K Number
K161920

Validate with FDA (Live)

Device Name
Comfort Cover
Date Cleared
2016-07-15

(2 days)

Product Code
Regulation Number
892.1710
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The indications for use of the product is to remove the cold from the bucky by placing this sheet between patient's breast and the image receptor plate allowing the patient to feel more comfortable and less cold without interfering with image quality.

Device Description

Comfort Cover is a single-use disposable cover placed between the patient's breast and the image receptor plate during mammography. The purpose of this product is to allow the patient to feel more comfortable and less cold during the mammographic examination without interfering with image quality. Comfort Cover will allow for artifact-free images in both Field Digital Mammography (FFDM) and Digital Breast Tomosynthesis (DBT), patient radiation dose is equivalent to that when using no Comfort Cover and the Image quality is equivalent to images exposed without Comfort Cover.

Comfort Cover is designed to fit small and large image receptor plates of all leading mammographic equipments. The product is constructed of a fabric material with a purple printed design and has adhesive backing with a paper liner. Both the material and adhesive backing of the product do not affect imaging.

The product is directly applied wrinkle free to the image receptor plate of the mammographic machine by the adhesive backing exposed by peeling off the liner. After the patient's mammogram, the product is peeled off the image receptor plate and disposed. As the product removes cleanly without leaving behind any residue, the image receptor plate does not need to be cleaned between Comfort Covers nor between patients.

AI/ML Overview

The provided document, a 510(k) Premarket Notification for the "Comfort Cover," describes the device, its intended use, and comparative testing against a predicate device. However, it does not contain the level of detail typically found in a clinical study report or a regulatory submission for a device involving AI or complex performance metrics. The Comfort Cover is a disposable cover for mammography plates to enhance patient comfort without affecting image quality.

Below is an attempt to extract and synthesize the requested information based on the document, noting where specific details are not provided or are not applicable to this type of device.

1. Table of Acceptance Criteria and Reported Device Performance

Performance MetricAcceptance Criteria (Stated or Implied)Reported Device Performance
Biocompatibility- Cytotoxicity: Pass ISO 10993-5 (Biological evaluation of medical devices - Part 5: Tests for in vitro Cytotoxicity).- Irritation and Skin Sensitization: Pass ISO 10993-10 (Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization).- The subject device passed all tests (ISO 10993-5 and ISO 10993-10).
Mechanical/Physical Performance- Visual and Dimensional: No visible defects, correct dimensions.- Thermal Conductivity: Performance in line with material properties (implied comparison to predicate).- Ink Smear: No ink smearing as per ASTM D5264-98.- Peel Adhesion: Adequate peel adhesion as per ASTM D6252-98 (90° Angle).- Removes Cleanly: Product removes cleanly without leaving residue (in-house method).- The subject device passed all visual and dimensional examinations.- Thermal Conductivity Test was performed.- Ink Smear Test was performed and passed.- Peel Adhesion Test was performed and passed.- Removes Cleanly Test was performed and passed (removes cleanly from mammographic equipment without leaving residue).
Radiological Performance- Artifact-free images: Allow for artifact-free images in both Full Field Digital Mammography (FFDM) and Digital Breast Tomosynthesis (DBT) when using Comfort Cover.- Patient Radiation Dose: Equivalent to that when using no Comfort Cover.- Image Quality: Equivalent to images exposed when using no Comfort Cover.- Radiolucent: Does not image, no artifacts, quality of images not affected.- Comfort Cover allows for artifact-free images in both FFDM and DBT.- Patient radiation dose is equivalent to that when using no Comfort Cover.- Image quality is equivalent to images exposed when using no Comfort Cover.- The device is radiolucent.
Usability- Fulfill design requirements of making mammography examination more comfortable.- Not negatively impact patient positioning.- The Comfort Cover fulfilled its design requirements of making the mammography examination more comfortable.- The Comfort Cover did not negatively impact patient positioning.
Comparison to Predicate (Overall)- Similar Indications for Use, Target Population, Where Used, Product Size, Packaging and Labeling, Biocompatibility, Sterility, Dispensing Method, and Environment Compatibility.- Minor material differences (fabric type, ink color) do not affect safety and effectiveness.- Provides comfort and removes cold without affecting image quality.- Allows for artifact-free images and image quality equivalent to images exposed without Comfort Cover.- Removes cleanly without residue.- Printed ink does not smear.- Remains where placed.- Radiolucent.- All "None" in the "Significant Differences" column of Table 5-1 indicate similarity.- Stated: "Safety and performance testing results have shown that the slight differences in material composition do not affect safety and effectiveness."- Stated: "Provides comfort and removes cold during a mammographic examination without affecting quality of the images."- Stated: "Allows for artifact-free images and image quality equivalent to Comfort Cover."- Stated: "Removes cleanly from the mammographic equipment without leaving behind any residue."- Stated: "Printed ink does not smear."- Stated: "Remains where placed."- Stated: "Does not image, no artifacts, quality of images not affected, radiolucent."

