K Number
K162418

Validate with FDA (Live)

Manufacturer
Date Cleared
2016-12-02

(94 days)

Product Code
Regulation Number
870.1250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Pioneer Plus catheter is intended to facilitate placement and positioning of catheters within the peripheral vasculature. The Pioneer Plus catheter also provides an intraluminal cross-sectional ultrasound image of interest to facilitate placement of guidewires beyond stenotic lesions (e.g., sub-total, total occlusions) prior to additional intervention (i.e. PTA, stent, etc.). The Pioneer Plus catheter is not indicated for use in the coronary or cerebral vasculature.

Device Description

The Pioneer Plus catheter is a dual lumen catheter that is inserted through a commercially available 6F introducer sheath and placed percutaneously into a peripheral vessel. The Pioneer Plus catheter tracks to its intended site in the vasculature over a standard, commercially available 0.014" (0.36 mm) tracking guidewire. The Pioneer Plus catheter utilizes an extendable, hollow Nitinol needle - minimum internal needle diameter: 0.016" (0.43 mm): it is compatible with floppy 0.014" (0.36 mm) over the wire (OTW) guidewires to facilitate the redirection and placement of the guidewire into peripheral vessels. The needle orientation placement is achieved through the use of an intravascular ultrasound (IVUS) transducer mounted at the distal end of the Pioneer Plus catheter. The Pioneer Plus catheter operator navigates the peripheral vasculature under fluoroscopic guidance and uses IVUS imaging to direct the needle to achieve optimal placement. The operator uses the Needle Deployment Ring incorporated into the handle of the Pioneer Plus catheter to extend the needle from the catheter. The needle can be extended up to 7mm in length, or to a pre-determined distance using the Stop Ring. A 0.014" (0.36 mm) OTW guidewire (needle guidewire) can then be advanced through the needle and used to facilitate the placement of subsequent catheters once the Pioneer Plus device is withdrawn.

AI/ML Overview

The provided text describes an FDA 510(k) summary for a medical device called the "Pioneer Plus Catheter." The acceptance criteria and the study proving the device meets these criteria are outlined in the "PERFORMANCE DATA" section.

Here's a breakdown of the requested information:

1. Table of Acceptance Criteria and the Reported Device Performance:

Acceptance Criteria (Test)Reported Device Performance
Visual InspectionMet acceptance criteria
Outer Diameter (OD) MeasurementMet acceptance criteria
Braid DurabilityMet acceptance criteria
Bend Radius (Kink)Met acceptance criteria
Tensile TestingMet acceptance criteria

The document states that "All testing met the acceptance criterial with 95/95 confidence/reliability."

2. Sample size used for the test set and the data provenance:

  • Sample Size: The document does not explicitly state the sample size used for each specific test (e.g., number of catheters tested for Bend Radius). It only mentions that "Non-clinical device testing was conducted to confirm the performance of the modified device."
  • Data Provenance: Not specified. The tests are described as "Bench testing," indicating they were performed in a lab setting, likely by the manufacturer (Philips Volcano). There is no information regarding country of origin of data, and it is inherently a prospective study as it's testing a modified device's performance.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This information is not provided in the document. The testing appears to be focused on engineering performance criteria rather than human interpretation or clinical outcomes that would require expert ground truth.

4. Adjudication method for the test set:

This information is not applicable and therefore not provided. The tests are objective engineering measurements (e.g., OD measurement, tensile strength) rather than subjective assessments requiring adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This information is not provided and is not applicable to the type of device and testing described. The Pioneer Plus Catheter is a physical medical device (catheter with ultrasound imaging), not an AI-driven diagnostic or assistive tool for human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

This information is not provided and is not applicable. The device is not an algorithm, but a physical catheter. While it incorporates an intravascular ultrasound (IVUS) transducer, the performance evaluation described focuses on the physical integrity and characteristics of the catheter itself, not the standalone performance of an ultrasound imaging algorithm.

7. The type of ground truth used:

The ground truth for the performance tests appears to be known standards or product specifications. The document states, "Bench testing was conducted against known standards or product specification."

8. The sample size for the training set:

This information is not provided. The document describes testing of a physical medical device against engineering specifications. There is no mention of a "training set" as would be relevant for machine learning or AI algorithms.

9. How the ground truth for the training set was established:

This information is not provided as there is no mention of a "training set."

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, representing medicine and healing. The symbol is enclosed within a circular border. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper portion of the circle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 2, 2016

Philips Volcano Mary Stanners Regulatory Affairs Specialist 3721 Valley Centre Drive, Suite 500 San Diego, CA 92130

Re: K162418

Trade/Device Name: Pioneer Plus Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: PDU, ITX Dated: November 02, 2016 Received: November 04, 2016

Dear Ms. Stanners:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Image /page/1/Picture/7 description: The image shows the text "Brian D. Pullin -S" in a clear, sans-serif font. The text is horizontally aligned and appears to be part of a document or presentation. The letters are uniformly sized and spaced, contributing to the legibility of the text.

