K Number
K161814

Validate with FDA (Live)

Manufacturer
Date Cleared
2016-09-27

(88 days)

Product Code
Regulation Number
866.3980
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Solana® Influenza A+B Assay is a qualitative in vitro diagnostic test for the detection and differentiation of influenza A and influenza B viral RNA in nasal and nasopharyngeal swabs from patients with signs and symptoms of respiratory infection. This test is intended for use as an aid in the differential diagnosis of influenza A and influenza B viral infections in humans in conjunction with clinical and epidemiological risk factors. The assay does not detect the presence of influenza C virus.

Negative results do not preclude influenza virus infection and should not be used as the sole basis for diagnosis, treatment or other patient management decisions.

Performance characteristics for the Solana® Influenza A+B Assay were established during the spring of 2016 when influenza A/H3 and 2009 H1N1 influenza were the predominant influenza viruses in circulation. When other influenza viruses are emerging, performance characteristics may vary.

If infection with a novel influenza virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent Influenza viruses and sent to state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.

Device Description

The Solana Influenza A+B Assay amplifies and detects viral RNA present in viral transport media containing nasopharyngeal or nasal swab specimens obtained from symptomatic patients.

The assay consists of two major steps: 1) specimen preparation, and 2) amplification and detection of target sequences specific to influenza A and/or influenza B using isothermal Reverse Transcriptase - Helicase-Dependent Amplification (RT-HDA) in the presence of target-specific fluorescence probes.

A patient nasal or nasopharyngeal swab specimen in viral transport media is transferred to a Process Buffer Tube, subjected to heat treatment at 95°C for 5 minutes and mixed. The processed sample is transferred to a Reaction Tube. The Reaction Tube contains lyophilized RT-HDA reagents, dNTPs, primers and probes. Once rehydrated with the processed sample, the Reaction Tube is placed in Solana for amplification and detection of influenza A and influenza B-specific target sequences. In Solana, the target sequences are amplified by influenza A and influenza B specific primers and detected by influenza A and influenza B specific fluorescence probes, respectively. A competitive process control (PRC) is included in the Process Buffer Tube to monitor sample processing, inhibitory substances in clinical samples, reagent failure or device failure. The PRC target is amplified by influenza B specific primers and detected by a PRC specific fluorescence probe.

The two target probes and PRC probe are labeled with a quencher on one end and a fluorophore on the other end. In addition, the two target probes and PRC probe have one or more bases that are comprised of ribonucleic acid. Upon annealing to influenza A, influenza B or PRC amplicons, the fluorescence probes are cleaved by RNaseH2 and the fluorescence signal increases due to physical separation of fluorophore from quencher. Solana measures and interprets the fluorescent signal, using on-board method-specific algorithms. Solana then reports the test results to the user on its display screen, and it can print out the results via an integrated printer.

AI/ML Overview

Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

Acceptance Criteria and Device Performance for Solana® Influenza A+B Assay

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state "acceptance criteria" as a separate, quantitative target. Instead, it presents the results of clinical studies comparing the Solana® Influenza A+B Assay to established methods (culture with DFA and DSFA, and an FDA-cleared molecular assay). The implied acceptance is based on demonstrating substantial equivalence through high agreement percentages (sensitivity/PPA and specificity/NPA).

Here's a table summarizing the primary clinical performance data:

MetricTarget (Implied Acceptance)Reported Device Performance (Influenza A)Reported Device Performance (Influenza B)
Comparison to Culture & DSFAHigh Agreement (demonstrate substantial equivalence)
Sensitivity (Influenza A)High (implied >95%)98.6% (95% CI: 96.8 to 99.4)N/A
Specificity (Influenza A)High (implied >95%)95.1% (95% CI: 93.7 to 96.3)N/A
Sensitivity (Influenza B)High (implied >95%)N/A100% (95% CI: 95.7 to 100)
Specificity (Influenza B)High (implied >95%)N/A99.3% (95% CI: 98.7 to 99.6)
Comparison to FDA-Cleared Molecular AssayHigh Agreement (demonstrate substantial equivalence)
Positive Percent Agreement (PPA) (Influenza A)High (implied >95%)97.2% (95% CI: 95.0 to 98.4)N/A
Negative Percent Agreement (NPA) (Influenza A)High (implied >95%)96.7% (95% CI: 95.4 to 97.7)N/A
Positive Percent Agreement (PPA) (Influenza B)High (implied >95%)N/A100% (95% CI: 95.4 to 100)
Negative Percent Agreement (NPA) (Influenza B)High (implied >95%)N/A98.9% (95% CI: 98.2 to 99.4)

Note: The "Target (Implied Acceptance)" values are inferred based on typical expectations for diagnostic test performance in regulatory submissions aiming for substantial equivalence. The document itself does not provide explicit numerical "acceptance criteria" as a separate heading.

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set Size:
    • Clinical Study vs. Culture & DSFA: 1473 specimens initially collected, 1408 included in the final analysis after exclusions (15 contaminated/toxic in cell culture, 50 invalid in Solana Assay). (709 fresh, 699 frozen for each analysis after exclusions).
    • Clinical Study vs. FDA-Cleared Molecular Assay: 1473 specimens initially collected, 1393 included in the final analysis after exclusions (31 invalid in comparator assay, 50 invalid in Solana Assay, 1 invalid in both). (710 fresh, 683 frozen for each analysis after exclusions).
  • Data Provenance: Prospective collection from February to April 2016 at five (5) sites across the United States.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document does not specify the number of experts used to establish the ground truth or their specific qualifications (e.g., "radiologist with 10 years of experience").

For the comparison against culture and DSFA:

  • The comparator methods were "culture for influenza A and B using the R-Mix Too mixed cells and direct specimen DFA (DSFA)".
  • A specimen was recorded as positive if either comparator test was positive.
  • The interpretation of DFA results typically involves trained laboratory personnel, but details on their number or explicit qualifications are not provided.

For the comparison against an FDA-cleared molecular assay:

  • The ground truth was established by an "FDA-cleared Influenza A+B molecular assay" run according to its package insert. This implies the results of that assay served as the reference standard, rather than a panel of human experts directly interpreting the primary data.

4. Adjudication Method for the Test Set

The document describes adjudication for discordant results in the comparison studies (Solana vs. culture/DSFA and Solana vs. FDA-cleared molecular assay).

