K Number
K163288

Validate with FDA (Live)

Date Cleared
2018-01-18

(422 days)

Product Code
Regulation Number
876.5320
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Pelvifine Pelvic Muscle Trainer (Model: KM518) is a non-implanted muscle stimulator designed to treat stress, urge and/or mixed urinary incontinence in women. It applies stimulation to the pelvic floor muscles and surrounding structures to improve strength and support.

Device Description

The Pelvifine Pelvic Muscle Trainer is a product designed to give women the opportunity to strengthen their pelvic floor muscles. This device is a dual-channel stimulator combining several treatment programs into one unit. Channel B is for the Junior User (who use the device no more than 5 times) and Channel A is for the Senior User (who use the device more than 5 times). It offers full control of Pulse Widths, Rates, Ramp up times, and Work/Rest cycles.

AI/ML Overview

The provided text is a 510(k) Summary for the Pelvifine Pelvic Muscle Trainer. It describes the device, its intended use, and a comparison to predicate devices to demonstrate substantial equivalence for regulatory approval.

However, this document does not contain information about clinical study acceptance criteria, device performance results from a clinical study, sample sizes for test or training sets, ground truth establishment, expert qualifications, or MRMC studies. The "Test Summary" section explicitly states that the device was evaluated by lab bench testing and usability testing, not clinical studies assessing its efficacy in treating urinary incontinence.

Therefore, I cannot fulfill your request for the comprehensive information about acceptance criteria and a study proving the device meets them, as the provided document does not contain this type of data.

The relevant section, "6. Test Summary," states:
"The Pelvifine Pelvic Muscle Trainer (Model: KM518) has been evaluated the safety and performance by lab bench testing and usability testing as following:

  • Electrical safety test according to IEC 60601-1 and IEC 60601-2-10 standards
  • Electromagnetic compatibility test according to IEC 60601-1-2 standard
  • Biocompatibility test according to ISO 10993-5 and ISO 10993-10 standards
  • Usability test according to IEC 62366 standard
  • Software verification and validation test according to the requirements of the FDA "Guidance for Pre Market Submissions and for Software Contained in Medical Devices"
  • The waveform test report has also been conducted to characterize the output specifications of the device according to Guidance for Transcutaneous Electrical Nerve Stimulator for Pain Relief Intended for Over the Counter Use and Guidance for Powered Muscle Stimulator for Muscle Conditioning"

This clearly indicates a focus on technical and safety performance, not clinical efficacy. The substantial equivalence argument relies on the similarity of the new device's technical specifications and intended use to those of already approved predicate devices, rather than on proving its therapeutic effectiveness through a new clinical trial.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION".

January 18, 2018

Shenzhen Konmed Technology Co., Ltd. % Cassie Lee Guangzhou Glomed Biological Technology Co., Ltd. Suite 306, Kecheng Mansion, No.121 Science Road Guangzhou Science Park Guangzhou, 510663 CN

Re: K163288

Trade/Device Name: Pelvifine Pelvic Muscle Trainer Regulation Number: 21 CFR§ 876.5320 Regulation Name: Nonimplanted Electrical Continence Device Regulatory Class: II Product Code: KPI Dated: December 19, 2017 Received: January 16, 2018

Dear Cassie Lee:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Image /page/1/Picture/6 description: The image shows the name "Benjamin R. Fisher -S" in a large, bold font. The text is black and appears to be against a white background. The letters are clearly legible and evenly spaced.

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K163288

Device Name

Pelvifine Pelvic Muscle Trainer, Model: KM518

Indications for Use (Describe)

The Pelvifine Pelvic Muscle Trainer (Model: KM518) is a non-implanted muscle stimulator designed to treat stress, urge and/or mixed urinary incontinence in women. It applies stimulation to the pelvic floor muscles and surrounding structures to improve strength and support.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Sponsor:Shenzhen Konmed Technology Co., Ltd.
Subject Device:Pelvic Muscle Trainer, Model: KM518
Document Name:510(k) Summary

Chapter 5. 510(k) Summary

510(k) Summary

This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.

