(286 days)
The QT Ultrasound Breast Scanner-1 is for use as an ultrasonic imaging system to provide reflection-mode and transmission-mode images of a patient's breast. The device is not intended to be used as a replacement for screening mammography.
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Small Organ (Breast) - Reflection, Transmission (Speed of Sound)
QT Ultrasound Breast Scanner-1 is intended for ultrasonic breast exams
The QT Ultrasound Breast Scanner-1 is an automated software-controlled ultrasonic imaging system that performs a standardized scan of the whole breast. The OT Ultrasound Breast Scanner-1 is comprised of a Patient Scanning System, Operator Console and Viewer Console.
The Patient Scanning System consists of a patient support table, scan tank, water management system, ultrasound transducer arrays and all associated image processing electronics. The scan tank is centered below a patient's breast and contains the ultrasound transducer arrays. The transducer arrays include a set of three reflection transducers that transmit pulsed ultrasound plane waves into targeted tissues using the water bath in the scan tank as coupling medium. An additional transmitter and receiver array pair collect the ultrasound energy to provide speed of sound values.
During scanning, a patient lies prone on the examination table with the breast suspended in a warm water bath maintained near skin temperature. Images are automatically acquired on a pendant breast positioned with the nipple as a point of reference. The transducer arrays rotate about a vertical axis to circle the breast in the coronal plane. The array is then translated vertically and the scanning process is repeated until the entire breast is scanned, allowing B-scan images to be constructively combined into tomographic, speed of sound and reflection ultrasound images.
The QT Ultrasound Breast Scanner-1 outputs the images to the QTviewer which allows the images to be stored until they are reviewed on a Viewer Console. Coronal, axial and sagittal images are generated for review by the radiologist. Speed of sound images may be queried by the Probe and Region of Interest (ROI) tools provided in the Viewer Console. These tools provide speed of sound values in meters/sec. to aid in diagnostic evaluation of the breast.
Here's a breakdown of the acceptance criteria and study details for the QT Ultrasound Breast Scanner-1, based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
The document doesn't explicitly state "acceptance criteria" in a numerical or pass/fail format per se for performance. Instead, it describes comparative evaluation results against predicate devices (X-ray Mammography and Handheld Ultrasound) to demonstrate "clinical usefulness" and "equivalent or better" image quality. The criteria effectively revolve around demonstrating that the QT Ultrasound Breast Scanner-1 images provide diagnostic information comparable to or superior to existing modalities for visualizing breast anatomy.
Since a numerical acceptance criterion table with specific targets isn't present, I'll describe the performance findings based on the Visual Grading Assessment (VGA) studies conducted.
| Acceptance Criteria (Implied/Demonstrative) | Reported Device Performance (QT Ultrasound Breast Scanner-1) |
|---|---|
| Clinical Usefulness for Visualizing Breast Anatomy (vs. XRM) | For QT Ultrasound vs. XRM: The analysis demonstrated that the QT Ultrasound Breast Scanner-1 produces clinically-useful depictions of patient anatomy. Readers scored the image quality on the QT Ultrasound images as equivalent to or better than on XRM for each feature in more than 90% of breasts. |
| Image Quality for Breast Anatomy Visualization (vs. HHUS) | For QT Ultrasound vs. HHUS: Except for epidermis and pectoralis muscle, readers scored the image quality on the QT Scan as equivalent or better than HHUS on each feature in more than 80% of breasts. Readers scored the image quality of epidermis and muscle as better on the QT Scan in more than 70% of breasts. The findings suggest that QT images are at least equivalent or better in defining the anatomical components of the breast than HHUS. |
| Safety - Acoustic Output | Meets all Track 1 acoustic output requirements. Examples: Max. Mechanical Index (MI) of 0.145-1.186 (vs. 1.9 limit), Max. ISPTA (mW/cm²) of 0.79-1.63 (vs. 94 mW/cm² limit). All reported as "Pass". |
| Safety - Electrical, EMC, Usability, Biocompatibility, Cleaning, Software | All applicable testing conducted, requirements met, and no new issues of safety or effectiveness compared to the predicate device were raised. |
| Performance - System Verification (Measurement Range/Accuracy, Spatial/Contrast Resolution, SoS Uniformity/Accuracy) | System verification testing conducted to ensure design requirements were met. All requirements were met and no new issues of safety or effectiveness compared to the predicate device were raised. |
2. Sample Sizes Used for the Test Set and Data Provenance
- Test Set (VGA Review with QT Ultrasound vs. X-ray Mammography - XRM):
- Sample Size: 22 cases from 20 subjects.
- Data Provenance: Not explicitly stated, but given the context of FDA submission, it refers to clinical data used for evaluation. The "QT Library cases" mention suggests historical/retrospective data. No country of origin is specified.
