K Number
K163031

Validate with FDA (Live)

Manufacturer
Date Cleared
2017-07-27

(269 days)

Product Code
Regulation Number
870.5150
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Bunegin-Albin Air Aspiration Set is intended to aspirate venous air emboli.

Device Description

The Bunegin-Albin Air Aspiration Set is comprised of an air aspiration catheter, wire guide, entry access needle, dilator, syringe, stopcock, and scalpel. The air aspiration catheter is manufactured from radiopaque extruded polyethylene tubing and is designed with a pre-molded winged fitting. The catheter is 5.8 French in size with an endhole diameter of 0.035 inches. The wire guide is manufactured from stainless steel coils, a safety wire, and a mandril wire. The access needle is manufactured from stainless steel. The dilator is manufactured from polyethylene and has a pre-molded proximal hub. The set also includes a 12 cc syringe, a low pressure three way stopcock, and a scalpel. The Bunegin-Albin Air Aspiration Set is sterilized by ethylene oxide and intended for one-time use.

AI/ML Overview

This documentation outlines the acceptance criteria and the results of various studies conducted for the Bunegin-Albin Air Aspiration Set to demonstrate its performance and substantial equivalence to a predicate device.

1. Table of Acceptance Criteria and Reported Device Performance

TestAcceptance CriteriaReported Device Performance
Performance (Zero Time)
Catheter Shaft TensilePeak load of catheter shaft section ≥ 15 N (BS EN ISO 10555-1)Acceptance criterion met.
Catheter Sideport TensilePeak load of catheter sideport section ≥ 15 N (BS EN ISO 10555-1)Acceptance criterion met.
Catheter Hub-to-Shaft TensilePeak load of catheter hub-to-shaft section ≥ 15 N (BS EN ISO 10555-1)Acceptance criterion met.
Catheter Air AspirationNo air shall enter the hub (Annex D of BS EN ISO 10555-1)Acceptance criterion met.
Catheter Liquid LeakageNo part of the catheter shall leak liquid (Annex C of BS EN ISO 10555-1)Acceptance criterion met.
Catheter Kink LengthCatheter shall not kink (Annex B of BS EN 13868)Acceptance criterion met.
Wire Guide CorrosionNo evidence of corrosion that could affect functional performance (Annex B of BS EN ISO 11070)Acceptance criterion met.
Wire Guide FlexNo signs of defects or damage, including flaking or material loss (Annex G of BS EN ISO 11070)Acceptance criterion met.
Wire Guide FractureWire guide shall not fracture when wound around an approximate former (Annex F of BS EN ISO 11070)Acceptance criterion met.
Wire Guide TensilePeak load to failure ≥ 10 N (Annex H of BS EN ISO 11070)Acceptance criterion met.
Wire Guide Rotations to FailureCharacterized for the wire guide.The number of rotations to failure was characterized. (Specific values not provided in this document but implied to be sufficient for meeting overall acceptance related to design input requirements.)
Wire Guide Tip DeflectionCharacterized for the wire guide.The tip deflection side loads were characterized. (Specific values not provided in this document but implied to be sufficient for meeting overall acceptance related to design input requirements.)
Dilator Shaft TensilePeak load of dilator shaft section ≥ 15 N (BS EN ISO 11070)Acceptance criterion met.
Dilator Hub-to-Shaft TensilePeak load of dilator hub-to-shaft section ≥ 15 N (BS EN ISO 11070)Acceptance criterion met.
Dimensional, Compatibility, Surface AnalysisAll measurements and dimensional requirements within listed tolerances. Set components compatible. External surface of effective length free of extraneous matter.Acceptance criterion met.
Particulate TestingMeet USP 788 thresholds for small volume injections.Acceptance criterion met.
Resistance to Overriding TestingTesting conducted in accordance with ISO 594-2.Acceptance criterion met.
Separation Force TestingTesting conducted in accordance with ISO 594-2.Acceptance criterion met.
Unscrewing Torque TestingTesting conducted in accordance with ISO 594-2.Acceptance criterion met.
Performance (3-Year Accelerated Aging)
Catheter Shaft TensilePeak load of catheter shaft section ≥ 15 N (BS EN ISO 10555-1)Acceptance criterion met.
Catheter Sideport TensilePeak load of catheter sideport section ≥ 15 N (BS EN ISO 10555-1)Acceptance criterion met.
Catheter Hub-to-Shaft TensilePeak load of catheter hub-to-shaft section ≥ 15 N (BS EN ISO 10555-1)Acceptance criterion met.
Catheter Air AspirationNo air shall enter the hub (Annex D of BS EN ISO 10555-1)Acceptance criterion met.
Catheter Liquid LeakageNo part of the catheter shall leak liquid (Annex C of BS EN ISO 10555-1)Acceptance criterion met.
Catheter Kink LengthCatheter shall not kink (based on a 20 mm kink radius) (Annex B of BS EN 13868)Acceptance criterion met.
Dilator Shaft TensilePeak load of dilator shaft section ≥ 15 N (BS EN ISO 11070)Acceptance criterion met.
Dilator Hub-to-Shaft TensilePeak load of dilator hub-to-shaft section ≥ 15 N (BS EN ISO 11070)Acceptance criterion met.
Animal Testing
Acute performance of aspiration catheter in animal modelCatheter shall receive a grading of "adequate" or "good" in each of the performance parameters, and it shall be able to remove air from the target sites.Acceptance criterion met.
Biocompatibility
Cytotoxicity, Sensitization, Intracutaneous Irritation, Acute Systemic Toxicity, Material-Mediated Pyrogenicity, Hemocompatibility (Hemolysis, Complement Activation, In Vivo Thrombogenicity)Testing performed per ISO 10993-1 and FDA guidance to ensure biocompatibility.Acceptance criterion met.