2. Sample size used for the test set and the data provenance
The document does not provide specific sample sizes (e.g., number of covers, images, or participants) for the test sets. It states that "a number of non-clinical performance tests" were completed.

  • Biocompatibility: Testing was conducted per ISO 10993-5 and ISO 10993-10 on the subject device.
  • Mechanical Testing: Performed on "both the subject and the predicate devices."
  • Radiological Studies: Phantom image testing was performed "as per MQSA regulations and with relevant International Electro technical Commission (IEC) standards (62220-1-2:2007 and 61223-3-2)." This implies the use of standardized phantoms, not patient data.
  • Usability Studies: Included "radiologists, radiology technologists and patients." The number of each group is not specified.

Data provenance (country of origin, retrospective/prospective) is not specified, but given the nature of the tests (lab-based, phantom studies, usability feedback), it is likely all internal data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This information is not provided. For the usability studies, "radiologists" and "radiology technologists" were involved, but their number and specific qualifications (e.g., years of experience) are not detailed. For radiological studies, phantom images were used, implying objective measurements rather than expert human interpretation for ground truth.

4. Adjudication method for the test set
Not applicable or not described. The tests are primarily objective measurements (e.g., passing a specific ISO standard, conforming to ASTM methods, measuring radiation dose, image quality metrics from phantom images). For usability, it simply states that user needs were fulfilled, but no adjudication method for consensus is mentioned.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No MRMC study was done, nor is this device related to Artificial Intelligence (AI). The Comfort Cover is a physical accessory for patient comfort during mammography.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
Not applicable. This device is not an algorithm or AI.

7. The type of ground truth used

  • Biocompatibility: In-vitro and in-vivo (skin irritation/sensitization) test results against ISO standards.
  • Mechanical Testing: Objective measurements (e.g., ASTM standards, in-house developed test for residue).
  • Radiological Studies: Objective measurements from phantom images (e.g., presence/absence of artifacts, radiation dose measurements, image quality metrics) compared to baselines (without Comfort Cover) and regulatory/IEC standards.
  • Usability Studies: User feedback (radiologists, technologists, patients) on comfort and impact on positioning.

8. The sample size for the training set
Not applicable. This device is not an AI/machine learning model and does not have a "training set."

9. How the ground truth for the training set was established
Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, with flowing lines representing hair or clothing.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 15, 2016

Precision Dynamics Corporation % Mr. Dave Yungvirt CEO Third Party Review Group, LLC The Old Station House 24 Lackawanna Place MILLBURN NJ 07041

Re: K161920

Trade/Device Name: Comfort Cover Regulation Number: 21 CFR 892.1710 Regulation Name: Mammographic x-ray system Regulatory Class: II Product Code: IZH Dated: July 10, 2016 Received: July 13, 2016

Dear Mr. Yungvirt:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Michael D.'Hara

Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health For

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known)

K161920

Device Name Comfort Cover

Indications for Use (Describe)

The indications for use of the product is to remove the cold from the bucky by placing this sheet between patient's breast and the image receptor plate allowing the patient to feel more comfortable and less cold without interfering with image quality.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for PDC, which is a Brady business. The logo features a series of vertical bars in shades of blue and green, followed by the letters "pdc" in a bold, blue font. Below the letters, the words "A Brady Business" are written in a smaller font, followed by the Brady logo.

