K Number
K162023

Validate with FDA (Live)

Device Name
CardioDay V2.5
Date Cleared
2017-03-24

(245 days)

Product Code
Regulation Number
870.1425
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CardioDay V2.5 Holter Analysis Software is designed for the acquisition, analysis, edit, review, report, and storage of ambulatory and multi-parameter ECG data.

Results of the automated analysis are intended to assist the physician in the interpretation of the recorded data. This information is not intended to serve as a substitute for the physician overread of the recorded ECG data.

The CardioDay V2.5 system is intended to be used by trained operators under the direct supervision of a licensed healthcare practitioner in a hospital or clinic environment.

Patient population includes both adult and pediatric human patients.

CardioDay V2.5 provides the user arrhythmia study and Holter analysis capabilities.

Data acquired may be used for the following indications:

  • · Evaluation of symptoms that may be caused by cardiac arrhythmia and/or conduction disturbances,
  • · Evaluation of symptoms that may be due to myocardial ischemia,
  • · Detection of ECG events that alter prognosis in certain forms of heart disease.
  • · Detection and analysis of pacemaker function and failure,
  • · Determination of cardiac response to lifestyle,
  • · Evaluation of therapeutic interventions,
  • · Investigations in epidemiology and clinical trials.

Prescription Device Statement: United States federal law restricts CardioDay to sale by or on the order of a physician.

Device Description

CardioDay® is a software package that allows a trained physician or health care professional knowledgeable in Holter interpretation, after having performed a long-term continuous electrocardiographic (ECG) recording, to download and analyze the data, review it and produce printed reports. CardioDay® does not perform any diagnosis of data by itself but only displays ECG morphologies and associated, calculated graphs such as heart rate trends, RR variability, and other statistical values in graphical form. The physician will be able to review, edit, and print the data collected.

AI/ML Overview

The provided document does not contain information about specific acceptance criteria, a study proving device performance against those criteria, or details regarding ground truth establishment, sample sizes for test and training sets, expert qualifications, or adjudication methods for CardioDay V2.5.

The document states:

  • "No clinical testing was performed in order to support safety or effectiveness."
  • "Non-clinical testing that has been conducted includes: a. ECG performance testing according to AAMI / ANSI / IEC 60601-2-47; b. Software development life cycle according to IEC 62304; c. Risk management process according to AAMI / ANSI / ISO 14971 and IEC /TR80002-1; d. Usability engineering process according to AAMI / ANSI / IEC 62366-1 and IEC 60601-1-6; e. Security tests to fulfill FDA Guidance for cybersecurity; f. Interface tests to fulfill FDA Guidance for interoperability."

Therefore, I cannot provide the requested table or detailed information as it is not present in the given text.

Summary of what's missing from the document based on your request:

  1. Acceptance criteria and reported device performance table: Not provided. The document states "No clinical testing was performed."
  2. Sample size for test set and data provenance: No clinical test set mentioned.
  3. Number of experts and qualifications for ground truth: No ground truth establishment for a test set mentioned.
  4. Adjudication method for test set: Not applicable as no clinical test set is mentioned.
  5. MRMC comparative effectiveness study: Not applicable as "No clinical testing was performed."
  6. Standalone performance study: Not applicable as "No clinical testing was performed."
  7. Type of ground truth used: Not applicable as no clinical testing or ground truth is mentioned.
  8. Sample size for training set: Not applicable as no machine learning training set is mentioned; the device is described as "Holter Analysis Software" that "does not perform any diagnosis of data by itself."
  9. How ground truth for training set was established: Not applicable for the same reason as above.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized image of an eagle with three heads, representing the department's commitment to health, human services, and science. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle image.

March 24, 2017

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

GETEMED Medizin- und Informationstechnik AG Bert Schadow Regulatory Affairs Manager Oderstrasse 77 Teltow, 14513 DE

Re: K162023

Trade/Device Name: CardioDay V2.5 Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II Product Code: DQK Dated: July 19, 2016 Received: July 22, 2016

Dear Bert Schadow:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

Mude Jellman
for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K162023

Device Name CardioDay V2.5

Indications for Use (Describe)

The CardioDay V2.5 Holter Analysis Software is designed for the acquisition, analysis, edit, review, report, and storage of ambulatory and multi-parameter ECG data.