- Test Set (VGA Review with QT Ultrasound vs. HandHeld Ultrasound - HHUS):
- Sample Size: 17 cases.
- Data Provenance: Not explicitly stated, but also likely retrospective/historical "QT Library cases." No country of origin is specified.
- Representative Clinical Cases (No Grading):
- Sample Size: 16 clinical cases.
- Data Provenance: Not explicitly stated, but implied to be from the clinical setting. "Historical QT Library cases" are mentioned for the VGA studies, suggesting a retrospective collection.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications
The document describes "readers" who performed the image quality assessment, not necessarily those establishing "ground truth" for disease state.
- VGA Review with QT Ultrasound vs. XRM:
- Number of Readers: 4
- Qualifications: "independent board-certified radiology readers."
- VGA Review with QT Ultrasound vs. HHUS:
- Number of Readers: 5
- Qualifications: "independent board-certified radiology readers."
- Representative Clinical Cases:
- Number of Reviewers: 1 (Dr. Elaine Iuanow, M.D.)
- Qualifications: "board certified radiologist, Dr. Elaine Iuanow, M.D. and Chief Medical Officer." This review was for demonstration, not graded assessment.
4. Adjudication Method for the Test Set
The studies described are Visual Grading Assessments (VGAs).
- The document states, "The four readers independently scored the image quality" and similarly for the HHUS comparison, "The five readers independently scored the image quality."
- Statistical analysis was then performed on these independent scores (e.g., median image quality score, proportion of breasts rated better/equivalent).
- There is no mention of an adjudication method (e.g., 2+1, 3+1 consensus) for establishing a single 'ground truth' score for image quality from the readers. The analysis appears to aggregate or summarize the independent reader scores.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size
- Type of Study: The VGA studies described are MRMC studies focused on image quality comparison, not necessarily diagnostic accuracy or human reader improvement with AI assistance. They compare the inherent image quality of QT Ultrasound to XRM and HHUS.
- Effect Size of Human Readers Improvement with AI vs. Without AI Assistance:
- Not Applicable. This device is an imaging system (QT Ultrasound Breast Scanner-1) generating images, not an AI interpretative software providing assistance to human readers. Therefore, the study design does not involve comparing human reader performance with and without AI assistance from this device. The focus is on the device's image quality compared to other imaging modalities.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
- This is an imaging device, not an algorithm for diagnosis. The "performance" in this context is primarily related to image quality and safety/technical specifications.
- The document describes non-clinical testing which can be considered "standalone" in the sense that it evaluates the device's technical performance (e.g., electrical safety, acoustic output, software verification, system verification like spatial resolution, contrast resolution, speed of sound uniformity). These tests are not "human-in-the-loop" studies but validate the device's operational characteristics.
7. The Type of Ground Truth Used
- For Image Quality Assessment (VGA studies): The "ground truth" for image quality was based on the subjective assessment of board-certified radiologists. They scored the visibility of various anatomical features using an ordinal scale (1=Excellent to 5=Poor).
- For Lesion Type in Representative Clinical Cases (demonstrative, not graded):
- Solid Benign and Solid Malignant Lesions: Ground truth was determined by histology.
- Cyst Lesions: Ground truth was determined by appearance on handheld ultrasound (HHUS).
- Breast density was determined by visual inspection on the QT Ultrasound images.
- Clinical history, previous biopsies, and other imaging studies were also used to confirm identification, size, and location of lesions (a form of clinical ground truth/consensus).
8. The Sample Size for the Training Set
- The document does not provide information on the sample size of a training set. This is expected as the submission is for a medical imaging device (hardware and software for image acquisition), not a specific AI diagnostic algorithm that requires a "training set" in the machine learning sense for its primary function. The images generated by the scanner are for human interpretation.
9. How the Ground Truth for the Training Set Was Established
- Not Applicable. As no training set data for an AI algorithm (in the context of image interpretation) is mentioned, the method for establishing its ground truth is not provided. The development and verification of the device's imaging capabilities were performed through engineering design and standard medical device verification and validation processes (e.g., non-clinical testing listed in Table 2-2).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines connecting them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 6, 2017
QT Ultrasound, LLC % John C. Klock, M.D. Chief Executive Officer and Managing Director 3 Hamilton Landing, Suite 160 NOVATO CA 94949
Re: K162372
Trade/Device Name: QT Ultrasound Breast Scanner-1 Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulatory Class: II Product Code: IYO, ITX Dated: May 31, 2017 Received: June 1, 2017
Dear Dr. Klock:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D. O'Hara
For
Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K162372
Device Name
QT Ultrasound Breast Scanner-1
Indications for Use (Describe)
The QT Ultrasound Breast Scanner-1 is for use as an ultrasonic imaging system to provide reflection-mode and transmission-mode images of a patient's breast. The device is not intended to be used as a replacement for screening mammography.
