K Number
K163429

Validate with FDA (Live)

Device Name
B6
Date Cleared
2017-01-04

(28 days)

Product Code
Regulation Number
892.1650
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The B6 is a mobile digital X-ray diagnostic system, which is intended to generate X-ray fluoroscopic image of a patient. The application includes:real-time positioning and monitoring operations in trauma surgery, orthopedics, spine surgery, and chest surgery. It is not intended to be used in interventional procedures.

The B6 permits a qualified doctor or technologist to take a range of diagnostic exposures of spinal column, chest, abdomen, extremities, and other body parts on the patients.

Device Description

There are two sets of X-ray tube assemblies and Image Intensifiers which are perpendicularly distributed on the G-Arm, acting as two sets of vertical X-ray source and receptor systems and providing fluoroscopy image of the patient. The two sets of X-ray tube assemblies and Image Intensifiers can operate simultaneously and separately.

The B6 is comprised of control unit and G-Arm, the control unit and G-Arm include below primary components.

Control Unit Major ComponentQty.
Viewing monitor2
Control monitor1
Keyboard1
Footswitch (3- or 2- pedal)1
Printer(optional)1
G-Arm Major ComponentQty.
Image intensifier assembly2
X-Ray tube assembly2
Laser light indicator2

The G-Arm system offers a group of motor-driven positioning features, including:

  • lmage Intensifier Adjustment: 300 mm±10mm
  • G-Arm Height Adjustment: 300 mm±10mm
  • G-Arm Orbital Rotation: +35°+2°
  • G-Arm Radial Rotation: ±45°±2°

The B6 can be operated in two different modes: Continuous fluoroscopy and Pulse fluoroscopy. Each of these two modes can be operated either manually or automatically.

So, the fluoroscopy mode can be categorized by the X-ray emitting pattern or by the control method of the imaging process, which consequently divides a fluoroscopy mode into the following four major types: Automatic Continuous Fluoroscopy, Automatic Pulse Fluoroscopy, Manual Pulse Fluoroscopy, Manual Continuous Fluoroscopy.

Besides the above four fundamental fluoroscopy modes, the G-Arm system offers two other features: ER Fluoroscopy and Half Dose Fluoroscopy (modifies only Automatic Pulse Fluoroscopy and Automatic Continuous Fluoroscopy modes)

AI/ML Overview

The document describes a 510(k) premarket notification for the B6 mobile digital X-ray diagnostic system. The focus of the substantiation is on the device's technical characteristics and compliance with recognized standards, rather than clinical performance metrics typically associated with AI-driven diagnostic devices. Therefore, many of the requested categories related to clinical study design, ground truth, and reader performance are not directly applicable or explicitly stated in this type of submission.

Here's an analysis based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for this device are primarily compliance with international and national safety and performance standards for medical electrical equipment and X-ray systems. The device's performance is reported as "Pass" for all tested items against these standards.

Acceptance Criteria CategorySpecific Criteria/StandardReported Device Performance
Electrical SafetyAAMI/ANSI ES60601-1Pass (34 specific items tested, e.g., Power input, Leakage currents, Dielectric strength, Mechanical strength test)
Electromagnetic Compatibility (EMC)IEC 60601-1-2Pass (8 specific items tested, e.g., Mains terminal continuous disturbance voltage, Radiated emission, Electrostatic discharge)
Radiation Protection in Diagnostic X-ray EquipmentIEC 60601-1-3Pass (4 specific items tested, e.g., Half-value layer, Leakage radiation in the loading state)
Specific Requirements for X-ray Equipment for Radiography and RadioscopyIEC 60601-2-54Pass (4 specific items tested, e.g., Test for dosimetric information, Accuracy of X-ray tube voltage)
Digital Imaging and Communications in Medicine (DICOM)DICOM provisionsMeets DICOM provisions
Software Functionality and ValidationInternal software testingPass (18 specific items tested, e.g., Function of Main Control Panel Software, Function of image process, CAN bus communication)
Substantial EquivalenceDemonstrates safety and effectiveness comparable to predicate deviceConcluded to be substantially equivalent based on non-clinical testing

2. Sample Size Used for the Test Set and the Data Provenance

This document describes non-clinical testing and verification, not a clinical study involving patient data. Therefore, the concepts of "test set sample size" and "data provenance" in the context of patient data do not apply. The testing involved the device (B6) itself as the "sample" for verification against technical standards.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

Not applicable. Ground truth, in the context of diagnostic accuracy, is not relevant for this type of technical performance and safety validation. The "ground truth" here is the adherence to established technical standards and specifications. The document implies that qualified engineers and technicians performed the tests, but their specific qualifications are not detailed.

