K Number
K163550

Validate with FDA (Live)

Device Name
Pobling MIITY 2
Date Cleared
2017-12-20

(366 days)

Product Code
Regulation Number
882.5890
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Pobling MITY 2 is intended for facial stimulation and is indicated for over-the counter cosmetic use.

Device Description

Pobling MIITY 2 is a facial toning device and it is composed of main unit and adaptor. The micro-current is delivered to healthy facial skin through head of main unit to stimulate the face. The device is a battery-powered and hand-held device. There have heating and cooling function by pressing the button additionally.

AI/ML Overview

The document describes the acceptance criteria and supporting studies for the Pobling MIITY 2, a Transcutaneous Electrical Nerve Stimulator for facial stimulation. The primary study presented is a comparison with predicate devices to demonstrate substantial equivalence, rather than a standalone clinical trial or MRMC study.

Here's an breakdown of the requested information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are implicitly defined by demonstrating substantial equivalence to the predicate devices (Carol Cole Company, NuFACE® Plus, K103472 and Nu Skin Enterprises, Inc., NuSkin Facial Spa, K122711) across various technical characteristics and safety standards. The reported device performance is presented as a direct comparison in the table on page 4.

FeatureAcceptance Criteria (Implied by Predicate Performance)Pobling MIITY 2 Performance
Regulation Number21 CFR § 882.589021 CFR § 882.5890
Regulation NameTranscutaneous Electrical Nerve StimulatorTranscutaneous Electrical Nerve Stimulator
Regulatory ClassClass IIClass II
Product CodeNFONFO
Intended UseStimulate the faceStimulate the face
Indication for UseOver-the-Counter Cosmetic UseOver-the-Counter Cosmetic Use
Power SourceBattery operatedBattery operated
Number output channels11
Automatic shut offYesYes
Indicator Display (ON/OFF)YesYes
Indicator Display (Low-Battery)YesYes
On Time (seconds)ConstantConstant
Off Time (seconds)NoneNone
Compliance with Voluntary StandardsIEC 60601-1 / IEC 60601-1-2 / IEC 60601-1-11IEC 60601-1 / IEC 60601-1-2 / IEC 60601-1-11
ModeOne to Five (Predicate NuSkin Facial Spa) / One (Predicate NuFACE® Plus)Two
WaveformPulsed Monophasic (Predicate NuFACE® Plus) / Direct Current (Predicate NuSkin Facial Spa)Direct
Maximum Output Voltage (@500Ω)137mV (NuFACE® Plus); 188mV-184mV (NuSkin Facial Spa)Mode1: 174mVMode2: 174mV
Maximum Output Voltage (@2kΩ)769mV (NuFACE® Plus); 722mV-741mV (NuSkin Facial Spa)Mode1: 710mVMode2: 710mV
Maximum Output Voltage (@10kΩ)3.82V (NuFACE® Plus); 2.626V-3.787V (NuSkin Facial Spa)Mode1: 3200mVMode2: 3200mV
Maximum Output Current (@500Ω)274uA (NuFACE® Plus); 376uA-369uA (NuSkin Facial Spa)Mode1: 350uAMode2: 350uA
Maximum Output Current (@2kΩ)387uA (NuFACE® Plus); 361uA-373uA (NuSkin Facial Spa)Mode1: 360uAMode2: 360uA
Maximum Output Current (@10kΩ)383uA (NuFACE® Plus); 362uA-379uA (NuSkin Facial Spa)Mode1: 320uAMode2: 320uA
Heating and CoolingNot specified in predicates as primary functionMode1: Heating (Temp. 41℃±3℃)Mode2: Cooling (Temp. 10℃±3℃)
Weight9oz (NuFACE® Plus); 120g (NuSkin Facial Spa)115g
Dimensions3"x5.25"x1.25" (NuFACE® Plus); 143x31.2x67.5mm (NuSkin Facial Spa - Large)150x500mm (unit)
Housing material and constructionThermo Plastic (NuFACE® Plus); ABS, Chrome plate, ABS, rubber (elastomer) (NuSkin Facial Spa)ABS

2. Sample size used for the test set and the data provenance:

  • Sample Size: Not explicitly stated for specific tests beyond the overall device. The testing described for electrical safety, EMC, biocompatibility, technical characteristic, and thermal safety would involve a sample of the Pobling MIITY 2 device(s) itself.
  • Data Provenance: The document does not specify the origin (e.g., country) of the data for the tests. It refers to "our testing method" and compliance with international standards (IEC, ISO). The comparison data for the predicate devices would be from publicly available information or submissions related to those devices. The study is non-clinical performance assessment.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This information is not provided in the document, as the study is a non-clinical performance assessment comparing the device to predicates and demonstrating compliance with technical standards. There is no mention of experts establishing ground truth for a test set in the context of diagnostic performance or clinical outcomes.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

