K Number
K162274

Validate with FDA (Live)

Manufacturer
Date Cleared
2016-10-25

(74 days)

Product Code
Regulation Number
866.2680
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Solana® Strep Complete Assay is a rapid in vitro diagnostic test, using isothermal amplification technology (helicase-dependent amplification, HDA), for the qualitative detection of Streptococcus pyogenes (Group A B-hemolytic Streptococcus) and Streptococcus dysgalactiae (pyogenic Group C and G ß-hemolytic Streptococus) nucleic acids isolated from throat swab specimens obtained from patients with signs and symptoms of pharyngitis, such as sore throat. The Solana® Strep Complete Assay is intended for use only with the the Solana® instrument.

Device Description

The Solana Strep Complete Assay amplifies, detects and differentiates Streptococcus pyogenes DNA and Streptococcus dysgalactiae DNA present in throat swab specimens obtained from symptomatic patients.

The assay consists of two major steps: 1) specimen preparation, and 2) amplification and detection of target sequence specific to S. pyogenes (GAS) and S. dysgalactiae (C/G) using isothermal Helicase-Dependent Amplification (HDA) in the presence of target-specific fluorescence probe.

Patient specimen on a throat swab is transferred to a Lysis Tube and subjected to heattreatment at 95±°C for 5 minutes. The heat-treated sample is added to a Dilution Tube, and then transferred to two Reaction Tubes, GAS Reaction Tube and Strep C/G Reaction Tube. GAS Reaction Tube contains white lyophilized HDA reagents, dNTPs, primers and probes specific for the amplification and detection of S. pyogenes target sequence, while C/G Reaction Tube contains blue lyophilized HDA reagents, dNTPs, primers and probes specific for the amplification and detection of S. dysgalactiae target sequence. Once rehydrated with the diluted sample, the Reaction Tubes are placed in a Solana Instrument for amplification and detection of the target sequences. In Solana, the target sequences are amplified by specific primers and detected by a specific fluorescence probe included in each Reaction Tube. Two competitive process controls (PRCs) are included in the Lysis Tube to monitor sample processing, inhibitory substances in clinical samples, reagent failure or device failure for each target. PRCs are amplified by the target-specific primers and detected by a PRC specific fluorescence probe.

The target and PRC probes are labeled with a quencher on one end and a fluorophore on the other end. Upon annealing to target or PRC amplicons, the fluorescence signal increases due to physical separation of the fluorophore from the quencher. Solana measures and interprets the fluorescent signal for each Reaction Tube, using on-board method-specific algorithms. Solana then reports the test results for each Reaction Tube to the user on its display screen, and optionally prints out the results via a printer.

AI/ML Overview

Here's a summary of the acceptance criteria and study details for the Solana® Strep Complete Assay, based on the provided text:

Solana® Strep Complete Assay Acceptance Criteria and Performance Study

Acceptance Criteria CategorySpecific MetricAcceptance Criteria (Implicit)Reported Device Performance
Clinical PerformanceGAS SensitivityNot explicitly stated, but typically high for medical devices.98.8% (95% CI: 97.3% to 99.4%)
GAS SpecificityNot explicitly stated, but typically high for medical devices.98.9% (95% CI: 98.3% to 99.2%)
S. dysgalactiae SensitivityNot explicitly stated, but typically high for medical devices.100% (95% CI: 95.3% to 100%)
S. dysgalactiae SpecificityNot explicitly stated, but typically high for medical devices.99.5% (95% CI: 99.1% to 99.7%)
Analytical Sensitivity (LOD)S. pyogenes LODLowest concentration at which 95% of replicates tested positive.8.5 x 10^4 CFU/mL
S. dysgalactiae LODLowest concentration at which 95% of replicates tested positive.7.1 x 10^5 CFU/mL
ReproducibilityConsistent results across sites, operators, and instruments for positive and negative controls/samples.Demonstrated reproducible results across three sites.
Cross-ReactivityNo cross-reactivity with common interfering organisms from in silico BLAST analysis.In silico analysis showed no evidence of cross-reactivity with 61 organisms.
Negligible cross-reactivity with a panel of common throat microorganisms during wet lab testing.Klebsiella pneumoniae, Serratia marcescens, and Enterococcus faecalis each cross-reacted once out of six times tested.
InterferenceNo interference from common biological and chemical substances found in throat samples.None of the 28 tested substances interfered with the assay.
InclusivityDetection of various strains of target organisms (S. pyogenes and S. dysgalactiae).Detected 7 S. pyogenes strains and 25 S. dysgalactiae strains at LOD concentrations.
Specimen StabilityStability of collected specimens under specified storage conditions.Specimens stable for 2 days at 25°C, then up to 6 days at 2-8°C, or up to 32 days at ≤-15°C or ≤-70°C.

Study Details:

  1. Sample Size for Test Set and Data Provenance:

    • Clinical Study: 2688 fresh throat swab specimens.
      • Provenance: Prospective study conducted from February to July 2016 at four (4) external and one (1) internal laboratories across the United States.
      • Additional data for reproducibility (analytical performance): 540 specimens (including controls) tested across three sites for reproducibility.
  2. Number of Experts Used to Establish Ground Truth and Qualifications:

    • The document does not explicitly state the number or qualifications of experts used to establish ground truth for the clinical test set. It mentions that "Cultured isolates were typed by latex agglutination" and "species were determined using an FDA-cleared MALDI TOF assay." It also states, "A specimen was recorded as positive for either Streptococcus pyoqenes or Streptococcus dysgalactiae if either the culture or the FDA-cleared NAAT was positive, respectively." This suggests that laboratory personnel performing these standard diagnostic methods established the ground truth, but specific expert qualifications (e.g., radiologist with 10 years of experience) are not provided as this is a molecular diagnostic assay.
  3. Adjudication Method for the Test Set:

    • For the clinical study, the ground truth was established by a composite reference method. A specimen was considered positive if either the culture result or an FDA-cleared Nucleic Acid Amplification Test (NAAT) was positive. This acts as a form of retrospective adjudication based on multiple gold standards.
  4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not explicitly described in the provided text. The study focused on the performance of the device itself against established laboratory methods, not on how human readers' performance might improve with AI assistance.
  5. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

    • Yes, the clinical performance study evaluated the "Solana® Strep Complete Assay" (the device/algorithm) against a composite reference standard (culture + FDA-cleared NAAT). The results presented are for the device's standalone performance in identifying the target pathogens. The device has an "on-board method-specific algorithms" that measures and interprets the fluorescent signal and reports results.
  6. Type of Ground Truth Used:

    • Composite Reference Standard: For the clinical study, the ground truth was established by combining the results of:
      • Directly cultured patients' throat swabs (followed by latex agglutination and FDA-cleared MALDI TOF for speciation).
      • Another FDA-cleared Nucleic Acid Amplification Test (NAAT) performed on leftover swab transport fluid.
  7. Sample Size for the Training Set:

    • The document does not specify a sample size for a training set. The descriptions provided are for analytical validation (LOD, reproducibility, cross-reactivity, interference, inclusivity) and a clinical performance study (test set). For molecular diagnostic assays, manufacturers typically perform extensive in-house development and optimization, which would involve internal training/validation sets, but these are not usually detailed in 510(k) summaries as separate "training sets" in the context of AI/ML device submissions.
  8. How the Ground Truth for the Training Set Was Established:

    • As no specific "training set" is detailed in the document, how its ground truth was established is not explicitly mentioned. For the analytical and clinical studies, ground truth was established using standard microbiological techniques (culture, latex agglutination, MALDI TOF) and comparison to another FDA-cleared NAAT.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle with three heads. The eagle is facing to the right.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 25, 2016

QUIDEL CORPORATION RONALD LOLLAR SENIOR DIRECTOR, CLINICAL REGULATORY, SCIENTIFIC AFFAIRS 2005 EAST STATE STREET, SUITE 100 ATHENS OH 45701

Re: K162274

Trade/Device Name: Solana Strep Complete Assay Regulation Number: 21 CFR 866.2680 Regulation Name: Streptococcus spp. nucleic acid-based assay Regulatory Class: II Product Code: PGX Dated: August 10, 2016 Received: August 12, 2016

Dear Mr. Lollar:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Ribhi Shawar -S

For Uwe Scherf, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K162274

Device Name Solana® Strep Complete Assay

Indications for Use (Describe)

The Solana® Strep Complete Assay is a rapid in vitro diagnostic test, using isothermal amplification technology (helicase-dependent amplification, HDA), for the qualitative detection of Streptococcus pyogenes (Group A B-hemolytic Streptococcus) and Streptococcus dysgalactiae (pyogenic Group C and G ß-hemolytic Streptococus) nucleic acids isolated from throat swab specimens obtained from patients with signs and symptoms of pharyngitis, such as sore throat. The Solana® Strep Complete Assay is intended for use only with the the Solana® instrument.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Applicant:

Quidel Corporation 12544 High Bluff Drive, Suite 200 San Diego, California 92130 Telephone: 858-552-7910 Fax: 858-646-8045

Contact Information:

Ronald H. Lollar, Senior Director Clinical and Regulatory Affairs 2005 East State Street, Suite 100 Athens, Ohio 45701 740-589-3300 – Corporate number 740-589-3373 – Desk phone 858-552-6451—Fax Ron.Lollar@quidel.com

Date of preparation of 510(k) summary:

August 10, 2016

A. 510(k) Number:

K162274

B. Purpose for Submission:

To obtain substantial equivalence for the Solana® Strep Complete Assay when performed on the Solana® instrument

C. Measurand:

DNase B (sdaB) sequence of Streptococcus pyogenes (Group A Streptococcus) Protein G gene of Streptococcus dysgalactiae (pyogenic Group C and G β-hemolytic Streptococcus)

D. Type of Test:

Helicase-dependent amplification (HDA)

{4}------------------------------------------------

E. Applicant:

Quidel Corporation

F. Proprietary and Established Names:

Solana® Strep Complete Assay

G. Regulatory Information:

Product CodeClassificationRegulation SectionPanel
PGX – Groups A, C and G Beta-HemolyticStreptococcusNucleic AcidAmplificationClass II (Non-exempt)21 CFR 866.2680 – Streptococcus spp. Nucleic Acid-Based AssayMicrobiology (83)

H. Intended Use:

1. Intended Use(s):

The Solana® Strep Complete Assay is a rapid in vitro diagnostic test, using isothermal amplification technology (helicase-dependent amplification, HDA), for the qualitative detection and differentiation of Streptococcus pyogenes (Group A β-hemolytic Streptococcus) and Streptococcus dysgalactiae (pyogenic Group C and G β-hemolytic Streptococcus) nucleic acids isolated from throat swab specimens obtained from patients with signs and symptoms of pharyngitis, such as sore throat. The Solana® Strep Complete Assay is intended for use only with the Solana® instrument.

    1. Indication(s) for Use:
      Same as intended Use
    1. Special conditions for use statement(s):
    • For in vitro diagnostic use only
    • . For prescription use only

{5}------------------------------------------------

4. Special instrument requirements:

Solana® instrument

Device Description: l.

The Solana Strep Complete Assay amplifies, detects and differentiates Streptococcus pyogenes DNA and Streptococcus dysgalactiae DNA present in throat swab specimens obtained from symptomatic patients.

The assay consists of two major steps: 1) specimen preparation, and 2) amplification and detection of target sequence specific to S. pyogenes (GAS) and S. dysgalactiae (C/G) using isothermal Helicase-Dependent Amplification (HDA) in the presence of target-specific fluorescence probe.

