Dario Blood Glucose Monitoring System
K163159 · Labstyle Innovations, Ltd. · NBW · Jul 19, 2017 · Clinical Chemistry
Device Facts
| Record ID | K163159 |
| Device Name | Dario Blood Glucose Monitoring System |
| Applicant | Labstyle Innovations, Ltd. |
| Product Code | NBW · Clinical Chemistry |
| Decision Date | Jul 19, 2017 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.1345 |
| Device Class | Class 2 |
| Attributes | Pediatric |
Intended Use
The Dario Blood Glucose Monitoring System consists of the Dario Blood Glucose Meter, Dario Glucose Test Strips, Dario Glucose Control Solutions and the Dario App as the display component of the Dario Blood Glucose Monitoring System. The Dario Blood Glucose Monitoring System is intended for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips. The Dario Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. The Dario Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home to monitor the effectiveness of diabetes control. The Dario Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. The Dario Blood Glucose Test Strips are for use with the Dario Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertip.
Device Story
Compact, all-in-one blood glucose meter; connects to smart mobile device (SMD) via 3.5mm audio plug for power. User applies fresh capillary whole blood to test strip; meter performs quantitative glucose measurement. SMD runs Dario App to display results; SMD provides no diagnostic or clinical analysis. Used at home by patients for diabetes monitoring. Device agnostic to SMD platform/model.
Clinical Evidence
User performance study conducted to evaluate expanded indications. Results confirmed substantial equivalence to the predicate device. Additional bench testing included repeatability, intermediate precision, linearity, operating conditions (altitude), and accuracy.
Technological Characteristics
Compact glucose meter; 3.5mm audio plug interface for power/data transfer to SMD. Software level of concern: moderate. Compliance with IEC 60601-1 (electrical safety) and electromagnetic interference standards (radiated emissions, ESD, RF immunity).
Indications for Use
Indicated for people with diabetes for home self-testing of blood glucose levels using fresh capillary whole blood from fingertips. Not for diagnosis, screening, or neonatal use. Single-patient use only.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
- Dario Blood Glucose Monitoring System (K150817)
Related Devices
- K172723 — Dario LC Blood Glucose Monitoring System · Labstyle Innovation, Ltd. · Mar 23, 2018
- K150817 — Dario Blood Glucose Monitoring System · Labstyle Innovations, Ltd. · Dec 21, 2015
- K052762 — LIBERTY BLOOD GLUCOSE MONITORING SYSTEM, MODEL 8000-0067 · Agamatrix, Inc. · Jan 23, 2006
- K223293 — CONTOUR® NEXT GEN Blood Glucose Monitoring System · Ascensia Diabetes Care U.S., Inc. · Nov 18, 2022
- K110074 — HL 568 SELF MONITORING BLOOD GLUCOSE SYSTEM, HL 568BA SELF-MONIROTING BLOOD GLUCOSE SYSTEM · Health & Life Co., Ltd. · May 25, 2012
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 19, 2017
LABSTYLE INNOVATIONS, LTD. C/O TRACEY WIELINSKI, PRESIDENT QUALTRA CONSULTING, INC. 5 CASIE LANE PEPPERELL, MA 01463
Re: K163159
Trade/Device Name: Dario Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose Test System Regulatory Class: Class II Product Code: NBW Dated: June 15, 2017 Received: June 19, 2017
Dear Tracey Wielinski:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Kellie B. Kelm -S
for Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K163159
Device Name Dario Blood Glucose Monitoring System
### Indications for Use (Describe)
The Dario Blood Glucose Monitoring System consists of the Dario Blood Glucose Test Strips, Dario Glucose Control Solutions and the Dario App as the display component of the Dario Blood Glucose Monitoring System. The Dario Blood Glucose Monitoring System is intended for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips. The Dario Blood Glucose Monitoring System is intended to be used by a single person and should not be shared.
The Dario Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home to monitor the effectiveness of diabetes control. The Dario Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use.
