Barrier Supreme Sterilization Wrapper

K162162 · Standard Textile Co., Inc. · FRG · Nov 4, 2016 · General Hospital

Device Facts

Record IDK162162
Device NameBarrier Supreme Sterilization Wrapper
ApplicantStandard Textile Co., Inc.
Product CodeFRG · General Hospital
Decision DateNov 4, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6850
Device ClassClass 2

Indications for Use

Barrier Supreme Sterilization Wrappers are intended to be used to enclose another medical device that is to be sterilized by a health care provider. They are intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used. Barrier Supreme Sterilization Wrappers will function as a sterilization wrap when processed according to instructions. Barrier Supreme Sterilization Wrappers are reusable through 75 wash, dry, and sterilization cycles. They are manufactured and distributed as non-sterilization wraps that are intended to be sterilized and processed by health care facilities and/or contract sterilization/laundry companies.

Device Story

Barrier Supreme Sterilization Wrappers are reusable textile wraps designed to enclose medical devices during steam sterilization and maintain sterility post-processing. The device consists of a woven 55/45 cotton/polyester blend fabric treated with a fluoropolymer finish. Healthcare staff use standard folding techniques (oblong or envelope) to wrap instrument packs, securing them with autoclave tape or elastomer closures. The wraps are compatible with prevacuum steam sterilization cycles (4 minutes at 270°F). They are designed for up to 75 wash, dry, and sterilization cycles. By providing a sterile barrier, the wrappers protect instruments from contamination, ensuring aseptic presentation for clinical use.

Clinical Evidence

Bench testing only. Performance validated through 75 wash, dry, and sterilization cycles. Testing included whole package challenge and 30-day shelf life study, demonstrating effective sterile barrier maintenance. Physical testing included strength (ASTM D-5034 >50 psi), linting (<5.0 log IPM), and cytotoxicity (MEM Elution, non-toxic). Sterilization efficacy confirmed via half-cycle exposure with no growth of indicator organisms. Colorfastness tested per AATCC 61-1993(4A).

Technological Characteristics

Woven 55/45 cotton/polyester blend fabric with fluoropolymer finish. Dimensions vary by model. Compatible with prevacuum steam sterilization (270°F, 4-minute exposure). Reusable for 75 cycles. Complies with ASTM D-5034 for strength and AATCC 61-1993 for colorfastness. Non-toxic per MEM Elution testing.

Indications for Use

Indicated for use by health care providers to enclose medical devices for sterilization and to maintain sterility of the enclosed devices until use. Suitable for use in health care facilities or by contract sterilization/laundry companies.