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K162418

Device Name Pioneer Plus Catheter

Indications for Use (Describe)

The Pioneer Plus catheter is intended to facilitate placement and positioning of catheters within the peripheral vasculature. The Pioneer Plus catheter also provides an intraluminal cross-sectional ultrasound image of interest to facilitate placement of guidewires beyond stenotic lesions (e.g., sub-total, total occlusions) prior to additional intervention (i.e. PTA, stent, etc.). The Pioneer Plus catheter is not indicated for use in the coronary or cerebral vasculature.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------


510(K) SUMMARY

SPONSOR:Philips Volcano3721 Valley Centre Drive, Suite 500San Diego, CA 92130
CONTACT/SUBMITTER:Mary StannersRegulatory Affairs SpecialistPhilips Volcano3721 Valley Centre Drive, Suite 500San Diego, CA 92130Tel: (858) 764-1296mary.stanners@philips.com
DATE OF SUBMISSION:November 2, 2016
DEVICE:
Trade Name:Pioneer Plus Catheter
Common Name:Diagnostic Ultrasound Transducer and Ultrasound Transducer

Classification and Product Codes:

CFR NumberClassProduct Code
21 CFR 870.1250 Percutaneous CatheterIIPDU
21 CFR 892.1570 Diagnostic Ultrasound TransducerIIITX

PREDICATE DEVICE:

510(k) Number:K101777
-------------------------

Trade Name: Pioneer Plus Catheter

Common Name: Diagnostic Ultrasound Transducer and Ultrasound Transducer

Classification and Product Codes:

{4}------------------------------------------------

CFR NumberClassProduct Code
21 CFR 870.1250 Percutaneous CatheterIIPDU
21 CFR 892.1570 Diagnostic Ultrasound TransducerIIITX

DEVICE DESCRIPTION:

The Pioneer Plus catheter is intended to facilitate placement and positioning of catheters within the peripheral vasculature. The Pioneer Plus catheter also provides an intraluminal cross-sectional ultrasound image of the area of interest to facilitate placement of guidewires beyond stenotic lesions (e.g., sub-total, total or chronic total occlusions) prior to additional intervention (i.e. PTA, stent, etc.). The Pioneer Plus catheter is not indicated for use in the coronary or cerebral vasculature.

The Pioneer Plus catheter is a dual lumen catheter that is inserted through a commercially available 6F introducer sheath and placed percutaneously into a peripheral vessel. The Pioneer Plus catheter tracks to its intended site in the vasculature over a standard, commercially available 0.014" (0.36 mm) tracking guidewire. The Pioneer Plus catheter utilizes an extendable, hollow Nitinol needle - minimum internal needle diameter: 0.016" (0.43 mm): it is compatible with floppy 0.014" (0.36 mm) over the wire (OTW) guidewires to facilitate the redirection and placement of the guidewire into peripheral vessels. The needle orientation placement is achieved through the use of an intravascular ultrasound (IVUS) transducer mounted at the distal end of the Pioneer Plus catheter. The Pioneer Plus catheter operator navigates the peripheral vasculature under fluoroscopic guidance and uses IVUS imaging to direct the needle to achieve optimal placement. The operator uses the Needle Deployment Ring incorporated into the handle of the Pioneer Plus catheter to extend the needle from the catheter. The needle can be extended up to 7mm in length, or to a pre-determined distance using the Stop Ring. A 0.014" (0.36 mm) OTW guidewire (needle guidewire) can then be advanced through the needle and used to facilitate the placement of subsequent catheters once the Pioneer Plus device is withdrawn.

INDICATIONS FOR USE:

The indications for Use remain the same and are as follows:

The Pioneer Plus catheter is intended to facilitate placement and positioning of catheters within the peripheral vasculature. The Pioneer Plus catheter also provides an intraluminal cross-sectional ultrasound image of the area of interest to facilitate placement of guidewires beyond stenotic lesions (e.g., sub-total, total or chronic total occlusions) prior to additional intervention (i.e. PTA, stent, etc.). The Pioneer Plus catheter is not indicated for use in the coronary or cerebral vasculature.

COMPARISON OF TECHNOLOGICAL CHARACTERISICS:

The modification made to the Pioneer Plus Catheter is a double wrap of Vestamid sheath material to the area located at the bonded joint of the braided stainless steel outer shaft and the Vestamid distal sleeve to improve braid containment. The Vestamid sheath material is already used on the Pioneer Plus Catheter,

{5}------------------------------------------------

there is just an additional wrap of that same material that did not change the Outer Dimension (OD). There was also tightening of a specification for the braided wire and an inspection method. The modifications do not affect the intended use of the device or technologies and it does not alter the fundamental scientific technologies. The proposed design change does not change the fundamental operating principles of the device. The indications for use are identical to those of the currently marketed device. These changes have no impact to the Pioneer Plus Catheter sterilization, materials, or packaging. The modified catheter is substantially equivalently to currently marketed device.

PERFORMANCE DATA:

Non-clinical device testing was conducted to confirm the performance of the modified device. Bench testing was conducted against known standards or product specification and evaluated Visual Inspection, Outer Diameter (OD) Measurement, Braid Durability, Bend Radius (Kink) and Tensile Testing. All testing met the acceptance criterial with 95/95 confidence/reliability.

The successful completion of performance testing concluded that the modified Pioneer Plus Catheter is substantially equivalent to the current Pioneer Plus Catheter.

The design change did not introduce new material and does not impact sterility or packaging. All packaging materials and processes remain the same. The sterile barrier system remains the same. The modified Pioneer Plus Catheter is packaged in the same configuration as the current Pioneer Plus.

BS EN ISO 13485:2012Medical Devices Quality Management System Requirements for Regulatory Purposes
ISO 14971:2012Medical Devices - Application of Risk Management to Medical Devices
10993-1Biological Evaluation of Medical Devices – Part 1: Evaluating and Testing within a Risk Management Process
10993-4Biological Evaluation of Medical Devices – Part 4: Selection of Tests for Interactions with Blood
10993-5Biological Evaluation of Medical Devices – Part 5: Tests for In Vitro Cytotoxicity

For this Pioneer Plus Catheter modification, the following standards as described in this submission's FDA 3654 Standards forms (Attachment F) were followed:

The following tests were completed; visual inspection, outer diameter (OD) measurement, braid durability, bend radius and tensile testing and the modified Pioneer Plus Double Layer Joint design change met all acceptance criteria.

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).