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the specific numerical sample sizes for each physical and chemical test (e.g., Catheter Shaft Tensile, Wire Guide Corrosion). Instead, the document refers to compliance with international standards (e.g., BS EN ISO 10555-1, BS EN ISO 11070, ISO 594-2, BS EN 13868, USP 788) which would implicitly define sample sizes for such tests.

For the Animal Testing, a specific sample size is not stated, but the testing evaluated the "acute performance" in an animal model. The provenance would be experimental ("prospective" if considered in the context of device development) rather than clinical human data. The country of origin for the animal study is not specified in the provided text.

The data for all engineering performance and biocompatibility tests is based on prospective testing conducted specifically for this device submission to demonstrate compliance with standards and design input requirements. The country of origin of this testing data is not explicitly stated, but it is for a device manufactured by Cook Incorporated, presumably in the US or one of their international facilities.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

For the engineering performance tests (tensile strength, kink resistance, leakage, etc.) and particulate testing, "ground truth" is established by adherence to quantitative specifications defined by international standards (e.g., ≥ 15 N, no air entry, no liquid leak, USP 788 thresholds). Therefore, these do not rely on human expert consensus in the same way clinical image interpretation might. The "experts" in this context would be the technicians and engineers performing and verifying the tests against the defined standards.

For Animal Testing, the grading of "adequate" or "good" for performance parameters and ability to remove air would imply evaluation by experts, likely veterinary surgeons or researchers with expertise in animal models and medical device performance. The number and specific qualifications of these experts are not provided.

For Biocompatibility testing, "ground truth" is established by the results of standardized tests per ISO 10993-1 and FDA guidance. Experts would be toxicologists and other scientific professionals interpreting these results, but their number and specific qualifications are not detailed.

4. Adjudication Method for the Test Set

For the physical, chemical, and mechanical performance tests, the adjudication method is direct comparison of measured values against predefined numerical or qualitative acceptance criteria outlined in international standards. There is no mention of a human consensus or specific adjudication for these quantitative evaluations beyond meeting the standard.

For the Animal Testing, the "grading of 'adequate' or 'good'" suggests an assessment, but the specific adjudication method (e.g., single expert, 2+1 consensus) is not detailed.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No MRMC comparative effectiveness study was done. The device is a physical medical instrument (an air aspiration set), not an AI or imaging diagnostic device that would typically involve human readers interpreting cases.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

Not applicable as this is a physical medical device. The "performance" studies are standalone in the sense that they evaluate the device's physical properties and functionality without human intervention being part of the device's performance mechanism, but rather as the operators and evaluators of the device.

