K Number
K162527

Validate with FDA (Live)

Date Cleared
2016-09-30

(21 days)

Product Code
Regulation Number
892.2050
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The 8MP Color LCD Monitor (C81W+) is intended to be used in displaying and viewing digital images, including standard and multi-frame digital mammography, for review, analysis and diagnosis by trained medical practitioners. It is specially designed for breast tomosynthesis applications.

Device Description

The C81W+ is an 8MP color LCD monitor for viewing medical images including breast tomosynthesis, digital mammography, ultrasound and general radiography. Its using the latest generation of LED back light panel, with high resolution 4096X2160, built-in brightness stabilization circuit, front sensor and ambient light sensor, stable brightness and persistent calibration can be guaranteed throughout its life. The anti-glare screen can prevent display from reflection under highlight conditions, make the image and display clearer.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study for the 8MP Color LCD Monitor (C81W+), structured to address your specific points.

Acceptance Criteria and Device Performance Study for 8MP Color LCD Monitor (C81W+)

The provided document describes a 510(k) premarket notification for the 8MP Color LCD Monitor (C81W+), comparing it to a predicate device, the EIZO RadiForce RX850. The core of the acceptance criteria and the study lies in demonstrating substantial equivalence to the predicate device, particularly in performance characteristics relevant to medical image display.

1. Table of Acceptance Criteria and Reported Device Performance

The document primarily relies on comparing the technological characteristics of the proposed device (C81W+) with the predicate device (RadiForce RX850) and then validating these characteristics through bench testing against established guidelines. The "acceptance criteria" are implicitly set by the performance of the predicate device and the standards outlined in the AAPM Task Group 18 (TG18 guideline).

Acceptance Criteria (Implicit from Predicate & TG18)Reported Device Performance (C81W+)
Intended Use Equivalence:Met: "totally same as that of the predicate device."
Display digital images for review, analysis, and diagnosis by trained medical practitioners, including standard and multi-frame digital mammography and breast tomosynthesis applications.(Same as predicate)
Display Performance/Specifications:Met/Equivalent:
Screen Technology (TFT Color LCD Panel (IPS))TFT Color LCD Panel (IPS)
Viewing Angle (H: 178°, V: 178°)H: 178°, V: 178°
Resolution (8MP: 4,096 x 2,160)8MP (4,096 x 2,160)
Aspect Ratio (17:9)17:9
Active Screen Size (697.9 mm x 368.0 mm)697.9 mm x 368.0 mm
Pixel Pitch (0.1704 mm x 0.1704 mm)0.1704 mm x 0.1704 mm
Maximum Luminance (850 cd/m²)850 cd/m²
DICOM Calibrated Luminance (500 cd/m²)500 cd/m²
Contrast Ratio (1450:1)1450:1
Response Time (typ.) (20 ms (On/Off))20 ms (On/Off)
Backlighting (LED)LED
Grayscale Tones (10-bit input: 1.07 billion colors max; 8-bit input: 16.77 million colors)(Same as predicate)
Luminance non-uniformity (Digital Uniformity Equalizer)Digital Uniformity Equalizer
Video Signal Input:Met/Equivalent:
Input video signals (DVI-D (dual link) x 2, DisplayPort x 2)DVI-D (dual link) x 2, DisplayPort x 2
Scanning Frequency (H, V) (31 - 140 kHz / 59 - 61 Hz)(Same as predicate)
Power Related Specifications:Different but assessed as not affecting safety/effectiveness:
Power Requirements (AC 100 - 120V, 200 - 240 V: 50 / 60 Hz)DC24V/9.0A (Difference in built-in vs. built-out power supply)
Power Consumption / Save Mode (227 W / Less than 6 W)180 W / Less than 5 W (Different design scheme)
Power Management (DVI DMPM, DisplayPort 1.1a)DVI DMPM, DisplayPort 1.2 (Different design scheme)
Miscellaneous Features/Specifications:Different but assessed as not affecting safety/effectiveness:
QC software (RadiCS)Beacon Monitor Manage (Different design scheme)
Sensor (Backlight Sensor, Integrated Front Sensor, Presence Sensor, Ambient Light Sensor)Backlight Sensor, Integrated Front Sensor, Ambient Light Sensor (Predicate has "Presence Sensor," proposed does not)
USB Ports/Standard (1. Upstream, 2. Downstream / Rev. 2.0)(Same as predicate)
Dimensions w/o stand (747 x 430 x 130 mm)755 x 433.8 x 99.1 mm (Different housing design due to panel size)
Image Quality Conformance to TG18 Guideline:Met: Bench tests performed and results showed equivalence.
DICOM GSDF conformancePassed
Angle dependency of luminance responseMeasured and deemed acceptable
Luminance non-uniformity characteristicsMeasured and deemed acceptable
Chromaticity non-uniformity characteristicsMeasured and deemed acceptable
Chromaticity at center (5%, 50%, 95% max luminance)Measured and deemed acceptable
Display reflections (specular, diffuse, haze)Measured and deemed acceptable
Small-spot contrast ratioMeasured and deemed acceptable
Spatial resolution (MTF)Measured and deemed acceptable
Noise (NPS)Measured and deemed acceptable
Pixel aperture ratioMeasured and deemed acceptable
Absence of miscellaneous artifactsVisually checked and deemed acceptable
Temporal responseMeasured and deemed acceptable
Luminance stabilityPerformance data available
Maximum number allowed for each type of pixel defects/faultsMet

