Intended Use

The device is a motor driven, indoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.

Device Description

The EQ10D scooter consists of an un-foldable platform which connects the two front wheels and two rear wheels, an adjustable tiller, two lead acid batteries with an off-board charger, a motor/electromagnetic brake assembly, a electric motor controller and a seat /backrest set.

The patient uses the tiller handle/handlebar for steering and a thumb operated potentiometer throttle control lever located at the top of the tiller to engage and disengage the scooter motion in both the forward and reverse directions. When the throttle control lever is released, the electromagnetic brake will be actuated and the scooter is slow to stop.

This scooter is designed for use on firm, even surfaces such as concrete, asphalt and indoor flooring.

AI/ML Overview

I am sorry, but the provided text does not contain information about the acceptance criteria or a study proving that a device meets those criteria. The document is a 510(k) premarket notification for a motorized three-wheeled vehicle (scooter) and primarily focuses on establishing substantial equivalence to a predicate device.

It details:

  • The FDA's decision regarding the 510(k) submission.
  • Indications for use of the KYMCO EQ10D Scooter.
  • A 510(k) summary, including device description, non-clinical testing performed (compliance with various ISO standards), and a statement of substantial equivalence.
  • A detailed table comparing the KYMCO EQ10D Scooter to its predicate device (KYMCO EQ 20C, K130940) across numerous specifications and features to demonstrate substantial equivalence.

While non-clinical testing for compliance with ISO standards is mentioned (e.g., ISO 7176-1 to -16, -25, ISO 10993-1, -5, -10, ISO 14971), specific acceptance criteria values for a device's performance (like sensitivity, specificity, accuracy, etc.) and a study showing the device meets those specific numerical criteria are not described. The document focuses on showing the new device is "substantially equivalent" to an already marketed device, meaning it performs similarly and raises no new questions of safety or effectiveness, rather than demonstrating performance against a set of predefined acceptance criteria through a specific study with statistical outcomes.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, layered on top of each other to create a sense of depth and unity.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 16, 2017

Kwang Yang Motor Co., Ltd. % Mrs. Junnata Chang No. 35. Wan Hsing Street, San Min Dist. Kaohsiung 803 Taiwan

Re: K163201

Trade/Device Name: KYMCO EO10D Scooter Regulation Number: 21 CFR 890.3800 Regulation Name: Motorized Three-Wheeled Vehicle Regulatory Class: Class II Product Code: INI Dated: February 5, 2017 Received: February 16, 2017

Dear Mrs. Junnata Chang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

Michael J. Hoffmann -S

Carlos L. Peña, PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K163201

Device Name KYMCO EQ10D Scooter

Indications for Use (Describe)

The device is a motor driven, indoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.

Type of Use (Select one or both, as applicable)

☑ Registration Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart G)

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

This summary of 510(k) safetv and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.

Applicant

Name: KWANG YANG MOTOR CO., LTD. Address: No. 35, Wan Hsing Street, San Min Dist., 803, Kaohsiung, Taiwan Registration number: 3003851898 Contact person: Wang Ching Yun Phone: +886-7-3822526 Fax: +886-7-3825834 e-mail: kymcoreg(@mail.kymco.com Date prepared: Jan. 15, 2017 Device Trade name: KYMCO EQ10D scooter Common name: Electrical scooter Classification name: Motorized three-wheeled vehicle Medical specialty (Panel): Physical Medicine Device Regulation number: 890.3800 Product Code: INI Classification: Class II Predicate devices Trade name: KYMCO Model: EQ 20C 510(k) number: K130940 Manufacture: KWANG YANG MOTOR CO., LTD. Regulation number: 890.3800 Product Code: INI Classification: Class II Intend use of device It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.

{4}------------------------------------------------

Device description:

The EQ10D scooter consists of an un-foldable platform which connects the two front wheels and two rear wheels, an adjustable tiller, two lead acid batteries with an off-board charger, a motor/electromagnetic brake assembly, a electric motor controller and a seat /backrest set.