7. Type of Ground Truth Used

The ground truth for the various tests is established through a combination of:

  • International Standards and Specifications: For most physical, mechanical, and chemical tests (tensile strength, kink length, leakage, particulate counts, etc.), the ground truth is defined by the acceptance criteria set forth in specific ISO and BS EN standards (e.g., BS EN ISO 10555-1, BS EN ISO 11070, ISO 594-2, BS EN 13868, USP 788 thresholds).
  • Empirical Observation/Expert Assessment: For animal testing, the ground truth is based on the observed "adequate" or "good" performance and the ability to remove air from target sites, likely evaluated by trained personnel.
  • Biocompatibility Standards: For biocompatibility, the ground truth is compliance with the requirements of ISO 10993-1 and FDA guidance, demonstrated through specific laboratory test results.

8. Sample Size for the Training Set

Not applicable in the context of device performance testing for substantial equivalence. There is no mention of a "training set" as this device does not involve machine learning or AI algorithms requiring data for training. The studies described are for verification and validation of the manufactured product.

9. How the Ground Truth for the Training Set was Established

Not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name encircling a symbol. The symbol consists of three stylized human profiles stacked on top of each other. The profiles are rendered in a dark color, contrasting with the white background.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 27, 2017

Cook Incorporated Johnathan Liu Regulatory Affairs Specialist 750 Daniels Way Bloomington, Indiana 47404

Re: K163031

Trade/Device Name: Bunegin-Albin Air Aspiration Set Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy Catheter Regulatory Class: Class II Product Code: DXE Dated: June 24, 2017 Received: June 27, 2017

Dear Johnathan Liu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

Kenneth J. Cavanaugh -S

for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

2

510(k) Number (if known)

K163031

Device Name Bunegin-Albin Air Aspiration Set

Indications for Use (Describe)

The Bunegin-Albin Air Aspiration Set is intended to aspirate venous air emboli.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Cook Medical. The word "COOK" is in large, white, sans-serif font on a red background. Below the word "COOK" is the word "MEDICAL" in a smaller, white, sans-serif font. The logo is simple and clean, and the red and white color scheme is eye-catching.

COOK INCORPORATED 750 DANIELS WAY, P.O. BOX OMINGTON, IN 474

2.0 510(k) SUMMARY

K163031 Bunegin-Albin Air Aspiration Set 21 CFR §870.5150 Date Prepared: July 25, 2017

Submitted By:

Submission:Traditional 510(k) Premarket Notification
Applicant:Cook Incorporated
Applicant Address:Cook Incorporated750 Daniels WayBloomington, IN 47404
Contact:Johnathan Liu
Email:regsubmissions@cookmedical.com
Contact Phone Number:(812) 335-3575 x104509
Contact Fax Number:(812) 332-0281

Device Information:

Trade Name:Bunegin-Albin Air Aspiration Set
Common Name:Catheter, Embolectomy
Classification Name:Embolectomy Catheter
Regulation:21 CFR §870.5150
Product Code:DXE
Device Class:II
Classification Panel:Cardiovascular

Predicate Device:

The Bunegin-Albin Air Aspiration Set is substantially equivalent to the following device: the Export Advance" Aspiration Catheter (K130536, Medtronic Inc.) cleared on July 16, 2013.

Device Description:

The Bunegin-Albin Air Aspiration Set is comprised of an air aspiration catheter, wire guide, entry access needle, dilator, syringe, stopcock, and scalpel. The air aspiration catheter is manufactured from radiopaque extruded polyethylene tubing and is designed

{4}------------------------------------------------

with a pre-molded winged fitting. The catheter is 5.8 French in size with an endhole diameter of 0.035 inches. The wire guide is manufactured from stainless steel coils, a safety wire, and a mandril wire. The access needle is manufactured from stainless steel. The dilator is manufactured from polyethylene and has a pre-molded proximal hub. The set also includes a 12 cc syringe, a low pressure three way stopcock, and a scalpel. The Bunegin-Albin Air Aspiration Set is sterilized by ethylene oxide and intended for onetime use.

Indications for Use:

The Bunegin-Albin Air Aspiration Set is intended to aspirate venous air emboli.

