K Number
K162518

Validate with FDA (Live)

Device Name
1012WCC, 1012WGC
Manufacturer
Date Cleared
2016-10-06

(27 days)

Product Code
Regulation Number
892.1680
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

1012WCC and 1012WGC Digital Flat Panel X-Ray Detector are indicated for digital imaging solution designed for human anatomy including head, neck, cervical spine, arm, leg and peripheral (foot, hand, wrist, fingers, etc.). They are intended to replace film based radio diagnostic systems and provide a case diagnosis and treatment planning for physicians and other health care professionals. Not to be used for mammography.

Device Description

1012WCC / 1012WGC is a wired/wireless digital solid state X-ray detector that is based on flat-panel technology. The wireless LAN(IEEE 802.11a/g/n/ac) communication signals images captured to the system and improves the user operability through high-speed processing. This radiographic image detector and processing unit consists of a scintillator coupled to an a-Si TFT sensor. This device needs to be integrated with a radiographic imaging system. It can be utilized to capture and digitalize X-ray images for radiographic diagnosis. The RAW files can be further processed as a DICOM compatible image file by a separate console SW program (K160579 / Xmaru View V1 and Xmaru PACS/ Rayence Co.,Ltd.) for a diagnostic analysis.

AI/ML Overview

The provided text describes the 510(k) premarket notification for the Rayence Co., Ltd.'s 1012WCC and 1012WGC Digital Flat Panel X-Ray Detectors. The document focuses on demonstrating substantial equivalence to a predicate device (1012WCA) and a reference device (1210SGA), rather than establishing novel acceptance criteria for a new device type.

Therefore, the acceptance criteria are implicitly defined by demonstrating equivalence to the performance of the predicate and reference devices. The "study" here refers to the performance testing conducted to prove this equivalence.

Here's an analysis of the provided information, framed to address your questions about acceptance criteria and the supporting study:

Acceptance Criteria and Reported Device Performance

The acceptance criteria are not explicitly stated as numerical thresholds to be met, but rather as demonstrating substantial equivalence to the predicate (1012WCA) and reference (1210SGA) devices in key performance metrics. These metrics are:

MetricAcceptance Criteria (Implied: Equivalent/Superior to Predicate/Reference)Reported Device Performance (1012WCC vs 1012WCA)Reported Device Performance (1012WGC vs 1210SGA)
Modulation Transfer Function (MTF)Equivalent to predicate/reference at various spatial frequencies.1012WCC shows similar/slightly higher MTF than 1012WCA, especially at higher frequencies (e.g., 3.93 lp/mm).1012WGC MTF performance "almost same" as 1210SGA.
Detective Quantum Efficiency (DQE)Equivalent to predicate/reference, particularly DQE(0).1012WCC DQE(0) = 0.778 (vs. 1012WCA DQE(0) = 0.753). "1012WCC has higher DQE performance at high spatial frequencies, especially from 2.5 lp/mm to 4 lp/mm, compared with 1012WCA."1012WGC DQE(0) = 0.437 (vs. 1210SGA DQE(0) = 0.470). Performance results "almost same".
Noise Power Spectrum (NPS)Equivalent to predicate/reference.Tested (results not explicitly detailed, but implied as satisfactory).Tested (results not explicitly detailed, but implied as satisfactory).
Clinical Image QualityDiagnostic image quality equivalent/superior to predicate/reference."images obtained with the 1012WCC/1012WGC were superior to the same view obtained from a similar patient with the 1012WCA/1210SGA, respectively." Specifically, "both the spatial and soft tissue contrast resolution are superior using the 1012WCC/1012WGC.""images obtained with the 1012WCC/1012WGC were superior to the same view obtained from a similar patient with the 1012WCA/1210SGA, respectively." Specifically, "soft tissues on extremity films were seen with better clarity."
Safety and Performance (Electrical, Mechanical, Environmental)Conform to IEC 60601-1:2005 and IEC 60601-1-2:2007 standards.All test results satisfactory.All test results satisfactory.

Study Information (Performance Testing)

The document describes "non-clinical test" and "clinical consideration test" to demonstrate substantial equivalence.

