K Number
K162758

Validate with FDA (Live)

Manufacturer
Date Cleared
2017-03-21

(172 days)

Product Code
Regulation Number
880.2800
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Verify® Steam Indicators are emulating indicators intended for use in steam sterilization. The Verify® Steam Indicators change color from yellow to blue/purple when exposed to the appropriate cycle temperature, type, and duration. The indicator models and their cycle temperatures, types, and times are:

MODELTEMPERATURESTERILIZATION TYPETIME
Verify 250F 30250°F (121°C)gravity steam30 minutes
Verify 270F 15270°F (132°C)gravity steam15 minutes
Verify 270F 3-10270°F (132°C)gravity flash steam3 or 10 minutes
Verify 270F 4270°F (132°C)Steam Flush Pressure Pulse (SFPP), pre-vacuum and Express steam4 minutes

The Verify® 275F 3 Indicator is a chemical indicator intended for use by health care providers to accompany products being sterilized through a sterilization procedure. The Verify® 275F 3 Indicator is an emulating indicator intended for use in steam sterilization. The indicator changes color from yellow to blue/purple when exposed to 275°F (135°C) pre-vacuum sterilization cycles for 3 minutes as indicated in the following table:

MODELTEMPERATURESTERILIZATION TYPETIME
Verify 275F 3275°F (135°C)Pre-vacuum steam3 minutes

The Verify® 275F Gravity Indicators are chemical integrators which meet ANSVAAMI 1140-1:2005 for emulating indicators intended for use in steam sterilization. The Verify® 275F Gravity Indicators change color from yellow to blue/purple when exposed to 275°F (135°C) and to the appropriate cycle type and duration. The Verify® 275F Indicators models and their cycle types and times are:

MODELTEMPERATURESTERILIZATION TYPETIME
Verify 275F 3-10275°F (135°C)Gravity flash steam3 and 10 minutes
Verify 275F 10275°F (135°C)Gravity steam10 minutes

The Verify® Challenge Packs -Version 2 - are test packs consisting of an emulating indicator surrounded by a steam penetration barrier, intended for use in steam sterilization. The Verify® Challenge Packs indicators change color from yellow to blue/purple when exposed to the appropriate cycle temperature, type, and duration. The challenge pack models and their cycle temperatures, types, and times are:

MODELTEMPERATURESTERILIZATION TYPETIME
Verify® 270F 4270°F (132°C)Pre-vacuum, Steam Flush Pressure Pulse (SFPP)4 minutes
Verify® 275F 3275°F (135°C)Pre-vacuum3 minutes

There is a process indicator outside of the packs which undergoes a visual color change when exposed to steam in a temperature range of 250°F (121°C) to 275°F (135°C).

The Verify® 270FP Challenge Pack is a test pack consisting of three emulating indicator inks situated on a test sheet surrounded by a steam penetration barrier, intended for use in SFPP (steam-flush pressure-pulse) and pre-vacuum steam sterilization for load release. The Verify® 270FP Challenge Pack indicators change color from yellow to blue/purple when exposed to the following conditions at 270°F (132°C):

  • · 4 minute indicator SFPP and pre-vacuum steam sterilization for 4 minutes.
  • · 10 minute indicator Pre-vacuum steam sterilization for 10 minutes.
  • · 20 minute indicator Pre-vacuum steam sterilization for 20 minutes.

The Browne TST Dual Use Test Pack is a chemical sterilization process monitor that can be used as either:

  • Bowie Dick Test in an empty chamber to monitor air removal and steam penetration during the vacuum stage of 132°C (270°F), 134°C (273°F), and 135 °C (275 °F) test cycles.
  • Challenge Pack in a loaded chamber to monitor steam penetration during the hold/plateau stage of 132°C (270°F), 134 °C (273 °F), and 135 °C (275 °F) steam sterilization cycles.
Device Description

The VERIFY® SixCess Chemical Indicators consist of a 22mm x 143 mm strip (long) (7/8" x 5.6") or a 22mm X 70 mm (short) polypropylene strip with a 12 mm circular chemical indicator ink spot (or two spots in the case of the 270°F 3-10 Indicator) located on one end, adjacent to a reference circle exhibiting the endpoint color. The indicator ink on the proposed VERIFY® Steam Indicators change from yellow to blue/purple color when the steam sterilization cycle is complete.