Comparison to Predicate Device:

The Bunegin-Albin Air Aspiration Set and the predicate device, the Export Advance™ Aspiration Catheter (K130536), are substantially equivalent in that these devices are similar in intended use and principle of operation. Additionally, the subject device has similar indications for use, materials, and dimensions as the predicate device. The differences between the subject device and the predicate device, including the materials, dimensions, and indications for use do not raise any new issues of safety and effectiveness. The substantial equivalence comparison of the subject device to the predicate is provided in the table below.

{5}------------------------------------------------

PREDICATE DEVICESUBJECT DEVICE
Export AdvanceTM Aspiration CatheterBunegin-Albin Air Aspiration Set
ManufacturerMedtronic IncorporatedCook Incorporated
510(k)K130536Subject of this submission
Regulation21 CFR §870.5150Identical
Product CodeDXEIdentical
RegulationDescriptionCatheter, EmbolectomyIdentical
ClassificationIIIdentical
Indications forUseThe Export AdvanceTM Aspiration Catheteris indicated for:- Removal/aspiration of embolicmaterial (thrombus/debris) fromvessels of the arterial system, and- To sub selectively infuse/deliverdiagnostics or therapeutics agentswith or without vessel occlusion.To aspirate venous air emboli.
Principle ofOperationManually aspirate emboli with syringeIdentical
VisualizationGuidanceFluoroscopyIdentical
Method of UseSingle useIdentical
CatheterPlacementPercutaneously using Seldinger techniqueunder fluoroscopyIdentical
CatheterMaterialPolytetrafluoroethylene (PTFE)Radiopaque extruded polyethylene
Catheter OuterDiameter6 French5.8 French
CatheterLength140 cm24, 60 cm
Catheter DistalEndOpenOpen with 6 side ports
CatheterLumensDual lumenSingle lumen
Wire GuideCompatibility0.014 inch, 0.035 inch0.035 inch
Accessory SetComponents2 Syringes1 Wire Guide1 Filter Cup1 Syringe1 Wire Guide1 Needle1 Dilator1 Stopcock1 Scalpel
SterilizationMethodEtOIdentical
SterilityAssuranceLevel10-6Identical
PackagingTyvek pouchTyvek Peel Pouch

{6}------------------------------------------------

Technological Characteristics:

The subject Bunegin-Albin Air Aspiration Set was subjected to applicable testing to assure reliable design and performance under the testing parameters. The tests are listed below:

Performance

  • Catheter Shaft Tensile (Zero time) The peak load of catheter shaft section shall be . greater than or equal to 15 N in accordance with BS EN ISO 10555-1. The acceptance criterion was met.
  • . Catheter Sideport Tensile (Zero time) – The peak load of catheter sideport section shall be greater than or equal to 15 N in accordance with BS EN ISO 10555-1. The acceptance criterion was met.
  • Catheter Hub-to-Shaft Tensile (Zero time) The peak load of catheter hub-to-shaft . section shall be greater than or equal to 15 N in accordance with BS EN ISO 10555-1. The acceptance criterion was met.
  • Catheter Air Aspiration (Zero time) No air shall enter the hub when tested in . accordance with Annex D of BS EN ISO 10555-1. The acceptance criterion was met.
  • Catheter Liquid Leakage (Zero time) No part of the catheter shall leak liquid when . tested in accordance with Annex C of BS EN ISO 10555-1. The acceptance criterion was met.
  • . Catheter Kink Length (Zero time) - Catheter shall not kink when tested in accordance with Annex B of BS EN 13868. The acceptance criterion was met.
  • . Wire Guide Corrosion (Zero time) – Wire guide shall show no evidence of corrosion that could affect the functional performance when tested in accordance with Annex B of BS EN ISO 11070. The acceptance criterion was met.
  • . Wire Guide Flex (Zero time) – Wire guide shall show no signs of defects or damage, including flaking or material loss when tested in accordance with Annex G of BS EN ISO 11070. The acceptance criterion was met.
  • Wire Guide Fracture (Zero time) Wire guide shall not fracture when wound around . an approximate former in accordance with Annex F of BS EN ISO 11070. The acceptance criterion was met.
  • . Wire Guide Tensile (Zero time) – The peak load to failure shall be greater than or equal to 10 N in accordance with Annex H of BS EN ISO 11070. The acceptance criterion was met.
  • Wire Guide Rotations to Failure Testing (Zero time) The number of rotations to . failure shall be characterized for the wire guide.

