K Number
K161694

Validate with FDA (Live)

Manufacturer
Date Cleared
2016-07-15

(25 days)

Product Code
Regulation Number
888.3070
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Threshold™ Pedicular Fixation System is intended for posterior, non-cervical fixation as an adjunct to fusion in skeletally mature patients for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis; trauma, (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis and/or lordosis); turnor; pseudoarthrosis; and/ or failed previous fusion.

Device Description

Spineology's Threshold Pedicular Fixation System consists of cannulated and solid titanium alloy screws and rods to allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Threshold cannulated and solid screws are available with or without a hydroxyapatite coating. The system is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine. The cannulated and solid screws can be placed in the pedicles in a variety of trajectories ranging from standard anatomic transpedicular path projected medially toward the ventral vertebral body, to a caudocephalad path sagittally and a laterally directed path in the transverse plane.

AI/ML Overview

This document describes acceptance criteria and studies for the Threshold™ Pedicular Fixation System, specifically for the addition of uncoated and hydroxyapatite (HA) coated Threshold solid pedicle screws.

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance Criteria CategorySpecific Test/RequirementAcceptance CriteriaReported Device Performance
Mechanical PerformanceStatic Fatigue Properties (ASTM F1798-97)Must meet or exceed properties of predicate devicesDemonstrated substantial equivalence to predicate devices (specific values not provided in this document, but implied by claiming "sufficient" demonstration).
Metallic Bone Screws (ASTM F543-13)Must meet or exceed properties of predicate devicesDemonstrated substantial equivalence to predicate devices.
Spinal Implant Constructs (ASTM F1717-13)Must meet or exceed properties of predicate devicesDemonstrated substantial equivalence to predicate devices.
BiocompatibilityBacterial Endotoxin Testing (BET) / Limulus amoebocyte lysate (LAL) test (ANSI/AAMI/ST72)< 20 EU/DevicePerformed (specific results of testing not provided but implied to be acceptable as no issues were noted).
Material EquivalenceBase MaterialsSame as predicate devicesConfirmed to be the same base materials.
Coating EquivalenceHydroxyapatite (HA) CoatingSame as predicate devicesConfirmed to be the same HA coating.

2. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size: Not explicitly stated as traditional "test sets" for a performance study were not used for this submission. The submission relies on non-clinical testing (mechanical testing, finite element analysis) and comparison to predicate devices, rather than a clinical study with human subjects.
  • Data Provenance: The mechanical testing was performed in accordance with ASTM standards, implying laboratory-based testing. The finite element analysis (FEA) is a computational method. This is retrospective in the sense that it relies on existing data and models or newly generated data in a lab setting, rather than prospective human clinical trials for this particular submission.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

  • Not applicable. The submission does not involve a clinical study with human patients and an associated ground truth established by experts for diagnostic or treatment outcomes. The evaluation is based on engineering and material science principles.

4. Adjudication Method for the Test Set:

  • Not applicable, as a clinical test set requiring expert adjudication was not part of this submission. The evaluation was based on meeting engineering standards and comparison to predicate devices.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

  • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This submission pertains to a pedicle fixation system, a physical implant, and not an AI-powered diagnostic or treatment assistance device.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

  • No, a standalone performance study in the context of an algorithm's performance was not done. This is a physical medical device.

7. Type of Ground Truth Used:

  • The "ground truth" for this device's evaluation is primarily established by engineering standards and material specifications. For mechanical testing, this means the defined limits and methodologies of the ASTM standards (F1798-97, F543-13, F1717-13) and for biocompatibility, the limits of ANSI/AAMI/ST72. Substantial equivalence to predicate devices serves as a comparative benchmark.

8. Sample Size for the Training Set:

  • Not applicable. This submission does not involve a machine learning algorithm requiring a "training set."

