K Number
K161624

Validate with FDA (Live)

Manufacturer
Date Cleared
2016-11-07

(147 days)

Product Code
Regulation Number
872.5570
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The MICRODENTAL, Inc. Micr02 device is intended to reduce night time snoring and mild to moderate obstructive sleep apnea (OSA) in adults.

Optionally, if the DentiTrac® micro-recorder is completely embedded in the MicrO2 device, the microrecorder is intended to measure patient compliance to oral device/appliance therapy in combination with the DentiTrac® System

Device Description

The MicrO2 OSA device series consists of maxillary and mandibular devices that when interfaced together reduce snoring and mild to moderate sleep apnea by holding the mandible forward during sleep, providing increased pharyngeal space. These separate upper and lower arch devices are designed with twin-mated posts and are CAD/CAM generated specifically for each prescription. Designed as a patient-specific device, the MicrO2 series consists of one or multiple lower device(s) together with one or multiple mated-post upper device(s) that are manufactured to the dentist prescriber's requested advancement positions to provide a selection of gentle adjustments according to patient comfort and need. As such, prescribed advancements can be achieved by simply removing the current upper or lower device and inserting the next upper or lower device in the mandibular advancing series. The MicrO2 does not have any adjustment mechanisms to modify or maintain the mandibular position such as pistons, screws, straps or repositioning elastics. The MicrO2's twin-mated post and lingualess design yields a small and a comfortable patient-specific mandibular advancement device. The design of the device maximizes tongue space and mandibular movement resulting in the ability to open and close during wear.

The Micr02 OSA device with Micro-recorder is identical to the primary predicate device, Somnomed with Micro-Recorder except for the difference in the adjustment mechanism. The MicrO2 uses twin mated posts to adjust whereas the Somnomed uses embedded screws. Without the compliance chip the subject device is identical in materials and manufacturing to the reference predicate MicrO2 OSA device. Any differences introduced by these modifications when compared to the predicate device, do not introduce any new concerns.

This submission adds the option for any clinician to decide to incorporate a DentiTrac® Microrecorder compliance chip embedded into a MicrO2 OSA device to record a patient's compliance to the prescribed oral appliance therapy. The MicrO2 OSA device with Micro-recorder is to be used in combination with the DentTrac® System which includes a base station at the provider's office used to upload the data from the chip to a web application for cloud based reporting and tracking. During scheduled visits, the data within the DentiTrac® can be uploaded to a web application for cloud-based reporting and tracking using a DentiTrac® Base Station at the clinician's office. The DentiTrac® micro-recorder monitors the wear time through the oral temperature, as well as tracks movements and head position. This is the identical chip that is included in the predicate device – the Somnodent with Micro-Recorder. The inclusion of the embedded DentiTrac® micro-recorder, provides additional information when used in combination with the DentiTrac® System, but does not impact the operating principles of the Micr02 OSA device as established by the reference device, the MicrO2 OSA without the DentiTrac® Micro-recorder.

AI/ML Overview

This FDA 510(k) summary describes the MicrO2 OSA Device with Micro-recorder. However, it explicitly states, "Clinical testing was not conducted on the subject device." Instead, the submission relies on demonstrating substantial equivalence to predicate devices through technical characteristics, material composition, and non-clinical performance testing. Therefore, the information requested in the prompt related to acceptance criteria for clinical performance, sample sizes for test sets, ground truth establishment, expert adjudication, or MRMC studies for an AI device is not applicable to this submission.

The document focuses on the mechanical performance and biocompatibility of the device and the embedded micro-recorder, not on the clinical efficacy of an AI-driven component.

Here's a breakdown of the available information, addressing the prompt as best as possible given the provided text:

1. A table of acceptance criteria and the reported device performance

The document does not provide a table of acceptance criteria in the traditional sense of numerical clinical performance targets for an AI device. Instead, it relies on demonstrating substantial equivalence to predicate devices through various comparisons and non-clinical testing. The "performance" reported is primarily in terms of meeting manufacturing specifications and maintaining structural integrity and biocompatibility.

Acceptance Criteria (Implied by Substantial Equivalence)Reported Device Performance
Manufacturing Quality ControlMicroDental Laboratories performs 100% visual and functional inspection to ensure the device meets manufacturing specifications for the MicrO2 OSA device with the inclusion of the DentiTrac® MicroRecorder Compliance Chip.
BiocompatibilityBiocompatibility assessment in accordance with ISO 10993-1 was conducted for the reference predicate (MicrO2 OSA device) and is considered supportive of the subject device as no new materials were added. Monomer leachability testing performed.
Structural Integrity / Mechanical PerformanceCompression and shear testing to approximate mechanical force applied during use was performed on the reference predicate. This testing is still applicable as the addition of the DentiTrac® Micro-Recorder does not impact structural integrity.
DentiTrac® Micro-recorder FunctionalityThe device can be repeatedly, and reliably embedded with the DentiTrac® and retain the same Quality Control functionality of the MicrO2 and DentiTrac®. The micro-recorder is completely sealed under a layer of PMMA.
Mode of Action (Mandibular Repositioning)Identical intended use and mode of action (mandibular repositioning to increase pharyngeal space) as predicate devices.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