1. Submitter's Information

510(k) Owner's Name: Shenzhen Konmed Technology Co., Ltd. Establishment Registration Number: Applying Address: 703B, Haide building A, Nanxin Road, Nanshan, Shenzhen, China Tel: +86-755-86704558 Fax: +86-755-86704558 Contact Person: Tony (Manager) Email: 2519021651@qq.com

Application Correspondent:

Contact Person: Ms. Cassie Lee Guangzhou GLOMED Biological Technology Co., Ltd. Address: Suite 306, Kecheng Mansion, No.121 Science Road, Guangzhou Science Park, Guangzhou 510663, China Tel: +86-20-61099984 Email: regulatory@glomed-info.com

2. Subject Device Information

Trade Name:Pelvifine Pelvic Muscle Trainer
Common Name:Pelvic Floor Muscle Stimulator
Classification name:Stimulator, Electrical, Non-Implanted for Incontinence
Review Panel:Gastroenterology/Urology
Product Code:KPI
Regulation Class:II
Regulation Number:21 CFR 876.5320

3. Predicate Device Information

SponsorInControl Medical, LLCInternational TradeGroup, IncorporatedInControl Medical, LLC
Device Name andModelInToneYARLAPApexM
510(k) NumberK110179K141643K150183

{4}------------------------------------------------

Sponsor:Shenzhen Konmed Technology Co., Ltd.
Subject Device:Pelvic Muscle Trainer, Model: KM518
Document Name:510(k) Summary
Product CodeKPIKPIKPI
Regulation Number21 CFR 876.532021 CFR 876.532021 CFR 876.5320
Regulation ClassIIIIII

2. Device Description

The Pelvifine Pelvic Muscle Trainer is a product designed to give women the opportunity to strengthen their pelvic floor muscles.

This device is a dual-channel stimulator combining several treatment programs into one unit. Channel B is for the Junior User (who use the device no more than 5 times) and Channel A is for the Senior User (who use the device more than 5 times). It offers full control of Pulse Widths, Rates, Ramp up times, and Work/Rest cycles.

5. Intended Use / Indications for Use

The Pelvifine Pelvic Muscle Trainer (Model: KM518) is a non-implanted muscle stimulator designed to treat stress, urge and/or mixed urinary incontinence in women. It applies stimulation to the pelvic floor muscles and surrounding structures to improve strength and support.

6. Test Summary

The Pelvifine Pelvic Muscle Trainer (Model: KM518) has been evaluated the safety and performance by lab bench testing and usability testing as following:

  • � Electrical safety test according to IEC 60601-1 and IEC 60601-2-10 standards
  • � Electromagnetic compatibility test according to IEC 60601-1-2 standard
  • � Biocompatibility test according to ISO 10993-5 and ISO 10993-10 standards
  • � Usability test according to IEC 62366 standard
  • � Software verification and validation test according to the requirements of the FDA "Guidance for Pre Market Submissions and for Software Contained in Medical Devices"
  • � The waveform test report has also been conducted to characterize the output specifications of the device according to Guidance for Transcutaneous Electrical Nerve Stimulator for Pain Relief Intended for Over the Counter Use and Guidance for Powered Muscle Stimulator for Muscle Conditioning

7. Comparison to predicate device and conclusion

The technological characteristics, features, specifications, materials, mode of operation, and intended use of subject device is substantially to the predicate devices identified above.

The differences between the subject device and predicate devices do not raise new issues of safety or effectiveness.

Elements ofComparisonSubject DevicePredicate Device EquivalentRemark
Device Name andPelvic Muscle Trainer,InToneYARLAPApexM

