K Number
K163525

Validate with FDA (Live)

Date Cleared
2017-09-05

(264 days)

Product Code
Regulation Number
866.5090
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

QUANTA Flash M2 (MIT3) is a chemiluminescent immunoassay for the semi-quantitative determination of IgG antimitochondrial antibodies in human serum. The presence of antibodies, in conjunction with clinical findings and other laboratory tests, is an aid in the diagnosis of primary biliary cholangitis.

QUANTA Flash M2 (MIT3) Calibrators are intended for use with the QUANTA Flash M2 (MIT3) chemiluminescent immunoassay for the determination of IgG anti-mitochondrial antibodies in human serum. Each calibrator establishes a point of reference for the working curve that is used to calculate unit values.

QUANTA Flash M2 (MIT3) Controls are intended for use with the QUANTA Flash M2 (MIT3) chemiluminescent immunoassay for quality control in the determination of IgG anti-mitochondrial antibodies in human serum.

Device Description

The QUANTA Flash M2 (MIT3) assay is designed to run on the BIO-FLASH® instrument. This platform is a fully automated closed system with continuous load and random access capabilities that automatically processes the samples, runs the assay and reports the results. It includes liquid handling hardware, luminometer and computer with software-user interface. The QUANTA Flash M2 (MIT3) assay utilizes a reagent cartridge format, which is compatible with the BIO-FLASH instrument.

Recombinant MIT3 antigen is coated onto paramagnetic beads. The bead suspension is lyophilized and stored in the bead tube. Prior to use in the BIO-FLASH system, the sealed reagent tubes are pierced with the reagent cartridge lid and the beads are rehydrated and resuspension buffer by pipetting up and down with a transfer pipette. The reagent cartridge is then loaded onto the BIO-FLASH instrument. Samples are also loaded onto the instrument in samples are diluted by the BIO-FLASH with system rinse in a small disposable plastic cuvette. Small amounts of the diluted patient serum, the beads, and assay buffer are combined into a second cuvette, and mixed. This cuvette is then incubated at 37°C. The beads are magnetized and washed several times. Isoluminol conjugated anti-human IgG antibodies are then added to the cuvette, and again incubated at 37°C. The beads are magnetized and washed repeatedly. The isoluminol conjugate is oxidized when Trigger 1 (Fe(II)coproporphyrin in sodium hydroxide solution) and Trigger 2 (urea-hydrogen peroxide in sodium chloride solution) are added to the cuvette, and the flash of light produced from this reaction is measured as Relative Light Units (RLU) by the BIO-FLASH optical system. The RLU are proportional to the amount of isoluminol conjugate that is bound to the human IgG, which is in turn proportional to the amount of anti-M2 (MIT3) antibodies bound to the corresponding beads.

For quantitation, the QUANTA Flash M2 (MIT3) assay utilizes a predefined lot specific Master Curve that is uploaded onto the instrument through the reagent cartridge barcode. Every new lot number of reagent cartridge must be calibrated before first use, with the QUANTA Flash M2 (MIT3) Calibrators. Based on the results obtained with the three Calibrators included in the Calibrator Set (sold separately), an instrument specific Working Curve is created, which is used to calculate chemiluminescent units (CU) from the instrument signal (RLU) obtained for each sample.

The QUANTA Flash M2 (MIT3) kit contains the following materials:

One (1) QUANTA Flash M2 (MIT3) Reagent Cartridge One (1) vial of Resuspension buffer One (1) Transfer pipette

The QUANTA Flash M2 (MIT3) reagent cartridge contains the following reagents for 50 determinations:

  • M2 (MIT3) coated paramagnetic beads, lyophilized. a.
  • b. Assay buffer - colored pink, containing Tris-buffered saline, Tween 20, protein stabilizers and preservatives.
  • Tracer IgG Isoluminol labeled anti-human IgG antibodies in buffer, containing protein C. stabilizers and preservative.

The QUANTA Flash M2 (MIT3) Calibrators kit contains two vials each of Calibrator 2, and Calibrator 3:

QUANTA Flash M2 (MIT3) Calibrators:

  • QUANTA Flash M2 (MIT3) Calibrator 1: Two (2) barcode labeled tubes containing 0.3 mL prediluted, ready to use reagent. Calibrators contain human anti-mitochondrial antibodies in stabilizer and preservative.
  • -QUANTA Flash M2 (MIT3) Calibrator 2: Two (2) barcode labeled tubes containing 0.3 mL prediluted, ready to use reagent. Calibrators contain human anti-mitochondrial antibodies in stabilizer and preservative.
  • -QUANTA Flash M2 (MIT3) Calibrator 3: Two (2) barcode labeled tubes containing 0.3 mL prediluted, ready to use reagent. Calibrators contain human anti-mitochondrial antibodies in stabilizer and preservative.

The QUANTA Flash M2 (MIT3) Controls kit contains two vials of Negative Control and two vials of Positive Control:

QUANTA Flash M2 (MIT3) Controls:

  • QUANTA Flash M2 (MIT3) Negative Control: Two (2) barcode labeled tubes containing 0.5 mL, ready to use reagent. Controls contain human anti-mitochondrial antibodies in stabilizer and preservative.
  • QUANTA Flash M2 (MIT3) Positive Control: Two (2) barcode labeled tubes containing 0.5 mL, ready to use reagent. Controls contain human anti-mitochondrial antibodies in stabilizer and preservative.
AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the QUANTA Flash M2 (MIT3) device:

1. Table of Acceptance Criteria and Reported Device Performance

Feature/MetricAcceptance CriteriaReported Device Performance
PrecisionTotal %CV: < 10%Within Run: CV(%) for samples varied from 2.6% to 5.5%. Between Runs: CV(%) for samples varied from 1.1% to 3.6%. Between Days: CV(%) for samples varied from 0.0% to 3.9%. Total: CV(%) for samples varied from 2.9% to 7.6%. (All within <10%)
Reproducibility (Between Sites)Total %CV: < 10%CV(%) for 9 samples varied from 2.1% to 9.1%. (All within <10%)
Reproducibility (Between Lots)All %CV values were within the acceptance limit, 10%.CV(%) for 6 samples varied from 4.0% to 8.7%. (All within <10%)
Linearity1. Best fitting polynomial is a linear one. 2. Otherwise, the difference between the best-fitting nonlinear and linear polynomial is less than 20%, or ± 4 CU, whichever is greater (allowable nonlinearity).All six specimens fulfilled the acceptance criteria, showing good linearity in their respective ranges. Slope (95% CI): Ranged from 0.96 to 1.08. Y-Intercept (95% CI): Ranged from -146.8 to 26.0. R2: All 1.00. Average % Recovery: Ranged from 98.5% to 109.2%. (All within acceptable limits).
Interference85% - 115% recovery, or ± 4 CU difference, whichever is greater.No interference was detected with: - Bilirubin up to 1 mg/mL (recovery: 93% to 100%) - Hemoglobin up to 2 mg/mL (recovery: 94% to 101%) - Triglycerides up to 1000 mg/dL (recovery: 94% to 102%) - Cholesterol up to 332.5 mg/dL (recovery: 97% to 104%) - Human IgG up to 35 mg/mL (recovery: 91% to 115%, or 1.8 CU to 3.7 CU) - Ursodeoxycholic acid up to 0.75 mg/mL (recovery: 97% to 104%) - Prednisone up to 0.3 mg/mL (recovery: 100% to 105%) - Methotrexate up to 9.1 mg/mL (recovery: 96% to 103%) - Cholestyramine up to 18 mg/mL (recovery: 97% to 100%) - Colestipol up to 18 mg/mL (recovery: 98% to 104%) - Hydroxyzine up to 1 mg/mL (recovery 100% to 108%) - Glutamate Oxacatate Transaminase up to 0.12 IU/mL (recovery: 98% to 115%) - HDL Cholesterol up to 3.5 mg/mL (recovery: 98% to 105%) - Rheumatoid factor up to 153.4 IU/mL (recovery: 96% to 111%). (All within acceptable ranges).
Sample Stability90-110% average recovery.All samples fulfilled the acceptance criteria at each time point for each condition (up to 21 days at 2-8°C, up to 48 hours at room temperature, and up to 3 freeze/thaw cycles).
Reagent Shelf Life (Beads)Lower 95% Cl interval of regression line is ≥85% and upper 95% Cl interval is ≤115% at day 14; no individual data point has ≤75% or ≥125% recovery at day 14.All three lots of beads retained between 85% and 115% reactivity (considering the 95% Cl) after two weeks at 37 ± 3℃.
Reagent Shelf Life (Controls/Calibrators)Lower 95% Cl interval of regression line is ≥90% and upper 95% Cl interval is ≤110% at day 14; no individual data point has ≤80% or ≥120% recovery at day 14.All Calibrators and Controls maintained between 90% and 110% reactivity (considering the 95% Cl) when stored at 37 ± 3°C for 2 weeks.
In-use Stability (Calibrators)1. All four calibrations performed in the 8 hour period are successful. 2. Mean Calibrator RLU recovery values for the first 4 calibrations are between 90% and 110% compared to the first use. 3. Control/patient panel CU recovery values are between 85% and 115% of those obtained on the first calibration curve.The first four calibrations performed in the 8 hour period were considered valid. Calibrators yielded average RLU recovery values from 100% to 110%. Control/patient panel CU recovery ranged from 87.8% to 96.6%. (Supports claim of up to 4 calibrations over an 8-hour period).
In-use Stability (Controls)1. All values run within their established range. 2. Linear regression line obtained by plotting %recovery values against the number of runs stays between 85% and 115% at run 15.All controls ran within their respective acceptable ranges for all runs. The regression line remained between 85% and 115% at run 15 for both Controls. (Supports claim of up to 15 uses, at 10 minutes onboard per use).
In-use Stability (Reagent Cartridge)Stability claim established at: - Actual measurement day preceding the 95% confidence interval of regression line reaches 85% or 115% recovery, OR - Actual measurement day preceding the day when 2 data points or ≥2% of recovery data (whichever is greater) is ≤ 75% or ≥ 125% recovery.RP0006 lot: 70 days 151003 lot: 60 days. In-use (onboard) stability was set at 60 days.
Real Time StabilityReagent Cartridge/Negative & Positive Controls: % recovery of each sample between 80-120% (compared to baseline). %CV of replicates <15%. Calibrators: % recovery of trilpicates average between 85% and 115%. %CV of triplicates <10%. Controls: results should fall within their acceptable ranges.All results to date (up to 9 months for reagent cartridge and controls, 3 months for calibrators at time of submission) were within the acceptance limits.
Clinical Sensitivity(Not explicitly stated as a separate acceptance criterion, but presented in the context of clinical performance evaluation against predicate.)84.1% (77.4 - 89.1% CI) for Primary Biliary Cholangitis (PBC)
Clinical Specificity(Not explicitly stated as a separate acceptance criterion, but presented in the context of clinical performance evaluation against predicate.)99.0% (97.6 - 99.6% CI) for Controls (non-PBC patient groups)

2. Sample Size and Data Provenance for Test Set (Clinical Validation Study)

  • Sample Size: A total of 586 characterized samples were included in the Validation Set for the QUANTA Flash M2 (MIT3).
    • For clinical sensitivity and specificity calculations, 568 samples were used (excluding 14 Limited cutaneous systemic sclerosis (ISSc) samples and 4 AIH/PBC overlap samples).
  • Data Provenance: Not explicitly stated regarding country of origin. The samples were "characterized" and included various patient groups (e.g., Autoimmune Hepatitis, Primary Sclerosing Cholangitis, Liver Cancer, Celiac Disease, Hepatitis B/C, Syphilis, Ulcerative Colitis, Crohn's Disease, Alcoholic liver disease, Idiopathic inflammatory myopathies, Systemic lupus erythematosus, Sjögren's syndrome, Sicca syndrome, Type 1 Diabetes, Osteoporosis, Chronic fatigue, Skin conditions, Drug-induced hepatotoxicity, Hypothyroidism, PBC). This suggests the data is retrospective, from existing sample banks of patients with known diagnoses.

3. Number of Experts and Qualifications for Ground Truth (Test Set)

  • The document states that the "cohort of characterized samples" were used for clinical validation. It also mentions "three diagnosed PBC patient specimens were assayed to aid in the determination of the cut-off."
  • However, the document does not explicitly state the number of experts used to establish the ground truth or their specific qualifications (e.g., radiologist with 10 years of experience) for the clinical validation set. The characterization implies that the samples had a pre-established clinical diagnosis, likely by medical professionals (e.g., physicians, pathologists), but the details of this process are not provided.

4. Adjudication Method (Test Set)

  • The document does not describe an adjudication method (e.g., 2+1, 3+1, none) for the clinical validation test set. The diagnoses for the "characterized samples" are presented as given.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done. This is an in vitro diagnostic device for semi-quantitative determination of antibodies, not an imaging device typically evaluated with MRMC studies involving human readers.

6. Standalone (Algorithm Only) Performance Study

  • Yes, a standalone study was done. The entire clinical validation and analytical performance sections evaluate the device (QUANTA Flash M2 (MIT3)), i.e., the algorithm/instrument system, without human interpretation beyond the initial sample collection and result reading from the instrument. The results presented for sensitivity, specificity, precision, linearity, etc., are all measures of the device's standalone performance.

7. Type of Ground Truth Used

  • The ground truth for the clinical validation study was based on clinical diagnoses of "characterized samples" from various patient groups, including confirmed Primary Biliary Cholangitis (PBC) patients and control groups for specificity. It also mentions "diagnosed PBC patient specimens" for cut-off determination. This falls under a combination of clinical diagnosis and presumed outcomes data (i.e., whether the patient actually had the disease).

8. Sample Size for Training Set

  • The document mentions a "reference population for establishing the reference interval for the M2 (MIT3) assay consisted of 180 subjects" for establishing the cut-off. This could be considered a "development set" or part of fine-tuning the algorithm's decision threshold (cut-off).
  • It also states "three diagnosed PBC patient specimens were assayed to aid in the determination of the cut-off."
  • Beyond this, the document does not explicitly describe a separate, distinct "training set" for the core algorithm in the way a machine learning model might. The analytical characterization and master curve development are part of the intrinsic design and calibration of the assay.