Regulatory Classification

Identification

A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's symbol, which is a stylized representation of a human figure. The symbol is composed of three overlapping profiles facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 November 4, 2016 Standard Textile Co., Inc. % Jonathan Kahan Partner Hogan Lovells US LLP 555 Thirteenth Street, NW Washington, District of Columbia 20004-1109 Re: K162162 Trade/Device Name: Barrier Supreme Sterilization Wrapper Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: FRG Dated: September 27, 2016 Received: September 27, 2016 Dear Jonathan Kahan: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in {1}------------------------------------------------ the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely. Michael J. Ryan -S for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### 510(k) Number (if known) K162162 #### Device Name Barrier Supreme Sterilization Wrappers (for models, colors, and sizes, see Table 1 below) #### Indications for Use (Describe) Barrier Sterilization Wrappers are intended to be used to enclose another medical device that is to be sterlized by a health care provider. They are intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used. Barrier Supreme Sterilization Wrappers will function wrap when processed according to instructions. Barrier Supreme Sterilization Wrappers are reusable through 75 wash, dry, and sterilization cycles. They are manufactured and distributed as non-sterily sterilization wraps that are intended to be sterlized and processed by health care facilities and/or contract sterilization/laundry companies. #### Pack Assembly: Pack assembly should follow established folding, assembly and wrapping procedures as defined by the clinical staff for aseptic presentation of the pack and its contents. The oblog and envelope folds have been validated for use on textile packs. Closure can include autoclave tape and elastomer closures. #### Sterilization: A prevacuum steam sterilization cycle (4 minute exposure at 270°F) has been validated for use on textile packs. The maximum pack size validated for use was 0.58 cubic feet with a pack weight of 23.2 pounds – pack density of 40.0 lbs./cu.ft. The maximum chamber loading configuration included 96 pounds of packs in a 9.1 cubic foot chamber – load density of 10.6 lbs./cu.ft. | | | | Table 1: Barrier Supreme Sterilization Wrappers Color and Model #'s | |--|--|------|---------------------------------------------------------------------| | | | A !! | | | Sizes | Misty | Jade | Ceil | |---------|----------|----------|----------| | 12"x12" | --- | -- | -- | | 18"x18" | 21621502 | 21631502 | 21651502 | | 23"x23" | 21622002 | 21632002 | 21652002 | | 30"x30" | 21622802 | 21632802 | 21652802 | | 35"x35" | 21623002 | 21633002 | 21653002 | | 45"x45" | 21625002 | 21635002 | 21655002 | | 54"x54" | 21626002 | 21636002 | 21656002 | | 54"x72" | 21628202 | 21638202 | 21658202 | | 60"x60" | 21628502 | -- | -- | Type of Use (Select one or both, as applicable) □ Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff #### PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number. {3}------------------------------------------------ ## 510K SUMMARY BARRIER SUPREME STERILIZATION WRAPPERS ## Contact Information | Standard Textile Co., Inc. | Contact Person: Brad Bushman | |----------------------------|---------------------------------------------------| | One Knollcrest Drive | Phone (513) 761-9255 Ext 2455; Fax (513) 679-8389 | | Cincinnati, Ohio 45237 | Summary Prepared on 10/27/2016 | ## Device Information | Device Name: | Barrier Supreme Sterilization Wrappers | |----------------------|--------------------------------------------------------------------| | Class: | Class II | | Common/Usual Name: | Non-sterile Sterilization Wrappers | | Classification Name: | Sterilization Wrap<br>Regulation 21 CFR 880.6850, Product Code FRG | # Device Description Barrier Supreme Sterilization Wrappers are made from Barrier Supreme fabric. This fabric uses intimately blended cotton and polyester yarns, is woven into fabric, dyed and treated with a fluoropolymer finish. The fabric comes in three colors - misty, ceil and jade. The dyes used to achieve the three shades include Teresil Navy, Novasol Blue, VatGreen1, Vat Yellow1, and Vat Red31. | Barrier Supreme Wrapper Product Line Table | | |--------------------------------------------|------------------------------------------| | Item | Description | | 21621502 | Barrier Supreme Wrapper, Misty 18x18 | | 21622002 | Barrier Supreme Wrapper, Misty 23X23 | | 21622802 | Barrier Supreme Wrapper, Misty 30x30 | | 21623002 | Barrier Supreme Wrapper, Misty 35x35 | | 21625002 | Barrier Supreme Wrapper, Misty 45x45 | | 21626002 | Barrier Supreme Wrapper, Misty 54x54 | | 21628202 | Barrier Supreme Wrapper, Misty 54x72 | | 21628502 | Barrier Supreme Wrapper, Misty 60x60 | | 21631502 | Barrier Supreme Wrapper, Jade 18x18 | | 21632002 | Barrier Supreme Wrapper, Jade 23x23 | | 21632802 | Barrier Supreme Wrapper, Jade 30x30 | | 21633002 | Barrier Supreme Wrapper, Jade 35x35 | | 21635002 | Barrier Supreme Wrapper, Jade 45x45 | | 21636002 | Barrier Supreme Wrapper, Jade 54x54 | | 21638202 | Barrier Supreme Wrapper, Jade 57x72 | | 21651502 | Barrier Supreme Wrapper, Ceil Blue 18x18 | | 21652002 | Barrier Supreme Wrapper, Ceil Blue 23x23 | | 21652802 | Barrier Supreme Wrapper, Ceil Blue 30x30 | | 21653002 | Barrier Supreme Wrapper, Ceil Blue 35x35 | | 21655002 | Barrier Supreme Wrapper, Ceil Blue 45x45 | | 21656002 | Barrier Supreme Wrapper, Ceil Blue 54x54 | | 21658202 | Barrier Supreme Wrapper, Ceil Blue 54x72 | {4}------------------------------------------------ # Intended Use/Indications for