27770 N. Entertainment Dr. Ste. 200 Valencia, CA 91355 800.847.0670 818.897.1111 Email info@pdcorp.com

pdcorp.com

SECTION 5: 510(k) SUMMARY

The following information is provided in accordance with 21 CFR 807.92, for the Traditional Premarket 510(k) Summary:

(1) SUBMITTER INFORMATION

Company Name and Address:

Contact Name: Contact Telephone: Contact E-mail: Date Prepared:

(2) DEVICE INFORMATION

Trade Name: Name of Device: Common Name: Classification Name: Regulatory Classification: Regulation Number: Product Code:

Precision Dynamics Corporation (PDC) 27770 N. Entertainment Drive, Suite 200, Valencia, California, 91355 Swarna Mukund Ph. D. (661) 481-8857 swarna_mukund@bradycorp.com June 10, 2016

Comfort Cover™ Comfort Cover Mammography Cover System, X-ray, mammographic Class II 892.1710 IZH

(3) PREDICATE DEVICE

K073262, Bella, Bella Blankets: Classification Name: Regulatory Classification: Regulation Number: Product Code:

Beekley Corporation System, X-ray, mammographic Class II 892.1710 IZH

(4) DEVICE DESCRIPTION

Comfort Cover is a single-use disposable cover placed between the patient's breast and the image receptor plate during mammography. The purpose of this product is to allow the patient to feel more comfortable and less cold during the mammographic examination without interfering with image quality. Comfort Cover will allow for artifact-free images in both Field Digital Mammography (FFDM) and Digital Breast Tomosynthesis (DBT), patient radiation dose is equivalent to that when using no Comfort Cover and the Image quality is equivalent to images exposed without Comfort Cover.

Comfort Cover is designed to fit small and large image receptor plates of all leading mammographic equipments. The product is constructed of a fabric material with a purple printed design and has adhesive backing with a paper liner. Both the material and adhesive backing of the product do not affect imaging.

{4}------------------------------------------------

The product is directly applied wrinkle free to the image receptor plate of the mammographic machine by the adhesive backing exposed by peeling off the liner. After the patient's mammogram, the product is peeled off the image receptor plate and disposed. As the product removes cleanly without leaving behind any residue, the image receptor plate does not need to be cleaned between Comfort Covers nor between patients.

(5) INDICATIONS FOR USE

The indications for use of the product is to remove the cold from the bucky by placing this sheet between patient's breast and the image receptor plate allowing the patient to feel more comfortable and less cold without interfering with image quality.

(6) COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

The following table compares the Comfort Cover to the predicate device with respect to indications for use, principles of operation, technological characteristics, materials, and performance testing. The comparison of the devices provides more detailed information regarding the basis for the determination of substantial equivalence. The subject device does not raise any new issues of safety or effectiveness based on the similarities to the predicate device.

Device CharacteristicsComfort CoverBellaSignificant Differences
Indications forUseThe indications for use of theproduct is to remove the coldfrom the bucky by placing thissheet between patient'sbreast and the imagingreceptor plate allowing thepatient to feel morecomfortable and less coldwithout interfering with imagequality.The intended use of theproduct is to remove the coldfrom the bucky by placing thissheet between patient'sbreast and the imagingreceptor plate allowing thepatient to feel morecomfortable and less coldwithout interfering with imagequality.None
TargetpopulationWomen over the age of 35having mammograms.Women over the age of 35having mammogramsNone
Where usedHospitals, radiology centersHospitals, radiology centersNone
Product Size22.9cm x 27.9cm designedfor use with 18cm x 24cmimage receptor plate29.2cm x 33.0cm designedfor use with 24cm x 30cmimage receptor plate22.9cm x 27.9cm designedfor use with 18cm x 24cmimage receptor plate29.2cm x 33.0cm designedfor use with 24cm x 30cmimage receptor plateNone
MaterialFabric (woven), adhesivebacking, purple ink.Fabric (non-woven), adhesivebacking, pink ink.Different. Safety andperformance testingresults have shownthat the slightdifferences inmaterial compositiondo not affect safetyand effectiveness