Not applicable. No clinical test set data is described for the subject device. Non-clinical testing refers to lab-based assessments of materials and mechanical properties.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. No clinical test set with expert-established ground truth is described.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. No clinical test set with adjudication is described.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is not an AI algorithm and no MRMC study was conducted.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This device is not an AI algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

Not applicable. For the non-clinical tests described, the "ground truth" would be established engineering specifications, material standards, and validated testing methods.

8. The sample size for the training set

Not applicable. The device is not an AI algorithm and does not have a training set.

9. How the ground truth for the training set was established

Not applicable. The device is not an AI algorithm and does not have a training set.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, arranged in a way that resembles a bird in flight.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 7, 2016

Microdental, Inc. David Kuhns, PhD Senior Director of Product Development And Technical Sales 5601 Arnold Rd. Dublin, California 94568

Re: K161624

Trade/Device Name: MicrO2 OSA Device with Micro-Recorder Regulation Number: 21 CFR 872.5570 Regulation Name: Intraoral Devices For Snoring And Intraoral Devices For Snoring And Obstructive Sleep Apnea Regulatory Class: Class II Product Code: PLC Dated: August 9, 2016 Received: August 9, 2016

Dear Dr. Kuhns:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

Michael J. Ryan -S

for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number : K161624

Device Name: MicrO2 OSA Device with Micro-recorder

Indications for Use:

The MICRODENTAL, Inc. Micr02 device is intended to reduce night time snoring and mild to moderate obstructive sleep apnea (OSA) in adults.

Optionally, if the DentiTrac® micro-recorder is completely embedded in the MicrO2 device, the microrecorder is intended to measure patient compliance to oral device/appliance therapy in combination with the DentiTrac® System

Prescription Use _YES (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use NO (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

{3}------------------------------------------------

510(k) Summary 3

1. Date Summary Prepared:October 10, 2016
Submitter Information:MICRODENTAL, Inc.5601 Arnold RdDublin, CA 94568
Contact Person:David Kuhns, PhDSenior Director of Product Development and Technical Sales
2. Device Name:
Proprietary/Trade name:MicrO2 OSA Device with Micro-recorder
Common name:Mandibular Advancement Device
Classification name:Intraoral devices for snoring and sleep apnea and obstructivesleep apnea
Device Classification:II

21 CFR 872.5570

Product code PLC

Device Classification.
Regulation number:

3. Identification of Predicate Device(s)

Primary Predicate:

Device Name: SomnoDent with Micro-Recorder Manufacturer: SomnoMed, Ltd. 510(k) Number: K150369

Reference Predicate:

Device Name: Micr02 OSA Device Manufacturer: MicroDental Laboratories, Inc. 510(k) Number: K133683

4. Device Description:

The MicrO2 OSA device series consists of maxillary and mandibular devices that when interfaced together reduce snoring and mild to moderate sleep apnea by holding the mandible forward during sleep, providing increased pharyngeal space. These separate upper and lower arch devices are designed with twin-mated posts and are CAD/CAM generated specifically for each prescription. Designed as a patient-specific device, the MicrO2 series consists of one or multiple lower device(s) together with one or multiple mated-post upper device(s) that are manufactured to the dentist prescriber's requested advancement positions to provide a selection of gentle adjustments according to patient comfort and need. As such, prescribed advancements can be achieved by simply removing the current upper or lower device and inserting the next upper or lower device in the mandibular advancing series. The MicrO2 does

{4}------------------------------------------------

not have any adjustment mechanisms to modify or maintain the mandibular position such as pistons, screws, straps or repositioning elastics. The MicrO2's twin-mated post and lingualess design yields a small and a comfortable patient-specific mandibular advancement device. The design of the device maximizes tongue space and mandibular movement resulting in the ability to open and close during wear.

The Micr02 OSA device with Micro-recorder is identical to the primary predicate device, Somnomed with Micro-Recorder except for the difference in the adjustment mechanism. The MicrO2 uses twin mated posts to adjust whereas the Somnomed uses embedded screws. Without the compliance chip the subject device is identical in materials and manufacturing to the reference predicate MicrO2 OSA device. Any differences introduced by these modifications when compared to the predicate device, do not introduce any new concerns.

