K Number
K163388

Validate with FDA (Live)

Manufacturer
Date Cleared
2017-07-19

(229 days)

Product Code
Regulation Number
892.1710
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Xpress BCD is a reusable device indicated for use with stereotactic biopsy platforms to apply temporary focal mechanical compression to a stereotactic breast biopsy site to achieve hemostasis. Rx Only. Federal (USA) law restricts this device to sale by or on the order of a physician.

Device Description

The XPress BCD device is an accessory used in conjunction stereotactic breast biopsy devices. The XPress BCD is comprised of Delrin® DuPont (acetal homopolymer resin, Delrin 150NC010), or similar material. There are three (3) independent XPress BCD device configurations that are provided as a set: configuration 1 (large flat), configuration 2 (modified pyramidal), and configuration 3 (modified spherical). These different configurations are used based on the biopsy site being treated. The Xpress design includes a mechanical slot for attaching to compression paddles of mammographic x-ray systems. The XPress BCD is used in conjunction with the XPress BCD Cover (K863853; classification JAA; manufactured by Preferred Medical Products, Ducktown, TN). It is used to enclose the XPress BCD for use as intended. The XPress BCD Cover, is a sterile, single-use covering. The XPress BCD Cover is a 4 X 5 inches covering of a 0.002 mm thick polyethylene blend material, and its closure system is a round or oblong synthetic rubber band made of a polyisoprene polymer, measuring 2 X 1/8 X 1/16 inches. The XPress BCD Cover is packaged with the XPress BCD or may be purchased separately.

AI/ML Overview

1. Acceptance Criteria and Reported Device Performance

The document describes a clinical study to evaluate the performance of the XPress BCD Breast Compression Device for achieving hemostasis after stereotactic-guided breast biopsy. The primary clinical performance measure/adverse effect was hematoma formation, defined as a new three-dimensional mass (> 0.52 cm³) detected on standard post-biopsy full-field mammography.

The document does not explicitly state pre-defined acceptance criteria with specific numerical thresholds for hematoma formation rate. However, the study results are presented, demonstrating the device's performance in this aspect.

Table of Acceptance Criteria (Implied) and Reported Device Performance:

Performance Measure (Implied Acceptance Criterion)Reported Device Performance (XPress BCD)
Rate of Immediate Post-Compression Hematoma7.5% (9/120 biopsies) by primary investigators
Rate of Delayed Hematoma (up to one week)3.3% (4/120 biopsies)
Rate of Fat Necrosis0% (0/65 evaluable sites at one year post-biopsy)
Rate of Hematoma (Independent Reviewer A)5.8% (7/120 biopsies)
Rate of Hematoma (Independent Reviewer B)7.5% (9/120 biopsies)
Rate of Fat Necrosis (Independent Reviewers)0% (0/65 evaluable sites)

Note: The document focuses on demonstrating that the device is "as safe and effective as the predicate device" through a clinical study, rather than specific acceptance thresholds.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: The study included 120 biopsies performed on 118 female subjects.
  • Data Provenance: The study was a "prospective historical control single arm clinical study (NCT02327598)" conducted at "one investigational site". The country of origin is not explicitly stated, but the FDA submission suggests it is likely the USA. The study design is prospective.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

  • Number of Experts: The study involved primary investigators and two "independent mammographers" for review.
    • Five investigators and six assistants participated in the study for immediate post-compression hematoma detection.
    • Two independent mammographers reviewed the imaging for hematoma and fat necrosis.
  • Qualifications of Experts: The qualifications of the "primary investigators" are not specified beyond their role in the study. The "independent mammographers" are referred to as such, suggesting they are qualified radiologists specializing in mammography, but their specific years of experience or board certifications are not provided.

