K Number
K163109

Validate with FDA (Live)

Date Cleared
2017-03-22

(135 days)

Product Code
Regulation Number
868.5470
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The intended use of the OxyHeal® 4000 Cylindrical Multiplace Hyperbaric Chamber System Product Family is to administer hyperbaric oxygen therapy (HBOT) to treat patients with any of the below listed indications.

The following indications which are listed on the Undersea & Hyperbaric Medical Society (UHMS) web site: www.uhms.org.are approved uses of hyperbaric oxygen therapy as defined by the Hyperbaric Oxygen Therapy Committee.

    1. Air or Gas Embolism
    1. Carbon Monoxide Poisoning
  • a. Carbon Monoxide Poisoning Complicated by Cyanide Poisoning
    1. Clostridial Myositis and Myonecrosis (Gas Gangrene)
    1. Crush Injury, Compartment Syndrome and Other Acute Ischemias
    1. Decompression Sickness
    1. Arterial Insufficiencies
  • a. Central Retinal Artery Occlusion
  • b. Enhancement of Healing in Selected Problem Wounds
    1. Severe Anemia
    1. Intracranial Abscess
    1. Necrotizing Soft Tissue Infections
    1. Osteomyelitis (Refractory)
    1. Delayed Radiation Injury (Soft Tissue and Bony Necrosis)
    1. Compromised Grafts and Flaps
    1. Acute Thermal Burn Injury
Device Description

The OxyHeal® 4000 Cylindrical Multiplace Hyperbaric Chamber System Product Family is comprised of a multiplace hyperbaric chamber of major subsystems that support the overall system operation, control, and monitoring.

The OxyHeal® 4000 Cylindrical Multiplace Hyperbaric Chamber is a pressure vessel for human occupancy (PVHO) that is designed in a horizontally orientated cylindrical geometry. Chamber configurations vary based on the needs of the end user, and may be designed and manufactured in one (1), two (2), or three (3), compartment configurations. Patient capacities may range anywhere from four (4) to twenty-four (24) dependent on chamber size, number of compartments, or the direction provided by the customer to meet their needs. Lastly, maximum operating pressures range from 3ATA (~30psi) to 6ATA (~73.5psi), with each of the compartments designed to operate independently.

The OxyHeal® 4000 Cylindrical Multiplace Hyperbaric Chamber System design also complies with the hyperbaric facilities requirements specified in the FDA recognized consensus standard NFPA 99 and satisfies the requirements for protection against electrical, explosive, and fire hazards and associated facilities used for medical procedures at gauge pressures within the ranges: 0psi to 100psi.

The OxyHeal® 4000 Cylindrical Multiplace Hyperbaric Chambers consists of the hyperbaric chamber itself and the major subsystems briefly described below. Each substantially equivalent to that which is contained in the OxyHeal® 5000 Rectangular Multiplace Hyperbaric Chamber System Product Family predicate device (K152223).

AI/ML Overview

The provided document is a 510(k) premarket notification for a hyperbaric chamber system. It focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study proving that a new, AI-based device meets acceptance criteria.

Therefore, the requested information about acceptance criteria, device performance, sample sizes, expert ground truth, MRMC studies, standalone performance, and training set details for an AI/ML medical device cannot be extracted from this document.

This document describes a change in the shape of a hyperbaric chamber (from rectangular to cylindrical) and demonstrates that it adheres to existing safety and manufacturing standards, mirroring the predicate device. It is a traditional medical device submission, not one pertaining to AI/ML.

No information regarding AI-specific acceptance criteria or a study proving an AI device meets them is present in the provided text.

The document details:

  • Device: OxyHeal® 4000 Cylindrical Multiplace Hyperbaric Chamber System Product Family
  • Predicate Device: OxyHeal® 5000 Rectangular Multiplace Hyperbaric Chamber System Product Family (K152223)
  • Key Modification: Change from rectangular to cylindrical geometry.
  • Demonstration of Equivalence: Through adherence to recognized industry standards (ASME PVHO-1, NFPA 99, ISO 14971), non-clinical testing (Non-destructive Testing, Hydrostatic Testing, Fire Suppression System Testing, First Operational System Test, Software Validation Testing - note: "software validation" in this context refers to confirming the chamber's control software functions as intended, not AI/ML inference), and factory acceptance testing.

The document explicitly states: "The fundamental technology for this minor modification consists of a change in dimensional specifications; i.e. the predicate device (K152223) is a multiplace hyperbaric chamber with a rectangular geometry. The modified device is a multiplace hyperbaric chamber with a cylindrical geometry. This modification does not change the intended use, indications for use, and product labeling." It also says, "The validation / verification efforts performed for the OxyHeal® 4000 Cylindrical Multiplace Hyperbaric Chamber System Product Family are identical to those performed for the OxyHeal® 5000 Rectangular Multiplace Hyperbaric Chamber System Product Family (K152223) predicate device."

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the United States Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 22, 2017

OxyHeal Medical Systems, Inc. Edward J. Chomas VP. Regulatory Affairs 3224 Hoover Ave National City, California 91950

Re: K163109

Trade/Device Name: OxyHeal® 4000 Cylindrical Multiplace Hyperbaric Chamber System Product Family Regulation Number: 21 CFR 868.5470 Regulation Name: Hyperbaric Chamber Regulatory Class: Class II Product Code: CBF Dated: February 22, 2017 Received: February 23, 2017

Dear Edward Chomas:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Image /page/1/Picture/8 description: The image contains a signature and a name. The name is "Tina Kiang-S". The signature is to the left of the name and is illegible.

Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K163109

Device Name

OxyHeal® 4000 Cylindrical Multiplace Hyperbaric Chamber System Product Family

Indications for Use (Describe)

The following indications which are listed on the Undersea & Hyperbaric Medical Society (UHMS) web site: www.uhms.org are approved uses of hyperbaric oxygen therapy as defined by the Hyperbaric Oxygen Therapy Committee.

    1. Air or Gas Embolism
    1. Carbon Monoxide Poisoning
  • a. Carbon Monoxide Poisoning Complicated by Cyanide Poisoning
    1. Clostridial Myositis and Myonecrosis (Gas Gangrene)
    1. Crush Injury, Compartment Syndrome and Other Acute Ischemias
    1. Decompression Sickness
    1. Arterial Insufficiencies
  • a. Central Retinal Artery Occlusion
  • b. Enhancement of Healing in Selected Problem Wounds
    1. Severe Anemia
    1. Intracranial Abscess
    1. Necrotizing Soft Tissue Infections
    1. Osteomyelitis (Refractory)
    1. Delayed Radiation Injury (Soft Tissue and Bony Necrosis)
    1. Compromised Grafts and Flaps
    1. Acute Thermal Burn Injury
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Section 008- 510(k) Summary for

OxyHeal® 4000 Cylindrical Multiplace Hyperbaric Chamber System Product Family

Device Name: OxyHeal® 4000 Cylindrical Multiplace Hyperbaric Chamber System Product Family

1. Submission Sponsor

OxyHeal® Medical Systems, Inc. 3224 Hoover Avenue National City, CA 91950 Phone: 619.336.2022 Fax: 619.336.2017 Contact: W. T. 'Ted' Gurneé, President & CEO