This submission adds the option for any clinician to decide to incorporate a DentiTrac® Microrecorder compliance chip embedded into a MicrO2 OSA device to record a patient's compliance to the prescribed oral appliance therapy. The MicrO2 OSA device with Micro-recorder is to be used in combination with the DentTrac® System which includes a base station at the provider's office used to upload the data from the chip to a web application for cloud based reporting and tracking. During scheduled visits, the data within the DentiTrac® can be uploaded to a web application for cloud-based reporting and tracking using a DentiTrac® Base Station at the clinician's office. The DentiTrac® micro-recorder monitors the wear time through the oral temperature, as well as tracks movements and head position. This is the identical chip that is included in the predicate device – the Somnodent with Micro-Recorder. The inclusion of the embedded DentiTrac® micro-recorder, provides additional information when used in combination with the DentiTrac® System, but does not impact the operating principles of the Micr02 OSA device as established by the reference device, the MicrO2 OSA without the DentiTrac® Micro-recorder.

5. Indications for Use:

The MICRODENTAL, Inc. MicrO2 device is intended to reduce night time snoring and mild to moderate obstructive sleep apnea (OSA) in adults.

Optionally, if the DentiTrac® micro-recorder is completely embedded into the MicrO2 appliance, the micro-recorder is intended to measure patient compliance to the oral appliance therapy in combination with the DentiTrac® System.

The Indications for Use Statement for the MicroDental, Inc. MicrO2 device with Micro-recorder is identical to the Somnodent with Micro-Recorder with the exception of the name. Both the subject and predicate devices have the same intended use for the treatment of OSA by mandibular repositioning.

6. Comparison of Technological Characteristics with the Predicate Device

Mandibular repositioning is the technological principle for both the subject and predicate devices. It is based on the movement of the mandible to increase the pharyngeal space. At a high level, the subject and predicate devices are based on the following elements:

{5}------------------------------------------------

  • A mandible tray
  • . An tray on the upper jaw
  • Use of a protrusion mechanism for the mandible
  • Optional use of the compliance chip to monitor patient compliance with the therapy
  • Material composition - PMMA

The following technological differences exist between the subject and predicate devices:

  • . The protrusion mechanism in the subject device is by use of a fin that is a part of the lower tray
  • . The digital milling process

From a technical design perspective, the subject device and the predicate devices improve the flow of air through the patient's pharyngeal space during sleep by repositioning the mandible and the use of upper and lower trays. While the device mechanics are slightly different between the subject device and the Somnomed with Micro-Recorder, they both advance the lower jaw forward in order to increase the pharyngeal space. The Somnomed with Micro-Recorder uses screws to advance the lower tray whereas the MicrO2 with Micro-recorder and the MicrO2 reference device both use a twin mated post for advancement.

The mechanism of the subject device and the predicate devices allows for adjustment of the amount of advancement based on dentist prescription and are patient specific. The dentist prescribes the amount of advancement increments for the MicroO2 with Microrecorder subject device and the MicrO2 OSA reference device according the physician professional assessment. The dentist prescribes these increments in the same way that the dentist adjusts the mandibular position with the pistons, screws, straps or repositioning elastics of the Somnomed with Micro-Recorder predicate device.

The MicrO2 OSA with Micro-recorder subject devices and both predicate devices use primarily the same material – PMMA – to provide a rigid tray.

The MicrO2 OSA with Micro-recorder is digitally milled identically to the MicrO2 reference device whereas the Somnomed with Micro-Recorder is manually manufactured.

Both the MicrO2 OSA with Micro-recorder and the SomnoMed with Micro-Recorder have an additional pocket to encase the Microrecorder. They both follow the same pocket process to include the DentiTrac® micro-recorder chip into the base device. The MicrO2 reference device does not have this pocket for the Microrecorder, thus the subject device – the MicrO2 OSA with Micro-recorder compliance chip is wider to include the pocket and the chip.

7. Performance Testing: Non-clinical Testing

{6}------------------------------------------------

Risk assessment, biocompatibility assessment in accordance with ISO 10993-1 and process validation for manufacturing was generated by MicroDental Laboratories to ensure that the MicrO2 OSA devices can be repeatedly and reliably embedded with the DentiTrac® and retain the same Quality Control functionality of the MicrO2 and DentiTrac®.

MicroDental Laboratories has generated process validation for the manufacturing process. Throughout the manufacturing process, MicroDental Laboratories performs 100% visual and functional inspection to ensure that the device meets manufacturing specifications for the MicrO2 OSA device with the inclusion of the DentiTrac® MicroRecorder Compliance Chip.

There are no new materials that have been added to the subject device from the reference predicate. As such, the biocompatibility assessment that was performed for the reference predicate MicrO2 OSA device is supportive of the subject device.

Clinical Data

Clinical testing was not conducted on the subject device

8. Substantial Equivalence

MICRODENTAL, Inc. has chosen the traditional 510(k) route of submission for marketing clearance of the Micr02 OSA Device.