9. How Ground Truth for Training Set Was Established

  • For the reference interval/cut-off establishment:
    • The 180 subjects included: 100 apparently healthy blood donors, 30 infectious disease patients, 30 rheumatoid arthritis patients, and 20 celiac disease patients.
    • The cut-off was established in accordance with CLSI EP28-A3c: Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline - Third Edition.
    • The Analyse-it for Excel software was used for calculations.
    • The non-parametric percentile method was used due to non-normal distribution of results.
    • Three diagnosed PBC patient specimens were also assayed to aid in determining the cut-off, ensuring optimal differentiation.
    • The "Master Curve" for quantitation is generated during manufacturing using 7 "Master Curve Standards" with assigned CU values. These assigned values serve as the ground truth for calibrating the system's quantitative output.

{0}------------------------------------------------

Public Health Service

Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling a stylized caduceus or a series of human profiles facing right, suggesting health and human services.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 5, 2017

Inova Diagnostics, Inc. Roger Albesa Supervisor, Research and Development 9900 Old Grove Road San Diego. California 92131-1638

Re: K163525

Trade/Device Name: OUANTA Flash M2 (MIT3), OUANTA Flash M2 (MIT3) Calibrators, QUANTA Flash M2 (MIT3) Controls Regulation Number: 21 CFR 866.5090 Regulation Name: Antimitochondrial antibody immunological test system Regulatory Class: Class II Product Code: DBM, JIT, JJX Dated: December 14, 2016 Received: December 15, 2016

Dear Roger Albesa:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR

{1}------------------------------------------------

Part 807): labeling (21 CFR Part 801 and Part 809): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and Part 809), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Kelly Oliner -S

For Lea Carrington, M.S., MBA, MT Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K163525

Device Name

QUANTA Flash M2 (MIT3), QUANTA Flash M2 (MIT3) Calibrators, QUANTA Flash M2 (MIT3) Controls

Indications for Use (Describe)

QUANTA Flash M2 (MIT3) is a chemiluminescent immunoassay for the semi-quantitative determination of IgG antimitochondrial antibodies in human serum. The presence of antibodies, in conjunction with clinical findings and other laboratory tests, is an aid in the diagnosis of primary biliary cholangitis.

QUANTA Flash M2 (MIT3) Calibrators are intended for use with the QUANTA Flash M2 (MIT3) chemiluminescent immunoassay for the determination of IgG anti-mitochondrial antibodies in human serum. Each calibrator establishes a point of reference for the working curve that is used to calculate unit values.

QUANTA Flash M2 (MIT3) Controls are intended for use with the QUANTA Flash M2 (MIT3) chemiluminescent immunoassay for quality control in the determination of IgG anti-mitochondrial antibodies in human serum.

Type of Use (Select one or both, as applicable):

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains the logo for Inova Diagnostics, a Werfen Company. The logo consists of a stylized graphic with two overlapping shapes in red and blue, followed by the text "Inova" and "Diagnostics" stacked vertically. Below the company name, the text "A Werfen Company" is displayed in a smaller font size.

510(k) Summary

QUANTA Flash® M2 (MIT3) QUANTA Flash® M2 (MIT3) Calibrators QUANTA Flash® M2 (MIT3) Controls

Table of Contents

Administrative data……………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………
Predicate device
Device description
Intended use(s)
Indications for use……………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………
Substantial equivalence
Comparison to predicate device
Analytical performance characteristics
Quantitation and units of measure
Value assignment and traceability of Calibrators and Controls
Precision
Reproducibility
Lot to lot comparison
Limit of Blank (LoB), Limit of Detection (LoD), and Limit of Quantitation (LoQ)
Analytical Measuring Range (AMR)
Auto-rerun function and reportable results
High concentration hook effect
Linearity
Interference
Cross-reactivity
Sample Stability and Handling
Reagent Stability
Cut-off, reference range
Clinical performance characteristics
Expected values22
Comparison with predicate device22

{4}------------------------------------------------

510(k) Summary QUANTA Flash® M2 (MIT3) Page 2 of 23

{5}------------------------------------------------

This summary of the 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

Administrative data

Submitter:Inova Diagnostics, Inc.9900 Old Grove Road,San Diego, CA, 92131
Purpose of submission:New device(s)
Devices in the submission:QUANTA Flash® M2 (MIT3)QUANTA Flash® M2 (MIT3) CalibratorsQUANTA Flash® M2 (MIT3) Controls
Scientific contact:Roger Albesa, Supervisor, Research and DevelopmentInova Diagnostics, Inc.9900 Old Grove Road, San Diego, CA, 92131Phone: 858-586-9900 x 1391Fax: 858-863-0025email: ralbesa@inovadx.com
Quality Systems contact:Ronda Elliott, VP, Quality Systems and RAInova Diagnostics, Inc.9900 Old Grove Road, San Diego, CA, 92131Phone: 858-586-9900 x 1381Fax: 858-863-0025email: relliot@inovadx.com
Device name (assay kit):Proprietary name: QUANTA Flash® M2 (MIT3)Common name: Anti-M2 (MIT3) Chemiluminescent ImmunoassayClassification name: antibodies, anti-mitochondrial
Regulation Description:Anti-mitochondrial antibody immunological test system
Regulation Medical Specialty:Immunology
Review Panel:Immunology
Product Code:DBM
Regulation Number:866.5090
Device Class:2
Device name (Calibrators):Proprietary name:Common name:Classification name:QUANTA Flash® M2 (MIT3) CalibratorsM2 (MIT3) CalibratorsCalibrator, secondary
Regulation Description:Calibrator
Regulation Medical Specialty:Clinical Chemistry
Product Code:JIT
Regulation Number:862.1150
Device Class:2
Device name (Controls):Proprietary name:Common name:Classification name:QUANTA Flash® M2 (MIT3) ControlsM2 (MIT3) Controlssingle (specified) analyte controls (assayed andunassayed)
Regulation Description:Quality control material (assayed and unassayed)
Regulation Medical Specialty:Clinical Chemistry
Product Code:JJX
Regulation Number:862.1660
Device Class:1 (reserved)

{6}------------------------------------------------

510(k) Summary QUANTA Flash® M2 (MIT3)

Predicate device

QUANTA Lite® M2 EP (MIT3) ELISA, 510(k) number: K052262

{7}------------------------------------------------

Device description

The QUANTA Flash M2 (MIT3) assay is designed to run on the BIO-FLASH® instrument. This platform is a fully automated closed system with continuous load and random access capabilities that automatically processes the samples, runs the assay and reports the results. It includes liquid handling hardware, luminometer and computer with software-user interface. The QUANTA Flash M2 (MIT3) assay utilizes a reagent cartridge format, which is compatible with the BIO-FLASH instrument.

