K Number
K161888

Validate with FDA (Live)

Manufacturer
Date Cleared
2016-09-12

(63 days)

Product Code
Regulation Number
888.3080
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

JULIET® LL Lumbar Interbody Device is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These spinal implants are to be used with autogenous bone graft. JULIET® LL Lumbar Interbody Device is to be used with supplemental fixation. Patients should have at least six (6) months of non-operative treatment prior to treatment

Device Description

The JULIETLL Titanium Lateral Lumbar cages are rectangle-shaped intervertebral body fusion devices with a central cavity that can be filled with bone graft (autograft) to facilitate fusion. The JULIETLL intervertebral body fusion spacer comes in various sizes and footprints in order to fulfill surgeons' needs and accommodate different patient anatomies.
The JULIETLL system is made of Titanium Ti 6Al-4V ELI conforming to ASTM F136 and ISO 5832-3.
The JULIETLL Titanium Lateral Lumbar cages are supplied with dedicated surgical instruments (reusable - provided non sterile).

AI/ML Overview

This document describes the FDA 510(k) clearance for the JULIET® LL Lateral Lumbar Cage, an intervertebral body fusion device. The acceptance criteria and supporting studies are primarily related to the mechanical performance and biocompatibility of the device, rather than diagnostic accuracy as would be presented for an AI/ML medical device.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific Test/StandardAcceptance CriteriaReported Device PerformanceStudy Type
Mechanical PerformanceStatic axial compression (ASTM F2077)Not explicitly stated (implied to be equivalent to predicate)Not explicitly stated (implied to be equivalent to predicate)Non-clinical (bench testing)
Mechanical PerformanceStatic shear compression (ASTM F2077)Not explicitly stated (implied to be equivalent to predicate)Not explicitly stated (implied to be equivalent to predicate)Non-clinical (bench testing)
Mechanical PerformanceSubsidence testing (ASTM F2267)Not explicitly stated (implied to be equivalent to predicate)Not explicitly stated (implied to be equivalent to predicate)Non-clinical (bench testing)
BiocompatibilityBacterial Endotoxin Testing (USP <85>)Endotoxin limit of 20 EU / deviceNot explicitly stated (implied to meet 20 EU/device)Non-clinical (laboratory testing)
Design & Material EquivalenceSimilar material (Titanium Ti 6Al-4V ELI conforming to ASTM F136 and ISO 5832-3)Equivalent to predicateMeets ASTM F136 and ISO 5832-3Design/Material Comparison
Intended Use EquivalenceSame indicated useEquivalent to predicateSame indications for useClinical/Regulatory Comparison

2. Sample Size Used for the Test Set and Data Provenance

This document describes a medical device (spinal implant), not a diagnostic AI/ML system. Therefore, the concepts of "test set" and "data provenance" in the context of clinical images or patient data are not directly applicable.

  • Mechanical Testing: For mechanical tests (static axial compression, static shear compression, subsidence testing), the sample size would refer to the number of physical devices tested. This information is not explicitly stated in the provided document. The tests were performed on the predicate device (JULIET®LL Titanium Lateral Lumbar cages, K141135), and a rationale of "verification activity and validation activity" was used to demonstrate equivalence for the modified devices.
  • Bacterial Endotoxin Testing: The sample size would refer to the number of devices or material samples tested for endotoxins. This information is not explicitly stated.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

Not applicable for a physical medical device clearance based on mechanical testing and biocompatibility. Ground truth in this context is established by adherence to engineering standards and material specifications, not by expert interpretation of data.

4. Adjudication Method for the Test Set

Not applicable for a physical medical device clearance. Adjudication methods are typically used for establishing ground truth in clinical studies involving interpretation (e.g., by radiologists).

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

No, an MRMC comparative effectiveness study was not conducted. This type of study is relevant for evaluating the impact of AI on human reader performance in diagnostic tasks, which is not the subject of this 510(k) clearance for a spinal implant.

6. If a Standalone Study (i.e., algorithm only without human-in-the-loop performance) was done

No, a standalone study in the context of an AI algorithm was not done. This 510(k) is for a physical medical device. The "standalone" performance here refers to the device's mechanical integrity and biocompatibility, which were evaluated through non-clinical bench testing.

7. The Type of Ground Truth Used

The ground truth for this device clearance is based on:

  • Engineering Standards: Adherence to established ASTM (American Society for Testing and Materials) and ISO (International Organization for Standardization) standards for mechanical testing (ASTM F2077, ASTM F2267) and material specification (ASTM F136, ISO 5832-3).
  • Biocompatibility Standards: Adherence to USP (United States Pharmacopeia) standard for bacterial endotoxin testing (USP <85>).
  • Predicate Device Equivalence: The primary ground truth for the modified device is its substantial equivalence to the performance and safety of the previously cleared predicate device (JULIET® LL, K141135).