The patient uses the tiller handle/handlebar for steering and a thumb operated potentiometer throttle control lever located at the top of the tiller to engage and disengage the scooter motion in both the forward and reverse directions. When the throttle control lever is released, the electromagnetic brake will be actuated and the scooter is slow to stop.

This scooter is designed for use on firm, even surfaces such as concrete, asphalt and indoor flooring.

Summary of non-clinical testing

The EQ10D scooter complied with the requirements of ISO 7176-1, ISO 7176-2, ISO 7176-3, ISO 7176-4, ISO 7176-5, ISO 7176-6, ISO 7176-7, ISO 7176-8, ISO 7176-9, ISO 7176-10, ISO 7176-11, ISO 7176-13, ISO 7176-14, ISO 7176-15, ISO 7176-16, ANSI/RESNA WC Vol. 2 Sec. 21, ISO 7176-25, ISO 10993-1, ISO 10993-5, ISO 10993-10 and ISO 14971.

Statement of substantial equivalence

The EQ10D scooter is substantially equivalent to the EQ 20C (K130940). They have same intended use of a motor driven, indoor and outdoor transportation vehicle to provide mobility to a disabled or elderly person limited to a seated position.

The design and technological characteristics of this device is basically similar to the predicate device chosen. Both are un-foldable scooters and have the same user interface. While there are minor differences between the devices, do not alter the intended use function and use of the device. Moreover, the non-clinical tests and the predicate comparisons demonstrate that these differences in their technological characteristics do not raise any questions as to the safety and effectiveness.

The EQ10D is substantially equivalent to the KYMCO EQ 20C (K130940) manufactured by KWANG YANG MOTOR CO. LTD.. They have same intended use of a motor driven, indoor and outdoor transportation vehicle to provide mobility to a disabled or elderly person limited to a seated position.

The design and technological characteristics of this device is basically similar to the predicate device. They are 4-wheeled scooters with basic conventional rear wheel drive which with free-wheel mode. The motor of both devices is equipped with differential mechanism, which the differential rate is 19.7:1. The anti-tip wheels are equipped

{5}------------------------------------------------

behind the rigid frame of both devices. The power supply of both devices is self-contained lead-acid batteries to provide power that can be recharged by a charger that can be plugged into an AC outlet when the devices are not in use. The horn and foldable armrests are also equipped on both devices.

Both devices have the same user interface, the patient uses the foldable tiller handle/handlebar for steering and a thumb operated potentiometer throttle control lever located at the top of the tiller to engage and disengage the scooter motion in both the forward and reverse directions. When the throttle control lever is released, the electromagnetic brake will be actuated and the scooter is slow to stop. The speed mode of the both devices was single mode and the speed control dial can control the maximum speed. Both seats of the devices are equipped with removable flip-up armrest and there is no headrest on backrest. Seat cushions and backrests of both devices used the same materials, which is PU foam covered by nylon fabric cloth. The seat, seat cushion, backrest, handlebar and throttle control lever of the device will contact to the patient. The materials of the patient contacting parts used for the EQ10D scooter and predicate device are exactly same and no additional processing different from predicate device, so no new safety and efficacy concerning.

The predicate device is equipped with head light, tail light, signal light and warning light, which EQ 10D does not. There are warn labels in the EQ10D user manual that remind users do not ride the scooter at night in case of danger, because the scooter does not equipped head light, tail light, signal light and warning light, so do not raise any questions as to the safety.

The main difference for the two devices is the rated power of the used batteries, the rated power of the motor, battery charger type, lights and the maximum speed. While there are minor differences between the devices including dimensions, wheels size and pre-charging distance, do not alter the intended use function and use of the device.

Moreover, the non-clinical tests and the predicate comparisons demonstrate that these differences in their technological characteristics do not raise any questions as to the safety and effectiveness.