  • Discordant Analysis (Solana vs. Culture & DSFA):

    • For Influenza A:
      • 51 discordant specimens (Solana Positive/Culture and DSFA Negative) were analyzed, with 28 confirmed positive by an alternate molecular assay.
      • 5 discordant specimens (Solana Negative/Culture and DSFA Positive) were analyzed, with 2 confirmed positive by an alternate FDA cleared molecular assay.
    • For Influenza B:
      • 9 discordant specimens (Solana Positive/Culture and DSFA Negative) were analyzed, with 2 confirmed positive by an alternate FDA cleared molecular assay.
    • The term "alternate molecular assay" or "alternate FDA cleared molecular assay" indicates that a third, independent and cleared method was used to resolve discrepancies.
  • Discordant Analysis (Solana vs. FDA-Cleared Molecular Assay):

    • For Influenza A:
      • 44 total discordant specimens.
      • 33 discordant (Solana Positive/Comparator Negative), 9 of which were positive by culture/DSFA.
      • 11 discordant (Solana Negative/Comparator Positive), 2 of which were positive by culture/DSFA.
    • For Influenza B:
      • 14 total discordant specimens.
      • 14 discordant (Solana Positive/Comparator Negative), 7 of which were positive by culture/DSFA.
    • This indicates a form of 2+1 adjudication where the two primary methods (Solana and the comparator assay) were adjudicated by a third method (culture/DSFA) in cases of disagreement.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, a multi-reader multi-case (MRMC) comparative effectiveness study was NOT done. This device is an in vitro diagnostic (IVD) assay for detecting viral RNA, not an AI-assisted imaging device or a tool that aids human interpretation of complex visual data. Therefore, the concept of "human readers improve with AI vs. without AI assistance" does not apply to this type of device.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

Yes, a standalone performance study was done. The clinical performance characteristics (Sensitivity, Specificity, PPA, NPA) presented in the tables (Tables 11, 12, 13, 14) are for the Solana® Influenza A+B Assay operating as a standalone device. The assay performs "amplification and detection of target sequences... using on-board method-specific algorithms" and "reports the test results to the user on its display screen." This indicates an automated, algorithm-driven standalone performance.

7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

Two types of ground truth/comparator methods were used for the clinical studies:

  • Virological Culture with Direct Specimen Immunofluorescence Assay (DFA): This is a laboratory-based method for detecting viable virus (culture) and viral antigens (DFA) in clinical specimens. For this study, if either culture or DSFA was positive, the specimen was considered positive.
  • FDA-cleared Influenza A+B Molecular Assay: This refers to a legally marketed and established molecular diagnostic test that serves as a reference standard for detecting influenza viral RNA.

In cases of discordance, an "alternate molecular assay" or "culture/DSFA" was used for adjudication, supporting the robustness of the ground truth establishment.

8. The Sample Size for the Training Set

The document does not explicitly state a sample size for a "training set." This is likely because the Solana® Influenza A+B Assay is an RT-HDA based in vitro diagnostic test rather than a machine learning/AI model that requires a distinct training phase on clinical data. The assay's "on-board method-specific algorithms" (page 26) are likely pre-defined based on analytical validation and principles of molecular biology, not learned from a clinical training dataset in the AI sense.

The "Analytical Reactivity (Inclusivity)" study (Tables 9 & 10) can be seen as part of the developmental data, demonstrating the assay's ability to detect various strains, which is analogous to ensuring a model's robustness to variations, but it is not a "training set" in the context of supervised machine learning.

9. How the Ground Truth for the Training Set Was Established

As noted above, there's no explicitly defined "training set" in the context of machine learning. For the analytical studies (e.g., Limit of Detection, Inclusivity), the ground truth was established by:

  • Quantified viral cultures (TCID50/mL): This involves precisely measuring the concentration of infectious virus in a sample, which serves as a known, quantitative input for analytical sensitivity studies.
  • Known strains and concentrations: For inclusivity and cross-reactivity studies, specific well-characterized viral and bacterial strains at known concentrations were used as ground truth to evaluate the assay's performance.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

September 27, 2016

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Ronald H. Lollar Senior Director, Clinical and Regulatory Affairs Quidel Corporation 2005 East State Street, Suite 100 Athens, OH 45701

Re: K161814

Trade/Device Name: Solana® Influenza A+B Assay Regulation Number: 21 CFR 866.3980 Regulation Name: Respiratory viral panel multiplex nucleic acid assay Regulatory Class: Class II Product Code: OCC. OZE Dated: June 30, 2016 Received: July 1, 2016

Dear Mr. Lollar:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of

{1}------------------------------------------------

medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Steven R. Gitterman -S

for Uwe Scherf, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K161814

Device Name Solana® Influenza A+B Assay

Indications for Use (Describe)

The Solana® Influenza A+B Assay is a qualitative in vitro diagnostic test for the detection and differentiation of influenza A and influenza B viral RNA in nasal and nasopharyngeal swabs from patients with signs and symptoms of respiratory infection. This test is intended for use as an aid in the differential diagnosis of influenza B viral infections in humans in conjunction with clinical and epidemiological risk factors. The assay does not detect the presence of influenza C virus.

Negative results do not preclude influenza virus infection and should not be used as the sole basis for diagnosis, treatment or other patient management decisions.

Performance characteristics for influenza A were established during the spring of 2016 when influenza A/H3 and 2009 H1N1 influenza were the predominant influenza A viruses in circulation. When other influenza A viruses are emerging, performance characteristics may vary.

If infection with a novel influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Applicant:

Quidel Corporation 12544 High Bluff Drive, Suite 200 San Diego, California 92130 Telephone: 858-552-7910 Fax: 858-646-8045

Contact Information:

Ronald H. Lollar, Senior Director Clinical and Regulatory Affairs 2005 East State Street, Suite 100 Athens, Ohio 45701 740-589-3300 - Corporate number 740-589-3373 – Desk phone 858-552-6451—Fax Ron.Lollar@quidel.com

Date of preparation of 510(k) summary:

June 30, 2016

A. 510(k) Number:

K161814

B. Purpose for Submission:

To obtain substantial equivalence for the Solana Influenza A+B Assay when performed on the Solana

C. Measurand:

Influenza A: Matrix Gene; Influenza B: Matrix Gene

D. Type of Test:

Reverse Transcriptase - Helicase-Dependent Amplification (RT-HDA)

{4}------------------------------------------------

E. Applicant:

Quidel Corporation

F. Proprietary and Established Names:

Solana Influenza A+B Assay

G. Regulatory Information:

Table 1.Regulatory Information
Product CodeClassificationRegulation SectionPanel
OCCClass II21 CFR 866.3980 Respiratoryviral panel multiplex nucleic acidassayMicrobiology(83)

H. Intended Use:

1. Intended Use(s):

The Solana® Influenza A+B Assay is a qualitative in vitro diagnostic test for the detection and differentiation of influenza A and influenza B viral RNA in nasal and nasopharyngeal swabs from patients with signs and symptoms of respiratory infection. This test is intended for use as an aid in the differential diagnosis of influenza A and influenza B viral infections in humans in conjunction with clinical and epidemiological risk factors. The assay does not detect the presence of influenza C virus.

