K Number
K163428

Validate with FDA (Live)

Manufacturer
Date Cleared
2017-02-02

(58 days)

Product Code
Regulation Number
888.3050
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The HA Minuteman G3-R MIS Fusion Plate is a posterior, non-pedicle supplemental fixation device, intended for use at a single level in the non-cervical spine (T1-S1). It is intended for plate fixation/attachment to spinous processes for the purpose of achieving supplemental fusion in the following conditions:

  • degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies);
  • spondylolisthesis;
  • . trauma (i.e., fracture or dislocation); and/or
  • . tumor.

The HA Minuteman G3-R MIS Fusion Plate is intended for use with bone graft material and is not intended for stand-alone use. The HA Minuteman G3-R MIS Fusion Plate may be implanted via a minimally invasive lateral approach (L1-S1).

Device Description

The HA Minuteman G3-R MIS Fusion Plate consists of bilateral Plates and a Body/Post that connects the Plates. The Plate components include several spikes on the face of each Plate for attachment to the spinous processes. The HA Minuteman G3-R device is available in multiple sizes to accommodate varying patient anatomy. The HA Minuteman G3-R is made from Ti6Al4V and Ti6Al4V ELI and has a hydroxyapatite coating on a portion of the implant.

AI/ML Overview

The provided text describes a medical device, the HA Minuteman G3-R MIS Fusion Plate, and its submission for 510(k) clearance to the FDA. The document focuses on establishing substantial equivalence to a predicate device through technological characteristics and performance data. However, it does not contain information typically found in a study demonstrating how a device meets specific acceptance criteria based on clinical or extensive human-in-the-loop performance. Instead, it describes mechanical and simulated deployment tests.

Here's an analysis of the provided text in relation to your request, highlighting what information is available and what is not:

1. A table of acceptance criteria and the reported device performance

The document does not provide a formal table of acceptance criteria with corresponding performance metrics like those found in a clinical study. It lists several mechanical tests performed, and for these, it states a general conclusion:

Acceptance Criteria CategoryReported Device Performance
Mechanical Performance
Static Axial Compression Test"functioned as intended and the results observed were as expected."
Static Torsion Test"functioned as intended and the results observed were as expected."
Dynamic Axial Compression Test"functioned as intended and the results observed were as expected."
Static Axial Pull-Out Test"functioned as intended and the results observed were as expected."
Static Plate Dissociation Test"functioned as intended and the results observed were as expected."
Simulated Deployment Testing"functioned as intended and the results observed were as expected."
Custom Body Connection Feature Tests (Torque, Axial Locking, Axial Disengagement)"functioned as intended and the results observed were as expected."
Cadaveric Insertion Testing"functioned as intended and the results observed were as expected."
Biocompatibility
Bacterial endotoxin testing"will also be performed on all batches of sterile packed devices."

Important Note: The document lacks specific quantitative acceptance criteria (e.g., "withstanding X N-m of torque without failure") and the exact quantitative results from these tests. The statement "functioned as intended and the results observed were as expected" is a qualitative summary, implying that the device met internal design specifications or standards comparable to the predicate.

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

  • Sample Size: The document does not specify sample sizes for the mechanical tests. For "Cadaveric Insertion Testing," it does not state how many cadavers were used. For "bacterial endotoxin testing," it mentions "all batches of sterile packed devices," which refers to manufacturing quality control rather than a specific test set for initial device validation.
  • Data Provenance: Not mentioned. The tests appear to be conducted by the manufacturer, Spinal Simplicity. The context suggests these are internal validation tests for the 510(k) submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

This information is not provided. The listed tests are primarily mechanical and engineering assessments, not human interpretation tasks that would require "experts to establish ground truth" in the sense of clinical diagnosis or image labeling. The "Cadaveric Insertion Testing" would have involved medical professionals, but their specific qualifications and role in establishing "ground truth" (e.g., successful implantation, stability assessment) are not detailed.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. The tests described are objective mechanical tests or simulated procedures, not subjective assessments requiring adjudication by multiple experts.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. The HA Minuteman G3-R MIS Fusion Plate is a physical surgical implant, not an AI-powered diagnostic or assistive tool for human readers. Therefore, an MRMC comparative effectiveness study involving human readers and AI is not relevant to this device.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Not applicable. This is a physical medical device, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

For the mechanical tests, the "ground truth" would be the established engineering specifications, material properties, and expected mechanical responses determined through design and industry standards for spinal implants. For the "Cadaveric Insertion Testing," the "ground truth" would be the successful, unhindered placement of the device according to surgical protocol within the cadaveric anatomy. These are performance benchmarks rather than diagnostic "ground truths."

8. The sample size for the training set

Not applicable. This is a physical medical device, not a machine learning model that requires a training set.

9. How the ground truth for the training set was established

Not applicable, as there is no training set for this type of device.