Recombinant MIT3 antigen is coated onto paramagnetic beads. The bead suspension is lyophilized and stored in the bead tube. Prior to use in the BIO-FLASH system, the sealed reagent tubes are pierced with the reagent cartridge lid and the beads are rehydrated and resuspension buffer by pipetting up and down with a transfer pipette. The reagent cartridge is then loaded onto the BIO-FLASH instrument. Samples are also loaded onto the instrument in samples are diluted by the BIO-FLASH with system rinse in a small disposable plastic cuvette. Small amounts of the diluted patient serum, the beads, and assay buffer are combined into a second cuvette, and mixed. This cuvette is then incubated at 37°C. The beads are magnetized and washed several times. Isoluminol conjugated anti-human IgG antibodies are then added to the cuvette, and again incubated at 37°C. The beads are magnetized and washed repeatedly. The isoluminol conjugate is oxidized when Trigger 1 (Fe(II)coproporphyrin in sodium hydroxide solution) and Trigger 2 (urea-hydrogen peroxide in sodium chloride solution) are added to the cuvette, and the flash of light produced from this reaction is measured as Relative Light Units (RLU) by the BIO-FLASH optical system. The RLU are proportional to the amount of isoluminol conjugate that is bound to the human IgG, which is in turn proportional to the amount of anti-M2 (MIT3) antibodies bound to the corresponding beads.

For quantitation, the QUANTA Flash M2 (MIT3) assay utilizes a predefined lot specific Master Curve that is uploaded onto the instrument through the reagent cartridge barcode. Every new lot number of reagent cartridge must be calibrated before first use, with the QUANTA Flash M2 (MIT3) Calibrators. Based on the results obtained with the three Calibrators included in the Calibrator Set (sold separately), an instrument specific Working Curve is created, which is used to calculate chemiluminescent units (CU) from the instrument signal (RLU) obtained for each sample.

The QUANTA Flash M2 (MIT3) kit contains the following materials:

One (1) QUANTA Flash M2 (MIT3) Reagent Cartridge One (1) vial of Resuspension buffer One (1) Transfer pipette

{8}------------------------------------------------

The QUANTA Flash M2 (MIT3) reagent cartridge contains the following reagents for 50 determinations:

  • M2 (MIT3) coated paramagnetic beads, lyophilized. a.
  • b. Assay buffer - colored pink, containing Tris-buffered saline, Tween 20, protein stabilizers and preservatives.
  • Tracer IgG Isoluminol labeled anti-human IgG antibodies in buffer, containing protein C. stabilizers and preservative.

The QUANTA Flash M2 (MIT3) Calibrators kit contains two vials each of Calibrator 2, and Calibrator 3:

QUANTA Flash M2 (MIT3) Calibrators:

  • QUANTA Flash M2 (MIT3) Calibrator 1: Two (2) barcode labeled tubes containing 0.3 mL prediluted, ready to use reagent. Calibrators contain human anti-mitochondrial antibodies in stabilizer and preservative.
  • -QUANTA Flash M2 (MIT3) Calibrator 2: Two (2) barcode labeled tubes containing 0.3 mL prediluted, ready to use reagent. Calibrators contain human anti-mitochondrial antibodies in stabilizer and preservative.
  • -QUANTA Flash M2 (MIT3) Calibrator 3: Two (2) barcode labeled tubes containing 0.3 mL prediluted, ready to use reagent. Calibrators contain human anti-mitochondrial antibodies in stabilizer and preservative.

The QUANTA Flash M2 (MIT3) Controls kit contains two vials of Negative Control and two vials of Positive Control:

QUANTA Flash M2 (MIT3) Controls:

  • QUANTA Flash M2 (MIT3) Negative Control: Two (2) barcode labeled tubes containing 0.5 ၊ mL, ready to use reagent. Controls contain human anti-mitochondrial antibodies in stabilizer and preservative.
  • । QUANTA Flash M2 (MIT3) Positive Control: Two (2) barcode labeled tubes containing 0.5 mL, ready to use reagent. Controls contain human anti-mitochondrial antibodies in stabilizer and preservative.

{9}------------------------------------------------

Intended use(s)

QUANTA Flash M2 (MIT3) is a chemiluminescent immunoassay for the semi-quantitative determination of IgG anti-mitochondrial antibodies in human serum. The presence of anti-mitochondrial antibodies, in conjunction with clinical findings and other laboratory tests, is an aid in the diagnosis of primary biliary cholangitis.

QUANTA Flash M2 (MIT3) Calibrators are intended for use with the QUANTA Flash M2 (MIT3) chemiluminescent immunoassay for the determination of IgG anti-mitochondrial antibodies in human serum. Each calibrator establishes a point of reference for the working curve that is used to calculate unit values.

QUANTA Flash M2 (MIT3) Controls are intended for use with the QUANTA Flash M2 (MIT3) chemiluminescent immunoassay for quality control in the determination of IgG anti-mitochondrial antibodies in human serum.

Indications for use

Same as Intended use.

Substantial equivalence

The QUANTA Flash M2 (MIT3) Reagent, the QUANTA Flash M2 (MIT3) Calibrators and the QUANTA Flash M2 (MIT3) Controls have the same intended use and assay principle as the predicate device.

Comparison to predicate device

Similarities
ItemQUANTA Flash M2 (MIT3)Predicate Device
Intended useQUANTA Flash M2 (MIT3) is achemiluminescent immunoassay forthe semi-quantitative determinationof IgG anti-mitochondrial antibodiesin human serum. The presence ofanti-mitochondrial antibodies, inconjunction with clinical findings andother laboratory tests, is an aid in thediagnosis of primary biliarycholangitis.QUANTA Lite M2 EP (MIT3) ELISA is anenzyme-linked immunosorbent assay(ELISA) for the semi-quantitativedetection of mitochondria antibodiesin human serum. The presence ofmitochondria antibodies can be usedin conjunction with clinical findingsand other laboratory tests to aid in thediagnosis of primary biliary cirrhosis.
Assay methodologySolid phase (heterogeneous)immunoassaySolid phase (heterogeneous)immunoassay
Antigenrecombinantrecombinant
Sample typeSerumSerum
Shelf lifeOne yearOne year

QUANTA Flash M2 (MIT3) reagent kit

{10}------------------------------------------------

Differences
ItemQUANTA Flash M2 (MIT3)Predicate Device
Detection/Operating principleChemiluminescent immunoassayEnzyme-linked immunosorbent assay
Solid phaseParamagnetic microparticles (beads)96-well polystyrene plate
ConjugateIsoluminol conjugated anti-humanIgGHRP conjugated anti-human IgG
CalibrationLot specific Master Curve + threecalibrators (sold separately)M2 EP (MIT3) ELISA Low Positive(single calibrator) - (Included in the kit)

QUANTA Flash M2 (MIT3) Calibrators

ItemQUANTA Flash M2 (MIT3) CalibratorsPredicate Device
Intended useQUANTA Flash M2 (MIT3) Calibrators are intended for use with the QUANTA Flash M2 (MIT3) chemiluminescent immunoassay for the determination of IgG anti-mitochondrial antibodies in human serum. Each calibrator establishes a point of reference for the working curve that is used to calculate unit values.No separate intended use; calibrator is part of the kit.
AnalyteAnti-mitochondrial antibodiesAnti-mitochondrial antibodies
MethodQUANTA Flash M2 (MIT3) chemiluminescent immunoassayQUANTA Lite M2 EP (MIT3) ELISA
MatrixHuman serum, stabilizer, and preservativeHuman serum, buffer, protein stabilizer, and preservative
UnitCU (Chemiluminescent units) (arbitrary)units (arbitrary)
Physico-chemical characteristicsLiquid, prediluted, ready to useLiquid, prediluted, ready to use
Storage2-8 °C2-8 °C
Shelf lifeOne yearOne year

