K Number
K162681

Validate with FDA (Live)

Date Cleared
2017-06-02

(249 days)

Product Code
Regulation Number
882.1400
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

ENOBIO is an EEG portable monitoring device of 8, 20, 32 channels intended for the use in clinical patient monitoring for use in hospitals and other medical environments.

The Enobio is intended to acquire, store, transmit and display electrophysiological signals in wireless mode as an aid in diagnostics. The system digitizes analogue EEG signals collected by a cap with electrodes, amplifies them, and uses WiFi connectivity to transmit the EEG data to a dedicated host computer with the software.

This device is intended to be used by trained health-care personnel. It is restricted to sale by or on order of a physician.

Device Description

The Neuroelectrics device, Enobio Wireless EEG, performs electroencephalographic measurements with non-invasive technology.

Enobio is a wireless, battery-operated and portable electrophysiology sensor system for the recording of the electroencephalogram (EEG). It has been designed for use in a clinical environment or hospital.. Enobio use must be controlled by specialized medical personnel able to guarantee the correct recording.

ENOBIO WIRELESS EEG, consists of three models designated as the 8, 20 and 32 channels. ENOBIO WIRELESS EEG has a sampling rate of 500 SPS, it works in a bandwidth from 0 to 125Hz (DC coupled) and it provides a resolution of 24 bits - 0.05μV. Enobio is capable of acquiring a variety of electrophysiological signals. These signals include EEG, EOG, ECG and EMG.

The Enobio system is primarily made up of the following three items:

  • Necbox (amplifier, transmitter and control box),
  • the neoprene cap with the electrodes and
  • the NIC software.

The Necbox, which has a data amplifier, connects directly via Wi-Fi to the Neuroelectrics Instrument Controller (NIC) software running on a computer. The EEG data is streamed through the standard Wi-Fi band, and the standard Wi-Fi operating distance range is 10 meters. In all situations the amplifiers store a copy of the data locally to allow for data back-up. This amplifier provides storage and subsequent transmission of data that is not transferred live when the amplifier is in out of range situations.

AI/ML Overview

The provided document is a 510(k) summary for the Enobio Wireless EEG device. It does not present a study proving the device meets acceptance criteria in the typical sense of a clinical trial or performance study with detailed metrics against defined acceptance criteria. Instead, it focuses on demonstrating substantial equivalence to a predicate device through non-clinical performance testing and a comparison of characteristics.

Therefore, many of the requested sections about acceptance criteria, sample sizes, ground truth, expert adjudication, and comparative effectiveness studies are not applicable directly from this document, as the submission strategy relies on non-clinical data and equivalence.

Here's a breakdown of the available information:

1. A table of acceptance criteria and the reported device performance

The document doesn't provide a table of acceptance criteria with quantitative performance metrics against them. Instead, it states that the device "passed all the testing in accordance with internal requirements, national standards, and international standards" and lists the types of tests conducted. These tests implicitly serve as the performance evaluation.

Test CategoryReported Device Performance/Compliance
BiocompatibilityPassed testing in accordance with: - ISO 10993-1: 2009 (Biological Evaluation of Medical Devices Part 1: Evaluation and Testing) - ISO 10993-5: 2009 (Test for In Vitro Cytotoxicity) - ISO 10993-10: 2010 (Tests for Irritation and Delayed-Type Hypersensitivity)
Electrical PerformancePassed testing in accordance with: - IEC 60601-1: 2005 + CORR. 1 (2006) + CORR 2 (2007) (General Requirements for Basic Safety and Essential Performance) - IEC 60601-1-2:2007 (Third Edition) (EMC Requirements and Tests) - IEC 60601-2-26:2012 (Particular requirements for electroencephalographs) - IEC 60601-1-11: 2010 (Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Compatibility) - IEC 61000-4-3 (Radiated, radio-frequency, electromagnetic field immunity test)
Software V&VTesting conducted in accordance with IEC 62304 and FDA Guidance
DisinfectionTesting completed
Electrode CompatibilityTesting completed
Product Specification Compliance TestsPassed for: - Bandwidth - Sampling Rate (Fixed 500 Hz, within predicate's range) - Input Referred Noise - Dynamic Range - Operating Hours - Resolution (24 bits - 0.05μV) - Input Impedance

2. Sample sized used for the test set and the data provenance

Not applicable for non-clinical performance testing of a hardware device. The tests were in a lab setting, not on patient data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. The "ground truth" for non-clinical hardware performance testing is established by engineering specifications and national/international standards, not by expert medical interpretation.

4. Adjudication method

Not applicable.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The submission explicitly states: "There was no human clinical testing required to support the medical device as the indications for use is equivalent to the predicate device."