Summary of what the document implies about meeting acceptance criteria:

The document states that the device "functioned as intended and the results observed were as expected" for all mechanical and simulated deployment tests. This general statement indicates that the device met the manufacturer's performance specifications, which would have been established based on industry standards, predicate device data, and risk analysis. The purpose of this 510(k) submission is to demonstrate substantial equivalence to an already marketed predicate device, not necessarily to set new, independently validated clinical acceptance criteria using large studies. The FDA's review confirms that these internal tests, along with the device's technological characteristics, are sufficient to support the claim of substantial equivalence and do not raise new questions of safety or effectiveness.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human figure in profile, repeated three times to create a sense of depth and community. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the figure.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Febraury 2, 2017

Spinal Simplicity Mr. Douglas B. Snell Director of Engineering, Quality, and Regulatory 10995 Quivira Road Overland Park, Kansas 66210

Re: K163428

Trade/Device Name: HA Minuteman G3-R MIS Fusion Plate Regulation Number: 21 CFR 888.3050 Regulation Name: Spinal interlaminal fixation orthosis Regulatory Class: Class II Product Code: PEK Dated: January 6, 2017 Received: January 9, 2017

Dear Mr. Snell:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page

510(k) Number (if known) K163428

Device Name

HA Minuteman G3-R MIS Fusion Plate

Indications for Use (Describe)

The HA Minuteman G3-R MIS Fusion Plate is a posterior, non-pedicle supplemental fixation device, intended for use at a single level in the non-cervical spine (T1-S1). It is intended for plate fixation/attachment to spinous processes for the purpose of achieving supplemental fusion in the following conditions:

  • degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies);
  • spondylolisthesis;
  • . trauma (i.e., fracture or dislocation); and/or
  • . tumor.

The HA Minuteman G3-R MIS Fusion Plate is intended for use with bone graft material and is not intended for stand-alone use. The HA Minuteman G3-R MIS Fusion Plate may be implanted via a minimally invasive lateral approach (L1-S1).

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.qov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/14)

{3}------------------------------------------------

510(k) SUMMARY Spinal Simplicity's HA Minuteman G3-R MIS Fusion Plate

Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared

  • Spinal Simplicity 10995 Quivira Road Overland Park, KS 66210 Phone: (913) 451 4414 Facsimile: (913) 888 0075 Contact Person: Douglas B Snell Date Prepared: January 6th, 2016

Name of Device

HA Minuteman G3-R MIS Fusion Plate

Common or Usual Name

Spinous Process Plate, 21 CFR 888.3050, Class II Product Code PEK

Primary Predicate Device

Spinal Simplicity HA Minuteman G3 MIS Fusion Plate (K151741)

Indications for Use

The HA Minuteman G3-R MIS Fusion Plate is a posterior, non-pedicle supplemental fixation device, intended for use at a single level in the non-cervical spine (T1-S1). It is intended for plate fixation/attachment to spinous processes for the purpose of achieving supplemental fusion in the following conditions:

  • degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies);
  • . spondylolisthesis;
  • o trauma (i.e., fracture or dislocation); and/or
  • o tumor.

The HA Minuteman G3-R MIS Fusion Plate is intended for use with bone graft material and is not intended for stand-alone use. The HA Minuteman G3-R MIS Fusion Plate may be implanted via a minimally invasive lateral approach (L1-S1).

{4}------------------------------------------------

Technological Characteristics

The HA Minuteman G3-R MIS Fusion Plate consists of bilateral Plates and a Body/Post that connects the Plates. The Plate components include several spikes on the face of each Plate for attachment to the spinous processes. The HA Minuteman G3-R device is available in multiple sizes to accommodate varying patient anatomy. The HA Minuteman G3-R is made from Ti6Al4V and Ti6Al4V ELI and has a hydroxyapatite coating on a portion of the implant.

Purpose of 510(k)

The purpose of this 510(k) is to:

  • Introduce a two-part Body/Post component that allows for reduction of the non-load ● bearing lateral profile of the implant remaining in the patient. The outer threaded portion of the Body/Post remains nearly identical to the Body/Post of the predicate device, with the exception of several geometric features that allow the two halves of the Body/Post to be reversibly locked together.
  • Introduce a feature that joins the Spike Plate and Hex Nut together as a single unit. ● This allows the user to hold the implant vertically with the nose of the device pointing towards the ground without the Spike Plate sliding down the Body/Post. This makes loading bone graft material into the device easier, as the user does not need to hold the Spike Plate out of the way.
  • Update Wing geometry for manufacturing and to improve Wing profile in the closed ● and open positions.
  • . Make minor instrumentation updates.

Performance Data

The following tests have been performed on the HA Minuteman G3-R MIS Fusion Plate to evaluate the mechanical performance of the device:

  • Static Axial Compression Test ●
  • o Static Torsion Test
  • Dynamic Axial Compression Test
  • Static Axial Pull-Out Test ●
  • Static Plate Dissociation Test ●
  • Simulated Deployment Testing
  • Custom Body Connection Feature Tests (Torque Test, Axial Locking Feature Test, ● Axial Disengagement Test)
  • . Cadaveric Insertion Testing

In all instances, the HA Minuteman G3-R MIS Fusion Plate functioned as intended and the results observed were as expected.

Bacterial endotoxin testing will also be performed on all batches of sterile packed devices.

{5}------------------------------------------------

Substantial Equivalence

The HA Minuteman G3-R MIS Fusion Plate has the same indications and similar technological characteristics and principles of operation as its predicate device. The minor technological differences between the HA Minuteman G3-R MIS Fusion Plate and its predicate device do not raise different issues of safety or effectiveness, as confirmed through risk analysis and design control activities. Thus, we believe the HA Minuteman G3-R MIS Fusion Plate is substantially equivalent.

Conclusions

The HA Minuteman G3-R MIS Fusion Plate is substantially equivalent to the predicate device.

§ 888.3050 Spinal interlaminal fixation orthosis.

(a)
Identification. A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.(b)
Classification. Class II.