  1. Sample sizes used for the test set and the data provenance:

    • Clinical Consideration Test: The document states "sample radiographs of similar age groups and anatomical structures." It does not specify the exact number of images or patients (sample size) in the clinical test set.
    • Data Provenance: Not explicitly stated regarding the origin (e.g., country) of the clinical data. It is a "clinical consideration test," meaning it's likely a small comparative review rather than a large clinical trial. The images are "taken from both subject devices" (1012WCC/WGC) and compared to images from the predicate/reference devices (1012WCA/1210SGA). This suggests a prospective collection of comparison images, not necessarily a large retrospective dataset.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Only one expert was used: "a licensed US radiologist."
    • Qualifications: "a licensed US radiologist." No further details on years of experience or subspecialty.
  3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • There was no formal adjudication method described. The review was conducted by a single "licensed US radiologist" who rendered an "expert opinion." This falls into the "none" category for a multi-expert adjudication process.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No MRMC study was done. This study solely focuses on the image quality of the device itself (hardware) compared to predicate devices, not on the interaction of human readers with AI assistance. The devices are flat-panel X-ray detectors, not AI algorithms for interpretation.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Yes, in essence, standalone technical performance was done. The non-clinical tests (MTF, DQE, NPS) are standalone evaluations of the device's image quality metrics, independent of human interpretation.
    • The "clinical consideration test" is a standalone evaluation of the image quality from the devices by an expert, focusing on diagnostic utility, rather than an "algorithm only" performance. The device itself is the "algorithm" in terms of image generation.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For the non-clinical tests (MTF, DQE, NPS): The ground truth is based on physical measurements and established international standards (IEC 62220-1) for image performance.
    • For the clinical consideration test: The ground truth for image quality assessment was the expert opinion/review of a single licensed US radiologist. There's no mention of pathology or outcomes data to establish clinical ground truth.
  7. The sample size for the training set:

    • Not applicable / Not explicitly mentioned. The document describes a comparison between new devices and predicate devices, demonstrating substantial equivalence for hardware. It does not describe a machine learning model that would require a "training set." The closest analogy might be the development data used for the original predicate devices, but that's not detailed here.
  8. How the ground truth for the training set was established:

    • Not applicable. As there's no mention of a machine learning model or training set, this question is not addressed.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes entwined around it, overlaid on a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper half of the circle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 6, 2016

Rayence Co., Ltd. % Mr. Dave Kim Medical Device Regulatory Affairs Mtech Group 8310 Buffalo Speedway HOUSTON TX 77025

Re: K162518 Trade/Device Name: 1012WCC / 1012WGC Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MQB Dated: September 1, 2016 Received: September 9, 2016

Dear Mr. Kim:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Robert Oals

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K162518

Device Name

1012WCC 1012WGC

Indications for Use (Describe)

1012WCC and 1012WGC Digital Flat Panel X-Ray Detector are indicated for digital imaging solution designed for human anatomy including head, neck, cervical spine, arm, leg and peripheral (foot, hand, wrist, fingers, etc.). They are intended to replace film based radio diagnostic systems and provide a case diagnosis and treatment planning for physicians and other health care professionals. Not to be used for mammography.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

1. 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.

Date 510k summary prepared: October 6, 2016

Submitter's Name, address, telephone number, a contact person:

Submitter's Name :Rayence Co., Ltd.
Submitter's Address:14, Samsung 1-ro 1-gil, Hwaseong-si, Gyeonggi-do, Korea
Submitter's Telephone:+82-31-8015-6459
Contact person:Mr. Kee Dock Kim / RA Team Manager / +82-31-8015-6459
Official Correspondent:Dave Kim (davekim@mtech-inc.net)
(U.S. Designated agent)
Address:8310 Buffalo Speedway, Houston, TX 77025
Telephone:+713-467-2607
Fax:+713-583-8988

Name of the device, including the trade or proprietary name if applicable, the common or usual name and the classification name, if known:

Trade/proprietary name:1012WCC / 1012WGC
Common Name:Digital Flat Panel X-ray Detector
Classification Name :21CFR892.1680 / Stationary x-ray system
Product Code:MQB
Predicate Device :
Manufacturer :Rayence Co., Ltd.
Device :1012WCA
510(k) Number :K123345
Common Name:Digital Flat Panel X-ray Detector
Classification Name :21CFR892.1680 / Stationary x-ray system
Product Code:MQB

{4}------------------------------------------------

Reference Device :
Manufacturer :Rayence Co., Ltd.
Device :1210SGA
510(k) Number :K113630
Common Name:Digital Flat Panel X-ray Detector
Classification Name :21CFR892.1680 / Stationary x-ray system
Product Code:MQB

2. Device Description

1012WCC / 1012WGC is a wired/wireless digital solid state X-ray detector that is based on flat-panel technology. The wireless LAN(IEEE 802.11a/g/n/ac) communication signals images captured to the system and improves the user operability through high-speed processing. This radiographic image detector and processing unit consists of a scintillator coupled to an a-Si TFT sensor. This device needs to be integrated with a radiographic imaging system. It can be utilized to capture and digitalize X-ray images for radiographic diagnosis. The RAW files can be further processed as a DICOM compatible image file by a separate console SW program (K160579 / Xmaru View V1 and Xmaru PACS/ Rayence Co.,Ltd.) for a diagnostic analysis.

3. Indication for use

1012WCC and 1012WGC Digital Flat Panel X-Ray Detector is indicated for digital imaging solution designed for human anatomy including head, neck, cervical spine, arm, leg and peripheral (foot, hand, wrist, fingers, etc.). It is intended to replace film based radiographic diagnostic systems and provide a case diagnosis and treatment planning for physicians and other health care professionals. Not to be used for mammography.

4. Summarv of Design Control Risk management

The 1012WCC / 1012WGC digital X-ray detector is a modification of 1012WCA(K123345), the predicate device which was developed for the purpose of portable imaging. 1012WCC / 1012WGC is slightly thinner, smaller and lighter than 1012WCA.

The risks and the hazardous impact of the device modification were analyzed with FMEA method. The specific risk control and protective measures to mitigate the risks from the modification were reviewed and implemented in the new product design phase. The overall assessment concluded that all risks and hazardous conditions identified arising from the design change were successfully mitigated and accepted.

{5}------------------------------------------------

5. Summary of the technological characteristics of the device compared to the predicate device:

1012WCC and 1012WGC detector described in this 510(k) has the same indications for use and similar technical characteristics as its predicate device, 1012WCA.

5.1Comparison table
CharacteristicProposedRayence Co.,Ltd.1012WCC / 1012WGCPredicateRayence Co.,Ltd.1012WCA
FeatureImage: Rayence Co.,Ltd. 1012WCC / 1012WGCImage: Rayence Co.,Ltd. 1012WCA
510(k) numberK123345
Intended Use1012WCC and 1012WGC Digital FlatPanel X-Ray Detector are indicated fordigital imaging solution designed forhuman anatomy including head, neck,cervical spine, arm, leg and peripheral(foot, hand, wrist, fingers, etc.). They areintended to replace film based radiographicdiagnostic systems and provide a casediagnosis and treatment planning forphysicians and other health careprofessionals. Not to be used formammography.1012WCA Digital Flat Panel X-RayDetector is indicated for digital imagingsolution designed for human anatomyincluding head, neck, cervical spine,arm, leg and peripheral (foot, hand,wrist, fingers, etc.). It is intended toreplace film based radiographicdiagnostic systems and provide a casediagnosis and treatment planning forphysicians and other health careprofessionals. Not to be used formammography.
Detector TypeAmorphous Silicon, TFTAmorphous Silicon, TFT
Scintillator1012WCC CsI:Tl1012WGC Gd2O2S:TbCsI:Tl
Imaging Area10 x 12 inches11 x 13 inches
Pixel matrix2304 X 17922560 X 2080
Pixel pitch127 μm127 μm
Resolution3.9 lp/mm3.9 lp/mm
A/D conversion14 / 16 bit14 / 16 bit
Preview time≤2≤3
Data outputRAW*The RAW files are convertible into DICOM3.0 by console S/WRAW*The RAW files are convertible into DICOM3.0 by console S/W
Dimensions328 X 268 X 15 mm395 x 337 x 18 mm
Weight3.5 kg (incl. battery)3.15 kg
ApplicationWireless/Wired portable systemWireless portable system
stand.stand
AddedBattery & Battery Charger- Battery & Battery Charger
OptionalInterface BoxInterface Box
ComponentsIrDA module