4. Adjudication Method for the Test Set

  • The document implies that immediate post-compression hematoma detection was performed by primary investigators.
  • For delayed hematomas and fat necrosis, the "primary study" and "independent mammographers" assessed the mammographic evidence.
  • There is no explicit mention of an adjudication method (like 2+1 or 3+1 consensus) if there were disagreements between the independent mammographers or between the primary study findings and the independent reviewers. Both reviewers' findings for hematoma and fat necrosis are reported separately, suggesting independent assessments without a formal reconciliation process mentioned.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

No, an MRMC comparative effectiveness study was not explicitly stated or described. The study was a "single arm clinical study" focusing on the XPress BCD device's performance after biopsy. There is no comparison of human readers with vs. without AI assistance. The study evaluates a physical device, not an AI algorithm.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

This question is not applicable. The XPress BCD is a physical breast compression device, not an AI algorithm. Therefore, no standalone algorithm performance study would be conducted.

7. The type of ground truth used

The ground truth for hematoma formation was based on:

  • Clinical observation: Primary investigators detected immediate post-compression hematoma.
  • Imaging review: Mammograms were used to detect hematomas and fat necrosis, reviewed by primary study staff and two independent mammographers.
  • Patient follow-up/outcomes data: Phone contact after 24 hours and 7 days, and clinical/ultrasound evaluation for returning patients, were part of the follow-up for adverse events. Hematoma was defined as a "new three dimensional mass (> 0.52 cm') detected on standard post-biopsy full-field mammography." This suggests an imaging-based "ground truth" for the primary outcome.

8. The Sample Size for the Training Set

This question is not applicable. The XPress BCD is a physical device, not a machine learning algorithm. Therefore, there is no "training set" in the context of AI.

9. How the Ground Truth for the Training Set was Established

This question is not applicable as there is no training set for a physical device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features the department's name encircling a symbol of three stylized human profiles. The profiles are stacked and connected by flowing lines, creating a sense of unity and collaboration. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 19, 2017

XPress BCD, LLC Peter Kremers, M.D. 6024 White Flint Drive Frederick, MD 21702

Re: K163388

Trade/Device Name: XPress BCD Breast Compression Device Regulation Number: 21 CFR 892.1710 Regulation Name: Mammographic x-ray System Regulatory Class: Class II Product Code: POY Dated: June 9, 2017 Received: June 12, 2017

Dear Dr. Kremers:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours. Jennifer R. Stevenson -23

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K163388

Device Name XPress BCD®

Indications for Use (Describe)

The Xpress BCD is a reusable device indicated for use with stereotactic biopsy platforms to apply temporary focal mechanical compression to a stereotactic breast biopsy site to achieve hemostasis.

Rx Only. Federal (USA) law restricts this device to sale by or on the order of a physician.

Type of Use (Select one or both, as applicable)
[x] Prescription Use (Part 21 CFR 801 Subpart D)[ ] Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K163388 510(K) SUMMARY

July 17, 2017

1.0 SPONSOR

Peter Kremers, MD (Contact person) 6024 White Flint Drive Frederick, MD 21702 T: 301.401.5853 E: peterkremers@comcast.net

NAME OF DEVICE 2.0

Trade name: XPress BCD® Breast Compression Device Common name: Post Breast Biopsy Hemostatic Breast Compression Device

CLASSIFICATION OF DEVICE 3.0

Classification: Class II Product Code: POY

Regulation Number: 892.1710

Classification Panel: Radiology

4.0 PREDICATE DEVICE

Device Trade Name: Biopsy Digit S and Biopsy Digit SL 510(k) Control Number: K113607 Product Code: IZH Manufacturer: Giotto USA, LLC, Wichita, KS

5.0 INDICATIONS FOR USE

The Xpress BCD is a reusable device indicated for use with stereotactic biopsy platforms to apply temporary focal mechanical compression to a stereotactic breast biopsy site to achieve hemostasis. Rx Only. Federal (USA) law restricts this device to sale by or on the order of a physician.

DEVICE DESCRIPTION 6.0

The XPress BCD device is an accessory used in conjunction stereotactic breast biopsy devices.

