K Number
K161950

Validate with FDA (Live)

Date Cleared
2017-05-11

(300 days)

Product Code
Regulation Number
880.5570
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Common Type Insulin Pen Needle is intended for use with pen injector device for subcutaneous injection of insulin.

The Safety Type Insulin Pen Needle is intended for use with pen injector device for subcutaneous injection of insulin. Additionally, after withdrawal of the Safety Type Insulin Pen Needle from the body, the attached needle safety shield automatically covers the needle to minimize the risk of accidental needlestick.

Device Description

The proposed product, insulin pen needle, is manufactured by Promisemed Hangzhou Meditech Co., Ltd, which is designed for use with a pen injector for the subcutaneous injection of insulin. The products have two types, common type and safety type. Both of the two types are sterile with a Sterility Assurance Level (SAL) of 10-6, non-pyrogenic and single-use devices. Each type has several models. Different models are distinguished by needle gauge and length.

The Common Type Insulin Pen Needle consists of needle shield, needle shield, needle tube, needle hub, UV glue and silicone oil. UV glue is used to glue needle tube and needle hub and the silicone oil is used to needle tube lubrication.

The Safety Type Insulin Pen Needle consists of needle shield, spring, housing, needle hub, needle container, UV glue and silicone oil. UV glue is used to glue needle tube and needle hub and the silicone oil is used to needle tube lubrication.

The user proceeds with inserting the needle into the skin manually. Safety Type Insulin Pen Needle is designed to reduce occurrence of accidental needle sticks from patient end of the needle by providing a shield that covers and locks the needle after use. As the user proceeds with inserting the needle into the skin the shield will retract. After the injection is completed and needle is removed from the skin, the shield will automatically extend to cover the needle and lock in place. Once the Safety Type Insulin Pen Needle is in the locked mode, it can no longer be used.

Both the Common Type Insulin Pen Needle and the Safety Type Pen Needle are for OTC use, and they are external communication, blood indirect devices. The contact duration for both subjected devices is within 24h, and they belong to limited contact device according ISO 10993-1.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and the study proving the device meets them:

Device: Verifine® Common Type Insulin Pen Needle and Verifine® Safety Type Insulin Pen Needle

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are generally implied by the conformance to various ISO standards and specific performance aspects. The document states "Passed" for all tests, indicating the device met the requirements.

Test CategoryAcceptance Criteria (Implied by Standards/Description)Reported Device Performance
MaterialMade of tubing materials specified in ISO 9626.Passed
DimensionsNeedles fit the test apparatus specified in item 7.3 of ISO 11608-2.Passed
Flow RateTested in accordance with Annex A to ISO 11608-2. (Implies meeting specified flow rates for different gauges/lengths)Passed
Bond (Hub & Needle Tube)Union of hub and needle tube does not break when tested in accordance with Clause 9 of ISO 11608-2.Passed
Needle PointsFulfill the requirements of 4.5 of ISO 11608-2.Passed
Freedom from DefectsFulfill the requirements of ISO 7864:1993, 11.3.Passed
LubricationNeedle tube lubricated at both patient end and cartridge end, fulfilling 4.7 of ISO 11608-2.Passed
Dislocation of Measuring Point Patient EndDislocation of the cannula point at the patient end in accordance with Table 2 (ISO 11688-2) when tested in accordance with Clause 8 of ISO 11608-2.Passed
Functional Compatibility with Needle-Based Injected SystemsCompatibility with any NIS claimed only after testing in accordance with Clause 11 of ISO 11608-2.Passed
Ease of Assemble and DisassemblyAttachment of the needle possible without removing the needle from its opened unit packaging. Compliance checked according to the requirements of Clause 11 of ISO 11608-2.Passed
BiocompatibilityCytotoxicity: No cytotoxicity (ISO 10093-5)Sensitization: No evidence of sensitization (ISO 10993-10)Intracutaneous Reactivity/Skin Irritation: No evidence of skin irritation (ISO 10993-10)Hemocompatibility: No evidence of hemolysis (ISO 10993-4, ASTM F765)Systemic Toxicity (acute): No systemic toxicity (ISO 10993-11)Pyrogen: No pyrogen (ISO 10993 series)Passed
Safety Feature (Safety Type Insulin Pen Needle)No needlestick injuries occurred and zero failure of the protective feature during simulated clinical use testing (per FDA guidance document of Medical Devices with Sharps Injury Prevention Features).Passed

2. Sample Size Used for the Test Set and Data Provenance

The document primarily refers to "nonclinical testing" and "bench tests."