2. Submission Correspondent

OxyHeal® Medical Systems, Inc. 3224 Hoover Avenue National City, CA 91950 Phone: 619.336.2022 Fax: 619.336.2017 Contact: Edward J. Chomas, VP Regulatory Affairs Email: echomas@oxyheal-international.com

3. Date Prepared

28 October 2016

4. Device Name

Trade/Proprietary Name: OxyHeal® 4000 Cylindrical Multiplace Hyperbaric Chamber System Product Family Common/Usual Name: Multiplace Hyperbaric Chamber Classification Name: Chamber, Hyperbaric Classification Regulation: 21 CBF 868.5470 Classification Panel: Anesthesiology Product Code: CBF Device Class: II

FDA Establishment Registration #: 1000519737

5. Predicate Device

OxyHeal® 5000 Rectangular Multiplace Hyperbaric Chamber System Product Family (K15223)

6. Device Description

The OxyHeal® 4000 Cylindrical Multiplace Hyperbaric Chamber System Product Family is comprised of a multiplace hyperbaric chamber of major subsystems that support the overall system operation, control, and monitoring.

{4}------------------------------------------------

The OxyHeal® 4000 Cylindrical Multiplace Hyperbaric Chamber is a pressure vessel for human occupancy (PVHO) that is designed in a horizontally orientated cylindrical geometry. Chamber configurations vary based on the needs of the end user, and may be designed and manufactured in one (1), two (2), or three (3), compartment configurations. Patient capacities may range anywhere from four (4) to twenty-four (24) dependent on chamber size, number of compartments, or the direction provided by the customer to meet their needs. Lastly, maximum operating pressures range from 3ATA (~30psi) to 6ATA (~73.5psi), with each of the compartments designed to operate independently.

The OxyHeal® 4000 Cylindrical Multiplace Hyperbaric Chamber System is regulated by the same codes and standards as the OxyHeal® 5000 Rectangular Multiplace Hyperbaric Chamber System, predicate device K152223.

    1. NFPA 99 (2012 Edition): National Fire Protection Agency Standard for Health Care Facilities Chapter 14 - Hyperbaric Chambers
    • FDA recognized consensus standard: (Recognition Number 1-67) .
    1. ANSI/ASME PVHO-1 (2012 Edition): American National Standards Institute/American Society of Mechanical Engineers - Safety Standard for Pressure Vessels for Human Occupancy.
    • FDA recognized consensus standard: (Recognition Number 1-78) ●
    1. ISO 14971 (2012 Edition): International Standard Organization. Medical Devices Application of Risk Management to Medical Devices
    • FDA recognized consensus standard: (Recognition Number 5-40) .

OxyHeal® Medical Systems, Inc. complies with the ASME/PVHO-1 the FDA recognized consensus code and standard requirements for materials, design, fabrication, and testing of the OxyHeal® 4000 Cylindrical Multiplace Hyperbaric Chamber. This includes: overall PVHO design, joint design, welding, non-destructive examination (NDE), viewports, penetrations, material reinforcement, pressure relief devices, piping, electrical outfitting, inspections, testing, risk analysis, documentation, and marking (labeling).

The OxyHeal® 4000 Cylindrical Multiplace Hyperbaric Chamber System design also complies with the hyperbaric facilities requirements specified in the FDA recognized consensus standard NFPA 99 and satisfies the requirements for protection against electrical, explosive, and fire hazards and associated facilities used for medical procedures at gauge pressures within the ranges: 0psi to 100psi.

OxyHeal® Medical Systems, Inc.'s design control processes for the OxyHeal® 4000 Cylindrical Multiplace Hyperbaric Chamber System are the same as those employed for the OxyHeal® 5000 Rectangular Multiplace Hyperbaric Chamber System, predicate device K152223. These processes conform to the requirements of 21 CFR 820.30 and comply with the ISO 14971 FDA recognized consensus standard.

The OxyHeal® 4000 Cylindrical Multiplace Hyperbaric Chambers consists of the hyperbaric chamber itself and the major subsystems briefly described below. Each substantially equivalent to that which is contained in the OxyHeal® 5000 Rectangular Multiplace Hyperbaric Chamber System Product Family predicate device (K152223).

{5}------------------------------------------------

6.1 Compressed Air System.

The compressed air system consists of two (2) rotary screw compressors capable of producing pressurized air that is then stored in an air receiver, which in turn is used to pressurize the hyperbaric chamber. Air is filtered prior to entering the hyperbaric chamber, resulting in a breathable quality. Compressed Gas Association (CGA) Grade "E" air as required by NFPA 99. Details of the parts and components comprising the compressed air system are identified and described in Table 1, Sub-table 1-3, Compressed Air System.

6.2 Fire Suppression System.

The fire suppression system (FSS) consists of both a fire deluge system (primary) and hand line system (secondary). Water (potable) for both systems is stored in pressure vessels manufactured to ASME standards. The fire deluge system is activated in the event of a fire in the hyperbaric chamber, while the hand line system is activated manually. This complies with NFPA 99:2012, Standard for Health Care Facilities Chapter 14 - Hyperbaric Facilities. Details of the parts and components comprising the FSS are identified and described in Table 1, Sub-table 1-4, Fire Suppression System.

6.3 Oxygen Delivery System

An oxygen (O2) delivery system is the primary source for supplying O2 to patients' breathing hoods inside the chamber. OxyHeal® Medical Systems, Inc. (OMS) provides the end user with the requirements needed for an OxyHeal® 4000 hyperbaric chamber system, and it is the end user's responsibility for supplying a system meeting these requirements. Examples of two types of O2 delivery systems and associated piping are identified and described in Table 1-5, Oxygen Delivery Requirements.

6.4 Built-in Breathing System

The built-in breathing system (BIBS) is capable of supplying each individually seated patient with breathing gas via standard oxygen hoods or free-flow masks. Breathing gasses can be O2, medical air, or a gas mixture. Details of the parts and components comprising the BIBS are identified and described in Table 1. Sub-table 1-6. Built-in Breathing System (BIBS).

6.5 Environmental Control System

The environmental control system (ECS) is used to manage the temperature (heating and cooling) and relative humidity (RH) of the hyperbaric chamber. Details of the parts and components comprising the ECS are identified and described in Table 1, Sub-table 1-7, Environmental Control System (ECS).

6.6 Control Console

The control console serves as the central location where a qualified hyperbaric chamber technician (CHT) is capable of controlling and monitoring an OxyHeal® 4000 product family hyperbaric chamber system. The Human-Machine Interface (HMI) touch screen control system installed in the operator control console is the primary location from which a hyperbaric chamber operator is able to initiate and monitor patient hyperbaric oxygen therapy (HBOT) treatments. Manual back-up control systems are built into the system for control of pressurization and depressurization from both inside and outside the hyperbaric chamber in the event that the automatic feature is inoperable for any reason.