Type of Use (Select one or both, as applicable)
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) |
|---|
| ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Diagnostic Ultrasound Indications for Use
System: QT Ultrasound Breast Scanner-1
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | Other | ||||||
|---|---|---|---|---|---|---|---|---|
| General(Track 1 only) | Specific (Tracks 1 & 3) | B | M | PWD | CWD | ColorDoppler | Combined | |
| Ophthalmic | Ophthalmic | |||||||
| Fetal | ||||||||
| Abdominal | ||||||||
| Intra-operative (Specify) | ||||||||
| Intra-operative (Neuro) | ||||||||
| Laparoscopic | ||||||||
| Pediatric | ||||||||
| Fetal Imaging& Other | Small Organ (Breast) | N1 | N2 | |||||
| Neonatal Cephalic | ||||||||
| Adult Cephalic | ||||||||
| Trans-rectal | ||||||||
| Trans-vaginal | ||||||||
| Trans-urethral | ||||||||
| Trans-esoph. (non-Card.) | ||||||||
| Musculo-skeletal(Conventional) | ||||||||
| Musculo-skeletal(Superficial) | ||||||||
| Intravascular | ||||||||
| Other (Specify) | ||||||||
| Cardiac | Cardiac Adult | |||||||
| Cardiac Pediatric | ||||||||
| Intravascular (Cardiac) | ||||||||
| Trans-esoph. (Cardiac) | ||||||||
| Intra-cardiac | ||||||||
| Other (Specify) | ||||||||
| Peripheral | Peripheral vessel | |||||||
| Vessel | Other (Specify) |
N=New Indication; P = previously cleared by FDA; E = added under this appendix
1 – Reflection
2 – Transmission (Speed of Sound)
Additional Comments: QT Ultrasound Breast Scanner-1 is intended for ultrasonic breast exams
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for QT ultrasound. The logo consists of the letters "QT" in a light green color, with the word "ultrasound" in a light gray color to the right of the letters. The logo is simple and modern, and the colors are soft and calming.
510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
In accordance with 21 CFR 807.92 the following summary of information is provided:
1.0 SUBMITTER INFORMATION
| Submitted By | QT Ultrasound LLC®3 Hamilton Landing, Suite 160Novato, CA 94949 |
|---|---|
| Contact Information | John Klock, MDPhone: (415) 842-7242Fax: (415) 234-6511Email: john.klock@qtultrasound.com |
| Date of Submission | August 23, 2016 |
| 510(k) Number | K162372 |
| DEVICE INFORMATION | |
| Trade / Proprietary Name | QT Ultrasound Breast Scanner-1 |
| Common Name | System, Imaging, Pulsed Echo UltrasonicTransducer, Ultrasonic, Diagnostic |
| Classification Name andRegulation Number | 21CFR §892.1560 Ultrasonic pulsed echo imaging system21CFR §892.1570 Diagnostic ultrasonic transducer |
| Product Codes | 90-IYO, 90-ITX |
3.0 PREDICATE DEVICE
2.0
The predicate device is identified as the SoftVue System manufactured by Delphinus Medical Technologies. SoftVue received market clearance under 510(k) number K142517.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for "QT ultrasound". The logo is in a light green and gray color scheme. The letters "QT" are in green, while the word "ultrasound" is in gray.
4.0 DEVICE DESCRIPTION
The QT Ultrasound Breast Scanner-1 is an automated software-controlled ultrasonic imaging system that performs a standardized scan of the whole breast. The OT Ultrasound Breast Scanner-1 is comprised of a Patient Scanning System, Operator Console and Viewer Console.
The Patient Scanning System consists of a patient support table, scan tank, water management system, ultrasound transducer arrays and all associated image processing electronics. The scan tank is centered below a patient's breast and contains the ultrasound transducer arrays. The transducer arrays include a set of three reflection transducers that transmit pulsed ultrasound plane waves into targeted tissues using the water bath in the scan tank as coupling medium. An additional transmitter and receiver array pair collect the ultrasound energy to provide speed of sound values.
During scanning, a patient lies prone on the examination table with the breast suspended in a warm water bath maintained near skin temperature. Images are automatically acquired on a pendant breast positioned with the nipple as a point of reference. The transducer arrays rotate about a vertical axis to circle the breast in the coronal plane. The array is then translated vertically and the scanning process is repeated until the entire breast is scanned, allowing B-scan images to be constructively combined into tomographic, speed of sound and reflection ultrasound images.
