K Number
K162595

Validate with FDA (Live)

Device Name
Trex_HD
Date Cleared
2016-11-04

(50 days)

Product Code
Regulation Number
882.1400
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Trex HD is an electroencephalograph intended to acquire, display, store and archive electroencephalographic signals, intended for electroencephalographic (EEG) or level 1-2 polysomnographic (PSG) recordings. The Trex HD amplifier is designed to be used with Natus NeuroWorks™ or Natus SleepWorks™ software.

Device Description

The Natus Medical Incorporated DBA Excel-Tech Ltd. (XLTEK) Trex_HD is a complete data acquisition system that incorporates built-in amplifiers, A/D converters, digital signal processors, and storage devices. The Trex_HD incorporates the following patient inputs: 24 referential inputs; DC inputs; Differential inputs; Pulse oximeter/photic input; and Patient event switch input. The Trex_HD also incorporates the following connection to a XLTEK DT computer running NeuroWorks or SleepWorks software applications: USB connection; External battery pack connection; and Bluetooth (BT) wireless connection. The Trex_HD is powered by two (2) AA batteries, or may utilize an external battery pack for longer studies.

AI/ML Overview

This document describes a 510(k) premarket notification for the Trex HD electroencephalograph. The information provided outlines the device's technical specifications and the testing performed to demonstrate its substantial equivalence to a predicate device. However, it does not contain information regarding diagnostic performance or a clinical study that proves the device meets specific performance acceptance criteria related to disease detection or diagnosis. Instead, the document focuses on technical verification against established standards.

Here's an analysis of the provided text in relation to your request:

1. A table of acceptance criteria and the reported device performance:

The document describes performance in terms of compliance with various technical and safety standards, rather than diagnostic accuracy or clinical metrics. The acceptance criteria are essentially meeting these standards.

Acceptance Criteria (Standard Compliance)Reported Device Performance
Software:- Robust software development process- Verification and validation according to internal requirements and FDA guidance documents (e.g., "The content of premarket submissions for software contained in medical devices," "Off-the-shelf software use in medical devices," "General principles of software validation") and standards (IEC 62304: 2006)."Results indicate that the XLTEK Trex_HD firmware complies with its predetermined specifications, the applicable guidance documents, and the applicable standards."
Electrical Safety:- Compliance with IEC 60601-1: 2005, Am1: 2012 (Medical electrical equipment – Part 1: General requirements for basic safety and essential performance).- Compliance with IEC 60601-1-11: 2011 (Medical electrical equipment Part 1-11: General requirements for basic safety and essential performance -Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment)."Results indicate that the XLTEK Trex_HD complies with the applicable standards."
Electromagnetic Compatibility:- Compliance with IEC 60601-1-2: 2007 (Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral standard: Electromagnetic compatibility – Requirements and tests)."Results indicate that the XLTEK Trex_HD complies with the applicable standards."
Performance Testing – Bench:- Compliance with internal requirements and applicable clauses of: - IEC 60601-1-6: 2010 (Medical electrical equipment – Part 1-6: General requirements for basic safety and essential performance – Collateral standard: Usability) - IEC 60601-2-26: 2012 (Medical electrical equipment – Part 2-26: Particular requirements for the basic safety and essential performance of electroencephalographs) - IEC 62366: 2007 (Medical devices – Application of usability engineering to medical devices) - ISO 80601-2-61: 2011 (Medical electrical equipment – Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment)."Results indicate that the XLTEK Trex_HD complies with its predetermined specifications and the applicable standards.""The results of these activities demonstrate that the XLTEK Trex_HD is as safe, as effective, and performs as well as or better than the predicate devices. Therefore, the XLTEK Trex_HD is considered substantially equivalent to the predicate devices."