{11}------------------------------------------------

ItemQUANTA Flash M2 (MIT3) ControlsPredicate Device
Intended useQUANTA Flash M2 (MIT3) Controlsare intended for use with theQUANTA Flash M2 (MIT3)chemiluminescent immunoassay forquality control in the determinationof IgG anti-mitochondrial antibodiesin human serum.No separate intended use; controlsare part of the kit.
AnalyteAnti-mitochondrial antibodiesAnti-mitochondrial antibodies
MethodQUANTA Flash M2 (MIT3)chemiluminescent immunoassayQUANTA Lite M2 EP (MIT3) ELISA
MatrixHuman serum, stabilizers, andpreservativeHuman serum, buffer, proteinstabilizer, and preservative
UnitCU (Chemiluminescent units)(arbitrary)units (arbitrary)
Physico-chemicalcharacteristicsLiquid, ready to useLiquid, prediluted, ready to use
Levels2 (negative and positive)2 (ELISA negative and High positive)
Storage2-8 °C2-8 °C
Shelf lifeOne yearOne year

QUANTA Flash M2 (MIT3) Controls

Analytical performance characteristics

Quantitation and units of measure

For quantitation, the QUANTA Flash M2 (MIT3) assay utilizes a lot specific Master Curve that is uploaded onto the instrument through the reagent cartridge barcode. The Master Curve for QUANTA Flash M2 (MIT3) consists of 7 Standards. These Master Curve Standards are used to create the lot specific Master Curve during the manufacturing procedure.

List of M2 (MIT3) Standards:

MaterialAssigned Value
M2 (MIT3) Master Curve Standard 10.0 CU
M2 (MIT3) Master Curve Standard 26.7 CU
M2 (MIT3) Master Curve Standard 327.1 CU
M2 (MIT3) Master Curve Standard 4108.6 CU
M2 (MIT3) Master Curve Standard 5434.3 CU
M2 (MIT3) Master Curve Standard 61737.3 CU
M2 (MIT3) Master Curve Standard 73474.6 CU

{12}------------------------------------------------

Value assignment and traceability of Calibrators and Controls

The QUANTA Flash M2 (MIT3) Calibrators and Controls are manufactured by diluting human serum that contains high titer of anti-mitochondrial antibodies with stabilizer and preservative. The human serum is obtained from commercial sources and it is tested for markers of infectious substances.

The target CU is achieved through trial dilutions on small scale. Once a dilution is selected, the Calibrators and Control are manufactured to final scale, tested, and adjusted. Upon completion of the manufacturing process, the Calibrators and Controls are tested on at least two instruments, on at least two lots of reagent cartridge, in replicates of 10 to determine final value assignment.

Calibrator and Control values are directly traceable to the in-house Standards that are used to create the Master Curves for the QUANTA Flash M2 (MIT3) assay.

MaterialManufacturingTarget ValueManufacturingTarget Range
M2 (MIT3) Calibrator 112 CU10 - 14 CU
M2 (MIT3) Calibrator 2400 CU360 - 440 CU
M2 (MIT3) Calibrator 32550 CU2450 - 2650 CU
M2 (MIT3) Negative Control10 CU8 - 12 CU
M2 (MIT3) Positive Control50 CU40 - 60 CU

M2 (MIT3) Calibrators and Controls with target manufacturing values:

Precision

The precision of the QUANTA Flash M2 (MIT3) assay was evaluated on 5 samples, along with negative and positive controls, containing various concentrations of anti-mitochondrial antibodies in accordance with CLSI EP05-A3, Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline. Samples were run in duplicates, twice a day, for 20 days.

Data were analyzed with the Analyse-it for Excel method evaluation software, and within run, between run, between day and total precision were calculated.

Acceptance criteria: Total %CV: < 10%

Results are summarized in the Table below.

QUANTA Flash M2 (MIT3)Within RunBetween RunsBetween DaysTotal
Sample IDNumber ofreplicatesMean(CU)SD(CU)CV(%)SD(CU)CV(%)SD(CU)CV(%)SD(CU)CV(%)
N808.50.44.20.22.40.00.30.44.8
P8045.41.22.71.02.20.61.31.73.7
18021.20.73.10.52.30.52.31.04.5
28018.30.52.60.21.10.20.80.52.9
380277.37.42.75.31.90.00.09.13.3
4801007.539.33.925.32.515.01.549.14.9
5802183.7119.65.578.73.685.23.9166.67.6

{13}------------------------------------------------

Reproducibility

Reproducibility between sites (instruments)

Nine samples were tested according to CLSI EP05-A3 Evaluation of Quantitative Measurement Procedures, at three different sites. Samples were run in replicates of 5, once a day, for 5 days, to generate 25 data points per site. Data were analyzed with the Analyse-it for Excel method evaluation software to calculate between site precision.

Acceptance criteria: Total %CV: < 10%

Results are summarized in the Table below.

QUANTA Flash M2 (MIT3)Between Site Precision (Reproducibility)
Sample IDNumber of replicatesMean (CU)SD (CU)CV (%)
1759.30.89.1
27511.10.43.4
37522.91.46.2
47530.51.13.5
57553.72.85.2
675404.331.47.8
775617.850.48.2
875939.459.16.3
9752182.744.92.1

Lot to lot comparison

Reproducibility between lots

Lot to lot reproducibility study was performed according to CLSI EP05-A3 Evaluation of Precision of Quantitative Measurement Procedures, by testing six samples with three different lots of reagents in five replicates for 5 days, to generate 25 data points per lot, 75 data points total for each sample.

Data were analyzed with the Analyse-it for Excel method evaluation software, between lots imprecision was calculated, and the results are summarized in the Table below. All %CV values were within the acceptance limit, 10%.

Results are summarized in the Table below.

QUANTA Flash M2 (MIT3)Between Lot Precision (Reproducibility)
Sample IDNMean (CU)SD (CU)CV (%)
17519.70.94.7
27518.91.26.4
375869.437.94.4
4752186.7191.08.7
5758.30.56.5
67542.41.74.0

{14}------------------------------------------------

Limit of Blank (LoB), Limit of Detection (LoD), and Limit of Quantitation (LoQ)

The LoD of the QUANTA Flash M2 (MIT3) assay is 0.61 CU, which is below the analytical measuring range (AMR) of the assay. It was determined by using two reagent lots, consistent with CLSI EP17-A2 guideline with proportions of false positives (alpha) less than 5% and false negatives (beta) less than 5%; based on 240 determinations; 60 measurements on blank samples and 60 measurements of low level samples with 2 reagent lots. The LoB is 0.43 CU.

The data generated from the LoD testing was used to calculate the LoQ for the QUANTA Flash M2 (MIT3) assay. The LoQ was determined by calculating the total error (TE) of each sample per reagent lot using the Westgard model: TE = |bias|+ 2S, where S= standard deviation.