The QT Ultrasound Breast Scanner-1 outputs the images to the QTviewer which allows the images to be stored until they are reviewed on a Viewer Console. Coronal, axial and sagittal images are generated for review by the radiologist. Speed of sound images may be queried by the Probe and Region of Interest (ROI) tools provided in the Viewer Console. These tools provide speed of sound values in meters/sec. to aid in diagnostic evaluation of the breast.
5.0 INTENDED USE
The OT Ultrasound Breast Scanner – 1 is for use as an ultrasonic imaging system to provide reflection-mode and transmission-mode images of a patient's breast. The device is not intended to be used as a replacement for screening mammography.
PREDICATE DEVICE COMPARISON 6.0
The OT Ultrasound Breast Scanner-1 is substantially equivalent to the Delphinus Medical Technologies SoftVue System cleared by the FDA in K142517. QT Ultrasound claims
{6}------------------------------------------------
substantial equivalence because the proposed device has an equivalent intended use, operating principles, and physical and operational specifications as compared to the predicate device. The QT Ultrasound Breast Scanner - 1 and the predicate device utilize B-mode grayscale ultrasound images to achieve their intended use. Both the OT Ultrasound Breast Scanner-1 and the predicate SoftVue device are table-top systems that have automatic scanning transducers to image breast tissue.
The specific details regarding similarities and differences between the QT Ultrasound Breast Scanner-1 and the SoftVue device have been identified and explained in Comparison Tables provided in Section 5.0 of this submission. A brief summary of the similarities and differences between the OT Ultrasound Breast Scanner-1 and the SoftVue device is included below.
Similarities
- Both systems use an automated transducer to acquire images of a patient's breast. ●
- Both systems use broadband transducers.
- o Both systems acquire and process B-mode grayscale images of a patient's breast.
- o Both systems acquire and process grayscale speed of sound images of a patient's breast.
- Both systems position the patient in a prone position lying on their examination table with the patient's breast in a pendulous position within an imaging chamber.
- Both systems position the patient's breast in a fluid environment to eliminate the need for breast compression and facilitate the transmission of ultrasound waves.
Differences
The differences between the QT Ultrasound Breast Scanner-1 and SoftVue System are listed in Table 2-1.
| Table 2-1. Differences Between the QT Ultrasound Breast Scanner-1 and SoftVue System | |
|---|---|
| TechnologicalCharacteristic | Discussion |
| Display of Speed ofSound Information | The QT Ultrasound Breast Scanner-1 provides grayscale speed ofsound images that may be queried by the Probe and/or ROI toolsavailable in the QTviewer. Both systems use this information as anaid / reference information for diagnostic evaluation of the breast. |
| 3-D volume ImageAcquisition | Although, QT image acquisition is performed in 3D and SoftVue isacquired in 2D, this technological characteristic does not raise any |
{7}------------------------------------------------
| Table 2-1. Differences Between the QT Ultrasound Breast Scanner-1 and SoftVue System | |
|---|---|
| TechnologicalCharacteristic | Discussion |
| different questions of safety and effectiveness compared to thepredicate device. | |
| Image Reconstruction | Although, QT image reconstruction is performed in 3D andSoftVue is reconstructed in 2D, this technological characteristicdoes not raise any different questions of safety and effectivenesscompared to the predicate device. |
The differences noted between the QT Ultrasound Breast Scanner-1 and the predicate device do not present any new or different questions related to safety and effectiveness.
7.0 SUMMARY OF NON-CLINICAL TESTING
The function and performance of the QT Ultrasound Breast Scanner-1 has been evaluated through non-clinical design verification and validation testing. Testing included system performance and simulated use tests. When applicable, non-clinical testing was conducted per the standards listed in Table 2-2.