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

This document details bench testing and verification/validation activities for hardware and software compliance with standards, not a clinical study involving patient data. Therefore, there's no mention of a "test set" in the context of patient data, sample size, or data provenance. The testing was likely conducted in a lab environment.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. This was not a clinical study with a "ground truth" derived from expert interpretation of patient data. The "ground truth" here is compliance with technical and safety standards, which would be assessed by engineers and quality assurance professionals.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. No clinical test set.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not an AI-assisted diagnostic device, nor is it a multi-reader multi-case study. It's an Electroencephalograph (EEG) system.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. The device itself is an ECG system, not an algorithm. The software mentioned is part of the system's operation and data management, and its performance was verified against its specifications and relevant standards.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For the technical and safety performance, the "ground truth" is defined by the specific requirements and limits set forth in the international and FDA-recognized standards (e.g., IEC 60601-1, IEC 60601-2-26) and the device's predetermined specifications.

8. The sample size for the training set

Not applicable. This document does not describe a machine learning or AI algorithm with a training set.

9. How the ground truth for the training set was established

Not applicable. No training set.

In summary, the provided document focuses on demonstrating technical and safety substantial equivalence of the Trex HD electroencephalograph to a predicate device, as required for a 510(k) submission. It details compliance with manufacturing, electrical safety, EMC, and performance standards through bench testing and verification/validation activities. It does not include information about clinical performance, diagnostic accuracy, or studies involving patient data interpreted by experts.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, with three human profiles facing right, representing health and human services.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 4, 2016

Natus Medical Incorporated Dba Excel-tech Ltd. (xltek) Sanjay Mehta Senior Manager, Ouality Assurance & Regulatory Affairs 2568 Bristol Circle Oakville, CA

Re: K162595

Trade/Device Name: Trex HD Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: Class II Product Code: GWQ, OLV Dated: September 14, 2016 Received: September 15, 2016

Dear Sanjay Mehta:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

Michael J. Hoffmann -A

Carlos L. Peña, PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K162595

Device Name Trex HD

Indications for Use (Describe)

The Trex HD is an electroencephalograph intended to acquire, display, store and archive electroencephalographic signals, intended for electroencephalographic (EEG) or level 1-2 polysomnographic (PSG) recordings. The Trex HD amplifier is designed to be used with Natus NeuroWorks™ or Natus SleepWorks™ software.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)
-----------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Submission Date:14 September 2016
Submitter:Natus Medical Incorporated DBA Excel-Tech Ltd. (XLTEK)2568 Bristol CircleOakville, Ontario, L6H 5S1Canada
Submitter andApplicationCorrespondentMr. Sanjay MehtaPhone: +1 (905) 287-5055Fax: +1 (905) 829-5304Email: sanjay.mehta@natus.com
Manufacturing Site:Natus Medical Incorporated DBA Excel-Tech Ltd. (XLTEK)2568 Bristol CircleOakville, Ontario, L6H 5S1Canada
Trade Name:Trex_HD
Common andClassificationName:Electroencephalograph
ClassificationRegulation:21 CFR §882.1400
Product Code:GWQ, OLV
SubstantiallyEquivalent Devices:New ModelPredicate 510(k)NumberPredicateManufacturer / Model
Trex_HDK131266XLTEK / Trex_HD
Device Description:The Natus Medical Incorporated DBA Excel-Tech Ltd. (XLTEK)Trex_HD is a complete data acquisition system that incorporates built-inamplifiers, A/D converters, digital signal processors, and storage devices
The Trex_HD incorporates the following patient inputs:
● 24 referential inputs;● DC inputs;● Differential inputs;● Pulse oximeter/photic input; and● Patient event switch input.

{4}------------------------------------------------

Device Description:(continued)The Trex_HD also incorporates the following connection to a XLTEK DT computer running NeuroWorks or SleepWorks software applications:USB connection; External battery pack connection; and Bluetooth (BT) wireless connection. The Trex_HD is powered by two (2) AA batteries, or may utilize an external battery pack for longer studies.
Intended Use:The Trex_HD is an electroencephalograph intended to be used to acquire, display, store and archive electroencephalographic signals, intended for electroencephalographic (EEG) or level 1-2 polysomnographic (PSG) recordings. The Trex_HD amplifier is designed to be used with Natus NeuroWorks™ or Natus SleepWorks™ software.
TechnologyThe XLTEK Trex HD employs the same technological characteristics as

Comparison:

The XLTEK Trex_HD employs the same technological characteristics as the predicate device.