Comparison table 5.1

{6}------------------------------------------------

5.2 Scintillator layer

1012WCC and 1012WGC have the same hardware, software and components. The type of a scintillator layer is different. ( * scintillator : a phosphor that produces scintillations)

ProposedPredicateReference
CsI(Cesium Iodide)1012WCC1012WCA
Gd2O2S:Tb(Gadolinium Oxysulfide)1012WGC1210SGA

5.3 Added Optional Components (Comparison with Predicate device)

ComponentsDescription
Battery & Battery ChargerImage: Battery and battery chargerSources of electricity.
Mobile Battery ChargerImage: Mobile battery charger
Interface BoxImage: Interface box1) Connector to synchronize the detector and the generator.2) Data transfer and battery charge while the detector is in use(Connect between the detector and Interface Box), Up to threedetectors can be connected.3) Transmitting an image/command between the detector and PC.4) Wireless AP.
IrDA moduleImage: IrDA moduleSharing function for PC and the detector.

{7}------------------------------------------------

Recommended Generator Specification

ModelManufacture32kW40kW50kW
CMP 200Communications & PowerIndustrieskVp40-12540-150
mA10-40010-50010-630
EDITOR HFe 501Rontgenwerk BochumkVp40-150
mA10-630
UD150L-40E/40FShimadzukVp40-150
mA@100 kVp- 500(320)
mA@80 kVp- 630(400)
PXR-321BPoskom Co.,Ltd.kVp125/150
mA500

Image /page/7/Picture/3 description: The image shows a caution sign. The sign is a yellow triangle with a black border and a black exclamation point in the center. Below the triangle, the word "CAUTION" is written in black, bold letters.

To our best knowledge, the detector is compatible with the X-ray generators with the specifications described above. If you have questions regarding the compatibility issue for other generators which are not listed above, please contact vour Ravence representative.

6. Summary of Performance Testing

1012WCC and 1012WGC Digital Flat Panel X-Ray Detectors have the same indications for use, material, form factor, performance, and safety characteristics compared to the predicate devices, 1012WCA.

The non-clinical test and clinical consideration test for each subject device were performed to demonstrate the substantial equivalency of the subject devices compared to the predicate device. The nonclinical test report contains the MTF, DQE and NPS test results of 1012WCC by using the identical test equipment and same analysis method described by IEC 62220-1.

The comparative result of the DQE test for 1012WCC detector with respect to the predicate device demonstrated that the DQE of the both subject devcies performed with the predicate device. The DQE represents the ability to visualize object details of a certain size and contrast. 1012WCC has higher DQE performance at high spartial frequencies, especially from 2.5 lp/mm to 4 lp/mm, compared with 1012WCA. The comparison of the MTF and DQE for 1012WGC detector

{8}------------------------------------------------

demonstrated that the MTF and DQE performance results for the subject device were almost same as those of 1210SGA (K113630), the reference device.

SpatialFrequencyMTF Value
1.0 lp/mm0.5600.575
2.0 lp/mm0.2590.278
3.0 lp/mm0.1430.143
3.5 lp/mm0.1080.101
3.93 lp/mm0.0860.074
SpatialFrequencyMTF Value
1.0 lp/mm0.5880.585
2.0 lp/mm0.2750.264
3.0 lp/mm0.1310.119
3.5 lp/mm0.0870.076
3.93 lp/mm0.0620.053
1012WCC1012WCA
DQE(0)0.7780.753
1012WGC1210SGA
DQE(0)0.4370.470

To further demonstrate the substantial equivalency of two devices, clinical images are taken from both subject devices and reviewed by a licensed US radiologist to render an expert opinion. Both the test subjects.1012WCC and 1012WGC, have been evaluated and compared to 1012WCA and 1210SGA, respectively, by taking sample radiographs of similar age groups and anatomical structures in accordance with the test protocol of diagnostic radiography evaluation procedure.

