K Number
K162046

Validate with FDA (Live)

Date Cleared
2017-04-11

(260 days)

Product Code
Regulation Number
872.4600
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The MINNE TIES™ MMF Suture System is an adjustable flexible plastic band that wraps between and around a tooth to create an anchorage point for maxilla-mandibular fixation and immobilization. MINNE TIES™ MMF Suture System is suitable for:

  • Pre-operative fixation
  • Per-operative fixation
  • Short-term (up to 3 weeks) fixation for minimally displaced fractures
  • Splintage post jaw dislocation
Device Description

The MINNE TIES™ MMF Suture System contains:

  • Four (4) 1.0mm sized sutures
  • Seven (7) 0.7mm sized sutures
  • One (1) 0.5mm Sized suture
    and are used to establish maxilla-mandibular fixation (MMF) and immobilization by applying independently balanced series of sutures bilaterally between and/or around the teeth in the interdental space (i.e. triangular apical embrasure). The sutures provide a stabilizing occlusive force allowing for short term (up to 3 weeks) fixation for minimally displaced fractures for pre-operative and per-operative fixation and splintage post jaw dislocation. An Introducer is used to advance sutures between and/or around the teeth. A cheek retractor is included with the system to organize the ties during the procedure.
    The MINNE TIES™ MMF Sutures are made of polypropylene, polyester and stainless steel. The cheek retractor is made of polypropylene.
AI/ML Overview

The provided text describes a 510(k) premarket notification for the "MINNE TIES™ MMF Suture System". This document indicates the device's substantial equivalence to a legally marketed predicate device rather than a ground-up study proving its performance against acceptance criteria for a novel AI/ML device.

Therefore, the information required to answer the prompt for an AI/ML device (e.g., acceptance criteria for AI performance, sample sizes for AI training/test sets, expert ground truth establishment, MRMC studies, standalone performance, etc.) is not present in the provided document.

The document focuses on:

  • Device Description: The components and general use of the MINNE TIES™ MMF Suture System.
  • Intended Use/Indications for Use: What the device is intended for (maxilla-mandibular fixation).
  • Predicate Device Comparison: A detailed comparison to a previously cleared device (Rapid IMF™ - K030605) to demonstrate substantial equivalence.
  • Non-Clinical Testing: Physical and material tests (tensile strength, fatigue, biocompatibility, sterilization, packaging, shelf life, corrosion resistance, system handling/cadaver test) to show that the new device performs similarly and safely to the predicate.
  • Clinical Testing: Explicitly states that clinical data was not required for this device, which further indicates the absence of human-in-the-loop performance studies or deep learning-based performance evaluations.

In summary, this document is for a medical device that does not involve artificial intelligence or machine learning, and thus the specific metrics and study methodologies requested in the prompt related to AI/ML device acceptance criteria are not applicable or provided.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized representation of three human profiles facing right, arranged in a stacked formation. The profiles are black and have a flowing, abstract design. The seal is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 11, 2017

Summit Medical, Inc. Nicole Dove QA/RA Manager 815 Northwest Parkway, Suite 100 St. Paul. Minnesota 55121

Re: K162046

Trade/Device Name: MINNE TIES MMF Suture System Regulation Number: 21 CFR 872.4600 Regulation Name: Intraoral Ligature And Wire Lock Regulatory Class: Class II Product Code: DYX Dated: March 16, 2017 Received: March 17, 2017

Dear Nicole Dove:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Michael J. Ryan -S

for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K162046

Device Name MINNE TIES MMF Suture System

Indications for Use (Describe)

The MINNE TIES MMF Suture System is an adjustable flexible plastic band that wraps between and around a tooth to create an anchorage point for maxilla-mandibular fixation and immobilization. MINNE TIES MMF System is suitable for:

  • Pre-operative fixation

  • Per-operative fixation

  • Short-term (up to 3 weeks) fixation for minimally displaced fractures

  • Splintage post jaw dislocation
    Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Summit Medical, which includes the word "SUMMIT" in large, stylized letters above the word "medical" in smaller letters. Above the word "SUMMIT" are three dark blue diamonds. Below the logo is the text "510(k) Summary" followed by "K162046/S001".