This information is not provided and is not applicable, as the study is a non-clinical performance assessment and not a reader study.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This device is not an AI-assisted diagnostic tool, but rather a direct-action therapeutic (stimulating) device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

Yes, a standalone performance assessment was conducted for the Pobling MIITY 2 device, focusing on its technical characteristics, electrical safety, EMC, biocompatibility, and thermal safety. The device's performance was evaluated independently to ensure it met relevant standards and its characteristics were comparable to predicates. There is no algorithm involved; it's a hardware device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

The "ground truth" for the non-clinical performance tests was based on:

  • Compliance with international standards: IEC 60601-1, IEC 60601-1-2, ISO 10993.
  • Technical specifications: Measured output voltage and current values compared against the reported values of the predicate devices.
  • Functional verification: For heating and cooling properties.

There is no "expert consensus," "pathology," or "outcomes data" in the typical sense as this is a non-clinical technical and safety assessment for a device without a diagnostic or complex analytical function.

8. The sample size for the training set:

This information is not applicable as the device is a hardware product and does not involve AI or machine learning algorithms that require a training set.

9. How the ground truth for the training set was established:

This information is not applicable as there is no training set for this device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 20, 2017

HABALAN Med & Beauty Co., Ltd c/o Peter Chung President Plus Global 300 Atwood Pittsburgh, Pennsylvania 15213

Re: K163550

Trade/Device Name: Pobling MIITY 2 Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: NFO Dated: December 7, 2017 Received: December 13, 2017

Dear Peter Chung:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Vivek J. Pinto -S

for Carlos L. Peña. PhD. MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K163550

Device Name Pobling MIITY 2

Indications for Use (Describe)

The Pobling MITY 2 is intended for facial stimulation and is indicated for over-the counter cosmetic use.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for HABALAN, with the words "Different Insight" underneath. The logo is in a bold, sans-serif font. The letters are all capitalized and connected to each other. The words "Different Insight" are in a smaller, sans-serif font.

510(k) Summary

Prepared in accordance with the content and format regulatory requirements of 21 CFR Part 807.92

1. Submitter's Information

    1. Name : HABALAN Med & Beauty Co., Ltd
    1. Address : 907, 219, Gasandigital 1-ro, Geumcheon-gu, Seoul, KOREA
    1. Contact person : Mr. Peter Chung

300 Atwood Street Pittsburgh, PA 15213, USA

Tel : 412-512-8802

    1. Date prepared : December, 05, 2016

2. Device Information

    1. Trade/Proprietary Name : Pobling MIITY 2
    1. Common Name : Facial Toning Device
    1. Classification Name : Transcutaneous Electrical Nerve Stimulator
    1. Device Class : Class II
    1. Product Code : NFO
    1. Regulation Number : 21 CFR 882.5890

3. Predicate Device

    1. Carol Cole Company, NuFACE® Plus, K103472
    1. Nu Skin Enterprises, Inc., NuSkin Facial Spa, K122711

4. Intended Use

The Pobling MIITY 2 is intended for facial stimulation.

5. Indication for Use

The Pobling MIITY 2 is indicated for over-the counter cosmetic use.

6. Device Description

Pobling MIITY 2 is a facial toning device and it is composed of main unit and adaptor. The micro-current is delivered to healthy facial skin through head of main unit to stimulate the face. The device is a battery-powered and hand-held device. There have heating and cooling function by pressing the button additionally.

7 Comparison with Predicate Device

The MITY 2 is identical with predicate devices for intended use and indications for use. It is

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows a stylized wordmark for "HABALAN" in a bold, sans-serif font. The letters are interconnected, creating a continuous, flowing design. The color of the text is black, and the background is white, providing a high contrast.

Different Insight

used as same technological principle with predicate devices to accomplish its intended use. all are hand-held, battery powered device and the there is no new questions of safety and/or effectiveness.

The determination of substantial equivalence for the MITY2 is based on an assessment of non-clinical performance. This assessment included a comparison of the MIITY2 to that of predicate. The output performance testing included:

  • Maximum output voltaget
  • Maximum output current
FeaturesSubject DevicePredicate DeviceJustification
MIITY 2NuFACE® PlusNuSkin Facial Spa
510(k) NumberTo be assignedK103472K122711
Regulation Number21 CFR § 882.5890Same
Regulation NameTranscutaneous Electrical Nerve StimulatorSame
Regulatory ClassClass IISame
Product CodeNFOSame
Intended UseStimulate the faceSame
Indication for UseOver-the-Counter Cosmetic UseSame
Power SourceBattery operatedSame
Number outputchannels1Same
Automatic shut offYesSame
Indicator Display
1) ON / OFF?YesSame
2) Low-Battery?YesSame
On Time (seconds)ConstantSame
Off Time (seconds)NoneSame
Compliance withVoluntary StandardsIEC 60601-1 / IEC 60601-1-2 / IEC 60601-1-11Same
ModeTwoOneOne to FiveSimilar
Wave formDirectPulsed MonophasicDirect CurrentSame
Maximum OutputVoltage@500ΩMode1: 174mVMode2: 174mV@500Ω137mV@500ΩMode1: 188mV(214mV)Mode2: 189mV(225mV)Mode3: 185mV(220mV)Mode4: 184mV(212mV)Mode5: 184mV(213mV)Similar
@2kΩMode1: 710mVMode2: 710mV@2kΩ769mV@2kΩMode1: 722mV(1527mV)Mode2: 746mV(812mV)Mode3: 735mV(815mV)Mode4: 740mV(802mV)Mode5: 741mV(801mV)Similar
Subject DevicePredicate DeviceJustification
FeaturesMIITY 2NuFACE® PlusNuSkin Facial Spa
Maximum Output@10kΩ@10kΩ@10kΩSimilar
VoltageModel: 3200mV3.82VModel: 2.626V(2.978V)
Mode2: 3200mVMode2: 3.787V(4.011V)
Mode3: 3.747V(4.014V)
Mode4: 3.503V(3.980V)
Mode5: 3.729V(3.980V)
Maximum Output@500Ω@500Ω@500ΩSimilar
CurrentModel: 350uA274uAModel: 376uA(427uA)
Mode2: 350uAMode2: 379uA(449uA)
Mode3: 371uA(440uA)
Mode4: 368uA(424uA)
Mode5: 369uA(427uA)
@2kΩ@2kΩ@2kΩSimilar
Model: 360uA387uAModel: 361uA(763uA)
Mode2: 360uAMode2: 373uA(406uA)
Mode3: 367uA(407uA)
Mode4: 370uA(401uA)
Mode5: 370uA(400uA)
@10kΩ@10kΩ@10kΩSimilar
Model: 320uA383uAModel: 362uA(398uA)
Mode2: 320uAMode2: 379uA(401uA)
Mode3: 375uA(401uA)
Mode4: 350uA(398uA)
Mode5: 373uA(398uA)
Heating and CoolingModel: HeatingMinor
(Temp. 41℃±3℃)function
Mode2: Cooling
(Temp. 10℃±3℃)
Weight115g9oz120g (approx. 4oz)
Facial Spa with Large
Dimensions150 x 500unit : mm3" x 5.25" x 1.25"143 x 31.2 x 67.5 mm
(5.6" x 1.2" x 2.7")
Facial Spa with Small
136 x 31.2 x 67.5 mm
(5.4" x 1.2" x 2.7")
Housing materialABSThermo PlasticABS, Chrome plate-
and constructionABS, rubber (elastomer)

TABLE : Comparison of MIITY 2 with predicate devices

{5}------------------------------------------------

The heating and cooling are not a major function to be used as accomplished its intended use as stimulator. That are minor for distinguishing the mode. And also conducted the performance data. Therefore, all the difference don't affect the safety and effectiveness which is concluded after all the required testing, So no safety and effectiveness issues relating to the MIITY2

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for HABALAN. The logo is black and consists of the word "HABALAN" in a stylized font. Below the logo, the words "Different Insight" are written in a smaller, sans-serif font.

8. Performance Data

The following performance data were provided in support of the substantial equivalence determination 1) Electrical Safety and Electromagnetic compatibility(EMC)

Electrical safety and EMC testing were conducted on MITY2, it comples with the IEC 60601-1:2005 +am1:2002 for safety and the IEC 60601-1-2:2007 standard for EMC.

    1. Biocompatibility Testing
      The biocompatibility evaluation for the contact part with healthy facial skin of MIITY2 was conducted in accordance with the ISO 10993, Cytotoxicity, Skin Sensitization and Skin Irritation testing, it was considered contacting part and duration (Tissue contacting, Duration of less than 24hours)
    1. Technical characteristic testing
      To show the technical characteristic testing about Transcutaneous Electrical Nerve Stimulator, the test was conducted in our testing method.
    1. Thermal Safety testing
      The device have the heating and cooling function with stimulation. To verify the heating and cooling properties, we performed the testing in our testing method.
    1. Clinical data
      Not applicable

9. Conclusion

The fundamental technology to be accomplished its intended use in view of technological and operational characteristics between MITY2 and predicate device is same. Non-clinical performance as documented support the safety and effectiveness are similar to the predicate device. In other words, the MIITY2 is substantially equivalent to the predicate devices.

§ 882.5890 Transcutaneous electrical nerve stimulator for pain relief.

(a)
Identification. A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.(b)
Classification. Class II (performance standards).