4. Adjudication Method for the Test Set

Not applicable, as this was not a clinical study requiring adjudication of diagnostic outcomes. The "adjudication" in this context would be the determination of "Pass" or "Fail" against the specified technical test procedures and criteria by the testing personnel/organization.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

No, an MRMC comparative effectiveness study was not done. This submission focuses on the technical safety and performance of the X-ray system itself, not on the diagnostic effectiveness with human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This refers to an X-ray imaging device, not an AI algorithm. Therefore, "standalone (algorithm only)" performance is not applicable in the typical sense. The device's performance stands alone in its ability to meet the technical standards for image generation and system functionality.

7. The Type of Ground Truth Used

The "ground truth" used for this submission is the established international and national technical standards for medical electrical equipment and X-ray systems (e.g., AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-54) and the device's design specifications. "Pathology" or "outcomes data" are not relevant here.

8. The Sample Size for the Training Set

Not applicable. This is not an AI-driven device that requires a training set.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for this device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 4, 2017

Beijing East Whale Imaging Technology Co., Ltd. % Ms. Sharon Fan Regulatory Affairs Manager Unit 701, Building 1, Yongchang Industrial Park No. 3 Yongchang North Rd., BDA Beijing, 100176 CHINA

Re: K163429 Trade/Device Name: B6 Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: OXO, JAA Dated: November 22, 2016 Received: December 7, 2016

Dear Ms. Fan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Michael D'Hara

For

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K163429

Device Name B6

Indications for Use (Describe)

The B6 is a mobile digital X-ray diagnostic system, which is intended to generate X-ray fluoroscopic image of a patient. The application includes:real-time positioning and monitoring operations in trauma surgery, orthopedics, spine surgery, and chest surgery. It is not intended to be used in interventional procedures.

The B6 permits a qualified doctor or technologist to take a range of diagnostic exposures of spinal column, chest, abdomen, extremities, and other body parts on the patients.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) □ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the word "WHALE" in large, bold, blue letters. To the right of the word is a small, white square with a rounded corner. Inside the square are two horizontal green bars. The logo is simple and clean, with a focus on the company name and a subtle graphic element.

013_510(k) Summary of Safety and Effectiveness

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the word "WHALE" in large, blue, sans-serif font. To the right of the word is a small, white square with rounded corners. Inside the square are two horizontal green bars, stacked on top of each other.

510(k) Summary of Safety and Effectiveness

[As required by 21 CFR 807.92]

1. Date Prepared [21 CFR807.92 (a) (1)]

Nov 22th, 2016

2. Submitter's Information [21 CFR807.92 (a) (1)]

Name of Sponsor:Beijing East Whale Imaging Technology Co., Ltd.
Address:Unit 701, Building 1, Yongchang Industrial Park, No. 3 Yongchang North Rd., BDA,100176, Beijing, China.
Contact Name:Sharon Fan
Telephone No.:+ 86 (10) 67896701- 8058
Fax No.:+ 86 (10) 67082218
Email Address:fanxiaoyan@whaleimaging.com

3. Trade Name, Common Name, Classification [21 CFR807.92(a)(2)]

Trade Name:B6
Common Name:MultiScan G-Arm System
Classification:Image-intensified fluoroscopic X-ray system
Product code:OXO&JAA
Regulation Number:21 CFR 892.1650
Classification Panel:Radiology
Device Class:II

4. Identification of Predicate Device(s) [21 CFR 807.92(a)(3)[

Table 1- Predicate Devices

510(k) NumberK160984
ApplicantBeijing East Whale Imaging Technology Co., Ltd.
Device NameMultiScan G-Arm System

{5}------------------------------------------------

5. Description of the Device [21 CFR 807.92(a)(4)]

There are two sets of X-ray tube assemblies and Image Intensifiers which are perpendicularly distributed on the G-Arm, acting as two sets of vertical X-ray source and receptor systems and providing fluoroscopy image of the patient. The two sets of X-ray tube assemblies and Image Intensifiers can operate simultaneously and separately.