Negative results do not preclude influenza virus infection and should not be used as the sole basis for diagnosis, treatment or other patient management decisions.

Performance characteristics for the Solana® Influenza A+B Assay were established during the spring of 2016 when influenza A/H3 and 2009 H1N1 influenza were the predominant influenza viruses in circulation. When other influenza viruses are emerging, performance characteristics may vary.

If infection with a novel influenza virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent Influenza viruses and sent to state or local health department for testing. Viral culture

{5}------------------------------------------------

should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.

    1. Indication(s) for Use:
      Same as Intended Use.

3. Special conditions for use statement(s):

  • For in vitro diagnostic use only
  • For prescription use only ●

4. Special instrument requirements:

Solana® Instrument

l. Device Description:

The Solana Influenza A+B Assay amplifies and detects viral RNA present in viral transport media containing nasopharyngeal or nasal swab specimens obtained from symptomatic patients.

The assay consists of two major steps: 1) specimen preparation, and 2) amplification and detection of target sequences specific to influenza A and/or influenza B using isothermal Reverse Transcriptase - Helicase-Dependent Amplification (RT-HDA) in the presence of target-specific fluorescence probes.

A patient nasal or nasopharyngeal swab specimen in viral transport media is transferred to a Process Buffer Tube, subjected to heat treatment at 95°C for 5 minutes and mixed. The processed sample is transferred to a Reaction Tube. The Reaction Tube contains lyophilized RT-HDA reagents, dNTPs, primers and probes. Once rehydrated with the processed sample, the Reaction Tube is placed in Solana for amplification and detection of influenza A and influenza B-specific target sequences. In Solana, the target sequences are amplified by influenza A and influenza B specific primers and detected by influenza A and influenza B specific fluorescence probes, respectively. A competitive process control (PRC) is included in the Process Buffer Tube to monitor sample processing, inhibitory substances in clinical samples, reagent failure or device failure. The PRC target is amplified by influenza B specific primers and detected by a PRC specific fluorescence probe.

The two target probes and PRC probe are labeled with a quencher on one end and a fluorophore on the other end. In addition, the two target probes and PRC probe have one or more bases that are comprised of ribonucleic acid. Upon annealing to influenza A,

{6}------------------------------------------------

influenza B or PRC amplicons, the fluorescence probes are cleaved by RNaseH2 and the fluorescence signal increases due to physical separation of fluorophore from quencher. Solana measures and interprets the fluorescent signal, using on-board method-specific algorithms. Solana then reports the test results to the user on its display screen, and it can print out the results via an integrated printer.

Materials Provided:

Solana® Influenza A+B Assay Kit: M300 48 Tests per kit

Table 2.Kit Components
ComponentQuantityStorage
Process Buffer48 tubes/kit 1.55 mL2°C to 8°C
Reaction Tubes48 tubes/kit2°C to 8°C

Materials required but not provided:

  • External controls for Influenza A and Influenza B (e.g. Quidel Molecular Influenza ● A+B Control Set, which contains positive and negative controls, serves as an external processing control)
  • Sterile DNAse-free filter-blocked positive displacement micropipettor tips ●
  • Micropipettor
  • Stopwatch or timer ●
  • Scissors or a blade
  • Workflow tray
  • . Transfer Rack
  • Heat block capable of 95 ± 2°C temperature ●
  • Thermometer
  • Solana instrument
  • Transport Media (BD/Copan UTM, Remel M4, Remel M4RT, Remel M5, Remel M6, ● or Copan eSwab)

J. Substantial Equivalence Information:

    1. Predicate device name(s):
      Lyra® Influenza A+B Assay
    1. Predicate 510(k) number(s): K131728

{7}------------------------------------------------

3. Comparison with predicate:

Table 3. Similarities
ItemSolana® Influenza A+B AssayLyra® Influenza A+B Assay(K131728)
Intended UseThe Solana® Influenza A+B Assay is aqualitative in vitro diagnostic test for thedetection and differentiation of influenzaA and influenza B viral RNA in nasal andnasopharyngeal swabs from patients withsigns and symptoms of respiratoryinfection. This test is intended for use asan aid in the differential diagnosis ofinfluenza A and influenza B viralinfections in humans in conjunction withclinical and epidemiological risk factors.The assay does not detect the presenceof influenza C virus.Negative results do not precludeinfluenza virus infection and should notbe used as the sole basis for diagnosis,treatment or other patient managementdecisions.Performance characteristics for influenzaA were established during the spring2016 when influenza A/H3 and 2009H1N1 influenza were the predominantinfluenza A viruses in circulation. Whenother influenza A viruses are emerging,performance characteristics may vary.If infection with a novel influenza A virus issuspected based on current clinical andepidemiological screening criteriarecommended by public healthThe Lyra® Influenza A+B assayis a multiplex Real Time RT-PCRassay for the in vitro qualitativedetection and differentiation ofinfluenza A and influenza Bviral RNA in nasal andnasopharyngeal swabs frompatients with signs andsymptoms of respiratoryinfection. This test is intendedfor use as an aid in thedifferential diagnosis ofinfluenza A and influenza Bviral infections in humans inconjunction with clinical andepidemiological risk factors.The assay does not detect thepresence of influenza C virus.Negative results do notpreclude influenza virusinfection and should not beused as the sole basis fordiagnosis, treatment or otherpatient management decisions.Performance characteristics forinfluenza A were establishedduring the 2011 and 2013influenza seasons wheninfluenza A/H3 and 2009 H1N1
Table 3.Similarities
ItemSolana® Influenza A+B AssayLyra® Influenza A+B Assay (K131728)
authorities, specimens should be collectedwith appropriate infection controlprecautions for novel virulent Influenzaviruses and sent to state or local healthdepartment for testing. Viral cultureshould not be attempted in these casesunless a BSL 3+ facility is available toreceive and culture specimens.influenza were thepredominant influenza Aviruses in circulation. Whenother influenza A viruses areemerging, performancecharacteristics may vary.If infection with a novelinfluenza A virus is suspectedbased on current clinical andepidemiological screeningcriteria recommended bypublic health authorities,specimens should be collectedwith appropriate infectioncontrol precautions for novelvirulent Influenza viruses andsent to state or local healthdepartment for testing. Viralculture should not beattempted in these casesunless a BSL 3+ facility isavailable to receive and culturespecimens.The assay can be performedusing either the LifeTechnologies QuantStudio® Dx;the Applied Biosystems® 7500Fast Dx, or the CepheidSmartCycler® II.
Sample Typesnasal swab and nasopharyngeal swabSame
DetectionAutomated multiplex assay using differentSame
Techniquesreporter dyes for each target
Table 4. Differences
ItemSolana® Influenza A+B AssayLyra® Influenza A+BAssay (K131728)
Viral TargetInfluenza A: Matrix Gene;Influenza B: Matrix GeneInfluenza A: MatrixGene;Influenza B:conserved influenzaB sequence withinthe neuraminidasegene
AmplificationTechnologyReverse Transcriptase - Helicase-DependentAmplification (RT-HDA)Real Time PCR-based system fordetecting thepresence or absenceof viral RNA inclinical specimens
ExtractionMethodsNonebioMérieuxeasyMAG®AutomatedMagnetic ExtractionReagents
InstrumentSolanaLife TechnologiesQuantStudio® Dx,the AppliedBiosystems® 7500Fast Dx, or theCepheidSmartCycler® II
PerformanceCharacteristicsThe analytical sensitivity (limit of detection or LOD) ofthe Solana® Influenza A+B Assay was determinedusing quantified (TCID50/mL) cultures of three (3)influenza A strains and two (2) influenza B strains,serially diluted in negative nasopharyngeal matrix.The analyticalsensitivity (limit ofdetection or LoD)of the QuidelMolecularInfluenza A+B assay
Influenza A+B assay. Analytical sensitivity (LOD) iswas determined
defined as the lowest concentration at which at leastusing quantified
95% of all replicates tested positive. The(TCID50/mL)
demonstrated LOD for each strain tested is showncultures of five (5)
below:influenza A strains,
LOD Valuesthree (3) influenza
VirusTCID50/mLB strains, serially
diluted in negative
Influenza ASubtypenasopharyngeal
matrix. Each
A/Taiwan/42/06H1N17.5x102dilution was
extracted using the
A/California/07/2009H1N1p4.7x102NucliSENS
A/Texas/50/2012H3N26.3x10°easyMAG System
and tested in
Influenza BLineagereplicates of 20 per
concentration of
B/Brisbane/60/08Victoria8.5x10+virus on the Life
Technologies
B/Massachusetts/2/2012Yamagata3.3x104QuantStudio® Dx;
the Applied
Biosystems® 7500
Fast Dx, or the
Cepheid
SmartCycler® II.
Analytical sensitivity
(LoD), as defined as
the lowest
concentration at
which 95% of all
replicates tested
positive, ranged
from 10 to 100
TCID50/mL.