| Table 2-2. Testing Performed | |
|---|---|
| Type of Testing | Tests Performed |
| Electrical SafetyAAMI ES60601-1:2005/(R)2012 AndA1:2012 Medical electrical equipment -Part 1: General requirements for basicsafety and essential performance | All applicable electrical, basic safety and essentialperformance tests.Testing was conducted by Intertek, an independenttesting laboratory, located in Menlo Park, CA. |
| Electromagnetic CompatibilityIEC 60601-1-2 Edition 3: 2007-03Medical electrical equipment - Part 1-2:General requirements for basic safety andessential performance - CollateralStandard: Electromagnetic disturbances -Requirements and tests | All applicable testing pertaining toelectromagnetic compatibility.Testing was conducted by Intertek, an independenttesting laboratory, located in Menlo Park, CA. |
| Table 2-2. Testing Performed | |
| Type of Testing | Tests Performed |
| UsabilityIEC 62366 Edition 1.1 2014-01 -Medical devices- Application of usabilityengineering to medical devicesIEC 60601-1-6 Edition 3.1 2013-10 -Medical electrical equipment Part 1-6General requirements for safety -Collateral Standard: Usability | All applicable testing pertaining to usability.Testing was conducted by Intertek, an independenttesting laboratory, located in Menlo Park, CA. |
| Acoustic OutputIEC 60601-2-37 Edition 2.0 2007Medical electrical equipment - Part 2-37:Particular requirements for the basicsafety and essential performance ofultrasonic medical diagnostic andmonitoring equipment.NEMA UD 2-2004 (R2009) AcousticOutput Measurement Standard forDiagnostic Ultrasound Equipment –Revision 3 | All applicable testing pertaining to therequirements for the safety of ultrasonic medicaldiagnostic and monitoring equipment and todemonstrate compliance with the "AcousticOutput Measurement Standard for DiagnosticUltrasound Equipment".Testing was conducted by QT Ultrasound andwitnessed by Intertek, an independent testinglaboratory, located in Menlo Park, CA. The QTUltrasound Breast Scanner-1 meets all Track 1acoustic output requirements. The results ofacoustic output testing are listed in Table 2-3below. |
| Software DevelopmentIEC 62304:2006 (First Edition) - Medicaldevice software - Software life cycleprocesses | Internal procedures for software life cyclemanagement were used for software development,verification / validation and configuration control |
| Software Verification and Validation | Software was tested at the module and systemlevels to ensure that it met the software's designand intended use requirements. All requirementswere met and no new issues of safety oreffectiveness compared to the predicate devicewere raised. |
| System Verification and Performance | System verification testing was conducted toensure that the QT Ultrasound Breast Scanner-1met design requirements. In addition, thefollowing system performance characteristics arereported:• Measurement Range / Accuracy• Spatial Resolution• Contrast Resolution / Contrast to Noise Ratio• Speed of Sound Uniformity and Accuracy |
| Table 2-2. Testing Performed | |
| Type of Testing | Tests Performed |
| All requirements were met and no new issues ofsafety or effectiveness compared to the predicatedevice were raised. | |
| BiocompatibilityISO 10993-1:2009/(R)2013 - Biologicalevaluation of medical devices - Part 1:Evaluation and testing within a riskmanagement process | Biocompatibility testing was conducted to ensurethat the patient contacting materials in the QTUltrasound Breast Scanner-1 met designrequirements. All requirements were met and nonew issues of safety or effectiveness compared tothe predicate device were raised. |
| ISO 10993-5:2009/(R)2014 - Biologicalevaluation of medical devices - Part 5:Tests for in vitro cytotoxicity | |
| ISO 10993-10:2010/(R)2014 - Biologicalevaluation of medical devices - Part 10:Tests for irritation and skin sensitization | |
| ISO 10993-11:2006/(R)2010 - Biologicalevaluation of medical devices - Part 11:Tests for systemic toxicity | |
| Cleaning ProceduresScan Tank | Microbial analysis and particulate testing of thewater in the scan tank was conducted. Allrequirements were met and no new issues of safetyor effectiveness compared to the predicate devicewere raised. |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for QT Ultrasound. The logo consists of the letters "QT" in a light green color, followed by the word "ultrasound" in a light gray color. The letters "QT" are larger than the word "ultrasound".
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for QT ultrasound. The logo is in a light green color. The letters "QT" are larger than the word "ultrasound".
Acoustic output testing per IEC 60601-2-37 Edition 2.0 2007 was conducted by QT Ultrasound LLC and witnessed by Intertek, an independent testing laboratory, located in Menlo Park, California. The QT Ultrasound Breast Scanner-1 meets all Track 1 acoustic output requirements. The results of acoustic output testing are listed in Table 2-3. A copy of the full test report is provided in Section 6.0 of this submission.
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for QT ultrasound. The logo consists of the letters "QT" in a light green color, with the word "ultrasound" in gray next to it. The font is sans-serif and the overall design is simple and modern.
| Table 2-3. Global Maximum Acoustic Output Values | ||||||
|---|---|---|---|---|---|---|
| Feature | Track 1ExposureLevel | QT Ultrasound BreastScanner-1 Level | Pass / Fail | |||
| Trans | 0 | 1 | 2 | |||
| Max. Mechanical Index (MI) | 1.9 | 0.145 | .98 | 1.07 | 1.186 | Pass |
| Max. ISPTA (mW/cm²) | 94 mW/cm2 | 0.79 | .713 | 0.995 | 1.63 | Pass |
The results of safety, performance and verification / validation testing demonstrate that the QT Ultrasound Breast Scanner-1 successfully meets the requirements of its intended use.