The TE was then compared to the predefined accuracy goal (<25%). The LoQ for the assay is 1.2 CU, which has been set as the lower limit of the AMR.

Analytical Measuring Range (AMR)

QUANTA Flash M2 (MIT3): 1.2 CU - 3000.0 CU

Auto-rerun function and reportable results

The BIO-FLASH software has an Auto-rerun option available. If this option is selected, the instrument will automatically rerun any sample that has a result of >3000.0 CU after further diluting it by 20 fold, thereby bringing the measured value within the AMR. The final result will be calculated by the software by taking into account the additional dilution factor. As the highest value that can be directly measured is 3000.0 CU, the highest value that can be reported is 60000.0 CU.

{15}------------------------------------------------

High concentration hook effect

To assess hook effect, measurement signal (RLU) was examined by performing serial dilutions of two high positive samples (with results above the AMR when tested as neat samples). RLU values showed increase with increasing analyte concentrations, that high positive specimens above the AMR do not show hook effect up to 256,113.1 CU (theoretical value calculated using the highest value in the Working Curve and its dilution factor) in the QUANTA Flash M2 (MIT3) assay. It is worth to remark that when a sample is run using the auto-rerun function it is diluted 1:20 prior to be in contact with the bead or the tracer, so at that point it is not different from another sample yielding results in the AMR, except that at the end the result is multiplied by 20 before reported.

Sample 1Sample 2
RLUCalculated ResultRLUCalculated Result
DilutionAverageCUAverageCU
1:1799996256113.1732727147096.3
1:2715723128056.671097873548.1
1:461892364028.369965636774.1
1:85117524622.267779618387.0
1:163840641811.46315289193.5
1:32244576797.75111804596.8
1:64152935429.03942281928.0
1:12879728208.5251350831.0
1:25640556105.8149460417.4
1:5121999553.694047248.1
1:1024974327.144318115.4
1:2048515214.62417364.3
1:409626187.41096130.3

Linearity

The linearity of the AMR was evaluated by a study according to CLSI EPG-A, Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline. Six serum samples with various mitochondrial antibody concentrations were diluted with negative serum in 10% increments (from 0% to 90% negative serum) to obtain values that cover the AMR. The dilutions were assayed in duplicates. Results were analyzed according to the guideline performing regression analysis and identifying the best fitting polynomial.

{16}------------------------------------------------

Acceptance criteria:

  • Best fitting polynomial is a linear one, otherwise, the difference between the best-fitting nonlinear and linear polynomial is less than 20%, or ± 4 CU, whichever is greater (allowable nonlinearity). All six specimens fulfilled the acceptance criteria, showing good linearity in their respective ranges.
SampleTest Range (CU)Slope(95% CI)Y-Intercept(95% CI)R2Average % Recovery
13373.5 to 481.91.08(0.92 to 1.24)-146.8(-486.0 to 192.5)0.9898.5%
21496.5 to 149.71.01(0.96 to 1.06)26.0(-21.7 to 73.7)1.00105.2%
3281.5 to 28.11.03(1.00 to 1.06)-1.6(-7.2 to 4.0)1.00100.8%
452.5 to 5.31.00(0.98 to 1.03)0.4(-0.5 to 1.2)1.00102.0%
522.7 to 2.30.98(0.94 to 1.02)1.0(0.4 to 1.6)1.00109.2%
611.8 to 1.20.96(0.92 to 1.00)0.1(-0.2 to 0.4)1.00100.3%
All samples3373.5 to 1.21.02(1.00 to 1.04)-2.2(-18.7 to 14.4)1.00102.9%

These data demonstrate the linearity of the analytical measuring range of the QUANTA Flash M2 (MIT3) assay.

Interference

The interference study was performed according to CLSI EPO7-A2, Interference Testing in Clinical Chemistry; Approved Guideline - Second Edition. Five specimens, one high positive (1422.5 CU), one moderately positive (392.4 CU), two near the cutoff (13.9 and 22.1 CU), and one negative (9.9 CU) sample were tested. Interfering substances (hemoglobin, bilirubin, triglycerides, cholesterol, and human IgG) were spiked into every specimen at three different concentrations in 10% of total specimen volume, and the resulting samples were assessed in triplicates with the QUANTA Flash M2 (MIT3) assay. Additionally, four samples, one negative (11.8 CU), two near the cutoff (18.3 and 23.8 CU) and one positive (195.8 CU) samples were tested for interference with Ursodeoxycholic acid, corticosteroids (Prednisone), methotrexate, cholestipol, hydroxyzine, Aminotransferases (Glutamate Oxacatate Transaminase) and HDL cholesterol (high density lipoprotein). Moreover, 6 additional samples were tested for RF interference by combining them with different concentrations of a high positive RF IgM serum sample (1534 IU/mL). Recovery of the unit values was calculated compared to control samples spiked with the same volume of diluents (10% of total sample volume). Acceptance criteria for the interference studies were 85% - 115% recovery, or ± 4 CU difference, whichever is greater.

No interference was detected with bilirubin up to 1 mg/mL (recovery: 93% to 100%), hemoglobin up to 2 mg/mL (recovery: 94% to 101%), triglycerides up to 1000 mg/dL (recovery: 94% to 102%), cholesterol up to 332.5 mg/dL (recovery: 97% to 104%), human IgG up to 35 mg/mL (recovery: 91% to 115%, or 1.8 CU to 3.7 CU), Ursodeoxycholic acid up to 0.75 mg/mL (recovery: 97% to 104%), Prednisone up to 0.3

{17}------------------------------------------------

510(k) Summary QUANTA Flash® M2 (MIT3)

mg/mL (recovery: 100% to 105%), methotrexate up to 9.1 mg/mL (recovery: 96% to 103%), cholestyramine up to 18 mg/mL (recovery: 97% to 100%), colestipol up to 18 mg/mL (recovery: 98% to 104%), Hydroxyzine up to 1 mg/mL (recovery 100% to 108%), Glutamate Oxacatate Transaminase up to 0.12 IU/mL (recovery: 98% to 115%), HDL Cholesterol up to 3.5 mg/mL (recovery: 98% to 105%) and rheumatoid factor up to 153.4 IU/mL (recovery: 96% to 111%).

{18}------------------------------------------------

Cross-reactivity

To test potential cross-reactivity with autoantibodies and infection-induced antibodies, results obtained on all 417 control samples plus 14 ISSc samples that were included in the clinical validation study were assessed. These samples were from patients with autoimmune diseases that are characterized with disease specific autoantibodies, or from patients with positive infectious disease serology. The composition of the cohort and the anti-mitochondrial positivity rate is shown in the Table below:

Patient groupNNumber positive% positive
Autoimmune Hepatitis type 1 (AIH1)4124.9%
LKM-1 antibody positive AutoimmuneHepatitis 2 (AIH2)2800.0%
Primary Sclerosing Cholangitis (PSC)2100.0%
Celiac Disease1900.0%
Hepatitis B virus (HBV)1000.0%
Hepatitis C virus (HCV)2500.0%
Syphilis1000.0%
Ulcerative Colitis2600.0%
Crohn's Disease1000.0%
Liver Cancer (LC)1000.0%
Alcoholic liver disease (ALD)2000.0%
Idiopathic inflammatory myopathies (IIM)800.0%
Systemic lupus erythematosus (SLE)800.0%
Sjögren's syndrome (SS)400.0%
Sicca syndrome2000.0%
Type 1 Diabetes3000.0%
Osteoporosis2813.6%
Chronic fatigue3013.3%
Skin3000.0%
Drug-induced hepatotoxicity900.0%
Hypothyroidism3000.0%
Limited cutaneous systemic sclerosis(ISSc)*14535.7%
Total43192.1%
  • Literature indicates a relatively high prevalence of PBC in SSc patients, as compared to the general population.

