K Number
K161994

Validate with FDA (Live)

Date Cleared
2016-09-22

(65 days)

Product Code
Regulation Number
878.4400
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Aerin Medical Stylus is indicated for use in otorhinolaryngology (ENT) surgery for the coagulation of soft tissue in the nasal airway, to treat nasal airway obstruction by shrinking submucosal tissue.

Device Description

The Aerin Medical Stylus is a handheld bipolar radiofrequency (RF) probe designed for use in otorhinolaryngology (ENT) surgery. The Stylus comprises a handle, shaft and treatment tip. The treatment tip consists of an array of bipolar electrodes and a temperature sensor that allows for monitoring of tissue temperature during radiofrequency energy delivery. The stylus is connected, via a flexible cable, to a commercially-available RF generator (the ORA-50 S, cleared under K993854) meeting the following requirements: RF operating frequency of 460 kHz (± 5 kHz); bipolar low power (3 watts to 5 watts) RF energy delivery; and sensitive temperature control with low overshoot (50°C to 70°C).

AI/ML Overview

This document describes a 510(k) premarket notification for the Aerin Medical Stylus, a radiofrequency wand intended for use in otorhinolaryngology (ENT) surgery to coagulate soft tissue in the nasal airway to treat nasal airway obstruction. The submission asserts substantial equivalence to the predicate device, Aerin Medical Wand (K150637).

Based on the provided text, the following information can be extracted regarding the acceptance criteria and the study that proves the device meets them:

1. A table of acceptance criteria and the reported device performance

The document does not provide a formal table of specific quantitative acceptance criteria for device performance. Instead, it describes general performance testing conducted and states that the device "met all performance testing requirements." The relevant performance aspects mentioned are:

Acceptance Criteria (General Description)Reported Device Performance
BiocompatibilityDemonstrated through testing per AAMI / ANSI / ISO 10993-1.
Sterilization and Packaging ValidationValidated per AAMI / ANSI / ISO 11137-1 and 11137-2. Packaging met acceptance criteria per ASTM standards.
Electrical SafetyDemonstrated through testing to IEC 60601-1 and 60601-2-2.
Electromagnetic CompatibilityPerformed according to IEC 60601-1-2.
Usability (ergonomics, visualization, orientation, force application)Evaluated as part of a cadaver study, implying met acceptance.
Shaft Strength (force load testing)Conducted, implying adequate strength and met acceptance.
Thermocouple Accuracy and Response TimeTested via water bath immersion, implying met accuracy and response criteria.
Tissue Heating TimeTested using representative tissues, implying met heating time criteria.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Test Set Sample Size:
    • Cadaver Study (for usability): The document mentions a "cadaver study" but does not specify the number of cadavers or individual tests performed.
    • Tissue Heating Time: The number of tissue samples or tests is not specified.
    • Other tests (Biocompatibility, Sterilization, Electrical Safety, EMC, Packaging): These tests typically involve a defined number of device units or samples according to the relevant standards, but specific numbers are not provided in this summary.
  • Data Provenance: Not specified in the provided text. The submission is from Aerin Medical, Inc. based in Sunnyvale, CA, USA, so it is likely that the testing was conducted in the USA or by facilities adhering to relevant international standards. The document does not indicate whether data was retrospective or prospective for any of the non-clinical tests.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable for this device. The testing described is primarily non-clinical engineering and performance testing. "Ground truth" in the traditional sense of expert consensus on diagnostic images or clinical outcomes is not relevant here. Decisions on usability would likely have been made by engineers or clinical specialists involved in the cadaver study, but their qualifications and numbers are not specified.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. As noted above, this summary focuses on non-clinical performance and safety testing, not clinical studies requiring adjudication of outcomes or diagnoses.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. The Aerin Medical Stylus is a surgical device, not an AI-powered diagnostic or assistive tool for human readers. No MRMC study was conducted or is relevant here.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is a physical medical device, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

Not applicable. For the non-clinical studies described:

  • Biocompatibility and Sterilization were validated against established ISO/AAMI standards.
  • Electrical Safety and EMC were validated against IEC standards.
  • Usability was evaluated in a cadaver study (presumably by observing ease of use, visualizing the treatment, etc.).
  • Thermocouple accuracy and response time were measured against known temperatures in a water bath.
  • Tissue heating time was measured directly in representative tissues.

The "ground truth" for these tests are the objective measurements against predefined engineering specifications and regulatory standards.

8. The sample size for the training set

Not applicable. This is a physical device, not an AI algorithm requiring a training set.

9. How the ground truth for the training set was established

Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized eagle with three lines forming its body and wings. The eagle is positioned to the right of a circular emblem that contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 22, 2016

Scott Wolf, M.D. President and Chief Medical Officer Aerin Medical, Inc. 690 W. Fremont Avenue, Unit 3 Sunnyvale, CA 94087

Re: K161994

Trade/Device Name: Aerin Medical Stylus Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: August 19, 2016 Received: August 23, 2016

Dear Dr. Wolf:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Page 2 - Scott Wolf, M.D.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Eric A. Mann -S

for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K161994

Device Name Aerin Medical Stylus

Indications for Use (Describe)

The Aerin Medical Stylus is indicated for use in otorhinolaryngology (ENT) surgery for the coagulation of soft tissue in the nasal airway, to treat nasal airway obstruction by shrinking submucosal tissue.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Aerin Medical. The word "Aerin" is in a large, blue, sans-serif font, with the word "MEDICAL" in a smaller, gray, sans-serif font below it. To the right of the word "Aerin" are several green circles that are arranged in a gradient pattern.