The Verify® 275F Indicator consist of: A 22 mm x 143 mm (long) or a 22mm X 70 mm (short) polypropylene strip with one 12 mm chemical indicator ink spot. The indicator ink spots are located on each end of the strip (or one end if it is just one ink spot), adjacent to a reference block exhibiting the endpoint color. The indicator ink on the proposed Verify® 275F Indicator changes from yellow to blue/purple color when the steam sterilization cycle is complete.

The Verify® 275F Gravity Indicators consist of: A 22 mm x 143 mm polypropylene strip with two 12 mm chemical indicator ink spots (for Verify® 275F 3-10 Indicator). A 22 mm x 143 mm (long) or a or a 22mm X 70 mm (short) polypropylene strip with one 12 mm chemical indicator ink spots (for Verify® 275F 10 Indicator). The indicator ink spots are located on each end of the strip (or one end if it is just one ink spot), adjacent to a reference block exhibiting the endpoint color. The indicator ink on the proposed Verify® 275F Gravity Indicators changes from yellow to blue/purple color when the steam sterilization cycle is complete.

The VERIFY® Challenge Packs Version 2 consists of an emulating indicator surrounded by a steam penetration barrier. The indicator ink inside the proposed Verify® Challenge Packs changes from yellow to blue/purple color when the steam sterilization cycle is complete. The process indicator outside of the packs undergoes a visual color change from pink to dark purple when exposed to steam in a temperature range of 250°F (121°C) to 275°F (135°C).

The VERIFY® 270FP Challenge Pack, Model LCC019, consists of three emulating indicator inks situated on a test sheet surrounded by a steam penetration barrier. The indicator inks on the proposed Verify® 270FP Challenge Pack test sheet change from yellow to blue/purple color when exposed to saturated steam at 270°F (132°C) for the following times. The process indicator outside of the packs undergoes a visual color change from pink to dark purple when exposed to steam in a temperature range of 250°F (121°C) to 275°F (135°C).

The TST Test Pack (models EQC009 and EQC010, VERIFY Bowie Dick Test Pack) consists of a chemical integrator surrounded by a steam penetration barrier. The indicator ink used for the Integrator Test Pack integrator is identical to the ink used for the TST Control Integrator for Steam Autoclave (K965154).

AI/ML Overview

This document describes the premarket notification (510(k)) for VERIFY® Steam Indicators and related products. It focuses on demonstrating substantial equivalence to previously cleared devices rather than providing a standalone study with acceptance criteria for a novel device.

The study presented here is an effectiveness comparison between the proposed devices with new REACh compliant inert components in the ink formulation and their predicate devices. The goal is to show that changing these inert components does not negatively impact the performance of the sterilization indicators.

Here's a breakdown of the requested information based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

The document doesn't explicitly list "acceptance criteria" in a typical quantitative clinical trial sense (e.g., sensitivity/specificity thresholds). Instead, the acceptance criteria are implicitly that the proposed device must perform identically to the predicate device for all relevant performance tests. The "reported device performance" refers to the results of these comparative tests.

TestAcceptance Criteria (Implicit)Reported Device Performance (Proposed Device)Comparison
Performance in a sterilizerMust "Pass" (perform equivalently to the predicate device) in a sterilizer, validating the indicator's color change accurately reflects sterilization conditions for its specified cycle temperature, type, and duration.PassSame
Performance in a resistometerMust "Pass" (perform equivalently to the predicate device) in a resistometer, validating the indicator's color change accurately reflects sterilization conditions under controlled laboratory conditions.PassSame
No delamination of indicators following exposure to steamMust "Pass" (perform equivalently to the predicate device), ensuring the integrity of the indicator material and ink after steam exposure.PassSame
Meets ISO 11140 requirements for Type 6 indicatorMust meet ISO 11140 Type 6 requirements (emulating indicator, responding to all critical variables of the sterilization process at stated values).Meets ISO 11140 requirements for Type 6 indicatorSame
Meets ISO 11140 requirements for Bowie Dick indicatorMust meet ISO 11140 requirements for Bowie Dick indicator (specifically for the TST Dual Use Test Pack, to monitor air removal and steam penetration).Meets ISO 11140 requirements for Bowie Dick indicatorSame
FeaturesMust be identical to the predicate device.IdenticalSame
Principle of operationMust be identical to the predicate device.IdenticalSame
Types/modelsMust be identical to the predicate device.IdenticalSame
Intended UseNo change to Intended Use or Indications for Use compared to the predicate device.No changeSame
Mechanism of ActionNo change to Mechanism of Action compared to the predicate device.No changeSame

