(265 days)
The Poseidon Ultrasound System is indicated for use by qualified and trained healthcare professionals to enable diagnostic ultrasound imaging and measurement of anatomical structures and fluids of adult and pediativ for the following clinical applications: Peripheral Vessel (including carotid and arterial studies), Procedural Guidance, Small Organs (including thyroid), Cardiac, Abdominal, Urology, Fetal/Obstetric, Gynecological, Musculoskeletal (conventional) and Musculoskeletal (superficial).
The Poseidon Ultrasound System is a general-purpose diagnostic imaging system for use by qualified and trained healthcare professionals enabling visualization and measurement of anatomical structures and fluid on adult and pediatric patients. The system consists of a single transducer with broad imaging capabilities connected via a USB cable to a standard handheld commercial off the shelf (COTS) mobile device. In addition to M-mode and B-mode imaging the instrument also supports Color Flow Doppler imaging.
The user interface includes touchscreen menus, buttons, controls, indicators and navigation icons that allow the operator to control the system and to view and measure ultrasound imagery.
The provided text is a 510(k) summary for the Butterfly Network Poseidon Ultrasound System. It describes the device, its intended use, and its equivalence to a predicate device. However, it does not contain information about acceptance criteria or a study that proves the device meets those criteria in the context of AI/ML performance.
This document focuses on the mechanical, electrical, and physical safety and performance of the ultrasound system itself, and its substantial equivalence to a previously cleared device (Philips Healthcare, Lumify Ultrasound System). It cites compliance with standards like IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, ISO 10993:2009, and NEMA UD-2.
There is no mention of an AI/ML component in the Poseidon Ultrasound System, nor any data or study design related to algorithmic performance as would be required for an AI/ML-driven medical device. Therefore, I cannot provide the requested information regarding acceptance criteria and a study proving an AI component meets those criteria.
The information you asked for (acceptance criteria for an AI component, sample sizes, expert qualifications, ground truth, etc.) is typically found in submissions for AI/ML-enabled devices, which this document does not appear to describe.
{0}------------------------------------------------
Public Health Service
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 6, 2017
Butterfly Network, Inc. % Mr. Brian Sawin Sr. Regulatory Affairs Manager 530 Old Whitfield Street GUILFORD CT 06437
Re: K163510
Trade/Device Name: Poseidon Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic Pulsed Doppler Imaging System Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: August 9, 2017 Received: August 10, 2017
Dear Mr. Sawin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Michael D. O'Hara
Robert A. Ochs, Ph.D Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K163510
Device Name Poseidon Ultrasound System
Indications for Use (Describe)
The Poseidon Ultrasound System is indicated for use by qualified and trained healthcare professionals to enable diagnostic ultrasound imaging and measurement of anatomical structures and fluids of adult and pediativ for the following clinical applications: Peripheral Vessel (including carotid and arterial studies), Procedural Guidance, Small Organs (including thyroid), Cardiac, Abdominal, Urology, Fetal/Obstetric, Gynecological, Musculoskeletal (conventional) and Musculoskeletal (superficial).
Type of Use (Select one or both, as applicable)
| ☑ Prescription Use (Part 21 CER 801 Subpart D) |
|---|
| ☐ Over-The-Counter Use (21 CER 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Diagnostic Ultrasound Indications For Use Table
System: Poseidon Ultrasound System
Transducer: Poseidon Ultrasound System transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | |||||||
|---|---|---|---|---|---|---|---|---|
| General (Track 1Only) | Specific (Tracks 1 & 3) | B | M | PWD | CWD | ColorDoppler | Combined(Specify) | Other*(Specify) |
| Ophthalmic | Ophthalmic | |||||||
| Fetal Imaging &Other | Fetal/Obstetric | N | N | N | B mode + Mmode | |||
| Abdominal | N | N | N | B mode + Mmode | ||||
| Intra-operative (Specify) | ||||||||
| Intra-operative (Neuro) | ||||||||
| Laparoscopic | ||||||||
| Pediatric | N | N | N | B mode + Mmode | ||||
| Small Organ (includingthyroid) | N | N | N | B mode + Mmode | ||||
| Neonatal Cephalic | ||||||||
| Adult Cephalic | ||||||||
| Trans-rectal | ||||||||
| Trans-vaginal | ||||||||
| Trans-urethral | ||||||||
| Trans-esoph. (non-Card.) | ||||||||
| Musculo-skeletal(Superficial) | N | N | N | B mode + Mmode | ||||
| Intravascular | ||||||||
| Other (Musculo-skeletal -Conventional) | N | N | N | B mode + Mmode | ||||
| Other (Gynecological) | N | N | N | B mode + Mmode | ||||
| Other (Urology) | N | N | N | B mode + Mmode | ||||
| Other (Specify) | ||||||||
| Cardiac | Cardiac Adult | N | N | N | B mode + Mmode | |||
| Cardiac Pediatric | N | N | N | B mode + Mmode | ||||
| Intravascular (Cardiac) | ||||||||
| Trans-esoph. (Cardiac) | ||||||||
| Intra-cardiac | ||||||||
| Other (Specify) | ||||||||
| Peripheral Vessel | Peripheral vessel | N | N | N | B mode + Mmode | |||
| Other (Carotid, arterialstudies) | N | N | N | B mode + Mmode | ||||
| Other (Procedural Guidance) | N | N | N | B mode + Mmode |
{4}------------------------------------------------
N = new indication; P = previously cleared by FDA; E = added under this appendix
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Butterfly Network. The logo consists of a blue butterfly icon on the left and the words "Butterfly Network" in black on the right. The word "Butterfly" is in a larger font than the word "Network", and there is a trademark symbol next to the word "Butterfly".