9. How the Ground Truth for the Training Set Was Established:

  • Not applicable. There is no training set for this device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of the department's name encircling a stylized image of three faces in profile. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The faces are stacked vertically, with the top face being the most prominent.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 25, 2016

Spineology Incorporated Ms. Jacqueline Hauge Regulatory Affairs Manager 7800 3rd Street North, Suite 600 Saint Paul, Minnesota 55128-5455

Re: K161694

Trade/Device Name: Threshold" Pedicular Fixation System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, MNH, MNI Dated: June 17, 2016 Received: June 20, 2016

Dear Ms. Hauge:

This letter corrects our substantially equivalent letter of July 15, 2016.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803); good manufacturing practice requirements as

{1}------------------------------------------------

set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K161694

Device Name Threshold™ Pedicular Fixation System

Indications for Use (Describe)

The Threshold™ Pedicular Fixation System is intended for posterior, non-cervical fixation as an adjunct to fusion in skeletally mature patients for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis; trauma, (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis and/or lordosis); turnor; pseudoarthrosis; and/ or failed previous fusion.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image contains the word "Spineology" in a bold, sans-serif font. To the right of the word is a graphic of a cluster of dots, with one larger dot in the center and several smaller dots surrounding it. The dots are all the same color, a shade of purple.

510(k) Summary

Date Prepared:July 8, 2016
------------------------------

Submitter:

Spineology Inc. 7800 3rd Street North Suite 600 Saint Paul, MN 55128 Establishment Registration Number: 2135156

Contact Person:Jacqueline A. Hauge
Regulatory Affairs Manger
Phone:651.256.8534
Fax:651.256.8505
Email:jhauge@spineology.com

Device Name and Classification

Trade Name:Threshold™ Pedicular Fixation System
Common Name:Pedicle Screw System
Classification Name:Orthosis, Spinal Pedicle Fixation
Product Codes:NKB, MNI, MNH
Regulatory Class:NKB: Class III MNI, MNH: Class II
Regulation Number:NKB: 888.3070(b)(2) MNI, MNH: 888.3070(b)(1)
Panel:Orthopedic

Predicate Devices

Primary:K143403Threshold™ Pedicular Fixation System
Additional:K140010Fortress™ Pedicle Screw System
K160980Threshold™ Pedicular Fixation System
K152148Fortress™ Pedicle Screw System

Purpose

Obtain 510(k) clearance for the addition of uncoated and hydroxyapatite (HA) coated Threshold solid pedicle screws to Spineology's Threshold Pedicular Fixation System.

Device Description

Spineology's Threshold Pedicular Fixation System consists of cannulated and solid titanium alloy screws and rods to allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Threshold cannulated and solid screws are available with or without a hydroxyapatite coating. The system is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine. The cannulated and solid screws can be placed in the pedicles in a variety of trajectories ranging from standard anatomic transpedicular path projected medially toward the ventral vertebral body, to a caudocephalad path sagittally and a laterally directed path in the transverse plane.

{4}------------------------------------------------

Indications for Use

The Threshold Pedicular Fixation System is intended for posterior, non-cervical fixation as an adjunct to fusion in skeletally mature patients for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis; trauma, (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis and/or lordosis); tumor; pseudoarthrosis; and/or failed previous fusion.

Comparison to Predicate

When compared to the predicate devices, uncoated and HA-coated Threshold solid pedicle screws have the same:

  • Intended Use
  • Indications for Use
  • Fundamental Scientific Technology
  • Principle of Operation ●
  • . Biological Safety
  • Biological Safety
  • Base Materials
  • Size Offering
  • Hydroxyapatite (HA) Coating ●

Non-Clinical Testing

Testing was not required to support the addition of uncoated and HA-coated Threshold solid pedicle screws. Spineology's risk analysis, a finite element analysis (FEA), and previous mechanical testing conducted on the predicate devices is sufficient to demonstrate the substantial equivalence of the uncoated and HA-coated Threshold solid pedicle screws to the predicate devices.

The previous mechanical testing was performed in accordance with:

  • ASTM F1798-97 Standard Guide for Evaluating Static and Fatigue Properties of ● Interconnection Mechanisms and Subassemblies Used in Spinal Arthrodesis Implants
  • Standard Specification and Test Methods for Metallic Medical bone ● ASTM F543-13 Screws
  • Standard Specification and Test Methods for Spinal Implant Constructs . ASTM F1717-13 in a Vertebrectomy Model

Bacterial Endotoxin testing (BET), also known as the Limulus amebocyte lysate (LAL) test, was performed per ANSI/AAMI/ST72 to a limit of <20 EU/Device.

Conclusion

Spineology has demonstrated that the uncoated and HA-coated Threshold solid pedicle screws are substantially equivalent to the predicate devices. The fundamental scientific principle, primary technological characteristics, and intended use are unchanged from the predicate devices.

§ 888.3070 Thoracolumbosacral pedicle screw system.

(a)
Identification. (1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
(b)
Classification. (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.