K Number
K161708
Date Cleared
2016-11-22

(154 days)

Product Code
Regulation Number
870.1130
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

It is a device intended to measure the systolic, diastolic blood pressure and pulse rate of an adult individual by using a non-invasive oscillometric technique in which an inflatable cuff (available sizes: 22cm44cm (8.7in17.3in)) is wrapped around the single upper arm.

Device Description

Arm automatic blood pressure monitor is based on pressure vibration method. Blood pressure cuff use the air pump to inflate, then the arteries are extruded by the cuff with pressure. Pressure sensor collects the pressure in the cuff, and then converts it to digital signal to the CPU. Then the software calculates the systolic and diastolic blood pressure and pulse rate.

AI/ML Overview

This document is a 510(k) Summary for a medical device called an "Arm automatic blood pressure monitor" (various models). The purpose of this summary is to demonstrate that the new device is substantially equivalent to a legally marketed predicate device.

Here's an analysis of the acceptance criteria and study information provided, extracting the requested details:

Acceptance Criteria and Device Performance

The acceptance criteria for accuracy are based on the ANSI/AAMI/ISO 81060-2:2009 standard for non-invasive sphygmomanometers. While the document mentions that a repeated clinical test for the subject devices was "unnecessary" due to technical similarity with the predicate, the predicate device's performance, which the subject devices claim substantial equivalence to, is detailed in a comparison table.

Table of Acceptance Criteria and Reported Device Performance

ItemAcceptance Criteria (Predicate Device Claimed Performance)Reported Device Performance (Subject Device Claimed Performance)Basis for Acceptance/Performance
Accuracy - Pressure±3 mmHg (±0.4 kPa)±3 mmHg (±0.4 kPa)Claimed identical to predicate
Accuracy - Pulse±5%±5%Claimed identical to predicate
Measurement Range - Pressure0 mmHg ~ 280 mmHg (0 kPa ~ 37.3 kPa)0 mmHg ~ 280 mmHg (0 kPa ~ 37.3 kPa)Claimed identical to predicate
Measurement Range - Pulse40 bpm ~ 180 bpm40 bpm ~ 180 bpmClaimed identical to predicate
Operating Environment - Temperature+5℃ ~ +40℃+5℃ ~ +40℃Claimed identical to predicate
Operating Environment - Humidity15% ~ 93%15% ~ 93%Claimed identical to predicate
Operating Environment - Pressure70.0 kPa ~ 106.0 kPa70.0 kPa ~ 106.0 kPaClaimed identical to predicate
Operating Environment - Altitude≤ 3000 m≤ 3000 mClaimed identical to predicate
Storage/Transport Environment - Temperature-25℃ ~ +70℃-25℃ ~ +70℃Claimed identical to predicate
Storage/Transport Environment - Humidity10% ~ 95%10% ~ 95%Claimed identical to predicate
Storage/Transport Environment - Pressure50.0 kPa ~ 106.0 kPa50.0 kPa ~ 106.0 kPaClaimed identical to predicate

Study Information:

  1. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • The document states that a repeated clinical test for the subject devices was unnecessary because the subject devices are identical to the predicate device from a technical point of view, and the differences (Power Source, Weight, Size, Broadcast Function) do not relate to blood pressure measurement technology. Therefore, no new clinical test data for the subject devices is provided in this summary.
    • The provenance of the data for the predicate device's clinical validation (which established the acceptance criteria for these equivalence claims) is not detailed in this document. It is implied that the predicate device (K151258) met the ANSI/AAMI/ISO 81060-2:2009 standard, but the specific details of that study are not provided in this submission.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Since a new clinical test for the subject device was deemed "unnecessary," there would be no new ground truth established for the subject device in this submission. The ground truth for the predicate device's accuracy would have been established during its original clinical validation against the ANSI/AAMI/ISO 81060-2:2009 standard, which typically involves comparing automated measurements to those taken by trained observers using a reference sphygmomanometer. Details on the number or qualifications of these "experts" (trained observers) are not provided in this document.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • No new clinical test was performed for the subject device, so no adjudication method is mentioned for its test set. The ANSI/AAMI/ISO 81060-2 standard outlines procedures for obtaining reference blood pressure measurements, often involving multiple observers to minimize bias, but the specific method used for the predicate device's validation is not detailed here.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • This is not applicable. The device is an automatic blood pressure monitor, not an AI-assisted diagnostic tool that would involve human "readers" or an MRMC study.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • The core functionality of an "Arm automatic blood pressure monitor" is inherently standalone. It uses an oscillometric technique to calculate blood pressure and pulse rate without human intervention during the measurement process. The comparison tables indicate the subject device and predicate device "use the same oscillometric method within the software algorithm to determine the systolic and diastolic blood pressure and pulse rate." This implies a standalone algorithm performance.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • For blood pressure monitors, the "ground truth" during clinical validation (following standards like ANSI/AAMI/ISO 81060-2) is typically established by simultaneous auscultatory measurements performed by trained observers using a calibrated reference sphygmomanometer. This is considered an expert comparison or reference measurement.
  7. The sample size for the training set

    • This document does not specify a training set as it focuses on demonstrating substantial equivalence rather than detailing the development and initial validation of the algorithm itself. The oscillometric method used is a well-established technique for blood pressure measurement, and any "training" (e.g., development of the algorithm's parameters) would have occurred during the initial design and validation of the predicate device's core technology, which is not described here.
  8. How the ground truth for the training set was established

    • As no training set is explicitly discussed in this document for the subject device, the method for establishing its ground truth is not provided. For oscillometric devices, the algorithm's parameters are typically derived and refined based on clinical data where the device's readings are compared against standard reference measurements.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal for the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is an abstract symbol consisting of three stylized human profiles facing to the right, with a ribbon-like element flowing beneath them.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 22, 2016

Shenzhen Pump Medical System Co., Ltd Ms. Migo Yang Clinical Registration Specialist 2/f. M-7 Sinosteel Building. Maqueling Estate. Hi-tech Industrial Park Nanshan District, Shenzhen, 518057 CN