Conclusion

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification, KWANG YANG MOTOR CO., LTD. concludes that, EQ10D scooter is substantially equivalent to predicate devices as described herein.

{6}------------------------------------------------

EQ 20C (K130940)EQ10DComparison to the predicate device
Indications for useThey are motor driven, indoor and outdoor transportation vehiclewith the intended use to provide mobility to a disabled or elderlyperson limited to a seated position.Same indications for use
Maximum load capacity127 kg (280 lbs)125 kg (275 lbs)Smaller load capacity
ISO 7176-5
Overall length1080 mm (43.5")1020 mm (40.2")Smaller overall length
Overall width520 mm (20.5")377 ~ 500 mm(14.8 ~ 19.7")Smaller overall width
Handgrip height88 cm(34.6")89 cm(35")Larger handgrip height
Stowage length106 cm(41.7")102 cm(40.2")Smaller stowage length
Stowage width510 mm(20.1")487 mm(19.2")Smaller stowage width
Stowage height450 mm(17.7")940 mm(37")Larger stowage height
Rising150 mm(5.9")105 mm(4.1")Smaller rising
Total masswith battery : 55.5 kg (122 lbs)without battery : 45 kg(99 lbs)with battery : 43 kg(94.6 lbs)without battery : 33 kg (72.6 lbs)Lighter mass
Mass of heaviest part20.5 kg (45.2 lbs) body framewith motor14 kg(30.8 lbs) body frame withmotorLighter mass of heaviest part
Pivot width250 cm(98.4")150.5 cm(59.3")Smaller pivot width
Reversing width160 cm(63")233 cm(91.7")Larger reversing width
(Continuous) The substantial equivalence comparison of the EQ10D and EQ 20C (K130940)
EQ 20C (K130940)EQ10DComparison to thepredicate device
Turning diameter250 cm(98.4")242 cm(95.2")Smaller turning diameter
Ground clearance60 mm(2.4")50 mm(2")Smaller ground clearance
Required width of angled corridor120 cm(47.2")106.5 cm(41.9")Smaller required width ofangled corridor
Required doorway entry depth106 cm(41.7")60 cm(23.6")Smaller required doorwayentry depth
Required corridor width for sideopening190 cm(74.8")110 cm(43.3")Smaller required corridorwidth for side opening
ISO 7176-7
Seat plane angleSmaller seat plane angle
Effective seat depth350 mm(13.8")483 mm (19")Larger effective seat depth
Seat width350 mm(13.8")400 mm (15.7")Larger seat width
Effective seat width307 mm ~ 430 mm(12" ~ 17")327 mm ~ 450 mm(12.8" ~ 18")Larger effective seat width
Seat surface height at front edge440 mm~516 mm(17.3"~20.3")445 ~ 521 mm(17.5" ~ 20.5")Larger seat surface height atfront edge
Back support angle12°15°Larger back support angle
Back support height395 mm(15.6")290 mm(11.4")Smaller back support height
Back support width350 mm(13.8")400 mm (15.7")Larger back support width
(Continuous)The substantial equivalence comparison of the EQ10D and EQ 20C (K130940)
EQ 20C (K130940)EQ10DComparison to the predicate device
Foot support to seat distance345 ~ 421 mm(13.6" ~ 16.6")370 ~ 446 mm(14.6" ~ 17.6")Both devices are adjustable
Foot support clearance70 mm(2.8")50 mm(2")Smaller foot support clearance
Arm support height155 mm(6.1")180 mm(7.1")Larger arm support height
Front of arm support to back support250 mm(9.8")350 mm(13.8")Larger front of arm support to back support
Arm support length260 mm(10.2")300 mm(11.8")Larger arm support length
Arm support width50 mm(2")Same arm support width
Arm support angleSame arm support angle
Distance between arm supports420 mm(16.5")377 mm ~ 500 mm(14.8"~19.7")Distance between arm supports is adjustable
(Continuous)The substantial equivalence comparison of the EQ10D and EQ 20C (K130940)
