K Number
K161776

Validate with FDA (Live)

Manufacturer
Date Cleared
2016-11-29

(153 days)

Product Code
Regulation Number
886.4300
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Lenstec's LC Injection System is intended for use in the implantation of the Softec I, Softec HDO, Softec HDM and any approved IOL in which labeling specifies use of this injector (s) to insert the capsular bag following extracapsular extraction.

Device Description

The proposed system consists of the following components: Cartridge with silicone cushion, IOL, and Injector. Three types of injectors are provided: they are syringe based and reusable and autoclavable. One type of lens loader is provided, which is reusable and autoclavable. The cartridges/silicone cushion are single-use and are provided sterile.

AI/ML Overview

The provided document is a 510(k) premarket notification for a medical device (Lenstec LC Injection System) and does not contain the detailed information required to fill out the table regarding acceptance criteria and device performance as it would be presented in a study report. This document focuses on demonstrating substantial equivalence to a predicate device rather than presenting a full clinical or non-clinical study report with detailed acceptance criteria, sample sizes, and expert evaluations.

However, I can extract the relevant information that is present in the document to address some of your points:

1. Table of Acceptance Criteria and Reported Device Performance:

The document mentions non-clinical tests but does not define specific quantitative acceptance criteria or provide detailed performance metrics in a tabular format. The general statement is that tests were "satisfactory."

Acceptance CriteriaReported Device Performance
Biocompatibility testsSatisfactory results
System performance with Lenstec IOL models (Softec HD, Softec HDO, Softec HDM, Softec I)Satisfactory results
Particulate Matter in Ophthalmic Solutions (USP <789>)Cartridges tested passed the USP <789> Light Obscuration Tests

2. Sample Size Used for the Test Set and Data Provenance:

  • Test Set (Non-clinical): The document does not specify exact sample sizes for the non-clinical tests performed on the cartridges or with the IOL models. It refers to "the system was tested with each of the following Lenstec intraocular lens models" and "the cartridges tested."
  • Data Provenance: Not specified, but standard for device manufacturers to conduct these tests in-house or through certified labs. This would be considered non-clinical, not patient data.

3. Number of Experts Used to Establish Ground Truth and Qualifications:

Not applicable. This is a submission for a medical device (intraocular lens injection system), not an AI or diagnostic device that requires expert-established ground truth from images or other patient data. The "ground truth" here is the physical and biological performance of the device components, validated by standard engineering and biocompatibility testing.

4. Adjudication Method for the Test Set:

Not applicable as it's not a study involving human interpretation or clinical adjudication.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

No. This is a submission for an ophthalmic medical device, not an AI or diagnostic tool.

6. Standalone (Algorithm Only) Performance Study:

No. This document pertains to a physical medical device, not an algorithm.

7. Type of Ground Truth Used:

  • For biocompatibility: Established biological safety standards and test methodologies.
  • For system performance with IOLs: Successful delivery of the IOL without damage or complications, likely assessed visually and functionally. The term "satisfactory results" implies meeting predefined engineering and functional specifications.
  • For particulate matter: USP <789> standards.

8. Sample Size for the Training Set:

Not applicable. This is not an AI/machine learning device that requires a training set.

9. How Ground Truth for the Training Set Was Established:

Not applicable.

Summary of what the document implies about the study (non-clinical):

The document describes non-clinical testing conducted to support the substantial equivalence claim for the Lenstec LC Injection System. This testing focused on:

  • Biocompatibility: Ensuring safety of contact materials.
  • Performance with IOLs: Verifying the system can implant various Lenstec IOL models (Softec HD, Softec HDO, Softec HDM, Softec I) satisfactorily.
  • Particulate Matter: Ensuring cartridges meet USP standards for ophthalmic solutions.

The basis for equivalence extends to changes in cartridge models, introduction of LubriMATRIX™ surface treatment, and a minor material change in polypropylene, with tests confirming these changes do not negatively impact device performance or safety. No clinical studies were required for this 510(k) submission.

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines beneath them that resemble water or fabric.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 29, 2016

Lenstec Inc. Mr. Jimmy Chacko Vice President, Regulatory Affairs 1765 Commerce Ave. N. St. Petersburg, Florida 33716

Re: K161776

Trade/Device Name: Lenstec LC Injection System Regulation Number: 21 CFR 886.4300 Regulation Name: Intraocular Lens Guide Regulatory Class: Class I Product Code: MSS Dated: November 2, 2016 Received: November 3, 2016

Dear Mr. Chacko:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing

{1}------------------------------------------------

(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"

(21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

Kesia Alexander

for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose, and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

Device Name Lenstec LC Injection System

Indications for Use (Describe)

Lensted's LC Injection System is intended for use in the implantation of the Softec I, Softec HDO, Softec HDM and any approved IOL in which labeling specifies use of this injector (s) to insert the capsular bag following extracapsular extraction.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{3}------------------------------------------------

Section 6: 510(k) Summary

Date Summary Prepared: June 17th, 2016

Lenstec Inc 510(k) Summary for Premarket Notification Submission for Lenstec LC Injection System

Labeling:

Federal (United States) Law restricts this device to sale by or on the order of a physician