Table 5-1: Comparison of Technological Characteristics with Predicate Device

{5}------------------------------------------------

DeviceCharacteristicsComfort CoverBellaSignificantDifferences
Packaging andLabelingNot individually wrapped.Fifty covers packaged in onebox of Comfort cover.Not individually wrapped.Fifty covers packaged in onebox of Comfort cover.None
Prescription use only.Prescription use only.None
Labeling same as Bella.NANA
BiocompatibilityYesYesNone
PerformanceTestingNo visible defects.No visible defects.None
Repositionable AdhesiveRepositionable AdhesiveNone
Provides comfort andremoves cold during amammographic examinationwithout affecting quality of theimages.Provides comfort andremoves cold during amammographic examinationwithout affecting quality of theimages.None
Allows for artifact-free imagesand image quality equivalentto images exposed withoutComfort Cover.Not tested side-by-side withComfort Cover. Bella testingresults can be found in their510k (K073262).None
Removes cleanly from themammographic equipmentwithout leaving behind anyresidue.Removes cleanly from themammographic equipmentwithout leaving behind anyresidue.None
Printed ink does not smear.Printed ink does not smear.None
Remains where placed.Remains where placed.None
SterilityNon-sterileNon-sterileNone
DispensingmethodDispenser BoxDispenser BoxNone
EnvironmentCompatibilityDisposable, Non-recyclableDisposable, Non-recyclableNone
RadiationSafetyDoes not image, no artifacts,quality of images notaffected, radiolucent.Not tested side by side withComfort Cover. Bella testingresults can be found in their510k (K073262).None

(7) PERFORMANCE DATA

As part of demonstrating safety and effectiveness of Comfort Cover and in showing substantial equivalence to the predicate device, Precision Dynamics Corporation completed a number of non-clinical performance tests. The Comfort Cover met all the requirements for overall design, biocompatibility, and radiologic safety confirming that the design output meets the design inputs and specifications for the device.

Summary and Conclusions of Biocompatibility Testing

Biocompatibility testing was conducted in accordance with ISO 10993-5, Biological evaluation of medical devices - Part 5: Tests for in vitro Cytotoxicity, and ISO 10993-10, Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization.

The subject device passed all of the above tests.

{6}------------------------------------------------

Summary and Conclusions of Comparative Testing

The following mechanical tests were conducted on both the subject and the predicate devices.

  • Visual and Dimensional examination ●
  • Thermal Conductivity Test
    • This testing was performed per ASTM D5470-12 which is a test method for O measurement of thermal conductivity of materials.
  • Ink Smear Test
    • This test was performed as per ASTM D5264-98 which is a standard test method o for abrasion resistance of printed materials by the Sutherland Rub Tester to test if the ink from the printed graphics smears.
  • . Peel Adhesion Test
    • This test was performed as per ASTM D6252-98 which is a standard test method O for peel adhesion of pressure sensitive label Stocks at a 90° Angle to test the peel adhesion of the device.
  • Removes Cleanly Test
    • o This test was an in-house developed method to test if the product removes cleanly from the mammographic equipment without leaving behind any residue.

The subject device passed all of the above tests.

Summary and Conclusions of Radiological Studies

Phantom image testing was performed as per MQSA regulations and with relevant International Electro technical Commission (IEC) standards (62220-1-2:2007 and 61223-3-2) to verify that,

  • Comfort Cover will allow for artifact-free images in both FFDM and DBT.
  • . Patient radiation dose is equivalent to that when using no Comfort Cover.
  • . Image quality is equivalent to images exposed when using no Comfort Cover.

The subject device passed all of the above tests.

Summary and Conclusions of Usability Studies

Usability studies included radiologists, radiology technologists and patients testing all of the user need requirements. The Comfort Cover fulfilled its design requirements of making the mammography examination more comfortable and did not negatively impact patient positioning.

Summary of Animal Studies

No animal testing was necessary.

Summary of Clinical Studies

No clinical testing was necessary.

{7}------------------------------------------------

8) CONCLUSIONS

The subject device Comfort Cover and the predicate device Bella have the same technological characteristics and intended use.

Both devices have similar design of having rectangular shape with adhesive backing which allows easy application and removal from the image receptor plate and with floral graphics printed on the top of the sheet using FDA approved ink. Both devices are presented as single use, non-sterile and non-compressible.

Both devices remove the cold from the image receptor plate of the mammographic equipment and provide comfort during routine mammographic examinations without affecting image quality. Both devices do not negatively affect patient positioning.

Minor difference in the Comfort Cover material as compared to the predicate device. Bella does not affect safety and effectiveness of the device when used as labeled as shown by safety and efficacy studies. Therefore, Precision Dynamics Corporation believes that Comfort Cover is substantially equivalent to the existing legally marketed predicate device, Bella.

§ 892.1710 Mammographic x-ray system.

(a)
Identification. A mammographic x-ray system is a device intended to be used to produce radiographs of the breast. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.