K Number
K163258

Validate with FDA (Live)

Manufacturer
Date Cleared
2017-01-25

(65 days)

Product Code
Regulation Number
888.3040
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CrossFT™ Knotless Suture Anchor with Disposable Driver is intended to reattach soft tissue to bone in orthopedic surgical procedures. The device may be used in either arthroscopic or open surgical procedures. After the suture is anchored to the bone, it may be used to reattach soft tissue, such as ligaments, tendons, or joint capsules to the bone. The suture anchor system thereby stabilizes the damaged soft tissue, in conjunction with appropriate postoperative immobilization, throughout the healing period.

Device Description

The CrossFT™ Knotless Suture Anchors with Disposable Driver are sterile, single use devices. The CrossFT™ Knotless Suture Anchors are manufactured from PolyEtherEtherKetone (PEEK™). The anchors are provided sterile, single use and preloaded on a disposable driver. The anchors are available in three sizes and nine configurations.

AI/ML Overview

The provided document describes the "CrossFT™ Knotless Suture Anchor with Disposable Driver" and its substantial equivalence to a predicate device (K091549). This is a medical device, not an AI/ML device, and therefore the concepts of acceptance criteria related to algorithmic performance (like sensitivity, specificity, MRMC studies, standalone performance, ground truth establishment by experts, training/test set sizes, and data provenance) are not applicable in this context.

The document focuses on demonstrating substantial equivalence for a medical device through engineering and biological performance testing.

Here's an interpretation of the relevant information provided, focusing on what is available:

1. Table of Acceptance Criteria and Reported Device Performance

The document doesn't explicitly state quantitative acceptance criteria in a table format with corresponding reported performance values for each criterion. It broadly mentions completed performance testing to demonstrate the device performs as intended and is substantially equivalent.

However, based on the types of testing listed, we can infer general areas of acceptance. The "Reported Device Performance" is stated generally as having "met the endotoxin limits" and that the testing "demonstrates that the CrossFT™ Knotless Suture Anchor with Disposable Driver performs as intended and is substantially equivalent to the predicate device."

Acceptance Criteria (Inferred from testing types)Reported Device Performance
ReliabilityPerformed as intended
Packaging IntegrityPerformed as intended
Ultimate Fixation StrengthPerformed as intended
Cyclic Loading PerformancePerformed as intended
Sterilization EfficacyPerformed as intended
Pyrogenicity (Bacterial Endotoxin)Met endotoxin limits
Verification Testing (General Device Function)Performed as intended
Transportation StabilityPerformed as intended
BiocompatibilityPerformed as intended
User Validation (Usability)Performed as intended
Shelf-lifePerformed as intended

2. Sample Size for Test Set and Data Provenance

This information is not provided in the document. For a physical medical device, "test set" typically refers to the number of units or samples subjected to each performance test. The document also does not specify the country of origin of data or whether tests were retrospective/prospective.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

This information is not applicable for this type of medical device submission. "Ground truth" established by experts is a concept central to evaluating diagnostic or AI/ML device performance. For this physical device, "ground truth" relates to engineering specifications, material properties, and biological safety standards.

4. Adjudication Method for the Test Set

This information is not applicable for this type of medical device submission. Adjudication methods (like 2+1, 3+1) are used to resolve disagreements among human readers/experts when establishing ground truth for diagnostic or AI/ML performance.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

This is not applicable. MRMC studies are used to evaluate the comparative effectiveness of diagnostic methods (e.g., human readers with and without AI assistance) on a set of cases. This document describes a physical surgical implant.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) was Done

This is not applicable. This refers to algorithmic performance without human intervention, which is relevant for AI/ML devices.

7. The Type of Ground Truth Used

The "ground truth" for this device's performance would be defined by engineering specifications, material science standards, and established biological safety standards. For example, the "ultimate fixation strength" would have a specific quantifiable standard that the device must meet, derived from biomechanical requirements. The "met the endotoxin limits" is an example of meeting a specific biological safety standard, which acts as a "ground truth" for endotoxin levels.

8. The Sample Size for the Training Set

This is not applicable. "Training set" refers to data used to train an AI/ML algorithm. This document describes a physical medical device.

