K Number
K163587

Validate with FDA (Live)

Manufacturer
Date Cleared
2017-03-24

(94 days)

Product Code
Regulation Number
880.2805
Age Range
12 - 120
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Use the VERIFY® Incubator for Assert™ Self Contained Biological Indicators SCBI (Incubator) to incubate and automatically read VERIFY Assert Self-Contained Biological Indicators at 57 °C for a fluorescent result within 40 minutes.

Device Description

The VERIFY® Incubator for Assert™ Self Contained Biological Indicators (Incubator) is an incubator/reader designed for use specifically with the VERIFY Assert Self-Contained Biological Indicator (SCBI), see K162701, currently under review. The incubator/reader provides a constant temperature range to allow for activation and outgrowth of Geobacillus stearothermophilus leading to production of a fluorescent moiety. The presence of an increasing fluorescence signal due to increasing concentrations of this fluorescent moiety in the SCBI is detected by the incubator/reader and indicates the presence of viable test microorganisms.

AI/ML Overview

The document provided describes the "VERIFY® Incubator for Assert™ Self Contained Biological Indicators" and its nonclinical testing to demonstrate substantial equivalence to a predicate device. This device is an incubator/reader used to incubate and automatically read self-contained biological indicators for sterilization processes.

Here's a breakdown of the acceptance criteria and study proving the device meets them:

1. Table of Acceptance Criteria and the Reported Device Performance:

TestAcceptance CriteriaReported Device Performance (Conclusion)
Maintenance of Incubation TemperatureMaintain 55-60 °C for a minimum of 40 minutes (incubation time of Verify Assert SCBI as stated in K162701)PASS
Qualification testing with Verify Assert SCBI40-minute fluorescent read meets >97% alignment with 7-day growth results per FDA guidance on reduced incubation time.PASS
Qualification testing with Verify Assert SCBIPass testing with SCBI exposed to full cycle exposure and negative growth result in incubator.PASS
Qualification testing with Verify Assert SCBIFail testing with SCBI exposed in abbreviated cycle exposure and positive result in incubator.PASS
Alarm, LED and Print function TestDemonstrate proper function of alarms, LED and print outputs.PASS

Study Proving the Device Meets Acceptance Criteria:

The document describes nonclinical performance testing.

2. Sample size used for the test set and the data provenance:

  • Sample Size: The document does not explicitly state the specific number of biological indicators (sample size) used for the qualification testing. It refers to "testing with SCBI exposed to full cycle exposure" and "testing with SCBI exposed in abbreviated cycle exposure," implying multiple indicators were tested.
  • Data Provenance: The document doesn't specify the country of origin for the data. Given it's a submission to the FDA (United States), it is highly likely the testing was conducted in the USA or by a facility adhering to US regulations. The testing is described as "performance testing," which typically refers to prospectively collected data for validation purposes.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This device is an incubator/reader for biological indicators. The "ground truth" for this type of device is the actual growth or non-growth of microorganisms in the biological indicator, determined through traditional microbiological culturing methods (e.g., 7-day growth results for biological indicators). There is no mention of human experts (like radiologists) being involved in establishing the ground truth for this specific device, as its function is to automate the reading of a biological indicator. The "7-day growth results" serve as the objective ground truth.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

Not applicable. The ground truth is based on the biological end-point (microbial growth/no growth), not subjective human interpretation requiring adjudication.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

Not applicable. This device is an automated incubator/reader for biological indicators, not an AI-assisted diagnostic imaging device that involves human reader interpretation. No human readers are involved in the "reading" process of this device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

Yes, this is effectively a standalone performance study. The device itself (the incubator/reader) provides the fluorescent read, which is then compared against the established biological "truth" (7-day growth results). Its primary function is automated reading without human interpretation of its output beyond acknowledging the alarm or visual indicator.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

The ground truth used is the microbiological growth status of the biological indicators, specifically the "7-day growth results." This is an objective measurement of whether the microorganisms in the biological indicator proliferated or not, which is the definitive indicator of sterilization effectiveness.

