K Number
K162875

Validate with FDA (Live)

Date Cleared
2017-02-17

(126 days)

Product Code
Regulation Number
870.3375
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Sentry IVC Filter is indicated for the prevention of recurrent pulmonary embolism via percutaneous placement in the inferior vena cava in patients with a transient high risk of PE, in the following situations:

· Pulmonary thromboembolism when anticoagulants are contraindicated;

· Failure of anticoagulant therapy in thromboembolic diseases;

· Emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced.

Device Description

The Sentry IVC Filter is a bioconvertible IVC filter intended for percutaneous implantation in the IVC and designed to provide protection against PE in patients at transient risk of PE.

The Sentry IVC Filter is designed for use in inferior vena cavae with diameters between 16mm and 28mm and has a maximum deployed length of 57.7mm. It is preloaded in a Loading Tool that can be orientated for either left/right femoral vein or a right jugular vein approach and is delivered through a 7 French ID Introducer Sheath (max OD 9.75Fr).

The Sentry IVC Filter consists of a cylindrical Nitinol frame and a Filter Cone formed by 6 Filter Arms held together in the center of the IVC by means of a bioabsorbable filament. The Filter Cone that is designed to trap emboli and thereby reduce the risk of PE while maintaining caval patency after it has converted. The Sentry IVC Filter converts into a non-filtering configuration, the Filter Cone opens and the arms retract towards the IVC wall.

AI/ML Overview

The provided text describes the Sentry Inferior Vena Cava Filter and summarizes the performance data used to support its 510(k) submission. Here's an analysis of the acceptance criteria and study information:

1. Table of Acceptance Criteria and Reported Device Performance

The document states that "All tests met the pre-determined acceptance criteria" but does not explicitly list the quantitative acceptance criteria for each test. It only lists the types of tests performed. Without specific numerical or qualitative targets, a detailed table cannot be created.

Acceptance Criteria (Implied / Not Quantified in Document)Reported Device Performance
• Delivery System visual and dimensionalMet pre-determined acceptance criteria
• Delivery System integrityMet pre-determined acceptance criteria
• Filter visual and dimensionMet pre-determined acceptance criteria
• Filter radio-opacityMet pre-determined acceptance criteria
• Fatigue resistanceMet pre-determined acceptance criteria
• Filter CorrosionMet pre-determined acceptance criteria
• Packaging Visual InspectionMet pre-determined acceptance criteria
• Labeling Visual InspectionMet pre-determined acceptance criteria
• Tyvek Pouch PeelMet pre-determined acceptance criteria
• Foil Pouch PeelMet pre-determined acceptance criteria
• Bubble Leak TestingMet pre-determined acceptance criteria
• Transportation testingMet pre-determined acceptance criteria
• Clot Trapping EfficiencyMet pre-determined acceptance criteria
• Migration resistanceMet pre-determined acceptance criteria
• Magnetic resonance testingMet pre-determined acceptance criteria
• Filter Migration PressureMet pre-determined acceptance criteria
• Radial ForceMet pre-determined acceptance criteria
• Ease of device useMet pre-determined acceptance criteria
• Filter Time-to-OpeningMet pre-determined acceptance criteria
• Simulated DeploymentMet pre-determined acceptance criteria
• Accuracy of DeploymentMet pre-determined acceptance criteria
• Finite Element AnalysisMet pre-determined acceptance criteria
• Computational Flow DynamicsMet pre-determined acceptance criteria
Clinical Success (Composite Endpoint): technical success; freedom from symptomatic PE; and freedom from IVC filter-related complications to 6 monthsAchieved in 97% of subjects, demonstrating the primary objective was met.
Specific Clinical Outcomes: Incidences of symptomatic PE, Filter tilting, Filter migration, Filter embolization, Filter fracture, Filter perforation or Filter-related death at 1, 2, and 6 months.No incidence reported for any of these specific adverse events.

2. Sample Size Used for the Test Set and Data Provenance

  • Clinical Study Test Set Sample Size: 129 subjects.
  • Data Provenance: Prospective, multi-center, single-arm clinical study. The document does not specify the country of origin, but generally, FDA submissions involve studies conducted in various countries, potentially including the US or Ireland (where the submitter is based).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

The document does not specify the number of experts used to establish the ground truth for the clinical study's endpoints or their specific qualifications. Clinical studies typically involve physicians and other healthcare professionals at investigational centers who assess patients and determine clinical outcomes. These would be considered the "experts" for clinical ground truth.