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

The device is an EEG acquisition system, not an AI algorithm. Its performance is evaluated based on its ability to accurately acquire and transmit electrophysiological signals, which were tested via non-clinical performance tests.

7. The type of ground truth used

For the non-clinical performance data, the "ground truth" is defined by the technical specifications and accepted international standards for medical electrical equipment and electroencephalographs (e.g., IEC 60601 series, ISO 10993 series). Performance is measured against these established technical benchmarks.

8. The sample size for the training set

Not applicable. This is a hardware device; no "training set" in the context of machine learning is involved.

9. How the ground truth for the training set was established

Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the symbol. The logo is black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 2, 2017

Neuroelectrics Barcelona S.L.U. % Deirdre Barrow Senior Consultant, Regulatory Emergo Global Consulting, LLC 2500 Bee Cave Road Building 1, Suite 300 Austin. Texas 78746

Re: K162681

Trade/Device Name: Enobio Wireless EEG Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: Class II Product Code: GWQ, GWL Dated: April 21, 2017 Received: April 27, 2017

Dear Deirdre Barrow:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

Michael J. Hoffmann -S

for

Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K162681

Device Name Enobio Wireless EEG

Indications for Use (Describe)

ENOBIO is an EEG portable monitoring device of 8, 20, 32 channels intended for the use in clinical patient monitoring for use in hospitals and other medical environments.

The Enobio is intended to acquire, store, transmit and display electrophysiological signals in wireless mode as an aid in diagnostics. The system digitizes analogue EEG signals collected by a cap with electrodes, amplifies them, and uses WiFi connectivity to transmit the EEG data to a dedicated host computer with the software.

This device is intended to be used by trained health-care personnel. It is restricted to sale by or on order of a physician.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

K162681

1. Submission Sponsor

Neuroelectrics Barcelona SLU

Avenida Tibidabo 47 bis

08035 Barcelona

SPAIN

Phone number: + 34 932540370

Contact: Ana Maiques

Title: CEO

2. Submission Correspondent

Emergo Global Consulting, LLC

2500 Bee Cave Road

Building 1, Suite 300

Austin, TX 78746

Cell Phone: +353 86 8733815

Office Phone: (512) 327.9997

Contact: Deirdre Barrow, Senior Consultant, RA

Email: project.management@emergogroup.com

3. Date Prepared

September 21, 2016

4. Device Identification

Trade/Proprietary Name:ENOBIO WIRELESS EEG
Common/Usual Name:Electroencephalograph
Classification Name:Full-Montage Standard Electroencephalograph
Regulation Number:882.1400 & 882.1835

{4}------------------------------------------------

Product Code:GWQ, Full-Montage Standard Electroencephalograph &GWL, Amplifier, Physiological Signal
Device Class:Class II
Classification Panel:Neurology

5. Legally Marketed Predicate Device

DeviceName510(k)No.Product CodeClassificationRegulationSponsor
NicoletWirelessEEGK103140GWL & GWQ21 CFR 882.1835 &21 CFR 882.1400Carefusion 209, Inc.

6. Indication for Use Statement

ENOBIO is an EEG portable monitoring device of 8, 20, 32 channels intended for the use in clinical patient monitoring for use in hospitals and other medical environments.

The Enobio is intended to acquire, store, transmit and display electrophysiological signals in wireless mode as an aid in diagnostics. The system digitizes analogue EEG signals collected by a cap with electrodes, amplifies them, and uses WiFi connectivity to transmit the EEG data to a dedicated host computer with the software.

This device is intended to be used by trained health-care personnel. It is restricted to sale by or on order of a physician.

7. Device Description

The Neuroelectrics device, Enobio Wireless EEG, performs electroencephalographic measurements with non-invasive technology.

Enobio is a wireless, battery-operated and portable electrophysiology sensor system for the recording of the electroencephalogram (EEG). It has been designed for use in a clinical environment or hospital.. Enobio use must be controlled by specialized medical personnel able to guarantee the correct recording.

ENOBIO WIRELESS EEG, consists of three models designated as the 8, 20 and 32 channels. ENOBIO WIRELESS EEG has a sampling rate of 500 SPS, it works in a bandwidth from 0 to 125Hz (DC coupled) and it provides a resolution of 24 bits - 0.05μV. Enobio is capable of acquiring a variety of electrophysiological signals. These signals include EEG, EOG, ECG and EMG.

The Enobio system is primarily made up of the following three items:

{5}------------------------------------------------

  • Necbox (amplifier, transmitter and control box),
  • the neoprene cap with the electrodes and
  • the NIC software.

The Necbox, which has a data amplifier, connects directly via Wi-Fi to the Neuroelectrics Instrument Controller (NIC) software running on a computer. The EEG data is streamed through the standard Wi-Fi band, and the standard Wi-Fi operating distance range is 10 meters. In all situations the amplifiers store a copy of the data locally to allow for data back-up. This amplifier provides storage and subsequent transmission of data that is not transferred live when the amplifier is in out of range situations.

