K Number
K163530

Validate with FDA (Live)

Date Cleared
2017-05-03

(138 days)

Product Code
Regulation Number
876.5860
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The PrisMax control unit is intended for:
• Continuous Renal Replacement Therapy (CRRT) for patients weighing 20 kilograms or more with acute renal failure and/or fluid overload.

All treatments administered via the PrisMax control unit must be prescribed by a physician.

Device Description

The PrisMax System is intended for Continuous Renal Replacement Therapy (CRRT) for patients with acute renal failure and/or fluid overload. The PrisMax System offers four Continuous Renal Replacement Therapy (CRRT) options: Slow Continuous Ultrafiltration (SCUF), Continuous Veno-Venous Hemofiltration (CVVH), Continuous VenoVenous Hemodialysis (CVVHD), and Continuous Venovenous Hemodialfiltration (CVVHDF). The proposed device PrisMax consists of PrisMax Control Unit and accessories for removing effluent.

AI/ML Overview

The provided text describes a 510(k) premarket notification for the PrisMax device, which is a high-permeability hemodialysis system. The core of this submission is to demonstrate the substantial equivalence of the PrisMax to a predicate device, the Prismaflex.

Here's an analysis of the acceptance criteria and the study proving the device meets them, based solely on the provided text. It's important to note that a 510(k) summary often summarizes extensive testing, so not all details of the actual studies might be present in this brief document.

First and foremost, this document describes a device comparison for substantial equivalence to a predicate, not a study evaluating clinical performance against a specific disease outcome or a MRMC study. Therefore, some of the requested information (like effect size of human readers improving with AI, or number of experts for ground truth) is not applicable to this type of regulatory submission. The acceptance criteria here are about equivalence to the predicate device's established performance parameters and safety.


1. Table of Acceptance Criteria and the Reported Device Performance

The acceptance criteria are generally that the PrisMax device performs at least as well as or equivalently to the predicate Prismaflex device for various technical specifications. The reported device performance is the PrisMax's specification for that feature. The table below is extracted directly from the "Table 3. Substantial Equivalence Table Device Comparison" in the provided document. The "Acceptance Criteria" column reflects the predicate device's performance, as the goal is to demonstrate equivalence or improvement without raising new safety/effectiveness concerns.