SUMMARY OF CLINICAL TESTING 8.0
VGA Review with QT Ultrasound vs. Xray Mammography (XRM)
QT Ultrasound LLC performed a Visual Grading Assessment (VGA) review of 22 cases from 20 subjects using four independent board-certified radiology readers. This was a paired-reader, paired-subject evaluation comparing the image quality of QT Ultrasound to X-ray mammography (XRM). The four readers independently scored the image quality of ten anatomical breast structures with XRM and OT Ultrasound during separate reading sessions, along with an overall image quality rating. The readers were provided the mammograms (CC and MLO views) and corresponding QT Ultrasound DICOM studies (Speed of Sound and Reflection images displayed in the coronal, axial, and sagittal planes) for each case.
VGA Review with OT Ultrasound vs. HandHeld Ultrasound (HHUS)
A similar VGA review of 17 cases was performed using 5 independent board-certified radiology readers. This was a paired-reader, paired-subject evaluation comparing the image quality of QT Ultrasound to HHUS. The five readers independently scored the image quality of ten anatomical breast structures with HHUS and QT Ultrasound during separate reading sessions, along with an overall image quality rating. The readers were provided the available targeted HHUS images (static DICOM images) and corresponding QT Ultrasound DICOM studies (Speed of Sound and Reflection images displayed in the coronal, axial, and sagittal planes) for each case.
The statistical analyses for both studies were performed under the direction of Nancy Obuchowski, PhD, Vice Chair of Quantitative Health Sciences at the Cleveland Clinic Foundation.
For the VGA study comparing QT Ultrasound and XRM, the results are as follows.
{11}------------------------------------------------
The median image quality score for each of the 10 anatomical features, as well as the overall image quality, was reported for each reader. The proportion of breasts where the image quality was rated better on QT Ultrasound than XRM or equivalent to XRM was reported for each feature. A 95% Confidence Interval (CI) for the proportion of breasts rated as equivalent or better image quality on QT Ultrasound was constructed for each feature using methods for clustered binary data, treating subject as the cluster. Similarly, the proportion of breasts where the image quality was rated better on QT Ultrasound than XRM was reported for each feature, along with its 95% CI.
| Table 2-4. Median Image Quality Scores by Modality, Reader, and Feature | ||||||||
|---|---|---|---|---|---|---|---|---|
| (1=excellent, 5=poor)* | ||||||||
| QT Ultrasound | XRM | |||||||
| Anatomical Feature | R1 | R2 | R3 | R4 | R1 | R2 | R3 | R4 |
| Skin (Overall) | 1.5 | 1.0 | 1.0 | 2.0 | 2.0 | 4.0 | 3.0 | 2.0 |
| Epidermis | 1.0 | 1.0 | 1.0 | 1.0 | 2.0 | 4.0 | 2.0 | 1.0 |
| Dermis | 2.0 | 2.0 | 2.0 | 2.0 | 5.0 | 5.0 | 5.0 | 5.0 |
| Hypodermis | 2.0 | 1.0 | 1.0 | 1.0 | 3.0 | 4.0 | 2.0 | 1.0 |
| Cooper's Ligament | 1.5 | 1.0 | 1.0 | 1.0 | 4.0 | 4.0 | 4.0 | 4.0 |
| Superficial Veins | 2.0 | 1.0 | 1.0 | 2.0 | 3.5 | 4.0 | 4.0 | 4.0 |
| Central Ducts EnteringNipple | 2.0 | 2.0 | 2.0 | 3.0 | 3.5 | 5.0 | 4.5 | 4.0 |
| Intermediate orperipheral Ducts | 2.0 | 2.0 | 1.0 | 2.0 | 5.0 | 5.0 | 5.0 | 5.0 |
| Terminal Duct LobularUnits | 2.0 | 2.0 | 2.0 | 2.0 | 5.0 | 5.0 | 5.0 | 5.0 |
| Pectoralis Muscle(Chest Wall) | 3.0 | 4.0 | 3.0 | 3.0 | 3.0 | 2.0 | 2.0 | 2.0 |
| Overall visualizationof breast anatomy | 2.0 | 2.0 | 1.0 | 2.0 | 4.0 | 4.0 | 4.0 | 4.0 |
- Ordinal Step Scale indicates how well the reader can see the anatomical breast structures. The definitions for the step scale are as follows:
- 1 = Excellent No limitations
- 2 = Good Minimal limitations
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image contains the logo for QT Ultrasound. The logo consists of the letters "QT" in a light green color, with the word "ultrasound" in a light gray color next to it. The font is sans-serif and the overall design is simple and modern.