8. Substantial Equivalence Discussion

The following table compares the ENOBIO WIRELESS EEG to the predicate device with respect to indications for use, principles of operation, technological characteristics, materials, and performance testing. The comparison of the devices provides more detailed information regarding the basis for the determination of substantial equivalence. The subject device does not raise any new issues of safety or effectiveness based on the similarities to the predicate device.

ManufacturerNeuroelectricsCarefusionSignificant Differences
Trade NameENOBIO WIRELESS EEGNicolet Wireless EEG
510(k) NumberK162681K103140 (predicate)n/a
Product CodeGWL & GWQGWL & GWQSame
RegulationNumber882.1835 & 882.1400882.1835 & 882.1400Same
Regulation NameAmplifier, PhysiologicalSignal & Full-MontageStandardElectroencephalographAmplifier, PhysiologicalSignal & Full-MontageStandardElectroencephalographSame
Indications for UseENOBIO is an EEGportable monitoringdevice of 8, 20, 32channels intended forthe use in clinical patientmonitoring for use inhospitals and othermedical environments.The Enobio is intended toacquire, store, transmitThe Nicolet WirelessEEG Amplifier isintended to be used as afront end amplifier toacquire, store, andtransmitelectrophysiologicalsignals in a wired orwireless mode for theNicolet NeurodiagnosticNo significant difference aseach device acquires,stores, amplifies andtransmits EEG signals tohost software forneurological applications.
ManufacturerNeuroelectricsCarefusionSignificant Differences
Trade NameENOBIO WIRELESS EEGand displayelectrophysiologicalsignals in wireless modeas an aid in diagnostics.The system digitizesanalogue EEG signalscollected by a cap withelectrodes, amplifiesthem, and uses WiFiconnectivity to transmitthe EEG data to adedicated host computerwith the software.This device is intended tobe used by trainedhealth-care personnel. Itis restricted to sale by oron order of a physicianNicolet Wireless EEGsystem.
Mechanism of ActionThe Enobio Wireless EEGis a wireless andwearableelectrophysiology sensorsystem for the recordingof electroencephalogram(EEG) measurementsThe device is intendedto be used as a front endNicoletEEGwireless32/64amplifier with theNicolet Neurodiagnosticsystem to record,measure, store, analyzeand display cerebral andextra cerebralphysiologic data for EEGand Sleep studies withor without synchronousdigital video. The deviceis wearable on the torsoor headNo Significant Difference; the differences between them are:● The Enobio Wireless EEG is only available in a wireless assembly whilst the Nicolet Wireless EEG is available in a wired and wireless assembly.● The associated software systems are obviously proprietary and are therefore different but the fundamental technology of acquiring, storing, amplifying and transmitting EEG signals to host software is very similar between the two
ManufacturerNeuroelectricsCarefusionSignificant Differences
Trade NameENOBIO WIRELESS EEGNicolet Wireless EEG
TechnologyOverviewSystem components:• cap with electrodes• control box(amplifier,transmitter)• software for PCSystem components:• patient cable• amplifier, CPU anddisplay monitor,telemetry unit• access pointInterface to amplifier:Wired Ethernet(10/100baseT)orWirelessEthernet (802.11b/g)devices• The Enobio Wireless EEGis only available in abattery operatedconfiguration whereasthe predicate device,Nicolet Wireless EEG,can be battery or ACpowered.No significant difference aseach device consists of amethod of capturing thesignal, storing, amplifyingand transmitting the signalNo significant difference asthe predicate device,Nicolet Wireless EEG offerswired or wireless signaltransmission to the hostcomputer used to displayand manage acquired data.The proposed device,Enobio Wireless EEG, offerswireless signal transmissiononly.
A/D conversion:24 bitsA/D conversion:24 bitsSame
Sampling rate:500 HzSampling rate:125, 250, 500, ... 48000HzNo significant difference asthe predicate device,Nicolet Wireless EEG offersthe user a choice ofsampling rates that the usermay select. The proposeddevice, Enobio Wireless
ManufacturerNeuroelectricsCarefusionSignificant Differences
Trade NameENOBIO WIRELESS EEGNicolet Wireless EEGEEG, provides a fixedsampling rate of 500 Hzincluded within the rangeoffered by the predicatedevice.
AnatomicalLocationWorn on the head via aheadcap or headbandWorn on the torsoand/or via a headcapNo significant difference asthe only option available forthe proposed device, EnobioWireless EEG, is a headcapor headband which isincluded with the optionsavailable with the NicoletWireless EEG
MaterialABS 40%c, ABS, gel ,neopreneNot specifiedNo significant difference aswhile a potential differenceexists between the devicesin terms of patientcontacting materials,appropriatebiocompatibility testing hasbeen conducted on theproposed device.
Number ofChannels8, 20, 32 channels32 or 64 channelNo significant difference asthe number of channelsreflects the simpler natureof the Enobio device.Testing has been conductedto ensure each modelperforms in accordancewith the intended use.
TemperatureControl RangeUse Temperature Range:+5 to 45 °CStorage in BoxTemperature Range: -25to +70 °CNot defined but as theuse has been confinedto research institutions,clinic, hospital,operating room andepilepsy evaluationNo significant difference asthe operating environmentis very similar to that of theproposed device thereforethe operating range in
ManufacturerNeuroelectricsCarefusionSignificant Differences
Trade NameENOBIO WIRELESS EEGNicolet Wireless EEG
Pressure RangeAtmospheric Pressure:700 - 1.000 hPaenvironments it wouldbe expected to be withinthe same range as thatof the proposed deviceterms of temperature,pressure and humiditywould be expected to havesimilar parameters.
Humidity Range15 - 93 %
SterilenonoSame
Single-UsenonoSame
Shelf Life5 yearsUnknownUnknown; predicate deviceshelf life is not available
SoftwareMicroprocessorATSAM4SD32CGWLUnknownN/A
Battery OperatedyesyesSame
AC PowerednoyesNo significant difference:The Enobio Wireless EEG isonly available in a batteryoperated configurationwhereas the predicatedevice, Nicolet WirelessEEG, can be battery or ACpowered
Complies with ISO10993-1yesyesSame
Electrical SafetyTesting PassedyesyesSame