The B6 is comprised of control unit and G-Arm, the control unit and G-Arm include below primary components.

Control Unit Major ComponentQty.
Viewing monitor2
Control monitor1
Keyboard1
Footswitch (3- or 2- pedal)1
Printer(optional)1

Table 2-1 control unit Primary components list

Table 2-2 G-Arm Primary components list

G-Arm Major ComponentQty.
Image intensifier assembly2
X-Ray tube assembly2
Laser light indicator2

The G-Arm system offers a group of motor-driven positioning features, including:

  • lmage Intensifier Adjustment: 300 mm±10mm
  • G-Arm Height Adjustment: 300 mm±10mm

G-Arm Orbital Rotation: +35°+2°

  • G-Arm Radial Rotation: ±45°±2°
  • 510(k) Summary of Safety and Effectiveness

{6}------------------------------------------------

The B6 can be operated in two different modes: Continuous fluoroscopy and Pulse fluoroscopy. Each of these two modes can be operated either manually or automatically.

So, the fluoroscopy mode can be categorized by the X-ray emitting pattern or by the control method of the imaging process, which consequently divides a fluoroscopy mode into the following four major types: Automatic Continuous Fluoroscopy, Automatic Pulse Fluoroscopy, Manual Pulse Fluoroscopy, Manual Continuous Fluoroscopy.

Besides the above four fundamental fluoroscopy modes, the G-Arm system offers two other features: ER Fluoroscopy and Half Dose Fluoroscopy (modifies only Automatic Pulse Fluoroscopy and Automatic Continuous Fluoroscopy modes)

6. Intended Use [21 CFR 807.92(a)(5)]

The B6 is a mobile digital X-ray diagnostic system, which is intended to generate X-ray fluoroscopic image of a patient. The application includes: real-time positioning and monitoring operations in trauma surgery, orthopedics, spine surgery, and chest surgery. It is not intended to be used in interventional procedures.

The B6 permits a qualified doctor or technologist to take a range of diagnostic exposures of spinal column, chest, abdomen, extremities, and other body parts on the patients.