Re: K161708

Trade/Device Name: Arm automatic blood pressure monitor, Model: BF3202, BF3212, BF6002, BF6102, BF1205, BF1215, BF1213, BF6051, BF6151, BF6043, BF6143, BF1204,BF1214, BF6023L, BF6123L, BF6048L, BF6148L, BF6053, BF6153, BF6055 Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: May 23, 2016 Received: June 21, 2016

Dear Ms. Migo Yang,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act

{1}------------------------------------------------

Page 2 - Ms. Migo Yang

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Mada Yellen
for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K161708

Device Name

Arm automatic blood pressure monitor, Model: BF3202, BF3212, BF1205,BF1215, BF1213, BF6051,BF6151, BF6043,BF6143, BF1204,BF1214, BF6023L,BF6123L, BF6148L, BF6053,BF6153, BF6055;

Indications for Use (Describe)

It is a device intended to measure the systolic, diastolic blood pressure and pulse rate of an adult individual by using a non-invasive oscillometric technique in which an inflatable sizes: 22cm44cm (8.7in17.3in)) is wrapped around the single upper arm.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

The assigned 510(k) number is: _K161708

Submitter's Identification: 1.

SHENZHEN PUMP MEDICAL SYSTEM CO., LTD. 2/F, M-7 Sinosteel Building, Maqueling Estate, Hi-Tech Industrial Park, Nanshan District, Shenzhen, PEOPLE'S REPUBLIC OF CHINA

Contact: Xie Qiongyu Tel:86-0755-26710795 Fax:86-0755-26498210 E-mail:xieqy@bpump.com.cn

2. Name of the Device:

Trade Name: Arm automatic blood pressure monitor Model: BF3202,BF3212, BF6002,BF6102, BF1205,BF1215, BF1213, BF6051,BF6151, BF6043,BF6143, BF1204,BF1214, BF6023L,BF6123L, BF6048L,BF6148L, BF6053,BF6153,BF6055; Regulation Description: Non-invasive blood pressure measurement system. Regulation Number: 21 CFR 870.1130 Regulation Class: II Product Code: DXN Review Panel: Cardiovascular

3. Information for the 510(k) Cleared Device(Predicate device):

Arm automatic blood pressure monitor, Model: BE6034, BE6134, K151258, SHENZHEN PUMP MEDICAL SYSTEM CO., LTD.

4. Description of the device

Arm automatic blood pressure monitor is based on pressure vibration method. Blood pressure cuff use the air pump to inflate, then the arteries are extruded by the cuff with

{4}------------------------------------------------

pressure. Pressure sensor collects the pressure in the cuff, and then converts it to digital signal to the CPU. Then the software calculates the systolic and diastolic blood pressure and pulse rate.

The arm automatic blood pressure monitor, Model: BF3202,BF3212, BF6002,BF6102, BF1205.BF1215. BF1213. BF6051.BF6151. BF6043.BF6143. BF1204.BF1214. BF6023L,BF6123L, BF6048L,BF6148L, BF6053,BF6153, BF6055; has the same basic principles, main function, performance and intended use, and they are consistent in product structure and material.

5. Intended Use:

Arm automatic blood pressure monitor, model: BF3202,BF3212, BF6002,BF6102, BF1205.BF1215. BF1213. BF6051.BF6151. BF6043.BF6143. BF1204.BF1214. BF6023L,BF6123L, BF6048L,BF6148L, BF6053,BF6153, BF6055; is intended for measuring systolic, diastolic blood pressure and pulse rate of an adult over the counter.

6. Indications for Use:

Arm automatic blood pressure monitor, Model: BF3202,BF3212, BF6002,BF6102, BF1205,BF1215, BF1213, BF6051,BF6151, BF6043,BF6143, BF1204,BF1214, BF6023L,BF6123L, BF6048L,BF6148L, BF6053,BF6153, BF6055; is intended to measure the systolic, diastolic blood pressure and pulse rate of an adult individual by using a non-invasive oscillometric technique in which an inflatable cuff (available sizes: 22cm to 44cm (8.7in to 17.3in)) is wrapped around the single upper arm.

7. Comparison to the 510(k) Cleared Device (Predicate device):

The subject device, Arm automatic blood pressure monitor, Model: BF3202,BF3212, BF6002.BF6102. BF1205.BF1215. BF1213. BF6051.BF6151. BF6043.BF6143. BF1204,BF1214, BF6023L,BF6123L, BF6048L,BF6148L, BF6053,BF6153, BF6055; and the predicate device: Arm automatic blood pressure monitor, Model: BE6034, BE6134 uses the same oscillometric method within the software algorithm to determine the systolic and diastolic blood pressure and pulse rate. Upper arm cuff is inflated automatically by air pump and the pressures are transferred via tubing to a sensor in the unit.

They differ by the appearance, the way of energy supply as well as PCB. The appearance of subject device is different from predicate device; The only way of energy supply of predicate device is battery while the way of energy supply of subject device can be both battery and adapter. The PCB of subject device has been updated. But those differences do not affect the accuracy or normal use of this device because they use the same fundamental scientific technology based on clinical declaration of identity and the declaration of clinical identity.

The subject device: Arm automatic blood pressure monitor, Model: BF3202,BF3212, BF6002,BF6102,BF1205,BF1215,BF1213,BF6051,BF6151,BF6043,BF6143, BF1204,BF1214,BF6023L,BF6123L,BF6048L,BF6148L,BF6053,BF6153,BF6055;

{5}------------------------------------------------

uses the same oscillometric method as the predicate device: Arm automatic blood pressure monitor, Model: BE6034, BE6134 to determinate the systolic and diastolic blood pressure and pulse rate. They have the same function and intended use. Based upon the aforementioned information, these devices are substantially equivalent. Please refer to table below for comparison.