FeatureAcceptance Criteria (Predicate Prismaflex 7.10, K131516)Reported Device Performance (Proposed PrisMax 1.0.6.0)
Indications for useContinuous Renal Replacement Therapy (CRRT) for patients weighing 20 kilograms or more with acute renal failure and/or fluid overload. All treatments administered via the Prismaflex control unit must be prescribed by a physician.Continuous Renal Replacement Therapy (CRRT) for patients weighing 20 kilograms or more with acute renal failure and/or fluid overload. All treatments administered via the PrisMax control unit must be prescribed by a physician.
Dedicated Disposable Sets Available in U.S.For CRRT: M60/M100/M150, HF1000 & HF1400For CRRT: M60/M100/M150, HF1000 & HF1400
Syringe Sizes20, 30 & 50 ml20 & 50 ml
AnticoagulationUser-controllable as continuous or bolusUser-controllable as continuous or bolus
Dialysate Flow Rate CVVH & CVVHDF (Range)0 to 8000 ml/hr0 to 8000 ml/hr
Dialysate Flow Rate CVVH & CVVHDF (Increment)50 ml/hr10 ml/hr
Dialysate Flow Rate Accuracy± 30 ml/hr± 30 ml/hr
Replacement solution / Fluid Flow Rate CVVH & CVVHDF (Range)0 to 8000 ml/hr0 to 8000 ml/hr
Replacement solution / Fluid Flow Rate CVVH & CVVHDF (Increment)50 ml/hr10 ml/hr
Replacement Flow Rate Accuracy± 30 ml/hr± 30 ml/hr
Blood Flow Rate (Range)10-450 ml/min10-450 ml/min
Blood Flow Rate Accuracy±10 % of user-set rate at nominal blood flow of 450 ml/min or the highest achievable disposable blood flow, having 37 °C, at an access pressure of -200 mmHg and without any PBP flow±10 % of user-set rate at nominal blood flow of 450 ml/min or the highest achievable disposable blood flow, having 37 °C, at an access pressure of -200 mmHg and without any PBP flow
Pre-Blood Pump Flow Rate SCUF (Range)0 to 1000 ml/hr (but actual was 0 to 2000 ml/hr)0 to 2000 ml/hr
Pre-Blood Pump Flow Rate CVVH, CVVHD, CVVHDF (Range)0 to 4000 ml/hr0 to 4000 ml/hr
Pre-Blood Pump Accuracy± 30 ml/hr± 30 ml/hr
Effluent Pump Flow Rate (Range)0 to 10,000 ml/h, Depending on the therapy selected.0 to 10,000 ml/h, Depending on the therapy selected.
ECG DischargerYESYES
TherapiesSCUF, CVVH, CVVHD, CVVHDFSCUF, CVVH, CVVHD, CVVHDF
PumpsPBP solution, Replacement solution, Dialysate solution, Effluent, BloodPBP solution, Replacement solution, Dialysate solution, Effluent, Blood
ScalesDialysate, Replacement, Effluent, Pre blood (PBP)Dialysate, Replacement, Effluent, Pre blood (PBP)
Trans Membrane Pressure Alarms (TMP CRRT)User settable: +70 to +350 mmHg, Default: +350 mmHgDefault: +300 mmHg
Dialysate Conductivity and TemperatureNot controlled by PrismaflexNot controlled by PrisMax
Patient Fluid Removal Performance (Range)0 to 2000 ml/hr0 to 2000 ml/hr
Patient Fluid Removal Performance (Increment)10 ml/hr5 ml/hr
Patient Fluid Removal Performance (Accuracy)± 30 ml/hr, ± 70 ml/3hr, ± 300 ml/24hr. Scales calibrated at ambient temperature at which they will be used. Ambient temperature change less than ±3 °C (5.4 °F) during treatment.± 30 ml/hr, ± 70 ml/3hr, ± 300 ml/24hr. Scales calibrated at ambient temperature at which they will be used. Ambient temperature change less than ±3 ℃ (5.4 °F) during treatment.
Access Pressure Sensor (Range)-250 to +450 mmHg-250 to +450 mmHg
Access Pressure Sensor (Accuracy)±15 mmHg±15 mmHg
Return Pressure Sensor (Range)-50 to +350 mmHg-50 to +350 mmHg
Return Pressure Sensor (Accuracy)±5 mmHg±5 mmHg
Filter Pressure Sensor (Range)-50 to +450 mmHg-50 to +450 mmHg
Filter Pressure Sensor (Accuracy)±15 mmHg±15 mmHg
Effluent Pressure Sensor (Range)-350 to +400 mmHg (CRRT)-350 to +400 mmHg (CRRT)
Effluent Pressure Sensor (Accuracy)±15 mmHg±15 mmHg
Control Unit SoftwareVersion 7.10Version 1.0.6.0

Study Proving Device Meets Acceptance Criteria

The document states: "Performance testing was conducted on the PrisMax System to evaluate the functional performance of the system. The performance testing confirms PrisMax remains as safe and effective as Prismaflex and is substantially equivalent."

The nature of the "study" demonstrating this is predominantly nonclinical performance testing, verification, and validation against engineering and regulatory standards, and comparison to the predicate device's specifications.

Specifically, the document explicitly mentions:

  • Design validation: The PrisMax design validation meets user needs and intended use and is substantially equivalent to the predicate.
  • Compliance with IEC 60601-2-16 Hemodialysis Equipment: Testing confirmed by CSA, a recognized test laboratory, for essential performance.
  • Electrical safety testing: According to IEC 60601-1 Edition 3.1. This includes reports for software, alarms, usability, safety, and performance.
  • Electromagnetic compatibility (EMC) testing.
  • Risk Assessment and risk control measures: Hazard analysis (therapy level, product level, process level) confirming the device does not perform in an unexpected or unsafe manner.
  • Labeling, Software including cybersecurity, Human Factors: These have been successfully implemented.
  • Verification and validation tests: These were performed subsequent to risk analysis and include Human Factors and Software Validation.

