K Number
K161885

Validate with FDA (Live)

Manufacturer
Date Cleared
2016-10-05

(86 days)

Product Code
Regulation Number
878.4790
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Cellfina™ System is intended for long term improvement in the appearance of cellulite in the buttocks and thigh areas of adult females as supported by clinical data demonstrating no significant reduction in treatment benefits up to 3 years of observation.

Device Description

The Cellfina System is intended to provide precise focal release of subcutaneous tissue for improvement in the appearance of cellulite. The system consists of a sterile, single-use, disposable kit (CK1) and an electromechanical, non-sterile, 50-use motor module (CM1).

AI/ML Overview

Here's an analysis of the provided text to extract the acceptance criteria and study details for "The Cellfina System":

Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Study Endpoint/Measure)Reported Device Performance at 3 Years
Primary Endpoint: Achievement of ≥1 point average reduction in the 0-5 point Cellulite Severity Scale as determined by independent physician assessment of subject photographs.The primary endpoint was met: The average improvement was 2.0 points (p<0.0001) with the 97.5% confidence limit of a 1.7 point improvement.
Secondary Efficacy Endpoint: Improvement of one grade or more in severity (none, mild, moderate, severe) in >60% of treated subjects as determined by independent physician assessment of subject photographs.The powered secondary endpoint was met: 91.1% of treated subjects had improvement of 1 grade with a 95% confidence limit of 81.6%.
Primary Safety Endpoint: 100% of subjects free from Serious Adverse Events (SAE) directly attributable to the Cellfina System or procedure.The safety endpoint was met: 100% of subjects were free from Serious Adverse Events (SAE) directly attributable to the Cellfina System or procedure.
Additional Secondary Measure (Rate of Correct Selection): (Implicit, likely referring to identification of affected areas/response to treatment)The average rate of correct selection was 97% with a 95% confidence limit of 92.6%.
Additional Secondary Measure (Noticeable Improvement by GAIS):100% of the subjects had noticeable improvement by the GAIS.
Additional Secondary Measure (Marked Improvement or Better by GAIS):56% were characterized as having marked improvement or better.
Additional Secondary Measure (Subject Satisfaction): Percentage of subjects either satisfied or very satisfied.93% of the subjects were either satisfied or very satisfied as evaluated by a 5-point Likert type scale.
Additional Secondary Measure (Pain Level): Percentage of subjects rating pain ≤3/10.71% of subjects rated pain ≤ 3/10 at 3 days, and over 95% rated pain ≤ 3/10 thereafter. There were no subjects who reported pain at the 3-year follow-up.

Study Details

  1. Sample size used for the test set and the data provenance:

    • Sample Size: 55 subjects participated in the baseline assessment and treatment.
    • Follow-up at 3 years: 45 subjects completed the 3-year follow-up.
    • Data Provenance: The study was a "Prospective, multi-center, non-randomized open label, safety and effectiveness study." This indicates it was a prospective study, typically conducted in multiple clinical sites (country of origin is not explicitly stated, but often such trials for FDA approval are largely US-based unless otherwise specified).
    • Imputation: For the 10 subjects lost to follow-up at 3 years, "last observation carried forward (LOCF)" was used for both primary and 1st secondary endpoints, and "multiple imputation" for the primary effectiveness endpoint.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • The primary and secondary endpoints were determined by "independent physician assessment of subject photographs." The exact number of independent physicians (experts) is not specified in the provided text. Their specific qualifications (e.g., "radiologist with 10 years of experience") are also not detailed, beyond being "independent physicians."
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • The adjudication method is not explicitly described in the provided text. It only mentions "independent physician assessment." This could imply a single independent assessor or a consensus method, but the specific process (e.g., if multiple physicians' assessments were averaged or if a tie-breaking mechanism was used) is not detailed.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC comparative effectiveness study was not done. This study evaluates a device (The Cellfina System) for treating cellulite, not an AI or imaging diagnostic tool. Therefore, the concept of "human readers improve with AI vs without AI assistance" is not applicable to this submission.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • No, this device is a powered surgical instrument, not an algorithm. The performance detailed here is for the physical device and procedure, with human physicians performing the assessment.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The ground truth for effectiveness was established primarily through expert assessment/consensus based on photographs ("independent physician assessment of subject photographs").
    • Safety was based on outcomes data (absence of Serious Adverse Events reported).
    • Subject satisfaction and pain levels were based on patient-reported outcomes.
  7. The sample size for the training set:

    • Not applicable. This study is for a medical device (a powered surgical instrument), not a machine learning model. There is no concept of a "training set" in this context. The clinical study described served as the validation and effectiveness study for the device itself.
  8. How the ground truth for the training set was established:

    • Not applicable, as there is no training set for a medical device of this nature.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing to the right, stacked on top of each other. The profiles are black and are set against a white background. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the image.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 5, 2016

Ulthera, Inc. Ms. Suzon Lommel Vice President, Regulatory & Quality Affairs 1840 S Stapley Drive. Suite 200 Mesa, Arizona 85204

Re: K161885 Trade/Device Name: The Cellfina System Regulation Number: 21 CFR 878.4790 Regulation Name: Powered Surgical Instrument for Improvement in the Appearance of Cellulite Regulatory Class: Class II Product Code: OUP Dated: July 8, 2016 Received: July 11, 2016

Dear Ms. Lommel:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerelv.

Jennifer R. Stevenson -A

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) TBD KJ 67886

Device Name Cellfina™ System

Indications for Use (Describe)

The Cellfina™ System is intended for long term improvement in the appearance of cellulite in the buttocks and thigh areas of adult females as supported by clinical data demonstrating no significant reduction in treatment benefits up to 3 years of observation.

Type of Use (Select one or both, as applicable)
Remediation Use (Part 21 CFR 201 Subpart D)
Source Tissue Contracting Use (21 CFR 361 Subpart C)

Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

5. 510(k) Summary

This 510(k) Summary for the Cellfina System is submitted in accordance with the requirements of the Safe Medical Device Act (SMDA) of 1990 and follows the Office of Device Evaluation (ODE) guidance concerning the organization and content of a 510(k) summary.

Applicant:Ulthera, Inc.
Address:1840 South Stapley DriveSuite 200Mesa, AZ 85204
Contact Person:Kristen DeJeu, Regulatory Affairs Specialist
Telephone:(408) 822-6507
Fax:(480) 619-4071
Submission Date:July 08, 2016
Device Trade Name:The Cellfina System
Common Name:Powered surgical instrument for improvement in the appearanceof cellulite.
Classification:Regulatory Class II
Classification Name:Powered surgical instrument for improvement in the appearanceof cellulite.
Regulation Number:21CFR 878.4790
Product Code:OUP
Legally Marketed Predicate:The Cellfina System, 510(k): K153677
Applicable Guidance:The following guidance special controls are applicable to theCellfina System:

Class II (special controls). The special controls for this device are:

(1) Non-clinical testing must be performed to demonstrate that the device meets all design specifications and performance requirements, and to demonstrate durability and mechanical integrity of the device.

(2) In vivo evaluation of the device must demonstrate device performance, including the safety of the release methodology and blood loss at the treatment sites.

(3) All elements of the device that may contact the patient must be demonstrated to be biocompatible.

(4) Electrical safety and electromagnetic compatibility of the device must be demonstrated.

Image /page/3/Picture/9 description: The image shows the word "ulthera" in a sans-serif font, with the "u" and "l" in a light blue color and the rest of the letters in a darker shade of blue. To the right of the word, there are three yellow circles arranged horizontally, followed by the trademark symbol. The overall design is clean and modern.

{4}------------------------------------------------

(5) The labeling must include a summary of in vivo evaluation data and all the device specific warnings, precautions, and/or contraindications.

(6) Sterility and shelf-life testing for the device must demonstrate the sterility of patient contacting components and the shelf life of these components.

  • Device Description: The Cellfina System is intended to provide precise focal release of subcutaneous tissue for improvement in the appearance of cellulite. The system consists of a sterile, single-use, disposable kit (CK1) and an electromechanical, non-sterile, 50-use motor module (CM1).
  • Indications for Use: The Cellfina System is intended for long term improvement in the appearance of cellulite in the buttocks and thigh areas of adult females as supported by clinical data demonstrating no significant reduction in treatment benefits up to 3 year of observation.