Patient specimen on a throat swab is transferred to a Lysis Tube and subjected to heattreatment at 95±°C for 5 minutes. The heat-treated sample is added to a Dilution Tube, and then transferred to two Reaction Tubes, GAS Reaction Tube and Strep C/G Reaction Tube. GAS Reaction Tube contains white lyophilized HDA reagents, dNTPs, primers and probes specific for the amplification and detection of S. pyogenes target sequence, while C/G Reaction Tube contains blue lyophilized HDA reagents, dNTPs, primers and probes specific for the amplification and detection of S. dysgalactiae target sequence. Once rehydrated with the diluted sample, the Reaction Tubes are placed in a Solana Instrument for amplification and detection of the target sequences. In Solana, the target sequences are amplified by specific primers and detected by a specific fluorescence probe included in each Reaction Tube. Two competitive process controls (PRCs) are included in the Lysis Tube to monitor sample processing, inhibitory substances in clinical samples, reagent failure or device failure for each target. PRCs are amplified by the target-specific primers and detected by a PRC specific fluorescence probe.

The target and PRC probes are labeled with a quencher on one end and a fluorophore on the other end. Upon annealing to target or PRC amplicons, the fluorescence signal increases due to physical separation of the fluorophore from the quencher. Solana measures and interprets the fluorescent signal for each Reaction Tube, using on-board method-specific algorithms. Solana then reports the test results for each Reaction Tube to the user on its display screen, and optionally prints out the results via a printer.

{6}------------------------------------------------

Quidel Corporation

510(k) Summary

Materials Provided:

● 48 Tests per Kit

ComponentQuantityStorage
Strep Complete Lysis Buffer48 tubes/kit 0.5 mL2°C to 8°C
Strep Dilution Buffer48 tubes/kit 0.5 mL2°C to 8°C
GAS Reaction Tubes48 tubes/kit2°C to 8°C
Strep C/G Reaction Tubes48 tubes/kit2°C to 8°C

Materials required but not provided:

  • . External controls for Group A Streptococcus (e.g. Quidel Molecular A + G Streptococci Control Set #M111, which contains positive and negative controls, serves as an external processing and extraction control)
  • Sterile DNase-free filter-blocked or positive displacement micropipettor tips
  • Micropipettor ●
  • Stopwatch or timer
  • Scissors or a blade
  • Heat block capable of 95° C ± 2° C temperature
  • . Solana workflow tray and transfer rack
  • Solana Instrument
  • Thermometer ●

Substantial Equivalence Information: J.

    1. Predicate device name(s):
      Lyra® Direct Strep
    1. Predicate 510(k) number(s): K133883
    1. Comparison with predicate:
Similarities
ItemSolana® Strep Complete AssayLyra™ Direct Strep Assay(K133883)
Intended UseThe Solana® Strep Complete Assay is a rapid in vitro diagnostic test, usingThe Lyra Direct Strep Assay is a Real-Time PCR in vitro diagnostic test for
Similarities
ItemSolana® Strep CompleteAssayLyra™ Direct Strep Assay(K133883)
isothermal amplificationtechnology (helicase-dependent amplification,HDA), for the qualitativedetection anddifferentiation ofStreptococcus pyogenes(Group A β-hemolyticStreptococcus) andStreptococcus dysgalactiae(pyogenic Group C and G β-hemolytic Streptococcus)nucleic acids isolated fromthroat swab specimensobtained from patients withsigns and symptoms ofpharyngitis, such as sorethroat. The Solana® StrepComplete Assay is intendedfor use only with theSolana® instrument.the qualitative detectionand differentiation ofGroup A β-hemolyticStreptococcus( Streptococcus pyogenes )and pyogenic Group Cand G β-hemolyticStreptococcus nucleicacids isolated fromthroat swab specimensobtained from patientswith signs and symptomsof pharyngitis, such assore throat. The assaydoes not differentiatebetween pyogenicGroups C and G β-hemolytic Streptococcus.All negative test resultsshould be confirmed bybacterial culture,because negative resultsdo not preclude Group A,C or G Strep infectionand should not be usedas the sole basis fortreatment.The assay is intended foruse in hospital, reference,or state laboratorysettings. The device is notintended for point-of-care use.
Similarities
ItemSolana® Strep CompleteAssayLyra™ Direct Strep Assay(K133883)
Sample TypesThroat swab specimensSame
Sample Heat LysisManualSame
Detection TechniqueAutomatically detectsfluorescence afterdissociation of fluorophorefrom quencher duringamplificationSame

{7}------------------------------------------------

{8}------------------------------------------------

Differences
ItemSolana® Strep Complete AssayLyra™ Direct Strep Assay (K133883)
DNA Amplification TechnologyHelicase-dependent amplification (HDA); self-containedReal time polymerase chain reaction
InstrumentSolana™ABI 7500 Fast DX Thermocycler
Target Sequence Detected78 base pair (bp) sequenceS. pyogenes genome, resident in the DNase B ( sdaB ) gene67 base pair (bp) sequenceProtein G gene of S. dysgalactiae (pyogenic Group C and G β-hemolytic Streptococcus)GAS – 99bp product in the putative competence ( comX1.1 ) genePyo GCS/GGS – 188bp product in the tagatose-6-phosphate kinase ( lacC ) gene