510(k) Summary of Safety and Effectiveness
Submitter Information
Submitter Name and Address
Butterfly Network, Inc. 530 Old Whitfield St. Guilford, CT 06437 USA (tel.) 203.204.6600 (fax) 203.458.2514 www.butterflynetwork.com
Contact Person
Brian Sawin Sr. Requlatory Affairs Manager 203.204.6600 bsawin@butterflynetinc.com
Date Prepared
August 9, 2017
Subject Device - Proprietary/Trade Name
Poseidon Ultrasound System
Subject Device - Common Name
Ultrasound Imaging System
Classification Name
| Regulation Number | Product Code | |
|---|---|---|
| Ultrasonic Pulsed Doppler Imaging System | 892.1550 | 90-IYN |
| Ultrasonic Pulsed Echo Imaging System | 892.1560 | 90-IYO |
| Diagnostic Ultrasound Transducer | 892.1570 | 90-ITX |
Classification
Class II
Predicate Device:
K162549 - Philips Healthcare, Lumify Ultrasound System (Clearance Date:
{6}------------------------------------------------
10/03/2016)
Device Summary:
The Poseidon Ultrasound System is a general-purpose diagnostic imaging system for use by qualified and trained healthcare professionals enabling visualization and measurement of anatomical structures and fluid on adult and pediatric patients. The system consists of a single transducer with broad imaging capabilities connected via a USB cable to a standard handheld commercial off the shelf (COTS) mobile device. In addition to M-mode and B-mode imaging the instrument also supports Color Flow Doppler imaging.
The user interface includes touchscreen menus, buttons, controls, indicators and navigation icons that allow the operator to control the system and to view and measure ultrasound imagery.
Indications for Use:
The Poseidon Ultrasound System is indicated for use by qualified and trained healthcare professionals to enable diagnostic ultrasound imaging and measurement of anatomical structures and fluids of adult and pediatric patients for the following clinical applications: Peripheral Vessel (including carotid and arterial studies), Procedural Guidance, Small Organs (including thyroid), Cardiac, Abdominal, Urology, Fetal/Obstetric, Gynecological, Musculoskeletal (conventional) and Musculoskeletal (superficial).
Summary of Technological Characteristics
There are no technological characteristics, features or indications for use in this submission that are not previously evaluated and cleared in the predicate device, with the exception of utilizing a capacitive micromachined ultrasonic transducer (CMUT) array rather than a traditional crystalline piezoelectric array. This technology meets the same intended use and performs the same actions as piezoelectric material.
Summary of Safety and Performance
Verification and validation activities were designed and performed to demonstrate that the Poseidon Ultrasound System meets predetermined performance specifications. The following standards were used to determine appropriate methods for evaluating the performance of the device:
IEC 60601-1: Medical Electrical Equipment – Part 1: General Requirements for Safety.
IEC 60601-1-2: Medical Electrical Equipment - Part 1-2: General Requirements for Safety – Collateral Standard: Electromagnetic Compatibility - Requirements and Tests.
IEC 60601-2-37: Medical Electrical Equipment – Part 2-37: Particular
{7}------------------------------------------------
Requirements for the Safety of Ultrasonic Medical Diagnostic and Monitoring Equipment.
ISO 10993:2009 Biological Evaluation of Medical Devices. Part 1
NEMA UD-2: Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment.
Summary of Substantial Equivalence:
Based on the indications for use, technological characteristics, and safety and performance testing, the subject device meets the requirements that are considered adequate for its intended use and is substantially equivalent in design, principles of operation and indications for use to the predicate device.
N/A