2. Sample Size Used for the Test Set and Data Provenance

The document explicitly states: "No animal or clinical testing is needed for C81W+." This means there was no "test set" of patient data in the traditional sense involving medical images for diagnostic performance evaluation. The "test set" here refers to the device itself undergoing various technical bench tests.

  • Test Set Sample Size: N/A (as it refers to the device undergoing technical tests, not a dataset of medical images).
  • Data Provenance: The data comes from bench testing performed on the C81W+ device, following guidelines like AAPM TG18. The country of origin for the device manufacturer is Shenzhen, China. The testing itself is implied to be conducted by or for Shenzhen Beacon Display Technology Co., Ltd.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

Since no clinical or retrospective patient data was used, there were no experts establishing "ground truth" for a medical image test set in this context. The "ground truth" for the device's technical specifications is established by internationally recognized standards (like DICOM GSDF, AAPM TG18 guideline) and the specifications of the legally marketed predicate device. The experts involved would be engineers or technicians with expertise in display technology and medical imaging standards, who conducted the bench tests and validated conformance. Their specific number and qualifications are not detailed in this premarket notification.

4. Adjudication Method for the Test Set

Not applicable. There was no clinical test set requiring adjudication by medical experts.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not performed. The submission explicitly states "No animal or clinical testing is needed for C81W+." This type of study would involve human readers evaluating cases with and without AI (or in this case, a new display) and comparing their diagnostic performance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Not applicable. The device is a display monitor, not an AI algorithm. Therefore, a "standalone algorithm performance" study is irrelevant here. The performance evaluated is the technical display characteristics.

7. The Type of Ground Truth Used

The "ground truth" for the device's performance is based on:

  • Technical specifications derived from established standards (e.g., DICOM GSDF, AAPM TG18).
  • Performance data of a legally marketed predicate device (EIZO RadiForce RX850).
  • Physical measurements obtained during bench testing of the C81W+ device.

8. The Sample Size for the Training Set

Not applicable. This device is a display monitor, not a machine learning or AI algorithm, so there is no "training set."

9. How the Ground Truth for the Training Set Was Established

Not applicable. As there is no training set for an AI algorithm, there is no ground truth to be established for it.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, with flowing lines above them that resemble a bird in flight. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the logo in a circular fashion.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Shenzhen Beacon Display Technology Co., Ltd. % Miss Fu Ailing Document Engineer 12F, Block B1, Nanshan Zhiyuan, No.1001 Xueyuan Road Shenzhen, Guangdong 518055 CHINA

October 6, 2016

Re: K162527 Trade/Device Name: 8MP Color LCD Monitor (C81W+) Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: PGY Dated: August 1, 2016 Received: September 9, 2016

Dear Miss Ailing:

This letter corrects our substantially equivalent letter of September 30, 2016.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{1}------------------------------------------------

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Jeff Rodriquez

for Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known)

Device Name

8MP Color LCD Monitor (C81W+)

Indications for Use (Describe)

The 8MP Color LCD Monitor (C81W+) is intended to be used in displaying and viewing digital images, including standard and multi-frame digital mammography, for review, analysis and diagnosis by trained medical practitioners. It is specially designed for breast tomosynthesis applications.

Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)
---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

008_510 (k) Summary

{4}------------------------------------------------

510(k) Summary

[As required by 21 CFR 807.92]

1. Date Prepared [21 CFR807.92 (a) (1)]

July 6, 2016

2. Submitter's Information [21 CFR807.92 (a) (1)]

Name of Sponsor:Shenzhen Beacon Display Technology Co., Ltd.
Address:12F, Block B1, Nanshan Zhiyuan, No.1001 Xueyuan Road, NanshanDistrict, Shenzhen, 518055 China
Contact Name:Fu Ailing
Telephone No.:+86-0755-33985118-5150
Fax No.:+86-0755-26995755
Email Address:fuailing@beacon-display.cn

3. Trade Name, Common Name, Classification [21 CFR807.92 (a) (2)]

Trade Name/Model:8MP Color LCD Monitor (C81W+)
Common Name:8MP Color LCD Monitor
Classification Name:Display, Diagnostic Radiology
Regulation Number:21 CFR 892.2050
Product code:PGY
Classification Panel:Radiology
Device Class:II

4. Identification of Predicate Device(s) [21 CFR 807.92(a) (3)]

The identified predicates within this submission are as follows:

EIZO Corporation, 8MP Color LCD Monitor, RadiForce RX850, RX850-AR has been cleared by FDA through 510(k) No. K152713 (Decision Date - November 20, 2015).

{5}------------------------------------------------

5. Description of the Device [21 CFR 807.92(a) (4)]

The C81W+ is an 8MP color LCD monitor for viewing medical images including breast tomosynthesis, digital mammography, ultrasound and general radiography. Its using the latest generation of LED back light panel, with high resolution 4096X2160, built-in brightness stabilization circuit, front sensor and ambient light sensor, stable brightness and persistent calibration can be guaranteed throughout its life. The anti-glare screen can prevent display from reflection under highlight conditions, make the image and display clearer.

6. Intended Use [21 CFR 807.92(a)(5)]

The 8MP Color LCD Monitor (C81W+) is intended to be used in displaying and viewing digital images, including standard and multi-frame digital mammography, for review, analysis and diagnosis by trained medical practitioners. It is specially designed for breast tomosynthesis applications.

Panel31", TFT, color, LCD screen, anti-glare, hard coating
Brightness (Max.)850 cd/m²
CR (Typ.)1450:1
Viewing angleR/L 178°, U/D 178° Typ. (CR > 10)
Pixel Pitch(H) 0.1704 × (V) 0.1704 mm
Native resolution4,096 x 2,160
Display area697.9mm(H)x368.0mm (V)
Compatible videosignals4096 x 2160@60Hz; DP/DVI
Horizontal resolution4096
Aspect ratio17:9
Screen size31.1" real diagonal
PowerDC 24V/9A

7. Technological Characteristics [21 CFR 807.92(a)(6)]

{6}------------------------------------------------

Power consumptionMax. 180 W
Input signalsDVI-D, Display Port
Digital inputTMDS (dual link) * 2
Plug and playVESA DDC 2B
Dimension755 mm (W) x 600.6 mm (H) x 251.4mm (D) (with stand) 755 mm (W) x 433.8 mm (H) x 99.1 mm (D) (without stand)
Weight13 kg (without Stand) 17.5 kg (with Stand)
Operating temperature and humidityTemperature: 0°C ~ 40°C Humidity: 20% ~80%
Storage temperature and humidityTemperature: -20°C ~ 60°C Humidity:10% ~90%