Results of the automated analysis are intended to assist the physician in the interpretation of the recorded data. This information is not intended to serve as a substitute for the physician overread of the recorded ECG data.

The CardioDay V2.5 system is intended to be used by trained operators under the direct supervision of a licensed healthcare practitioner in a hospital or clinic environment.

Patient population includes both adult and pediatric human patients.

CardioDay V2.5 provides the user arrhythmia study and Holter analysis capabilities.

Data acquired may be used for the following indications:

  • · Evaluation of symptoms that may be caused by cardiac arrhythmia and/or conduction disturbances,
  • · Evaluation of symptoms that may be due to myocardial ischemia,
  • · Detection of ECG events that alter prognosis in certain forms of heart disease.
  • · Detection and analysis of pacemaker function and failure,
  • · Determination of cardiac response to lifestyle,
  • · Evaluation of therapeutic interventions,
  • · Investigations in epidemiology and clinical trials.

Prescription Device Statement: United States federal law restricts CardioDay to sale by or on the order of a physician.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

DHF - Technical Documentation@getemed
CardioDay V2.51450-TD-0033-00
510(K) SUMMARY - Assigned 510(k) number: K162023Rev 04

1 510(K) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

The assigned 510(k) number is: K162023

Submitted By:GETEMED Medizin- und Informationstechnik AGOderstrasse 77, 14513 Teltow, GermanyTel: + 49 - 3328 3942 0
Contact:Dr. Bert SchadowRegulatory Affairs ManagerGETEMED Medizin- und Informationstechnik AGOderstrasse 77, 14513 Teltow, GERMANYTel: + 49 – 3328 3942 70Fax: + 49 - 3328 3942 99
Manufacturing Facility:GETEMED Medizin- und Informationstechnik AGOderstrasse 77, 14513 Teltow, GERMANY
Date of Preparation:March 22, 2017
Name of the Device:CardioDay® V2.5
Common or Usual Name:Holter evaluation software
Classification Name:Computer, Diagnostic, Programmable
Product Classification:21 CFR 870.1425, Class II
Product Code:DQK

Predicate Device Information:

DeviceManufacturer510(k) Number
Primary Predicate Device:CardioDay® version 2.4GETEMED Medizin- undInformationstechnik AGK130516
Secondary Predicate Device:MARS Holter AnalysisWorkstationGE Medical SystemsInformation TechnologiesK132437

2 Reason for Submission

Premarket notification for CardioDay® V2.5, a New Device, seeking authority to market the device under Section 510(k) as a device that is substantially equivalent to the evaluation software CardioDay version 2.4 (K130516, GETEMED Medizin- und Informationstechnik AG) and MARS (K132437, GE Medical Systems Information Technologies).

{4}------------------------------------------------

DHF - Technical Documentation@getemed
CardioDay V2.51450-TD-0033-00
510(K) SUMMARY - Assigned 510(k) number: K162023Rev 04

3 Device Description

CardioDay® is a software package that allows a trained physician or health care professional knowledgeable in Holter interpretation, after having performed a long-term continuous electrocardiographic (ECG) recording, to download and analyze the data, review it and produce printed reports. CardioDay® does not perform any diagnosis of data by itself but only displays ECG morphologies and associated, calculated graphs such as heart rate trends, RR variability, and other statistical values in graphical form. The physician will be able to review, edit, and print the data collected.

The following Interfacing Devices (not part of this 510(k)) are supported by CardioDay V2.5:

  • CardioMem CM 3000 (identical to CardioDay V2.4) ●
  • . CardioMem CM 3000Z (identical to CardioDay V2.4)
  • CardioMem CM 3000-12BT (identical to CardioDay V2.4) ●
  • CardioMem CM 4000(B) (identical to CardioDay V2.4) ●
  • SEER 12 (identical to CardioDay V2.4) ●
  • SEER Light (new in CardioDay V2.5)
  • SEER Light Extend (new in CardioDay V2.5) ●
  • SEER 1000 (24h, 48h and 7-day models) (identical to CardioDay V2.4) ●
  • . Care Scape Gateway (new in CardioDay V2.5)

ব Intended Use

The CardioDav V2.5 Holter Analysis Software is designed for the acquisition, analysis, edit, review, report, and storage of ambulatory and multi-parameter ECG data.