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify the sample size used for the tests (e.g., number of indicators tested in sterilizer runs).

The data provenance is retrospective in comparison to predicate devices. The testing was conducted by STERIS Corporation (Mentor, OH, USA), implying the data is from the United States. The study focuses on comparing the new formulation with existing, cleared formulations.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

This type of device (chemical sterilization indicator) does not typically involve human expert evaluation for "ground truth" in the same way clinical imaging or diagnostics might. The "ground truth" is based on the physical and chemical response of the indicator to controlled sterilization conditions. Therefore, there were no "experts" in the sense of clinicians establishing ground truth by interpreting visual data. The performance is assessed against established ISO standards (ISO 11140) and comparison to predicate devices, which are objective measures.

4. Adjudication Method for the Test Set

Not applicable. As described above, the "ground truth" is determined by objective physical and chemical properties and adherence to ISO standards, not by human interpretation requiring adjudication.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

No, a multi-reader, multi-case comparative effectiveness study was not done. This type of study is relevant for diagnostic devices where human readers interpret medical images or data. The VERIFY® Steam Indicators are chemical indicators for sterilization processes, where the "reading" (color change) is designed to be unambiguous and objective.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

This question is not directly applicable to a chemical indicator. Chemical indicators are inherently "standalone" in their function; they change color based on exposure to sterilization parameters. There is no "algorithm" or software component. Their performance is their inherent physical-chemical reaction, without human intervention in the reaction itself. The "human-in-the-loop" is the interpretation of the color change, which is designed to be a clear qualitative assessment (yellow to blue/purple).

7. The Type of Ground Truth Used

The ground truth is established by:

  • Controlled sterilization parameters: Exposure to specific temperature, time, and steam conditions in sterilizers and resistometers, against which the indicator's color change is measured.
  • Objective performance criteria: Adherence to ISO 11140 requirements for Type 6 and Bowie Dick indicators, and demonstrating equivalent performance to previously cleared predicate devices under these controlled conditions. The expected color change (yellow to blue/purple) for successful sterilization is the ground truth.

8. The Sample Size for the Training Set

Not applicable. This device is a chemical indicator, not an AI or machine learning algorithm. Therefore, there is no "training set" in the context of data-driven model development. The development of the indicator chemistry itself would involve experimentation and formulation, but not a "training set" like in AI/ML.

9. How the Ground Truth for the Training Set was Established

Not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, overlaid on three human profiles. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 21, 2017

STERIS Corporation Mr. Anthony Piotrkowski Senior Manager, Regulatory Affairs 5960 Heisley Rd. Mentor, Ohio 44060

Re: K162758

Trade/Device Name: VERIFY® Steam Indicators (VERIFY® SixCess® Chemical Indicators) VERIFY 275F 3 Indicators (VERIFY® SixCess Indicators) VERIFY 275F Gravity Indicators (VERIFY® SixCess Indicators) VERIFY Challenge Packs - Version 2 (VERIFY® SixCess Challenge Pack) VERIFY 270FP Challenge Pack Browne TST Dual Use Test Pack (VERIFY® Bowie Dick Test Pack) Regulation Number: 21 CFR 880.2800(b) Regulation Name: Sterilization Process Indicator Regulatory Class: Class II Product Code: JOJ Dated: February 15, 2017 Received: February 16, 2017