The Dario Blood Glucose Test Strips are for use with the Dario Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertip.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
> Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| Submitter: | LabStyle Innovations<br>9 Halamish Street<br>North Industrial Park<br>Caesarea 3890000<br>Israel |
|-------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Tara N. Turney<br>Regulatory Consultant<br>Qualtra Consulting, Inc<br>Telephone: +852-5964-8211<br>E-mail: tara@mydario.com |
| Date Prepared: | July 18, 2017 |
| Trade Name: | Dario® Blood Glucose Monitoring System |
| Common Name: | Glucose Test System |
| Regulation: | 21 § 862.1345 |
| Product Code: | NBW |
| Classification<br>Name: | System, Test, Blood Glucose, Over The Counter |
| Predicate Device: | Dario Blood Glucose Monitoring System<br>K150817 |
| Device<br>Description: | The Dario Blood Glucose Monitoring System is a compact,<br>all-in-one blood glucose meter that is used in conjunction with<br>a smart mobile device (SMD). The Dario meter is connected<br>to a SMD through a standard 3.5mm audio plug in order to<br>provide the glucose meter with power. The SMD also<br>provides an external display for the user to review blood<br>glucose testing results via a mobile application downloaded to<br>the SMD. The full process of blood glucose measurement is<br>done on the glucose meter; the SMD does not provide any<br>diagnostic or clinical analysis function. When connected to a<br>SMD, the Dario glucose meter is agnostic both to the platform<br>of the smart mobile device and to the model of the individual<br>SMD within a platform. |
| Statement of<br>Intended Use: | The Dario Blood Glucose Monitoring System is intended for<br>the quantitative measurement of glucose (sugar) in fresh |
| | capillary whole blood. The Dario is intended for self-testing<br>outside the body (in vitro diagnostic use) by people with<br>diabetes at home to monitor the effectiveness of their diabetes<br>control. |
| Statement of<br>Indications for Use: | The Dario Blood Glucose Monitoring System consists of the<br>Dario Blood Glucose Meter, Dario Glucose Test Strips, Dario<br>Glucose Control Solutions and the Dario App as the display<br>component of the Dario Blood Glucose Monitoring System.<br>The Dario Blood Glucose Monitoring System is intended for<br>the quantitative measurement of glucose (sugar) in fresh<br>capillary whole blood samples drawn from the fingertips. The<br>Dario Blood Glucose Monitoring System is intended to be<br>used by a single person and should not be shared. |
| | The Dario Blood Glucose Monitoring System is intended for<br>self-testing outside the body (in vitro diagnostic use) by<br>people with diabetes at home to monitor the effectiveness of<br>diabetes control. The Dario Blood Glucose Monitoring<br>System should not be used for the diagnosis of or screening of<br>diabetes or for neonatal use. |
| | The Dario Blood Glucose Test Strips are for use with the<br>Dario Blood Glucose Meter to quantitatively measure glucose<br>(sugar) in fresh capillary whole blood samples drawn from the<br>fingertip. |
| Summary of<br>Technological<br>Characteristics: | The proposed device has the same technological<br>characteristics and is identical in design and configuration as<br>compared to the predicate device. |
| Summary of Non-<br>Clinical Data: | The following performance and safety testing has confirmed<br>the proposed device to be substantially equivalent to the<br>predicate device: |
| | • In Vitro: the proposed Dario Blood Glucose<br>Monitoring System has been tested for in vitro<br>performance and meets all of its associated<br>specifications; |
| | • Software: documentation was prepared and submitted<br>for a moderate level of concern device in accordance<br>with FDA's Guidance for the Content of Premarket<br>Submissions for Software Contained in Medical<br>Devices; |
| | • Electrical Safety: the proposed Dario Blood Glucose<br>Monitoring System has been tested and successfully |
| Summary of<br>Clinical Data: | A User Performance study was completed for the expanded<br>indications of the Dario Blood Glucose Monitoring System.<br>This evaluation confirmed the proposed device to be<br>substantially equivalent to the predicate device. |
| Conclusion from<br>Data: | LabStyle Innovations has demonstrated that the proposed<br>Dario Blood Glucose Monitoring System is substantially<br>equivalent to the predicate device based upon indications for<br>use, design, test results and the same fundamental scientific<br>technology. |
{4}------------------------------------------------
{5}------------------------------------------------
passed all of the relevant sections of IEC 60601-1 Medical electrical equipment, General requirements for Safety;
- . Electromagnetic Interference: the proposed Dario Blood Glucose Monitoring System has been tested and successfully met all of the relevant sections (Radiated emissions, Electrostatic discharge immunity test, radiated radio frequency, electromagnetic field immunity, and power frequency magnetic field immunity test) to satisfy compliance;
The following categories of performance were assessed and have confirmed the proposed device to be substantially equivalent to the predicate device:
- Repeatability Precision
- . Intermediate Precision
- Linearity
- Operating Conditions .
- . Altitude
- Accuracy ●