Use Barrier Supreme Sterilization Wrappers are intended to be used to enclose another medical device that is to be sterilized by a health care provider. They are intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used. Barrier Supreme Sterilization Wrappers will function as a sterilization wrap when processed according to instructions. Barrier Supreme Sterilization Wrappers are reusable through 75 wash, dry, and sterilization cycles. They are manufactured and distributed as non-sterilization wraps that are intended to be sterilized and processed by health care facilities and/or contract sterilization/laundry companies. ## Pack Assembly: Pack assembly should follow established folding, assembly and wrapping procedures as defined by the clinical staff for aseptic presentation of the pack and its contents. The oblong and envelope folds have been validated for use on textile packs. Closure can include autoclave tape and elastomer closures. ## Sterilization: A prevacuum steam sterilization cycle (4 minute exposure at 270°F) has been validated for use on textile packs. The maximum pack size validated for use was 0.58 cubic feet with a pack weight of 23.2 pounds - pack density of 40.0 lbs./cu.ft. The maximum chamber loading configuration included 96 pounds of packs in a 9.1 cubic foot chamber - load density of 10.6 lbs./cu.ft. # Substantial Equivalence Comparison Barrier Supreme Sterilization Wrappers have been shown to be "substantially equivalent" to WrapPel "T" sterilization wraps K923408. All fabrics used are woven, dyed and treated with a fluoropolymer finish. Fabrics are spread, cut and sewn into the similar if not the exact wrapper size. All products can use color coded thread around the edges to provide easy size identification. Performance through 75 processings demonstrate that Barrier Supreme Sterilization Wrappers and WrapPel "T" Sterilization Wrappers have the strength for reuse. Both subject and predicate wrappers can be steam sterilized. In practice Barrier Supreme Sterilization Wrappers differ from WrapPel "T" Sterilization Wrappers in that it is a 55/45 blended cotton/polyester fabric versus WrapPel "T" Sterilization Wrappers which are 100% polyester. # Substantial Equivalence Conclusion Statement Based on intended use, bench testing and technological characteristics, Barrier Supreme Sterilization Wrappers are substantially equivalent to the predicate device. {5}------------------------------------------------ ## Performance Barrier Supreme Sterilization Wrappers have successfully completed the non-clinical performance tests listed below. - a. Whole Package Challenge and 30 Day Shelf Life Study Barrier Supreme Sterilization Wrappers demonstrated their ability in these tests to be an effective sterile barrier after 75 cycles of soiling, washing, drying and sterilization when used as a sequential double wrap for textile packs. - b. Strength through 75 wash, dry and sterilization cycles ASTM #D-5034-2013 >50 psi - c. Linting < 5.0 log IPM. Resists the rigors of linting associated with flexing and abrasion. - d. Cytotoxicity MEM Elution (MG023) Non-Toxic - e. Sterilization Sterilization efficiency/penetration for the Barrier Supreme Sterilization Wrap was demonstrated with a half-cycle exposure condition showing no growth of the indicator organism in worst case challenge load configurations. A prevacuum steam sterilization cycle (4 minute exposure at 270°F) has been validated for use on textile packs. The maximum pack size validated for use was 0.58 cubic feet with a pack weight of 23.2 pounds - pack density of 40.0 lbs./cu.ft. The maximum chamber loading configuration included 96 pounds of packs in a 9.1 cubic foot chamber - load density of 10.6 lbs./cu.ft. - f. Use Life - 75 processing (wash, dry and sterilization). - Colorfastness to Commercial Laundering AATCC #61-1993(4A) >3.5. q. The non-clinical tests demonstrate that Barrier Supreme Sterilization Wrappers are substantially equivalent to the predicate device. {6}------------------------------------------------ | Substantial Equivalence Table | | | |---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------|----------------------------------------| | Barrier Supreme Sterilization Wrappers | | | | vs.<br>WrapPel "T" Sterilization Wrappers (K923408) | | | | | Barrier Supreme Sterilization<br>Wrappers | WrapPel "T" Sterilization<br>Wrappers | | <b>Manufacturing</b> | | | | Quality System | 21 CFR 820 | 21 CFR 820 | | Fabric | Woven (1:1) | Woven (1:1) | | Finishing | Vat Dyed | Disperse Dyed &<br>Flurocarbon Treated | | Cut-n-Sew | Yes | Yes | | Finished Dimensions | Similar/Same | Similar/Same | | <b>Performance</b> - Non-clinical tests were used to evaluate the physical and performance<br>characteristics of Barrier Supreme Sterilization Wrappers. Performance through 75<br>processings demonstrate that Barrier Supreme Sterilization Wrappers and WrapPel “T”<br>Sterilization Wrappers have the strength for reuse. Both wrappers can be steam sterilized are<br>colorfast, low lint and non-toxic. | | | | Strength | >20 psi | >20 psi | | Colorfastness | Yes | Yes | | Linting | Completed | Completed | | Toxicity | Non-Toxic | Non-Toxic | | Use Life | 75 wash, dry, & sterilizations | 75 wash, dry, & sterilizations | | <b>Sterilization – Prevacuum Steam Cycles</b> | | | | Cycle – 4 minute<br>exposure at 270°F. | Yes | Yes | | Intended Use, i.e., traditional wrapping practices using oblong and envelope folds | | | | Fabric Packs | Yes | Yes | | Substantially Equivalent Conclusion Statement - Based on intended use, bench testing,<br>and technological characteristics, the subject device is substantially equivalent to the<br>predicate device. | | | Г
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...