{9}------------------------------------------------

ItemSolana® Strep Complete AssayLyra™ Direct Strep Assay (K133883)
Testing Time25 minutes60 -70 minutes
Clinical SensitivityGAS* Sensitivity:98.8%[95% CI: 97.3% to99.4%]Pyo GCS/GGS* S.dysgalactiae Sensitivity:100%[95% CI: 95.3% -100%]GAS* Sensitivity:96.5%[95% CI: 91.3% -98.6%]Pyo GCS/GGS* Sensitivity:95.7%[95% CI: 88.1% -98.5%]
Clinical SpecificityGAS* Specificity:98.9%[95% CI: 98.3% to99.2%]Pyo GCS/GGS* S.dysgalactiae Specificity:99.5%[95% CI: 99.1% -99.7%]GAS* Specificity:98.0%[95% CI: 97.0% -98.6%]Pyo GCS/GGS* Specificity:98.3%[95% CI: 97.4% -98.9%]

*GAS = Group A Streptococcus; Pyo GCS/GGS = Pyogenic Group C/G Streptococcus

K. Standard/Guidance Document Referenced (if applicable):

Not applicable

L. Test Principle:

Patient specimen on a throat swab is transferred to a Lysis Tube and subjected to heattreatment at 95°±2°C for 5 minutes. The heat-treated sample is added to a Dilution Tube, and then transferred to two Reaction Tubes, GAS Reaction Tube and Strep C/G Reaction Tube. GAS Reaction Tube contains white lyophilized HDA reagents, dNTPs, primers and probes specific for the amplification and detection of S. pyogenes target sequence, while C/G Reaction Tube contains blue lyophilized HDA reagents, dNTPs, primers and probes specific for the amplification and detection of S. dysgalactiae target sequence. Once rehydrated with the diluted sample, the Reaction Tubes are placed in a Solana Instrument for amplification and detection of the target sequences. In Solana, the target sequences are

{10}------------------------------------------------

amplified by specific primers and detected by a specific fluorescence probe included in each Reaction Tube. Two competitive process controls (PRCs) are included in the Lysis Tube to monitor sample processing, inhibitory substances in clinical samples, reagent failure or device failure for each target. PRCs are amplified by the target-specific primers and detected by a PRC specific fluorescence probe.

The target and PRC probes are labeled with a quencher on one end and a fluorophore on the other end. Upon annealing to target or PRC amplicons, the fluorescence signal increases due to physical separation of the fluorophore from the quencher. Solana measures and interprets the fluorescent signal for each Reaction Tube, using on-board method-specific algorithms. Solana then report the test results for each Reaction Tube to the user on its display screen, and optionally prints out the results via a printer.

M. Performance Characteristics:

1. Analytical performance:

  • a. Precision/Reproducibility:

Reproducibility

In order to confirm the reproducibility of the Solana Strep Complete Assay a blinded and randomized study panel containing both Streptococcus pyoqenes and Streptococcus dysgalactiae negative and positive samples (3x, 1x, 0.3x LOD) were tested at three (3) test sites (one in-house laboratory and two (2) clinical sites) with three (3) instruments. Each site tested a reproducibility panel and Assay Controls for five (5) days in triplicate. Testing was done by two operators at each site. Each operator ran the panel once a day using one lot of Solana Strep Complete Assay. A total of five hundred forty (540) specimens were tested (including controls). The Solana Strep Complete Assay generated reproducible results in this study.

{11}------------------------------------------------

StreptococcuspyogenesCategorySITEOverall PercentPositive95%ConfidenceInterval
Site #1Site #2Site #3
Detected:#positive /#tested%PositiveDetected:#positive /#tested%PositiveDetected:#positive /#tested%Positive
GAS High Negative13/3043%10/3033%13/3043%36/9040%27% to 47%
GAS Low Positive30/30100%30/30100%30/30100%90/90100%96% to 100%
GAS Moderate Positive30/30100%30/30100%30/30100%90/90100%96% to 100%
GAS Negative0/300%0/300%0/300%0/900%0% to 4%
GAS Positive Control30/30100%30/30100%30/30100%90/90100%96% to 100%
GAS Negative Control0/30100%0/300%0/300%0/900%0% to 4%
StreptococcusdysgalactiaeCategorySITESite #1Site #2Site #3Overall PercentPositive95%ConfidenceInterval
Detected:#positive /#tested%PositiveDetected:#positive /#tested%PositiveDetected:#positive /#tested%Positive
C/G High Negative10/3033%6/3020%5/3017%21/9023%16% to 33%
C/G Low Positive30/30100%30/30100%30/30100%90/90100%96% to 100%
C/G Moderate Positive30/30100%30/30100%30/30100%90/90100%96% to 100%
C/G Negative0/300%0/300%0/300%0/900%0% to 4%
C/G Positive Control30/30100%30/30100%30/30100%90/90100%96% to 100%
C/G Negative Control0/300%0/300%0/300%0/900%0% to 4%

The results suggest that there are no significant differences between different users using different instruments at different sites on different days.

  • b. Linearity/assay reportable range:
    Not applicable – This assay is qualitative.

  • c. Traceability, Stability, Expected values (controls, calibrators, or methods):
    Traceability:

Not applicable. This assay is qualitative.

{12}------------------------------------------------

Specimen Stability:

A study was performed to determine the stability of samples collected in a number of routinely used swab systems: nylon flocked swabs in Amies media, Rayon swab in Amies media, polyester swab in Amies media, Rayon swab in Stuart media, polyester swab in Stuart media, and Rayon in Amies gel.

Freshly grown stocks of Streptococcus pyogenes and Streptococcus dysqalactiae, of known titer, were used to spike the swabs listed above. The spiked samples were stored at 25°C ± 2°C for 2 days and then at 2° to 8°C for up to 6 more days prior to being tested in the Solana® Strep Complete Assay. A separate study was performed where the spiked samples were stored at ≤-15℃ or ≤-70℃ for a minimum of 32 days before testing.

Based on this study, specimens collected using various collection/transport systems listed above can be stored at 25°C ± 2°C for 2 days and then at 2 to 8°C for up to 6 more days before testing or at ≤-15℃ or ≤-70℃ for up to 32 days before testing.