From the HMI touchscreen, the operator is also able to control the following:

{6}------------------------------------------------

  • Administer BIBS gasses a.
  • b. Analyze / monitor O2
  • Analyze/ monitor carbon dioxide (CO2) [option] c.
  • Analyze / monitor relative humidity inside the hyperbaric chamber d.
  • Control and monitor the temperature in the hyperbaric environment e.
  • f. Open and close doors in any hyperbaric chamber compartment
  • Turn ON/OFF and adjust the intensity of hyperbaric chamber lighting; and g.
  • h. Perform administrative functions.

The FSS is activated from the control console. A communications system is installed at the control console allowing a CHT to communicate to patients and inside attendants within the hyperbaric chamber. The equipment used for the analysis of chamber and BIBS line gasses is mounted at the control console. Chamber lighting is managed at the control console. The control console also contains equipment used to visually monitor patients inside of the hyperbaric chamber from a CCTV, and initiate and adjust patient audio and visual entertainment (radio, CD, DVD, and TV). Details of the parts and components comprising the control console are identified and described in Table 1-8, Control Console.

{7}------------------------------------------------

Comparison Table 1. OxyHeal® 4000 Cylindrical Multiplace Hyperbaric Chamber System Product Family Compared to the OxyHeal® 5000 Rectangular Multiplace Hyperbaric Chamber System Product Family

1-1 Regulatory Info
ManufacturerOxyHeal® MedicalSystems, Inc.OxyHeal® MedicalSystems, Inc.OxyHeal® 4000 FamilySummary Comparison toPredicate (K152223)
Trade NameOxyHeal® 4000 Family(K163109)OxyHeal® 5000 Family(K152223)
510(k) NumberK163109K152223Unique K numbers for eachproduct family have beenassigned.
Product CodeCBFCBFIdentical
Regulation Number21 CFR 868.547021 CFR 868.5470Identical
Regulation NameHyperbaric ChamberHyperbaric ChamberIdentical
Indications for use:As defined in theHyperbaric OxygenTherapy Committee Report,dated 2008As defined in theHyperbaric OxygenTherapy CommitteeReport, dated 2008Identical
1-2 Hyperbaric Chamber
ManufacturerOxyHeal® Medical Systems, Inc.OxyHeal® Medical Systems, Inc.OxyHeal® 4000 Family Summary Comparison to Predicate (K152223)
Trade NameOxyHeal® 4000 Family (K163109)OxyHeal® 5000 Family (K152223)
Hyperbaric Chamber Code Design1. ASME: Boiler and Pressure Code2. ASME PVHO-1: Safety Standard for Pressure Vessels for Human Occupancy.FDA consensus standard recognition no. 1-78.1. ASME: Boiler and Pressure Code2. ASME PVHO-1: Safety Standard for Pressure Vessels for Human Occupancy.FDA consensus standard recognition no. 1-78.Regardless of PVHO geometry:Rectangular: OxyHeal® 5000 [non-uniform distribution of pressure (worst case)] Cylindrical: OxyHeal® 4000 [uniform distribution of pressure] Codes and standards defining requirements for PVHO materials, design, fabrication, and testing are identical.
CalculationsASME Section VIII, Div. 1ASME Section VIII, Div. 1Identical
1-2 Hyperbaric Chamber
ManufacturerTrade NameOxyHeal® MedicalSystems, Inc.OxyHeal® 4000 Family(K163109)OxyHeal® MedicalSystems, Inc.OxyHeal® 5000 Family(K152223)OxyHeal® 4000 FamilySummary Comparison toPredicate (K152223)
Joint Design-Design and fabricationshall be in accordancewith specified Divisions ofSection VIII of the ASMECode and the followingrequirements ofASME/PVHO-1, para. 1-7.1 (a) through (c)Design and fabrication shallbe in accordance withspecified Divisions ofSection VIII of the ASMECode and the followingrequirements ofASME/PVHO-1, para. 1-7.1 (a) through (c)Identical
ASME PVHO-1-MaterialSA 516 Grade 70SA 516 Grade 70Identical
Stress Allowance-Per ASME Section IIPer ASME Section IIIdentical
Non-Destructive-Examinations(NDE)NDE shall be performed inaccordance withASME/PVHO-1, para. 1-7.3NDE shall be performed inaccordance withASME/PVHO-1, para. 1-7.3Identical
Hydrostatic Test-Hydrostatic testing to beperformed in accordancewith ASME Boiler andPressure Vessel Code,Section UG-99 @ 1.3times max. allowableworking pressure.Hydrostatic testing to beperformed in accordancewith ASME Boiler andPressure Vessel Code,Section UG-99 @ 1.3 timesmax. allowable workingpressure.Identical
HyperbaricChamber SystemFire Safety DesignNFPA 99, Chapter 14 -Hyperbaric Facilities.FDA consensus standardrecognition no. 1-76.NFPA 99, Chapter 14 -Hyperbaric Facilities. FDAconsensus standardrecognition no. 1-76.Regardless of PVHOgeometry:• Rectangular: OxyHeal®5000• Cylindrical: OxyHeal®4000This standard definingrequirements for protectionagainst electrical, explosive,and fire hazards inhyperbaric facilities isidentical
Operating Pressure3.0ATA – 6.0ATA3.0ATA – 6.0ATAIdentical
OperatingTemperature50°F – 125°F50°F – 125°FIdentical
Design Temperature50°F – 125°F50°F – 125°FIdentical
1-2 Hyperbaric Chamber
ManufacturerOxyHeal® Medical Systems, Inc.OxyHeal® Medical Systems, Inc.OxyHeal® 4000 Family Summary Comparison to Predicate (K152223)
Trade NameOxyHeal® 4000 Family (K163109)OxyHeal® 5000 Family (K152223)
Design Pressure30psig – 75psig30psig – 75psig
Design Life90,000 cycles or 60 years, which ever happens first90,000 cycles or 60 years, which ever happens firstIdentical
Hydrostatic Pressure39psi - 97.5psi39psi - 97.5psiIdentical
Inspection AuthorityIndependent 3rd Party ASME Authorized Inspector (AI).Affix ASME Stamp on chamber data plateIndependent 3rd Party ASME Authorized Inspector (AI).Affix ASME Stamp on chamber data plateIdenticalReference FDA recognized consensus standard: ASME PVHO-1 (Recognition Number 1-78)
Weight (lbs.)12,600lbs to 100,000lbs15,000lbs to 120,000lbsSubstantially equivalent
DimensionsFor all compartments, the following min/max applyFor all compartments, the following min/max applySubstantially equivalent
Main Compartment (Lock) (ML)Min: 60" Diameter x 10'LMax: 120" Diameter x 20'LMin: 8' W x 7' H x 10'LMax: 11'W x 8'H x 20'L
Transfer Compartment (Lock) (EL)Flat Heads, Rectangular Shell, and Rectangular Door Frames.
Inner Compartment (Lock) (IL)Semi-elliptical or flat heads, cylindrical or semi-cylindrical shell, and circular or rectangular door frames
Total VolumeFrom 567 ft3 to 2434 ft3600 ft3 to 2,600 ft3Substantially equivalent
Medical LockCylindricalMin: 10 inch diameterMax: 16 inch diameterCylindricalMin: 10 inch diameterMax: 16 inch diameterIdentical
Main Doorway SizeRectangular DoorMin: 39.5" W x 65" HMax: 47.75" W x 75.5" HRectangular DoorMin: 44" W x 80" HMax: 52" W x 80" HSubstantially equivalent
1-2 Hyperbaric Chamber
ManufacturerOxyHeal® MedicalSystems, Inc.OxyHeal® 4000 FamilySummary Comparison toPredicate (K152223)
OxyHeal® MedicalSystems, Inc.
Trade NameOxyHeal® 4000 Family(K163109)
OxyHeal® 5000 Family(K152223)
Min: 33.5" DMax: 90" D
PenetratorsMaximum of 30Penetrations of 2" x 12"blocks.Identical
Viewports(PVHO-1)Minimum: One (1) perhyperbaric chamber.Maximum: Six (6) percompartment.Minimum: 8" Diameter,Maximum: 30" Diameter.Identical
Compartment ReliefOne (1) ASME certifiedpressure relief valve percompartment.30 psig to 75 psigIdentical
Compartment DrainMinimum One (1) manualdrain in each compartmentIdentical
Finish - ChamberSandblasted and finishedwith 2-part high qualityepoxy paint with glossyfinish.Identical
Capacity MainCompartment4 Patients Up to 24PatientsIdentical
Hyperbaric Chamber Interior
Lighting SubsystemLED lightsMin: 4Max: 15Chamber illuminationcomplies with paragraph14.2.3 of NFPA 99 (FDAconsensus std. recognitionno. 1-67).Identical
BIBS withOverboard DumpFour (4) to Twenty-Four(24), on demand gasdelivery.Identical
Hoods withOverboard DumpFour (4) to Twenty-Four(24), 1-100LPM deliveryIdentical
1-2 Hyperbaric Chamber
ManufacturerOxyHeal® Medical Systems, Inc.OxyHeal® Medical Systems, Inc.OxyHeal® 4000 Family Summary Comparison to Predicate (K152223)
Trade NameOxyHeal® 4000 Family (K163109)flow for hoods: 40-lpm.OxyHeal® 5000 Family (K152223)flow for hoods: 40-lpm.
Depth MeasurementDigital with analog backupDigital with analog backupIdentical
Pressure TransmitterRanges:Minimum = 0 psigMaximum = 75 psigRanges:Minimum = 0 psigMaximum = 75 psigIdentical
Temperature Sensor75°F ± 5°F75°F ± 5°FIdentical
Relative Humidity Sensor0% - 100%0% - 100%Identical
Television (TV) System24VDC LED TV Monitors. Customer specified. Two (2) or four (4) per treatment compartment24VDC LED TV Monitors. Customer specified. Two (2) or four (4) per treatment compartmentIdentical

