoelectric System (K100410, cleared April 28, 2010) and PIEZOTOME 2 (K091331, cleared December 11, 2009). The Technical Characteristics for Dental Application of PIEZOTOME M+ (Piezoelectric Performances and Irrigation Performances) are similar as the Predicate Device PIEZOTOME 2 (K091331, cleared December 11, 2009). The Bone Surgery Tips for PIEZOTOME M+ are similar as the Bone Surgery Tips of the Predicate Device Piezoelectric System (K100410, cleared April 28, 2010).

Sterilization Validation: The Sterilizability Tests have been performed according to applicable Standards for Re-Usable accessories and Single Use Accessories provided under Sterile State.

Biocompatibility Validation: The Biocompatibility Tests have been performed according to applicable Standards for Accessories in contact to the Patient. Tests were conducted according to the ISO 10993-1 Standard and 510(k) Memorandum - #G95-1 Table 1 "Initial Evaluation Tests for Consideration requirements (Body Contact: External communicating Device - Type of Tissue / Bone Dentin Communicating - Contact Duration: A limited less than 24 hours).

VIII - CONCLUSION

Following all information contained in this Pre-Market Notification File, we can declare that the SATELEC PIEZOTOME M+ is Substantially Equivalent (SE) to the SATELEC Predicate Devices Piezoelectric System (K100410, cleared April 28, 2010) and PIEZOTOME 2 (K091331, cleared December 11, 2009).

End of Section (No other information after this line expect Appendix if applicable)

Pre-Market Notification 510(k) Submission for PIEZOTOME M+ By SATELEC K163610 - Traditional 510(k) - Confidential Document - April 10, 2017 - Version 02

Image /page/11/Picture/15 description: The image shows the logo for ACTEON. The logo is in a dark gray color and is in all caps. There is a blue swoosh that goes from the bottom left of the image to the top left.

006 - 9

§ 888.4580 Sonic surgical instrument and accessories/attachments.

(a)
Identification. A sonic surgical instrument is a hand-held device with various accessories or attachments, such as a cutting tip that vibrates at high frequencies, and is intended for medical purposes to cut bone or other materials, such as acrylic.(b)
Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 888.9.