Controls:

External Controls (Quidel Molecular A + G Streptococci Control Set #M111, which contains positive and negative controls, serves as an external processing and extraction control) were run on the Solana® Strep Complete Assay each day of testing.

The assay controls are described as follows:

  • a. The process control is used to monitor sample processing, to detect HDA inhibitory specimens and to confirm the integrity of assay reagents and Solana instrument functionality. The process control is included in the Lysis Buffer tube.
  • b. The external positive control may be treated as a patient specimen. The control should be sampled and tested as if it were a specimen and processed as described in the Assay Procedure. The external positive control is intended to monitor substantial reagent and instrument failure.

{13}------------------------------------------------

  • c. The external neqative control may be treated as a patient specimen. The control should be sampled and tested as if it were a specimen and processed as described in the Assay Procedure. The external negative control is used to detect reagent or environmental contamination (or carry-over) by Streptococcus pyogenes or Streptococcus dysgalactiae DNA or amplicon.
  • d. Detection limit:

The analytical sensitivity (limit of detection or LOD) of the Solana® Strep Complete Assay was determined using quantified (CFU/mL) cultures of two (2) Streptococcus pyogenes and two (2) Streptococcus dysqalactiae strains by serially dilution. Analytical sensitivity (LOD) is defined as the lowest concentration at which 95% of all replicates tested positive.

The LOD for the two (2) Streptococcus pyogenes strains tested were 1.5x10" CFU/mL (ATCC #19615) and 8.5x10 CFU/mL (ATCC #12344). The LOD for the two (2) Streptococcus dysgalactiae strains were 5.7x10 CFU/mL (ATCC #12394) and 7.1x10 ر CFU/mL (ATCC #10009).

Based on this data the reported LOD for Streptococcus pyogenes and Streptococcus dysgalactiae using the Solana® Strep Complete Assay is 8.5x10* CFU/mL and 7.1x105 CFU/mL, respectively.

  • e. Analytical specificity:

Cross Reactivity:

An in silico BLAST analysis of primers used in the Solana® Strep Complete Assay against sixty-one (61) potential interfering organisms (see below) did not show evidence of cross-reactivity.

Arcanobacterium sp.Human adenovirus FLactobacillus sp.1
Bacillus sp.Human adenovirus GLegionella pneumophila
Bacteroides sp.2Human coronavirus 229EMeasles virus
Bordetella sp.Human coronavirus HKU1Human Metapneumovirus

Includes L. acidophilus

2 Includes B. ovatus

{14}------------------------------------------------

Branhamella sp.Human coronavirus NL63Moraxella sp.
Burkholderia sp.Human enterovirus AMumps virus
Campylobacter sp.3Human enterovirus BMycoplasma pneumoniae
Candida sp.Human enterovirus CNeisseria sp.
Corynebacterium sp.Human enterovirus DPeptostreptococcus sp.
CytomegalovirusHuman herpesvirus 1Proteus sp.
Enterobacterio phage MS2Human herpesvirus 2Pseudomonas sp.
Enterococcus sp.Human herpesvirus 4Respiratory syncytial virus Type B
Escherichia coliHuman parainfluenza virus 1Saccharomyces cerevisiae
Fusobacterium sp.Human parainfluenza virus 2Serratia sp.
Haemophilus sp.Human parainfluenza virus 3Staphylococcus sp.
Human adenovirus AHuman parainfluenza virus 4a and 4bTreponema sp.
Human adenovirus BInfluenza virus AVeillonella sp.
Human adenovirus CInfluenza virus BYersinia sp.
Human adenovirus DInfluenza virus CPrevotella oralis4
Human adenovirus EKlebsiella sp.Parvimonas micra5
Veillonella parvula

A study was performed to evaluate the performance of the Solana® Strep Complete Assay in the presence of forty-five (45) microorganisms commonly found in throat specimens. Each potentially interfering microorganism was tested in the presence of 2 x LOD Streptococcus pyogenes and Streptococcus dysgalactiae (2 strains each) in the presence of clinically relevant levels of viruses (10 pfu/ml) and bacteria (10 °cfu/mL) or higher. All strain combinations were spiked on to swabs. The strains included in the cross-reactivity study are shown in the table below.

Acinetobacter IwoffiiStaphylococcus epidermidis MRSE
Arcanobacterium haemolyticumStenotrophomonas maltophilia
Bacillus cereusStreptococcus agalactiae
Bordetella pertussisStreptococcus anginosus

3 Includes C. rectus

4 In NCBI, Bacteroides oralis is Prevotella oralis.

్ In NCBI, Peptostreptococcus micros is Parvimonas micra.

{15}------------------------------------------------

Burkholderia cepaciaStreptococcus bovis
Corynebacterium diphtheriaStreptococcus canis
Enterococcus faecalisStreptococcus gordonii (Virdans type)
Escherichia coliStreptococcus intermedius
Fusobacterium necrophorumStreptococcus mitis
Haemophilus influenza type AStreptococcus mutans
Klebsiella pneumoniaStreptococcus oralis
Lactobacillus acidophilusStreptococcus pneumoniae
Lactococcus lactisStreptococcus salivarius
Legionella jordanisStreptococcus sanguinis
Legionella micdadeiStreptococcus suis
Legionella pneumophilaCandida albicans
Moraxella cartarrhalisAdenovirus Type 1
Neisseria gonorrhoeaeAdenovirus Type 11 (Slobitski)
Neisseria subflavaInfluenza A
Peptostreptococcus micros (aka Parvimonasmicra)Influenza B
Pseudomonas aeruginosaParainfluenza Type 4B (VR-1377)
Serratia marcescensRhinovirus Type 15 (1734)
Staphylococcus aureus MRSA

Of the 45 microorganisms tested that might be found in throat specimens, Klebsiella pneumoniae, Serratia marcescens and Enterococcus faecalis each cross-reacted once out of six times tested (triplicate testing was repeated for each cross-reactive strain) with the Solana® Strep Complete Assay.