Dear Mr. Anthony Piotrkowski:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Michael J. Ryan -S

for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K162758

Device Name

VERIFY® Steam Indicators (VERIFY® SixCess® Chemical Indicators)

Indications for Use (Describe)

The Verify® Steam Indicators are emulating indicators intended for use in steam sterilization. The Verify® Steam Indicators change color from yellow to blue/purple when exposed to the appropriate cycle temperature, type, and duration. The indicator models and their cycle temperatures, types, and times are:

MODELTEMPERATURESTERILIZATION TYPETIME
Verify 250F 30250°F (121°C)gravity steam30 minutes
Verify 270F 15270°F (132°C)gravity steam15 minutes
Verify 270F 3-10270°F (132°C)gravity flash steam3 or 10 minutes
Verify 270F 4270°F (132°C)SFPP, pre-vacuum and Express steam4 minutes

Type of Use (Select one or both, as applicable)

ا Prescription Use (Part 21 CFR 801 Subpart D)

മ Over-The-Counter Use (21 CFR 801 Subpart C)

Form Approved: OMB No. 0910-0120

Expiration Date: January 31, 2017

See PRA Statement below.

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Number (if known) K162758

Device Name

VERIFY 275F 3 Indicator (VERIFY® SixCess Indicators)

Indications for Use (Describe)

The Verify® 275F 3 Indicator is a chemical indicator intended for use by health care providers to accompany products being sterilized through a sterilization procedure. The Verify® 275F 3 Indicator is an emulating indicator intended for use in steam sterilization. The indicator changes color from yellow to blue/purple when exposed to 275ºF (135ºC) prevacuum sterilization cycles for 3 minutes as indicated in the following table:

MODELTEMPERATURESTERILIZATION TYPETIME
Verify 275F 3275°F (135°C)Pre-vacuum steam3 minutes

Type of Use (Select one or both, as applicable)

ا Prescription Use (Part 21 CFR 801 Subpart D)

മ Over-The-Counter Use (21 CFR 801 Subpart C)

Form Approved: OMB No. 0910-0120

Expiration Date: January 31, 2017

See PRA Statement below.

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Number (if known) K162758

Device Name

VERIFY 275F Gravity Indicators (VERIFY® SixCess Indicators)

Indications for Use (Describe)

The Verify® 275F Gravity Indicators are chemical integrators which meet ANSI/AAMI 1140-1:2005 for emulating indicators intended for use in steam sterilization. The Verify® 275F Gravity Indicators change color from yellow to blue/ purple when exposed to 275°F (135°C) and to the appropriate cycle type and duration. The Verify® 275F Indicators models and their cycle types and times are:

MODELTEMPERATURESTERILIZATION TYPETIME
Verify 275F 3-10275°F (135°C)Gravity flash steam3 and 10 minutes
Verify 275F 10275°F (135°C)Gravity steam10 minutes

Type of Use (Select one or both, as applicable)

ا Prescription Use (Part 21 CFR 801 Subpart D)

മ Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{5}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K162758

Device Name

VERIFY Challenge Packs - Version 2 (VERIFY® SixCess Challenge Pack)

Indications for Use (Describe)

The Verify® Challenge Packs -Version 2 - are test packs consisting of an emulating indicator surrounded by a steam penetration barrier, intended for use in steam sterilization. The Verify® Challenge Packs indicators change color from yellow to blue/purple when exposed to the appropriate cycle temperature, type, and duration. The challenge pack models and their cycle temperatures, types, and times are:

  • Verify® 270F 4 270ºF (132ºC) Pre-vacuum, Steam Flush Pressure Pulse (SFPP), 4 minutes
  • Verify® 275F 3 275°F (135°C) Pre-vacuum 3 minutes

There is a process indicator outside of the packs which undergoes a visual color change when exposed to steam in a temperature range of 250°F (121°C) to 275°F (135°C).