Comparison of Characteristics

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

9. Non-Clinical Performance Data

As part of demonstrating safety and effectiveness of Enobio Wireless EEG and in showing substantial equivalence to the predicate devices that are subject to this 510(k) submission, Neuroelectrics completed a number of non-clinical performance tests. The Enobio Wireless EEG meets all the requirements for overall design, biocompatibility, and electrical safety, confirming that the design output meets the design inputs and specifications for the device.

The Enobio Wireless EEG passed all the testing in accordance with internal requirements, national standards, and international standards shown below to support substantial equivalence of the subject device:

{10}------------------------------------------------

  • Biocompatibility; ●
    • ISO 10993-1: 2009, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing o
    • O ISO 10993-5: 2009, Biological Evaluation of Medical Devices - Part 5: Test for In Vitro Cytotoxicity
    • ISO 10993-10: 2010, Biological Evaluation of Medical Devices Part 10: Tests for Irritation O and Delayed-Type Hypersensitivity
    • . Electrical Performance;
      • o IEC 60601-1: 2005 + CORR. 1 (2006) + CORR 2 (2007), Medical Electrical Equipment -Part 1 General Requirements for Basic Safety and Essential Performance
      • o IEC 60601-1-2:2007 (Third Edition), Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements And Tests. (General II (ES/EMC))
      • IEC 60601-2-26:2012, Medical electrical equipment Part 2-26: Particular requirements o for the basic safety and essential performance of electroencephalographs
      • o IEC 60601-1-11: 2010, Medical Electrical Equipment - Part 1-2: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements and Tests; all applicable requirements were met
      • o IEC 61000-4-3, Radiated, radio-frequency, electromagnetic field immunity test
    • . Software verification and validation testing per IEC 62304 and FDA Guidance
    • . Disinfection testing
    • Compatibility of electrodes with Enobio Wireless EEG ●
    • . Product specification compliance tests:
      • Bandwidth O
      • O Sampling Rate
      • Input Referred Noise O
      • O Dynamic Range
      • O Operating Hours
      • o Resolution
      • Input Impedance o

10. Clinical Performance Data

There was no human clinical testing required to support the medical device as the indications for use is equivalent to the predicate device. These types of devices, including the predicate device, have been on the market for many years with proven safety and efficacy for the device. The non-clinical testing detailed in this submission supports the substantial equivalence of the device.

11. Statement of Substantial Equivalence

By definition, a device is substantially equivalent to a predicate device has the same intended use and the same technological characteristics as the previously cleared predicate device. Or the device has the same intended use and different technological characteristics that can be

{11}------------------------------------------------

demonstrated that the device is substantially equivalent to the predicate device, and that the new device does not raise additional questions regarding its safety and effectiveness as compared to the predicate device.

The Enobio Wireless EEG, as designed and manufactured, is determined to be substantially equivalent to the referenced predicate device, Nicolet Wireless EEG.

§ 882.1400 Electroencephalograph.

(a)
Identification. An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.(b)
Classification. Class II (performance standards).