This 510(k) Summary is being submitted in accordance with the requirements of 21CFR 807.92.

General Information

Submitter Information
Company:Aerin Medical, Inc.
Submitter's Address:690 W. Fremont Avenue, Unit 3Sunnyvale, CA 94087
Contact Person:Scott Wolf, M.D.President and Chief Medical OfficerPhone: 650-434-3247Fax: 408-716-2438
EstablishmentRegistration Number3011625895
Date Prepared:September 21, 2016
Name of the Device
Proprietary Name:Aerin Medical Stylus
Common Name:Radiofrequency wand
Classification Name:Electrosurgical cutting and coagulation device andaccessories
Classification Panel:General and Plastic Surgery
Device Class:Class II
Product Code:GEI
CFR Section:21 CFR 878.4400
Predicate Device:Aerin Medical Wand (K150637)

Device Description

The Aerin Medical Stylus is a handheld bipolar radiofrequency (RF) probe designed for use in otorhinolaryngology (ENT) surgery. The Stylus comprises a handle, shaft and treatment tip. The treatment tip consists of an array of bipolar electrodes and a temperature sensor that allows for monitoring of tissue temperature during radiofrequency energy delivery. The stylus is connected, via a flexible cable, to a commercially-available RF generator (the ORA-50 S, cleared under K993854) meeting the following requirements: RF operating frequency of 460 kHz (± 5 kHz); bipolar low power (3 watts to 5 watts) RF energy delivery; and sensitive temperature control with low overshoot (50°C to 70°C).

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Aerin Medical. The word "Aerin" is in a large, blue, sans-serif font. Below "Aerin" is the word "MEDICAL" in a smaller, sans-serif font. To the right of "Aerin" are several green circles of varying sizes, arranged in a pattern.

The Aerin Medical Stylus is used in the treatment of patients experiencing chronic nasal airway obstruction. During a treatment procedure, the clinician inserts the tip of the Aerin Medical Stylus into a patient's nostril to deliver low power RF energy to the target tissue of the nasal airway. Radiofrequency treatment of tissue creates a coaqulative lesion which fibroses and retracts as it heals, thereby shrinking the tissue to lessen the degree of obstruction.

Indications for Use

The Aerin Medical Stylus is indicated for use in otorhinolaryngology (ENT) surgery for the coagulation of soft tissue in the nasal airway, to treat nasal airway obstruction by shrinking submucosal tissue.

Summary of the technological characteristics of the device compared to the predicate device

The Aerin Medical Stylus was found to be equivalent to the predicate device in design and intended use to deliver bipolar RF energy to coagulate tissue in otorhinolaryngology procedures.

CharacteristicAerin Medical Stylus(Models FG011 & FG174)Subject DeviceAerin Medical Wand(Model FG011)Predicate DeviceK150637
Indications for UseSameIndicated for use inotorhinolaryngology (ENT)surgery for the coagulationof soft tissue in the nasalairway, to treat nasalairway obstruction byshrinking submucosaltissue.
Design configurationSameIntegrated cable, handleand electrode
Energy typeSameBipolar radiofrequency
Tissue temperatureSame50 — 70 °C(temperature controlled)

Summary of non-clinical tests

Device safety is supported through biocompatibility testing, sterilization and packaging validation activities, electrical safety testing and electromagnetic Biocompatibility of the Aerin Medical Stylus was compatibility testing. demonstrated through testing performed in accordance with AAMI / ANSI / ISO standards 10993-1 (2009/(R) 2013). Radiation sterilization was validated per AAMI / ANSI / ISO 11137-1 (2006/(R) 2010) and 11137-2 (2013). Electrical safety was demonstrated through testing to IEC 60601-1 (2005/(R) 2012) and

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Aerin Medical. The word "Aerin" is in a large, blue, sans-serif font. Below "Aerin" is the word "MEDICAL" in a smaller, sans-serif font. To the right of "Aerin" are several green circles of varying sizes.

60601-2-2 (2009) and electromagnetic compatibility testing was performed according to IEC 60601-1-2 (2007). The packaging systems were subjected to transit testing, visual inspection, accelerated aging, gross leak detection and seal strength testing in accordance with the applicable ASTM standards (ASTM D4169-14, ASTM F1980-07, ASTM F2096-11, and ASTM F88/F88M-15), and met all acceptance criteria.

Device performance testing included usability factors as well as factors related to system responsiveness and effectiveness for use in a clinical setting. Usability factors, such as ergonomics, visualization, orientation and force application, were evaluated as part of a cadaver study, Additionally, force load testing was conducted to verify adequate shaft strength. The efficacy of the subject device is supported by thermocouple accuracy and response time testing via water bath immersion and tissue heating time testing. Tissue heating time testing was performed using tissues representative of the submucosal tissues of the nasal airway. The subject device met all performance testing requirements, thereby supporting its equivalency to the predicate device in terms of delivery of RF energy.

Summary of clinical tests

Not applicable. No clinical tests were necessary.

Conclusion

Testing demonstrates that the Aerin Medical Stylus is substantially equivalent to the predicate in terms of both indications for use and delivered RF treatment and is as safe and effective for its intended use.

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.