{6}------------------------------------------------

(Subject Devices:BE6034,BE6134,BF3202,BF3212,BF6002,BF6102,BF1213)
ModelComparison
NoItemsBE6034, BE6134BF3202BF3212BF6002BF6102BF1213ResultNote
1ManufacturerSHENZHENPUMPMEDICALSYSTEMCO.,LTD.SHENZHENPUMPMEDICALSYSTEMCO.,LTD.SHENZHENPUMPMEDICALSYSTEMCO.,LTD.SHENZHENPUMP MEDICALSYSTEMCO.,LTD.SHENZHENPUMP MEDICALSYSTEMCO.,LTD.SHENZHENPUMPMEDICALSYSTEMCO.,LTD.IdenticalNA
2Intended UseIt is a deviceintended tomeasure thesystolic, diastolicblood pressureand pulse rate ofan adult individualby using a non-invasiveoscillometrictechnique inwhich aninflatable cuff(available sizes:22cm to 44cm(8.7in to 17.3in))is wrapped aroundthe single upperarm.It is a deviceintended tomeasure thesystolic,diastolic bloodpressure andpulse rate of anadult individualby using a non-invasiveoscillometrictechnique inwhich aninflatable cuff(availablesizes: 22cm to44cm (8.7in to17.3in)) iswrappedaround thesingle upperarm.It is a deviceintended tomeasure thesystolic, diastolicblood pressureand pulse rate ofan adultindividual byusing a non-invasiveoscillometrictechnique inwhich aninflatable cuff(available sizes:22cm to 44cm(8.7in to 17.3in))is wrapped aroundthe single upperarm.It is a deviceintended tomeasure thesystolic, diastolicblood pressureand pulse rate ofan adult individualby using a non-invasiveoscillometrictechnique inwhich aninflatable cuff(available sizes:22cm to 44cm(8.7in to 17.3in))is wrapped aroundthe single upperarm.It is a deviceintended tomeasure thesystolic, diastolicblood pressure andpulse rate of anadult individual byusing a non-invasiveoscillometrictechnique in whichan inflatable cuff(available sizes:22cm to 44cm(8.7in to 17.3in))is wrapped aroundthe single upperarm.It is a deviceintended tomeasure thesystolic,diastolic bloodpressure andpulse rate of anadult individualby using a non-invasiveoscillometrictechnique inwhich aninflatable cuff(available sizes:22cm to 44cm(8.7in to17.3in)) iswrapped aroundthe single upperarm.IdenticalNA
3Indications forUseIt is intended formeasuringsystolic, diastolicblood pressureand pulse rate ofan adult over thecounter.It is intendedfor measuringsystolic,diastolic bloodpressure andpulse rate of anadult over thecounter.It is intended formeasuringsystolic, diastolicblood pressureand pulse rate ofan adult over thecounter.It is intended formeasuringsystolic, diastolicblood pressureand pulse rate ofan adult over thecounter.It is intended formeasuringsystolic, diastolicblood pressure andpulse rate of anadult over thecounter.It is intendedfor measuringsystolic,diastolic bloodpressure andpulse rate of anadult over thecounter.IdenticalNA
ModelComparison
NoItemsBE6034, BE6134BF3202BF3212BF6002BF6102BF1213ResultNote
4PatientpopulationAdultAdultAdultAdultAdultAdultIdenticalNA
5Environment ofuseHomeHomeHomeHomeHomeHomeIdenticalNA
6MeasuringPrincipleOscillometricMethodOscillometricMethodOscillometricMethodOscillometricMethodOscillometricMethodOscillometricMethodIdenticalNA
7MeasurementRangePressure:0mmHg280mmHg(0kPa37.3kPa)Pulse: 40 bpm~180 bpmPressure:0mmHg280mmHg(0kPa37.3kPa)Pulse: 40 bpm~180 bpmPressure:0mmHg280mmHg(0kPa37.3kPa)Pulse: 40 bpm~180 bpmPressure:0mmHg280mmHg(0kPa37.3kPa)Pulse: 40 bpm~180 bpmPressure:0mmHg280mmHg (0kPa37.3kPa)Pulse: 40 bpm~180 bpmPressure:0mmHg280mmHg(0kPa37.3kPa)Pulse: 40 bpm~180 bpmIdenticalNA
8AccuracyPressure:±3mmHg(±0.4kPa)Pulse: ±5%Pressure:±3mmHg(±0.4kPa)Pulse: ±5%Pressure:±3mmHg(±0.4kPa)Pulse: ±5%Pressure:±3mmHg(±0.4kPa)Pulse: ±5%Pressure:±3mmHg(±0.4kPa)Pulse: ±5%Pressure:±3mmHg(±0.4kPa)Pulse: ±5%IdenticalNA
9DisplayTN-LCD DigitalDisplayTN-LCDDigital DisplayTN-LCD DigitalDisplayTN-LCD DigitalDisplayTN-LCD DigitalDisplayTN-LCDDigital DisplayIdenticalNA
10Memory2 Memory sets, 60readings each set.2 Memory sets,60 readingseach set.2 Memory sets, 60readings each set.2 Memory sets, 60readings each set.2 Memory sets, 60readings each set.2 Memory sets,60 readingseach set.IdenticalNA
11PowerSource4 AA Alkalinebattery4 AA Alkalinebattery orAC Adapter(Input: AC100-240 V,50/60Hz,400mA; Output :D.C. 6.0V,0.5A;)4 AA Alkalinebattery orAC Adapter(Input: AC 100-240 V,50/60Hz,400mA;Output : D.C.6.0V, 0.5A;)4 AA Alkalinebattery orAC Adapter(Input: AC 100-240 V,50/60Hz,400mA;Output : D.C.6.0V, 0.5A;)4 AA Alkalinebattery orAC Adapter(Input: AC 100-240 V,50/60Hz,400mA;Output : D.C.6.0V, 0.5A;)4 AA Alkalinebattery orAC Adapter(Input: AC 100-240 V,50/60Hz,400mA; Output :D.C. 6.0V,0.5A;)DifferentNA
ItemsModelComparison
NoBE6034, BE6134BF3202BF3212BF6002BF6102BF1213ResultNote
12BroadcastFunctionBE6034: YES;BE6134: NoNoNoNoNoNoDifferentNA
13OperatingEnvironmentTemperature:+5℃~+40℃;Humidity: 15%-93%Pressure: 70.0kPa~106.0kPaAltitude: ≤ 3 000mTemperature:+5℃+40℃Humidity :15%-93%Pressure:70.0kPa106.0kPaAltitude: < 3000 mTemperature:+5℃~+40℃;Humidity: 15%-93%Pressure: 70.0kPa~106.0kPaAltitude: ≤ 3 000mTemperature:+5℃~+40℃;Humidity: 15%-93%Pressure: 70.0kPa~106.0kPaAltitude: ≤ 3 000mTemperature:+5℃~+40℃;Humidity: 15%-93%Pressure: 70.0kPa~106.0kPaAltitude: ≤ 3 000mTemperature:+5℃+40℃Humidity:15%-93%Pressure:70.0kPa106.0kPaAltitude: ≤ 3000 mIdenticalNA
14StorageandTransportEnvironmentTemperature: -25℃~+70℃;Humidity: 10%~95%Pressure:50.0kPa~106.0kPaTemperature:-25°C ~ +70°CHumidity :10%95%Pressure:50.0kPa106.0kPaTemperature: -25°C ~ +70°C ;Humidity: 10%~95%Pressure:50.0kPa~106.0kPaTemperature: -25°C ~ +70°C ;Humidity: 10%~95%Pressure:50.0kPa~106.0kPaTemperature: -25℃+70℃;Humidity: 10%95%Pressure:50.0kPa~106.0kPaTemperature: -25℃~+70℃;Humidity:10%95%Pressure:50.0kPa106.0kPaIdenticalNA
15Weight360g (Withoutbatteries)275g(withoutbatteries)275g(withoutbatteries)293g(withoutbatteries)293g(withoutbatteries)314g(withoutbatteries)DifferentNA
16Size140mm×116mm×81mm138mm×110mm×68mm138mm×110mm×68mm140mm×110mm×70mm140mm×110mm×70mm150mm×110mm×77mmDifferentNA