Following is a summary of 510(k) safety and effectiveness information in accordance with 21 CFR 807.92

Date Prepared:March 16, 2017
Applicant:Summit Medical, Inc.815 Northwest Parkway, Suite 100St. Paul, MN 55121 USA
Official Correspondent:Nicole DoveQA/RA ManagerTel: (651) 789-3921 / Fax: (651) 789-3961ndove@summitmedicalusa.com
Subject Device
Trade/Device Name:MINNE TIES™ MMF Suture System
Common or Usual Name:Intraoral ligature and wire lock
Regulation Number:21 CFR 872.4600
Regulation Name:Intraoral Ligature and Wire Lock
Regulatory Class:Class II
Product Code:DYX
Indications for UseThe MINNE TIES™ MMF Suture System is an adjustable flexible plasticband that wraps between and around a tooth to create an anchorage pointfor maxilla-mandibular fixation and immobilization. MINNE TIESTMMMF Suture System is suitable for:- Pre-operative fixation- Per-operative fixation- Short-term (up to 3 weeks) fixation for minimally displacedfractures- Splintage post jaw dislocation
Device DescriptionThe MINNE TIESTM MMF Suture System contains:- Four (4) 1.0mm sized sutures- Seven (7) 0.7mm sized sutures- One (1) 0.5mm Sized sutureand are used to establish maxilla-mandibular fixation (MMF) andimmobilization by applying independently balanced series of suturesbilaterally between and/or around the teeth in the interdental space (i.e.triangular apical embrasure). The sutures provide a stabilizing occlusiveforce allowing for short term (up to 3 weeks) fixation for minimallydisplaced fractures for pre-operative and per-operative fixation and
splintage post jaw dislocation. An Introducer is used to advance suturesbetween and/or around the teeth. A cheek retractor is included with thesystem to organize the ties during the procedure.
The MINNE TIES™ MMF Sutures are made of polypropylene, polyesterand stainless steel. The cheek retractor is made of polypropylene.

510(k) Premarket Notification – MINNE TIES™ MMF Suture System Summit Medical, Inc.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Summit Medical. The logo features the word "SUMMIT" in a stylized font, with the word "medical" in a smaller font underneath. Above the word "SUMMIT" are three blue diamonds arranged horizontally. The logo is simple and professional, and the blue color gives it a sense of trust and reliability.