The key changes and their justification for not raising new safety/effectiveness concerns are footnoted in Table 3:

  • Syringe Sizes (30ml removed): Due to infrequent use, device behavior unchanged.
  • Dialysate Flow Rate / Replacement Solution Flow Rate Increment (50ml/hr to 10ml/hr): Allows for more precise settings, verified and validated to comply to updated specifications without new risks.
  • Pre-Blood Pump Flow Rate (SCUF range clarification): Corrected the predicate's stated range to its actual operational range, showing PrisMax is equivalent. Verified and validated.
  • Trans Membrane Pressure Alarms (Fixed default of +300 mmHg, removed user settable): Removes infrequent use case, reduces use-error risk, increases usability, lowers default to reduce risk. Verified and validated.
  • Patient Fluid Removal Performance Increment (10ml/hr to 5ml/hr): Increases setting range, allows more precise setting. Verified and validated.
  • The software update (version 1.0.6.0) was verified and validated subsequent to risk analysis and includes Human Factors and Software Validation.

Details based on the provided text:

  • 2. Sample size used for the test set and the data provenance:

    • Sample Size: Not explicitly stated in terms of patient count or device unit count for the validation tests. These are typically engineering performance tests, not clinical trials with "patients." The performance tests would use a relevant number of devices or test conditions to ensure specifications are met.
    • Data Provenance: The testing appears to be internal "performance testing" and "verification and validation" conducted by Baxter Healthcare Corporation and by a recognized test laboratory (CSA) for specific standards like IEC 60601-2-16. This is non-clinical, laboratory-based data, not patient data from a specific country.
  • 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable in the context of a 510(k) submission based on technical performance and equivalence to a predicate device, rather than a clinical diagnostic study requiring expert interpretation of results. The "ground truth" here is adherence to engineering specifications and safety standards.
  • 4. Adjudication method for the test set:

    • Not applicable, as this is not a study requiring human adjudication for diagnostic categorization. The "adjudication" is through engineering verification and validation processes against predefined technical specifications and standards.
  • 5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This is a medical device for continuous renal replacement therapy, not an AI/diagnostic imaging device that would typically involve human readers. Therefore, an MRMC study is not relevant here.
  • 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable in the way this question typically refers to AI algorithms. The device itself is a "standalone" system in its function, but its performance is verified through engineering tests against objective criteria, not an "algorithm only" performance separate from the device's integrated function. Its operation relies on physician prescription (human-in-the-loop for clinical application, but not for the device's technical performance evaluation).
  • 7. The type of ground truth used:

    • Engineering Specifications and Performance Standards: The "ground truth" for this submission is compliance with established engineering specifications for flow rates, accuracy, pressure ranges, alarm settings, and adherence to international safety standards (e.g., IEC60601 series). Additionally, the predicate device's cleared performance serves as a comparative ground truth for equivalence.
  • 8. The sample size for the training set:

    • Not applicable. This is not a machine learning or AI device that requires a "training set" in the conventional sense. The "training" for the device's development would be iterative engineering design and testing.
  • 9. How the ground truth for the training set was established:

    • Not applicable, as there is no "training set" in the AI/ML context. The functional requirements for the device are derived from clinical needs, regulatory standards, and the performance of previous generations of such devices.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized image of an eagle with three heads, all facing to the right. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circle around the eagle image.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 3, 2017

Baxter Healthcare Corporation Chris Scavotto Senior Manager Regulatory Affairs 32650 N. Wilson Road Round Lake, IL 60073

K163530 Re:

Trade/Device Name: PrisMax Regulation Number: 21 CFR§ 876.5860 Regulation Name: High Permeability Hemodialysis System Regulatory Class: II Product Code: KDI Dated: March 31, 2017 Received: April 4, 2017

Dear Chris Scavotto:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Joyce M. Whang -S

for

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K163530

Device Name PrisMax

Indications for Use (Describe)

The PrisMax control unit is intended for:

· Continuous Renal Replacement Therapy (CRRT) for patients weighing 20 kilograms or more with acute renal failure and/or fluid overload.