The XPress BCD is comprised of Delrin® DuPont (acetal homopolymer resin, Delrin 150NC010), or similar material. There are three (3) independent XPress BCD device configurations that are provided as a set: configuration 1 (large flat), configuration 2 (modified pyramidal), and configuration 3 (modified spherical). These different configurations are used based on the biopsy site being treated. The Xpress design includes a mechanical slot for attaching to compression paddles of mammographic x-ray systems

{4}------------------------------------------------

The XPress BCD is used in conjunction with the XPress BCD Cover (K863853; classification JAA; manufactured by Preferred Medical Products, Ducktown, TN). It is used to enclose the XPress BCD for use as intended. The XPress BCD Cover, is a sterile, singleuse covering. The XPress BCD Cover is a 4 X 5 inches covering of a 0.002 mm thick polyethylene blend material, and its closure system is a round or oblong synthetic rubber band made of a polyisoprene polymer, measuring 2 X 1/8 X 1/16 inches. The XPress BCD Cover is packaged with the XPress BCD or may be purchased separately.

7.0 BIOCOMPATIBILITY

The Xpress BCD device does not come in direct or indirect contact with the patient as intended for use. Biocompatibility testing has not been performed on the XPress BCD.

8.0 STERILIZATION AND CLEANING

The XPress BCD is a non-sterile, reusable device. The Xpress BCD device does not come in direct or indirect contact with patients during breast biopsy procedures. The Xpress BCD compression device is to be covered by a sterile, single-use XPress BCD Cover prior to use as intended.

The XPress BCD device, without protective covering, must be cleaned, disinfected, and inspected prior to use on a patient, including between biopsies on the same patient, using an EPA-registered disinfectant that achieves intermediate level of disinfection with a tuberculocidal claim.

9.0 CLINICAL TESTING

A prospective historical control single arm clinical study (NCT02327598) was conducted at one investigational site to evaluate the performance of the XPress BCD breast compression device to obtain hemostasis after stereotactic-guided breast biopsy following 120 biopsies in 118 female subjects with mammographic findings that warranted stereotactic biopsy. Five investigators and six assistants participated in this study; the 118 enrolled and treated subjects met the following inclusion criteria:

  • Adult women: eighteen years of age and older. ●
  • . Sites of calcification (size not restricted) and/or breast lesions (masses) ≤ 1.5 cm in diameter deemed suspicious by radiologist/ breast surgeon.
  • Provided written informed consent.

None of the 118 enrolled and treated subjects met the study's criteria for exclusion:

  • Pregnancy.
  • Lactation.
  • Cognitive impairment.
  • . On active blood thinners including, but not limited to, warfarin, clopidogrel and heparin. Low dose aspirin or non-steroidal medication are not subject to this requirement.
  • A known bleeding disorder or uncorrected coagulopathy with a platelet count of ≤ . 80,000 and/or an international normalized ratio of ≥ 1.5.

{5}------------------------------------------------

  • Note : Coagulation parameters were not routinely performed unless suspicion of . coagulopathy including known deficiencies such as Factor VIII and XI, Diffuse Intravascular Coagulation, Uremia, Liver Failure, Myeloproliferative Disorder, Alpha 2 Antiplasmin deficiency, Monoclonal Gammopathy, and Lupus Anticoagulant).
    Follow-up after discharge included phone contact after 24 hours and 7 days. Patients were asked to return if they developed significant bruising greater than 3 cm in diameter, a new lump at the biopsy site, increasing pain, or signs of infection including redness, increased warmth, or increased temperature. Returning patients out to one week were evaluated clinically as well as with ultrasound of the biopsy site.

Additionally, all 65 of the 120 original biopsy sites deemed evaluable for evidence of fat necrosis were reviewed at one year. Evaluable sites required the absence of interval surgical biopsy/lumpectomy and mammograms available for review at one year following biopsy and mechanical compression.

The primary clinical performance measure / adverse effect were hematoma formation. Hematoma was defined as a new three dimensional mass (> 0.52 cm') detected on standard post-biopsy full- field mammography. Incidence and size of delayed hematomas were also noted. Unanticipated adverse events were defined as any serious adverse effect on health, safety, or any life-threatening problem or death caused by or associated with the device. No unanticipated adverse events were noted during the course of the study and follow-up evaluation.