4. Adjudication Method for the Test Set

The document does not specify the adjudication method (e.g., 2+1, 3+1, none) for the clinical study's endpoints. In multi-center clinical trials, independent adjudication committees are common for critical endpoints, but this is not mentioned.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, an MRMC comparative effectiveness study involving human readers with and without AI assistance was not done. This device is an IVC filter, not an AI-powered diagnostic tool. The clinical study focused on the device's physical performance, safety, and effectiveness in patients.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

No, a standalone algorithm performance study was not conducted as this is a medical device (an IVC filter), not an algorithm or AI software.

7. Type of Ground Truth Used (Clinical Study)

For the clinical study, the ground truth was based on clinical outcomes data from patients, assessed by healthcare professionals (investigators) at 23 investigational centers. This included:

  • Technical success of placement.
  • Occurrence of symptomatic pulmonary embolism (PE).
  • Occurrence of IVC filter-related complications (e.g., tilting, migration, embolization, fracture, perforation, death).

8. Sample Size for the Training Set

The document does not specify a training set size. This is a medical device, and the evaluation involved pre-clinical (bench and animal) testing and a clinical trial. The term "training set" is typically used for machine learning models, which is not applicable here.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there was no "training set" in the context of machine learning. The clinical performance was assessed prospectively based on patient outcomes. For pre-clinical studies (bench and animal), "ground truth" would be established by direct observation, measurement, and histological analysis of the devices and biological responses according to established experimental protocols and standards.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle with three human profiles incorporated into its design. The eagle is positioned to the right of a circular text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 17, 2017

Novate Medical Ltd. Gordon Crowley Regulatory & Ouality Director Block 11, Galway Technology Park, Parkmore Galway, Ireland

Re: K162875

Trade/Device Name: Sentry Inferior Vena Cava Filter Regulation Number: 21 CFR 870.3375 Regulation Name: Cardiovascular Intravascular Filter Regulatory Class: Class II Product Code: DTK Dated: January 17, 2017 Received: January 19, 2017

Dear Gordon Crowley:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K162875

Device Name Sentry Inferior Vena Cava Filter

Indications for Use (Describe)

The Sentry IVC Filter is indicated for the prevention of recurrent pulmonary embolism via percutaneous placement in the inferior vena cava in patients with a transient high risk of PE, in the following situations:

· Pulmonary thromboembolism when anticoagulants are contraindicated;

· Failure of anticoagulant therapy in thromboembolic diseases;

· Emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)
-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K162875

510(k) Summary

Date Prepared:February 13, 2016
510(k) Submitter:Novate Medical Ltd., Block 11, Galway Technology Park,Parkmore, Galway, Ireland.
Contact:Gordon Crowley, Quality and Regulatory Director
Tel:+353 91 787716
Trade Name:Sentry Inferior Vena Cava Filter
Common Name:Vena Cava Filter
ClassificationInformation:Cardiovascular Intravascular Filter (21 CFR 870.3375; Class II)
Product Code:DTK
Panel:Cardiovascular
Predicate Devices:Cordis Optease® IVC Filter (K034050)
B Braun VenaTech Convertible IVC Filter (K152765)

The Sentry IVC Filter is a bioconvertible IVC filter intended for percutaneous implantation in the IVC and designed to provide protection against PE in patients at transient risk of PE.

The Sentry IVC Filter is designed for use in inferior vena cavae with diameters between 16mm and 28mm and has a maximum deployed length of 57.7mm. It is preloaded in a Loading Tool that can be orientated for either left/right femoral vein or a right jugular vein approach and is delivered through a 7 French ID Introducer Sheath (max OD 9.75Fr).

{4}------------------------------------------------

The Sentry IVC Filter consists of a cylindrical Nitinol frame and a Filter Cone formed by 6 Filter Arms held together in the center of the IVC by means of a bioabsorbable filament. The Filter Cone that is designed to trap emboli and thereby reduce the risk of PE while maintaining caval patency after it has converted. The Sentry IVC Filter converts into a non-filtering configuration, the Filter Cone opens and the arms retract towards the IVC wall.

Intended Use / Indications

The Sentry IVC Filter is indicated for the prevention of recurrent pulmonary embolism via percutaneous placement in the inferior vena cava in patients with a transient high risk of PE, in the following situations:

  • Pulmonary thromboembolism when anticoagulants are contraindicated; .
  • Failure of anticoagulant therapy in thromboembolic diseases; .
  • Emergency treatment following massive pulmonary embolism where anticipated . benefits of conventional therapy are reduced.