Type of Use (Select one or both, as applicable)

ا Prescription Use (Part 21 CFR 801 Subpart D)

മ Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{6}------------------------------------------------

510(k) Number (if known) K162758

Device Name VERIFY 270FP Challenge Pack

Indications for Use (Describe)

The Verify® 270FP Challenge Pack is a test pack consisting of three emulating indicator inks situated on a test sheet surrounded by a steam penetration barrier, intended for use in SFPP (steam-flush pressure-pulse) and pre-vacuum steam sterilization for load release. The Verify® 270FP Challenge Pack indicators change color from vellow to blue/purple when exposed to the following conditions at 270 F (132C):

  • 4 minute indicator SFPP and pre-vacuum steam sterilization for 4 minutes.
  • 10 minute indicator Pre-vacuum steam sterilization for 10 minutes.
  • 20 minute indicator Pre-vacuum steam sterilization for 20 minutes.

Type of Use (Select one or both, as applicable)

  • Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{7}------------------------------------------------

510(k) Number (if known) K162758

Device Name

Browne TST Dual Use Test Pack (VERIFY® Bowie Dick Test Pack)

Indications for Use (Describe)

The Browne TST Dual Use Test Pack is a chemical sterilization process monitor that can be used as either:

  • Bowie Dick Test - in an empty chamber to monitor air removal and steam penetration during the vacuum stage of 132°C (270°F) , 134°C (273°F) , and 135 °C (275 °F) test cycles.

  • Challenge Pack - in a loaded chamber to monitor steam penetration during the hold/plateau stage of 132°C (270°F), 134 °C (273 °F), and 135 °C (275 °F) steam sterilization cycles.

Type of Use (Select one or both, as applicable)

ا Prescription Use (Part 21 CFR 801 Subpart D)

മ Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for STERIS. The word STERIS is in all caps and in a serif font. Below the word STERIS is a graphic of several horizontal wavy lines that are blue.

510(k) Summary For K162758 VERIFY® SixCess® Chemical Indicators and Chemical Indicator Challenge Packs

STERIS Corporation 5960 Heisley Road Mentor, OH 44060 Phone: (440) 354-2600 Fax No: (440) 357-9198

Anthony Piotrkowski Contact: Senior Manager, Regulatory Affairs Tel: 440-392-7437 Fax: 440-357-9198 e-mail: tpiotrko@steris.com

Submission Date: March 20, 2017

Premarket Notification Number: K162758

STERIS Corporation ■ 5960 Heisley Road ■ Mentor, OH 44060-1834 USA ■ 440-354-2600

{9}------------------------------------------------

Device Name 1.

VERIFY® Steam Indicators (VERIFY® SixCess® Chemical Trade Name: Indicators) VERIFY 275F 3 Indicators (VERIFY® SixCess Indicators) VERIFY 275F Gravity Indicators (VERIFY® SixCess Indicators) VERIFY Challenge Packs - Version 2 (VERIFY® SixCess Challenge Pack) VERIFY 270FP Challenge Pack Browne TST Dual Use Test Pack (VERIFY® Bowie Dick Test Pack)

Device Class:Class II
Common/usual Name:Chemical Indicator
Classification Name:Physical/chemical sterilization process indicators
Classification Number:21 CFR 880.2800 (b)
Product Code:JOJ

Predicate Devices 2.

510(k) #Device Name
K002741Browne TST Dual Use Test Pack
K070461VERIFY Steam Indicators
K071895VERIFY 275F 3 Indicator
K073683VERIFY Challenge Packs - Versions 2
K083643VERIFY 275F Gravity Indicator
K103053VERIFY 270FP Challenge Pack

The proposed devices are identical to the predicates in performance and composition with the following exceptions:

  • . Solvent is changing to a REACh compliant base - the solvent evaporates during the printing process.
  • . Plasticizer is changing to a REACh compliant chemical - the plasticizer is an inert component and makes the ink easier to work with during the printing process.
  • . Adhesive is changing to a REACh compliant chemical - the adhesive is inert and used to adhere laminate to substrate.
  • . Labeling references the (most recent) 2014 revision of ISO 11140-1 and uses the newly introduced "Type 6" in place of "Class 6".