3 = Sufficient - Moderate limitations with no substantial loss of information
-
4 = Restricted Relevant limitations with clear loss of information
Image /page/12/Figure/3 description: This line graph compares the readers' median score of QT Ultrasound and XRM across different features. The x-axis lists the features: Sk, Epi, Der, Hy, Coop, Ve, CenD, XLob, TerD, and Mus. The y-axis represents the readers' median score, ranging from 0 to 5. The XRM line generally has higher scores than the QT Ultrasound line, indicating a higher median score for XRM across most features. -
5 = Poor Significant loss of information
Figure 1: Readers' median image quality score by modality (QT Ultrasound in blue and XRM in red) and breast feature: skin (overall), epidermis, dermis, Cooper's ligaments, superficial veins, central ducts entering nipple, intermediate or peripheral ducts (extra-lobular ducts), terminal duct lobular units and pectoralis muscle (chest wall).
Image quality was scored on an ordinal rating scale: 1=Excellent, 2=Good, 3=Sufficient, 4=Restricted, 5=Poor.
The analysis of anatomical breast structures comparing QT Ultrasound images to XRM, demonstrated that the OT Ultrasound Breast Scanner-1 produces clinically-useful depictions of patient anatomy. The readers scored the image quality on the OT Ultrasound images as equivalent to or better than on XRM for each feature in more than 90% of breasts.
For the VGA review comparing OT Ultrasound vs. HHUS, the results are as follows:
Table 1 summarizes the median image quality scores from the results pooled over all readers, as well as the lowest median by any reader and highest median by any reader for each of the breast features. Note that readers used a score of 99 (i.e. anatomy not included on image) on both HHUS and QT Scan, but more often on HHUS. Figure 2 illustrates the readers' pooled median score for the two modalities. Except for epidermis where the median scores are equivalent, when the anatomy was visible on the image, the readers' median scores indicated superior image quality on the QT Scan.
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image contains the logo for QT Ultrasound. The logo consists of the letters "QT" in a light green color, followed by the word "ultrasound" in a light gray color. The text is simple and modern.
| HHUS | QTUS | |||||
|---|---|---|---|---|---|---|
| Median* | Min** | Max** | Median* | Min** | Max** | |
| Skin (Overall) | 3.0***(80% not scoredbecause not included onimage) | 3.0 | 3.0 | 2.0 | 1.0 | 2.0 |
| Epidermis | 1.0 | 1.0 | 2.0 | 1.0 | 1.0 | 2.0 |
| Dermis | 3.0 | 1.0 | 4.0 | 1.0 | 1.0 | 3.0 |
| Hypodermis | 2.0 | 1.0 | 3.0 | 1.0 | 1.0 | 2.0 |
| Cooper's Ligament | 3.0 | 2.0 | 4.0 | 1.0 | 1.0 | 1.0 |
| Superficial Vessels | 5.0(18.8% not scored becausenot included on image) | 2.0 | 5.0 | 1.0 | 1.0 | 2.0 |
| Central DuctsEntering Nipple | 5.0***(68.2% not scoredbecause not included onimage) | 2.0 | 5.0 | 2.0 | 1.0 | 5.0 |
| Intermediate orperipheral Ducts | 5.0(23.5% not scored becausenot included on image) | 1.0 | 5.0 | 2.0 | 1.0 | 5.0 |
| Terminal DuctLobular Units | 5.0(21.2% not scored becausenot included on image) | 4.0 | 5.0 | 2.0(3.5% not scoredbecause not included onimage) | 1.0 | 3.0 |
| Median* | Min** | Max** | Median* | Min** | Max** | |
| Pectoralis Muscle(Chest Wall) | 3.0(7.1% not scored because notincluded on image) | 3.0 | 5.0 | 2.0(8.2% not scoredbecause not included onimage) | 1.0 | 4.0 |
| Overall visualizationof breast anatomy | 3.0***(61.2% not scoredbecause not included onimage) | 3.0 | 4.0 | 2.0 | 1.0 | 3.0 |
- median over 85 observations pooled from 5 readers
** minimum and maximum median of the 5 readers
***For HHUS, overall skin, central ducts and overall visualization of breast anatomy are features that have a majority of cases not scored due to targeted HHUS.
{14}------------------------------------------------
Image /page/14/Figure/1 description: The image is a line graph comparing the median scores of two different categories, QTUS and HHUS, across various features. The x-axis represents the features, labeled as Sk, Epi, Der, Hy, Coop, Ve, CenD, XLob, TerD, and Mus. The y-axis represents the median score, ranging from 0 to 5. The QTUS line is mostly flat at a score of 1, with increases at Ve and CenD to a score of 2, while the HHUS line fluctuates between 1 and 5.
Figure 2: Readers' median image quality score by modality (QT Scan in blue and HHUS in red) and breast feature: overall skin (SK), epidermis (Epi), dermis (Hy), Cooper's ligaments (Coop), superficial vessels (Ve), central ducts entering nipple (CenD), intermediate or peripheral ducts (XLob), terminal duct lobular units(TerD), and pectoralis muscle (Mus).