8. The sample size for the training set:

Not applicable. This document describes a traditional medical device (an incubator/reader) validated through performance testing, not a machine learning or AI-based device that typically requires a separate training set. The device's "knowledge" is built into its design (temperature control, fluorescence detection) rather than learned from a data set.

9. How the ground truth for the training set was established:

Not applicable, as there is no training set mentioned for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a series of human profiles facing to the right.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 24, 2017

Steris Corporation Anthony Piotrkowski Senior Regulatory Affairs Manager 5976 Heisley Rd Mentor, Ohio 44060

Re: K163587

Trade/Device Name: Verify Incubator For Assert Self Contained Biological Indicators Regulation Number: 21 CFR 880.2805 Regulation Name: Recombinant Biological Indicator Regulatory Class: Class II Product Code: OWP Dated: February 17, 2017 Received: February 21, 2017

Dear Anthony Piotrkowski:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Michael J. Ryan -S

for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K163587

Device Name

VERIFY® Incubator for Assert™ Self Contained Biological Indicators

Indications for Use (Describe)

Use the VERIFY® Incubator for Assert™ Self Contained Biological Indicators SCBI (Incubator) to incubate and automatically read VERIFY Assert Self-Contained Biological Indicators at 57 °C for a fluorescent result within 40 minutes.

Type of Use (Select one or both, as applicable)Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for STERIS. The word STERIS is in all caps and in a serif font. Below the word STERIS is a graphic of several horizontal wavy lines that are blue.

510(k) Summary For K163587 VERIFY® Incubator for Assert™ Self-Contained Biological Indicators

Sponsor Facility

STERIS Corporation 5960 Heisley Road Mentor, OH 44060 Phone: (440) 354-2600 Fax No: (440) 357-9198

Manufacturing Facility

STERIS Corporation 9325 Pinecone Drive Mentor, OH 44060 Phone: (440) 392-7800 Fax No: (440) 392-7896

Contact: Anthony Piotrkowski

Telephone: (440) 392-7437 Fax No: (440) 357-9198 e-mail: tony piotrkowski(@steris.com

Submission Date: March 22, 2017

510(k) Number: K163587

STERIS Corporation ■ 5960 Heisley Road ■ Mentor, OH 44060-1834 USA ■ 440-354-2600

{4}------------------------------------------------

1. Device Name

Trade Name:VERIFY® Incubator for Assert™ Self ContainedBiological Indicators
Common/usual Name:Incubator/Reader (accessory to Biological Indicator)
Device Classification:Class II
Classification Name:Recombinant Biological Indicator(21 CFR 880.2805, OWP)

2. Predicate Device

VERIFY CRONOS SCBI K102469/DEN110006

Reference Device: 3M Attest 390 Auto-reader K123546

3. Description of Device

The VERIFY® Incubator for Assert™ Self Contained Biological Indicators (Incubator) is an incubator/reader designed for use specifically with the VERIFY Assert Self-Contained Biological Indicator (SCBI), see K162701, currently under review. The incubator/reader provides a constant temperature range to allow for activation and outgrowth of Geobacillus stearothermophilus leading to production of a fluorescent moiety. The presence of an increasing fluorescence signal due to increasing concentrations of this fluorescent moiety in the SCBI is detected by the incubator/reader and indicates the presence of viable test microorganisms.

4. Intended Use/ Indications for Use

Use the VERIFY® Incubator for Assert™ Self Contained Biological Indicators SCBI (Incubator) to incubate and automatically read VERIFY Assert Self-Contained Biological Indicators at 57 °C for a fluorescent result within 40 minutes.

5. Summary of Technical Characteristics

A comparison of technical characteristics are summarized in Tables 1 and 2.