Table 1. Comparison of Product Specification (Subject Deviceses:BF3202,BF3212,BF6002,BF6102,BF1213)

{7}------------------------------------------------

{8}------------------------------------------------

Table 2. Comparison of Product Specification (Subject Deviceses:BF1205,BF1215,BF6051,BF6151,BF6055)

NoItemsModelComparison
BE6034, BE6134BF1205BF1215BF6051BF6151BF6055ResultNote
1ManufacturerSHENZHENPUMPMEDICALSYSTEMSHENZHENPUMPMEDICALSYSTEMSHENZHENPUMPMEDICALSYSTEMSHENZHENPUMP MEDICALSYSTEMCO.,LTD.SHENZHENPUMP MEDICALSYSTEMCO.,LTD.SHENZHENPUMPMEDICALSYSTEMIdenticalNA
ModelComparison
NoItemsBE6034, BE6134BF1205BF1215BF6051BF6151BF6055ResultNote
CO.,LTD.CO.,LTD.CO.,LTD.CO.,LTD.
2Intended UseIt is a deviceintended tomeasure thesystolic, diastolicblood pressureand pulse rate ofan adult individualby using a non-invasiveoscillometrictechnique inwhich aninflatable cuff(available sizes:22cm to 44cm(8.7in to 17.3in))is wrapped aroundthe single upperarm.It is a deviceintended tomeasure thesystolic,diastolic bloodpressure andpulse rate of anadult individualby using a non-invasiveoscillometrictechnique inwhich aninflatable cuff(availablesizes: 22cm to44cm (8.7in to17.3in)) iswrappedaround thesingle upperarm.It is a deviceintended tomeasure thesystolic, diastolicblood pressureand pulse rate ofan adultindividual byusing a non-invasiveoscillometrictechnique inwhich aninflatable cuff(available sizes:22cm to 44cm(8.7in to 17.3in))is wrapped aroundthe single upperarm.It is a deviceintended tomeasure thesystolic, diastolicblood pressureand pulse rate ofan adult individualby using a non-invasiveoscillometrictechnique inwhich aninflatable cuff(available sizes:22cm to 44cm(8.7in to 17.3in))is wrapped aroundthe single upperarm.It is a deviceintended tomeasure thesystolic, diastolicblood pressure andpulse rate of anadult individual byusing a non-invasiveoscillometrictechnique in whichan inflatable cuff(available sizes:22cm to 44cm(8.7in to 17.3in))is wrapped aroundthe single upperarm.It is a deviceintended tomeasure thesystolic,diastolic bloodpressure andpulse rate of anadult individualby using a non-invasiveoscillometrictechnique inwhich aninflatable cuff(available sizes:22cm to 44cm(8.7in to17.3in)) iswrapped aroundthe single upperarm.IdenticalNA
3Indications for UseIt is intended formeasuringsystolic, diastolicblood pressureand pulse rate ofan adult over thecounter.It is intendedfor measuringsystolic,diastolic bloodpressure andpulse rate of anadult over thecounter.It is intended formeasuringsystolic, diastolicblood pressureand pulse rate ofan adult over thecounter.It is intended formeasuringsystolic, diastolicblood pressureand pulse rate ofan adult over thecounter.It is intended formeasuringsystolic, diastolicblood pressure andpulse rate of anadult over thecounter.It is intendedfor measuringsystolic,diastolic bloodpressure andpulse rate of anadult over thecounter.IdenticalNA
4PatientpopulationAdultAdultAdultAdultAdultAdultIdenticalNA
5Environment ofHomeHomeHomeHomeHomeHomeIdenticalNA
Comparison
NoItemsBE6034, BE6134BF1205BF1215BF6051BF6151BF6055ResultNote
use
6MeasuringPrincipleOscillometricMethodOscillometricMethodOscillometricMethodOscillometricMethodOscillometricMethodOscillometricMethodIdenticalNA
7MeasurementRangePressure:0mmHg280mmHg(0kPa37.3kPa)Pulse: 40 bpm~180 bpmPressure:0mmHg280mmHg(0kPa37.3kPa)Pulse: 40 bpm~180 bpmPressure:0mmHg280mmHg(0kPa37.3kPa)Pulse: 40 bpm~180 bpmPressure:0mmHg280mmHg(0kPa37.3kPa)Pulse: 40 bpm~180 bpmPressure:0mmHg280mmHg(0kPa37.3kPa)Pulse: 40 bpm~180 bpmPressure:0mmHg280mmHg(0kPa37.3kPa)Pulse: 40 bpm~180 bpmIdenticalNA
8AccuracyPressure:±3mmHg(±0.4kPa)Pulse: ±5%Pressure:±3mmHg(±0.4kPa)Pulse: ±5%Pressure:±3mmHg(±0.4kPa)Pulse: ±5%Pressure:±3mmHg(±0.4kPa)Pulse: ±5%Pressure:±3mmHg(±0.4kPa)Pulse: ±5%Pressure:±3mmHg(±0.4kPa)Pulse: ±5%IdenticalNA
9DisplayTN-LCD DigitalDisplayTN-LCDDigital DisplayTN-LCD DigitalDisplayTN-LCD DigitalDisplayTN-LCD DigitalDisplayTN-LCDDigital DisplayIdenticalNA
10Memory2 Memory sets, 60readings each set.2 Memory sets, 60readings each set.2 Memory sets, 60readings each set.2 Memory sets, 60readings each set.2 Memory sets, 60readings each set.2 Memory sets, 60readings each set.IdenticalNA
11PowerSource4 AA Alkalinebattery4 AA Alkalinebattery orAC Adapter(Input: AC100-240 V,50/60Hz,400mA;Output : D.C.6.0V, 0.5A;)4 AA Alkalinebattery orAC Adapter(Input: AC 100-240 V,50/60Hz,400mA;Output : D.C.6.0V, 0.5A;)4 AA Alkalinebattery orAC Adapter(Input: AC 100-240 V,50/60Hz,400mA;Output : D.C.6.0V, 0.5A;)4 AA Alkalinebattery orAC Adapter(Input: AC 100-240 V,50/60Hz,400mA;Output : D.C.6.0V, 0.5A;)4 AA Alkalinebattery orAC Adapter(Input: AC 100-240 V,50/60Hz,400mA;Output : D.C.6.0V, 0.5A;)DifferentNA
12BroadcastFunctionBE6034: YES;BE6134: NoNoNoNoNoNoDifferentNA
Comparison
NoItemsBE6034, BE6134BF1205BF1215BF6051BF6151BF6055ResultNote
13OperatingEnvironmentTemperature:+5°C~+40℃;Humidity: 15%-93%Pressure: 70.0kPa~106.0kPaAltitude: ≤ 3 000mTemperature:+5°C~+40℃;Humidity:15%-93%Pressure:70.0kPa~106.0kPaAltitude: ≤ 3000 mTemperature:+5°C~+40℃;Humidity: 15%-93%Pressure: 70.0kPa~106.0kPaAltitude: ≤ 3 000mTemperature:+5°C~+40℃;Humidity: 15%-93%Pressure: 70.0kPa~106.0kPaAltitude: ≤ 3 000mTemperature:+5°C~+40℃;Humidity: 15%-93%Pressure: 70.0kPa~106.0kPaAltitude: ≤ 3 000mTemperature:+5°C~+40℃;Humidity:15%-93%Pressure:70.0kPa~106.0kPaAltitude: ≤ 3000 mIdenticalNA
14StorageandTransportEnvironmentTemperature:-25°C~+70°C;Humidity: 10%~95%Pressure:50.0kPa~106.0kPaTemperature:-25℃~+70℃;Humidity:10%~95%Pressure:50.0kPa~106.0kPaTemperature:25°C~+70℃;Humidity: 10%~95%Pressure:50.0kPa~106.0kPaTemperature:25°C~+70℃;Humidity: 10%~95%Pressure:50.0kPa~106.0kPaTemperature:25°C~+70℃;Humidity: 10%~95%Pressure:50.0kPa~106.0kPaTemperature:25°C~+70℃;Humidity:10%~95%Pressure:50.0kPa~106.0kPaIdenticalNA
15Weight360g (Withoutbatteries)315g(withoutbatteries)313g(withoutbatteries)297g(withoutbatteries)285g(withoutbatteries)357g(withoutbatteries)DifferentNA
16Size140mm×116mm×81mm140mm×116mm×79mm140mm×116mm×79mm154mm×124mm×65mm154mm×124mm×65mm157mm×113mm×60mmDifferentNA