Image quality was scored on an ordinal rating scale: 1=Excellent, 2=Good, 3=Sufficient, 4=Restricted, 5=Poor.
Except for epidermis and pectoralis muscle, the readers scored the image quality on the QT Scan as equivalent or better than HHUS on each feature in more than 80% of breasts. Readers scored the image quality of epidermis and muscle as better on the QT Scan in more than 70% of breasts. On this initial study utilizing historical QT Library cases, the findings from the 5 study readers suggest that QT images are at least equivalent or better in defining the anatomical components of the breast than HHUS.
Representative Clinical Cases
Sixteen clinical cases of QT Ultrasound images representative of different breast densities and lesion types, such as cancer in fatty breast, cancer in dense breast and cyst in fatty breast, were reviewed by board certified radiologist, Dr. Elaine Iuanow, M.D. and Chief Medical Officer. There was no grading performed on the Representative Clinical Cases. The cases were representative of the different types of cases seen in the clinical setting. Each case, consisting of Mammography, Hand Held Ultrasound and QT Ultrasound imaging, for each subject was reviewed.
The OT Ultrasound images were displayed in two modes. Speed of Sound (Transmission) and B-Mode (Reflection) images. Each imaging mode included coronal, axial and sagittal views. The cases included the span of clinical scenarios that the device would typically be used for.
{15}------------------------------------------------
Image /page/15/Picture/0 description: The image shows the logo for QT Ultrasound. The logo consists of the letters "QT" in a light green color, followed by the word "ultrasound" in gray. The font is sans-serif and the overall design is simple and modern.
The breast density patterns include fatty, scattered, heterogeneously dense and extremely dense. The density was determined by visual inspection on the QT Ultrasound images. The clinical history for each case was provided. Clinical information, such as previous biopsies and/or imaging studies were used to confirm the identification of lesions in the QT Ultrasound images using clock position, appearance, size and location. The speed of sound values for any identified lesions were also provided. The lesion type was determined by histology (ground truth) for solid benign and solid malignant lesions. The ground truth for cyst lesions was determined by its appearance on handheld ultrasound (HHUS).
| Case # | Case ID | Breast Density Determined Visually | Lesion Type |
|---|---|---|---|
| 1 | 901-065-L | Extremely Dense | Solid-Benign |
| 2 | 102007V | Heterogeneously Dense | * |
| 3 | 901-005-L | Fatty | Cyst |
| 4 | 102019V | Fatty | * |
| 5 | 901-020-R | Scattered | Cyst |
| 6 | 901-022-L | Scattered | Cyst |
| 7 | 901-008-R | Heterogeneously Dense | Cyst |
| 8 | 901-037-L | Scattered | Solid-Benign |
| 9 | 901-043-L | Heterogeneously Dense | Solid-Malignant |
| 10 | 901-044-L | Heterogeneously Dense | Solid-Malignant |
| 11 | 901-001-L | Scattered | Cyst |
| 12 | 901-029-R | Heterogeneously Dense | Cyst |
| 13 | 901-041-L | Fatty | Solid-Malignant |
| 14 | 901-019-R | Extremely Dense | Cyst |
| 15 | 102002V | Fatty | * |
| 16 | 901-034-R | Scattered | Solid-Benign |
Table 2-6. Case summary by breast density and lesion type
- No lesion types are provided for volunteers as they were scanned for training only.
{16}------------------------------------------------
Image /page/16/Picture/0 description: The image shows the logo for QT Ultrasound. The logo consists of the letters "QT" in a light green color, followed by the word "ultrasound" in gray. The letters "QT" are larger than the word "ultrasound", and the "Q" is stylized with a curved tail.
The clinical cases demonstrated the QT Ultrasound Breast Scanner-1 is capable of imaging different breast densities and lesion types over the span of clinical scenarios that the device would typically be used for.
9.0 CONCLUSION
The QT Ultrasound Breast Scanner-1 is substantially equivalent to the Delphinus SoftVue System with respect to intended use, principle of operation, design, performance and safety features. Both devices are intended as an adjunct to mammography and are not intended for screening purposes. The primary difference between the QT Ultrasound Breast Scanner-1 and the SoftVue System is the manner in which the qualitative speed of sound information is displayed. Although the method of display is different, the fundamental intended use is the same.
QT Ultrasound has demonstrated through verification and validation testing that the different technological characteristics between the SoftVue device and the QT Ultrasound Breast Scanner-1 do not raise new or different questions of safety and effectiveness. Therefore, it is the opinion of QT Ultrasound LLC that the QT Ultrasound Breast Scanner-1 is substantially equivalent to the predicate device identified in this submission that is currently cleared for market in the United States.
§ 892.1560 Ultrasonic pulsed echo imaging system.
(a)
Identification. An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.