7. Technological Characteristics [21 CFR 807.92(a)(6)]

ItemPredicate DeviceB5 SSubject DeviceB6SERemark
K#K160984In process---
ApplicableStandardAAMI / ANSI ES60601-1IEC60601-1-2IEC60601-1-3IEC60601-2-54AAMI / ANSI ES60601-1IEC60601-1-2IEC60601-1-3IEC60601-2-54Same
Power apply110/120VAC±10%, 50/60Hz110/120VAC±10%, 50/60HzSame
EnergydeliveredX-Ray and laserX-Ray and lasersame
PrimarycomponentsImage gantry + X-ray tube + imageintensifier +view monitor + controlmonitor + footswitchImage gantry + X-ray tube + imageintensifier +view monitor + controlmonitor + footswitchSame
Mobile orstationaryMobileMobileSame
FluoroscopyModeManual Continuous FluoroscopyAutomatic Continuous FluoroscopyManual Pulse FluoroscopyAutomatic Pulse FluoroscopyER Fluoroscopy (ER)Half Dose FluoroscopyManual Continuous FluoroscopyAutomaticContinuousFluoroscopyManual Pulse FluoroscopyAutomatic Pulse FluoroscopyER Fluoroscopy (ER)Half Dose FluoroscopySame
Design
AppearancedesignAppearance of B5 S see the figure 1of "012_ Substantial EquivalenceDevice Comparison.pdf"Appearance of B6 see figure 2of "012_ Substantial EquivalenceDevice Comparison.pdf"Note 1
ElectricalstructureThe circuits regarding to Inverter,switch power supply, a part of PCB inthe control unit side.The control panel only in the controlunit side .The circuits regarding to Inverter,switch power supply, a part of PCBin the G-Arm side.The control panel both in thecontrol unit and G-Arm, but thecontrol panel in the G-Arm sidecannot control exposure.Note 2
Constitute and structure
Trackingwheel systemOptionalTracking wheels is integrated inthe base of G-Arm.Note 3
Function and performance
Manual Continuous FluoroscopykV: 40–120, step: 1kVmA: 0.1–8.0, step: 0.1 mAManual Continuous FluoroscopykV: 40–120, step: 1kVmA: 0.2–8.0, step: 0.1 mA
Automatic Continuous FluoroscopykV: 40–120, step: 1kVmA: 7.0mAAutomatic ContinuousFluoroscopykV: 40–120mA: 0.2–7.0
Manual Pulse FluoroscopykV: 40–120, step: 1kVmA: 0.1–8.0, step: 0.1 mAManual Pulse FluoroscopykV: 40–120, step: 1kVmA: 0.1–15.0 (1-planeimaging)/8.0mA (2-planeimaging), step: 0.1 mA
Ranges ofloadingfactorsAutomatic Pulse FluoroscopykV: 40–120, step: 1kVmA: 8.0mAAutomatic Pulse FluoroscopykV: 40–120, step: 1kVmA: 0.1–15.0 (1-planeimaging)/8.0mA (2-plane imaging)Note 4
ER FluoroscopykV: min40, max120mA : 8.0mAER FluoroscopykV: min40, max120mA : 8.0mA
Half automatic continuousfluoroscopy:kV: 40–120, step: 1kVmA: 3.5mAHalf dose automatic continuousfluoroscopy:kV: 40-120, step: 1kVmA: 0.2-3.5mA
Half automatic pulse fluoroscopy:kV: 40–120, step: 1kVmA: 7.5mAHalf dose automatic pulsefluoroscopy:kV: 40–120, step: 1kVmA: 0.2–7.5mA
movementG-Arm Orbital rotation
G-Arm radial rotation
G-Arm lift distance300 mm±10 mm300 mm±10 mm
I.I. unit travel distance300 mm±5mm300 mm±10mm
Movement controlControl keys on the control panel in the control unit.Control keys on the line control in the G-Arm.
The G-Arm's orbital rotation is driven by DC motor.The G-Arm's orbital and radial rotation are driven by servo motor.
Physical Characteristics
SID890-1190mm800-1100mm
dimensionG-Arm:2116(min L)/ 2324(max L) x 910 (W)x1950(min H)/ 2550(max H)G-Arm:2520(min L)/ 2820(max L) x 910 (W) x 1876(min H)/ 2476(max H)Note 6
Control unit:864 (L) x 1110 (W) x 1873(H)Control unit:904(L) x 1010 (W) x 1760(H)
WeightG-arm: ≈365 kg (with tracking wheel system)G-Arm: ≈ 520 kg
Control Unit≈ 235 kgControl Unit: ≈ 150 kg
Whole Product Gross Weight: 600 kg(with tracking wheels)Whole Product Gross Weight: ≈ 670 kg

Table 3- Major similarities and differences between subject and predicate devices

{7}------------------------------------------------

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the word "WHALE" in large, bold, blue letters. To the right of the word is a white square with rounded corners. Inside the square are two horizontal green bars, stacked on top of each other.

{9}------------------------------------------------

510(k) Submission

Image /page/9/Picture/1 description: The image shows the word "WHALE" in large, bold, blue letters. To the right of the word is a white square with a blue outline. Inside the square are two horizontal green bars. The logo appears to be for a company or organization named "WHALE".

Note1 - Appearance design

B6's appearance is more elegant and suitable for the electrical structure. This difference will not affect safety and effectiveness.

Note 2 - Electrical structure

In B6's electrical structure, only the circuit layout is changed, while the essential principle and components are the same as the predicate device. Adding additional control panel to the G-Arm base helps in better observation of the exposure conditions during surgery and more convenient adjustment of the loading factors. The image-intensifier has not been modified.