9. How the Ground Truth for the Training Set Was Established

This is not applicable for the reasons stated above.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 25, 2017

ConMed Corporation Ms. Nyrobia Freeman Regulatory Affairs Specialist 525 French Road Utica, New York 13502

Re: K163258

Trade/Device Name: CrossFT10 Knotless Suture Anchor with Disposable Driver Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: MBI Dated: December 29, 2016 Received: January 5, 2017

Dear Ms. Freeman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known) K163258

Device Name

CrossFTTM Knotless Suture Anchor with Disposable Driver

Indications for Use (Describe)

The CrossFT™ Knotless Anchor with Disposable Driver is intended to reattach soft tissue to bone in orthopedic surgical procedures. The device may be used in either arthroscopic or open surgical procedures. After the sutured to the bone, it may be used to reattach soft tissue, such as ligaments, tendons, or joint capsules to the bone. The suture anchor system thereby stabilizes the damaged soft tissue, in conjunction with appropriative immobilization, throughout the healing period.

Type of Use (Select one or both, as applicable)

❌ Prescription Use (Part 21 CFR 801 Subpart D)
❍ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Conmed Corporation. The logo features a blue square with a white curved shape inside on the left. To the right of the square is the word "CONMED" in black, with the word "CORPORATION" in smaller, blue letters underneath.

510(k) SUMMARY

In accordance with the requirements of the Safe Medical Device Act of 1990 and 21 CFR 807.92, CONMED Corporation is hereby submitting the 510(k) Summary of Safety and Effectiveness for 510(k) Number K163258.

I. SUBMITTER

CONMED Corporation 11311 Concept Blvd Largo, Florida 33773

Phone: 727-399-5416 Fax: 727-399-5264

Contact Person: Nyrobia Freeman Date Prepared: November 18, 2016

II. DEVICE NAME

Device Name:CrossFT™ Knotless Suture Anchor with Disposable Driver
Common Name:Nonabsorbable Suture Anchor
Classification Name:Fastener, Fixation, Non-degradable, Soft-Tissue
Regulatory Class:Class II, per 21 CFR Part 888. 3040
Product Codes:MBI

III. PREDICATE/ LEGALLY MARKET DEVICE

Device Name:CrossFTTM Suture Anchor with Disposable Drive
Company Name:CONMED Linvatec
510(k) #:K091549

IV. DEVICE DESCRIPTION

The CrossFT™ Knotless Suture Anchors with Disposable Driver are sterile, single use devices. The CrossFT™ Knotless Suture Anchors are manufactured from PolyEtherEtherKetone (PEEK™). The anchors are provided sterile, single use and preloaded on a disposable driver. The anchors are available in three sizes and nine configurations. (See table 1)

Catalog NumberDescription
CFK-40CrossFTTM Knotless 4.0 mm Suture Anchor
CFK-40TACrossFTTM Knotless 4.0 mm Suture Anchor with one 2 mm Hi-Fi ® Tape (Blue)

Table 1: CrossFT™ Knotless Suture Anchors with Disposable Driver sizes and configurations

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for CONMED Corporation. The logo consists of a blue square with a white abstract shape on the left, and the word "CONMED" in black capital letters on the right. Below the word "CONMED" is the word "CORPORATION" in smaller black capital letters, underlined by a thin blue line. The logo is clean and professional, and the colors are simple and effective.

CFK-40TBCrossFTTM Knotless 4.0 mm Suture Anchor with one 2 mm Hi-Fi® Tape (White/Black)
CFK-475SCrossFTTM Knotless 4.75 mm Suture Anchor
CFK-475STACrossFTTM Knotless 4.75 mm Suture Anchor with one 2 mm Hi-Fi Tape (Blue)
CFK-475STBCrossFTTM Knotless 4.75 mm Suture Anchor with one 2 mm Hi-Fi Tape (White/Black)
CFK-55SCrossFTTM Knotless 5.5 mm Suture Anchor
CFK-55STACrossFTTM Knotless 5.5 mm Suture Anchor with one 2 mm Hi-Fi® Tape (Blue)
CFK-55STBCrossFTTM Knotless 5.5 mm Suture Anchor with one 2 mm Hi-Fi® Tape (White/Black)

V. INTENDED USE/ INDICATIONS FOR USE

The CrossFT™ Knotless Suture Anchor with Disposable Driver is intended to reattach soft tissue to bone in orthopedic surgical procedures.