After a broad review of plain radiographic images taken with the 1012WCC/1012WGC and the 1012WCA/1210SGA, the images obtained with the 1012WCC/1012WGC were superior to the same view obtained from a similar patient with the 1012WCA/1210SGA, respectively. In general, both the spatial and soft tissue contrast resolution are superior using the 1012WCC/1012WGC. Specifically, the soft tissues on extremity films were seen with better clarity. There is little difficulty in evaluating a wide range of anatomic structures necessary to provide a correct conclusion.

Based on the non-clinical and clinical consideration test and the outcome of a comparative review by an expert for both devices, the sponsor claims the substantial equivalency between the subject devices and their predicate device in terms of diagnostic image quality.

{9}------------------------------------------------

The manufacturing facility is in conformance with the design control procedure requirements and the relevant EPRC standards as specified in 21 CFR 802.30 and the records are available for review.

7. Summary for any testing in the submission:

Electrical, mechanical, environmental safety and performance testing according to standard IEC 60601-1:2005 (3rd Edition) + CORR. 1:2006 + CORR. 2:2007 + A1:2012 (Medical electrical equipment Part 1:General requirements for basic safety and essential performance) was performed, and EMC testing were conducted in accordance with standard IEC 60601-1-2: 2007.

Non-clinical & Clinical considerations according to FDA Guidance for the Submissions of 510(k)'s for Solid State X-ray Imaging Devices" was performed.

All test results were satisfactory.

8. Summary for risk management activities in the submission For 1012WCC vs 1012WCA

Added functionRisk factorDiscussion
1Cassettesize- Essentialperformance(Image size)Implemented the cassette size to change total pixel matrix(pixels): 2304x1792, total pixel area: 292.6x227.6, andeffective pixel matrix (pixels): 264x1752. The imagequality was tested through SSXI Nonclinical/ Clinical testand to be equivalent compared to the predicate device.
The revised RM table (Att3-1) include the following risk: for1012WCC - Operational hazard (Incorrect image data)Highlighted in yellow (HI-601-1-4.3)

For 1012WGC vs 1012SGA

Added functionRisk factorDiscussion
1Cassette size-. Essential performance (Image size)Implemented the cassette size to change total pixel matrix (pixels): 2304x1792, total pixel area: 292.6x227.6, and effective pixel matrix (pixels): 264x1752. The image quality was tested through SSXI Nonclinical/ Clinical test and to be equivalent compared to the predicate device.The revised RM table (Att3-1) include the following risk: for 1012WGC - Operational hazard (Incorrect image data) Highlighted in yellow (HI-601-1-4.3)
2Wireless-. BatteryThe risk related to a wireless product that uses a battery has been identified and mitigated according to IEC 60601-1, IEC62366 test, and SW validation.The revised RM table; Att3-1 and Att 3-2 include the following risk: Operation hazard (Use error)-Incorrect replacement. Hazard ID (HI-601-1-7.3.3) and Operation

{10}------------------------------------------------

hazard-Batter connection, Hazard ID (HI-601-1-15.4.3.2)Att3-3 SW validation also includes HZ 0017-powerinstability caused by low battery power which is highlightedin yellow.
WirelesssignalAtt 3-3. SW validation also identified the following riskand risk management activities regarding the wirelesssignal; (HZ 0017-power instability caused by low batterypower.(HZ 0019)-Misdiagnosis
Not possible to check the data validity of all data packets,
including CRC value, based on the IrDA communication
transmission in the PC and the detector/HZ 0020)-Misdiagnosis
Image not transmitted to the User PC due to the network
transmission failure.
Check the connection between the detector and the User-PC

9. Conclusions:

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification Rayence Co., Ltd. concludes that 1012WCC and 1012WGC are substantially equivalent in comparison with 1012WCA, the predicate device as described herein.

§ 892.1680 Stationary x-ray system.

(a)
Identification. A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.