{5}------------------------------------------------

FeatureVERIFY Incubator forAssert SCBI(proposed)VERIFY CRONOSSCBI(K102469/DEN110006)PredicateComparison
Intended UseUse the VERIFY®Incubator for Assert™ SelfContained BiologicalIndicators SCBI(Incubator) to incubate andautomatically readVERIFY Assert Self-Contained BiologicalIndicators at 57 °C for afluorescent result within40 minutes.The Verify Cronos Self-Contained BiologicalIndicator for SteamSterilization Processes isintended for use by healthcare providers as astandard method tomonitor steam sterilizationprocesses. The VerifyCronos is validated to beused in the followingsteam sterilization cycles:1.) Standard Prevac270°F/132°C for 4 minutes2.) Standard Prevac275°F/135°C for 3 minutes3.) Standard Gravity250°F/121°C for 30 minutes4.) Standard Gravity270°F/132°C for 15 minutes5.) Standard Gravity275°F/135°C for 10 minutes6.) Standard SFPP270°F/132°C for 4minutesThe predicate incubator wasan accessory to theBiological Indicator and didnot have its own intendeduse or indications for use.The proposed device is anaccessory to the VERIFYAttest SCBI.
Basis ofReadoutPhotodiode detectsfluorescence produced byenzymatic activity thatresults from growingbiological indicatororganismsPhotodiode detectsfluorescence produced byenzymatic activity thatresults from growingbiological indicatororganismsSame
IncubationTemperatureRange55 - 60 °C55 - 60 °CSame
Voltage Range90 to 264 VAC with 12VDC conversion.90 to 264 VAC with 12VDC conversion.Same
Test capacity8 wells8 wellsSame
CalibrationFactory calibration - nocalibration by customerFactory calibration - nocalibration by customerSame
IncubationTime40 minutes40 minutesSame
FeatureVERIFY Incubator forAssert SCBI(proposed)VERIFY CRONOSSCBI(K102469/DEN110006)PredicateComparison
FluorescenceDetectionActive enzyme from thespores cleaves thefluorogenic substratecontained within theSCBI's media to produce afluorescent molecule.Active enzyme from thespores cleaves thefluorogenic substratecontained within theSCBI's media to produce afluorescent molecule.Same
Indication ofResultsPositive - audible alarm,visual LED lights andscreenNegative - no alarm,visual indication withLED lights and LCDscreenUser must acknowledgeresultsPositive - audible alarm,visual LED lights andscreenNegative - no alarm,visual indication withLED lights and LCDscreenUser must acknowledgeresultsSame
ComplianceElectrical Safety and EMCTesting• IEC 61010-1 (2010)Third Ed• IEC 61010-2-010(2013) Third EdElectromagneticcompatibility• USA Title 47, Codeof FederalRegulations (2007)for:• Radiated Emissions(FCC Part 15,Subpart B, Class A)• ConductedEmissions (FCCPart 15, Subpart B,Class A)• IEC 61326:2013 -• EN 55011:2009, Inc.A1:2010• EN 61000-3-2:2006,Inc. A1:2009 andA2:2009EN 61000-3-3-2013Electrical Safety and EMCTesting• IEC 61010-1 secondEd• IEC 61010-2-010Second EdElectromagneticcompatibility• USA Title 47, Codeof FederalRegulations (2007)for:• Radiated Emissions(FCC Part 15,Subpart B, Class A)• ConductedEmissions (FCCPart 15, Subpart B,Class A)• IEC 61326:• EN 55011:• EN 61000-3-2:2006,Inc. A1:2009 andA2:2009EN 61000-3-3Both meet applicableelectrical safety standards
FeatureVERIFY Incubator forAssert SCBI(proposed)VERIFY CRONOSSCBI(K102469/DEN110006)PredicateComparison
SystemOperationThe reader/incubator wellsare arranged in 2 banks of4 wells and preset to 57°C.The measurement offluorescence is initiated byplacement of a VERIFYAssert SCBI into any ofthe incubation wells andpressing the adjacent"ACTION" button.When an SCBI is placedinto a well, the auto-readerdetects its presence. Uponpressing the buttonassociated with that well, ablinking yellowish lightindicates that incubation isin process and the readinitiated.A "positive" reading isinterpreted as anindication of a potentialsterilization cycle failure.A "positive" finding isindicated to the user byred light on the front paneladjacent to the well, by anaudible alarm, and by textdisplayed on the LCDscreen. The alarm must bemuted by the operatorwhen a positive result isobtained. The LCD screenprovides instructions forthe user to turn off thealarm for that specificSCBI.Should another BI become"positive", the alarm willsound again and the aboveactions are repeated.If the result is not positiveat the end of theincubation time, the resultis negative. Negativeresults are identified by agreen light on the frontpanel adjacent to the wellwith the "negative" BI andby text on the LCD.The reader/incubator wellsare arranged in 2 banks of4 wells and preset to 57°C.The measurement offluorescence is initiated byplacement of a VERIFYAssert SCBI into any ofthe incubation wells andpressing the adjacent"ACTION" button.When an SCBI is placedinto a well, the auto-readerdetects its presence. Uponpressing the buttonassociated with that well, ablinking yellowish lightindicates that incubation isin process and the readinitiated.A "positive" reading isinterpreted as anindication of a potentialsterilization cycle failure.A "positive" finding isindicated to the user byred light on the front paneladjacent to the well, by anaudible alarm, and by textdisplayed on the LCDscreen. The alarm must bemuted by the operatorwhen a positive result isobtained. The LCD screenprovides instructions forthe user to turn off thealarm for that specificSCBI.Should another BI become"positive", the alarm willsound again and the aboveactions are repeated.If the result is not positiveat the end of theincubation time, the resultis negative. Negativeresults are identified by agreen light on the frontpanel adjacent to the wellwith the "negative" BI andby text on the LCD.Same