{8}------------------------------------------------

1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

{9}------------------------------------------------

{10}------------------------------------------------

flow meters. Minimum

flow meters. Minimum

{11}------------------------------------------------

nsmitterMinimum = 0 psigMaximum = 75 psigMinimum = 0 psigMaximum = 75 psig
perature Sensor75°F ± 5°F75°F ± 5°FIdentical
ative Humiditysor0% - 100%0% - 100%Identical
vision (TV)em24VDC LED TVMonitors. Customerspecified.Two (2) or four (4) pertreatment compartment24VDC LED TV Monitors.Customer specified.Two (2) or four (4) pertreatment compartmentIdentical
1-3 Compressed Air System
ManufacturerOxyHeal® MedicalSystems, Inc.OxyHeal® MedicalSystems, Inc.OxyHeal® 4000 FamilySummary Comparison toPredicate (K152223)
Trade NameOxyHeal® 4000 Family(K163109)OxyHeal® 5000 Family(K152223)
Air CompressorsQty. two (2) rotary screwOperating range:Rated Power:208V ±10%, 3-ph, 60Hz -460V ±10%, 3-ph, 60HzMax Working Pressure:125psig - 217psigAir Delivery155 CFM - 288 CFMQty. two (2) rotary screwOperating range:Rated Power:208V ±10%, 3-ph, 60Hz -460V ±10%, 3-ph, 60HzMax Working Pressure:125psig - 217psigAir Delivery155 CFM - 288 CFMIdentical
Air ReceiverRanges: Qty. 1 – Qty.5depending on space andlocation within clientfacility.Ranges: Qty. 1 – Qty.5depending on space andlocation within clientfacility.Substantially equivalent
- Design CodeASME Boiler and PressureCodeASME Boiler and PressureCodeIdentical
- CalculationsASME Section VIII, Div. 1ASME Section VIII, Div. 1Identical
- Joint DesignJoint design and fabricationshall be performed inJoint design and fabricationshall be performed inIdentical
1-3 Compressed Air System
ManufacturerOxyHeal® MedicalSystems, Inc.OxyHeal® 4000 Family(K163109)OxyHeal® MedicalSystems, Inc.OxyHeal® 5000 Family(K152223)OxyHeal® 4000 FamilySummary Comparison toPredicate (K152223)
Trade Nameaccordance with UW-12and Table UW-12 ofASME Boiler and PressureVessel Code, Division, I,Section VIIIaccordance with UW-12and Table UW-12 ofASME Boiler and PressureVessel Code, Division, I,Section VIIIIdentical
Non--DestructiveExaminations(NDE)Radiographic examinationand ultrasonic examinationshall be performed inaccordance with UW-51 ofASME Boiler and PressureVessel Code, Division, I,Section VIIIRadiographic examinationand ultrasonic examinationshall be performed inaccordance with UW-51 ofASME Boiler and PressureVessel Code, Division, I,Section VIIIIdentical
Ultrasonic examinationshall be performed inaccordance with UW-53and Appendix 12 of ASMEBoiler and Pressure VesselCode, Division, I, SectionVIIIUltrasonic examinationshall be performed inaccordance with UW-53and Appendix 12 of ASMEBoiler and Pressure VesselCode, Division, I, SectionVIIIIdentical
Hydrostatic-TestHydrostatic testing to beperformed in accordancewith ASME Boiler andPressure Vessel Code,Section UG-99 @ 1.3 timesmaximum allowableworking pressure.Hydrostatic testing to beperformed in accordancewith ASME Boiler andPressure Vessel Code,Section UG-99 @ 1.3 timesmaximum allowableworking pressure.Identical
Pressure VesselReliefOne (1) pressure reliefsatisfying the requirementsof ASME Boiler andPressure Vessel Code,Section UG-125.Rated at 200 psigOne (1) pressure reliefsatisfying the requirementsof ASME Boiler andPressure Vessel Code,Section UG-125.Rated at 200 psigIdentical
FiltersIdentical
Liquid Separator-Removes moisture from airusing centrifugal forceRemoves moisture from airusing centrifugal forceIdentical
Prefilter Filter-5 micron element5 micron elementIdentical
Coalescing Filter-1 micron element1 micron elementIdentical
Active Carbon-FilterCharcoal elementCharcoal elementIdentical
1-3 Compressed Air System
ManufacturerOxyHeal® MedicalSystems, Inc.OxyHeal® MedicalSystems, Inc.OxyHeal® 4000 FamilySummary Comparison to
Trade NameOxyHeal® 4000 Family(K163109)OxyHeal® 5000 Family(K152223)Predicate (K152223)
Particulate Filter3 micron element3 micron elementIdentical
High EfficiencyFilter.01 micron element.01 micron elementIdentical
PipingASTM B88ASTM B88Identical
TestingOMS Factory AcceptanceTest (FAT)OMS Factory AcceptanceTest (FAT)Substantially Equivalent