{8}------------------------------------------------

Solana® Influenza A+B Assay 6/30/2016 Page 6 of 24

510(k) Summary

{9}------------------------------------------------

Solana® Influenza A+B Assay 6/30/2016 Page 7 of 24

510(k) Summary

{10}------------------------------------------------

K. Standard/Guidance Document Referenced (if applicable):

Guidance for Industry and FDA Staff: Statistical Guidance on Reporting Results from Studies Evaluating Diagnostic Tests (Final, 3/13/2007)

{11}------------------------------------------------

http://www.fda.gov/downloads/MedicalDevices/DeviceRegulationandGuidance/GuidanceD ocuments/ucm071287.pdf

Guidance for Industry and FDA Staff - Class II Special Controls Guidance Document: Respiratory Viral Panel Multiplex Nucleic Acid Assay -

http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocuments/ ucm180307.htm

Guidance on Informed Consent for In Vitro Diagnostic Device Studies Leftover Human Specimens that are Not Individually Identifiable (April 2006) – http://www.fda.gov/cdrh/oivd/guidance/1588.pdf.

Guidance for Industry and Food and Drug Administration Staff - eCopy Program for Medical Device (December 2012)

http://www.fda.gov/downloads/MedicalDevices/DeviceRegulationandGuidance/GuidanceD ocuments/UCM313794.pdf

L. Test Principle:

The Solana Influenza A+B Assay amplifies and detects viral RNA present in viral transport media containing nasopharyngeal or nasal swab specimens obtained from symptomatic patients.

The assay consists of two major steps: 1) specimen preparation, and 2) amplification and detection of target sequences specific to influenza A and/or influenza B using isothermal Reverse Transcriptase - Helicase-Dependent Amplification (RT-HDA) in the presence of target-specific fluorescence probes.

A patient nasal or nasopharyngeal swab specimen in viral transport media is transferred to a Process Buffer Tube, subjected to heat treatment at 95°C for 5 minutes and mixed. The processed sample is transferred to a Reaction Tube. The Reaction Tube contains lyophilized RT-HDA reagents, dNTPs, primers and probes. Once rehydrated with the processed sample, the Reaction Tube is placed in Solana for amplification and detection of influenza A and influenza B-specific target sequences. In Solana, the target sequences are amplified by influenza A and influenza B specific primers and detected by influenza A and influenza B specific fluorescence probes, respectively. A competitive process control (PRC) is included in the Process Buffer Tube to monitor sample processing, inhibitory substances in clinical samples, reagent failure or device failure. The PRC target is amplified by influenza B specific primers and detected by a PRC specific fluorescence probe.

{12}------------------------------------------------

The two target probes and PRC probe are labeled with a quencher on one end and a fluorophore on the other end. In addition, the two target probes and PRC probe have one or more bases that are comprised of ribonucleic acid. Upon annealing to influenza A, influenza B or PRC amplicons, the fluorescence probes are cleaved by RNaseH2 and the fluorescence signal increases due to physical separation of fluorophore from quencher. Solana measures and interprets the fluorescent signal, using on-board method-specific algorithms. Solana then reports the test results to the user on its display screen, and it can print out the results via an integrated printer.

M. Performance Characteristics:

1. Analytical performance:

  • a. Precision/Reproducibility:

Reproducibility

The reproducibility of the Solana® Influenza A+B Assay was evaluated at three laboratory sites. A four sample panel consisting of three levels of a combined influenza A and influenza B contrived sample and a negative contrived sample are tested in this study. Influenza A and influenza B viruses (Influenza A/California/07/2009 and Influenza B/Brisbane/60/08, respectively) are diluted in negative nasal matrix to 2 x LOD for moderate positive, 1 x LOD for low positive and diluted to C20 to C80 for high negative / low positive nasal matrix without spiked virus is used for the negative sample. The Solana Influenza A+B assay was used according to the instructions for use.

Panels and controls were tested at each site by two operators per instrument for five days, each sample tested in three (3) replicates, for a total of 90 results per level for each virus for each instrument (2 operators x 5 days x 3 sites x 3 replicates).