Method of Action and Physical Characteristics: SomnoDent with Micro-Recorder cleared by K150369.

Substantial Equivalent TableMicrO2 OSADevice withMicro-recorderK161624PrimaryPredicate:SomnoDent withMicro-RecorderK150369ReferencePredicate:MicrO2 OSA DeviceK133683
Intended Use
Intended as an intraoral deviceYESYESYES
Intended to be worn during sleep onlyYESYESYES
Intended to reduce snoring or helpalleviate snoringYESYESYES
Intended to reduce or help alleviatemild to moderate sleep apneaYESYESYES
For single patient, multi-useYESYESYES
For personal use at home or in sleeplaboratoriesYESYESYES
For adult patient populationsYESYESYES
Prescription deviceYESYESYES
Cleaned dailyYESYESYES

{7}------------------------------------------------

Option to monitor patient complianceYESYESNO
Design
Separate upper & lower tray piecesYESYESYES
All rigid tray pieces for more stable fitYESYESYES
Custom fit, unique per patientYESYESYES
Advances the lower jawYESYESYES
Advancement position maintained byramp designYESYESYES
Can be adjusted, repaired and/or refitYESYESYES
Ease of insertionYESYESYES
Ease of removalYESYESYES
Allows the patient to open & closemouth during wearYESYESYES
Permits patient to breathe through themouthYESYESYES
Material
Hard PMMA materialYESYESYES
Non-sterileYESYESYES
DentiTrac® Micro-recorder embeddedinto MicrO2YESYESNO

Substantial Equivalence Based on Mode of Action

The Micr02 OSA device with Micro-recorder has the same intended use as the Somnomed with Micro-Recorder in terms of providing increased pharyngeal space through mandibular repositioning for the relief of snoring and nighttime mild to moderate sleep apnea. The MicrO2 OSA and the predicate devices are both classified as mandibular positioning devices that are designed to increase the pharyngeal space. The subject device and the primary predicate maintain the option of embedding the DentiTrac® Micro-recorder compliance chip that monitors patient compliance with the device. Without the compliance chip the subject device is identical in mode of action to the reference predicate MicrO2 OSA device.

Substantial Equivalence Based on Bench Testing

The choice of reference predicate device, MicrO2 OSA is made from currently marketed devices which have identical intended use, and, which provide equivalent physical parameters, in terms of vertical shear, horizontal shear, horizontal compression and leachability.

The added compliance chip does not negatively impact the design and does not invalidate the original performance testing as performance testing as performed in support of clearance of the reference predicate device MicrO2 OSA

{8}------------------------------------------------

Both the Micr02 OSA device with the Micro-recorder as well as the Somnomed and MicrO2 predicates are prescribed and fitted by dentists or physicians. All of the devices have a mechanical/physical mode of action; the materials used for the fabrication of the subject device are identical to the materials used for the subject device.

Compression and shear testing to approximate the mechanical force applied during use of the device was performed on the reference predicate. Since the addition of the DentiTrac® Micro-Recorder compliance chip does not impact the structural integrity of the device, this testing is still applicable to demonstrate performance and support substantial equivalence.

Monomer leachability testing was performed on the reference predicate. The materials in the subject device have been changed, thus the leachability testing continues to support biocompatibility and substantial equivalence to the reference predicate.

Substantial Equivalence with the Addition of the DentiTrac® Micro-recorder Compliance Chip

The addition of the DentiTrac® Micro-Recorder to the MicrO2 device does not change the intended use of the MicrO2 device for the treatment of night time snoring and mild to moderate obstructive sleep apnea in adult patients 18 years of age or older. Based on the process of embedding the DentiTrac® Micro-recorder inside the MicrO2 device, the patient will not be physically exposed to the micro-recorder. The DentiTrac® is completely sealed under a layer of PMMA within the MicrO2 device in the identical fashion as the SomnoDent with Micro-recorder has embedded the DentiTrac®. Furthermore, the size, position and location of the embedded DentiTrac® Micro-recorder in the MicrO2 device as there is ample space in the patient's oral cavity to accommodate the additional volume.

9. Conclusion

Based on the comparison technologic characteristics and the indications for use, together with the non-clinical performance testing,

MicrO2 Obstructive Sleep Apnea Device with Micro-recorder is substantially equivalent to the predicate devices SomnoDent with Micro-Recorder and Micr02 OSA Device.

§ 872.5570 Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea.

(a)
Identification. Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea are devices that are worn during sleep to reduce the incidence of snoring and to treat obstructive sleep apnea. The devices are designed to increase the patency of the airway and to decrease air turbulence and airway obstruction. The classification includes palatal lifting devices, tongue retaining devices, and mandibular repositioning devices.(b)
Classification. Class II (special controls). The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA.”