K Number
K163432

Validate with FDA (Live)

Device Name
Stellar LEAP
Date Cleared
2017-06-27

(202 days)

Product Code
Regulation Number
890.3850
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Stellar LEAP Manual Tilt Wheelchair is intended to provide mobility to persons restricted to a seated position.

Device Description

The Stellar LEAP manual tilt wheelchair is an indoor / outdoor, manually operated, Tilt-in-Space wheelchair. Its intended function and use is to provide mobility to persons ages 16 and up (adolescents and adults). The Stellar LEAP manual tilt wheelchair consists of metal frame architecture with a backrest, seat, and rear wheels, which allow for occupant or attendant propelling of the device. Smaller caster wheels are also mounted on the front of the frame to facilitate steering and turning. The Stellar LEAP consists of three (3) metal frame subassemblies: a lower frame, upper seat frame, and sub-frame assembly. The manual Tilt-in-Space operation is achieved using a gas strut mechanism connected between the upper seat frame and sub-frame. The Stellar LEAP Manual Tilt Wheelchair is available in a standard configuration, which allows 20° Tilt-in-Space (posterior) tilt, with a maximum weight capacity of 250 lbs.

PDG offers optional features / options to be added to the Stellar LEAP wheelchair, including the two below features:

  • Stellar LEAP Anterior Tilt Feature (0° 30°): Functional reach extension & transfer . assist
  • Stellar LEAP Dynamic Recline Backrest Tilt Feature: 30° range of dynamic recline for the backrest combined with the anterior tilt

Other available options / accessories to complete the Stellar LEAP configuration include: seat width / depth / height, armrests, casters, frog leg suspension forks, rear wheel / tire, hand rim, wheel locks, anti-tippers, back post style, back options, back upholstery, headrest, tilt control, recline control, front rigging, foot plates, footrest / leg rest accessories, position straps, IV pole, 02 holder.

AI/ML Overview

The provided text describes the 510(k) summary for the Stellar LEAP manual tilt wheelchair (K163432). It details the device, its intended use, comparison to predicate devices, and performance data. However, the requested information focuses on performance criteria and studies for AI/ML-driven devices, which is not applicable to this submission for a mechanical wheelchair.

Therefore, most of the specific points in the request cannot be answered from the provided document.

Here's a breakdown of what can be extracted and what cannot:

1. Table of acceptance criteria and the reported device performance:

The acceptance criteria for the Stellar LEAP wheelchair are defined by its compliance with various ISO 7176 standards. The reported performance is that the device "meets the requirements" or "performed compliant" to these standards.

Acceptance Criteria (ISO Standard Requirement)Reported Device Performance
ISO 7176 – Part 1 – Static stabilityMeets requirements
ISO 7176 – Part 3 – Effectiveness of brakesMeets requirements
ISO 7176 Part 5 – Overall dimensions, mass, turning spaceMeets requirements
ISO 7176 Part 7 – Measurement of seating and wheel dimensionsMeets requirements
ISO 7176 Part 8 – Static impact and fatigue strengthsMeets requirements
ISO 7176 – Part 11 – Test dummiesMeets requirements
ISO 7176 – Part 13 – Coefficient of friction of test surfacesMeets requirements
ISO 7176 – Part 15 – Information disclosure, documentation, labelingMeets requirements
ISO 7176 – Part 16 – Resistance to ignition of postural support devicesMeets requirements
ISO 7176 – Part 19 – Wheeled mobility devices for use as seat in motor vehiclesMeets requirements
Biocompatibility (ISO 10993-1, FDA Guidance)Biocompatibility assessment carried out and deemed safe

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size for Test Set: Not applicable/not specified for a mechanical device. The testing described is based on engineering verification and validation according to international standards, typically involving a set number of units or prototypes to test specific characteristics.
  • Data Provenance: Not applicable for this type of device. The provenance is the testing conducted by the manufacturer against established ISO standards.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. "Ground truth" in the context of AI/ML is not relevant here. Compliance with ISO standards is verified through defined test methods and measurements, not expert consensus on data interpretation.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. This concept is for diagnostic performance assessment, not mechanical device testing.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is for AI/ML diagnostic devices.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is for AI/ML systems.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

Not applicable. The "ground truth" for a mechanical device is its adherence to specified engineering, safety, and performance standards as determined by standardized tests and measurements, as well as risk analysis and biocompatibility assessment.

8. The sample size for the training set

Not applicable. The Stellar LEAP is a mechanical medical device, not an AI/ML product that requires a training set for algorithm development.