Interference:

A study was conducted using two strains of Streptococcus pyogenes (ATCC 19615 and 12344) and Streptococcus dysgalactiae strains (ATCC 12394 and ATCC 10009) tested near LOD to evaluate the Solana® Strep Complete Assay for potential interference using a panel consisting of twenty-eight (28) common biological and chemical substances found in throat samples. Substances were introduced into the swabs at concentrations which were medically relevant. Each of the strains was tested for each substance. None of the substances tested were found to interfere with the Solana® Strep Complete Assay.

{16}------------------------------------------------

Substance NameTest ConcentrationInterference? (Y/N)
Children's Dimetapp DM Cold & CoughElixir25% v/vNo
Chloraseptic Max: Sore Throat Relief10% v/vNo
BreathSavers 3 Hour Mint-Spearmint10% w/vNo
Cepacol Sore Throat: Cherry Flavor5% w/vNo
Robitussin Cough & Cold-CF Max10% v/vNo
Ricola Mountain Herb Throat Drops-Sugar Free15% w/vNo
Human Saliva10% v/vNo
Robitussin Nighttime Cold, & Flu10% v/vNo
Crest Pro-Health Night Mint25% v/vNo
CVS Tussin CF15% v/vNo
Chloraseptic Throat Cherry lozenge10% w/vNo
Halls Cherry Mentholyptus15% w/vNo
Tic Tac Freshmints10% w/vNo
Zicam® Oral Mist0.625% v/vNo
Sucrets Complete-Vapor Cherry5% w/vNo
Acetaminophen19.5 mg/mLNo
Aspirin12.3 mg/mLNo
Ibuprofen15.6 mg/mLNo
Benadryl2.7 mg/mLNo
Crest® Complete Toothpaste5% w/vNo
Contac® Cold + Flu Caplets Night10% w/vNo
Children's Wal-Tap Elixir Cold & Allergy(Dimetap Children's Cold and Allergy)25% v/vNo
Children's Wal-Tap DM Elixir Cold &Cough25% v/vNo
Robitussin Nighttime Cough, Cold, & Flu(peak cold)10% v/vNo
Halls Mentholyptus (not cherry flavor)15% w/vNo
Listerine Cool Mint Antiseptic15% v/vNo
Whole Blood5% v/vNo
Mucin (Bovine Submaxillary Gland, type I-s)5.0 mg/mLNo

{17}------------------------------------------------

Analytical Reactivity (Inclusivity):

The inclusivity of the Solana® Strep Complete Assay was further evaluated by functional testing of organisms in addition to those strains used in the LOD study. Seven (7) strains of Streptococcus pyogenes (GAS) and twenty five (25) Streptococcus dysgalactiae (C/G) strains were tested at concentrations at a LOD of 8.5x104 CFU/mL and 7.1x10 CFU/mL, respectively.

Bacterial speciesBacterial StrainConcentration CFU/mLStrain Detected (Yes/No)
Streptococcus pyogenesATCC 123848.5x104Yes
Streptococcus pyogenesNCIMB 132858.5x104Yes
Streptococcus pyogenesCCUG 330618.5x104Yes
Streptococcus pyogenesCCUG 334098.5x104Yes
Streptococcus pyogenesCCUG 391588.5x104Yes
Streptococcus pyogenesATCC 493998.5x104Yes
Streptococcus pyogenesCCUG 535538.5x104Yes
S. dysgalactiae subspecies equisimilis group GATCC 66447.1x105Yes
S. dysgalactiae subspecies equisimilis group CATCC 95427.1x105Yes
S. dysgalactiae subspecies equisimilis group CATCC 123887.1x105Yes
S. dysgalactiae subspecies equisimilis group CATCC 356667.1x105Yes
S. dysgalactiae subspecies equisimilis group GCCUG 5027.1x105Yes
S. dysgalactiae subspecies equisimilis group CCCUG 14837.1x105Yes
S. dysgalactiae subspecies equisimilis group CCCUG 67137.1x105Yes
S. dysgalactiae subspecies equisimilis group GCCUG 156797.1x105Yes
S. dysgalactiae subspecies equisimilis group GCCUG 156807.1x105Yes
S. dysgalactiae subspecies equisimilis group GCCUG 215577.1x105Yes
S. dysgalactiae subspecies equisimilis group GCCUG 240707.1x105Yes
S. dysgalactiae subspecies equisimilis group GCCUG 261477.1x105Yes
S. dysgalactiae subspecies equisimilis group GCCUG 274777.1x105Yes
S. dysgalactiae subspecies equisimilis group CCCUG 274797.1x105Yes
S. dysgalactiae subspecies equisimilis group CCCUG 274807.1x105Yes
S. dysgalactiae subspecies equisimilis group GCCUG 274827.1x105Yes
S. dysgalactiae subspecies equisimilis group GCCUG 274837.1x105Yes
S. dysgalactiae subspecies dysgalactiae group CCCUG 276587.1x105Yes
S. dysgalactiae subspecies dysgalactiae group CCCUG 276597.1x105Yes
S. dysgalactiae subspecies dysgalactiae group CCCUG 276647.1x105Yes
S. dysgalactiae subspecies dysgalactiae group CCCUG 281157.1x105Yes
S. dysgalactiae subspecies dysgalactiae group CCCUG 281167.1x105Yes
S. dysgalactiae subspecies equisimilis group CCCUG 282387.1x105Yes
S. dysgalactiae subspecies equisimilis group GCCUG 338027.1x105Yes
S. dysgalactiae subspecies dysgalactiae group CCCUG 484777.1x105Yes
  • f. Assay cut-off:
    Not applicable.