    1. Applicant Information:
    • a. Name: Lenstec Inc.
    • b. Address: 1765 Commerce Avenue North St. Petersburg, FL 33716 Telephone Number: (727) 571-2272 Fax Number: (727) 571-1792
    • c. Contact Person: Jimmy Chacko, Vice President, Regulatory Affairs Email: JChacko@Lenstec.com Additional Contact Person: Sheila Baker, Clinical Research Manager Email: SBaker@Lenstec.com
    1. Name of Device
    • a. Trade name: Lenstec LC Injection System
    • Common name: Intraocular lens injection system b.
    • Regulation name: Intraocular lens guide ﻥ
    • d. Regulatory class: Class I, reserve
    • e. Product code: MSS
    • Regulation number: 21 CFR 886.4300 f.
    1. Substantially Equivalent legally-marketed device:
    • a. Lenstec Injection System for Intraocular Lens delivery (K122848)

{4}------------------------------------------------

4. Device Description

The proposed system consists of the following components:

CART SERIES CARTRIDGE CHART
Cartridge withsilicone cushionIOLInjectorTip Diameter(mm)Lenstec IOL Powerrange (D)
CART 20SSOFTEC HDI-9011SI-9012I-9012 FS1.65.0 - 26.0
SOFTEC I
SOFTEC HDOSOFTEC HDO5.0 - 22.0
CART 45SSOFTEC HDI-9011SI-9012I-9012 FS1.65.0-26.0
SOFTEC I
SOFTEC HDO
CART MSOFTEC HDMI-9011SI-9012I-9012 FS1.75.0-36.0

Three types of injectors are provided: they are syringe based and reusable and autoclavable. One type of lens loader is provided, which is reusable and autoclavable. The cartridges/silicone cushion are single-use and are provided sterile.

    1. Use:
      Lenstec's LC Injection System is intended for use in the implantation of the Softec HD (P090022), Softec I (P090022/S001), Softec HDO (P090022/S011), Softec HDM (P090022/S026) and any approved IOL in which the labeling specifies use of this injector to insert the lens into the capsular bag following extracapsular extraction. The predicate device was comprised of the nearly identical components.

Lenstec seeks to have cleared the addition of a surface treatment technology, LubriMATRIX™, to Cart 45S. Cart 20S and Cart M cartridges as well as a change of material due to the vendor discontinuing the current material. We also request the addition of two new lens injectors, which have barrels that are now threaded instead of smooth, allowing for a twisting motion to advance the plunger into the cartridge.

The only changes in this iniection system are

  • a) the modified indication for inclusion for use of the Softec HD, Softec I, both through K122848. Softec HDO and HDM through this current application.

{5}------------------------------------------------

  • b) additional new cartridges; Cart 45S, Cart 20S, Cart M
  • c) The addition of surface treatment technology (LubriMATRIX™) on all "Cart" models (Cart 45S, Cart 20S, Cart M)
  • d) A slight material change from the Atofina Polypropylene 7823MZ to the Total 7823M Polypropylene. The only reason for the change is that the vendor has discontinued the 7823MZ. The only difference is that the 7823M does not contain an antistat.
    1. Indications for use:

Lenstec's LC Injection System is intended for use in the implantation of the Softec HD, Softec I, Softec HDO, Softec HDM and any approved IOL in which the labeling specifies use of this injector to insert the lens into the capsular bag following extracapsular extraction.

    1. Technological characteristics:
      The system has three major components: a reusable injector, a reusable lens loader and one of three disposable cartridges (Cart 45S, Cart 20S, Cart M) with a silicone cushion (SIC-01-02). Discussed here are the characteristics of the changes only. The lens loader and silicone cushion have not had any changes and are still identical to those cleared under K122848.
  • a. The proposed injectors (1-9012 and I-9012 FS) both have threaded barrels instead of the smooth barrel ("syringe like") injection of the I-9011S. Both the I-9012 and I-9012 FS advance the IOL through the cartridge once the user twists the injector knob until the IOL is delivered into the eye. The I-9011S injector advances the IOL through the cartridge once the user pushes the injector knob until the IOL is delivered into the eye.

  • b. The surface treated cartridges (Cart 45S, Cart 20S, Cart M) have LubriMATRIXTM applied for lubricity.

  • The proposed cartridge is manufactured of medical grade polypropylene c. and is for single use as in the cleared K122848. however, a material change is necessary due to the vendor discontinuing the polypropylene we currently purchase (7823MZ). The proposed material is now 7823M without the antistat that the 7823MZ had.

    1. Performance data:
    • a. Non clinical tests

All contact materials have been tested for biocompatibility, with satisfactory results. Also, the system was tested with each of the following Lenstec intraocular lens models with satisfactory results: Softec HD, Softec HDO and Softec HDM. Due to the similarities of the Softec HD and Softec 1 IOL design, the Softec 1 will not undergo performance testing of its own. Based on the overall design shape of the Softec HD to the Softec 1, equivalence is expected and we can therefore include the Softec 1 as part of the indications for use as in the previously cleared K122848.

{6}------------------------------------------------

The Cart M was successfully tested with the Softec HDM and can also be included as part of the indications for use statement. Furthermore, the cartridges tested passed the USP <789> Particulate Matter in Ophthalmic Solutions- Light Obscuration Tests.

    1. Clinical tests: Not required
    1. Conclusions:

The Lenstec LC Injection System is substantially equivalent to the legally marketed predicate device, and the included testing validates the expanded indication for use.

17 June 16

Timmy Chacko Vice President, Regulatory Affairs Lenstec Inc.

Date

§ 886.4300 Intraocular lens guide.

(a)
Identification. An intraocular lens guide is a device intended to be inserted into the eye during surgery to direct the insertion of an intraocular lens and be removed after insertion is completed.(b)
Classification. Class I (general controls). Except when used as folders or injectors for soft or foldable intraocular lenses, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 886.9.