K Number
K162153

Validate with FDA (Live)

Date Cleared
2017-07-14

(346 days)

Product Code
Regulation Number
888.3040
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The proposed indications for use for the Integra® CAPTURE™ Screw System are as follows:
"The CAPTURE™ Screw System implants (screws) are intended for fixation of fractures, and osteotomies of the small bones in the hand and foot. The implants are intended for single use only.

The proposed indications for use for the Integra® Ti6® Internal Fixation System are as follows: "The Ti6@ Internal Fixation System implants (screws) are intended for fixation of fractures, non-unions, arthrodeses and osteotomies of the small bones in the hand and foot. The implants are intended for single use only.

Device Description

The Integra® CAPTURETM Screw System and Integra® Ti6® Internal Fixation System consists of bone screws of various designs and sizes intended to fixate bones in cases of fractures, osteotomies, or fusions. The screws are self- drilling and self-tapping, and are manufactured from Ti-6Al-4V titanium alloy. This submission presents new cannulated (Digital Fusion) and non-cannulated (QuickSnap) screw designs that are line extensions to the current Integra® CAPTURE™ Screw System and Integra® Ti6® Internal Fixation System.

AI/ML Overview

This document is a 510(k) premarket notification for the Integra® CAPTURE™ Screw System & Integra® Ti6® Internal Fixation System. It describes the device, its intended use, and the basis for substantial equivalence to predicate devices. Crucially, as a 510(k) submission, it focuses on demonstrating that the new device is as safe and effective as a legally marketed predicate device, rather than proving novel effectiveness through clinical trials.

The document states:
"Clinical Performance Data: Clinical performance data was not necessary to demonstrate substantial equivalence."

Therefore, the specific information requested about acceptance criteria for device performance studies, sample sizes, expert ground truth, adjudication methods, MRMC studies, standalone algorithm performance, and training set details will not be available in this document as it explicitly states clinical performance data was not required or provided.

However, I can provide the non-clinical performance data and its findings, which were used to establish substantial equivalence.

1. Table of Acceptance Criteria and Reported Device Performance

TestAcceptance Criteria (Implied by Substantial Equivalence to Predicate)Reported Device Performance
Torsional StrengthSubstantially equivalent or better than predicate devicePerformed substantially equivalent or better than predicate device
Driving TorqueSubstantially equivalent or better than predicate devicePerformed substantially equivalent or better than predicate device
Axial Pullout StrengthSubstantially equivalent or better than predicate devicePerformed substantially equivalent or better than predicate device

2. Sample size used for the test set and the data provenance:

  • Test Set Sample Size: Not specified in the document for the non-clinical tests.
  • Data Provenance: The tests were conducted "per ASTM F543," which is a standard for metallic bone screws. This indicates laboratory testing, not human patient data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable. The ground truth for mechanical testing is based on established engineering standards (ASTM F543) and physical measurements, not expert human interpretation.

4. Adjudication method for the test set:

  • Not applicable as the testing involves direct physical measurements against engineering standards rather than subjective interpretations requiring adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No. A MRMC comparative effectiveness study was not done. This device is a bone fixation screw system, not an AI or imaging diagnostic device that would involve human readers or AI assistance in interpretation.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

  • No. This device is a physical medical device (screws), not an algorithm or software. Standalone performance as an algorithm is not applicable.

7. The type of ground truth used:

  • Non-clinical performance data: Ground truth was established by adherence to ASTM F543 standards for mechanical properties (Torsional Strength, Driving Torque, Axial Pullout Strength) and comparison to predicate devices.

8. The sample size for the training set:

  • Not applicable. This device is a physical medical implant, not an AI or machine learning model that requires a training set. The "training" for the device's design and manufacturing would come from engineering principles and standards.

9. How the ground truth for the training set was established:

  • Not applicable. See point 8.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, with three human profiles facing to the right.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 14, 2017

Ascension Orthopedics Blesson Abraham Senior Regulatory Affairs Specialist 8700 Cameron Road Austin, Texas 78754

Re: K162153

Trade/Device Name: Integra® CAPTURE™ Screw System & Integra® Ti6® Internal Fixation System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: HWC Dated: June 14, 2017 Received: June 15, 2017

Dear Blesson Abraham:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Vincent J. Devlin -S

for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K162153

Device Name

Integra® CAPTURETM Screw System & Integra® Ti6® Internal Fixation System

Indications for Use (Describe)

The proposed indications for use for the Integra® CAPTURE™ Screw System are as follows:

"The CAPTURE™ Screw System implants (screws) are intended for fixation of fractures, and osteotomies of the small bones in the hand and foot. The implants are intended for single use only.