{12}------------------------------------------------

{13}------------------------------------------------

1-4 Fire Suppression System
ManufacturerOxyHeal® MedicalSystems, Inc.OxyHeal® MedicalSystems, Inc.OxyHeal® 4000 FamilySummary Comparison to
Trade NameOxyHeal® 4000 Family(K163109)OxyHeal® 5000 Family(K152223)Predicate (K152223)
Fire SuppressionSystem DesignCodeFire Protection complieswith paragraph 14.2.5 ofNFPA 99 (FDA consensusstd. recognition no. 1-67).Fire Protection complieswith paragraph 14.2.5 ofNFPA 99 (FDA consensusstd. recognition no. 1-67).Identical
Fire Deluge TankQuantity ranges based onnumber of chambercompartments:1 Compartment: Qty. 12 Compartment: Qty. 23 Compartment: Qty. 3Fire deluge systemcomplies with paragraph14.2.5.2 of NFPA 99 (FDAconsensus std. recognitionno. 1-67).Quantity ranges based onnumber of chambercompartments:1 Compartment: Qty. 12 Compartment: Qty. 23 Compartment: Qty. 3Fire deluge systemcomplies with paragraph14.2.5.2 of NFPA 99 (FDAconsensus std. recognitionno. 1-67).Identical
Design Code-ASME Boiler and PressureCode, Section VIII DivisionIASME Boiler and PressureCode, Section VIII DivisionIIdentical
Calculations-ASME Section VIII, Div. 1ASME Section VIII, Div. 1Identical
Joint Design-Joint design and fabricationshall be performed inaccordance with UW-12and Table UW-12 ofJoint design and fabricationshall be performed inaccordance with UW-12and Table UW-12 ofIdentical
1-4 Fire Suppression System
ManufacturerTrade NameOxyHeal® MedicalSystems, Inc.OxyHeal® 4000 Family(K163109)OxyHeal® MedicalSystems, Inc.OxyHeal® 5000 Family(K152223)OxyHeal® 4000 FamilySummary Comparison toPredicate (K152223)
ASME Boiler and PressureVessel Code, Division, I,Section VIIIASME Boiler and PressureVessel Code, Division, I,Section VIII
Non-DestructiveExaminations(NDE)Radiographic examinationand ultrasonic examinationshall be performed inaccordance with UW-51 ofASME Boiler and PressureVessel Code, Division, I,Section VIIIRadiographic examinationand ultrasonic examinationshall be performed inaccordance with UW-51 ofASME Boiler and PressureVessel Code, Division, I,Section VIIIIdentical
Ultrasonic examinationshall be performed inaccordance with UW-53and Appendix 12 of ASMEBoiler and Pressure VesselCode, Division, I, SectionVIIIUltrasonic examinationshall be performed inaccordance with UW-53and Appendix 12 of ASMEBoiler and Pressure VesselCode, Division, I, SectionVIII
HydrostaticTestHydrostatic testing to beperformed in accordancewith ASME Boiler andPressure Vessel Code,Section VIII, UG-99 @ 1.3times max. allowableworking pressure.Hydrostatic testing to beperformed in accordancewith ASME Boiler andPressure Vessel Code,Section VIII, UG-99 @ 1.3times max. allowableworking pressure.Identical
Pressure VesselReliefOne (1) pressure reliefsatisfying the requirementsof ASME Boiler andPressure Vessel Code,Section UG-125.Rated at 200 psigOne (1) pressure reliefsatisfying the requirementsof ASME Boiler andPressure Vessel Code,Section UG-125.Rated at 200 psigIdentical
Hand Line TankHand Line system complieswith paragraph 14.2.5.3 ofNFPA 99 (FDA consensusstd. recognition no. 1-67).Hand Line system complieswith paragraph 14.2.5.3 ofNFPA 99 (FDA consensusstd. recognition no. 1-67).Identical
-Design CodeASME Boiler and PressureCode, Section VIII DivisionIASME Boiler and PressureCode, Section VIII DivisionIIdentical
-CalculationsASME Section VIII, Div. 1ASME Section VIII, Div. 1Identical
-Design CodeJoint design and fabricationshall be performed inJoint design and fabricationshall be performed inIdentical
1-4 Fire Suppression System
ManufacturerOxyHeal® Medical Systems, Inc.OxyHeal® 4000 FamilyOxyHeal® Medical Systems, Inc.OxyHeal® 5000 FamilyOxyHeal® 4000 Family Summary Comparison to Predicate (K152223)
Trade Name(K163109)(K152223)
accordance with UW-12and Table UW-12 of ASME Boiler and Pressure Vessel Code, Division, I, Section VIIIaccordance with UW-12and Table UW-12 of ASME Boiler and Pressure Vessel Code, Division, I, Section VIII
- Joint DesignRadiographic examination and ultrasonic examination shall be performed in accordance with UW-51 of ASME Boiler and Pressure Vessel Code, Division, I, Section VIIIUltrasonic examination shall be performed in accordance with UW-53 and Appendix 12 of ASME Boiler and Pressure Vessel Code, Division, I, Section VIIIRadiographic examination and ultrasonic examination shall be performed in accordance with UW-51 of ASME Boiler and Pressure Vessel Code, Division, I, Section VIIIUltrasonic examination shall be performed in accordance with UW-53 and Appendix 12 of ASME Boiler and Pressure Vessel Code, Division, I, Section VIIIIdentical
- Non-Destructive Examinations (NDE)NDE shall be performed in accordance with ASME/PVHO-1, para. 1-7.3NDE shall be performed in accordance with ASME/PVHO-1, para. 1-7.3Identical
- Hydrostatic TestHydrostatic testing to be performed in accordance with ASME Boiler and Pressure Vessel Code, Section UG-99 @ 1.3 times max. allowable working pressure.Hydrostatic testing to be performed in accordance with ASME Boiler and Pressure Vessel Code, Section UG-99 @ 1.3 times max. allowable working pressure.Identical
- Pressure Vessel ReliefOne (1) pressure relief satisfying the requirements of ASME Boiler and Pressure Vessel Code, Section UG-125.Rated at 200 psigOne (1) pressure relief satisfying the requirements of ASME Boiler and Pressure Vessel Code, Section UG-125.Rated at 200 psigIdentical
- PipingASTM B88ASTM B88Identical
- TestingOMS Factory Acceptance Test (FAT)OMS Factory Acceptance Test (FAT)Substantially Equivalent