ECG SystemCharacteristicXLTEK Trex_HD PredicateDevice(K131266)XLTEK Trex_HD(Proposed Device)
Intended UseThe Trex_HD is anelectroencephalograph intended tobe used to acquire, display, storeand archiveelectroencephalographic signals,intended forelectroencephalographic (EEG) orlevel 1-2 polysomnographic(PSG) recordings. The Trex_HDamplifier is designed to be usedwith Natus NeuroWorks™ orNatus SleepWorks™ software.Same
24 Referential Inputs± 10Same
Four (4) DifferentialInputs± 20Same
Electrode Connections(including commoninput)Safety TouchSame
Four (4) Non-isolatedDC Inputs± 5 VSame
Oximeter/PhoticStimulator ConnectionYes (either/or)Same
Patient Event ButtonYesSame
BatteriesTwo (2) AASame

{5}------------------------------------------------

ECG SystemCharacteristicXLTEK Trex_HD PredicateDevice(K131266)XLTEK Trex_HD(Proposed Device)
Compatible withRechargeable ExternalBattery PackNoYes
Bluetooth UseTransmission of time stamp fromvideo recorder to amplifier toallow time synchronizationbetween video images recorded ina separate camcorder and thephysiological signals recorded onthe Trex_HD amplifer.Transmission of streaming EEGdata over Bluetooth fromTrex_HD to acquisition PCduring the study to allow formonitoring of study data in realtime.

Summary of Performance Testing:

Software The XLTEK Trex_HD firmware was designed and developed according to a robust software development process, and was rigorously verified and validated. Firmware information is provided in accordance with internal requirements and the following FDA guidance documents and standards:

  • . The content of premarket submissions for software contained in medical devices, 11 May 05.
  • . Off-the-shelf software use in medical devices, 09 Sep 99.
  • . General principles of software validation; Final guidance for industry and FDA staff, 11 Jan 02.
  • . Content of premarket submissions for management of cybersecurity in medical devices, 02 Oct 14.
  • IEC 62304: 2006, Medical device software – Software life cycle processes

Results indicate that the XLTEK Trex_HD firmware complies with its predetermined specifications, the applicable guidance documents, and the applicable standards.

The XLTEK Trex_HD was verified for performance in accordance with Electrical Safety the following standard:

  • IEC 60601-1: 2005, Am1: 2012, Medical electrical equipment – Part 1: General requirements for basic safety and essential performance.
  • IEC 60601-1-11: 2011, Medical electrical equipment Part 1-11: ● General requirements for basic safety and essential performance -Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.

Results indicate that the XLTEK Trex_HD complies with the applicable standards.

{6}------------------------------------------------

Electromagnetic The XLTEK Trex HD was verified for performance in accordance with Compatibility the following standard: IEC 60601-1-2: 2007, Medical electrical equipment – Part 1-2: . General requirements for basic safety and essential performance – Collateral standard: Electromagnetic compatibility – Requirements and tests. Results indicate that the XLTEK Trex_HD complies with the applicable standards. Performance The XLTEK Trex_HD was verified for performance in accordance with Testing – Bench internal requirements and the applicable clauses of the following standards: IEC 60601-1-6: 2010, Medical electrical equipment – Part 1-6: . General requirements for basic safety and essential performance – Collateral standard: Usability . IEC 60601-2-26: 2012, Medical electrical equipment – Part 2-26: Particular requirements for the basic safety and essential performance of electroencephalographs. ● IEC 62366: 2007, Medical devices – Application of usability engineering to medical devices. ● ISO 80601-2-61: 2011, Medical electrical equipment – Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment. Results indicate that the XLTEK Trex_HD complies with its predetermined specifications and the applicable standards. Verification and validation activities were conducted to establish the Conclusion performance and safety characteristics of the device modifications made to the XLTEK Trex HD. The results of these activities demonstrate that the XLTEK Trex_HD is as safe, as effective, and performs as well as or better than the predicate devices. Therefore, the XLTEK Trex_HD is considered substantially equivalent to the predicate devices.

§ 882.1400 Electroencephalograph.

(a)
Identification. An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.(b)
Classification. Class II (performance standards).