{10}------------------------------------------------

The B6 has passed all the tests in accordance to AAMI/ ANSI ES60601-1, IEC60601-1-2, IEC60601-3 and IEC60601-2-54. Therefore, this difference does not raise different questions of safety and effectiveness and that the device is as safe and effective as the predicate device.

Note 3 - Tracking wheel system

B5 S and B6 use different connection types in their tracking wheel systems. The tracking wheels of B6 are mechanically integrated with the G-Arm base, i.e., it is a part of base. The tracking wheel system of B5 S is optional. But the control principle and control process between B5 S and B6 are the same. The B6 has passed all the tests in accordance to AAMI/ ANSI ES60601-1, IEC60601-1-2, IEC60601-3 and IEC60601-2-54. Therefore, this difference does not raise different questions of safety and effectiveness and that the device is as safe and effective as the predicate device.

Note 4 - Ranges of loading factors

Loading factors of the B6 are controlled by the software, the software system has been well tested and validated. The B6 has passed all the tests in accordance to AAMI/ ANSI ES60601-1, IEC60601-1-2, IEC 60601-3 and IEC60601-2-54. Therefore, this change does not raise different questions of safety and effectiveness and that the device is as safe and effective as the predicate device.

Note 5 - Movement

Comparing with the movements of the predicate device, the B6 introduces the radial rotation of the G-Arm and the angle of the orbital rotation movement is changed from ±13° (B5 S) to ±35°±2°.

The G-Arm's orbital and radial rotations are driven by servo motor, which meets the requirements of the associated clauses of AASI/AAMI 60601-1, making the rotation control more accurately.

The movements control keys of B5 S are incorporated into the control panel, which is located on the control unit, while the B6's movements control keys are located on a separate movement control board (line control) on the G-Arm base. The control principle and electrical circuits between B5 S and B6 are the same. No new technology is introduced.

{11}------------------------------------------------

The B6 has passed all the tests in accordance to AAMI/ ANSI ES60601-1, IEC60601-1-2, IEC 60601-3 and IEC60601-2-54. Therefore, this change does not raise different questions of safety and effectiveness and that the device is as safe and effective as the predicate device.

Note 6 - SID & Dimension & Weight

The primary mechanical structure between predicate device and proposed device is the same. The differences with regard to the SID & Dimensions & Weight are resulted from different mechanical characteristic settings. No new technology is introduced. The B6 has passed all the tests in accordance to AAMI/ ANSI ES60601-1, IEC60601-1-2, IEC 60601-3, and IEC60601-2-54. Therefore, this difference does not raise different questions of safety and effectiveness.

The software modifications of proposed device have been verified and validated, the following table include the major changed elements.

No.Modified elementssectionNote
1Separated the G-Arm'smovement from Main controlpanel software2.2.2&4.2.1&4.2.5&4.2.6&4.5&5.3.1&5.5&6.5&9.1&9.4Better control
2Single exposure time limit33# of 3.2User need
3Use CAN bus communication36# of3.2&5.1&6.1&9.2.4More stable and accuratecommunication
4Increase a G-Arm control panel inthe G-Arm side.4.2.3&4.2.4&5.4&6.4&6.5&9.5help doctor to better see theexposure conditions andadjust the loading factorsmore convenience
5Update dose report icon andBMP image save icon4.3.1.3&4.3.1.7&9.1.2Enhance user experience

Table 4- Major modifications of software

{12}------------------------------------------------

The B6 have the same technological characteristics as the predicate device except items in table3-Major similarities and differences between subject and predicate devices. However, the B6 employs the same imaging concepts and fundamental scientific technology with the predicate device and has passed all the tests in according to relevant standards. The differences will not impact the safety and effectiveness of the device.

8. Substantial Equivalence [21 CFR 807.92(b)(1) and 807.92(b)(2)]

Results of performance and compliance testing conducted on B6 indicate conformance to all applicable standards recognized by FDA for this device.

Testing result from non-clinical demonstrates that the proposed device B6 is as safe and effective as the predicate devices.