{9}------------------------------------------------

{10}------------------------------------------------

{11}------------------------------------------------

Table 3. Comparison of Product Specification (Subject Deviceses:BF6043,BF6143,BF1204,BF1214,BF6023L)

NoItemsModelComparison
BE6034, BE6134BF6043BF6143BF1204BF1214BF6023LResultNote
1ManufacturerSHENZHENPUMPMEDICALSYSTEMCO.,LTD.SHENZHENPUMPMEDICALSYSTEMCO.,LTD.SHENZHENPUMPMEDICALSYSTEMCO.,LTD.SHENZHENPUMP MEDICALSYSTEMCO.,LTD.SHENZHENPUMP MEDICALSYSTEMCO.,LTD.SHENZHENPUMPMEDICALSYSTEMCO.,LTD.IdenticalNA
NoItemsBE6034, BE6134BF6043BF6143BF1204BF1214BF6023LComparison
2Intended UseIt is a deviceintended tomeasure thesystolic, diastolicblood pressureand pulse rate ofan adult individualby using a non-invasiveoscillometrictechnique inwhich aninflatable cuff(available sizes:22cm to 44cm(8.7in to 17.3in))is wrapped aroundthe single upperarm.It is a deviceintended tomeasure thesystolic,diastolic bloodpressure andpulse rate of anadult individualby using a non-invasiveoscillometrictechnique inwhich aninflatable cuff(available sizes:22cm to 44cm(8.7in to17.3in)) iswrapped aroundthe single upperarm.It is a deviceintended tomeasure thesystolic, diastolicblood pressureand pulse rate ofan adultindividual byusing a non-invasiveoscillometrictechnique inwhich aninflatable cuff(available sizes:22cm to 44cm(8.7in to 17.3in))is wrappedaround the singleupper arm.It is a deviceintended tomeasure thesystolic, diastolicblood pressureand pulse rate ofan adult individualby using a non-invasiveoscillometrictechnique inwhich aninflatable cuff(available sizes:22cm to 44cm(8.7in to 17.3in))is wrapped aroundthe single upperarm.It is a deviceintended tomeasure thesystolic, diastolicblood pressure andpulse rate of anadult individual byusing a non-invasiveoscillometrictechnique in whichan inflatable cuff(available sizes:22cm to 44cm(8.7in to 17.3in))is wrapped aroundthe single upperarm.It is a deviceintended tomeasure thesystolic, diastolicblood pressureand pulse rate ofan adultindividual byusing a non-invasiveoscillometrictechnique inwhich aninflatable cuff(available sizes:22cm to 44cm(8.7in to 17.3in))is wrapped aroundthe single upperarm.IdenticalNA
3Indications forUseIt is intended formeasuringsystolic, diastolicblood pressureand pulse rate ofan adult over thecounter.It is intended formeasuringsystolic,diastolic bloodpressure andpulse rate of anadult over thecounter.It is intended formeasuringsystolic, diastolicblood pressureand pulse rate ofan adult over thecounter.It is intended formeasuringsystolic, diastolicblood pressureand pulse rate ofan adult over thecounter.It is intended formeasuringsystolic, diastolicblood pressure andpulse rate of anadult over thecounter.It is intended formeasuringsystolic, diastolicblood pressureand pulse rate ofan adult over thecounter.IdenticalNA
4PatientpopulationAdultAdultAdultAdultAdultAdultIdenticalNA
5Environment ofuseHomeHomeHomeHomeHomeHomeIdenticalNA
6MeasuringPrinciplOscillometricMethodOscillometricMethodOscillometricMethodOscillometricMethodOscillometricMethodOscillometricMethodIdenticalNA
NoItemsModelComparison
eBE6034, BE6134BF6043BF6143BF1204BF1214BF6023LResultNote
7MeasurementRangePressure:0mmHg280mmHg(0kPa37.3kPa)Pulse: 40 bpm~180 bpmPressure:0mmHg280mmHg(0kPa37.3kPa)Pulse: 40 bpm~180 bpmPressure:0mmHg280mmHg(0kPa37.3kPa)Pulse: 40 bpm~180 bpmPressure:0mmHg280mmHg(0kPa37.3kPa)Pulse: 40 bpm~180 bpmPressure:0mmHg280mmHg(0kPa37.3kPa)Pulse: 40 bpm~180 bpmPressure:0mmHg280mmHg(0kPa37.3kPa)Pulse: 40 bpm~180 bpmIdenticalNA
8AccuracyPressure:±3mmHg(±0.4kPa)Pulse: ±5%Pressure:±3mmHg(±0.4kPa)Pulse: ±5%Pressure:±3mmHg(±0.4kPa)Pulse: ±5%Pressure:±3mmHg(±0.4kPa)Pulse: ±5%Pressure:±3mmHg(±0.4kPa)Pulse: ±5%Pressure:±3mmHg(±0.4kPa)Pulse: ±5%IdenticalNA
9DisplayTN-LCD DigitalDisplayTN-LCD DigitalDisplayTN-LCD DigitalDisplayTN-LCD DigitalDisplayTN-LCD DigitalDisplayTN-LCD DigitalDisplayIdenticalNA
10Memory2 Memory sets, 60readings each set.2 Memory sets, 60readings each set.2 Memory sets, 60readings each set.2 Memory sets, 60readings each set.2 Memory sets, 60readings each set.2 Memory sets, 60readings each set.IdenticalNA
11PowerSource4 AA Alkalinebattery4 AA Alkalinebattery orAC Adapter(Input: AC 100-240 V,50/60Hz,400mA;Output: D.C. 6.0V,0.5A;)4 AA Alkalinebattery orAC Adapter(Input: AC 100-240 V,50/60Hz,400mA;Output: D.C. 6.0V,0.5A;)4 AA Alkalinebattery orAC Adapter(Input: AC 100-240 V,50/60Hz,400mA;Output: D.C. 6.0V,0.5A;)4 AA Alkalinebattery orAC Adapter(Input: AC 100-240 V,50/60Hz,400mA;Output: D.C. 6.0V,0.5A;)4 AA Alkalinebattery orAC Adapter(Input: AC 100-240 V,50/60Hz,400mA;Output: D.C. 6.0V,0.5A;)DifferentNA
12BroadcastFunctionBE6034: YES;BE6134: NoNoNoNoNoNoDifferentNA
13OperatingEnvironmentTemperature:+5°C~+40°C;Humidity: 15%-93%Pressure: 70.0kPa~106.0kPaTemperature:+5°C~+40°C;Humidity: 15%-93%Pressure: 70.0kPa~106.0kPaTemperature:+5°C~+40°C;Humidity: 15%-93%Pressure: 70.0kPa~106.0kPaTemperature:+5°C~+40°C;Humidity: 15%-93%Pressure: 70.0kPa~106.0kPaTemperature:+5°C~+40°C;Humidity: 15%-93%Pressure: 70.0kPa~106.0kPaTemperature:+5°C~+40°C;Humidity: 15%-93%Pressure: 70.0kPa~106.0kPaIdenticalNA
NoItemsModelComparison
BE6034, BE6134BF6043BF6143BF1204BF1214BF6023LResultNote
Altitude: ≤ 3 000m106.0kPaAltitude: ≤ 3000 mAltitude: ≤ 3 000mAltitude: ≤ 3 000mAltitude: ≤ 3 000mAltitude: ≤ 3 000m
14StorageandTransportEnvironmentTemperature: -25℃~+70℃;Humidity: 10%~95%Pressure:50.0kPa~106.0kPaTemperature:25℃~+70℃;Humidity: 10%95%Pressure:50.0kPaa106.0kPaTemperature:25℃~+70℃;Humidity: 10%~95%Pressure:50.0kPa~106.0kPaTemperature:25℃~+70℃;Humidity: 10%~95%Pressure:50.0kPa~106.0kPaTemperature: -25℃+70℃;Humidity: 10%95%Pressure:50.0kPa~106.0kPaTemperature:25℃~+70℃;Humidity: 10%~95%Pressure:50.0kPa~106.0kPaIdenticalNA
15Weight360g (Withoutbatteries)305g(withoutbatteries)305g(withoutbatteries)354g(withoutbatteries)354g(withoutbatteries)336g(withoutbatteries)DifferentNA
16Size140mm×116mm×81mm141mm×114mm×67mm141mm×114mm×67mm140mm×116mm×81mm140mm×116mm×81mm154mm×124mm×65mmDifferentNA