9. How the ground truth for the training set was established

Not applicable. No training set is used for this device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 27, 2017

Pdg Product Design Group, Inc % Georgiann Keyport Senior Partner Canopy Medical LLC 1160 Vierling Drive Shakopee, Minnesota 55379

Re: K163432

Trade/Device Name: Stellar LEAP Regulation Number: 21 CFR 890.3850 Regulation Name: Mechanical Wheelchair Regulatory Class: Class I Product Code: IOR Dated: May 29, 2017 Received: June 2, 2017

Dear Georgiann Keyport:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the

{1}------------------------------------------------

electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and Part 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

Michael J. Hoffmann -S

for

Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K163432

Device Name Stellar LEAP

Indications for Use (Describe)

The Stellar LEAP Manual Tilt Wheelchair is intended to provide mobility to persons restricted to a seated position.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) SUMMARY - K163432

Date Prepared: June 26th, 2017

510(k) SubmitterContact
PDG Product Design Group Inc.Contact:
103-318 East Kent Ave. SouthTorr Brown, Chief Engineer
Vancouver, CA-BC CANADA V5X 4N6Tel: 604-323-9220; Fax: 604-323-9097
Email: tbrown@pdgmobility.com
General Information
Trade NameStellar LEAP
Common NameManual Tilt Wheelchair
ClassificationMechanical wheelchair (21 CFR 890.3850; Class I)
InformationProCode: IOR; Panel: Physical Medicine
Predicate Device(s)The Stellar Tilt (K990557)
Reference Device (s)PDG Elevation Manual Wheelchair (K140023)

1.1 Device Description

The Stellar LEAP manual tilt wheelchair is an indoor / outdoor, manually operated, Tilt-in-Space wheelchair. Its intended function and use is to provide mobility to persons ages 16 and up (adolescents and adults). The Stellar LEAP manual tilt wheelchair consists of metal frame architecture with a backrest, seat, and rear wheels, which allow for occupant or attendant propelling of the device. Smaller caster wheels are also mounted on the front of the frame to facilitate steering and turning. The Stellar LEAP consists of three (3) metal frame subassemblies: a lower frame, upper seat frame, and sub-frame assembly. The manual Tilt-in-Space operation is achieved using a gas strut mechanism connected between the upper seat frame and sub-frame. The Stellar LEAP Manual Tilt Wheelchair is available in a standard configuration, which allows 20° Tilt-in-Space (posterior) tilt, with a maximum weight capacity of 250 lbs.

PDG offers optional features / options to be added to the Stellar LEAP wheelchair, including the two below features:

  • Stellar LEAP Anterior Tilt Feature (0° 30°): Functional reach extension & transfer . assist
  • Stellar LEAP Dynamic Recline Backrest Tilt Feature: 30° range of dynamic recline for the backrest combined with the anterior tilt

{4}------------------------------------------------

Other available options / accessories to complete the Stellar LEAP configuration include: seat width / depth / height, armrests, casters, frog leg suspension forks, rear wheel / tire, hand rim, wheel locks, anti-tippers, back post style, back options, back upholstery, headrest, tilt control, recline control, front rigging, foot plates, footrest / leg rest accessories, position straps, IV pole, 02 holder.

1.2 Intended Use / Indications

The Stellar LEAP wheelchair is intended to provide mobility to persons restricted to a seated position.

The Indications for Use statement for the Stellar LEAP is identical to the predicate device, the Stellar Tilt. Both the subject and predicate devices have the same intended use for providing mobility to persons restricted to a seated position.

1.3 Substantial Equivalence Comparison

The Stellar LEAP and predicate Stellar Tilt wheelchairs have the same basic design, principle of use and intended use / indications. The subject and predicate devices share similar dimensions, weight, and both offer posterior Tilt-in-Space capability. The overall dimensions, seat width, seat depth, and adjustable seat to floor height for both the subject and predicate wheelchairs are very similar. The subject, predicate and reference device differ in their frame design, materials of construction, and tilt features, however the overall function of the wheelchairs are similar. The predicate device does not include the anterior tilt feature, thus the reference device K140023 has been included in the comparison below.