Non-clinical testing:

The proposed device has been tested to compliance to the following safety and performance standards:

a) AAMI / ANSI ES60601-1:2005/(R)2012 And A1:2012, C1:2009(R)2012 And A2:2010/(R)2012

No.Testing itemsResult
1Power inputPass
2Humidity preconditioning treatmentPass
3Determination of applied parts and accessible partsPass
4Legibility of markingsPass
5Durability of markingsPass
6ME equipment for connection to a power source by a plugPass
7Internal capacitive circuitsPass
8Protective earth connectionPass
9Leakage currentsPass
10Dielectric strengthPass
11Ball pressure testPass
12Measurement of creepage distance and air clearancePass
13Cord anchoragePass
14GapsPass
15Instability-overbalance in transport positionPass
16Instability-overbalance excluding transport positionPass
17Instability-overbalance from horizontal and vertical forcesPass
18Castors and wheels -force for propulsionPass
19Castors and wheels -movement over a thresholdPass
20Instability from unwanted lateral movement in transport positionPass
21Instability from unwanted lateral movement excluding transport positionPass
22Audible acoustic energyPass
23Support systemPass
24Maximum temperature during normal usePass
25SpillagePass
26Ingress of water or particulate matterPass
27CleaningPass
28Interruption of the power supply/supply mains to ME equipmentPass
29Single fault conditionsPass
30Mechanical strength testPass
31Transformer short circuitPass
32Transformer overload testPass
33Transformer dielectric strength testPass
34ME system-leakage measurementsPass

Table 5-- Performed testing items of ANSI/AAMI ES60601-1

Image /page/12/Picture/13 description: The image shows the word "WHALE" in large, bold, blue letters. To the right of the word, there is a white square with rounded corners. Inside the square, there are two horizontal green bars. The logo appears to be for a company or organization named Whale.

{13}------------------------------------------------

WHAL u

{14}------------------------------------------------

Image /page/14/Picture/1 description: The image shows the word "WHALE" in large, bold, blue letters. To the right of the word is a small, white rectangle with rounded corners. Inside the rectangle are two horizontal green bars. The logo is simple and clean, with a focus on the company name.

  • b) IEC 60601-1-2 Edition 3:2007-03
No.Testing itemsResult
1Mains terminal continuous disturbance voltagePass
2Radiated emissionPass
3Electrostatic dischargePass
4RF electromagnetic field immunity testPass
5Fast transients on AC power line, signal line and interconnecting linePass
6Injected current into AC power line, signal line and interconnecting linePass
7Surges to AC power port, signal line and interconnecting linePass
8Variations of power frequencyPass

c) IEC 60601-1-3 Edition 2.1:2013-04

Table 7--Performed testing items of IEC60601-1-3

No.Testing itemsResult
1Half-value layerPass
2Focal spot to image receptor distancePass
3Leakage radiation in the loading statePass
4Leakage radiation when not in the loading statePass
  • d) IEC 60601-2-54 Edition 1.1: 2015-04
    Table 8—Performed testing items of IEC 60601-2-54
No.Testing itemsResult
1Test for dosimetric informationPass
2Accuracy of X-ray tube voltagePass
3Accuracy of X-ray tube currentPass
4Determining the attenuation of residual radiationPass

{15}------------------------------------------------

Also the proposed device meets the provisions of Digital Imaging communications in Medicine

(DICOM)

The proposed device has performed the following software testing to demonstrates that the proposed

device B6 is as safe and effective as the predicate devices

No.Testing itemsResult
1Function of Main Control Panel SoftwarePass
2Function of main interfacePass
3Function of patient listPass
4Function of image processPass
5Function of save asPass
6Function of printPass
7Function of system settingsPass
8Function of on-line processing settingsPass
9Function of DICOM settingsPass
10Function of DICOM serverPass
11Others functionPass
12Long Time ExposurePass
13High Dose ExposurePass
14Pulse ExposurePass
15CAN bus communicationPass
16Laser Aimer Control TestPass
17Function of Movement control Panel SoftwarePass
18Function of G-ARM Control Panel SoftwarePass

Table 9- software testing items

{16}------------------------------------------------

9. Conclusion [21 CFR 807.92(b) (3)]

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the non-clinical result and relative information provided in this premarket notification, we concludes that B6

is substantially equivalent to predicate devices with regard to safety and effectiveness.

N/A