Substantial Equivalence Comparison:

The subject device and the predicate device are similar in the following ways: same regulation number and product code, anatomical use area, clinical use group (doctor's office, clinic), energy type with the same mechanism of action, packaging materials and configuration are unchanged, there are no differences in patient contacting materials, design of the CK1 Disposable Kit is the same, and the sterilization SAL is the same (EO, SAL 10th. Both devices are compliant to IEC 60601 standards and EMC standards.

The subject device differs from the predicate device in that the subject device is intended for 3 year efficacy whereas the predicate is intended for 2 year efficacy. In addition, the subject device's CK1 Disposable Kit has a shelf life of 24 months, whereas the predicate device had a shelf life of 12 months. The subject device CK1 is now qualified for a fully validated sterilization cycle, whereas the predicate went through a sterilization batch release process. The Subject Motor Module is intended for use up to 50 uses, whereas the predicate was intended for use up to 15 uses. The subject Motor Module is now validated for two types of cleaning methods (IPA based wipes and Chloride based wipes). The housing material for the enclosure and power button also changed from ABS and Elastomer to Makroblend and Wacker Elastosil, respectively in order to include a more robust material type that would withstand the new cleaning method. These are non-patient contacting materials and therefore additional biocompatibility testing was not necessary. The predicate device was only validated for one type of cleaning (Chloride based wipes). The subject device Motor Module includes minor tolerance changes to allow for injection molding, whereas the predicate was machined. Minor changes to location of epoxy, simplification of design to remove metal pin and using screws instead of heat staking were made to accommodate a 50 use "use life". The subject device will have a separate IFU for the CK1 and CM1, whereas the predicate utilized identical IFUs with repetitive information.

Non-clinical Performance Data:

Bench testing was performed to verify that the modifications to the subject device do not impact the safety or effectiveness of the device. Retesting for EMC and IEC 60601 was completed for the motor module to ensure that the changes in design would have no impact on safety or efficacy. Shelf life testing and use life testing were conducted on the CK1 and CM1 respectively for the changes in shelf and use life, which demonstrated full functionality. Cleaning and disinfection testing was conducted for the addition of the cleaning method for the motor module

Image /page/4/Picture/10 description: The image features the logo for Ulthera, a company specializing in medical aesthetic devices. The logo consists of the word "ulthera" in a lowercase, sans-serif font, with the "u" and "l" in a light blue color and the remaining letters in a darker shade of blue. Above the "thera" portion of the word, there are three orange circles arranged horizontally, adding a visual element to the brand identity.

{5}------------------------------------------------

with no impact to device performance. Sterilization was completed for the CK1 disposable kit in order to remove the requirement for batch release testing. The SAL level and sterilization method (EO) remains the same.

Clinical Performance Data (In-Vivo):

The safety and effectiveness of the Cellfina System was evaluated in a pivotal clinical study conducted under IDE G120116. The IDE was modified to allow for extended follow-up clinical performance data to be gathered for safety and efficacy at 3 years. No additional treatments were performed.

During the course of the 3 year Clinical study, 10 of the 55 subjects were lost to follow up. The reason for loss to follow up ranged from inability to contact the subject after multiple attempts to the subject removing herself from the study. Imputation analyses were performed for the 10 missing subjects from the 3 year report and carried out using the last observation carried forward (LOCF) for both the primary and 1st secondary endpoints using a paired t-test as well as multiple imputation for the primary effectiveness endpoint only, which met the study endpoint. The overall study success was met, despite the reduction in follow up participation at 3 years with a clinically significant, long term improvement in the appearance of cellulite in the treated subjects. The table below is based on the LOCF data and accounts for the 10 missing subjects:

Clinical Study Design Characteristics
Study Design:Prospective, multi-center, non-randomized open label, safety and effectiveness study
Sample Size55 Subjects participated in the baseline assessment and treatment
Follow up (N)3 Day (55), 14 Day (54), 1 Month(54), 3Month (55), 6 Month (52), 1 year (50), 2 year (52), 3 year (45)
Inclusion CriteriaFemale between the ages of 18 and 55, BMI less than 35
Exclusion CriteriaNo prior drug treatment for contouring or cellulite during prior 90 days; no prior liposuction of thighs or buttocks, subject has had >10 % weight loss in prior 6 months; subject has known difficulty with local anesthesia, subject is pregnant, other minor criteria
Primary Endpoint result at 3 yearsThe primary endpoint was met: achievement of ≥1 point average reduction in the 0-5 point Cellulite Severity Scale as determined by independent physician assessment of subject photographs taken before and (3 years) after treatment.• The average improvement was 2.0 points (p<0.0001) with the 97.5% confidence limit of a 1.7 point improvement
Secondary Efficacy result at 3 yearsThe powered secondary endpoint was met: improvement of one grade or more in severity (none, mild, moderate, severe) in >60% of treated subjects as determined by independent physician assessment of subject photographs taken before and 3 years after treatment.• 91.1% of treated subjects had improvement of 1 grade with a 95% confidence limit of 81.6%.
Primary Safety ResultThe safety endpoint was met: 100% of subjects were free from Serious Adverse Events (SAE) directly attributable to the Cellfina System or procedure at 3 years
AdditionalAll additional secondary measures were also achieved or acceptable at the 3

Image /page/5/Picture/6 description: The image shows the word "ulthera" in a sans-serif font. The letters are light blue, and there are three yellow dots above the "e" in "ulthera". To the right of the word is a small "TM" symbol, indicating that the word is a trademarked brand name. The overall impression is of a clean and modern logo.

{6}------------------------------------------------

Measuresyear time point:
The average rate of correct selection was 97% with a 95% confidence limit of 92.6%. 100% of the subjects had noticeable improvement by the GAIS and 56% were characterized as having marked improvement or better. 93% of the subjects were either satisfied or very satisfied as evaluated by a 5 point Likert type scale at the 3 year follow-up. 71% of subjects rated pain $\leq$ 3/10 at 3 days, and over 95% rated pain $\leq$ 3/10 thereafter. There were no subjects who reported pain at the 3 year follow-up.

Conclusion:

This Traditional 510(k) Premarket Notification is for an expansion to the indications for use (from 2 years to 3 years). Additionally, the motor module (CM1) use life has been extended from 15 uses to 50 uses and the shelf life of the disposable kit (CK1) has been extended from a 12 month shelf life to a 24 month shelf life. Labeling changes have also been implemented to reflect the above described changes. Additional testing was performed for compliance to Amendment 1:2012 (Edition 3.1) of 60601-1 and EMC 60601-1-2:2007 3rd Edition.

These changes do not pose any new questions of safety, product output or efficacy. As demonstrated through bench testing and clinical data, the subject device Cellfina System is as safe, as effective, and performs as well as the legally marketed predicate device (K153677), and as such, is substantially equivalent.

Image /page/6/Picture/5 description: The image shows the logo for Ulthera. The logo is in a sans-serif font and is a light blue color. There are three yellow dots to the right of the word "ulthera". The letters are all lowercase.

§ 878.4790 Powered surgical instrument for improvement in the appearance of cellulite.

(a)
Identification. A powered surgical instrument for improvement in the appearance of cellulite is a prescription device that is used for the controlled release of subcutaneous tissue for improvement in the appearance of cellulite. The device consists of a cutting tool powered by a motor and a means for instrument guidance to control the areas of subcutaneous tissue cutting underneath the cellulite depressions or dimples.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Non-clinical testing must be performed to demonstrate that the device meets all design specifications and performance requirements, and to demonstrate durability and mechanical integrity of the device.
(2) In vivo evaluation of the device must demonstrate device performance, including the safety of the release methodology and blood loss at the treatment sites.
(3) All elements of the device that may contact the patient must be demonstrated to be biocompatible.
(4) Electrical safety and electromagnetic compatibility of the device must be demonstrated.
(5) The labeling must include a summary of in vivo evaluation data and all the device specific warnings, precautions, and/or contraindications.
(6) Sterility and shelf-life testing for the device must demonstrate the sterility of patient contacting components and the shelf life of these components.