The proposed indications for use for the Integra® Ti6® Internal Fixation System are as follows: "The Ti6@ Internal Fixation System implants (screws) are intended for fixation of fractures, non-unions, arthrodeses and osteotomies of the small bones in the hand and foot. The implants are intended for single use only.

Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."


{3}------------------------------------------------

K162153

Page 1 of 2

510(k) Summary

Sponsor:Ascension Orthopedics8700 Cameron RoadAustin, TX 78754-3832, USA
Contact Person:Blesson AbrahamSenior Regulatory Affairs Specialist512-852-3942512-368-1423
Date:August 1, 2016
Establishment Registration Number:1651501
Trade Name:Integra® CAPTURE™ Screw System &Integra® Ti6® Internal Fixation System
Common Name:Bone Screw
Product Code:HWC – Screw, Fixation, Bone
Classification:21 CFR §888.3040 – Smooth or threaded metallic bonefixation fastener
Classification Panel:Orthopedic
Predicate Devices:Smith & Nephew, Inc. VLP Foot Plating, Screw System andAccessories, K090675, cleared on June 4, 2009.
Wright Medical Technology, Inc. WRIGHTTM CompressionScrew, K082320, cleared on November 5, 2008.
Koby Surgical. Koby Surgical Internal Fixation System,K060026, cleared on January 26, 2006.
Device Description:The Integra® CAPTURETM Screw System and Integra® Ti6®Internal Fixation System consists of bone screws of variousdesigns and sizes intended to fixate bones in cases offractures, osteotomies, or fusions. The screws are self- drillingand self-tapping, and are manufactured from Ti-6Al-4Vtitanium alloy. This submission presents new cannulated(Digital Fusion) and non-cannulated (QuickSnap) screwdesigns that are line extensions to the current Integra®CAPTURE™ Screw System and Integra® Ti6® Internal FixationSystem.
Intended Use:The proposed indications for use for the Integra®CAPTURE™ Screw System are as follows:"The CAPTURE™ Screw System implants (screws) areintended for fixation of fractures, non-unions, arthrodeses, andosteotomies of the small bones in the hand and foot. Theimplants are intended for single use only.

{4}------------------------------------------------

Image /page/4/Picture/3 description: The image shows the logo for INTEGRA. The word "INTEGRA" is in large, gray, sans-serif font. Below the word "INTEGRA" is the phrase "LIMIT UNCERTAINTY" in a smaller, green, sans-serif font. To the right of the word "INTEGRA" is a green graphic consisting of several squares of different sizes.

The proposed indications for use for the Integra® Ti6® Internal Fixation System are as follows:

"The Ti6" Internal Fixation System implants (screws) are intended for fixation of fractures, non-unions, arthrodeses and osteotomies of the small bones in the hand and foot. The implants are intended for single use only.

The additional Integra® CAPTURE™ Screw System and Basis of Substantial Equivalence: Integra® Ti6® Internal Fixation System QuickSnap and Digital Fusion screws have the same intended use and fundamental scientific technology as the predicate devices. Sizing differences between the subject and predicate devices have been assessed and do not present new issues of safety or efficacy. Anodization surface treatment of the subject devices does not present new issues to safety or efficacy. Therefore, the proposed Integra® CAPTURE™ Screw System and Integra® Ti6® Internal Fixation System QuickSnap and Digital Fusion screws, anodization, are substantially equivalent to the predicate devices.

Non-Clinical Performance Data: The following non-clinical performance analysis was conducted per ASTM F543 in order to establish substantial equivalence.

    1. Torsional Strength
    1. Driving Torque
  • Axial Pullout Strength 3.

The Digital Fusion and OuickSnap performed substantially equivalent or better when compared to the predicate device.

Clinical Performance Data: Clinical performance data was not necessary to demonstrate substantial equivalence.

§ 888.3040 Smooth or threaded metallic bone fixation fastener.

(a)
Identification. A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.