The device may be used in either arthroscopic or open surgical procedures. After the suture is anchored to the bone, it may be used to reattach soft tissue, such as ligaments, tendons, or joint capsules to the bone. The suture anchor system thereby stabilizes the damaged soft tissue, in conjunction with appropriate postoperative immobilization, throughout the healing period.

COMPARISON OF THE TECHNOLOGICAL CHARACTERISTICS WITH THE VI. PREDICATE DEVICE

The CrossFTTM Knotless Suture Anchor with Disposable Driver is substantially equivalent in design, materials, intended use, principles of operation, and technical characteristics to the predicate CrossFT™ Suture Anchors with Disposable Driver and raises no new issues of safety or effectiveness.

The similarities and difference between the predicate and proposed Nonabsorbable Suture Anchors are the following-

CrossFT TM Knotless Suture Anchor withDisposable DriverProposedConMed Linvatec Soft Tissue to Bone System(CrossFT TM Suture Anchor with Disposable Driver)Predicate
ManufacturerCONMED CorporationConMed Linvatec
510kPendingK091549
Product CodeMBIMBI
ClassificationIIII
Intended UseThe CrossFT TM Knotless Suture Anchor isintended to reattach soft tissue to bone inorthopedic surgical procedures.The CrossFT TM Suture Anchor is intended toreattach soft tissue to bone in orthopedic surgicalprocedures.
Indication for UseThe device may be used in either arthroscopicor open surgical procedures. After the suture isanchored to the bone, it may be used to reattachsoft tissue, such as ligaments, tendons, or jointcapsules to the bone. The suture anchor systemThe device may be used in either arthroscopic oropen surgical procedures. After the suture isanchored to the bone, it may be used to reattachsoft tissue, such as ligaments, tendons, or jointcapsules to the bone. The suture anchor system
CrossFTTM Knotless Suture Anchor withDisposable DriverProposedConMed Linvatec Soft Tissue to Bone System(CrossFTTM Suture Anchor with Disposable Driver)Predicate
Contraindicationsthereby stabilizes the damaged soft tissue, inconjunction with appropriate postoperativeimmobilization, throughout the healing period.*Pathological conditions of bone which wouldadversely affect the CrossFT Knotless SutureAnchor.*Pathological conditions in soft tissue to berepaired or reconstructed which wouldadversely affect suture fixation.*Physical conditions that would eliminate, ortend to eliminate adequate implant support orretard healing.*Conditions which tend to limit the patient'sability or willingness to restrict activities orfollow directions during the healing period.*Attachment of artificial ligaments or otherimplants.*Foreign body sensitivity, known or suspectedallergies to implant and/or instrument materials.*This device is not cleared for screwattachment or fixation to the posterior elements(pedicles) of the cervical, thoracic or lumbarspine.thereby stabilizes the damaged soft tissue, inconjunction with appropriate postoperativeimmobilization, throughout the healing period.*Pathological conditions of bone which wouldadversely affect the CrossFT Knotless SutureAnchor.*Pathological conditions in soft tissue to berepaired or reconstructed which would adverselyaffect suture fixation.*Physical conditions that would eliminate, or tendto eliminate adequate implant support or retardhealing.*Conditions which tend to limit the patient's abilityor willingness to restrict activities or followdirections during the healing period.*Attachment of artificial ligaments or otherimplants.*Foreign body sensitivity, known or suspectedallergies to implant and/or instrument materials.*This device is not cleared for screw attachment orfixation to the posterior elements (pedicles) of thecervical, thoracic or lumbar spine.
HandleTwo pieces that consist of a handle and gripOne piece handle
AnchorDimensions4.0 mm, 4.75 mm and 5.5 mm in width14.5" inches in length4.5mm, 5.5mm and 6.5mm in width9.5" inches in length
Suture and/orSuture TapesconfigurationsCrossFTTM Knotless 4.0 mm Suture AnchorCrossFTTM Knotless 4.0 mm Suture Anchorwith one 2 mm Hi-Fi Tape (White/Black)CrossFTTM Knotless 4.75 mm Suture AnchorCrossFTTM Knotless 4.75 mm Suture Anchorwith one 2 mm Hi-Fi Tape (Blue)CrossFTTM Knotless 4.75 mm Suture Anchorwith one 2 mm Hi-Fi Tape (White/Black)CrossFTTM Knotless 5.5 mm Suture AnchorCrossFTTM Knotless 5.5 mm Suture Anchorwith one 2 mm Hi-Fi Tape (Blue)CrossFTTM Knotless 5.5 mm Suture Anchorwith one 2 mm Hi-Fi Tape (White/Black)Cross FT Anchor with two #2 Hi-Fi Suture, 4.5mmCross FT Anchor with three #2 Hi-Fi Suture,4.5mmCross FT Anchor with two # 2 Hi-Fi Suture,5.5mmCross FT Anchor with three # 2 Hi-Fi Suture,5.5mmCross FT Anchor with two #2 Hi-Fi Suture,6.5mmCross FT Anchor with two #2 Hi-Fi Suture,6.5mm
ThreaderUsed to thread Hi-Fi® Tape limbs or Hi-Fi® #2suture limbs through the anchor and driverFor the 4.0 mm anchor, a maximum of two (2)Hi-Fi® Tape limbs can be loaded.For the 4.75 mm and 5.5 mm anchors, amaximum of two (2) Hi-Fi® Tape limbs plustwo (2) Hi-Fi® #2 suture limbs, or six (6) Hi-Fi® #2 suture limbs can be loaded.None
CleatAvailable with two (2) suture cleats. The firstcleat is part of the handle design and located ontop of the handle. This cleat is used to wrap thepreloaded suture and/or suture type standsaround for packaging purposes. The secondThe suture cleat is used to hold the suturestrands while tensioning the anchor and forpackaging purpose
CrossFTTM Knotless Suture Anchor withDisposable DriverProposedConMed Linvatec Soft Tissue to Bone System(CrossFTTM Suture Anchor with Disposable Driver)Predicate
cleat is grey in color and has slotted indents oneach side to wrap the suture strands around. Thesecond cleat is also used to maintain tension onthe suture strands while the anchor is driveninto the bone.
PackagingPouchTrayPouchPETG retaining clipTyvek and Nylon/PE Film pouch
Single Use/ReusableSterile, Single-UseSterile, Single-Use
SterilizationEtoEto
ComponentsBone PunchBone TapDrill BitBone PunchBone Tap
Shelf-Life18 months5 year