{10}------------------------------------------------

3. Description of Device

  • The VERIFY® SixCess Chemical Indicators consist of a 22mm x 143 mm 3.1 strip (long) (7/8" x 5.6") or a 22mm X 70 mm (short) polypropylene strip with a 12 mm circular chemical indicator ink spot (or two spots in the case of the 270°F 3-10 Indicator) located on one end, adjacent to a reference circle exhibiting the endpoint color. The indicator ink on the proposed VERIFY® Steam Indicators change from yellow to blue/purple color when the steam sterilization cycle is complete.
    • . The Verify® 250ºF 30 Indicator (models: PCC005, VERIFY SixCess 250F 30 Minute Indicator - Short, and PCC007, VERIFY SixCess 250F 30 Minute Indicator - Long) can be used to monitor a 30 minute 250°F/121°C gravity steam sterilization cycle.
    • . The Verify® 270ºF 15 Indicator (model PCC031, VERIFY SixCess 270F 15 Minute Indicator) can be used to monitor a 15 minute 270°F/132°C gravity steam sterilization cycle.
    • . The Verify® 270°F 3-10 Indicator (model LCC310, VERIFY SixCess 270F Flash Indicator) can be used to monitor a 3 minute or10 minute 270°F/132°C gravity flash steam sterilization cycle.
    • . The Verify® 270ºF 4 Indicator (models PCC006, VERIFY SixCess 270F 4 Minute Indicator - Short and PCC008, VERIFY SixCess 270F 4 Minute Indicator - Long) can be used to monitor a 4 minute 270°F/132°C SFPP. pre-vacuum and Express steam sterilization cycle.
  • 3.2 The Verify® 275F Indicator consist of:
    • . A 22 mm x 143 mm (long) or a 22mm X 70 mm (short) polypropylene strip with one 12 mm chemical indicator ink spot.

The indicator ink spots are located on each end of the strip (or one end if it is just one ink spot), adjacent to a reference block exhibiting the endpoint color. The indicator ink on the proposed Verify® 275F Indicator changes from yellow to blue/purple color when the steam sterilization cycle is complete.

The Verify® 275F Indicator can be used to monitor 275°F (135°C) sterilization cycles as follows:

  • . The Verify® 275F 3 Indicator (models PCC034, VERIFY SixCess 275F 3 Minute Indicator - Long and PCC038, VERIFY SixCess 275F 3 Minute Indicator - Short) can be used to monitor a 3 minute 275°F (135°C) prevacuum steam sterilization cycle.
  • The Verify® 275F Gravity Indicators consist of: 3.3

{11}------------------------------------------------

  • A 22 mm x 143 mm polypropylene strip with two 12 mm chemical indicator . ink spots (for Verify® 275F 3-10 Indicator).
  • . A 22 mm x 143 mm (long) or a or a 22mm X 70 mm (short) polypropylene strip with one 12 mm chemical indicator ink spots (for Verify® 275F 10 Indicator).

The indicator ink spots are located on each end of the strip (or one end if it is just one ink spot), adjacent to a reference block exhibiting the endpoint color. The indicator ink on the proposed Verify® 275F Gravity Indicators changes from yellow to blue/purple color when the steam sterilization cycle is complete.

The Verify® 275F Gravity Indicators can be used to monitor 275°F (135°C) sterilization cycles as follows:

  • . The Verify® 275F 3-10 Indicator (model LCC005, VERIFY SixCess 275F Flash Indicator) can be used to monitor a 3 and 10 minute 275°F (135°C) gravity flash steam sterilization cycle (including prevacuum, single wrapped trays).
  • . The Verify® 275F 10 Indicator can be used to monitor a 10 minute 275°F (135℃) gravity steam sterilization cycle.
  • 3.4 The VERIFY® Challenge Packs Version 2 consists of an emulating indicator surrounded by a steam penetration barrier. The indicator ink inside the proposed Verify® Challenge Packs changes from yellow to blue/purple color when the steam sterilization cycle is complete:
    • . The Verify® 270F 4 Challenge Pack -Version 2 (model LCCC003, VERIFY SixCess 270F 4 Challenge Pack) - can be used to monitor a 4 minute 270°F/132°C Steam-Flush Pressure-Pulse (SFPP) and pre-vacuum steam sterilization cycle.
    • The Verify® 275F 3 Challenge Pack - Version 2 - (model LCC014, VERIFY SixCess 275F 3 Challenge Pack) can be used to monitor a 3 minute 275°F/135°C pre-vacuum steam sterilization cycle.