{13}------------------------------------------------

Solana® Influenza A+B Assay 6/30/2016 Page 11 of 24

510(k) Summary

Table 5. Reproducibility Summary
SITESITESITE
Site #1Site #2Site #3Overall Percent Agreement95% Confidence Interval
#Detected positive/# tested% Agreement with Expected Result#Detected positive/# tested% Agreement with Expected Result#Detected positive/# tested% Agreement with Expected Result
Influenza A/California/07/2009 High Negative (1.4 x102 TCID50/mL)10/3033.325/3083.323/3076.758/9064.454.1 to 73.6
Influenza A/California/07/2009 Low Positive (4.7x102 TCID50/mL)30/3010030/3010030/3010090/9010096.5 to 100
Influenza A/California/07/2009 Moderate Positive (9.4x102 TCID50/mL)30/3010030/3010030/3010090/9010096.5 to 100
Negative0/301000/301000/301000/9010096.5 to 100
Influenza A Positive Control15/1510015/1510015/1510045/4510094.2 to 100
Influenza A Negative Control0/151000/151000/151000/4510094.2 to 100
SITESITESITE
Site #1Site #2Site #3Overall Percent Agreement With Expected Results95% Confidence Interval
#Detected positive/# tested% Agreement with Expected Result#Detected positive/# tested% Agreement with Expected Result#Detected positive/# tested% Agreement with Expected Result
Influenza B/Brisbane/60/08 High Negative (2.6 x101 TCID50/mL)9/30305/3016.710/3033.324/9026.718.6 to 36.6
Influenza B/Brisbane/60/08 Low Positive (8.5x102TCID50/mL)30/3010030/3010030/3010090/9010096.5 to 100
Influenza B/Brisbane/60/08 Moderate Positive (1.7x102 TCID50/mL)30/3010030/3010030/3010090/9010096.5 to 100
Negative0/301000/301000/301000/9010096.5 to 100
Table 5. Reproducibility Summary
SITE
Site #1Site #2Site #3
#Detectedpositive/#tested% AgreementwithExpectedResult#Detectedpositive/#tested%AgreementwithExpectedResult#Detectedpositive/#tested% Agreementwith ExpectedResultOverall PercentAgreement95%ConfidenceInterval
Influenza B PositiveControl15/1510015/1510015/1510045/4510094.2 to 100
Influenza B NegativeControl0/151000/151000/151000/4510094.2 to 100

{14}------------------------------------------------

The results suggest that there are no significant differences between different users and different sites on different days. Reproducibility studies are acceptable.

  • b. Linearity/assay reportable range:
    Not applicable - This assay is qualitative.

  • c. Traceability, Stability, Expected values (controls, calibrators, or methods):
    Traceability:

Not applicable. This assay is qualitative.

Specimen Stability:

Influenza A and B strains (Influenza A/California/07/2009 and Influenza B/Brisbane/60/08, respectively) were formulated in six (6) transport medium pooled negative matrix (Copan UTM, Remel M4, Remel M5, Remel M6, Remel M4RT or Copan ESwab transport media) at a final concentration of 2x LOD level.

The transport media systems containing the contrived samples were stored at 2° to 8°C up to 9 days. The samples were processed according to the instructions for use. Each transport media was tested in 3 replicates at Day 0, 24 hours, 48 hours, 72 hours, Day 7, and Day 9.

Influenza A and influenza B are stable in transport media BD UTM (1- and 3- mL), Remel M4 (3-mL), Remel M4RT (3-mL), Remel M5 (3-mL), and Remel M6 (3-mL) at 2° to 8°C for up to 9 days.

{15}------------------------------------------------

Influenza A and influenza B are stable in Copan eSwab transport media at 2 to 8°C for up to 48 hours.

Controls:

Controls (Quidel Molecular Influenza A+B Control Set), which contains positive and negative controls, serves as an external processing and extraction control) were run on the Solana Influenza A+B Assay each day of testing. These controls are described as follows:

  • a. The process control is used to monitor sample processing, to detect HDA inhibitory specimens, to confirm the integrity of assay reagents and the operation of the Solana instrument. The process control is included in the Reaction Mix tube.
  • b. The external positive control may be treated as a patient specimen. The control should be sampled and tested as if it were a patient specimen and processed as described above in the Assay Procedure. The external positive control is intended to monitor substantial reagent and instrument failure.
  • c. The external negative control may be treated as a patient specimen. The control should be sampled and tested as if it were a patient specimen and processed as described above in the Assay Procedure. The external negative control is used to detect reagent or environmental contamination (or carryover) by influenza A or B RNA or amplicon.

d. Detection limit:

The analytical sensitivity (limit of detection or LOD) of the Solana® Influenza A+B Assay was determined using quantified (TCID50/mL) cultures of three (3) influenza A strains and two (2) influenza B strains, serially diluted in negative nasopharyngeal matrix. Each dilution was run as 20 replicates in the Solana Influenza A+B assay. Analytical sensitivity (LOD) is defined as the lowest concentration at which at least 95% of all replicates tested positive. The demonstrated LOD for each strain tested is shown below:

{16}------------------------------------------------

Table 6.LOD Values
VirusTCID50/mL
Influenza ASubtype
A/Taiwan/42/06H1N1$7.5x10^2$
A/California/07/2009H1N1p$4.7x10^2$
A/Texas/50/2012H3N2$6.3x10^0$
Influenza BLineage
B/Brisbane/60/08Victoria$8.5x10^1$
B/Massachusetts/2/2012Yamagata$3.3x10^1$

e. Analytical specificity:

Cross Reactivity:

A study was performed to evaluate the cross-reactivity of the Solana® Influenza A+B Assay with forty-four (44) microorganisms (24 bacteria, 1 yeast, 19 viruses) potentially found in specimens that are collected from patients symptomatic for influenza. Each microorganism was diluted in negative nasal matrix to the desired concentration (106 or higher CFU/mL for bacteria, yeast and 105 or higher pfu/mL or TCID50/mL for viruses) and tested. The organisms and their concentrations included in the cross-reactivity study are shown in the table below.