8. The Sample Size for the Training Set

Not applicable. There is no AI model or "training set" in the context of this traditional medical device clearance. The "training" for such devices involves adherence to manufacturing processes and quality control systems.

9. How the Ground Truth for the Training Set was Established

Not applicable. As there is no "training set," there is no ground truth established in this manner for the submitted device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a circular logo for the U.S. Department of Health & Human Services. The logo features the department's name encircling an abstract symbol. The symbol is a stylized representation of a human figure, possibly suggesting care and well-being. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 12, 2016

SPINEART Mr. Franck Pennesi Director of Industry & Quality International Center Cointrin 20 Route De Pré-bois CP1813 1215 Geneva, 15 SWITZERLAND

Re: K161888

Trade/Device Name: JULIET® LL Lateral Lumbar Cage Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: August 12, 2016 Received: August 15, 2016

Dear Mr. Pennesi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling Parts 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

K161888

Page 1 of 1

510(k) Number (if known) K161888

Device Name JULIET® LL Lateral Lumbar Cage

Indications for Use (Describe)

JULIET®LL Lumbar Interbody Device is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These spinal implants are to be used with autogenous bone graft. JULIET®LL Lumbar Interbody Device is to be used with supplemental fixation. Patients should have at least six (6) months of non-operative treatment with an intervertebral cage.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for Spineart. The logo features the word "spineart" in a stylized font, with the "ea" in a lighter shade of purple. Above the word is a purple butterfly.

510(k) SUMMARY

Submitted bySPINEART
International Center Cointrin
20 route de pré-bois
CP1813
1215 GENEVA 15
SWITZERLAND
ContactsFranck PENNESI Director of Industry & Quality
Phone : +41 22 799 40 25 Fax : +41 22 799 40 26
Mail : fpennesi@spineart.com
Regulatory contact : Dr Isabelle DRUBAIX (Idée Consulting)idrubaix@nordnet.fr
Date PreparedSeptember 7th 2016
Common NameIntervertebral body fusion device
Trade NameJULIET® LL Lateral Lumbar Cage
Classification NameIntervertebral Fusion Device With Bone Graft, Lumbar
ClassII
Product CodeMAX
CFR section888.3080
Device panelORTHOPEDIC
Legally marketed predicate devicePrimary predicate: JULIET® LL (K141135) manufactured by Spineart
Indications for useJULIET® LL Lumbar Interbody Device is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These spinal implants are to be used with autogenous bone graft. JULIET® LL Lumbar Interbody Device is to be used with supplemental fixation. Patients should have at least six (6) months of non-operative treatment prior to treatment

{4}------------------------------------------------

Description of the deviceThe JULIETLL Titanium Lateral Lumbar cages are rectangle-shaped intervertebral body fusion devices with a central cavity that can be filled with bone graft (autograft) to facilitate fusion. The JULIETLL intervertebral body fusion spacer comes in various sizes and footprints in order to fulfill surgeons' needs and accommodate different patient anatomies.The JULIETLL system is made of Titanium Ti 6Al-4V ELI conforming to ASTM F136 and ISO 5832-3.The JULIETLL Titanium Lateral Lumbar cages are supplied with dedicated surgical instruments (reusable - provided non sterile).
Technological CharacteristicsThe modification to JULIET®LL Titanium Lateral Lumbar cages consists of the addition of lordotic version of the range of existing spacers having a width of 17 mm, a height ranging from 08 to 16 mm, and a length ranging from 40 to 60 mm.
Discussion of TestingThe following non-clinical tests were initially conducted on JULIET®LL Titanium Lateral Lumbar cages (K141135): Static axial compression, Static shear compression according to ASTM F2077 and subsidence testing according to ASTM F2267. A rationale based on verification activity and validation activity demonstrate that the devices added to the JULIET®LL Titanium Lateral Lumbar range of products are as safe, as effective, and performs at least as safely and effectively as predicate JULIET®LL Titanium Lateral Lumbar cages (K141135). No additional testing has been performed.Bacterial endotoxin testing as specified in USP <85> standard is used for pyrogenicity testing to achieve the Endotoxin limit of 20 EU / device.
ConclusionThe extended range of JULIET®LL Titanium Lateral Lumbar cages is substantially equivalent to its predicate devices in terms of intended use, labeling, material, design, mechanical properties and function. Verification Activity and Validation Activity demonstrate that the added devices are as safe, as effective, and perform at least as safely and effectively as its predicates (K141135).

N/A