  • Nonclinical Testing (General): The document does not specify a general sample size for all nonclinical tests for the proposed devices but states that "bench tests of Needle with 33G demonstrated conformances..." and "bench tests of Needles with 29G, 30G and 4mm, 6mm demonstrated conformances..." implying specific samples were used for these models.
  • Simulated Clinical Use Testing (for Safety Type Insulin Pen Needle): 600 subject devices were chosen for this specific test.
  • Data Provenance: The tests were conducted by Promisemed Hangzhou Meditech Co., Ltd. (China). The document does not specify if the testing was retrospective or prospective, but given it's for a 510(k) submission, it would be considered prospective testing for the purpose of demonstrating equivalence.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

  • None stated directly. The document refers to conformance with ISO standards and an FDA guidance document as the basis for evaluation,
    rather than expert adjudication for ground truth. For technical performance, the "ground truth" is adherence to the standard's requirements. For the safety feature, the ground truth is the absence of needlestick injuries and feature failures, as observed in the simulated clinical use.

4. Adjudication Method for the Test Set

  • None (in the sense of expert consensus on patient cases). The "adjudication" is by adherence to objective, measurable criteria defined by international standards (ISO) and observed performance in specific tests (e.g., the simulated clinical use test). The results are reported as "Passed," implying a direct comparison against the pre-defined criteria.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

  • No. This type of study (MRMC) is typically relevant for interpretative devices (e.g., imaging AI) where multiple human readers assess cases, with and without AI assistance, to measure diagnostic performance. The devices in question are insulin pen needles, which are mechanical devices, not interpretative. Therefore, an MRMC study is not applicable or mentioned.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

  • Not applicable. This device is a physical medical device (insulin pen needle), not an algorithm or AI system. Therefore, the concept of "standalone performance" for an algorithm without human-in-the-loop is not relevant. The device's performance is inherently "standalone" in its mechanical function.

7. The Type of Ground Truth Used

  • Adherence to International Standards and Observed Technical Performance. The ground truth for most tests is the objective fulfillment of requirements laid out in ISO standards (e.g., ISO 9626 for tubing materials, ISO 11608-2 for needle performance, ISO 10993 series for biocompatibility). For the safety feature, the ground truth is the direct observation of zero failures during the simulated clinical use testing as per FDA guidance.

8. The Sample Size for the Training Set

  • Not applicable. As a physical medical device, there is no "training set" in the context of an algorithm or AI. The manufacturing processes and design principles are based on engineering standards and iterative design, not machine learning model training.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable. See point 8. There is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 11, 2017

Promisemed Hangzhou Meditech Co., Ltd. % Field Fu Consultant Shenzhen Joyantech Consulting Co., Ltd No. 55 Shizhou Middle Road, Nanshan District Shenzhen, GD755 Guangdong CHINA

Re: K161950

Trade/Device Name: Verifine® Common Type Insulin Pen Needle and Verifine® Safety Type Insulin Pen Needle Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: Class II Product Code: FMI Dated: April 5, 2017 Received: April 10, 2017

Dear Field Fu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Image /page/1/Picture/8 description: The image contains a signature and the name "Tina Kiang". The signature is a complex, looping design to the left of the name. Below the name is the character sequence "-s", which may be part of the signature or a separate element.

Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K161950

Device Name

Verifine® Common Type Insulin Pen Needle and Verifine® Safety Type Insulin Pen Needle

Indications for Use (Describe)

The Common Type Insulin Pen Needle is intended for use with pen injector device for subcutaneous injection of insulin.

The Safety Type Insulin Pen Needle is intended for use with pen injector device for subcutaneous injection of insulin. Additionally, after withdrawal of the Safety Type Insulin Pen Needle from the body, the attached needle safety shield automatically covers the needle to minimize the risk of accidental needlestick.