Comparison of Technological characteristics with the Predicate Device and ReferenceDevice.
Subject DeviceStellar LEAPPredicate DeviceStellar Tilt(As cleared underK990557)ReferenceDeviceElevation(As cleared underK140023)
CharacteristicDimension/WeightDimension/WeightDimension /Weight
Seat width14" - 22"14" - 22"12" - 19"
Seat depth16" - 22"16" - 22"14" - 20"
Seat to Floor Height(adjustable)13" - 18"14" - 20"18" - 21"
Posterior Tilt0° to 20°45°15°±5°(depending on setup)
Comparison of Technological characteristics with the Predicate Device and ReferenceDevice.
Subject DevicePredicate DeviceReference Device
Stellar LEAPStellar Tilt(As cleared under K990557)Elevation(As cleared under K140023)
Back angle (non-dynamic adjustable)90°, 98°, 106°, 114°, 122°90°, 98°, 106°, 114°, 122°N/A
Option / Accessory: Back angle -dynamic or Dynamic Recline BackrestRange of 30°, @ neutral (90°- 120°)Range of 30°Range of 30°, @ neutral (90°- 120°)
Option / Accessory: Anterior Tilt30° (max)N/A25°±10°(depending on set up)
Back height20" or 25" or 31"25" or 31"(seat width +9.5")9.4"-14.6"
Overall width12" and16" rearwheels(seat width +9.5")N/A
20",22", 24"rearwheels(seat width +10.5")25" rear wheels
(seat width +10.5")Min 23.1"Max 27.1"
Overall length (without front rigging)30" – 37"30" – 37"With Legrest(32"-24.5")
Weight capacity250 lbs.250 lbs.250 lbs.
Chair weight,ISO7176-5 Part 88.983 lbs. (37.8kg)75 lbs. (34.2kg)26.5 lbs. (12.0 kg)
Static stability downhillISO7176-1 Part 88.2Tippingcommences atgreater than 10.5°Tippingcommences atgreater than 11.0°Tippingcommences atgreater than 10°
8.4Tippingcommences atgreater than 10°Tippingcommences atgreater than 11.0°Tippingcommences atgreater than 10°
Static stability uphill9.3Tippingcommences at 7.8°Tippingcommences at 7.8°Tippingcommences at
Subject DevicePredicate DeviceReference Device
Stellar LEAPStellar Tilt(As cleared under K990557)Elevation(As cleared under K140023)
ISO7176-1 Part 99.5Tipping commences at 8.2°onto anti-tipsgreater than 5.4°
Static stabilitysidewaysISO7176-1 Part10.2, 10.310.2Tipping commences atgreater than 10°Tipping commences atgreater than 12°Tipping commences atgreater than 10°
10.3Tipping commences atgreater than 10°Tipping commences atgreater than 12°Tipping commences atgreater than 10°
Static stabilityanti-tipISO7176-1 Part 1111.2Tipping commences atgreater than 10°Tipping commences atgreater than 14.0°Tipping commences atgreater than 10°
11.3Tipping commences atgreater than 17°Tipping commences atgreater than 14.0°Tipping commences atgreater than 10°
11.4N/ATipping commences atgreater than 14.0°N/A
Minimum turningradiusISO7176-5 Part 88.13830mm815 mm515mm

{5}------------------------------------------------

{6}------------------------------------------------

1.4 Performance Data

The Stellar LEAP has been evaluated through ISO 14971-compliant risk analysis and nonclinical testing performed in accordance with the below wheelchair standards:

  • . ISO 7176 – Part 1 – Determination of static stability
  • ISO 7176 – Part 3 – Determination of effectiveness of brakes
  • ISO 7176 Part 5 Determination of overall dimensions, mass and turning space •
  • ISO 7176 Part 7 Measurement of seating and wheel dimensions. .

{7}------------------------------------------------

  • ISO 7176 Part 8 Requirements and test methods for static impact and fatigue ● strengths
  • ISO 7176 - Part 11 - Test dummies
  • ISO 7176 - Part 13 - Determination of coefficient of friction of test surfaces.
  • . ISO 7176 - Part 15 - Requirements for information disclosure, documentation and, labeling.
  • ISO 7176 - Part 16 - Resistance to ignition of postural support devices
  • . ISO 7176 - Part 19 - Wheeled mobility devices for use as seat in motor vehicles

Results of the above ISO evaluations demonstrated that the Stellar LEAP meets the requirements of ISO 7176- 1, 3, 5, 7, 8, 11, 13, 15, 16, and 19.

Furthermore, a biocompatibility assessment was carried out for the Stellar LEAP by identifying the patient contacting materials and conducting a biomaterial evaluation in accordance with:

  • ISO 10993-1 Biological evaluation of medical devices Part 1: Evaluation and ● testing within a risk management process (2010)
  • FDA Guidance: Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" (June 16, 2016)

1.5 Conclusions

The Stellar LEAP has substantially equivalent indications and principle of operation to the currently marketed Stellar Tilt. While there are design differences in the frame configuration. materials of construction, joining methods and tilt features, the Stellar LEAP has been evaluated through ISO 14971 - Application of risk management to medical devices and design verification and validation (DV&V) testing was carried out and showed the Stellar LEAP performed compliant to ISO 7176-1, 3. 5. 7. 8. 11. 13. 15. 16 and 19. The subject device has the same intended use, principles of operation, and technological characteristics to the predicate device. Furthermore, similar to the reference device, Elevation, the Stellar LEAP utilizes gas struts to provide anterior tilting function capability. Additionally, the biocompatibility assessment of the Stellar LEAP was carried out in consideration of the safe history of use of these patient contacting materials with other marketed PDG wheelchairs. Thus the Stellar LEAP raises no new questions of safety or effectiveness compared to the predicate device and is, therefore, substantially equivalent.

  • End of 510(k) Summary -

§ 890.3850 Mechanical wheelchair.

(a)
Identification. A mechanical wheelchair is a manually operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.(b)
Classification. Class I (general controls).