Predicate Device
Trade/Device Name/510(k)Rapid IMF™ - K030605
Comparison of SubjectMINNE TIES™ MME
Predicate Device:
Comparison of SubjectMINNE TIES™ MMFPredicate Device:
Device to Predicate DeviceSuture System – K162046/S001Rapid IMF™ – K030605
(under review)
Indications for UseThe MINNE TIES™ MMF SutureSystem is an adjustable flexibleplastic band that wraps between andaround a tooth to create ananchorage point for maxilla-mandibular fixation andimmobilization. MINNE TIES™MMF Suture System is suitable for:- Pre-operative fixation- Per-operative fixation- Short-term (up to 3 weeks)fixation for minimallydisplaced fractures- Splintage post jawdislocationThe Rapid IMF™ is an adjustableflexible plastic band that wrapsaround a tooth to create ananchorage point for maxilla-mandibular fixation andimmobilization (similar to anorthodontic band). Rapid IMF™ issuitable for:- Pre-operative fixation- Per-operative fixation- Short-term (up to 3 weeks)fixation for minimallydisplaced fractures- Splintage post jawdislocation
Anatomical SiteOral cavityOral cavity
TechnologicalCharacteristics/PerformanceComparison of SubjectDevice to Predicate DeviceMINNE TIES™ MMFSuture System – K162046/S001(under review)Predicate Device:Rapid IMF™ – K030605
MaterialsPlastic (polypropylene, polyester)Plastic (nylon, elastic) and stainless
and stainless steelsteel
Performance
- Mechanical safetyBlunt device, less potential damage to oral mucosaBlunt device, less potential damage to oral mucosa
- Human factorsLess risk of glove puncture as blunt deviceLess risk of glove puncture as blunt device
- BiocompatibilityBiocompatible with respect to tests carried outBiocompatible; has history of clinical use
- Time UsedSingle Use Only, up to 3 weeksSingle Use Only, up to 3 weeks
SterilityTerminal sterilization by ethylene oxide, sterility assurance level 10-6Terminal sterilization by ethylene oxide, sterility assurance level 10-6
Clinical testingClinical data is not required.Clinical data was not required for the predicate device.
Non-Clinical TestingThe MINNE TIES™ MMF Suture System proved to be successful in all the non-clinical tests conducted. The non-clinical tests conducted included the tests listed below and demonstrated substantial equivalence of the subject device. Mechanical stability of construct (performance, tensile strength) Biocompatibility Corrosion resistance System handling, End-user (cadaver test) Packaging Shelf Life Sterilization, Ethylene Oxide Residuals, BioburdenThe Rapid IMF™ conducted the following non-clinical tests: Tensile Strength Test Biocompatibility Bioburden Ethylene Oxide Residuals
Non-Clinical Testing SummaryThe following Non-Clinical Testing was conducted using current standards and guidance and demonstrated substantial equivalence of the subject device. Biocompatibility per ISO 10993-1: Cytotoxicity per ISO 10993-5, Sensitization per ISO 10993-10 and Irritation per ISO 10993-10 Sterile barrier system testing per Distribution Cycle ASTM D4169, Visual Inspection ASTM F1886-09. Gross Leak Detection ASTM
F2096-11, Seal Strength ASTM F88-09 and Accelerated AgingASTM F1980-07
- Ethylene Oxide Sterilization validation per ANI/AAMI/ISO 11135-1:2007
- Mechanical stability of construct (fatigue, tensile, strength, floss-out)
- Corrosion resistance
- System handling, End-user (cadaver test)
Conclusion:In summary, the MINNE TIEST™ MMF Suture System has the followingsimilarities to the predicate device which has previously received 510(k)clearance:
- Has the same indications for use
- Has the same intended use
- Has the same target population
- Has the same implant duration
- Used in the same anatomical site
- Uses similar technological characteristics
- Uses similar biocompatible materials
- Has the same sterility assurance level
The differences in technological characteristics between the subjectdevice and predicate device are mitigated by the following bench tests:
- Tensile strength
- Fatigue strength
- Floss out
- Human Factors
Therefore, the MINNE TIEST™ MMF Suture System is substantiallyequivalent to the predicate device.

510(k) Premarket Notification – MINNE TIES™ MMF Suture System Summit Medical, Inc.

.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Summit Medical. The logo consists of the word "SUMMIT" in a sans-serif font, with the word "medical" in a smaller font below it. Above the word "SUMMIT" are three blue diamonds, arranged horizontally. The logo is simple and modern, and the blue color gives it a professional look.

510(k) Premarket Notification – MINNE TIES™ MMF Suture System Summit Medical, Inc.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Summit Medical. The logo consists of the word "SUMMIT" in a large, sans-serif font, with the word "medical" in a smaller font below it. Above the word "SUMMIT" are three blue diamonds, arranged horizontally. The logo is simple and modern, and the use of blue gives it a professional and trustworthy feel.

§ 872.4600 Intraoral ligature and wire lock.

(a)
Identification. An intraoral ligature and wire lock is a metal device intended to constrict fractured bone segments in the oral cavity. The bone segments are stabilized by wrapping the ligature (wire) around the fractured bone segments and locking the ends together.(b)
Classification. Class II.