8. Substantial Equivalence [21 CFR 807.92(b) (1) and 807.92]

8.1 Intended uses:

Table 1 Intended Use Comparison

IDComparison ItemProposed Device8MP Color LCD Monitor(C81W+)Predicate Device8MP Color LCD Monitor(RX850, RX850-AR)
1Intended UseThe 8MP Color LCD Monitor(C81W+) is intended to beused in displaying and viewingdigital images, includingstandard and multi-framedigital mammography, forreview, analysis and diagnosisby trained medicalpractitioners. It is speciallydesigned for breasttomosynthesis applications."8MP Color LCD Monitor(RX850, RX850-AR)" is intendedto be used in displaying andviewing digital images, includingstandard and multi-frame digitalmammography, for review,analysis and diagnosis by trainedmedical practitioners. It isspecially designed for breasttomosynthesis application.
IDComparison ItemProposed Device8MP Color LCDMonitor (C81W+)Predicate Device8MP Color LCDMonitor(RX850, RX850-AR)Explanation ofDifference
2Display Performance/Specifications
2.1ScreenTechnologyTFT Color LCD Panel(IPS)TFT Color LCD Panel(IPS)-
2.2Viewing angle(H, V)H: 178°, V: 178°H: 178°, V: 178°-
2.3Resolution8MP (4,096 x 2,160)8MP (4,096 x 2,160)(4MP: 2,048 x 2,160x2)-
2.4Aspect ratio17:9 (8.5 : 9 x 2)17:9 (8.5 : 9 x 2)-
2.5Active screensize697.9 mm x 368.0 mm697.9 mm x 368.0 mm-
2.6Pixel pitch0.1704 mm x 0.1704mm0.1704 mm x 0.1704mm-
2.7Maximumluminance850 cd/m²850 cd/m²-
2.8DICOMcalibratedluminance500 cd/m²500 cd/m²-
2.9Contrast ratio1450:11450:1-
2.10ResponseTime (typ.)20 ms (On/Off)20 ms (On/Off)-
2.11BacklightingLEDLED-
2.12GrayscaleTonesFrom a palette of 68billion colors:- 10-bit input(DisplayPort): 1.07billion colors(maximum)- 8-bit input: 16.77million colorsFrom a palette of 68billion colors:- 10-bit input(DisplayPort): 1.07billion colors(maximum)- 8-bit input: 16.77million colors-
2.13Luminancenon-uniformityDigital UniformityEqualizerDigital UniformityEqualizer-
3Video Signal Input
3.1Input video signalsDVI-D (dual link) x 2, DisplayPort x 2DVI-D (dual link) x 2, DisplayPort x 2 (two inputs are required)
3.2Scanning Frequency (H, V)31 - 140 kHz / 59 - 61 Hz (VGA Text: 69 - 71 Hz) Frame synchronous mode: 29.5 - 30.5 Hz, 59 - 61 Hz31 - 140 kHz / 59 - 61 Hz (VGA Text: 69 - 71 Hz) Frame synchronous mode: 29.5 - 30.5 Hz, 59 - 61 Hz
4Power Related Specifications
4.1Power RequirementsDC24V/9.0AAC 100 - 120V, 200 - 240 V: 50 / 60 HzThe difference between Built-in power supply and Built-out power supply.
4.2Power Consumption / Save Mode180 W / Less than 5 W227 W / Less than 6 WDifferent design scheme.
4.3Power ManagementDVI DMPM, DisplayPort 1.2DVI DMPM, DisplayPort 1.1aDifferent design scheme.
5Miscellaneous Features/Specifications
5.1QC softwareBeacon Monitor ManageRadiCSDifferent design scheme.
5.2SensorBacklight Sensor, Integrated Front Sensor, Ambient Light SensorBacklight Sensor, Integrated Front Sensor, Presence Sensor, Ambient Light Sensor
5.3USB Ports/Standard1. Upstream, 2. Downstream / Rev. 2.01.Upstream, 2.Downstream / Rev. 2.0
5.4Dimensions w/o stand (W x H x D)755 x 433.8 x 99.1 mm747 x 430 x 130 mmDifferent housing design due to the different panel size.

8.2 Comparison table

Table 2 General Comparison

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

It is clear that the technological characteristics differences discussed above do not affect the safety and the effectiveness of the C81W+.

8.3 Performance Testing

According to the instructions in Guidance for Industry and FDA Staff: Display Accessories for Full-Field Digital Mammography Systems-Premarket Notification (510(k)) Submissions, the bench tests were performed on C81W+ as below.

  • Verify the conformance to DICOM GSDF in accordance with Assessment of Display Performance for Medical Imaging Systems by AAPM Task Group 18 (TG18 guideline).
  • Measure the angle dependency of luminance response in horizontal, vertical and diaqonal directions.
  • Measure the luminance non-uniformity characteristics of the display screen in accordance with TG18 quideline.
  • Measure the chromaticity non-uniformity characteristics of the display screen in accordance with TG18 guideline.
  • Measure the chromaticity at the center of the display screen at 5%, 50% and 95% of the maximum luminance.
  • Measure the display reflections including specular, diffuse and haze components.
  • Measure the small-spot contrast ratio.
  • Measure the spatial resolution expressed as modulation transfer function (MTF).
  • Measure the noise expressed as noise power spectrum (NPS).
  • Measure the pixel aperture ratio.
  • Visually check the presence or absence of miscellaneous artifacts on the display screen in accordance with TG18 guideline.
  • Measure the temporal response.
  • The performance data on luminance stability
  • Maximum number allowed for each type of pixel defects/faults

The test results showed that C81W+ is with display characteristics equivalent to those of the predicate device, RadiForce RX850 except some items, each of which was determined that it would not affect observer's performance.

{10}------------------------------------------------

No animal or clinical testing is needed for C81W+.

9. Conclusion [21 CFR 807.92(b) (3)]

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification. Shenzhen Beacon Display Technology Co., Ltd. concludes that:

  • The intended use of C81W+ is totally same as that of the predicate device.
  • The technological characteristics differences between C81W+ and RadiForce RX850 do not affect the safety and effectiveness, no new risk is raised.
  • Demonstrated by the bench tests, the display characteristics of C81W+ are equivalent to those of the predicate device.

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).