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Conmed Corporation. The logo features a blue square with a white curved shape inside, followed by the word "CONMED" in black, bold letters. Below the word "CONMED" is the word "CORPORATION" in smaller, blue letters with a blue line above it.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Conmed Corporation. The logo consists of a blue square with a white curved shape inside on the left. To the right of the square is the word "CONMED" in black, with a blue line underneath, and the word "CORPORATION" in black below the line.

PERFORMANCE DATA

Testing has been completed to demonstrate that the CrossFT™ Knotless Suture Anchor with Disposable Driver performs as intended and is substantially equivalent to the predicate device. The bacterial endotoxin testing was conducted and met the endotoxin limits.

Completed testing includes the following:

  • . Reliability
  • Packaging .
  • Ultimate Fixation Strength ●
  • Cyclic ●
  • Sterilization ●
  • Pyrogenicity
  • Verification Testing ●
  • . Transportation
  • Biocompatibility ●
  • User Validation ●
  • Shelf-life ●

VIII. CONCLUSION

The CrossFTTM Knotless Suture Anchor with Disposable Driver is either substantially equivalent or identical in design, materials, intended use, principles of operation, and technical characteristics to the predicate CrossFT™ Suture Anchor with Disposable Driver. Based upon the findings of our performance testing, the differences present no new issues of safety and efficacy and the CrossFT™ Knotless Suture Anchor with Disposable Driver is substantially equivalent to the CrossFT™ Suture Anchor with Disposable Driver (K091549).

§ 888.3040 Smooth or threaded metallic bone fixation fastener.

(a)
Identification. A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.