All treatments administered via the PrisMax control unit must be prescribed by a physician.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

|X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Section 5. 510(k) Summary

December 15th, 2016

OWNER:

Baxter Healthcare Corporation One Baxter Parkway Deerfield, Illinois 60015

CONTACT PERSON:

Chris Scavotto Senior Regulatory Manager, Global Regulatory Affairs 32650 N Wilson Road Round Lake, IL 60073 Telephone: 763-463-4621 Fax: 763-463-4606

IDENTIFICATION OF THE DEVICE:

Common Name: Hemodialysis Delivery System Trade Name or Proprietary Name: PrisMax Classification Panel: 78 Gastroenterology and Urology Classification: High Permeability Hemodialysis System (876.5860) Class: Class II Product Code: 78KDI

Table 1. Product Code(s) for PrisMax System
---------------------------------------------------
Code NumberName
115369PrisMax control unit
115370PrisMax Accessory, Auto Effluent
115485PrisMax Accessory, Auto Effluent Extension

PREDICATE DEVICE:

Table 2. Predicate Device(s)

DeviceCompanyPredicate 510(k)Clearance Date
PrismaflexGambro Lundia ABK1315161/3/2014

{4}------------------------------------------------

DESCRIPTION OF THE DEVICE:

The PrisMax System is intended for Continuous Renal Replacement Therapy (CRRT) for patients with acute renal failure and/or fluid overload. Reference the PrisMax control unit in Figure 1.

The goals of acute renal failure treatments are removal of waste products, restoration of acid-base balance; correction of electrolyte imbalances (e.g., hyperkalemia), patient fluid balance, nutritional support, and other conditions in which fluid removal is needed. PrisMax System offers four Continuous Renal Replacement Therapy (CRRT) options: Slow Continuous Ultrafiltration (SCUF), Continuous Veno-Venous Hemofiltration (CVVH), Continuous VenoVenous Hemodialysis (CVVHD), and Continuous Venovenous Hemodialfiltration (CVVHDF).

Image /page/4/Picture/5 description: The image shows a dialysis machine with a screen on top. The screen displays various parameters and readings related to the dialysis process. The machine has several tubes, pumps, and filters that are essential for the dialysis procedure. There are also bags hanging from the machine, which likely contain fluids used during the treatment.

Figure 1. PrisMax control unit

The proposed device PrisMax, which is the subject of this Traditional 510(k) premarket notification, consists of PrisMax Control Unit. Also included are accessories for removing effluent. The proposed device PrisMax uses the current marketed device

{5}------------------------------------------------

Prismaflex as the predicate. The Prismaflex was previously cleared under 510(k) premarket notification K131516 (cleared on January 3rd, 2014).

INDICATIONS FOR USE:

The PrisMax control unit is intended for:

  • . Continuous Renal Replacement Therapy (CRRT) for patients weighing 20 kilograms or more with acute renal failure and/or fluid overload.
    All treatments administered via the PrisMax control unit must be prescribed by a physician.

TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE:

The differences between the predicate Prismaflex 7.10 and the proposed device PrisMax as reflected in the Substantial Equivalence (SE) Table are considered substantially equivalent. Changes included in the SE table are marked and discussed in Table 3.

Identification of Predicate Device:

  • Trade Name: Prismaflex, 7.10
  • Manufacturer: Gambro Lundia AB, Sweden
  • Common Name: Hemodialysis Delivery System.
  • Clearance: K0131516 cleared in January 2014.

Identification of Proposed Device:

  • Trade Name: PrisMax, 1.0.6.0.
  • Manufacturer: Gambro UF Solutions, Minneapolis Minnesota
  • Common Name: Hemodialysis Delivery System.
  • Clearance: Pending.

PrisMax will not utilize the TPE2000 Blood Tubing Set. It will utilize the CRRT Blood Tubing sets referenced in the Substantial Equivalence Table (M60/M100/M150 + HF1000/HF1400) all which have been previously cleared. The predicate device Prismaflex was cleared in 2011 to include TPE2000 Blood Sets. The PrisMax is not

{6}------------------------------------------------


requesting clearance for TPE Blood Sets. References to TPE have been removed from Table 3.