Table 7.Cross-reactive Organisms
OrganismConcentrationTested
Adenovirus 11.0x105 TCID50/mL
Adenovirus 111.0x105 TCID50/mL
Bordetella bronchiseptica1.0x106 CFU/mL
Bordetella pertussis1.0x106 CFU/mL
Candida albicans1.0x106 CFU/mL
Chlamydophila pneumoniae5.0x104 TCID50/mL*
Coronavirus 229E1.0x105 TCID50/mL
Corynebacterium diptheriae1.0x106 CFU/mL
Coxsackievirus B5/10/20061.0x10⁵ TCID₅₀/mL
Echovirus 111.0x10⁵ TCID₅₀/mL
Echovirus 61.0x10⁵ TCID₅₀/mL
Enterovirus 701.0x10⁵ TCID₅₀/mL
Enterovirus 712.0x10⁴ TCID₅₀/mL*
Epstein Barr virus1.0x10⁵ TCID₅₀/mL
Escherichia coli1.0x10⁶ CFU/mL
Haemophilus influenzae1.0x10⁶ CFU/mL
HSV 1 MacIntyre Strain1.0x10⁵ TCID₅₀/mL
HSV 2 G strain1.0x10⁵ TCID₅₀/mL
Human Rhinovirus1.0x10⁵ TCID₅₀/mL
Klebsiella pneumoniae1.0x10⁶ CFU/mL
Lactobacillus plantarum1.0x10⁶ CFU/mL
Legionella pneumophila1.0x10⁶ CFU/mL
Measles1.0x10⁵ TCID₅₀/mL
Metapneumovirus A11.0x10⁵ TCID₅₀/mL
Moraxella catarrhalis1.0x10⁶ CFU/mL
Mumps1.0x10⁵ TCID₅₀/mL
Mycobacterium avium1.0x10⁶ CFU/mL
Mycobacterium tuberculosis1.0x10⁶ CFU/mL
Mycoplasma pneumoniae1.0x10⁶ CFU/mL
Neisseria gonorrhoeae1.0x10⁶ CFU/mL
Neisseria meningitides1.0x10⁶ CFU/mL
Parainfluenza Type 11.0x10⁵ TCID₅₀/mL
Parainfluenza Type 21.0x10⁵ TCID₅₀/mL
Parainfluenza Type 31.0x10⁵ TCID₅₀/mL
Proteus mirabilis1.0x10⁶ CFU/mL
Proteus vulgaris1.0x10⁶ CFU/mL
Pseudomonas aeruginosa1.0x10⁶ CFU/mL
Respiratory syncytial virus1.0x10⁵ TCID₅₀/mL
Staphylococcus aureus1.0x10⁶ CFU/mL
Staphylococcus epidermidis1.0x10⁶ CFU/mL
Streptococcus mutans1.0x10⁶ CFU/mL
Streptococcus pneumoniae1.0x10⁵ CFU/mL*
Streptococcus pyogenes1.0x10⁶ CFU/mL
Streptococcus salivarius1.0x10⁶ CFU/mL

{17}------------------------------------------------

  • Due to low concentration of the stock organism, the concentration tested was below the target. The actual concentration tested is listed in the table.

{18}------------------------------------------------

No cross-reactivity was observed with the forty-four (44) microorganisms tested with the Solana® Influenza A+B Assay.

Interference:

The performance of Solana® Influenza A+B Assay was evaluated with potentially interfering substances that may be present in nasopharyngeal specimens. The potentially interfering substances were evaluated with influenza A (A/Mexico/ 4108/2009) and influenza B (Influenza B/Brisbane/60/08) at concentrations of 2x LOD. There was no evidence of interference caused by the substances tested at the concentrations shown below.

Table 8. Interfering Substances
SubstancesActive IngredientConcentration Tested
Purified mucin proteinMucin protein2.5 mg/mL
Blood (human)Blood5.0%
Afrin — nasal sprayOxymetazoline5.0%
Saline nasal spraySaline15.0%
PhenylephrinehydrochloridePhenylephrine hydrochloride15.0%
FlonaseFluticasone5.0%
Zicam Gentle Allergy ReliefNasalGelGalphimia glauca, Histaminumhydrochloricum, Luffa operculata,Sulfur5.0%
MupirocinMupirocin12.0 mg/mL
OseltamivirOseltamivir2.2 µg/mL
ZanamivirZanamivir282.0 ng/mL
TobramycinTobramycin2.5 mg/mL
ChlorasepticBenzocaine, Menthol0.68 g/mL
Amantadine hydrochlorideAmantadine hydrochloride282.0 ng/mL
Nasocort Allergy 24 hourTriamcinolone5.0%
Sinus Buster Nasal SprayCapsicum annuum (Capsaicin)5.0%
NasalCrom Nasal AllergyCromolyn Sodium5.0%

{19}------------------------------------------------

Table 8.Interfering Substances
SubstancesActive IngredientConcentration Tested
Spray
RhinocortBudesonide (Glucocorticoid)5.0%
Air-Vita Allergy Multi-Symptom ReliefAllium cepa, Ambrosiaartemisiaefolia, Apis mellifica,Chamomilla, Eucalyptol, Eucalyptusglobulus, Euphrasia officinalis,Galphimia glauca, Histaminumhydrochloricum, Natrummuriaticum, Nux vomica, Quercusrobur, Silicea, Wyethia helenioides5.0%
Ipratropium bromideIpratropium bromide10.0 mg/mL
Olopatadine hydrochlorideOlopatadine hydrochloride10.0 mg/mL
Amantadine hydrochlorideAmantadine hydrochloride282.0 ng/mL

Analytical Reactivity (Inclusivity):

The reactivity of the Solana® Influenza A+B Assay was evaluated against multiple strains of influenza A and influenza B viruses. The clinical influenza panel consisted of fourteen (14) influenza A strains, and eight (8) Influenza B strains at concentrations near the level of detection (LOD) of the assay.

Table 9. Inclusivity Strains
StrainSubtype/LineageTCID50/mLInclusive (Yes orNo)
Influenza A
A/Mexico/4108/2009H1N1p2.3x103Yes
A/Denver/1/57H1N12.3x103Yes
A/New Jersey/8/76H1N12.3x103Yes
A/PR/8/34H1N12.3x103Yes
A/FM/1/47H1N12.3x103Yes
A/Solomon Islands/3/06H1N12.3x103Yes
A/New Caledonia/20/1999H1N12.3x103Yes
A/Victoria/361/11H3N22.3x103Yes
A/Port Chalmers/1/73H3N21.4x104Yes
A/Aichi/2/68H3N22.3x103Yes
A/Victoria/3/75H3N22.3x103Yes

{20}------------------------------------------------

Table 9.Inclusivity Strains
StrainSubtype/LineageTCID50/mLInclusive (Yes orNo)
A/Hong Kong/8/68H3N22.3x103Yes
A/Wisconsin/67/2005H3N22.3x103Yes
A/WS/33H1N12.3x103Yes
Influenza B
B/Malaysia/2506/04Victoria2.6x102Yes
B/Florida/07/2004Victoria7.7x102Yes
B/Maryland/1/59Yamagata2.6x102Yes
B/Allen/45Yamagata2.6x102Yes
B/Lee/40Yamagata2.6x102Yes
B/Florida/04/2006Yamagata7.7x102Yes
B/Panama/45/90Yamagata2.6x102Yes
B/Hong Kong/5/72Victoria2.6x102Yes
B/Malaysia/25/06/04Victoria2.6x102Yes

Due to restrictions and availability of a number of influenza A strains, in silico analysis was performed for three additional strain designations:

  • A total of four (4) H3N2v (1 human strain and 3 swine) sequences were analyzed in silico. All four sequences demonstrated 100% homology.
  • . A total of three hundred forty (340) H5N1 strains were analyzed in silico. Three hundred thirty-nine (339) strains in the database demonstrated ≥95% overall homology and ≥88% homology to any individual primer or probe sequence. One H5N1 strain demonstrated an overall homology of 88% and ≥82% homology to any individual primer or probe sequence.
  • A total of one hundred sixty-four (164) H7N9 sequences were analyzed in silico. All 164 sequences demonstrated 100% homology.
  • Fourteen (14) non-clinical avian restricted influenza A viruses (Table 10) were analyzed in silico.
Table 10.Non-clinical Avian Restricted Influenza A Viruses
SubtypeStrain
H2N2A/Mallard/NY/6750/78 (H2N2)
H7N3A/Chicken/NJ/15086-3/94 (H7N3)
H9N2A/Chicken/NJ/12220/97 (H9N2)
H4N8A/Mallard/OH/338/86 (H4N8)
H6N2A/Chicken/CA/431/00 (H6N2)

{21}------------------------------------------------

Table 10.Non-clinical Avian Restricted Influenza A Viruses
SubtypeStrain
H8N4A/Blue Winged Teal/LA/B174/86 (H8N4)
H5N1A/Anhui/01/2005(H5N1)-PR8-IBCDC-RG5
H10N7A/GWT/LA/169GW/88 (H10N7)
H11N9A/Chicken/NJ/15906-9/96 (H11N9)
H12N5A/Duck/LA/188D/87 (H12N5)
H13N6A/Gull/MD/704/77 (H13N6)
H14N5A/Mallard/GurjevRussia/262/82 (H14N5)
H15N9A/Shearwater/Australia/2576/79 (H15N9)
H16N3A/Shorebird/DE/172/2006(H16N3)

A total of twenty-seven (27) sequences were available for analysis. The Solana™ FluA primers and probe are 90-100% conserved to the specified avian strains and to representative avian strains.

  • f. Assay cut-off:
    Not applicable.

2. Comparison studies:

  • a. Method comparison with predicate device:
    Not applicable

  • b. Matrix comparison:
    Not applicable

    1. Clinical studies:
    • a. Clinical Sensitivity:

{22}------------------------------------------------

Performance characteristics of the Solana Influenza A+B Assay were established during a prospective study with specimens collected between February and April 2016. One thousand four hundred seventy-three (1473) prospectively collected specimens have been included in this study at five (5) sites across the United States. A single nasal or nasopharyngeal swab specimen (302 and 1171, respectively) was collected per patient in viral transport media (BD/Copan UTM, Remel M5, Remel M6). All specimens were transported to a central location at 2℃ to 8℃ for testing by the comparator methods (culture for influenza A and B using the R-Mix Too mixed cells and direct specimen DFA (DSFA), and extraction with the NucliSENS® easyMAG® and testing with a FDA-cleared Influenza A+B molecular assay). The specimens were processed and tested with Solana Influenza A+B Assay (frozen (731) and fresh (742)) on the Solana instrument at the sites.

COMPARISON VERSUS CULTURE WITH DFA AND DSFA

One thousand four hundred seventy-three (1473) fresh specimens were included in this study. Each specimen was cultured for influenza A and B using the R-Mix Too mixed cells and stained with an FDA-cleared device and processed for direct specimen DFA (DSFA). All comparator testing was performed on fresh specimens within 72-hours of their collection. A specimen was recorded as positive for influenza A or B if either comparator test was positive.

Seven hundred and forty-two (742) of these specimens were tested fresh using the Solana Influenza A+B Assay for the presence influenza A or B. Seven hundred and thirty-one (731) specimens were frozen and stored at -70°C prior to testing with the Solana Influenza A+B Assay. Fifteen (15) specimens were contaminated or toxic in the cell culture (1.0%). Fifty (50) specimens were invalid in the Solana Assay (3.4%). These sixty-five (65) specimens have been excluded from further analysis. Tables 11 and 12 detail the performance of the Solana Assay for influenza A and influenza B respectively for the remaining one thousand four hundred eight (1408) specimens, across all testing sites combined, as compared to viral culture with DSFA results.

Table 11.Performance Characteristics of the Solana Influenza A+B Assay for Influenza ACompared to Culture and DSFA (Across all Sites Combined)
SourceCategoryNTPFPTNFNSensitivity%(95% CI)Specificity%(95% CI)
Fresh70918024503298.9(96.1 to 99.7)95.4(93.3 to 96.9)
Frozen69917627493398.3(95.2 to 99.4)94.8(92.6 to 96.4)
All140835651*9965**98.6(96.8 to 99.4)95.1(93.7 to 96.3)

*Of the fifty-one (51) discordant specimens (Solana Positive/Culture and DSFA Negative) reported in Table 11, twenty-eight (28) of these specimens were positive by an alternate molecular assay.

{23}------------------------------------------------

Table 12.Performance Characteristics of the Solana Influenza A+B Assay for Influenza B Compared toCulture and DSFA (Across all Sites Combined)
SourceCategoryNTPFPTNFNSensitivity%(95% CI)Specificity%(95% CI)
Fresh7096216460100(94.2 to 100)99.8(99.1 to 100)
Frozen6992386680100(85.7 to 100)98.8(97.7 to 99.4)
All1408859*13140100(95.7 to 100)99.3(98.7 to 99.6)

**Of the five (5) discordant specimens (Solana Negative/Culture and DSFA Positive) reported in Table 11, two (2) of these specimens were positive by an alternate FDA cleared molecular assay.

*Of the nine (9) discordant specimens (Solana Positive/Culture and DSFA Negative) reported in Table 12, two (2) of these specimens were positive by an alternate FDA cleared molecular assay.

COMPARISON WITH A FDA-CLEARED INFLUENZA A+B MOLECULAR ASSAY

One thousand four hundred seventy-three (1473) specimens were processed using the NucliSENS® easyMAG® and tested with a FDA-cleared Influenza A+B molecular assay according to the assay's package insert. The comparator testing was performed on fresh specimens within 72-hours of their collection.

Seven hundred and thirty-one (731) of the original specimens were frozen and stored at -70°C prior to testing with the Solana Influenza A+B Assay. Seven hundred and forty-two (742) of the original specimens were tested fresh using the Solana Influenza A+B Assay for the presence of influenza A or B. Thirty one (31) specimens were invalid in the comparator assay (2.1%). Fifty (50) specimens were invalid in the Solana Assay (3.4%)(one specimen was invalid in both assays). These eighty (80) specimens have been excluded from further analysis.