Table 1 Summary of Incubator Physical Description and Technological Properties vs Predicate

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

Table 2 Summary of Incubator Physical Description and Technological Properties vs Reference Device

FeatureVERIFY Incubator forAssert SCBI(proposed)Attest 390(K123546) ReferenceComparison
Intended UseUse the VERIFY®Incubator for Assert™ SelfContained BiologicalIndicators SCBI(Incubator) to incubate andautomatically readVERIFY Assert Self-Contained BiologicalIndicators at 57 °C for afluorescent result within40 minutes.The 3M Attest 390 Auto-reader is designed toincubate and automaticallyread the 3M Attest Rapidreadout Biologicalindicators for Steam 1291,1292, at 60 °C for a finalfluorescent result at 1 hourfor 1291 and 3 hours for1292.Both are intended forincubation and automaticreading of specific self-contained biologicalindicators.
Basis ofReadoutPhotodiode detectsfluorescence produced byenzymatic activity thatresults from growingbiological indicatororganismsPhotodiode detectsfluorescence produced byenzymatic activity thatresults from growingbiological indicatororganismsSame
IncubationTemperatureRange55 - 60 °C60 ± 2 °CBoth operate within thegrowth temperature range ofGeobacillusstearothermophilus
Voltage Range90 to 264 VAC with 12VDC conversion.100-240 VAC with 12VDC conversion.Similar range and sameoperating voltage. Electricalsafety testing demonstratessuitability.
Test capacity8 wells10 wellsBoth can accommodatemultiple samples
CalibrationFactory calibration - nocalibration by customerFactory calibration - nocalibration by customerSame
IncubationTime40 minutes1 hour (1291), 3 hours(1292)Reduced incubation timetesting in accordance withFDA guidance is provided inthis submission
FluorescenceDetectionActive enzyme from thespores cleaves thefluorogenic substratecontained within theSCBI's media to produce afluorescent molecule.UV LEDs are used toexcite the fluorescentmolecule produced byenzyme cleavage of thefluorogenic substratecontained in the SCBI'smedia. The emitted lightis detected by aphotodiode.Similar Detection Method -photodiode sensors
FeatureVERIFY Incubator forAssert SCBI(proposed)Attest 390(K123546) ReferenceComparison
SystemOperationThe reader/incubator wellsare arranged in 2 banks of4 wells and preset to 57°C.The measurement offluorescence is initiated byplacement of a VERIFYAssert SCBI into any ofthe incubation wells andpressing the adjacent"ACTION" button.When an SCBI is placedinto a well, the auto-readerdetects its presence. Uponpressing the buttonassociated with that well, ablinking yellowish lightindicates that incubation isin process and the readinitiated.A "positive" reading isinterpreted as anindication of a potentialsterilization cycle failure.A "positive" finding isindicated to the user byred light on the front paneladjacent to the well, by anaudible alarm, and by textdisplayed on the LCDscreen. The alarm must bemuted by the operatorwhen a positive result isobtained. The LCD screenprovides instructions forthe user to turn off thealarm for that specificSCBI.Should another BI become"positive", the alarm willsound again and the aboveactions are repeated.If the result is not positiveat the end of theincubation time, the resultis negative. Negativeresults are identified by agreen light on the frontpanel adjacent to the wellwith the "negative" BI andby text on the LCD.The measurement offluorescence is initiated byplacement of an AttestSCBI into any of theincubation wells. Thereader/incubator wells arecolor-coded to match thecolor of the top of theSCBIs. The incubatorblock may be color coded.When a BI is placed into awell, its presence isdetected. The LCD screenbelow the SCBI shows thetime remaining tocomplete the incubationperiod.Detection of a positiveresult is indicated by a '+'sign on the LCD screen,accompanied by anaudible alarm. A positiveresult is indicative of asterilization cycle failure.The alarm must be mutedby the operator when apositive result is obtained.Pressing the mute buttondisables the alarm only forthe specific SCBI whichwas just identified aspositive. Should anotherBI become "positive", thealarm will again sound.If the result is not positiveat the end of theincubation time, the resultis negative. Negativeresults are identified by a'- 'sign on the LCDscreen.Both have similar modes ofoperation only type ofbiological indicator isintended for use with theproposed device so no colorcoding is required and bothalarm for a positive BI.
FeatureVERIFY Incubator forAssert SCBI(proposed)Attest 390(K123546) ReferenceComparison
Indication ofResultsPositive - audible alarm,visual LED lights andscreenNegative - no alarm,visual indication withLED lights and LCDscreenUser must acknowledgeresultsPositive - audible alarm(if enabled), "+" indicationon LCD panel, andflashing of LCD panelbacklightingNegative - no alarm, "-"indication on LCD panelSimilar Indication of Results
ComplianceElectrical Safety and EMCTestingIEC 61010-1 (2010)Third Ed IEC 61010-2-010(2013) Third Ed Electromagneticcompatibility USA Title 47, Codeof FederalRegulations (2007)for: Radiated Emissions(FCC Part 15,Subpart B, Class A) ConductedEmissions (FCCPart 15, Subpart B,Class A) IEC 61326:2013 - EN 55011:2009, Inc.A1:2010 EN 61000-3-2:2006,Inc. A1:2009 andA2:2009 EN 61000-3-3:2013Electrical Safety and EMCTesting IEC 61010-1 (2001)Second Ed IEC 61010-2-010(2003) Second Ed Electromagneticcompatibility USA Title 47, Codeof FederalRegulations (2009)for: Radiated Emissions(FCC Part 15,Subpart B, ClassA) ConductedEmissions (FCCPart 15, Subpart B,Class A) IEC 61326:Both meet relevant electricalstandards for safety,emissions and compatibilitythat were active at time ofsubmission.