Substantial Equivalence Comparison

The Sentry IVC Filter is substantially equivalent in its intended use, principles of operation, design, materials, and sterile package configuration to the Cordis OptEase Vena Cava Filter (K034050) and the B.Braun VenaTech Convertible IVC Filter (K152765). The Sentry IVC Filter is designed to bioconvert from a Filtering Configuration to a Non-Filtering Configuration. The design differences, which allow the Filter to convert to the non-filtering configuration. were subjected to non-clinical and clinical performance evaluations.

Performance Data

Novate developed a design verification and validation program for the Sentry IVC Filter with reference to the FDA Guidance for Cardiovascular Intravascular Filter 510(k) Submissions and the international standard ISO 25539-3 Cardiovascular implants-Endovascular devices -Part 3: Vena cava filters. Bench and animal studies were

{5}------------------------------------------------

K162875

undertaken to demonstrate the performance and safety of the Sentry IVC Filter when used according to the Instructions for Use.

• Delivery System visual and dimensional• Clot Trapping Efficiency
• Delivery System integrity• Migration resistance
• Filter visual and dimension• Magnetic resonance testing
• Filter radio-opacity• Filter Migration Pressure
• Fatigue resistance• Radial Force
• Filter Corrosion• Ease of device use
• Packaging Visual Inspection• Filter Time-to-Opening
• Labeling Visual Inspection• Simulated Deployment
• Tyvek Pouch Peel• Accuracy of Deployment
• Foil Pouch Peel• Finite Element Analysis
• Bubble Leak Testing• Computational Flow Dynamics
• Transportation testing

The design verification and validation program included the evaluations listed below:

All tests met the pre-determined acceptance criteria. Results from the design verification and validation program demonstrate the Sentry IVC Filter achieves its intended use as a vena cava filter and demonstrates the mechanical integrity and performance of the device.

Biocompatibility Evaluations

A biological safety assessment was performed to evaluate the biological risks associated with the Sentry IVC Filter per FDA's Guidance Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". This risk assessment involved biocompatibility testing on the implant and delivery system/accessories. The Sentry IVC Filter was also subjected to chemical characterization evaluations performed in accordance with ISO 10993 Biological Evaluation of Medical Devices - Part 18: Chemical characterization of materials. The biocompatibility studies, required by ISO 10993 Biological Evaluation of Medical Devices - Part 1: Evaluation and testing were performed in accordance with Good Laboratory Practices (GLP; 21 CFR Part 58).

{6}------------------------------------------------

K162875

Animal Studies

GLP animal studies were performed in an ovine model during the development of the Sentry IVC Filter. A predicate device (Cordis OptEase IVC Filter [K034050]) was also evaluated as a control and comparison. Results of this animal testing demonstrated that the Sentry IVC Filter performed according to its intended use, and thus support the safety of the Sentry IVC Filter.

Clinical Evaluation

A prospective, multi-center, single-arm clinical study with a performance goal was conducted to assess the safety and effectiveness of the Sentry IVC Filter. A total of 129 subjects were enrolled at 23 investigational centers. Subjects were evaluated at 1 month, 2 months and 6 months. The primary composite endpoint of Clinical Success at 6 months comprising: technical success; freedom from symptomatic PE; and freedom from IVC filter-related complications to 6 months was achieved in 97% of subjects; therefore the primary objective, Clinical Success, was met. There was no incidence of symptomatic PE, Filter tilting, Filter migration, Filter embolization, Filter fracture, Filter perforation or Filter-related death at the 1 month, 2 months and 6 month follow-ups. Refer to the IFU for clinical study information.

Conclusions

The Sentry IVC Filter is similar in indications, principle of operation, and design to the currently marketed Cordis OptEase IVC Filter (K034050) and B.Braun VenaTech Convertible IVC Filter (K152765). The design differences, which allow the Sentry IVC Filter to convert to the non-filtering configuration, were evaluated by non-clinical and clinical performance studies. Based on the supportive data provided in this 510(k), it can be concluded that the Sentry IVC Filter raises no new questions of safety or effectiveness compared to the predicate devices and is, therefore, substantially equivalent.

§ 870.3375 Cardiovascular intravascular filter.

(a)
Identification. A cardiovascular intravascular filter is an implant that is placed in the inferior vena cava for the purpose of preventing pulmonary thromboemboli (blood clots generated in the lower limbs and broken loose into the blood stream) from flowing into the right side of the heart and the pulmonary circulation.(b)
Classification. Class II. The special controls for this device are:(1) “Use of International Standards Organization's ISO 10993 ‘Biological Evaluation of Medical Devices Part I: Evaluation and Testing,’ ” and
(2) FDA's:
(i) “510(k) Sterility Review Guidance and Revision of 2/12/90 (K90-1)” and
(ii) “Guidance for Cardiovascular Intravascular Filter 510(k) Submissions.”