The process indicator outside of the packs undergoes a visual color change from pink to dark purple when exposed to steam in a temperature range of 250°F (121°C) to 275°F (135°C).

  • 3.5 The VERIFY® 270FP Challenge Pack, Model LCC019, consists of three emulating indicator inks situated on a test sheet surrounded by a steam penetration barrier. The indicator inks on the proposed Verify® 270FP Challenge Pack test sheet change from yellow to blue/purple color when exposed to saturated steam at 270°F (132°C) for the following times.

{12}------------------------------------------------

  • . 4 minute indicator - The 4 minute indicator ink on the Verify® 270FP Challenge Pack can be used to monitor a 4 minute SFPP (steam-flush pressure-pulse) and pre-vacuum steam sterilization cycle.
  • . 10 minute indicator – The 10 minute indicator ink on the Verify® 270FP Challenge Pack can be used to monitor a 10 minute pre-vacuum steam sterilization cycle.
  • . 20 minute indicator - The 20 minute indicator ink on the Verify® 270FP Challenge Pack can be used to monitor a 20 minute pre-vacuum steam sterilization cycle.

The process indicator outside of the packs undergoes a visual color change from pink to dark purple when exposed to steam in a temperature range of 250°F (121°C) to 275°F (135°C).

  • 3.6 The TST Test Pack (models EQC009 and EQC010, VERIFY Bowie Dick Test Pack) consists of a chemical integrator surrounded by a steam penetration barrier. The indicator ink used for the Integrator Test Pack integrator is identical to the ink used for the TST Control Integrator for Steam Autoclave (K965154).

Intended Use 4.

4.1 Verify® Steam Indicators

The Verify® Steam Indicators are emulating indicators intended for use in steam sterilization. The Verify® Steam Indicators change color from yellow to blue/purple when exposed to the appropriate cycle temperature, type, and duration. The indicator models and their cycle temperatures, types, and times are:

MODELTEMPERATURESTERILIZATION TYPETIME
Verify 250F 30250°F (121°C)gravity steam30 minutes
Verify 270F 15270°F (132°C)gravity steam15 minutes
Verify 270F 3-10270°F (132°C)gravity flash steam3 or 10 minutes
Verify 270F 4270°F (132°C)Steam Flush Pressure Pulse (SFPP), pre-vacuum and Express steam4 minutes

4.2 Verify® 275F 3 Indicator

The Verify® 275F 3 Indicator is a chemical indicator intended for use by health care providers to accompany products being sterilized through a sterilization procedure. The Verify® 275F 3 Indicator is an emulating indicator intended for use in steam sterilization. The indicator changes color from yellow to blue/purple when exposed to 275°F (135°C) pre-vacuum sterilization cycles for 3 minutes as indicated in the following table:

{13}------------------------------------------------

MODELTEMPERATURESTERILIZATION TYPETIME
Verify 275F 3275°F (135°C)Pre-vacuum steam3 minutes

4.3 Verify® 275F Gravity Indicators

The Verify® 275F Gravity Indicators are chemical integrators which meet ANSVAAMI 1140-1:2005 for emulating indicators intended for use in steam sterilization. The Verify® 275F Gravity Indicators change color from yellow to blue/purple when exposed to 275°F (135°C) and to the appropriate cycle type and duration. The Verify® 275F Indicators models and their cycle types and times are:

MODELTEMPERATURESTERILIZATION TYPETIME
Verify 275F 3-10275°F (135°C)Gravity flash steam3 and 10 minutes
Verify 275F 10275°F (135°C)Gravity steam10 minutes

4.4 Verify® Challenge Packs

The Verify® Challenge Packs -Version 2 - are test packs consisting of an emulating indicator surrounded by a steam penetration barrier, intended for use in steam sterilization. The Verify® Challenge Packs indicators change color from yellow to blue/purple when exposed to the appropriate cycle temperature, type, and duration. The challenge pack models and their cycle temperatures, types, and times are:

MODELTEMPERATURESTERILIZATION TYPETIME
Verify® 270F 4270°F (132°C)Pre-vacuum, Steam Flush Pressure Pulse (SFPP)4 minutes
Verify® 275F 3275°F (135°C)Pre-vacuum3 minutes

There is a process indicator outside of the packs which undergoes a visual color change when exposed to steam in a temperature range of 250°F (121°C) to 275°F (135°C).