{7}------------------------------------------------

  • Wire Guide Tip Deflection (Zero time) The tip deflection side loads will be . characterized for the wire guide.
  • . Dilator Shaft Tensile (Zero time) – the peak load of dilator shaft section shall be greater than or equal to 15 N in accordance with BS EN ISO 11070. The acceptance criterion was met.
  • Dilator Hub-to-Shaft Tensile (Zero time) The peak load of dilator hub-to-shaft . section shall be greater than or equal to 15 N in accordance with BS EN ISO 11070. The acceptance criterion was met.
  • . Dimensional, Compatibility, and Surface Analysis (Zero-time) - All measurements and dimensional requirements shall be within the listed tolerances. The set components shall be compatible for each test specimen. The external surface of the effective length of the catheter, dilator, and wire guide shall appear free of extraneous matter for each test specimen. The acceptance criterion was met.
  • Particulate Testing (Zero-time) Each device must meet the USP 788 thresholds for . small volume injections. The acceptance criterion was met.
  • . Resistance to Overriding Testing (Zero-time) – Testing was conducted in accordance with ISO 594-2. The acceptance criterion was met.
  • . Separation Force Testing (Zero-time) - Testing was conducted in accordance with ISO 594-2. The acceptance criterion was met.
  • Unscrewing Torque Testing (Zero-time) Testing was conducted in accordance with . ISO 594-2. The acceptance criterion was met.
  • Catheter Shaft Tensile (3-year accelerated aging) The peak load of catheter shaft . section shall be greater than or equal to 15 N in accordance with BS EN ISO 1055-1. The acceptance criterion was met.
  • Catheter Sideport Tensile (3-year accelerated aging) The peak load of catheter . sideport section shall be greater than or equal to 15 N in accordance with BS EN ISO 10555-1. The acceptance criterion was met.
  • . Catheter Hub-to-Shaft Tensile (3-year accelerated aging) - The peak load of catheter hub-to-shaft section shall be greater than or equal to 15 N in accordance with BS EN ISO 10555-1. The acceptance criterion was met.
  • Catheter Air Aspiration (3-year accelerated aging) No air shall enter the hub when . tested in accordance with Annex D of BS EN ISO 10555-1. The acceptance criterion was met.

{8}------------------------------------------------

  • . Catheter Liquid Leakage (3-year accelerated aging) - No part of the catheter shall leak liquid when tested in accordance with Annex C of BS EN ISO 10555-1. The acceptance criterion was met.
  • . Catheter Kink Length (3-year accelerated aging) - Catheter shall not kink (based on a 20 mm kink radius) when tested in accordance with Annex B of BS EN 13868. The acceptance criterion was met.
  • Dilator Shaft Tensile (3-year accelerated aging) The peak load of dilator shaft . section shall be greater than or equal to 15 N in accordance with BS EN ISO 11070. The acceptance criterion was met.
  • Dilator Hub-to-Shaft Tensile (3-year accelerated aging) The peak load of dilator . hub-to-shaft section shall be greater than or equal to 15 N in accordance with BS EN ISO 11070. The acceptance criterion was met.

Animal Testing

  • Acute performance of aspiration catheter in animal model Catheter shall receive a grading of "adequate" or "good" in each of the performance parameters, and it shall be able to remove air from the target sites. The acceptance criterion was met.

Biocompatibility

  • Per ISO 10993-1 and FDA guidance, testing for cytotoxicity, sensitization, . intracutaneous irritation, acute systemic toxicity, material-mediated pyrogenicity, and hemocompatibility (hemolysis, complement activation, and in vivo thrombogenicity) were performed to ensure the biocompatibility of the subject device set. The acceptance criterion was met.

Conclusion:

The results of these tests support a conclusion that the Bunegin-Albin Air Aspiration Set met the design input requirements based on the intended use and support the conclusion that this device does not raise new questions of safety or effectiveness and is substantially equivalent to the predicate device, the Export Advance™ Aspiration Catheter (Medtronic Inc., K130536).

§ 870.5150 Embolectomy catheter.

(a)
Identification. An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.(b)
Classification. Class II (performance standards).