{18}------------------------------------------------

    1. Comparison studies:
    • a. Method comparison with predicate device:

Not applicable

  • b. Matrix comparison:
    A comparison study was conducted between negative clinical matrix and the contrived negative matrices used in the analytical studies in order to validate their use. The matrix comparison study results are shown in the table below.
Contrived NegativeMatrixPooled NegativeClinical Matrix
Detected% PosDetected% Pos
StreptococcuspyogenesATCC 196151 x LOD20/20100%20/20100%
StreptococcusdysgalactiaeATCC 123941 x LOD20/20100%20/20100%

These studies demonstrate that the contrived negative matrices do not alter the performance of the device in the context of these analytical studies.

3. Clinical studies:

a. Clinical Sensitivity:

Performance characteristics of the Solana Strep Complete Assay were established during a prospective study during the winter through summer of 2016 (February to July). Two thousand six hundred eighty-eight (2688) fresh throat swab specimens were included in this study at four (4) external and one (1) internal laboratories across the United States using the same swab that was plated for the culture. A single specimen was collected per patient. Samples were collected on Polyester or Rayon Swab with liquid Amie's, Polyester Swab or Rayon swab with liquid Stuart's or nylon swab with liquid Amies.

{19}------------------------------------------------

A composite result of directly cultured patients' throat swabs combined with the culture. Cultured isolates were typed by latex agglutination. β-hemolytic isolates that were typed as Group C or G were subcultured and the species were determined using an FDA-cleared MALDI TOF assay. Swab transport fluid was also tested using another FDA cleared nucleic acid amplification test (NAAT) and cultured at a central reference laboratory. Results from culture and NAAT were used to calculate assay sensitivity and specificity. Each site cultured the swabs prior to performing the Solana Strep Complete Assay. The swab specimens were processed and tested with Solana Strep Complete Assay. The leftover swab transport media was shipped to the central location for an additional culture and NAAT testing.

A specimen was recorded as positive for either Streptococcus pyoqenes or Streptococcus dysgalactiae if either the culture or the FDA-cleared NAAT was positive, respectively.

Two thousand six hundred eighty-eight (2688) fresh throat swab specimens were tested using the algorithm described above (dual culture, FDA-cleared NAAT and Solana Strep Complete Assay). Two (2) specimens were repeatedly invalid when tested with the Solana® Strep Complete Assay (0.07%). These specimens have been removed from additional analysis. The table below details the combined results for Streptococcus pyogenes for the remaining two thousand six hundred eighty-six (2686).

Combined Clinical Sites' Results for Streptococcus pyogenes
Combined Culture and NAAT Result
Solana Strep CompleteAssayPositiveNegativeTotal
Positive47525500
Negative621802186
Total48122052686
95% CI
Sensitivity475/48198.8%97.3% to 99.4%
Specificity2180/220598.9%98.3% to 99.2%
Site 1- Streptococcus pyogenes Results
Combined Culture and NAAT Result
Solana® Strep CompleteAssayPositiveNegativeTotal
Positive90494
Negative2679681
Total92683775
95% CI
Sensitivity90/9297.8%92.4% to 99.4%
Specificity679/68399.4%98.5% to 99.8%
Site 2 - Streptococcus pyogenes Results
Combined Culture and NAAT Result
Solana® Strep CompleteAssayPositiveNegativeTotal
Positive84690
Negative1510511
Total85516601
95% CI
Sensitivity84/8598.8%93.6% to 99.8%
Specificity510/51698.8%97.5% to 99.5%
Site 3- Streptococcus pyogenes Results
Combined Culture and NAAT Result
Solana® Strep CompleteAssayPositiveNegativeTotal
Positive1003103
Negative3492495
Total103495598
95% CI
Sensitivity100/10397.191.8% to 99.0%
Specificity492/49599.4%98.2% to 99.8%
Site 4- Streptococcus pyogenes Results
Combined Culture and NAAT Result
Solana® Strep CompleteAssayPositiveNegativeTotal
Positive831295
Negative0254254
Total83266349
95% CI
Sensitivity83/83100%95.6% to 100%
Specificity254/26695.5%92.3% to 97.4%
Site 5- Streptococcus pyogenes Results
Combined Culture and NAAT Result
Solana® Strep CompleteAssayPositiveNegativeTotal
Positive1180118
Negative0245245
Total118245363
95% CI
Sensitivity118/118100%96.8% to 100%
Specificity245/245100%98.5% to 100%

{20}------------------------------------------------

Quidel Corporation

510(k) Summary

Two thousand six hundred eighty-eight (2688) fresh throat swab specimens were tested using the algorithm described above (dual culture, FDA-cleared NAAT and Solana Strep

{21}------------------------------------------------

Complete Assay). Two (2) specimens were repeatedly invalid when tested with the Solana® Strep Complete Assay (0.07%). The table below details the combined results for Streptococcus dysgalactiae for the remaining two thousand six hundred eighty-six (2686).