{14}------------------------------------------------

{15}------------------------------------------------

-て . (

{16}------------------------------------------------


1-5 Oxygen Delivery Requirements
ManufacturerOxyHeal® MedicalSystems, Inc.OxyHeal® MedicalSystems, Inc.OxyHeal® 4000 FamilySummary Comparison toPredicate (K152223)
Trade NameOxyHeal® 4000 Family(K163109)OxyHeal® 5000 Family(K152223)
Facility SuppliedGround Storage andBulk Liquid O2(Customer Provided)Ranges:Gallons: 500-13,000SCF: 50K - 1.3M100psi - 150PSIPermitted by AuthorityHaving Jurisdiction (AHJ)(Customer Provided)Ranges:Gallons: 500-13,000SCF: 50K - 1.3M100psi - 150PSIPermitted by AuthorityHaving Jurisdiction (AHJ)Identical
or
Microbulk(Customer Provided)Ranges:Gallons: 230-715SCF: 5,658 - 66,592100psi - 150PSIPermitted by AuthorityHaving Jurisdiction (AHJ)(Customer Provided)Ranges:Gallons: 230-715SCF: 5,658 - 66,592100psi - 150PSIPermitted by AuthorityHaving Jurisdiction (AHJ)Identical
PipingASTM B819ASTM 819Identical
1-6 Built In Breathing System (BIBS)
ManufacturerOxyHeal® MedicalSystems, Inc.OxyHeal® MedicalSystems, Inc.OxyHeal® 4000 FamilySummary Comparison to
Trade NameOxyHeal® 4000 Family(K163109)OxyHeal® 5000 Family(K152223)Predicate (K152223)
HP OxygenCylinder Size: KPressure: 3000 psigCGA 346Cylinder Size: KPressure: 3000 psigCGA 346Identical
HP Medical AirCylinder Size: KPressure: 3000 psigCGA 346Cylinder Size: KPressure: 3000 psigCGA 346Identical
HP Mixed Gas (e.g.)- Nitrogen- Nitrox- HeliumCylinder Size: KPressure: 3000 psig- CGA 580- CGA 326- CGA 580Cylinder Size: KPressure: 3000 psig- CGA 580- CGA 326- CGA 580Identical
PipingASTM B819ASTM B819Identical
TestingOMS Factory AcceptanceTest (FAT)OMS Factory AcceptanceTest (FAT)Substantially Equivalent

{17}------------------------------------------------

1-7 Environmental Control System (ECS)
ManufacturerOxyHeal® MedicalSystems, Inc.OxyHeal® MedicalSystems, Inc.OxyHeal® 4000 FamilySummary Comparison toPredicate (K152223)
Trade NameOxyHeal® 4000 Family(K163109)OxyHeal® 5000 Family(K152223)
EnvironmentalControl SystemChamber temperature andhumidity control complieswith paragraph 14.2.4.3 ofNFPA 99 (FDA consensusstd. recognition no. 1-67).Chamber temperature andhumidity control complieswith paragraph 14.2.4.3 ofNFPA 99 (FDA consensusstd. recognition no. 1-67).Identical
EnvironmentalControl Unit (ECU)Quantity ranges based onnumber of chambercompartments:1 Compartment: Qty. 12 Compartment: Qty. 23 Compartment: Qty. 3Quantity ranges based onnumber of chambercompartments:1 Compartment: Qty. 12 Compartment: Qty. 23 Compartment: Qty. 3Identical
ECU PhysicalDimensions24"L X 24"W X 36"H24"L X 24"W X 36"HIdentical
ECU Heat Load(Heat Mode)~ 12000 BTU/HR~ 12000 BTU/HRIdentical
ECU Heat Load(Heat Mode)~ 16000 BTU/HR~ 16000 BTU/HRIdentical
ECU Coolant PumpMotor2.2 AMPS2.2 AMPSIdentical
ECU Coolant PumpFlow Rate4.2 GPM @ 24 FT HEAD4.2 GPM @ 24 FT HEADIdentical
TestingOMS Factory AcceptanceTest (FAT)OMS Factory AcceptanceTest (FAT)Substantially Equivalent
1-8 Control Console
ManufacturerOxyHeal® MedicalSystems, Inc.OxyHeal® MedicalSystems, Inc.OxyHeal® 4000 FamilySummary Comparisonto Predicate (K152223)
Trade NameOxyHeal® 4000 Family(K163109)OxyHeal® 5000 Family(K152223)
HMI TouchscreenMonitorQuantity ranges based onnumber of chambercompartments (Comp):1 Comp: Qty. 12 Comp: Qty. 1-23 Comp: Qty. 2-3Quantity ranges based onnumber of chambercompartments (Comp):1 Comp: Qty. 12 Comp: Qty. 1-23 Comp: Qty. 2-3Identical
Primary method ofcontrolling chamber Primary means formonitoring chambersystem statusPrimary method ofcontrolling chamber Primary means formonitoring chambersystem status
Life SupportControlsAutomatic pressurization &depressurization withmanual back-up from bothinside and outside eachchamber compartment.Automatic pressurization &depressurization withmanual back-up from bothinside and outside eachchamber compartment.Identical
EmergencyDepressurizationChamber depressurizationcomplies with paragraph14.2.4.5 of NFPA 99 (FDAconsensus std. recognitionno. 1-67).Chamber depressurizationcomplies with paragraph14.2.4.5 of NFPA 99 (FDAconsensus std. recognitionno. 1-67).Identical
VentilationAutomatic chamberventilation with manualback-up. Ranges are asfollows:- Min = 6 cfm/min- Max = 48 cfm/min ± 1FSW stabilityChamber ventilationcomplies with paragraph14.2.4.1 of NFPA 99 (FDAconsensus std. recognitionno. 1-67).Automatic chamberventilation with manualback-up. Ranges are asfollows:- Min = 6 cfm/min- Max = 48 cfm/min ± 1FSW stabilityChamber ventilationcomplies with paragraph14.2.4.1 of NFPA 99 (FDAconsensus std. recognitionno. 1-67).Identical
ProgrammableLogic Controller(PLC)Controls the operatorinitiated actions at the HMItouch screen bycommunicating with variousControls the operatorinitiated actions at the HΜΙtouch screen bycommunicating with variousIdentical
1-8 Control Console
ManufacturerOxyHeal® MedicalSystems, Inc.OxyHeal® MedicalSystems, Inc.OxyHeal® 4000 FamilySummary Comparisonto Predicate (K152223)
Trade NameOxyHeal® 4000 Family(K163109)OxyHeal® 5000 Family(K152223)
devices managing theperformance of chambersubsystemsdevices managing theperformance of chambersubsystems
Keyboard andMouse #1 and #2Alternate method forproviding input to controlhyperbaric chamberAlternate method forproviding input to controlhyperbaric chamberIdentical
Quantity ranges based onnumber of chambercompartments (Comp):1 Comp: Qty. 12 Comp: Qty. 1-23 Comp: Qty. 2-3Quantity ranges based onnumber of chambercompartments (Comp):1 Comp: Qty. 12 Comp: Qty. 1-23 Comp: Qty. 2-3
Patient Monitor,Quad ScreenVisual method formonitoring patientsundergoing HBOTVisual method formonitoring patientsundergoing HBOTIdentical
Quantity ranges based onnumber of chambercompartments (Comp):1 Comp: Qty. 12 Comp: Qty. 1-23 Comp: Qty. 2-3Quantity ranges based onnumber of chambercompartments (Comp):1 Comp: Qty. 12 Comp: Qty. 1-23 Comp: Qty. 2-3
Fire SuppressionSystem (FSS)ActivationPushbuttonWhen depressed, activatesFSS and illuminates greenWhen depressed, activatesFSS and illuminates greenIdentical
Quantity ranges based onnumber of chambercompartments (Comp):1 Comp: Qty. 12 Comp: Qty. 23 Comp: Qty. 3Quantity ranges based onnumber of chambercompartments (Comp):1 Comp: Qty. 12 Comp: Qty. 23 Comp: Qty. 3
CommunicationsSystemPrimary: Wireless telephoneSecondary: IntercomTertiary: (OPTION) Soundpowered backupPrimary: Wireless telephoneSecondary: IntercomTertiary: (OPTION) Soundpowered backupIdentical
Gas Analysis SystemDetermines the O2 and CO2from within a treatmentcompartment or from theBIBS gases.Determines the O2 and CO2from within a treatmentcompartment or from theBIBS gases.Identical
1-8 Control Console
ManufacturerOxyHeal® Medical Systems, Inc.OxyHeal® 4000 Family (K163109)OxyHeal® Medical Systems, Inc.OxyHeal® 5000 Family (K152223)OxyHeal® 4000 Family Summary Comparisonto Predicate (K152223)
Trade Name
Monitor w/SDA Carbon DioxideMonitor (OPTION add on)Monitor w/SDA Carbon DioxideMonitor (OPTION add on)
O₂ @ 0% - 100% ± 1%CO₂ @ 0 – 5250NOTE parts per million(ppm) ± 25ppm.O₂ @ 0% - 100% ± 1%CO₂ @ 0 – 5250NOTE parts per million(ppm) ± 25ppm.
Oxygen monitoringcomplies with paragraph14.2.8.4 of NFPA 99 (FDAconsensus std. recognitionno. 1-67).Oxygen monitoringcomplies with paragraph14.2.8.4 of NFPA 99 (FDAconsensus std. recognitionno. 1-67).
FlowmetersControls the flow of gas tothe gas analyzer so as not todamage the analyzersinternal parts.Controls the flow of gas tothe gas analyzer so as not todamage the analyzersinternal parts.Identical
Selector SwitchTwo (2) per compartmentUser adjustable and used forselecting between zero gas,calibration gas or gas flowfrom the chambercompartment.Two (2) per compartmentUser adjustable and used forselecting between zero gas,calibration gas or gas flowfrom the chambercompartment.Identical
One (1) per compartmentOne (1) per compartment
PatientEntertainmentSystemComponents as listed below.Components as listed below.Identical
Television (TV)—SystemDisplays digital video onchamber interior mountedTVsDisplays digital video onchamber interior mountedTVsIdentical
DVD Player—Provides digital video topatient entertainmentmonitor inside chamberProvides digital video topatient entertainmentmonitor inside chamberIdentical
CD Player—Provides digital audio tospeaker of patientheadphones inside chamberProvides digital audio tospeaker of patientheadphones inside chamberIdentical
AM/FM Tuner—Provides audio to patientheadphones inside chamberProvides audio to patientheadphones inside chamberIdentical
1-8 Control Console
ManufacturerOxyHeal® MedicalSystems, Inc.OxyHeal® Medical Systems, Inc.OxyHeal® 4000 FamilySummary Comparisonto Predicate (K152223)
Trade NameOxyHeal® 4000 Family(K163109)OxyHeal® 5000 Family(K152223)
4-Zone MixerIndividual up to 4-channels.Varies based on customerspecification for number ofcompartments and numberof patients to be treated.Individual up to 4-channels.Varies based on customerspecification for number ofcompartments and numberof patients to be treated.Identical
AmplifierAmplifies the outputs of theAM/FM tuner, CD/DVD tothe chamber interior and thepatient headsetsAmplifies the outputs of theAM/FM tuner, CD/DVD tothe chamber interior and thepatient headsetsIdentical
Audio / VideoRecorderPermits customer ability torecord patient video and / oraudio from within acompartment.Permits customer ability torecord patient video and / oraudio from within acompartment.Identical
UninterruptedPower Supply (UPS)OPTION - CustomerMeets NFPA-99, para.14.2.5.4.3 for automaticbattery back-up.See also NFPA-99:14.2.5.1.3 (3) for emergencycommunications andemergency lighting. 14.2.3.3for emergency lighting.14.2.5.4.3 for fire delugeOPTION - CustomerMeets NFPA-99, para.14.2.5.4.3 for automaticbattery back-up.See also NFPA-99:14.2.5.1.3 (3) for emergencycommunications andemergency lighting. 14.2.3.3for emergency lighting.14.2.5.4.3 for fire delugeIdentical
TestingOMS Factory AcceptanceTest (FAT)OMS Factory AcceptanceTest (FAT)Substantially Equivalent