{12}------------------------------------------------

{13}------------------------------------------------

{14}------------------------------------------------

Table 4. Comparison of Product Specification (Subject Deviceses:BF6123L,BF6048L,BF6148L,BF6053,BF6153)

NoItemsModelComparison
BE6034, BE6134BF6123LBF6048LBF6148LBF6053BF6153ResultNote
1ManufacturerSHENZHENPUMPMEDICALSYSTEMCO.,LTD.SHENZHENPUMPMEDICALSYSTEMCO.,LTD.SHENZHENPUMPMEDICALSYSTEMCO.,LTD.SHENZHENPUMP MEDICALSYSTEMCO.,LTD.SHENZHENPUMP MEDICALSYSTEMCO.,LTD.SHENZHENPUMPMEDICALSYSTEMCO.,LTD.IdenticalNA
2Intended UseIt is a deviceintended tomeasure thesystolic, diastolicblood pressureand pulse rate ofan adult individualby using a non-It is a deviceintended tomeasure thesystolic,diastolic bloodpressure andpulse rate of anadult individualIt is a deviceintended tomeasure thesystolic, diastolicblood pressureand pulse rate ofan adultindividual byIt is a deviceintended tomeasure thesystolic, diastolicblood pressureand pulse rate ofan adult individualby using a non-It is a deviceintended tomeasure thesystolic, diastolicblood pressure andpulse rate of anadult individual byIt is a deviceintended tomeasure thesystolic,diastolic bloodpressure andpulse rate of anadult individualIdenticalNA
NoItemsModelComparison
BE6034, BE6134BF6123LBF6048LBF6148LBF6053BF6153ResultNote
invasiveoscillometrictechnique inwhich aninflatable cuff(available sizes:22cm to 44cm(8.7in to 17.3in))is wrapped aroundthe single upperarm.by using a non-invasiveoscillometrictechnique inwhich aninflatable cuff(availablesizes: 22cm to44cm (8.7in to17.3in)) iswrappedaround thesingle upperarm.using a non-invasiveoscillometrictechnique inwhich aninflatable cuff(available sizes:22cm to 44cm(8.7in to 17.3in))is wrapped aroundthe single upperarm.invasiveoscillometrictechnique inwhich aninflatable cuff(available sizes:22cm to 44cm(8.7in to 17.3in))is wrapped aroundthe single upperarm.invasiveoscillometrictechnique in whichan inflatable cuff(available sizes:22cm to 44cm(8.7in to 17.3in))is wrapped aroundthe single upperarm.by using a non-invasiveoscillometrictechnique inwhich aninflatable cuff(available sizes:22cm to 44cm(8.7in to17.3in)) iswrapped aroundthe single upperarm.
3Indications forUseIt is intended formeasuringsystolic, diastolicblood pressureand pulse rate ofan adult over thecounter.It is intendedfor measuringsystolic,diastolic bloodpressure andpulse rate of anadult over thecounter.It is intended formeasuringsystolic, diastolicblood pressureand pulse rate ofan adult over thecounter.It is intended formeasuringsystolic, diastolicblood pressureand pulse rate ofan adult over thecounter.It is intended formeasuringsystolic, diastolicblood pressure andpulse rate of anadult over thecounter.It is intendedfor measuringsystolic,diastolic bloodpressure andpulse rate of anadult over thecounter.IdenticalNA
4PatientpopulationAdultAdultAdultAdultAdultAdultIdenticalNA
5Environment ofuseHomeHomeHomeHomeHomeHomeIdenticalNA