Type of Use (Select one or both, as applicable)
-------------------------------------------------
Production Use (Part 21 CFR 221.Subpart D)
Test Strip Control Use (21 CFR 221.Subpart C)

__ Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

Contact Details 1.

1.1 Applicant information

Applicant NamePromisemed Hangzhou Meditech Co.,Ltd.
AddressNo.18 Haishu Road Future Sci-Tech Park, Yuhang District, Hangzhou,China
Phone No.+86-571-86775619
Fax No.+86-571-88772985
Contact personKeshin Hu
Contact person's e-mailkeshin@promisemed.ca
Company e-mailadmin@promisemed.ca
Date PreparedApril 05, 2017
Websitehttp://www.promisemed.ca/

1.2 Consultant information

NameShenzhen Joyantech Consulting Co., LtdImage: [logo]Image: [Chinese characters]
AddressRoom 1122, International Mayors Communication Centre, NO. 55Shizhou middle road, Nanshan District, Shenzhen
Phone No.+86-755-86069197
Contact personElly Xu; Field Fu
Contact person's e-mailelly@cefda.com; info@cefda.com
Websitehttp://www.cefda.com

Device information 2.

Trade nameVerifine® Common Type Insulin Pen Needle; Verifine® Safety TypeInsulin Pen Needle
Common nameInsulin Pen Needle
ModelCommon Type:IPN-29-12;IPN-30-8;IPN-31-4;IPN-31-5;IPN-31-6;IPN-31-8;IPN-32-4;IPN-32-5; IPN-32-8; IPN-33-4Safety Type:SPN-29-6;SPN-29-8;SPN-30-4;SPN-30-5;SPN-30-6;SPN-30-8;SPN-31-4;SPN-31-5;SPN-31-6;SPN-31-8
ClassificationII
Classification nameNeedle, Hypodermic, Single Lumen
Product codeFMI
Regulation No.880.5570

3. Legally Marketed Predicate Device

{4}------------------------------------------------

Trade NameInsulin Pen Needle
510(k) NumberK133059
Product CodeFMI
ManufacturerWenzhou Beipo Science & Technology Co., Ltd
Trade NameSafe Block Pen Needle
510(k) NumberK151311
Product CodeFMI
ManufacturerArtsana S.p.A Unit Production N°30, Via Saldarini Catelli 6/10, 22070Grandate COMO - ITALY

Device Description 4.

The proposed product, insulin pen needle, is manufactured by Promisemed Hangzhou Meditech Co., Ltd, which is designed for use with a pen injector for the subcutaneous injection of insulin. The products have two types, common type and safety type. Both of the two types are sterile with a Sterility Assurance Level (SAL) of 104, non-pyrogenic and single-use devices. Each type has several models. Different models are distinguished by needle gauge and length.

The Common Type Insulin Pen Needle consists of needle shield, needle shield, needle tube, needle hub, UV glue and silicone oil. UV glue is used to glue needle tube and needle hub and the silicone oil is used to needle tube lubrication.

The Safety Type Insulin Pen Needle consists of needle shield, spring, housing, needle hub, needle container, UV glue and silicone oil. UV glue is used to glue needle tube and needle hub and the silicone oil is used to needle tube lubrication.

The user proceeds with inserting the needle into the skin manually. Safety Type Insulin Pen Needle is designed to reduce occurrence of accidental needle sticks from patient end of the needle by providing a shield that covers and locks the needle after use. As the user proceeds with inserting the needle into the skin the shield will retract. After the injection is completed and needle is removed from the skin, the shield will automatically extend to cover the needle and lock in place. Once the Safety Type Insulin Pen Needle is in the locked mode, it can no longer be used.

Both the Common Type Insulin Pen Needle and the Safety Type Pen Needle are for OTC use, and they are external communication, blood indirect devices. The contact duration for both subjected devices is within 24h, and they belong to limited contact device according ISO 10993-1.

The dimension for both Common Type Insulin Pen Needle and the Safety Type Pen Needle are shown as below.