Table 9-1 summarizes demographics and performance measures for 120 biopsy procedures performed on 118 enrolled and treated clinical study patients. Immediate post-compression hematoma was detected by primary investigators in 9/120 biopsies (7.5%). Additional delayed hematomas detected out to one week after biopsy were found following 4/120 (3.2%) of the biopsies. No mammographic evidence of fat necrosis was noted at one-year post-biopsy in a subgroup of 65 subjects who had not been subject to open biopsy or lumpectomy and were evaluated with mammography at one year post-biopsy. All biopsy sites meeting the criteria for evaluation were assessed for mammographic for evidence of fat necrosis by both the primary study and independent mammographers.

Table 9-1. Demographics and Performance Measures across 118 Patients and 120 Biopsy Sites

MeasureAge(yrs)BreastDensityBreastCompression(mm)Calcs/Other*CoresCompressionTime (min)HematomaHematomaSize (cm³)DelayedHematomaFatNecrosis
Total80/ 409/120(7.5%)4/120 (3.3%)0/65
Average572659.8114.4
Range34-791-430-1035-2410-201.0-12.1Occurred ondays 2, 5, 7, 7

*80 sites were clusters of calcification. 40 sites were other targets, including masses, densities, and sites of architectural distortion.

Table 9-2 provides the results from two independent mammographers' examinations; one found hematomas on immediate post-biopsy imaging following 7/120 (5.8%) of biopsy procedures and the other in 9/120 (7.5%). Neither noted any mammographic evidence of fat necrosis in the 65 subjects who did not proceed to open biopsy or lumpectomy and for whom mammography at one year post-biopsy was available for review.

{6}------------------------------------------------

ReviewerHematomaAverageHematomaSize (cm3)Fat Necrosis
A7/120(5.80%)5.60/65
B9/120(7.5%)5.70/65
Table 9-2. Independent Imaging Review across 118 Patients and 120 Biopsy Sites

Of the 120 original biopsy sites. 65 were evaluable for evidence of fat necrosis as surgical biopsy/lumpectomy had not been performed and mammograms were available for review at one year following biopsy and mechanical compression.

Table 9-3 presents assessments reported by study staff and the 118 subjects. Most assessments are reported per biopsy site (N=120), however, post-study satisfaction was reported per patient (N=118).

Table 9-3. Staff and Subject Reported Outcomes across 118 Patients and 120 Biopsy Sites

Staff-Reported EvaluationSubject-Reported PainSubject-Reported Satisfaction
ProcedureCompressionPost0-hour24-hour1-7 day
# BiopsySites120120120120120115/118patients113/118patients
Average520.90.31.31.31.2
CriteriaRang-5= extreme degradation5= extreme enhancement0 = no pain10 = sever pain1 = great experience5 = poor experience

Note: Staff evaluation compared their impressions on the use of mechanical compression on workflow relative to the use of standard manual compression using a 10 point scale (-5 extreme degradation to +5 extreme enhancement of workflow). The effect on workflow closely mirrored staff estimates of net time saved or lost (in minutes) relative to standard manual compression (assuming that they would have been responsible for performing manual breast compression).

SUBSTANTIAL EQUIVALENCE FEATURES 10.0

During breast biopsy procedures, the XPress BCD is used to apply direct mechanical compression to the biopsy site, as a replacement for standard manual compression. The XPress BCD is inserted into the stereo guide compression paddle, and pressure is applied mechanically by the paddle and attached XPress BCD device. Before insertion into the compression paddle, the XPress BCD is placed inside, and enclosed by, a sterile XPress BCD Cover, so that the device does not come in direct contact with the patient. Further, sterile 4 X 4 gauze pads are placed between the covered XPress BCD device and the patient, which is an additional barrier to direct contact with the patient.