{5}------------------------------------------------

Sponsor:Shenzhen Konmed Technology Co., Ltd.
Subject Device:Pelvic Muscle Trainer, Model: KM518
Document Name:510(k) Summary
Elements ofComparisonSubject DevicePredicate Device EquivalentRemark
ModelModel: KM518
510(k) NumberK163288K110179K141643K150183
Intended Use &Indications for UsePelvic Muscle Trainer(Model: KM518) is anon-implanted musclestimulator designed totreat stress, urge and/ormixed urinaryincontinence in women.It applies stimulation tothe pelvic floor musclesand surroundingstructures to improvestrength and support.The InControl InTonedevice is a non-implanted electricalstimulator indicated foruse in the treatment offemale urinaryincontinence. It applieselectrical stimulation tothe pelvic floormusculature andsurrounding structures.It is intended for acuteand ongoing treatmentof mixed urinaryincontinence where thefollowing results mayimprove urinary control:strengthening of pelvicfloor muscles andinhibition of thedletrusor musclethrough reflexivemechanisms. Thebiofeedback feature canbe used for muscle re-education purposes.The device is intendedto provide electricalstimulation andneuromuscular re-education for thepurpose of rehabilitationof weak pelvic floormuscles for thetreatment of stress, urgeand mixed urinaryincontinence in womenand to maintain urinarycontinence in women.ApexM is a non-implanted musclestimulator designed totreat stress, urge and/ormixed urinaryincontinence in women.It applies stimulation tothe pelvic floor musclesand surroundingstructures to improvestrength and support.SENote:IdenticaltoApexM
Power Source(s)9Vdc Battery4/5 AA nickel metalhydride battery--4 x AAA AlkalinebatterySENote1
Method of LineCurrent IsolationType BF Applied PartType BF Applied PartType BF Applied PartType BF Applied PartSE
PatientLeakageCurrentNCSFC< 100 $ μ $ A< 500 $ μ $ AN/A (battery)--N/A (battery)SENote 1
Number of OutputChannels2 (only the selected onechannel will work duringoperation)1--1SENote 2
Number of OutputModes121--1SENote 2
Output IntensityLevel100------SENote 2
Timer Range15~20 minutes12 minutes--10~15 minutesSENote 2
WeightMain Unit: 300gProbe: 200g------SENote 2
DimensionsMain Unit:150mm x 73mm x20mmProbe:145mm x 25mm xInsertion Unit: 12.2" x2.5" x 4.0"Inflatable Probe: 5.5" x2.2" x 2.5"--Insertion Unit :12.2"x 2.5"x 4.0"Inflatable Probe:5.5"x 2.2"x 2.5"SENote 2

{6}------------------------------------------------

Sponsor:Shenzhen Konmed Technology Co., Ltd
Subject Device:Pelvic Muscle Trainer, Model: KM518
Document Name:510(k) Summary
Elements ofComparisonSubject DevicePredicate Device EquivalentDevice EquivalentRemark
25mm
Electrode SurfaceArea7.85cm² (x 2)10.5cm² (x 2)6.4cm² (x 2)6.00cm² ± 0.5cm² (x 2) SE Note 3
Housing Materialsand ConstructionMain unit:ABS plasticProbe Material: ABSplastic, Stainless steelABS plastics--Main unit:ABS plasticInsertion Material:Silicone, Plastics SE Note 4
Waveform andShapeRectangular pulses,BiphasicRectangular pulses,BiphasicRectangular pulses,BiphasicRectangular pulses,Monophasic SE Note:Identical to InTone
Maximum OutputVoltage37Vdc @500Ω50Vdc @500Ω--40Vdc @500Ω SE Note 3
Maximum OutputCurrent<80 mA @500Ω100mA @500Ω80mA±10% @ 500Ω50mA±10% @ 2kΩ19mA±10% @10kΩ80mA @500Ω SE Note 3
Pulse Duration150μs, 200μs, 220μs200μs200-250μs200μs SE Note 3
Pulse frequency10~40Hz50Hz10~35Hz13, 50Hz SE Note 3
Net Charge (perpulse)0 µC @ 500Ω--0μC @ 500Ω-- SE Note 3
Maximum PhaseCharge19.8 µC @ 500Ω20μC @ 500Ω20μC @ 500Ω16μC @ 500Ω SE Note 3
Maximum CurrentDensity9.48 mA/cm²4.7 mA/cm²12.5mA/cm²(Vaginal)13.3 mA/cm² SE Note 3
Maximum AveragePower Density3.09 mW/cm²4.8 mW/cm²3.5 mW/cm²5.33 mW/cm² SE Note 3
BiocompatibilityAll user directlycontacting materials arecompliance withISO10993-5 andISO10993-10requirements.All user directlycontacting materials arecompliance withISO10993-5 andISO10993-10requirements.All user directlycontacting materials arecompliance withISO10993-5 andISO10993-10requirements.All user directlycontacting materials arecompliance withISO10993-5 andISO10993-10requirements. SE
Electrical SafetyComply with IEC 60601-1 and IEC 60601-2-10Comply with IEC60601-1 and IEC60601-2-10Comply with IEC 60601-1 and IEC 60601-2-10Comply with IEC60601-1 and IEC60601-2-10 SE
EMCComply with IEC 60601-1-2Comply with IEC60601-1-2Comply with IEC 60601-1-2Comply with IEC60601-1-2 SE
UtilityPredicate deviceK141643Subject deviceRate HzPulse usWork Sec.Rest Sec.Time Min.*Max. W/cm²80mAAve. W/cm²30mAAve. W/cm²45mAmaximum phase charge
---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