TypeModelNeedle ODNeedle availablelength
CommonTypeIPN-29-120.33mm12mm
IPN-30-80.30mm8mm
IPN-31-40.25mm4mm
IPN-31-50.25mm5mm
IPN-31-60.25mm6mm
IPN-31-80.25mm8mm
IPN-32-40.23mm4mm

{5}------------------------------------------------

IPN-32-50.23mm5mm
IPN-32-80.23mm8mm
IPN-33-40.20mm4mm
SafetyTypeSPN-29-60.33mm6mm
SPN-29-80.33mm8mm
SPN-30-40.30mm4mm
SPN-30-50.30mm5mm
SPN-30-60.30mm6mm
SPN-30-80.30mm8mm
SPN-31-40.25mm4mm
SPN-31-50.25mm5mm
SPN-31-60.25mm6mm
SPN-31-80.25mm8mm

5. Intended Use/Indication for Use

The Common Type Insulin Pen Needle is intended for use with pen injector device for subcutaneous injection of insulin.

The Safety Type Insulin Pen Needle is intended for use with pen injector device for subcutaneous injection of insulin.

Additionally, after withdrawal of the Safety Type Insulin Pen Needle from the body, the attached needle safety shield automatically covers the needle to minimize the risk of accidental needlestick.

6. Substantial Equivalence Comparison

6.1 Comparison between Common Type Insulin Pen Needle and Predicate Device(K133059)

ItemProposed Device:Common Type Insulin PenNeedlePredicate Device:Insulin Pen Needle(K133059)Comments
Product CodeFMIFMISame
Intended UseThe Common Type Insulin Pen Needle is intended for use with pen injector device for subcutaneous injection of insulin.The Insulin Pen Needle is intended for use with pen injector devices for the subcutaneous injection of insulin.Same
OperatingPrincipleThe user proceeds with inserting the needle into the skin manually. The patient end and the cartridge end of the tube are lubricated using a silicone based lubricant for ease of injection and rubber septum penetration.The user proceeds with inserting the needle into the skin manually. The patient end and the cartridge end of the tube are lubricated using a silicone based lubricant for ease of injection and rubber septum penetration.Same

{6}------------------------------------------------

SEC 005:001 510K Summary

Version:A/0

ConfigurationNeedle Tube, Needle shield (Tube Sheath), Needle hub (Hub), Needle container (Hub Sheath), Sealing dialysis paper (Sealed Paper).Needle Tube, Hub, Tube Sheath, Hub Sheath and Sealed Paper.Same
Length4mm, 5mm, 6mm, 8mm, 12mm4mm, 5mm, 6mm, 8mm, 10mm, 12mmSimilar (* 1)
Needle Gauge29G/30G/31G/32G/33G29G/30G/31G/32GSame
Tip configurationPer ISO 11608-2, section 4.5, visually sharp at 2.5X magnification, designed to minimize coring and fragmentationPer ISO 11608-2, section 4.5, visually sharp at 2.5X magnification, designed to minimize coring and fragmentationSame
MaterialTube: 304 Stainless SteelNeedle hub: PolypropyleneTube: 304 Stainless SteelHub: PolypropyleneSame
PerformanceComplied with ISO 7864, ISO 9626, and ISO 11608-2Complied with ISO 7864, ISO 9626, and ISO 11608-2Same
SterilizationSAL: 10-6Method: EO SterilizedSAL: 10-6Method: EO SterilizedSame
Shelf Life5 years5 yearsSame
Single UseYesYesSame
BiocompatibilityComplied with ISO10993 series standards, and the following tests are performed,- Cytotoxicity: No cytotoxicity- Skin Irritation: No evidence of skin irritation- Skin Sensitization: No evidence of sensitization- Acute Systemic Toxicity: No systemic toxicity- Hemolysis: No evidence of hemolysis- Pyrogen: No pyrogenComplied with ISO10993 series standards, and the following tests are performed,- Cytotoxicity: No cytotoxicity- Skin Irritation: No evidence of skin irritation- Skin Sensitization: No evidence of sensitization- Acute Systemic Toxicity: No systemic toxicity- Hemolysis: No evidence of hemolysis- Pyrogen: No pyrogenSame

*1: The needle tube lengths of the proposed device are covered by the predicate device. Compare to the predicate device, the proposed device has an additional 33G model insulin pen needle. 33G model for Insulin Pen Needle is a widely used in the market and had been approved in FDA (Such as K152410). On the other hand, the bench tests of Needle with 33G demonstrated conformances to ISO 7864, ISO 9626, and ISO 11608-2.