Tables 13 and 14 detail the positive percent agreement (PPA) and the negative percent agreement (NPA) of the Solana Influenza A+B Assay results for influenza A and influenza B respectively, as compared with an FDA cleared molecular comparator, for the remaining one thousand three hundred ninety-three (1393) specimens.

Table 13.Percent Agreement of the Solana® Influenza A+B Assay for Influenza ACompared to an FDA cleared Influenza A+B Molecular Assay (Across all Sites Combined)
SourceCategoryNTPFPTNFNPPA(95% CI)NPA(95% CI)
Fresh7101959499796.5(93.0 to 98.3)98.2(98.7 to 99.1)
Frozen68318024475497.8(94.5 to 99.2)95.2(92.9 to 96.7)

{24}------------------------------------------------

Table 13. Percent Agreement of the Solana® Influenza A+B Assay for Influenza A Compared to an FDA cleared Influenza A+B Molecular Assay (Across all Sites Combined)
Source CategoryNTPFPTNFNPPA (95% CI)NPA (95% CI)
All1393375339741197.2 (95.0 to 98.4)96.7 (95.4 to 97.7)

There were a total of forty-four (44) discordant specimens among the hundred ninety-three (1393) specimens evaluated. Of the thirty-three (33) discordant specimens (Solana Positive) Comparator Negative) reported in Table 13, nine (9) of these specimens were positive by culture/DSFA. Of the eleven (11) discordant specimens (Solana Negative/Comparator Positive) reported in Table 13, two (2) of these specimens were positive by culture/DSFA.

Percent Agreement of the Solana® Influenza A+B Assay for Influenza BCompared to an FDA-cleared Influenza A+B Molecular Assay (Across all Sites Combined)
Table 14.SourceCategoryNTPFPTNFNPPA(95% CI)NPA(95% CI)
Fresh7105766470100(93.7 to 100)99.1(98.0 to 99.6)
Frozen6832386520100(85.7 to 100)98.8(97.6 to 99.4)
All1393801412990100(95.4 to 100)98.9(98.2 to 99.4)

There were a total of fourteen (14) discordant specimens among the one thousand three hundred ninety-three (1393) specimens evaluated. Of the fourteen (14) discordant specimens (Solana Positive) Comparator Negative) reported in Table 14, seven (7) of these specimens were positive by culture/DSFA.

b. Clinical specificity:

See Section 3a.

Other clinical supportive data (when a. and b. are not applicable): C.

Not applicable

    1. Clinical cut-off:
      Not applicable
    1. Expected values:
      The expected values of the Solana Influenza A+B Assay were established during a prospective study conducted between February and April 2016. One thousand four hundred seventy-three (1473) specimens (fresh (742) and frozen (731)) have been included in this study at five (5) sites across the United States. A single specimen was

{25}------------------------------------------------

collected per patient. The specimens were processed and tested with Solana® Influenza A+B Assay on the Solana® instrument at the sites.

The expected value of influenza A and influenza B with the Solana Influenza A+B Assay has been calculated for the combined sites based on the age of the patient.

Fifty-three (53) of the one thousand four hundred seventy-three (1473) specimens were removed from analysis: (three (3) specimens did not have the age provided; fifty (50) specimens were invalid). Table 15 provides the percentage of influenza A and influenza B positive cases per specified age group, as determined by the Solana Influenza A+B Assay, for the remaining one thousand four hundred twenty (1420) specimens.

Table 15. Expected Values (N=1420)
Age GroupInfluenza AInfluenza B
Number of PatientsNumber of PositivesPrevalenceNumber of PatientsNumber of PositivesPrevalence
≤ 5 years3779124.1%377266.9%
6 to 21 years2978930.0%2974816.2%
22 to 59 years50419137.9%504173.4%
≥ 60 years2423715.3%24231.2%

The prospective clinical study had a dual infection rate for Influenza A and Influenza B of 0.2% (3/1420) using the Solana Influenza A+B Assay. All three (3) of these dual detections were only positive for influenza A by culture and DSFA and also by an alternate molecular comparator.

N. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:

Instrument: Solana Instrument

O. System Descriptions:

1. Modes of Operation:

The Solana instrument heats each reaction tube to 58°C. If present, the target RNA sequence is reverse transcribed into cDNA by the Reverse Transcriptase and influenza A or B specific primers that are present in the reaction mix. After the completion of the Reverse Transcriptase step, the Solana instrument heats each reaction tube to 65°C where the isothermal DNA polymerase amplifies the cDNA strands using the influenza A or B specific primers. Influenza A or B specific fluorescence probes are also included in the Reaction Tube. The target probes are labeled with a quencher on one end and a fluorophore specific for the recognition sequence on the other end. In addition, the

{26}------------------------------------------------

target probes carry a ribonucleic acid. Upon annealing to amplicons, the fluorescence probes are cleaved by RNaseH2 and the fluorescence signal increases due to physical separation of fluorophore from quencher. The Solana instrument measures and interprets the fluorescent signal, using on-board method-specific algorithms. Solana instrument will then report the test results to the user on its display screen, and it can print out the results via a printer.

2. Software:

FDA has reviewed applicant's Hazard Analysis and software development processes for this line of product types:

Yes No

P. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10, 21 CFR 801.109, and the special controls.

§ 866.3980 Respiratory viral panel multiplex nucleic acid assay.

(a)
Identification. A respiratory viral panel multiplex nucleic acid assay is a qualitative in vitro diagnostic device intended to simultaneously detect and identify multiple viral nucleic acids extracted from human respiratory specimens or viral culture. The detection and identification of a specific viral nucleic acid from individuals exhibiting signs and symptoms of respiratory infection aids in the diagnosis of respiratory viral infection when used in conjunction with other clinical and laboratory findings. The device is intended for detection and identification of a combination of the following viruses:(1) Influenza A and Influenza B;
(2) Influenza A subtype H1 and Influenza A subtype H3;
(3) Respiratory Syncytial Virus subtype A and Respiratory Syncytial Virus subtype B;
(4) Parainfluenza 1, Parainfluenza 2, and Parainfluenza 3 virus;
(5) Human Metapneumovirus;
(6) Rhinovirus; and
(7) Adenovirus.
(b)
Classification. Class II (special controls). The special controls are:(1) FDA's guidance document entitled “Class II Special Controls Guidance Document: Respiratory Viral Panel Multiplex Nucleic Acid Assay;”
(2) For a device that detects and identifies Human Metapneumovirus, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Human Metapneumovirus (hMPV) Using Nucleic Acid Assays;” and
(3) For a device that detects and differentiates Influenza A subtype H1 and subtype H3, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Detection and Differentiation of Influenza A Virus Subtypes Using Multiplex Nucleic Acid Assays.” See § 866.1(e) for the availability of these guidance documents.