Combined Clinical Sites' Results for Streptococcus dysgalactiae
Combined Culture and NAAT Result
Solana® Strep CompleteAssayPositiveNegativeTotal
Positive781492
Negative025942594
Total7826082686
95% CI
Sensitivity78/78100%95.3% to 100%
Specificity2594/260899.5%99.1% to 99.7%
Site 1 - Streptococcus dysgalactiae Results
Combined Culture and NAAT Result
Solana® Strep CompleteAssayPositiveNegativeTotal
Positive32436
Negative0739739
Total32743775
95% CI
Sensitivity32/32100%89.3% to 100%
Specificity739/74399.598.6% to 99.8%
Site 2 - Streptococcus dysgalactiae Results
Combined Culture and NAAT Result
Solana® Strep CompleteAssayPositiveNegativeTotal
Positive16521
Negative0580580
Total16585601
95% CI
Sensitivity16/16100%80.6% to 100%
Specificity580/58599.1%98.0% to 99.6%
Site 3- Streptococcus dysgalactiae Results
Combined Culture and NAAT Result
Solana® Strep CompleteAssayPositiveNegativeTotal
Positive26430
Negative0568568
Total26572598
95% CI
Sensitivity26/26100%87.1% to 100%
Specificity568/57299.3%98.2% to 99.7%
Site 4- Streptococcus dysgalactiae Results
Combined Culture and NAAT Result
Solana® Strep CompleteAssayPositiveNegativeTotal
Positive202
Negative0347347
Total2347349
95% CI
Sensitivity2/2100%34.2% to 100%
Specificity347/347100%98.9% to 100%
Site 5- Streptococcus dysgalactiae Results
Combined Culture and NAAT Result
Solana®Strep CompleteAssayPositiveNegativeTotal
Positive213
Negative0360360
Total2361363
95% CI
Sensitivity2/2100%34.2% to 100%
Specificity360/36199.7%98.4% to 100%

{22}------------------------------------------------

All pyogenic Group C and Group G streptococcal species found during this assay were Group C and G S. dysgalactiae.

  • b. Clinical specificity:
    See Section 3a.

  • Other clinical supportive data (when a. and b. are not applicable): ﻥ
    Not applicable

    1. Clinical cut-off:
      Not applicable
    1. Expected values:
      The prevalence of Streptococcus pyogenes (Group A β-hemolytic Streptococcus) and Streptococcus dysgalactiae (pyogenic Group C and G β-hemolytic Streptococcus) with the Solana Strep Complete Assay has been calculated based on the age of the patient.

{23}------------------------------------------------

Two (2) specimens were invalid when tested with the Solana® Strep Complete Assay (0.07%) (in both the initial and repeat test no process control was detected) and have been removed from the Expected Values table. The table below presents the data for the remaining two thousand six hundred eighty-six (2686) specimens.

The overall incidence of Streptococcus pyogenes or Streptococcus dysgalactiae in patients tested during this study based on culture results alone was 16.0% (431/2686) for Streptococcus pyoqenes and 2.4%(65/2686) for Streptococcus dysgalactiae. The overall incidence of Streptococcus pyogenes or Streptococcus dysgalactiae in patients tested during this study based on a combination of culture results and another FDAcleared NAAT assay was 17.9% (481/2686) for Streptococcus pyogenes and 2.9% (78/2686) for Streptococcus dysqalactiae.

Combined Study Prevalence (2686)
Streptococcus pyogenesStreptococcus dysgalactiae
AgeTotal #Total PositivePrevalenceTotal #Total PositivePrevalence
≤ 2 years158117.0%15831.9%
3 to 12 years118933628.3%1189121.0%
13 to 21 years556509.0%556386.8%
≥ 22 years78310313.2%783395.0%
Overall268648117.9%2686782.9%

N. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:

Instrument: Solana™ Instrument

O. System Descriptions:

    1. Modes of Operation:
      The Solana instrument heats each reaction tube to 64°C. If present, the target sequence is amplified by S. pyogenes and/or S. dysgalactiae specific primers and detected by S. pyogenes and/or S. dysgalactiae specific fluorescence probes included in the Reaction Tube. Each probe has a florescent dye of specific wavelength. The target probes are labeled with a quencher on one end and a fluorophore on the other end. In addition, the target probes carry a ribonucleic acid. Upon annealing to the respective amplicons, the fluorescence probes are cleaved by RNaseH2 and the fluorescence signal increases due

{24}------------------------------------------------

to physical separation of fluorophore from quencher. The Solana instrument measures and interprets the fluorescent signal, using on-board method-specific algorithms. Solana instrument will then report the test results to the user on its display screen, and it can print out the results via a printer.

2. Software:

FDA has reviewed applicant's Hazard Analysis and software development processes for this line of product types:

Yes __________________________________________________________________________________________________________________________________________________________________________ No ___________________________________________________________________________________________________________________________________________________________________________

P. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10, 21 CFR 801.109, and the special controls.

§ 866.2680

Streptococcus spp. nucleic acid-based assay.(a)
Identification. AStreptococcus spp. nucleic acid-based assay is a qualitative in vitro diagnostic device intended to simultaneously detect and identify variousStreptococcus spp. nucleic acids extracted directly from clinical specimens. The device detects specific nucleic acid sequences for organism identification. The identification aids in the diagnosis of diseases caused by bacteria belonging to the genusStreptococcus and provides epidemiological information on these diseases. Pathogenic streptococci are associated with infections, such as sore throat, impetigo (an infection characterized by small pustules on the skin), urinary tract infections, rheumatic fever, and kidney disease.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include detailed device description documentation, including the device components, ancillary reagents required but not provided, and a detailed explanation of the methodology including primer/probe sequence, design, and rationale for sequence selection.
(2) Premarket notification submissions must include detailed documentation from the following analytical and clinical performance studies: Analytical sensitivity (Limit of Detection), reactivity, inclusivity, precision, reproducibility, interference, cross reactivity, carry-over, and cross contamination.
(3) Premarket notification submissions must include detailed documentation from a clinical study. The study, performed on a study population consistent with the intended use population, must compare the device performance to results obtained from well-accepted reference methods.
(4) Premarket notification submissions must include detailed documentation for device software, including, but not limited to, software applications and hardware-based devices that incorporate software.
(5) Premarket notification submissions must include database implementation methodology, construction parameters, and quality assurance protocols, as appropriate.
(6) The device labeling must include limitations regarding the need for culture confirmation of negative specimens, as appropriate.
(7) A detailed explanation of the interpretation of results and acceptance criteria must be included in the device's 21 CFR 809.10(b)(9) compliant labeling.
(8) Premarket notification submissions must include details on an end user device training program that will be offered while marketing the device, as appropriate.