6MeasuringPrincipleOscillometricMethodOscillometricMethodOscillometricMethodOscillometricMethodOscillometricMethodOscillometricMethodIdenticalNA
7MeasurementRangePressure:0mmHg280mmHg(0kPa37.3kPa)Pulse: 40 bpm~180 bpmPressure:0mmHg280mmHg(0kPa37.3kPa)Pulse: 40 bpmPressure:0mmHg280mmHg(0kPa37.3kPa)Pulse: 40 bpm~180 bpmPressure:0mmHg280mmHg(0kPa37.3kPa)Pulse: 40 bpm~180 bpmPressure:0mmHg280mmHg (0kPa37.3kPa)Pulse: 40 bpm~180 bpmPressure:0mmHg280mmHg(0kPa37.3kPa)Pulse: 40 bpm~180 bpmIdenticalNA
NoItemsModelComparison
BE6034, BE6134BF6123LBF6048LBF6148LBF6053BF6153ResultNote
~180 bpm
8AccuracyPressure:$±3$ mmHg( $±0.4$ kPa)Pulse: $±5$ %Pressure:$±3$ mmHg( $±0.4$ kPa)Pulse: $±5$ %Pressure:$±3$ mmHg( $±0.4$ kPa)Pulse: $±5$ %Pressure:$±3$ mmHg( $±0.4$ kPa)Pulse: $±5$ %Pressure:$±3$ mmHg( $±0.4$ kPa)Pulse: $±5$ %Pressure:$±3$ mmHg( $±0.4$ kPa)Pulse: $±5$ %IdenticalNA
9DisplayTN-LCD DigitalDisplayTN-LCD DigitalDisplayTN-LCD DigitalDisplayTN-LCD DigitalDisplayTN-LCD DigitalDisplayTN-LCD DigitalDisplayIdenticalNA
10Memory2 Memory sets, 60readings each set.2 Memory sets, 60readings each set.2 Memory sets, 60readings each set.2 Memory sets, 60readings each set.2 Memory sets, 60readings each set.2 Memory sets, 60readings each set.IdenticalNA
11PowerSource4 AA Alkalinebattery4 AA Alkalinebattery orAC Adapter(Input: AC100-240 V,50/60Hz,400mA;Output: D.C. 6.0V,0.5A;)4 AA Alkalinebattery orAC Adapter(Input: AC 100-240 V,50/60Hz,400mA;Output: D.C.6.0V, 0.5A;)4 AA Alkalinebattery orAC Adapter(Input: AC 100-240 V,50/60Hz,400mA;Output: D.C.6.0V, 0.5A;)4 AA Alkalinebattery orAC Adapter(Input: AC 100-240 V,50/60Hz,400mA;Output: D.C.6.0V, 0.5A;)4 AA Alkalinebattery orAC Adapter(Input: AC 100-240 V,50/60Hz,400mA;Output: D.C. 6.0V,0.5A;)DifferentNA
12BroadcastFunctionBE6034: YES;BE6134: NoNoNoNoNoNoDifferentNA
13OperatingEnvironmentTemperature:+5℃~+40℃;Humidity: 15%-93%Pressure: 70.0kPa~106.0kPaAltitude: ≤ 3 000mTemperature:+5℃+40℃;Humidity: 15%-93%Pressure:70.0kPa106.0kPaAltitude: ≤ 3Temperature:+5℃~+40℃;Humidity: 15%-93%Pressure: 70.0kPa~106.0kPaAltitude: ≤ 3 000mTemperature:+5℃~+40℃;Humidity: 15%-93%Pressure: 70.0kPa~106.0kPaAltitude: ≤ 3 000mTemperature:+5℃~+40℃;Humidity: 15%-93%Pressure: 70.0kPa~106.0kPaAltitude: ≤ 3 000mTemperature:+5℃+40℃Humidity:15%-93%Pressure:70.0kPa106.0kPaAltitude: ≤ 3IdenticalNA
NoItemsModelComparison
BE6034, BE6134BF6123LBF6048LBF6148LBF6053BF6153ResultNote
000 m000 m
14StorageandTransportEnvironmentTemperature:-25°C ~+70°C;Humidity: 10%~95%Pressure:50.0kPa~106.0kPaTemperature:-25°C~+70°C:Humidity:10%95%Pressure:50.0kPa106.0kPaTemperature:25°C~+70°C;Humidity: 10%~95%Pressure:50.0kPa~106.0kPaTemperature:25°C ~+70°C;Humidity: 10%~95%Pressure:50.0kPa~106.0kPaTemperature:25°C~+70°C;Humidity: 10%~95%Pressure:50.0kPa~106.0kPaTemperature:25°C~+70°C;Humidity:10%~95%Pressure:50.0kPa~106.0kPaIdenticalNA
15Weight360g (Withoutbatteries)336g(withoutbatteries)339g(withoutbatteries)339g(withoutbatteries)358g(withoutbatteries)358g(withoutbatteries)DifferentNA
16Size140mm×116mm×81mm154mm×124mm×65mm141mm×114mm×67mm141mm×114mm×67mm140mm×116mm×81mm140mm×116mm×81mmDifferentNA