Results of the automated analysis are intended to assist the physician in the interpretation of the recorded data. This information is not intended to serve as a substitute for the physician overread of the recorded ECG data.

The CardioDay V2.5 system is intended to be used by trained operators under the direct supervision of a licensed healthcare practitioner in a hospital or clinic environment. Patient population includes both adult and pediatric human patients.

CardioDay V2.5 provides the user arrhythmia study and Holter analysis capabilities. Data acquired may be used for the following indications:

• Evaluation of symptoms that may be caused by cardiac arrhythmia and/or conduction disturbances,

  • · Evaluation of symptoms that may be due to myocardial ischemia.
  • · Detection of ECG events that alter prognosis in certain forms of heart disease.
  • · Detection and analysis of pacemaker function and failure,
  • · Determination of cardiac response to lifestyle,
  • · Evaluation of therapeutic interventions,
  • · Investigations in epidemiology and clinical trials.

Prescription Device Statement:

United States federal law restricts CardioDay® to sale by or on the order of a physician.

{5}------------------------------------------------

DHF – Technical Documentationgetemed
CardioDay V2.51450-TD-0033-00
510(K) SUMMARY - Assigned 510(k) number: K162023Rev 04

5 Comparison to Predicate Devices

The primary predicate device is CardioDay version 2.4 (K130516). There have been no changes implemented in the modifications to CardioDay® V2.5 that impact the fundamental technology. The indications for use from CardioDay® version 2.4 (K130516) was adapted to the indications for use from MARS (K132437) in order to clarify the indications without affecting the substance or meaning of the indications.

As all indications for use cover the same diagnostic scope, i.e. patients suffering from cardiac related illnesses or complaining of potentially having such illnesses, as well as patients wearing a pacemaker, the Holter evalution software CardioDay® V2.5 can be regarded as substantially equivalent to CardioDay® version 2.4 (K130516) from GETEMED Medizin- und Informationstechnik AG and MARS Holter Analysis Workstation (K132437) from GE Medical Systems Information Technologies.

The following incremental changes were also made:

  • Graphical user interface -> OEM Version for GE Healthcare;
  • MUSE Import/Export Option (order management implemented);
  • ECG Monitor Import Option (via Carescape Gateway from GE Healthcare) and support of SEER Light Holter recorders;
  • Software Licensing Concept with Single Sign-On;
  • Bluetooth Hook-Up Process for SEER 1000 Holter recorders;
  • Error Logging implemented

A detailed discussion of similarities and differences is provided as a table of comparison in this 510(k) submission.

6 Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows

Non-clinical testing that has been conducted includes:

  • a. ECG performance testing according to AAMI / ANSI / IEC 60601-2-47:
  • b. Software development life cycle according to IEC 62304;
  • c. Risk management process according to AAMI / ANSI / ISO 14971 and IEC /TR80002-1;
  • d. Usability engineering process according to AAMI / ANSI / IEC 62366-1 and IEC 60601-1-6;
  • e. Security tests to fulfill FDA Guidance for cybersecurity;
  • f. Interface tests to fulfill FDA Guidance for interoperability.

None of the testing demonstrated that CardioDay® V2.5 brought up any issues of safety or effectiveness.

7 Discussion of Clinical Tests Performed

No clinical testing was performed in order to support safety or effectiveness.

8 Conclusion

CardioDay® V2.5 is identical to its predicate device in intended use / indications for use and operating principle. Verification and validation of CardioDay V2.5 demonstrated that the incremental changes do not raise any new questions of safety and effectiveness to the subject device and the subject device is substantially equivalent to the predicate devices.

§ 870.1425 Programmable diagnostic computer.

(a)
Identification. A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.(b)
Classification. Class II (performance standards).