{18}------------------------------------------------

T

{19}------------------------------------------------

I

{20}------------------------------------------------


Г

Г

{21}------------------------------------------------

7. Intended Use and indications for Use

The intended use of the OxyHeal® 4000 Cylindrical Multiplace Hyperbaric Chamber System Product Family is to administer hyperbaric oxygen therapy (HBOT) to treat patients with any of the below listed indications.

The following indications which are listed on the Undersea & Hyperbaric Medical Society (UHMS) web site: www.uhms.org.are approved uses of hyperbaric oxygen therapy as defined by the Hyperbaric Oxygen Therapy Committee.

    1. Air or Gas Embolism
    1. Carbon Monoxide Poisoning
    • a. Carbon Monoxide Poisoning Complicated by Cyanide Poisoning
    1. Clostridial Myositis and Myonecrosis (Gas Gangrene)

{22}------------------------------------------------

    1. Crush Injury, Compartment Syndrome and Other Acute Ischemias
    1. Decompression Sickness
    1. Arterial Insufficiencies
    • a. Central Retinal Artery Occlusion
    • b. Enhancement of Healing in Selected Problem Wounds
    1. Severe Anemia
    1. Intracranial Abscess
    1. Necrotizing Soft Tissue Infections
    1. Osteomyelitis (Refractory)
    1. Delayed Radiation Injury (Soft Tissue and Bony Necrosis)
    1. Compromised Grafts and Flaps
    1. Acute Thermal Burn Injury

There is no difference between the OxyHeal® 4000 Cylindrical Multiplace Hyperbaric Chamber System Product Family and the OxyHeal® 5000 Rectangular Multiplace Hyperbaric Chamber System Product Family in regards to intended use or these indications for use.

8. Technological Characteristics and Substantial Equivalence

The OxyHeal® 4000 Cylindrical Multiplace Hyperbaric Chamber System Product Family in this Special 510(k) submittal consists of a minor modification of the OxyHeal® 5000 Rectangular Multiplace Hyperbaric Chamber System Product Family predicate device (K152223).