{7}------------------------------------------------

Sponsor:Shenzhen Konmed Technology Co., Ltd.
Subject Device:Pelvic Muscle Trainer, Model: KM518
Document Name:510(k) Summary
MixedMode 1--1220055151.20.170.3820 C @ 500Ω
MixedMode 2--2025088202.50.350.7920 C @ 500Ω
Mixed--P041022055201.10.310.4719.8 C @ 500Ω
Mixed--P051022055201.10.310.4719.8 C @ 500Ω
Mixed--P06_11022044201.020.310.4719.8 C @ 500Ω
Mixed--P06_220220441.20.630.9419.8 C @ 500Ω
Mixed--P07_12022055201.20.630.9419.8 C @ 500Ω
Mixed--P07_220220551.20.630.9419.8 C @ 500Ω
Mixed--P07_320220551.20.630.9419.8 C @ 500Ω
Mixed--P07_410220441.20.320.4719.8 C @ 500Ω
Mixed--P082022055201.20.630.9419.8 C @ 500Ω
UrgeMode 3--12200510151.20.170.3820 C @ 500Ω
UrgeMode 4--10200612201.00.140.3220 C @ 500Ω
Urge--P0310200612150.90.280.4219.8 C @ 500Ω
StressMode 5--12250515151.50.210.4720 C @ 500Ω
StressMode 6--35200618203.50.51.1120 C @ 500Ω
Stress--P0140220615202.31.231.8719.8 C @ 500Ω
Stress--P02102205152010.310.4619.8 C @ 500Ω
  • The treatment times listed for the subject device are for Channel B has a 3-minute treatment time.

Comparison in Detail(s):

Note 1:

Although the "Power Source(s)", "Patient Leakage Current", "Weight", and "Dimensions", a little different from the predicate devices, the differences will not raise any safety or effectiveness issue.

Note 2:

Although the "Number of Output Modes" "Output Intensity Level", "Timer Range", "Weight" and "Dimensions" of subject device are different from the predicate devices, the differences will not raise any safety or effectiveness issue.

Note 3:

Although the "Electrode Surface Area", "Maximum Output Voltage", "Maximum Output Current", "Pulse Duration", "Pulse Frequency", "Net Charge (per pulse)", "Maximum Phase Charge", "Maximum Current Density", "Maximum Average Power Density" are a little different from the predicate devices, the differences will not raise any safety or effectiveness issue.

Note 4:

Although the "Housing Materials and Construction" are a little different from the predicate devices, the differences will not raise any safety or effectiveness issue.

Note 5:

The stimulation parameters for "specific urinary incontinence type" of subject device and predicate device are very similar , so the differences will not significantly impact safety and effectiveness.

Finial Conclusion:

The subject device "Pelvifine Pelvic Muscle Trainer" is Substantial Equivalent to the predicate devices.

8. Date of the summary prepared: December 19, 2017

§ 876.5320 Nonimplanted electrical continence device.

(a)
Identification. A nonimplanted electrical continence device is a device that consists of a pair of electrodes on a plug or a pessary that are connected by an electrical cable to a battery-powered pulse source. The plug or pessary is inserted into the rectum or into the vagina and used to stimulate the muscles of the pelvic floor to maintain urinary or fecal continence. When necessary, the plug or pessary may be removed by the user. This device excludes an AC-powered nonimplanted electrical continence device and the powered vaginal muscle stimulator for therapeutic use (§ 884.5940).(b)
Classification. Class II (performance standards).