K Number
K161873

Validate with FDA (Live)

Manufacturer
Date Cleared
2016-11-22

(138 days)

Product Code
Regulation Number
870.3680
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Pacel™ Flow Directed Pacing Catheters are indicated for use in temporary, transvenous, right ventricular pacing.

Device Description

This catheter is designed to establish temporary right ventricular pacing with or without fluoroscopy guidance for placement. Catheters are depth marked as an aid in catheter placement under fluoroscopy. Electrodes at the catheter tip enable ECG monitoring and bipolar pacing. An inflatable balloon, located between two platinum electrodes, allows for flow directed positioning of the catheter in the right ventricle without fluoroscopy guidance. Once inflated, the balloon aids the operator in crossing the tricuspid valve. After the catheter has crossed the valve, the balloon is deflated and the catheter is advanced to the right ventricular apex in the usual manner. Electrograms are monitored to verify proper positioning in the right ventricular apex and pacing thresholds are obtained to confirm both proper location and reliable pacing. The catheters are available in two curve styles. The distal end of the catheter includes a platinum electrical connector ring and a tip electrode. The proximal end of the catheter includes a two pin electrical connector. The electrical signals can be transmitted to external equipment or from external equipment to the heart.

AI/ML Overview

This document describes the 510(k) summary for the Pacel™ Flow Directed Pacing Catheters, which pertains to a change in the adhesive used to bond the latex balloon to the catheter and seal around the metal electrode ring edges. The submission aims to demonstrate substantial equivalence to a predicate device (Pacel™ Flow Directed Pacing Catheters K914185).

The information provided is not for a device that uses AI or performs any kind of diagnostic or analytical function that would require a study with acceptance criteria related to accuracy, sensitivity, specificity, or human reader performance. This submission is for a physical medical device (a catheter) and the studies are primarily engineering design verification tests and biocompatibility studies related to a material change (adhesive).

Therefore, I cannot fulfill the request to provide information about the acceptance criteria and study proving the device meets those criteria in the context of AI/diagnostic device performance. The provided text does not contain any of the following details required:

  • Acceptance Criteria Table (related to performance metrics like accuracy, sensitivity, specificity): Not applicable for this type of device. The document mentions "passed all acceptance criteria per the DV requirements" for design verification, but these are related to physical properties of the catheter (e.g., surface appearance, inflation strength, inflation decay).
  • Sample size for test set and data provenance: No "test set" in the context of image analysis or algorithm performance. The design verification and animal study would use sample sizes relevant to engineering and animal studies, not AI validation.
  • Number of experts for ground truth, qualifications, adjudication method: This is irrelevant for a physical medical device and adhesive change.
  • MRMC comparative effectiveness study: Not applicable as there is no "human reader" component or effect size from AI assistance.
  • Standalone (algorithm only) performance: Not applicable as there is no algorithm.
  • Type of ground truth: Not applicable in the context of diagnostic accuracy.
  • Training set size and ground truth establishment for training set: Not applicable as there is no AI algorithm being trained.

The study described is focused on the physical and biological properties of the catheter due to a change in adhesive.

Summary of Stated Device Performance (from the document):

The document reports the following for the non-clinical testing:

  1. Design Verification:

    • Acceptance Criteria: "show that the proposed device will meet Product Specification through the labelled Shelf Life of the device (T=18 mo)." and "passed all acceptance criteria per the DV requirements."
    • Reported Performance:
      • Surface Appearance: Adhesive on the balloon and electrodes. Results: Conforming
      • Balloon Repeat Inflation Strength. Results: Conforming
      • Inflation Decay. Results: Conforming
    • Specifics on sample size or individual performance values are not detailed in this summary.
  2. Biocompatibility:

    • Acceptance Criteria: "demonstrated an acceptable biocompatibility profile." and "passed all acceptance criteria."
    • Reported Performance: "Biocompatibility testing was performed in accordance with ISO 10993. The proposed adhesive, Loctite 4311, used to secure the balloon to the catheter shaft demonstrated an acceptable biocompatibility profile."
    • Specifics on sample size or individual performance values are not detailed in this summary.
  3. Chronic GLP Animal Study:

    • Acceptance Criteria: To demonstrate safety comparable to the control device, and that both devices are "substantially equivalent regarding clinical pathology, thrombogenicity and pathology/histopathology."
    • Reported Performance: "Both the test and control devices were considered substantially equivalent regarding clinical pathology, thrombogenicity and pathology/histopathology. The use of both devices was well tolerated. This data demonstrates that the proposed Pacel Flow Direct catheter's safety is comparable to the control device."
    • Specifics on sample size or individual performance values are not detailed in this summary.

In conclusion, the document's purpose is to demonstrate substantial equivalence of a physical medical device (catheter) after a manufacturing material change, not the performance of an AI/diagnostic algorithm.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 22, 2016

St. Jude Medical Blair Schwartz Regulatory Affairs Specialist 14901 Deveau Place Minnetonka, Minnesota 55345

Re: K161873

Trade/Device Name: Pacel™ Flow Directed Pacing Catheter Regulation Number: 21 CFR 870.3680 Regulation Name: Cardiovascular Permanent or Temporary Pacemaker Electrode Regulatory Class: Class II Product Code: LDF Dated: October 18, 2016 Received: October 19, 2016

Dear Blair Schwartz:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

Mude Jellman
for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K161873

Device Name

Pacel™ Flow Directed Pacing Catheters

Indications for Use (Describe)

Pacel™ Flow Directed Pacing Catheters are indicated for use in temporary, transvenous, right ventricular pacing.

Type of Use (Select one or both, as applicable)

☒ Exemption Use (Part 21 CFR 601.21, Subpart E)☐ Compassionate Use (21 CFR 601.21, Subpart E)
-------------------------------------------------------------------------------------------------------------------------------------------------

|× | Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains the logo for St. Jude Medical. The logo consists of a grid of green squares with one square rotated, followed by the text "ST. JUDE MEDICAL" in a serif font. The text is aligned to the right of the grid.

510(K) SUMMARY

The 510(k) Summary of the Pacel™ Flow Directed Pacing Catheters, per 21 CFR 807.92, is provided below.