{9}------------------------------------------------

{10}------------------------------------------------

6. Summary of Nonclinical Tests

Performance testing to demonstrate substantial equivalence to the predicate has been completed and is summarized in Table 3 below.

TestAcceptance CriteriaConclusion
Maintenance ofIncubationTemperatureMaintain 55-60 °C for a minimum of 40 minutes(incubation time of Verify Assert SCBI as statedin K162701)PASS
Qualification testingwith Verify AssertSCBI40-minute fluorescent read meets >97%alignment with 7-day growth results per FDAguidance on reduced incubation time.PASS

Table 3. Summary of Non-clinical Testing

{11}------------------------------------------------

TestAcceptance CriteriaConclusion
Qualification testingwith Verify AssertSCBIPass testing with SCBI exposed to full cycleexposure and negative growth result in incubatorPASS
Fail testing with SCBI exposed in abbreviatedcycle exposure and positive result in incubatorPASS
Alarm, LED andPrint function TestDemonstrate proper function of alarms, LED andprint outputsPASS

7. Conclusion

The VERIFY® Incubator for Assert™ Self-Contained BI's has met the established performance criteria. The results of the studies demonstrate that the biological indicator performs as intended, and based on the nonclinical tests performed, the subject device is substantially equivalent and is as safe and as effective as the legally marketed predicate device Class II (21 CFR 880.2805, Product code OWP).

N/A