4.5 Verify® 270FP Challenge Pack

The Verify® 270FP Challenge Pack is a test pack consisting of three emulating indicator inks situated on a test sheet surrounded by a steam penetration barrier, intended for use in SFPP (steam-flush pressure-pulse) and pre-vacuum steam sterilization for load release. The Verify® 270FP Challenge Pack indicators change color from yellow to blue/purple when exposed to the following conditions at 270°F (132°C):

  • · 4 minute indicator SFPP and pre-vacuum steam sterilization for 4 minutes.
  • · 10 minute indicator Pre-vacuum steam sterilization for 10 minutes.
  • · 20 minute indicator Pre-vacuum steam sterilization for 20 minutes.

{14}------------------------------------------------

4.6 TST Dual Use Test Pack

The Browne TST Dual Use Test Pack is a chemical sterilization process monitor that can be used as either:

  • Bowie Dick Test in an empty chamber to monitor air removal and steam penetration during the vacuum stage of 132°C (270°F), 134°C (273°F), and 135 °C (275 °F) test cycles.
  • Challenge Pack in a loaded chamber to monitor steam penetration during the hold/plateau stage of 132°C (270°F), 134 °C (273 °F), and 135 °C (275 °F) steam sterilization cycles.

5. Description of Safety and Substantial Equivalence

5.1 Technology Comparison

  • Features - identical between predicate and proposed device
  • . Principle of operation - identical between predicate and proposed devices
  • . Types/models - identical between predicate and proposed devices
  • . Materials - identical between predicate and proposed devices with the exception of the three ink formulation changes described in section 2 above.
FeaturePredicateProposedComparison
SubstratePolypropylenePolypropyleneSame
InkProprietaryIdentical formulationas predicate withexception of inertcomponents - solvent,plasticizer andadhesiveData presented inthe submission todemonstratesimilarities
PerformanceSpecification(Type 6)Meets ISO 11140requirements forType 6 indicatorMeets ISO 11140requirements for Type6 indicatorSame
PerformanceSpecification(BD)Meets ISO 11140requirements forBowie Dick indicatorMeets ISO 11140requirements forBowie Dick indicatorSame
Intended UseNo change to intended use or indications foruse between proposed and predicateSame
Mechanism ofActionNo change to mechanism of action betweenproposed and predicateSame

{15}------------------------------------------------

5.2 Effectiveness Comparison

The ink formulation changes were to inert component of the chemical indicator ink. The testing described in the table below, over the 3-year shelflife, demonstrated equivalent performance of the proposed indicators to the predicates.

TestPredicateProposedComparison
Performance in a sterilizerPassPassSame
Performance in a resistometerPassPassSame
No delamination of indicatorsfollowing exposure to steamPassPassSame

Safety Comparison 5.3

The ink formulation changes were to inert component of the chemical indicator ink, and the further the solvent is not present in the final product. The safety of the indicator products is inherent to their design. Indicators are not intended to contact patients. All indicators are laminated to prevent contact of the indicator ink with items within a sterilization load. Challenge pack indicators are further contained within steam barrier materials.

6. Conclusion

Based on the intended uses, technological characteristics and non-clinical performance data, the subject devices are substantially equivalent and is as safe and as effective as the legally marketed predicate device Class II (21 CFR 880.2800, Product code JOJ).

§ 880.2800 Sterilization process indicator.

(a)
Biological sterilization process indicator —(1)Identification. A biological sterilization process indicator is a device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization.(2)
Classification. Class II (performance standards).(b)
Physical/chemical sterilization process indicator —(1)Identification. A physical/chemical sterilization process indicator is a device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.(2)
Classification. Class II (performance standards).