Based on the results, the QUANTA Flash M2 (MIT3) assay does not show cross-reactivity with autoantibodies that are present in various autoimmune diseases, or with antibodies against infectious agents.

{19}------------------------------------------------

Sample Stability and Handling

Four samples, encompassing negative, around the cut-off, and positive samples were tested in duplicates for up to 21 days while stored at 2-8°C, up to 48 hours while stored at room temperature, and after repeated freeze/thaw cycles up to 3 cycles. Results were compared to those obtained on control samples (day zero, stored at 2-8°C)

Acceptance criteria: 90-110% average recovery.

All samples fulfilled the acceptance criteria at each time point for each condition. Based on these results, we recommend that samples are stored up to 48 hours at room temperature, up to 14 days at 2-8°C, and can be subjected to up to 3 freeze/thaw cycles (when samples are stored at or below -20°C).

Reagent Stability

Shelf life

To establish the initial claim for shelf life, accelerated stability studies were performed for up to 4 weeks at 37°C ± 3°C, where one week is equal to six months at 5 ± 3°C.

Accelerated stability testing was performed on each of the following sealed components of the QUANTA Flash M2 (MIT3) to establish initial stabliity claim: the beads, the two Calibrators, and the Negative and Positive Controls. Each week a new sealed component was placed in the incubator, and all components were tested at the end of the experiment together with the one that was stored at 5 ± 3℃. The recovery of the measured values was calculated for each time point (compared to those obtained with 5 ± 3°C stored reagent). All calculations were performed by comparing results of sealed components stored at 5 ± 3℃ (control) to those stored at 37 ± 3℃ (test) for 1, 2, 3, and (optional) 4 weeks, where one week is equal to six months at 5 ± 3℃. Linear regression analysis was performed between recovery values and the number of days.

Acceptance criteria for one year preliminary expiration dating:

- Beads:

With regression analysis, the lower 95% Cl interval of the regression line is ≥85% and the upper 95% Cl interval is at ≤115% at day 14, and no individual data point has ≤75% or ≥125% recovery at day 14.

- Controls and Calibrators:

With regression analysis, the lower 95% Cl interval of the regression line is ≥90% and the upper 95% Cl interval is ≤110% at day 14, and no individual data point has ≤80% or ≥120% recovery at day 14.

Beads

Testing was performed on three lots of M2 (MIT3) coupled beads using up to 9 characterized samples with various reactivity levels.

All three lots of beads retained between 85% and 115% reactivity (considering the 95% Cl) after two weeks at 37 ± 3℃, and therefore pass the acceptance criteria for one year expiration date.

{20}------------------------------------------------

Calibrators and Controls

Testing was performed on three lots of M2 (MIT3) Calibrators and Controls. All Calibrators and Controls maintained between 90% and 110% reactivity (considering the 95% Cl) when stored at 37 ± 3°C for 2 weeks, and therefore pass the acceptance criteria for one year expiration dating.

In-use (onboard) stability

Calibrators

Onboard stability claim: 4 calibrations, or 8 hours onboard

During assessing on-board stability, Calibrators were placed uncapped, onboard the instrument, and calibration was performed altogether five times over 9 hours. Controls and a panel of characterized patient specimens were run on each calibration curve.

Calibrators are considered stable if all four calibrations performed in the 8 hour period are successful, mean Calibrator RLU recovery values for the first 4 calibrations are between 90% and 110% compared to the first use, and Control/patient panel CU recovery values are between 85% and 115% of those obtained on the first calibration curve.

The first four calibrations performed in the 8 hour period were considered valid by the software. The calibrators vielded average RLU recovery values ranging from 100% to 110%. The Control/patient panel CU recovery ranged from 87.8% to 96.6%. This supports the claim that calibrators can be used for up to 4 calibrations over an 8 hour period.

Controls

Onboard stability claim: up to 15 uses, at 10 minutes onboard per use

During assessing on-board stability, 2 vials of each Control were assayed once a day for a total of 20 runs. The first run was used to establish baseline value, by running each vial in duplicate, and then additional 19 runs were performed, by running each vial in singleton. During runs, the Controls were left uncapped, onboard the instrument for 15 minutes per run. When not in use, the controls were capped, and stored at 5 ± 3°C.

Percent recovery of each value was calculated compared to the baseline value. Controls are considered stable when all values run within their established range, and the linear regression line obtained by plotting %recovery values against the number of runs stays between 85% and 115% at run 15.

All controls ran within their respective acceptable ranges for all runs. Moreover, the regression line remained between 85% and 115% at run 15 for both Controls. These results support the claim that controls can be used for up to 15 times, at 10 minutes per use.

Reagent Cartridge

To establish the in-use stability of the QUANTA Flash M2 (MIT3) reagent cartridges were tested with up to 4 serum specimens (with different reactivity levels). The specimens were tested periodically up to 70 days. Percent recoveries were calculated compared to the day zero average values, and linear regression analysis was performed by plotting %recovery against the number of days. The claim was established using the following criteria (using the one that is fulfilled first):

{21}------------------------------------------------

Page 19 of 23

  • The stability claim is established at the actual measurement day proceeding the 95% confidence interval of the regression line reaches 85% or 115% recovery, or

  • At the actual measurement day preceding the day when 2 data points or ≥2% of the recovery data (whichever is greater) is ≤ 75% or ≥ 125% recovery.

The onboard stability results of the two lots are as follows:

RP0006: 70 days

151003: 60 days

Using these criteria, the in-use (onboard) stability of M2 (MIT3) reagent cartridge was set at 60 days.

Real time stability

Real time stability testing has been scheduled to be performed every three months on the reagent cartridge, Calibrators and Controls, to verify the one year expiration that was assigned based on accelerated stability studies. At the time of the submission, results were available up to 9 months for reagent cartridge and Controls, while 3 months data for Calibrators will be available in January 2017.

For reagent cartridge, negative and positive controls, along with one additional sample, were tested in replicates of 9 at 0 and 6 month time points, and replicates of 3 at 3 and 9 month time points. - Acceptance criteria: compared to the average baseline value, the percent recovery of each sample is between 80-120%. The % CV of replicates is <15% at each time point.

Calibrators will be tested in triplicates on a calibrated cartridge at each time point. Averages of the triplicates will be compared to the value that was assigned to the Calibrators at release. - Acceptance criteria: % recovery of the average of the triplicates is between 85% and 115%, and %CV of the triplicates is < 10%.