{15}------------------------------------------------

{16}------------------------------------------------

{17}------------------------------------------------

The product specification of subject device is identical in Intended Use, Indication for Use, Patient population, Environment of use, Measuring Principle, Measurement Range, Operating Environment, Storage and Transport Environment;

The subject device and predicate device are differing in Power Source, Weight, Size; Comparing to the predicate device, the subject device has additional power Adapter and Broadcast Finction: Besides, the subject device also has different appearance, which result in different weight and size; But the differences mentioned above do not affect the device. So, we concluded that there is no significant difference between the product device and predicate device: The product specification of subject device is substantially equivalent (SE) to predicate device which is US legally market device.

{18}------------------------------------------------

K161708 Page 16 of 16

8. Discussion of Non Clinical tests Performed for Determination of Substantial

Equivalence: are as follows:

Testing information demonstrating substantial equivalence to the predicate device and safety and effectiveness of the Arm automatic blood pressure monitor, model: BF3202,BF3212, BF6002.BF6102. BF1205.BF1215. BF1213. BF6051.BF6151. BF6043.BF6143. BF1204,BF1214, BF6023L,BF6123L, BF6048L,BF6148L, BF6053,BF6153, BF6055; in the intended environment of use is supported by testing that was conducted in accordance with the FDA guidance "Reviewer Guidance for Premarket Notification Submissions", which outlines Electrical, Mechanical and Environmental Performance requirements. The following testing was conducted:

The following testing was conducted:
IEC 60601-1 Test

a. IEC 60601-1 Test

b. IEC 60601-1-2 Test

c. IEC 60601-1-11 Test

d. IEC 60601-2-30 Test

None of the testing demonstrated any design characteristics that violated the requirements of the Reviewer Guidance or resulted in any safety hazards. It was our conclusion that Arm automatic blood pressure monitor, Model: BF3202,BF3212, BF6002,BF6102, BF1205,BF1215,BF1213,BF6051,BF6151,BF6043,BF6143,BF1204,BF1214, BF6023L,BF6123L, BF6048L,BF6148L, BF6053,BF6153, BF6055; tested met all relevant requirements of the aforementioned tests.

9. Discussion of Clinical Tests Performed

The subject device: Arm automatic blood pressure monitor, Model: BF3202,BF3212, BF6002.BF6102. BF1205.BF1215. BF1213. BF6051.BF6151. BF6043.BF6143. BF1204.BF1214.BF6023L.BF6123L. BF6048L.BF6148L. BF6053.BF6153. BF6055; are identical to the predicate device: Arm automatic blood pressure monitor, Model:BE6034, BE6134 from the technical point of view. The differences between them do not relate to blood pressure measurement technology, so the clinical accuracy in terms of blood pressure detection will not be affected. Therefore repeated clinical test in accordance with the standard ANSI/AAMI/ISO 81060-2:2009 Non-invasive sphygmomanometers - Part 2: Clinical validation of automated measurement type is unnecessary.

10. Software Information

Software validation was conducted in accordance with a moderate level of concern designation in accordance with the FDA November 2005 document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices".

11. Conclusion

We have demonstrated that there are no significant differences between the subject device: Arm automatic blood pressure monitor (Model: BF3202,BF3212, BF6002,BF6102, BF1205,BF1215, BF1213, BF6051,BF6151, BF6043,BF6143, BF1204,BF1214, BF6023L,BF6123L, BF6048L,BF6148L, BF6053,BF6153, BF6055;), and the predicate device: Arm automatic blood pressure monitor (Model: BE6034, BE6134), in terms of safety and effectiveness based on electrical, mechanical and environmental test results per the FDA guidance "Non-Invasive Blood Pressure (NIBP) Monitor Guidance", and the IEC 80601-2-30:2009 + A1:2013.

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).