The fundamental technology for this minor modification consists of a change in dimensional specifications; i.e. the predicate device (K152223) is a multiplace hyperbaric chamber with a rectangular geometry. The modified device is a multiplace hyperbaric chamber with a cylindrical geometry. This modification does not change the intended use, indications for use, and product labeling. The design of the modified device conforms to the same three FDA recognized standards as the predicate device; two (2) of which are the design and manufacturing standards / codes. The company's design control procedures are the same for the predicate and modified device and conform to the requirements as specified in 21 CFR 820.30.

Table 1 provides a substantially equivalent comparison between the OxyHeal® 4000 Cylindrical Multiplace Hyperbaric Chamber System Product Family and the OxyHeal® 5000 Rectangular Multiplace Hyperbaric Chamber System Product Family; thereby providing more detailed information regarding the basis for the determination of substantial equivalence.

9. Non-Clinical Testing

In accordance with the requirements of the ASME PVHO-1 standard (the FDA consensus standard recognition number 1-78), the only test methods and /or examinations required for verifying the modification of the pressure vessel for human occupancy [(PVHO) or hyperbaric chamber] from a rectangular to a cylindrical geometry consists of certifying the weld integrity of the joint design and the integrity of the entire PVHO consist of non-destructive testing and hydrostatic testing. These are the identical tests required by the aforementioned standard and also performed on the OxyHeal® 5000 hyperbaric chambers (predicate device K152223). A description of the non-destructive tests and the hydrostatic test conducted for all OxyHeal® 4000 hyperbaric chambers is described below.

{23}------------------------------------------------

9.1 Non-destructive Testing

There are four different types of non-destructive examinations (NDEs) which are used to verify the integrity of the structural welds comprising the OxyHeal® 4000 hyperbaric chamber structure.

  • Penetrant Examination (PE) is a color contrast procedure used to identify surface weld defects. a. OMS welders apply this liquid penetrant by spraying onto the surface of a root pass (1" pass) weld to ensure that the very base of the weld is without imperfections.
  • b. Ultrasonic Testing (UT) is a technique used to test the integrity of a weld using the propagation of ultrasonic sound waves into the steel material to detect internal flaws. This testing is contracted to an ASME qualified independent 3th party testing organization who provides a written copy of the results of this testing.
  • Radiographic Examinations (RT) is a method employed for weld testing which makes use of C. X-rays to verify the internal structure and integrity of the welded steel material. This testing is contracted to an ASME qualified independent 3th party testing organization who provides a written copy of the results of this testing.
  • d. Magnetic Particle Examination (MT) is process which propagates a magnetic field into the welded steel and is used for detecting surface and slightly subsurface discontinuities. This testing is contracted to an ASME qualified independent 3d party testing organization who provides a written copy of the results of this testing.

Hydrostatic Testing 9.2

A Standard Hydrostatic Test is conducted in accordance with the ASME Section VIII, Division 1 Code paragraph UG-99. This test consists of completely filling the OxyHeal 4000 hyperbaric chamber with water and pressure testing at 1.3 times the maximum allowable working pressure (MAWP) of the PVHO. Each viewport is installed in its location and hydrostatically tested as part of the overall structural test. This test is witnessed by an ASME qualified independent 3d party authorized inspector (AI). The AI will also review/ approve the nameplate to be affixed on the PVHO to ensure all applicable data, the certification mark, and designator are stamped into the nameplate. Upon successful completion of this testing and review of the nameplate data, the manufacturer (OxyHeal® Medical Systems, Inc.) and the AI will sign the Form U-1 Manufacturer's Data Report for Pressure Vessels as required by the provisions of the ASME Boiler and Pressure Vessel Code Rules, Section VIII, Division 1. Once the Form U-i1 has been signed, OMS is authorized to affix the nameplates to the PVHO.

In parallel, OxyHeal® Medical Systems, Inc. is responsible for signing the FORM PVHO-1 Manufacturer's Data Report for Pressure Vessels for Human Occupancy as required by the provisions of ASME PVHO-1 certifying the design and certification of compliance of the PVHO.

9.3 Other V&V Testing

The validation / verification efforts performed for the OxyHeal® 4000 Cylindrical Multiplace Hyperbaric Chamber System Product Family are identical to those performed for the OxyHeal® 5000 Rectangular Multiplace Hyperbaric Chamber System Product Family (K152223) predicate device.

The following V&V activities listed in this Special 510(k) provide an overview of the V&V activities performed for the predicate device (K152223) and apply to the OxyHeal® 4000 Cylindrical Multiplace Hyperbaric Chamber System Product Family.

{24}------------------------------------------------

9.3.1 Fire Suppression System Testing

A Fire Suppression System (FSS) test was conducted at the completion of the hydrostatic test to ensure that the fire deluge system water spray system and the hand line met the requirements of the FDA recognized consensus standard NFPA 99 (Recognition Number 1-67).

First Operational System Test 9.3.2

A first operational system test (FOST) was performed to verify that the system design met each of the specification requirements. This testing includes the following items specified in the OxyHeal® 4000 hyperbaric chamber system product family as defined in the User Design Specification satisfies the requirements of the FDA recognized consensus standard: ASME PVHO-1 (Recognition Number 1-78).

  • a. Testing of the minimum and maximum pressurization rates
  • Testing of the minimum and maximum depressurization rates b.
  • Testing of the minimum and maximum ventilation rates C.
  • Testing of the conditions under which these rates are to be maintained d.
  • Testing of the patient gas delivery systems and flow meter range e.
  • f. Testing of the chamber pressurization and ventilation gas for meeting requirements for CGA Grade E

9.3.3 Software Validation Testing.

A software validation test was conducted to validate that observed output of designated hyperbaric chamber control functions met the output that they were designed to perform. This testing includes the following items specified in the OxyHeal® 4000 hyperbaric chamber system product family as defined in the User Design Specification.

  • Testing of the minimum and maximum pressurization rates a.
  • Testing of the minimum and maximum depressurization rates b.
  • Testing of the minimum and maximum ventilation rates C.
  • d. Testing of the conditions under which these rates are to be maintained
  • e. Testing of the patient gas delivery systems and flow meter range

9.3.4 Factory Acceptance Test

A Factory Acceptance Test (FAT) was performed to verify that the system is able to perform all required operational functions. The FAT is witnessed by and signed off by an independent 3rd party Authorized Inspector (AI). This testing satisfies the requirements of the FDA recognized consensus standards: ASME PVHO-1 (Recognition Number 1-78).and NFPA 99 (Recognition Number 1-67).

10 Conclusion

It has been shown in this Special 510(k) submission that the OxyHeal® 4000 Cylindrical Multiplace Hyperbaric Chamber System Product Family as designed, manufactured, and tested does not raise any different questions regarding its safety and effectiveness, there is no difference in the indications and intended use, is designed to the same FDA recognized consensus standards, and is determined to be substantially equivalent to the OxyHeal® 5000 Rectangular Multiplace Hyperbaric Chamber System Product Family predicate device (K152223).

§ 868.5470 Hyperbaric chamber.

(a)
Identification. A hyperbaric chamber is a device that is intended to increase the environmental oxygen pressure to promote the movement of oxygen from the environment to a patient's tissue by means of pressurization that is greater than atmospheric pressure. This device does not include topical oxygen chambers for extremities (§ 878.5650).(b)
Classification. Class II (performance standards).