Controls were tested in triplicates on a calibrated cartridge at each time point. Individual values, and averages of the triplicates were compared to the values that were assigned to the Controls at release. - Acceptance criteria: results should fall within their acceptable ranges as were established at the release of the Controls.

All results to date were within the acceptance limits.

{22}------------------------------------------------

Cut-off, reference range

QUANTA Flash M2 (MIT3):

Negative<20 CU
Positive≥20 CU

The reference population for establishing the reference interval for the M2 (MIT3) assay consisted of 180 subjects:

Sample GroupN
Apparently healthy blood donors100
Infectious disease30
Rheumatoid arthritis30
Celiac disease20

All specimens were the same matrix (serum) as specified in the Intended Use. All specimens were unaltered. The cut-off was established in accordance to CLSI EP28-A3c: Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline - Third Edition. The Analyseit for Excel software was used to make the calculations. The distribution of the results was non-normal (Saphiro-Wilk p<0.0001), so the non-parametric percentile method was used. One high positive sample was excluded from the calculations as an outlier. This sample was confirmed to be true positive with the predicate device. The 99th percentile of the remaining obtained values was calculated as 3835 RLU.

Additionally, three diagnosed PBC patient specimens were assayed to aid in the determination of the cut-off. Based on the distribution of RLU values in these (known) positive samples, the cutoff was increased to 6000 RLU to ensure optimal differentiation between negatives and a 20 CU value was assigned to this RLU value. No reference patients tested positive at this cutoff level.

{23}------------------------------------------------

Clinical performance characteristics

Clinical sensitivity, specificity

A cohort of characterized samples, none of which were used for establishing the reference range, was used to validate the clinical performance of the QUANTA Flash M2 (MIT3). A total of 586 characterized samples were included in the Validation Set for the QUANTA Flash M2 (MIT3). All samples were run on the QUANTA Flash M2 (MIT3). The distribution of the cohort and the M2 (MIT3) positivity rate is in the Table below:

Patient groupNNumberpositive% positive
Autoimmune Hepatitis type 1 (AIH1)4124.9%
LKM1 antibody positive Autoimmune Hepatitis 2 (AIH2)2800.0%
Primary Sclerosing Cholangitis (PSC)2100.0%
Liver Cancer (LC)1000.0%
Celiac Disease1900.0%
Hepatitis B virus (HBV)1000.0%
Hepatitis C virus (HCV)2500.0%
Syphilis1000.0%
Ulcerative Colitis2600.0%
Crohn's Disease1000.0%
Alcoholic liver disease (ALD)2000.0%
Idiopathic inflammatory myopathies (IIM)800.0%
Systemic lupus erythematosus (SLE)800.0%
Sjögren's syndrome (SS)400.0%
Sicca syndrome2000.0%
Type 1 Diabetes3000.0%
Osteoporosis2813.6%
Chronic fatigue3013.3%
Skin conditions3000.0%
Drug-induced hepatotoxicity900.0%
Hypothyroidism3000.0%
Controls used for Specificity Calculations41741.0%
Other: Limited cutaneous systemic sclerosis (ISSc)*14535.7%
Other: AIH/PBC overlap**44100.0%
PBC15112784.1%
Total586
  • Literature indicates a relatively high prevalence of PBC in SSc patients, as compared to the general population. ISSc samples were therefore excluded from sensitivity and specificity calculations

** Four samples with AIH/PBC overlap were excluded for all sensitivity and specificity calculations.

{24}------------------------------------------------

510(k) Summary QUANTA Flash® M2 (MIT3)

Clinical Analysis including allPBC (N=568)QUANTA Flash M2 (MIT3)Analysis(95% confidence)
PositiveNegativeTotal
DiagnosisPBC12724151Sensitivity: 84.1% (77.4 - 89.1%)
DiagnosisControls4413417Specificity: 99.0% (97.6 - 99.6%)
Total131437568

Clinical sensitivity and specificity of the QUANTA Flash M2 (MIT3) were analyzed in the table below:

Expected values

The expected value in the normal population is "negative". Anti-mitochondrial antibody levels were analyzed in a cohort of 100 apparently healthy blood donors (50 females and 50 males, ages 17 to 57 years, with an average and median age of 34 years) using the QUANTA Flash M2 (MIT3). This patient population was different from the one that was used to establish the cutoff, and was only used to assess expected values. With the cut-off of 20 CU, two samples were positive on the QUANTA Flash M2 (MIT3). The mean concentration was 2.8 CU, and the values ranged from <1.2 to 91.4 CU.

Comparison with predicate device

Samples for method comparison analysis include 409 samples from the clinical validation study, plus 8 additional PBC samples that yield results around the cutoff of the assay to increase to 11.5% of the samples within the AMR, for a total of 417 samples were tested on both the QUANTA Flash M2 (MIT3) and on the predicate ELISA. Negative percent agreement (NPA), positive percent agreement (PPA), and total percent agreement (TPA) were calculated. The data are presented in four ways; first using all samples (EIA equivocal presented as positive, then negative), and next using only the 174 samples within the AMR (EIA equivocal presented as positive, then negative).

Method Comparison, all samples:

Method Comparison –all samples (N=417)ELISA equivocal=posQUANTA Flash M2 (MIT3)Percent Agreement(95% Confidence)
NegativePositiveTotal
Predicate ELISANegative2507257NPA: 97.3% (94.5% – 98.7%)
Positive24136160PPA: 85.0% (78.7% – 89.7%)
Total274143417TPA: 92.6% (89.6% – 94.7%)

{25}------------------------------------------------

510(k) Summary QUANTA Flash® M2 (MIT3)

Method Comparison –all samples (N=417)ELISA equivocal=negQUANTA Flash M2 (MIT3)Percent Agreement(95% Confidence)
Predicate ELISANegativePositiveTotal
Negative25512267NPA: 95.5% (92.3% – 97.4%)
Positive19131150PPA: 87.3% (81.1% – 91.7%)
Total274143417TPA: 92.6% (89.6% – 94.7%)

Method Comparison, samples within the AMR:

Method Comparison -Within AMR (N=174)ELISA equivocal=posQUANTA Flash M2 (MIT3)Percent Agreement(95% Confidence)
NegativePositiveTotal
Predicate ELISANegative71778NPA: 91.0% (82.6% – 95.6%)
Positive138396PPA: 86.5% (78.2% – 91.9%)
Total8490174TPA: 88.5% (82.9% – 92.4%)
Method Comparison -QUANTA Flash M2 (MIT3)Percent Agreement
Within AMR (N=174)NegativePositiveTotal(95% Confidence)
ELISA equivocal=neg
Predicate ELISANegative731285NPA: 85.9% (76.9% – 91.7%)
Positive117889PPA: 87.6% (79.2% – 93.0%)
Total8490174TPA: 86.8% (80.9% – 91.0%)

§ 866.5090 Antimitochondrial antibody immunological test system.

(a)
Identification. An antimitochondrial antibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the antimitochondrial antibodies in human serum. The measurements aid in the diagnosis of diseases that produce a spectrum of autoantibodies (antibodies produced against the body's own tissue), such as primary biliary cirrhosis (degeneration of liver tissue) and chronic active hepatitis (inflammation of the liver).(b)
Classification. Class II (performance standards).