The XPress BCD Breast Compression Device, is substantially equivalent to the legally marketed, commercially available Biopsy Digit S and Biopsy Digit SL (K113607). manufactured by Giotto USA, LLC, Wichita, KS 67226. The XPress device does not introduce any technological characteristic differences that raise questions of safety or effectiveness; to establish this, a list of comparative features and their similarities and differences is provided in the table below:

{7}------------------------------------------------

Table 10 -1. Comparison of Similarities and Differences between Biopsy Digit S and
Biopsy Digit SL (K113607) and XPress BCD
COMPARISONPREDICATE DEVICE:BIOPSY DIGIT S ANDBIOPSY DIGIT SLSUBJECT DEVICE:XPRESS BCDSUBSTANTIALLYEQUIVALENT
Intended UseThe Biopsy Digit S andSL are intended to beused for mammographicprocedures requiringstereotactic guidance,such as fine needleaspiration, needle biopsyand guide wire placement.The XPress BCD is intendedto be used for mammographicprocedures requiringstereotactic guidance. Thedevice is indicated for focalmechanical compression ofbiopsy sites to achieve andmaintain hemostasis.Yes. Intended for use inmammographicprocedures requiringstereotactic guidance.SE demonstrated bywell-controlled clinicaltrial.
IntendedPopulationThe target populationconsists of female patientsundergoing stereotacticguided mammographicbiopsy procedures.The target population consistsof female patients undergoingstereotactic guidedmammographic biopsyprocedures.Yes.
Prescription UseYesYesYes
Device ClassClass IIClass IIYes
DeviceSystem, X-Ray,MammographicPost Breast Biopsy HemostaticBreast Compression DeviceYes. RegulationDescription isMammographic x-raysystem. SEdemonstrated by well-controlled clinical trial.
Regulation Number21 CFR 892.1710(a) Identification. Amammographic x-raysystem is a deviceintended to be used toproduce radiographs ofthe breast. This generictype of device mayinclude signal analysisand display equipment,patient and equipmentsupports, componentparts, and accessories.(b) Classification. ClassII.21 CFR 892.1710(a) Identification. Amammographic x-ray systemis a device intended to be usedto produce radiographs of thebreast. This generic type ofdevice may include signalanalysis and displayequipment, patient andequipment supports,component parts, andaccessories.(b) Classification. Class II.Yes
Product CodeIZHPOYYes. Product Codesincluded in RegulationNumber 892.1710. SEdemonstrated by well-controlled clinical trial.
Used formammographicproceduresrequiringstereotacticguidanceYesYesYes
RegulationDescriptionMammographic x-raysystem.Mammographic x-ray system.Yes
Provided Sterile orNon-SterileNon-sterileNon-sterileYes
COMPARISONPREDICATE DEVICE:BIOPSY DIGIT S ANDBIOPSY DIGIT SLSUBJECT DEVICE:XPRESS BCDSUBSTANTIALLYEQUIVALENT
User interfacesoftwareYesNoYes. No differences insafety or effectivenessare introduced
Electro-mechanicalmechanismsYesYes*Yes.
Computer interfaceto store imagesYesNoYes. No differences insafety or effectivenessare introduced
Computer interfaceto determinecoordinates forneedle positioningYesNoYes. No differences insafety or effectivenessare introduced

{8}------------------------------------------------

"Though the subject device does not contain any electro-mechanisms, when attached to a stereotactic breast biopsy device it may be moved either manually or electro-mechanically via that device.

11.0 SUBSTANTIAL EQUIVALENCE

The XPress BCD, and the claimed predicate device, the Biopsy Digit S and Biopsy Digit SL device, are accessories used during stereotactic-guided breast biopsy procedures including breast biopsies. The XPress BCD device and the predicate device have the same general intended use – they are accessories for mammographic procedures requiring stereotactic guidance including breast biopsies. The XPress device does not introduce any technological characteristic differences that raise questions of safety or effectiveness. The XPress BCD device has been demonstrated to be as safe and effective as the predicate device for intended use in a prospective, single site single arm historical control clinical study (NCT02327598). There are no additional significant questions of safety or effectiveness.

§ 892.1710 Mammographic x-ray system.

(a)
Identification. A mammographic x-ray system is a device intended to be used to produce radiographs of the breast. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.