FeaturesSEProposedPrisMax 1.0.6.0PredicatePrismaflex 7.10, K131516
Indications for useSEThe PrisMax control unit is intended for:• Continuous Renal Replacement Therapy (CRRT) for patients weighing 20 kilograms or more with acute renal failure and/or fluid overload.All treatments administered via the PrisMax control unit must be prescribed by a physician.The Prismaflex control unit is intended for:• Continuous Renal Replacement Therapy (CRRT) for patients weighing 20 kilograms or more with acute renal failure and/or fluid overload.All treatments administered via the Prismaflex control unit must be prescribed by a physician.
Dedicated Disposable Sets Available in U.S.SEFor CRRT:M60/M100/M150HF1000 & HF1400For CRRT:M60/M100/M150HF1000 & HF1400
Syringe Sizes ASE20 & 50 ml20, 30 & 50 ml
AnticoagulationSEUser-controllable as continuous or bolusUser-controllable as continuous or bolus
Dialysate Flow RateBSECVVH & CVVHDF:Range: 0 to 8000 ml/hrIncrement: 10 ml/hrCVVH & CVVHDF:Range: 0 to 8000 ml/hrIncrement: 50 ml/hr
Dialysate Flow Rate AccuracySE± 30 ml/hr± 30 ml/hr
Replacement solution / Fluid Flow RateCSECVVH & CVVHDF:Range: 0 to 8000 ml/hrIncrement: 10 ml/hrCVVH & CVVHDF:Range: 0 to 8000 ml/hrIncrement: 50 ml/hr
Replacement Flow Rate AccuracySE± 30 ml/hr± 30 ml/hr
Blood Flow RateSERange: 10-450 ml/minRange: 10-450 ml/min
Blood Flow Rate AccuracySE±10 % of user-set rate at nominal blood flow of 450 ml/min or the highest achievable disposable blood flow, having 37 °C, at an access pressure of -200 mmHg and without any PBP flow±10 % of user-set rate at nominal blood flow of 450 ml/min or the highest achievable disposable blood flow, having 37 °C, at an access pressure of -200 mmHg and without any PBP flow
ProposedPredicate
FeaturesSEPrisMax 1.0.6.0Prismaflex 7.10, K131516
Pre-Blood PumpFlow RateDSESCUF:Range: 0 to 2000 ml/hrSCUF:Range: 0 to 1000 ml/hr
SECVVH, CVVHD, CVVHDF:Range: 0 to 4000 ml/hrCVVH, CVVHD, CVVHDF:Range: 0 to 4000 ml/hr
Pre-Blood PumpAccuracySE± 30 ml/hr± 30 ml/hr
Effluent PumpFlow RateSERange: 0 to 10,000 ml/hDepending on the therapy selected.Range: 0 to 10,000 ml/hDepending on the therapy selected.
ECG DischargerSEYESYES
TherapiesSESCUFCVVHCVVHDCVVHDFSCUFCVVHCVVHDCVVHDF
PumpsSEPBP solutionReplacement solutionDialysate solutionEffluentBloodPBP solutionReplacement solutionDialysate solutionEffluentBlood
ScalesSEDialysateReplacementEffluentPre blood (PBP)DialysateReplacementEffluentPre blood (PBP)
Trans MembranePressure AlarmsTMP (CRRT)ESECRRT TMP:Default: +300 mmHgCRRT TMP:User settable: +70 to +350 mmHgDefault: +350 mmHg
DialysateConductivity andTemperatureSEDialysate Conductivity andTemperature are not controlled byPrisMaxDialysate Conductivity andTemperature are not controlled byPrismaflex
Patient FluidFRemovalPerformance(Range)SE0 to 2000 ml/hrIncrement: 5 ml/hr0 to 2000 ml/hrIncrement: 10 ml/hr

Table 3. Substantial Equivalence Table Device Comparison

{7}------------------------------------------------


{8}------------------------------------------------

ProposedPredicate
FeaturesSEPrisMax 1.0.6.0Prismaflex 7.10, K131516
Patient FluidRemovalPerformance(Accuracy)SE± 30 ml/hr± 70 ml/3hr± 300 ml/24hrScales calibrated at ambienttemperature at which they will beused. Ambient temperature changeless than ±3 ℃(5.4 °F) during treatment.± 30 ml/hr± 70 ml/3hr± 300 ml/24hrScales calibrated at ambienttemperature at which they will beused. Ambient temperature changeless than ±3 °C(5.4 °F) during treatment.
Access PressureSensorSERange:-250 to +450 mmHgAccuracy:±15 mmHgRange:-250 to +450 mmHgAccuracy:±15 mmHg
Return PressureSensorSERange:-50 to +350 mmHgAccuracy:±5 mmHgRange:-50 to +350 mmHgAccuracy:±5 mmHg
Filter PressureSensorSERange:-50 to +450 mmHgAccuracy:±15 mmHgRange:-50 to +450 mmHgAccuracy:±15 mmHg
Effluent PressureSensorSERange:-350 to +400 mmHg (CRRT)Accuracy:±15 mmHgRange:-350 to +400 mmHg (CRRT)Accuracy:±15 mmHg
Control UnitSoftware GSEVersion 1.0.6.0Version 7.10