6.2 Comparison between Safety Type Insulin Pen Needle and Predicate Device(K151311)

{7}------------------------------------------------

SEC_005:001_510K Summary

ItemProposed Device:Safety Type Insulin PenNeedlePredicate Device:Insulin Pen Needle (K151311)Comments
Product CodeFMIFMISame
Intended UseThe Safety Type Insulin PenNeedle is intended for use withpen injector device forsubcutaneous injection ofinsulin.Additionally, after withdrawal ofthe Safety Type Insulin PenNeedle from the body, theattached needle safety shieldautomatically covers theneedle to minimize the risk ofaccidental needlestick.It is intended for use with peninjector devices for theinjection of fluids, includinginsulin.Additionally the attached safetyshield automatically locks inplace and reduces theoccurrence of accidental sticksfrom the patient end of theneedle. The shield also servesto conceal the needle beforeand after injection.Similar(* 2)
OperatingPrincipleAs the user proceeds withinserting the needle into theskin the shield will retract. Afterthe injection is completed andneedle is removed from theskin, the shield willautomatically extend to coverthe needle and lock in place.Once the Safety Type InsulinPen Needle is in the lockedmode, it can no longer beused.As the user proceeds withinserting the needle into theskin the shield will retract. Afterthe injection is completed andneedle is removed from theskin, the shield willautomatically extend to coverthe needle and lock in place.Once the Safety Type InsulinPen Needle is in the lockedmode, it can no longer beused.Same
Length4mm, 5mm, 6mm,8mm(tolerance as per ISO11608-2)5mm, 8mm(tolerance as perISO 11608-2)Similar (* 3)
GaugeSharps InjuryPreventionFeaturesG29, G30, G31Safety shieldG31Safety shieldSame
Configurationand MaterialTube: Stainless SteelNeedle Hub: PolypropyleneSpring: Stainless SteelNeedle container(PrimaryContainer):PPNeedle/Safety shield: ABSCannula: Stainless SteelNeedle Hub: PolypropyleneSpring: Stainless SteelPrimary Container:HDPEShield:HDPESameSameSameDifferent (* 4)
PerformanceComplied with ISO 7864, ISO 9626, ISO 11608-2 and ISO 23908Complied with ISO 7864, ISO 9626, ISO 11608-2 and ISO 23908Same
SterilizationSAL: 10-6Method: Irradiation SterilizedSAL: 10-6Method: EO SterilizedDifferent (* 5)
Shelf Life5 years5 yearsSame
Single UseYesYesSame
BiocompatibilityComplied with ISO10993 series standards, and the following tests are performed,• Cytotoxicity: No cytotoxicity• Skin Irritation: No evidence of skin irritation• Skin Sensitization: No evidence of sensitization• Acute Systemic Toxicity: No systemic toxicity• Hemolysis: No evidence of hemolysisPyrogen: No pyrogenComplied with ISO10993 series standards, and the following tests are performed,• Cytotoxicity: No cytotoxicity• Skin Irritation: No evidence of skin irritation• Skin Sensitization: No evidence of sensitization• Acute Systemic Toxicity: No systemic toxicity• Hemolysis: No evidence of hemolysisPyrogen: No pyrogenSame

{8}------------------------------------------------

Version:A/0

  • 2: The proposed device is intended to use with pen injector device for subcutaneous injection of insulin. Compare to proposed device, the predicate device is used for the injection of fluids, including insulin. The usage range of proposed device is included in the predicate device.

  • 3: The length and the gauge is the specification of needle tube that has no relationship with the safety prevention features. The Insulin Pen Needle (K151311) has the tube diameter of 29G and 30G, it also has the tube length of 4mm and 6mm. These models of insulin pen needle are widely used in the market. Moreover, the bench tests of Needles with 29G, 30G and 4mm, 6mm demonstrated conformances to ISO 7864. ISO 9626. ISO 11608-2 and ISO 23908. Therefore the differences do not raise new concerns to establish substantial equivalence to the predicate.