510(k) Summary
510(k) NumberK161873
Submitter Information:
Date Prepared:July 06, 2016
Submitter Name &Address:St. Jude Medical14901 DeVeau PlaceMinnetonka, MN 55345Establishment Registration Number: 2182269
Contact Person:Blair SchwartzRegulatory Affairs Specialist5050 Nathan Lane NPlymouth MN 55426Phone (651) 756-2706BSchwartz@sjm.com
Device Information:
Trade Name:Pacel™ Flow Directed Pacing Catheters
Common Name:Electrode, Pacemaker, Temporary
ClassII
Classification Name:21CFR § 870.3680 – Cardiovascular permanent or temporarypacemaker electrode
Predicate Device:Pacel™ Flow Directed Pacing Catheters (K914185)
510(k) Summary
Device Description:This catheter is designed to establish temporary right ventricular pacing with or without fluoroscopy guidance for placement. Catheters are depth marked as an aid in catheter placement under fluoroscopy. Electrodes at the catheter tip enable ECG monitoring and bipolar pacing. An inflatable balloon, located between two platinum electrodes, allows for flow directed positioning of the catheter in the right ventricle without fluoroscopy guidance. Once inflated, the balloon aids the operator in crossing the tricuspid valve. After the catheter has crossed the valve, the balloon is deflated and the catheter is advanced to the right ventricular apex in the usual manner. Electrograms are monitored to verify proper positioning in the right ventricular apex and pacing thresholds are obtained to confirm both proper location and reliable pacing.The catheters are available in two curve styles. The distal end of the catheter includes a platinum electrical connector ring and a tip electrode. The proximal end of the catheter includes a two pin electrical connector. The electrical signals can be transmitted to external equipment or from external equipment to the heart.
Intended Use:(Indications for Use)Pacelt™ Flow Directed Pacing Catheters are indicated for use in temporary, transvenous, right ventricular pacing.
Description of Change andComparison to PredicateDevicesThe Pacel™ Flow Directed Pacing Catheter that is the subject of this application remains substantially equivalent to the predicate device. There have been no changes to the packaging, sterilization, final device specifications or labeling. The proposed change is to switch the current moisture cured cyanoacrylate adhesives, Loctite 4013 which is used to bond the latex balloon to the catheter and Loctite 4031 which is used to seal/trim around the metal electrode ring edges to a visible light cured cyanoacrylate adhesive, Loctite 4311
510(k) Summary
Summary on Non-ClinicalTestingDesign Verification:Design verification testing has been completed for thePacel Flow Direct catheter to show that the proposeddevice will meet Product Specification through thelabelled Shelf Life of the device (T=18 mo). Designverification testing is complete and passed allacceptance criteria per the DV requirements.
Testing Performed:Surface Appearance: Adhesive on the balloonand electrodes.Results: Conforming Balloon Repeat Inflation Strength Results: Conforming Inflation Decay Results: Conforming
Biocompatibility:Biocompatibility testing was performed in accordancewith ISO 10993. The proposed adhesive, Loctite 4311,used to secure the balloon to the catheter shaftdemonstrated an acceptable biocompatibility profile.Biocompatibility testing passed all acceptance criteria.
Chronic GLP Animal Study:Both the test and control devices were consideredsubstantially equivalent regarding clinical pathology,thrombogenicity and pathology/histopathology. The useof both devices was well tolerated. This datademonstrates that the proposed Pacel Flow Directcatheter's safety is comparable to the control device.
510(k) Summary
Statement of EquivalenceThe Pacel TM Flow Directed Pacing Catheters that are subjectof this application have the same indications for use andtechnological characteristics as the predicate device. Based onthis and the data provided in this pre-market notification, thesubject device remains substantially equivalent to thepredicate device.

{4}------------------------------------------------

Pacel Pacing Catheter
Premarket Notification 510(k) K161873

Image /page/4/Picture/1 description: The image contains the logo for St. Jude Medical. The logo consists of a green square grid with one square rotated. To the right of the grid is the text "ST. JUDE MEDICAL" in a serif font. The text is aligned horizontally with the grid.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for St. Jude Medical. The logo consists of a green square grid on the left side and the text "St. Jude Medical" on the right side. The text is in a serif font, with "St." and "Jude" in a slightly larger font size than "Medical".

{6}------------------------------------------------

Pacel Pacing Catheter
Premarket Notification 510(k) K161873

Image /page/6/Picture/1 description: The image shows the logo for St. Jude Medical. The logo consists of a green square made up of nine smaller squares, with one of the squares rotated. To the right of the square is the text "St. Jude Medical" in a serif font.

§ 870.3680 Cardiovascular permanent or temporary pacemaker electrode.

(a)
Temporary pacemaker electrode —(1)Identification. A temporary pacemaker electrode is a device consisting of flexible insulated electrical conductors with one end connected to anexternal pacemaker pulse generator and the other end applied to the heart. The device is used to transmit a pacing electrical stimulus from the pulse generator to the heart and/or to transmit the electrical signal of the heart to the pulse generator.(2)
Classification. Class II (performance standards).(b)
Permanent pacemaker electrode —(1)Identification. A permanent pacemaker electrode is a device consisting of flexible insulated electrical conductors with one end connected to an implantable pacemaker pulse generator and the other end applied to the heart. The device is used to transmit a pacing electrical stimulus from the pulse generator to the heart and/or to transmit the electrical signal of the heart to the pulse generator.(2)
Classification. Class III (premarket approval).(c)
Date PMA or notice of completion of PDP is required. A PMA or notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before October 4, 2012, for any permanent pacemaker electrode device that was in commercial distribution before May 28, 1976, or that has, on or before October 4, 2012, been found to be substantially equivalent to any permanent pacemaker electrode device that was in commercial distribution before May 28, 1976. Any other pacemaker repair or replacement material device shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.