EQ 20C (K130940)EQ10DComparison to the
predicate device
Horizontal location of the wheel axle219 mm(8.6")283 mm(11.4")Larger horizontal location of the wheel axle
Vertical location of the wheel axle284 mm(11.2")277 mm(10.9")Smaller vertical location of the wheel axle
MaterialPU foam covered by nylon fabric clothSame materials
Seat
Seat cushion
Backrest
Differential mechanismDifferential rate: 19.7 : 1Same differential mechanism
Both devices have free- wheel
Free-wheel modeYesmode
Speed modeSingle mode, VariableSame speed mode
Lights
Head/Tail lightsYesNoWithout head/tail lights
Signal lightYesNoWithout signal light
Warning lightYesBoth devices have warning light
HornYesBoth devices have horn
(Continuous)The substantial equivalence comparison of the EQ10D and EQ 20C (K130940)
EQ 20C (K130940)EQ10DComparison to the predicate device
Anti-tip wheelsYesBoth devices have anti-tip wheels
Operation modeThumb operated potentiometer throttle control leverSame operation mode
ArmrestRemovable flip-up armrestSame type armrest
Tiller foldableYesSame type tiller
HeadrestNoBoth device without headrest
Motor
Rated power250 W270 WLarger rated power
Input voltageDC 24VSame input voltage
Amount1 PcSame amount
Brake systemIntelligent regenerative electromagnetic brake and hand brakeSame type brake system
ControllerPG S-Drive, 45ASame controller
Rear wheel driveSealed transaxle direct driveSame type drive
Braking distance(at max speed)Forward:1.0 m (39.4")Forward: 1.1 m (43.3")Larger braking distance
Per-charge distanceUp to 21 km (13 miles)Up to 16 km (10 miles)Smaller per-charge distance
Maximum speedUp to 6.5 km/h (4.1 mph), variableUp to 6.4 km/hr (4 mph), variableSmaller max. speed
Maximum curb height50 mm (2")Same maxi. curb height
SuspensionFront: No, Rear: YesSame
(Continuous)The substantial equivalence comparison of the EQ10D and EQ 20C (K130940)
EQ 20C (K130940)EQ10DComparison to the predicate device
Battery
TypeLead-acidSame type batteries
Rated power22 Ah24 AhLarger rated power
Output voltage24 VDCSame output voltage
AmountDC 12V x 2 PcsSame amount
Battery level indicatorYesBoth have battery level indicator
Charger
Typeon-board, Automatic Typeoff-board, Automatic TypeDifferent type
Input power requirement115-230 VACSame input power
Output voltage / currentDC 24V / 5ADC 24V / 2ASame output voltageSmaller output current
Front wheels
Diameter70 mm/62.5 mm – 100 mm(2.8"/2.5"-4")7.5" × 2.35"Larger diameter
typenon-pneumatic tiresSame type tires
MaterialPolyurethane (PU)Same material
Amount2Same amount
(Continuous)The substantial equivalence comparison of the EQ10D and EQ 20C (K130940)
EQ 20C (K130940)EQ10DComparison to the predicate device
Rear wheels
Diameter70 mm/62.5 mm – 100 mm(2.8"/2.5"-4")7.5" × 2.35"Larger diameter
typenon-pneumatic tiresSame type tires
MaterialPolyurethane (PU)Same material
Amount2Same amount

The substantial equivalence comparison of the EQ10D and EQ 20C (K130940)

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

{10}------------------------------------------------

{11}------------------------------------------------

{12}------------------------------------------------

§ 890.3800 Motorized three-wheeled vehicle.

(a)
Identification. A motorized three-wheeled vehicle is a gasoline-fueled or battery-powered device intended for medical purposes that is used for outside transportation by disabled persons.(b)
Classification. Class II (performance standards).