  • 4: Although the materials of safety shield and needle container between the proposed device and the predicate device are different, the ABS and PP are two kinds of materials that are widely used on medical device. Moreover, the biocompatibility test report of the proposed device demonstrate that subject device is biocompatible and the performance as intended.

  • 5: Although the sterilization methods between the proposed device and the predicate device are different, but the radiation sterilization effect of the proposed device is proved by the Microbiological validation report of gamma radiation sterilization dose audit test report in VOL 014. The validation report showed that the sterilization effect of the proposed device can achieve a Sterility Assurance Level (SAL) of 10°, and the radiation dose audit test reports showed that the minimum gamma radiation dose to achieve a 10° SAL was acceptable according to ISO 11737-2:2012.

{9}------------------------------------------------

Non-clinical Testing 7.

All nonclinical testing performed on new devices is to demonstrate the substantial equivalence to the predicate devices. Tests setup and execution are performed in accordance with applicable standards. Results of the testing are demonstrating the standards and matching the performance of new devices to the predicate devices.

The following performance data were provided in support of the substantial equivalence determination.

TestRequirementsResults
MaterialsThe needle shall be made of tubing materials specified in ISO 9626.Passed
DimensionsThe needles shall fit the test apparatus specified in item 7.3 of ISO 11608-2.Passed
Flow rate through the needleThe needles were tested in accordance with Annex A to ISO 11608-2.Passed
Bond between hub and needle tubeThe union of the hub and needle tube shall not break when tested in accordance with Clause 9 of ISO 11608-2.Passed
Needle pointsThe needle points shall fulfill the 4.5 of ISO 11608-2.Passed
Freedom from defectsThe needle points shall fulfil the requirements of ISO 7864:1993, 11.3.Passed
LubricationThe needle tube should be lubricated at both patient end and the cartridge end. The lubrication shall fulfil the 4.7 of ISO 11608-2.Passed
Dislocation of measuring point patient endDislocation of the cannula point at the patient end shall be in accordance with Table 2 (ISO 11688-2) when tested in accordance with Clause 8 of ISO 11608-2.Passed
Functional compatibility with needle-based injected systemsCompatibility with any NIS shall be claimed only after testing in accordance with Clause 11 of ISO 11608-2.Passed
Easy of assemble and disassemblyAttachment of the needle shall be possible without removing the needle from its opened unit packaging. Compliance is checked according to the requirements of Clause 11 of ISO 11608-2.Passed
BiocompatibilityCytotoxicityISO 10093-5 Biological evaluation of medical devices -Part 5: Tests for in vitro cytotoxicityPassed
SensitizationISO 10993-10 Biological evaluation of medical devices -Part 10:Tests for irritationPassed
IntracutaneousReactivityISO 10993-10 Biological evaluation ofmedical devices -Part 10:Tests for irritationand skin sensitizationPassed
and skin sensitization
HaemocompatibilityISO 10993-4 Biological evaluation of medicaldevices Part 4: Selection of tests forinteractions with bloodASTM F765 Standard Practice forAssessment of Hemolytic Properties ofMaterialsPassed
System toxicity(acute)ISO 10993-11:2006/(R)2010, Biologicalevaluation of medical devices - Part 11: Testsfor systemic toxicityPassed

{10}------------------------------------------------

Simulated Clinical Use Testing

As for Safety Type Insulin Pen Needle, a simulated clinical use testing was conducted to verify the safety feature per FDA guidance document of Medical Devices with Sharps Injury Prevention Features. Based on the guidance, 600 subject devices were chosen in the simulated clinical use testing. There was no needlestick injuries occurred and zero failure of the protective feature in the testing.

8. Clinical Testing

Substantial equivalence does not depend on the clinical test data.

Conclusions 9.

Based on device comparison information and non-clinical bench testing, the proposed device is substantially equivalent to legally marketed predicate devices (K133059, K151311).

§ 880.5570 Hypodermic single lumen needle.

(a)
Identification. A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.(b)
Classification. Class II (performance standards).