Table 3: Foot notes:

A. Syringe Sizes: 20, 30 & 50 ml. Removed 30 ml.

  • . The 30ml syringe is removed due to infrequent use in the market. The device behavior in conjunction with the specified syringes remains strictly unchanged.
  • . The updated specification does not affect safety or effectiveness of the device.

{9}------------------------------------------------

B. Dialysate Flow Rate: CVVH & CVVHDF: Range: 0 to 8000 ml/hr, Increment: 10 ml/hr. Updated the increment from 50ml/hr to 10ml/hr.

  • . The setting range was increased from 50 to 10ml/hr. This allows the practitioner to select a more precise setting and reduces limitation existing with a 50ml/hr increment.
    The updated specification does not introduce new or increased risks to the system and does not introduce new questions of safety and effectiveness. Compliance of the PrisMax System to the updated specifications has been verified successfully. Successful validation has substantiated that the system does not raise new questions of safety and effectiveness.

C. Replacement solution / Fluid Flow Rate: CVVH & CVVHDF: Range: 0 to 8000 ml/hr, Increment: 10 ml/hr. Updated the increment from 50ml/hr to 10ml/hr.

  • . The range was increased from 50 to 10ml/hr. This allows the practitioner to select a more precise setting and reduces limitation existing with a 50ml/hr increment.
    The updated specification does not introduce new or increased risks to the system and does not introduce new questions of safety and effectiveness. Compliance of the PrisMax System to the updated specifications has been verified successfully. Successful validation has substantiated that the system does not raise new questions of safety and effectiveness.

D. Pre-Blood Pump Flow Rate: SCUF: Range: 0 to 2000 ml/hr. Align with Predicate device Prismaflex from clearances K110823 and K131516.

  • . The Prismaflex SE table incorrectly states a range of 0 to 1000 ml/hr vs. the actual of 0 to 2000 ml/hr. As per Prismaflex requirement sss 1283, the predicate device has operated at the Range: 0 to 2000 ml/hr since the original clearance in 2005. The PrisMax is equivalent in practice and requirements. The Prismaflex SE table will be updated on the next submission.
    The updated specification does not introduce new or increased risks to the system and does not introduce new questions of safety and effectiveness. Compliance of the PrisMax System to the updated specifications has been verified successfully. Successful validation has substantiated that the system does not raise new questions of safety and effectiveness.

{10}------------------------------------------------

E. Trans Membrane Pressure Alarms TMP (CRRT): User settable: +70 to +350 mmHg, default +350 mmHg. Implemented fixed default of +300 mmHg.

  • . User Settable TMP was removed due to infrequent use in the predicate device and to reduce risk of use-error handling during set up.
  • Increase usability by removing steps during set up.
  • Implement lowered default to reduce risk.

The updated specification does not introduce new or increased risks to the system and does not introduce new questions of safety and effectiveness. Compliance of the PrisMax System to the updated specifications has been verified successfully. Successful validation has substantiated that the system does not raise new questions of safety and effectiveness.

F. Patient Fluid Removal Performance (Range): 0 to 2000 ml/hr Increment: 5 ml/hr.

Updated the increment from 10ml/hr to 5ml/hr.

  • . Increase setting range by improving increment range of setting from 10 to 5ml/hr. This allows the practitioner to select a more precise setting and reduces limitation existing with a 10ml/hr increment.
    The updated specification does not introduce new or increased risks to the system and does not introduce new questions of safety and effectiveness. Compliance of PrisMax System to the updated specifications has been verified successfully. Successful validation has substantiated that the system does not raise new questions of safety and effectiveness.

G. Control Unit Software: Software version 1.0.6.0 or Rev. A, from Prismaflex 7.10.

The device software on the PrisMax has been implemented correctly. The software has been verified and validated subsequent to risk analysis. The verification and validation tests including Human Factors and Software Validation. The software does not raise questions of safety and effectiveness. PrisMax is considered substantially equivalent to the predicate device Prismaflex.

Substantial Equivalence Summary

The differences between the PrisMax System and its predicate device do not introduce new questions of safety and effectiveness. All modifications have been verified and

{11}------------------------------------------------

validated per Design Controls Activities. As shown through successful verification and validation testing, the PrisMax Control Unit is considered substantially equivalent to its predicate.

DISCUSSION OF NONCLINICAL TESTS:

Performance testing was conducted on the PrisMax System to evaluate the functional performance of the system. The performance testing confirms PrisMax remains as safe and effective as Prismaflex and is substantially equivalent.

In summary, the PrisMax Control Unit has successfully implemented performance requirements and subsequent outputs verifying and validating:

  • . The PrisMax design validation meets the user needs and intended use and is substantially equivalent to the predicate.
  • . The device complies with IEC60601-2-16 Hemodialysis Equipment. Testing was confirmed by CSA, the recognized test laboratory as part of the testing to 60601-1 Edition3.1. The testing specifically confirms the device meets the requirements for Essential Performance according to the particular standard.
  • . Electrical safety testing according to the most recent IEC60601-1 Edition 3.1 standard. The standard includes reports for software, alarms, usability, safety and performance.
  • Electromagnetic compatibility (EMC) testing.
  • Risk Assessment and risk control measures. A therapy level, product level and process level hazard analysis confirms the device doesn't perform in an unexpected or unsafe manor.
  • Labeling, Software including cybersecurity, Human Factors, have been successfully implemented.

SUBSTANTIAL EQUIVALENCE DECISION

Based on the the information provided in this premarket notification, Baxter Healthcare Corporation believes the proposed PrisMax is substantially equivalent, for purposes of section 510(k) of the Federal Food, Drug and Cosmetic Act only, to the predicate device identified in this premarket notification.

§ 876.5860 High permeability hemodialysis system.

(a)
Identification. A high permeability hemodialysis system is a device intended for use as an artificial kidney system for the treatment of patients with renal failure, fluid overload, or toxemic conditions by performing such therapies as hemodialysis, hemofiltration, hemoconcentration, and hemodiafiltration. Using a hemodialyzer with a semipermeable membrane that is more permeable to water than the semipermeable membrane of the conventional hemodialysis system (§ 876.5820), the high permeability hemodialysis system removes toxins or excess fluid from the patient's blood using the principles of convection (via a high ultrafiltration rate) and/or diffusion (via a concentration gradient in dialysate). During treatment, blood is circulated from the patient through the hemodialyzer's blood compartment, while the dialysate solution flows countercurrent through the dialysate compartment. In this process, toxins and/or fluid are transferred across the membrane from the blood to the dialysate compartment. The hemodialysis delivery machine controls and monitors the parameters related to this processing, including the rate at which blood and dialysate are pumped through the system, and the rate at which fluid is removed from the patient. The high permeability hemodialysis system consists of the following devices:(1) The hemodialyzer consists of a semipermeable membrane with an in vitro ultrafiltration coefficient (K
uf ) greater than 8 milliliters per hour per conventional millimeter of mercury, as measured with bovine or expired human blood, and is used with either an automated ultrafiltration controller or anther method of ultrafiltration control to prevent fluid imbalance.(2) The hemodialysis delivery machine is similar to the extracorporeal blood system and dialysate delivery system of the hemodialysis system and accessories (§ 876.5820), with the addition of an ultrafiltration controller and mechanisms that monitor and/or control such parameters as fluid balance, dialysate composition, and patient treatment parameters (e.g., blood pressure, hematocrit, urea, etc.).
(3) The high permeability hemodialysis system accessories include, but are not limited to, tubing lines and various treatment related monitors (e.g., dialysate pH, blood pressure, hematocrit, and blood recirculation monitors).
(b)
Classification. Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Device—Part I: Evaluation and Testing,’ ”
(2) “Guidance for the Content of 510(k)s for Conventional and High Permeability Hemodialyzers,”
(3) “Guidance for Industry and CDRH Reviewers on the Content of Premarket Notifications for Hemodialysis Delivery Systems,”
(4) “Guidance for the Content of Premarket Notifications for Water Purification Components and Systems for Hemodialysis,” and
(5) “Guidance for Hemodialyzer Reuse Labeling.”